Medtronic 694875 Technical Manual

SPRINT FIDELIS™ 6948
Steroid eluting, quadripolar, ventricular lead with tined tip and RV/SVC defibrillation coil electrodes
Technical manual
Caution: Federal Law (USA) restricts this device to sale by or on the order of a physician (or properly licensed practitioner).
The following are trademarks of Medtronic: Medtronic, Sprint Fidelis, Sprint Quattro, CapSure Sense
Table of Contents
Device description 5
Contents of package 6 Accessory descriptions 6
Indications 6
Contraindications 6
Warnings and precautions 7
Potential adverse events 10
Clinical study 10
Model 6944 clinical study 11 Model 4074 (CapSure Sense tip electrode) clinical study 15
Directions for use 20
Opening the package 21 Inserting the lead 21 Positioning the lead 22 Taking electrical measurements and defibrillation efficacy
measurements 23 Anchoring the lead 24 Connecting the lead 25 Placing the device and leads into the pocket 26 Post-implant evaluation 26
Detailed device description 27
Specifications (nominal) 27 Specifications drawing (nominal) 28
Medtronic warranty 29
Service 29
6948 Technical manual 3

Device description

The Medtronic Sprint Fidelis Model 6948 lead is a steroid eluting, quadripolar, ventricular lead with tined tip and right ventricular (RV) and superior vena cava (SVC) defibrillation coil electrodes. The lead is designed for pacing, sensing, cardioversion, and defibrillation therapies.
The lead features a tined tip, silicone insulation with overlay, parallel conductors, titanium nitride coated platinum iridium tip and ring electrodes, and RV and SVC coil electrodes. See “Specifications drawing (nominal)” on page 28 for a lead drawing.
The tip electrode is common to the connector pin of the
1
bipolar leg.
IS-1
The ring electrode is common to the connector ring of the
1
bipolar leg.
IS-1
The RV coil electrode is common to the DF-12 leg of the trifurcation, labeled and marked with a red band.
The SVC coil electrode is common to the DF-12 leg of the trifurcation, labeled and marked with a blue band.
The RV and SVC coils deliver cardioversion and defibrillation therapies. Pacing and sensing occur between the tip and ring electrodes.
The IS-1 bipolar leg of the trifurcation features a lumen for stylet passage. The DF-1 connectors will not accept a stylet.
The steroid dexamethasone sodium phosphate is located on the tip electrode surface. The tip electrode also incorporates a steroid eluting plug containing dexamethasone acetate. The tip electrode contains a maximum of 1.0 mg of dexamethasone steroid. Exposure to body fluids elutes the steroid from the lead tip. The steroid is known to suppress the inflammatory response that is believed to cause threshold rises typically associated with implanted pacing electrodes.
1
IS-1 refers to the International Connector Standard (ISO 5841-3) whereby pulse generators and leads so designated are assured of a basic mechanical fit.
2
DF-1 refers to the International Connector Standard (ISO 11318) whereby pulse generators and leads so designated are assured of a basic mechanical fit.
6948 Technical manual English 5
Contents of package
The lead and accessories are provided sterile. Each package contains the following:
1 lead with 1 radiopaque anchoring sleeve1, stylet, and stylet guide
1 vein lifter
1 slit anchoring sleeve
2 pin caps
extra stylets
product literature
Accessory descriptions
Stylet – A stylet provides additional stiffness and controlled flexibility
for maneuvering the lead into position. Each stylet knob is labeled with the stylet diameter and length.
Stylet guide – A stylet guide facilitates stylet insertion into the lead.
Anchoring sleeve – An anchoring sleeve secures the lead from
moving and protects the lead insulation and conductors from damage caused by tight sutures.
Slit anchoring sleeve – A slit anchoring sleeve secures excess lead length in the device pocket.
Pin cap – A pin cap covers and insulates unused connector pins.
Vein lifter – A vein lifter facilitates lead insertion into a vessel.

Indications

The lead is intended for single, long-term use in the right ventricle.
This lead has application for patients for whom implantable cardioverter defibrillators are indicated.

Contraindications

Atrial use – The lead is contraindicated for the sole use of detection
and treatment of atrial arrhythmias.
Ventricular use – The lead is contraindicated for ventricular use in patients with tricuspid valvular disease or a tricuspid mechanical heart valve.
Transient ventricular tachyarrhythmias – The lead is contraindicated for patients with transient ventricular tachyarrhythmias due to reversible causes (drug intoxication, electrolyte imbalance, sepsis, hypoxia) or other factors (myocardial infarction, electric shock).
Steroid use – The lead is contraindicated in patients for whom a single dose of 1.0 mg of dexamethasone may be contraindicated.
1
Two anchoring sleeves are provided with leads 85 cm or longer.
6 English 6948 Technical manual

Warnings and precautions

For single use only – Do not resterilize and reimplant an
explanted lead.
Inspecting the sterile package – Inspect the package prior to opening.
If the seal or package is damaged, contact your local Medtronic representative.
Do not use the product after its expiration date.
Ethylene oxide resterilization – The lead has been sterilized with ethylene oxide prior to shipment. If the integrity of the sterile package has been compromised prior to the expiration date, resterilize using ethylene oxide. Avoid resterilization techniques that could damage the lead.
Refer to sterilizer instructions for operating instructions.
Use an acceptable method for determining sterilizer effectiveness, such as biological indicators.
Before resterilization, remove the disk tip protector, and place the device in an ethylene oxide permeable package.
Do not exceed temperatures of 55°C (131°F).
Do not resterilize more than 1 time.
After resterilization, allow the device to aerate ethylene oxide residues.
Electrophysiologic testing – Prior to lead implant, it is strongly recommended that patients undergo a complete cardiac evaluation, which should include electrophysiologic testing. Also, electrophysiologic evaluation and testing of the safety and efficacy of the proposed pacing, cardioversion, or defibrillation therapies are recommended during and after the implant of the system.
Steroid use – It has not been determined whether the warnings, precautions, or complications usually associated with injectable dexamethasone apply to the use of this highly localized, controlled-release device. For listing of potentially adverse effects, refer to the Physician’s Desk Reference.
Handling the steroid tip – Reducing the available amount of steroid may adversely affect low-threshold performance. Avoid reducing the amount of steroid available prior to lead implant.
Do not allow the electrode surface to come in contact with surface contaminants.
Do not wipe or immerse the electrode in fluid, except blood, at the time of implant.
Handling the lead – Handle the lead with great care at all times.
Protect the lead from materials shedding particles such as lint and dust. Lead insulators attract these particles.
Handle the lead with sterile surgical gloves that have been rinsed in sterile water or a comparable substance.
6948 Technical manual English 7
Do not severely bend, kink, or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pins.
Do not immerse leads in mineral oil, silicone oil, or any other liquid, except blood, at the time of implant.
Inserting the lead using a lead introducer that features a hemostasis valve may require a larger introducer than the size recommended. To avoid distortion of the coil electrode, do not withdraw the lead through a hemostasis valve.
Handling the stylets – Use care when handling stylets.
Do not use excessive force or surgical instruments when inserting a stylet.
Avoid overbending and kinking.
Use a new stylet when blood or other fluids accumulate on the stylet. Accumulated fluids may cause lead damage or difficulty in passing the stylet through the lead.
Do not use a sharp object to impart a curve to the distal end of the stylet.
Necessary hospital equipment – Keep external defibrillation equipment nearby for immediate use during the acute lead system testing, implant procedure, or whenever arrhythmias are possible or intentionally induced during post-implant testing.
Line-powered equipment – An implanted lead forms a direct current path to the myocardium. During lead implant and testing, use only battery-powered equipment or line-powered equipment specifically designed for this purpose, to protect against fibrillation that may be caused by alternating currents. Line-powered equipment used in the vicinity of the patient must be properly grounded. Lead connector pins must be insulated from any leakage currents that may arise from line-powered equipment.
Second anchoring sleeve – Leads 85 cm or longer feature 2 anchoring sleeves. Use both anchoring sleeves to assure adequate fixation, see the section “Anchoring the lead” on page 24.
Concurrent devices – Output pulses, especially from unipolar devices, may adversely affect device sensing capabilities. If a patient requires a separate stimulation device, either permanent or temporary, allow enough space between the leads of the separate systems to avoid interference in the sensing capabilities of the devices. Previously implanted pulse generators, implantable cardioverter defibrillators, and leads should generally be explanted. Refer to “Chronic repositioning or removal” on page 9, for further information on explanting leads.
8 English 6948 Technical manual
Diathermy – People with metal implants such as pacemakers, implantable cardioverter defibrillators (ICDs), and accompanying leads should not receive diathermy treatment. The interaction between the implant and diathermy can cause tissue damage, fibrillation, or damage to the device components, which could result in serious injury, loss of therapy, or the need to reprogram or replace the device.
Chronic repositioning or removal – Chronic repositioning or removal of leads may be difficult because of fibrotic tissue development. Return all removed leads, or lead segments, to Medtronic. If a lead must be removed or repositioned, proceed with extreme caution.
Lead removal may result in avulsion of the endocardium, valve, or vein.
Lead junctions may separate, leaving the lead tip and bare wire in the heart or vein.
Chronic repositioning may adversely affect the low-threshold performance of a steroid eluting lead.
Cap abandoned leads to avoid transmitting electrical signals.
For leads that have been severed, seal the remaining lead end and suture the lead to adjacent tissue.
Connector compatibility – Although the lead conforms to the International Connector Standards IS-1 and DF-1, do not attempt to use the lead with any device other than a commercially available implantable defibrillator system with which it has been tested and demonstrated to be safe and effective. The potential adverse consequences of using such a combination may include, but are not limited to, undersensing cardiac activity and failure to deliver necessary therapy.
6948 Technical manual English 9

Potential adverse events

The potential adverse events related to the use of transvenous leads include, but are not limited to, the following patient-related conditions:
cardiac perforation
cardiac tamponade
constrictive pericarditis
embolism
endocarditis
fibrillation or other arrhythmias
heart wall rupture
hemothorax
infection
pneumothorax
thrombosis
tissue necrosis
Other potential adverse events related to the lead include, but are not limited to, the following:
Insulation failure
Lead conductor or electrode fracture
Lead dislodgment
Poor connection to the device, which may lead to oversensing, undersensing, or a loss of therapy

Clinical study

Clinical data was not collected in the approval process for this lead. Clinical data from the Models 6944 and 4074 leads support the safety and efficacy of the Model 6948 lead.
The Model 6948 lead is a downsized version of the Medtronic Sprint Quattro Model 6944 lead. The Model 6948 lead includes a combination of components used in currently marketed Medtronic leads with some enhancements. All functional features of the Model 6948 lead have been approved in these currently marketed Medtronic leads. Previous clinical studies have shown titanium nitride coated electrodes (6948) do not significantly change clinical pacing thresholds or sensing amplitudes as compared to platinized electrodes (6944 and 4074). The overall surface of the defibrillation electrodes falls within the range of currently approved Sprint leads.
Based upon its similarity to the Model 6944 and 4074 leads, the clinical data from these lead models supports the safety and efficacy of the Model 6948 lead (Table 1).
10 English 6948 Technical manual
Tab le 1.
Sprint Fidelis Model 6948 features:
RV/SVC defibrillation electrodes RV/SVC defibrillation
Passive fixation - tines Passive fixation - tines
Ring electrode for true bipolar sensing
Steroid eluting Steroid eluting
Silicone insulation with polyurethane overlay
a
Standard
Sprint Fidelis Model 6948 features:
CapSure Sense electrode tip CapSure Sense electrode tip
a
electrode spacing
8 mm tip to ring spacing, 12 mm tip to RV coil spacing
Sprint Quattro Model 6944 clinical data supports:
electrodes
Ring electrode for true bipolar sensing
Silicone insulation with polyurethane overlay
a
Standard
CapSure Sense Model 4074 clinical data supports:
electrode spacing
Model 6944 clinical study
Note: The leads used for the Model 6944 clinical study contained
only dexamethasone acetate in a steroid plug.
The Model 6944 clinical study was a prospective, randomized, multicenter trial evaluating the safety and efficacy of the Sprint
Quattro
Model 6944 lead for the treatment of life threatening arrhythmias. The Model 6944 was compared to the market released Sprint Model 6942 lead.
The Model 6944 lead was implanted in 112 patients and the Model 6942 lead was implanted in 122 patients in 27 centers located in the United States and Canada between February 25, 1999 and July 30, 1999. All but 17 of the patients received the Model 7227 ICD, the Model 7271 ICD or the Model 7273 ICD.
Patients had to meet the following eligibility criteria: 1) able to receive a pectoral implant, and 2) survival of at least one episode of cardiac arrest due to a ventricular tachyarrhythmia or episodes of recurrent, poorly tolerated, sustained VT (spontaneous or induced). In addition, patients in Canada could be included in the study if they were able to receive a pectoral implant and met all of the following three conditions: 1) prior myocardial infarction, 2) a left ventricular ejection fraction of <
35%, and 3) had a documented history of nonsustained
VT with inducible ventricular tachyarrhythmia.
Patients studied
Patient demographics for both the Model 6944 lead and the Model 6942 lead are provided in Table 2.
6948 Technical manual English 11
Table 2. Model 6944 vs. Model 6942 patient demographics
Patient demographic 6944 (n = 112) 6942 (n =122)
Gender (N,%)
Male 93 (83.0%) 101 (82.8%)
Female 19 (17.0%) 21 (17.2%)
Age (years)
Mean 64.7 65.6
Range (32.0 - 84.6) (41.0 - 88.8)
Standard deviation 12.2 10.5
Primary indication (mutually exclusive, N,%)
SCD 13 (11.6%) 14 (11.5%)
VT 64 (57.1%) 78 (63.9%)
SCD/VT 34 (30.4%) 30 (24.6%)
MADIT 1 (0.9%) 0
Primary cardiovascular history (non-exclusive, N,%)
CAD and MI 77 (68.8%) 90 (73.8%)
CAD without MI 14 (12.5%) 18 (14.8%)
MI without CAD 9 (8.0%) 5 (4.1%)
Cardiomyopathy (dilated/ ischemic) 65 (58.0%) 85 (69.7%)
Primary electrical disease 1 (0.9%) 2 (1.6%)
Valvular heart disease 41 (36.6%) 29 (23.8%)
Pacemaker dependence 5 (4.5%) 5 (4.1%)
Congestive heart failure 58 (51.8%) 64 (52.5%)
Ejection fraction (%)
Mean 32.1 32.0
Range 10 - 68 4 - 70
Standard deviation 14.2 14.1
Not available 7 (6.3%) 2 (1.6%)
NYHA classification (N,%)
Class I 29 (25.9%) 26 (21.3%)
Class II 54 (48.2%) 63 (51.6%)
Class III 23 (20.5%) 28 (23.0%)
Class IV 4 (3.6%) 3 (2.5%)
Unknown 2 (1.8%) 2 (1.6%)
Objectives
The primary objectives of the study were to show equivalence in:
Ventricular lead-related event-free survival at three months
Pacing thresholds at one and three months
R-wave amplitudes at three months.
12 English 6948 Technical manual
Methods
Pulse width thresholds, R-wave amplitudes (EGM), pacing lead impedances, and subthreshold high voltage lead impedances were measured at implant and at one and three months post-implant. Adverse events were collected throughout the study.
Result addressing the primary objective (ventricular lead-related event-free survival at three months)
The Kaplan-Meier estimate of the ventricular lead-related event-free survival at three months was 92.7% (95% confidence interval of
86.0% to 96.3%) for the Model 6944 lead and 97.4% (95% confidence interval of 92.2% to 99.2%) for the Model 6942 lead. The upper 95% confidence limit on the difference of 4.7% between the Model 6944 and the Model 6942 leads was 9.4%, which is below the equivalence of 10%. Thus the ventricular lead-related event-free survival at three months for the Model 6944 lead was equivalent to that of the Model 6942 lead.
Result addressing the primary objective (pacing thresholds at one month and three months)
The upper 95% confidence limits on the differences in average pulse width thresholds at one volt pulse amplitude between the Model 6944 lead and the Model 6942 lead were 0.059 ms at one month and 0.043 ms at three months. Since the upper 95% confidence limits were below the equivalence bound of 0.10 ms, the pulse width thresholds at one and three months for the Model 6944 lead are equivalent to those of the Model 6942 lead.
Result addressing the primary objective (R-wave amplitudes at three months)
The difference in mean R-wave amplitudes between the Model 6944 and the Model 6942 leads at three months was 1.4 mV with an upper confidence limit of 2.16 mV. The objective was written with an equivalence bound of 2 mV. However, this equivalence bound did not take into account the differences in the sensing mechanisms of the Model 6944 (true bipolar) and the Model 6942 (integrated bipolar) leads. Even though the R-wave amplitudes of the Model 6944 lead are not equivalent to the Model 6942 lead (with equivalence defined as to within 2 mV), the R-wave amplitudes of the Model 6944 lead are on average consistent over time and of adequate magnitude for clinically acceptable sensing performance.
Pace/sense measurement results
The pace/sense measurements for the Model 6944 lead are summarized in Table 3 and the pace/sense measurements for the Model 6942 are summarized in Table 4.
6948 Technical manual English 13
Tab l e 3. Model 6944 pace/sense measurements at implant,
Pulse width thresholds at one volt
N 105 86 90
Median (ms) 0.10 0.20 0.20
Range (ms) (0.03 - 0.40) (0.03 - 0.90) (0.03 - 1.40)
R-wave amplitude (EGM)
N 110 100 99
Mean (mV) 9.1 9.5 9.2
Standard deviation 2.7 3.9 3.2
Pacing lead impedance
N 112 103 101
Median (ohms) 985 776 776
Range (ohms) (564 - 1722) (410 - 1389) (410 - 1156)
High voltage lead impedance (subthreshold)
N 105 97 96
Median (ohms) 14 17 17
Range (ohms) (10 - 25) (12 - 30) (12 - 32)
1 month, 3 months
Implant 1 month 3 months
Tab le 4. Model 6942 pace/sense measurements at implant,
Pulse width thresholds at one volt
N 111 107 96
Median (ms) 0.10 0.12 0.20
Range (ms) (0.03 - 2.00) (0.03 - 1.00) (0.03 - 1.00)
R-wave amplitude (EGM)
N 119 111 100
Mean (mV) 10.3 11.0 10.6
Standard deviation 3.7 4.0 3.5
Pacing lead impedance
N 120 112 105
Median (ohms) 412.5 410 410
Range (ohms) (275 - 776) (275 - 716) (275 - 661)
High voltage lead impedance (subthreshold)
N 110 101 98
Median (ohms) 14 18 17.5
Range (ohms) (10 - 21) (14 -27) (13 - 23)
1 month, 3 months
Implant 1 month 3 months
14 English 6948 Technical manual
Model 6944 conclusion
The Model 6944 lead is equivalent to the Model 6942 lead with respect to safety and pacing performance. The sensing performance of the Model 6944 lead is clinically appropriate.
The Model 6944 lead’s higher impedance values and low pacing threshold combine to reduce pacing system energy requirements ­potentially improving pacing system longevity. For specific pacing system longevity values, refer to the applicable implantable cardioverter defibrillator product literature.
The clinical experience with the Model 6944 lead demonstrates that the lead is safe and effective for human use.
Model 4074 (CapSure Sense tip electrode) clinical study
A multi-center, prospective, nonrandomized, historically controlled clinical study conducted at 19 investigational sites in the United States and 4 investigational sites in Canada compared the Model 4074 and Model 4574 pacing leads to the Medtronic Model 4092, Model 4592, and Model 5072 pacing leads (historical control leads).
Clinical experience from the market released Model 4092 and Model 4592 leads was used as the historical control for the safety, pacing threshold, ventricular sensing, and impedance objectives. The market released Model 5072 lead was used as the historical control for the atrial sensing objective because it has a tip to ring spacing of 10 mm which is similar to the 9 mm tip to ring spacing in the Model 4574 lead.
During the study, 132 patients received Model 4074 leads in the ventricle and 132 patients received Model 4574 leads in the atrium. A total of 132 patients participated in the clinical study.
Primary objectives
The clinical study has four primary objectives for safety and effectiveness.
Lead related events – Verify the safety of the Model 4074 lead as measured by
ventricular lead-related adverse events compared to the Model 4092.
– Verify the safety of the Model 4574 lead as measured by
atrial lead-related adverse events compared to the Model 4592.
6948 Technical manual English 15
Pacing performance – Verify pacing performance of the Model 4074 lead as
measured by ventricular pacing thresholds compared to the Model 4092.
– Verify pacing performance of the Model 4574 lead as
measured by atrial pacing thresholds compared to the Model 4592.
Sensing performance – Verify sensing performance of the Model 4074 lead as
measured by ventricular R-wave amplitudes compared to the Model 4092.
– Verify sensing performance of the Model 4574 lead as
measured by atrial P-wave amplitudes compared to the Model 5072.
Lead impedance – Verify pacing impedance of the Model 4074 lead as
measured by ventricular pacing impedance compared to the Model 4092.
– Verify pacing impedance of the Model 4574 lead as
measured by atrial pacing impedance compared to the Model 4592.
Results
For Model 4074 lead related adverse events, the 95% upper confidence bound on the difference between lead-related adverse event rates was 7.75%, which is below the 10% upper bound criteria (Table 5). Therefore, the objective concerning the equivalence of ventricular lead-related adverse event rates was met. For Model 4574 lead related adverse events, the 95% upper confidence bound on the difference between lead-related adverse event rates was 4.98%, which is below the 10% upper bound criteria (Table 6). Therefore, the objective concerning the equivalence of atrial lead-related adverse event rates was met.
Tab l e 5. Ventricular lead related events: number (rate per patient month)
Event Complication Observation
Failure to capture/Loss of capture 1 (0.002) 0
Lead dislodgement 6 (0.013) 0
a
Other
To ta l: 8 (0.017) 0 (0)
a
During atrial lead placement, the physician elected to reposition the ventricular lead, which required a longer lead length.
1 (0.002) 0
16 English 6948 Technical manual
Tab le 6. Atrial lead related events: number (rate per patient month)
Event Complication Observation
Elevated pacing thresholds 2 (0.004) 1 (0.002)
Failure to capture/Loss of capture 2 (0.004) 1 (0.002)
Lead dislodgement 4 (0.008) 0
To ta l: 8 (0.017) 2 (0.004)
For Model 4074 pacing performance, the 95% upper confidence bound on the difference between the two means was 0.011 ms, which was below the 95% upper bound criteria of 0.06 ms. Therefore, the objective concerning the equivalence of ventricular pulse width thresholds was met (Figure 1).
Figure 1. Ventricular pulse width thresholds at 2.5 V
6948 Technical manual English 17
For Model 4574 pacing performance, the 95% upper confidence bound on the difference between the two means was 0.008 ms, which was below the 95% upper bound criteria of 0.06 ms. Therefore, the objective concerning the equivalence of atrial pulse width thresholds was met (Figure 2).
Figure 2. Atrial pulse width thresholds at 2.5 V
For Model 4074 sensing performance, the 95% upper confidence bound on the difference between the two means was 0.997 mV, which was below the 95% upper bound criteria of 3.0 mV. Therefore, the objective concerning the equivalence of ventricular R-wave amplitudes was met (Figure 3).
Figure 3. Ventricular R-wave amplitude
18 English 6948 Technical manual
For Model 4574 sensing performance, the 95% upper confidence bound on the difference between the two means was 0.003 mV, which was below the 95% upper bound criteria of 1.5 mV. Therefore, the objective concerning the equivalence of atrial P-wave amplitudes was met (Figure 4).
4574
Figure 4. Atrial P-wave amplitude
For Model 4074 lead impedance, the 95% upper confidence bound on the difference between the two means was -59 , which was below the 95% upper bound criteria of 200 . Therefore, the objective concerning the equivalence of ventricular pacing impedances was met (Figure 5).
Figure 5. Ventricular pacing impedances
6948 Technical manual English 19
For Model 4574 lead impedance, the 95% upper confidence bound on the difference between the two means was -21 , which was below the 95% upper bound criteria of 200 . Therefore, the objective concerning the equivalence of atrial pacing impedances was met (Figure 6).
Figure 6. Atrial pacing impedances
Secondary objective
Lead performance with Ventricular Capture Management
Characterize the distribution of false negative increases in ventricular output due to the Ventricular Capture Management feature.
Results
In order to fully evaluate Ventricular Capture Management (VCM), a patient needs to have a continuum of data from each device interrogation. There were 122 patients with Ventricular Capture Management data available to analyze. Of those patients, six experienced at least one false negative in ventricular output due to the Ventricular Capture Management feature. The 1-sided lower confidence bound on the proportion of patients who did not experience a false negative (116/122 = 95.1%) was 90.5%.

Directions for use

Proper surgical procedures and sterile techniques are the responsibility of the medical professional. The following procedures are provided for information only. Each physician must apply the information in these instructions according to professional medical training and experience.
The implant procedure generally includes the following steps:
Opening the package
Inserting the lead
Positioning the lead
20 English 6948 Technical manual
Taking electrical measurements and defibrillation efficacy measurements
Anchoring the lead
Connecting the lead
Placing the device and leads into the pocket
Post-implant evaluation
Opening the package
Use the following steps to open the sterile package and inspect the lead:
1. Within the sterile field, open the sterile package and remove the lead and accessories.
2. Spread the slot in the disc-shaped tine protector at the distal end of the lead and carefully remove it from the lead.
3. Inspect the lead. Leads shorter than 85 cm should have 1 anchoring sleeve on the lead body. Leads 85 cm or longer should have 2 anchoring sleeves on the lead body.
Inserting the lead
Caution: Use care when handling the lead during insertion.
Do not severely bend, kink, or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pins.
Insert the lead using the following techniques:
1. Select a site for lead insertion. The lead may be inserted by venotomy through several different venous routes, including the right or left cephalic vein or the external or internal jugular vein. Use the cephalic vein whenever possible to avoid lead damage in the first rib or clavicular (thoracic inlet) space.
Caution: Certain anatomical abnormalities, such as thoracic outlet syndrome, may also precipitate pinching and subsequent fracture of the lead.
Caution: When using a subclavian approach, avoid techniques that may damage the lead.
Place the insertion site as far lateral as possible to avoid clamping the lead body between the clavicle and the first rib (Figure 7).
6948 Technical manual English 21
Figure 7.
Do not force the lead if significant resistance is encountered during lead passage.
Do not use techniques such as adjusting the patient’s posture to facilitate lead passage. If resistance is encountered, it is recommended that an alternate venous entry site be used.
2. Insert the tapered end of a vein lifter into the incised vein and gently push the lead tip underneath and into the vein (Figure 8).
Note: A percutaneous lead introducer (PLI) kit may be used to facilitate insertion. Refer to the technical manual packaged with an appropriate percutaneous introducer for further instructions.
Figure 8.
3. Advance the lead into the right atrium using a straight stylet to facilitate movement through the veins.
Positioning the lead
Caution: Use care when handling the lead during positioning.
Do not severely bend, kink or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pins.
Use the following steps to position the lead:
1. After the lead tip is passed into the atrium, advance the lead through the tricuspid valve. Replacing the straight stylet with a gently curved stylet may add control in maneuvering the lead through the tricuspid valve.
22 English 6948 Technical manual
Caution: Do not use a sharp object to impart a curve to the distal end of the stylet. Imparting a curve to the stylet can be accomplished with a smooth-surface, sterile instrument (Figure 9).
Figure 9.
Note: Passing the lead tip through the tricuspid valve or chordae tendineae may be difficult due to the tines and the flexible nature of the lead body. Rotating the lead body as the tip passes through the valve may facilitate passage.
2. After the lead tip is in the ventricle, the curved stylet may be replaced with a straight stylet. Withdraw the stylet slightly, to avoid using excessive tip force while achieving final electrode position. Avoid known infarcted or thin wall areas, to minimize the occurrence of perforation.
3. Proper positioning of the electrode is essential for stable endocardial pacing. A satisfactory position usually is achieved when the lead tip points straight toward the apex, or when the distal end dips or bends slightly. Use fluoroscopy (lateral position) to ensure that the tip is not in a retrograde position or lodged in the coronary sinus.
4. After final positioning, remove the stylet guide and stylet completely. When removing the stylet guide, grip the lead firmly just below the connector pin, to help prevent possible lead dislodgment.
5. Obtain final electrical measurements as defined in “Taking electrical measurements and defibrillation efficacy measurements” on page 23.
Taking electrical measurements and defibrillation efficacy measurements
Caution: Prior to taking electrical or defibrillation efficacy
measurements, move objects made of conductive materials, such as guide wires, away from all electrodes. Metal objects, such as guide wires, can short a lead and an active implantable device, causing electrical current to bypass the heart and possibly damage the implantable device lead.
Use the following steps to take electrical measurements:
1. Attach a surgical cable to the lead connector pin. A notch in the stylet guide allows connection of a surgical cable for obtaining electrical measurements.
6948 Technical manual English 23
2. Use a testing device, such as a pacing system analyzer, for obtaining electrical measurements. For information on the use of the testing device, consult the product literature for that device.
In order to demonstrate reliable defibrillation efficacy, obtain final defibrillation measurements for the lead system.
Tab l e 7. Recommended measurements at implant
(when using a pacing system analyzer)
Measurements required Acute
a
lead system
Chronic
b
lead system
Capture threshold (at 0.5 ms pulse width) 1.0 V 3.0 V
Pacing impedance 200 - 1000 ohms 200 - 1000 ohms
Filtered R-wave amplitude (during sinus rhythm)
5 mV (during sinus rhythm)
3 mV (during sinus rhythm)
Slew rate 0.75 V/s 0.45 V/s
a
< 30 days after implant.
b
> 30 days after implant.
If initial electrical measurements deviate from the recommended values, it may be necessary to repeat the testing procedure 15 minutes after final positioning. Initial electrical measurements may deviate from the recommended values:
Initial impedance values may exceed the measuring capabilities of the testing device, resulting in an error message.
Values may vary depending upon lead type, implantable device settings, cardiac tissue condition, and drug interactions.
If electrical measurements do not stabilize to acceptable levels, it may be necessary to reposition the lead and repeat the testing procedure.
In order to keep patient morbidity and mortality to a minimum, patients should be rescued promptly with an external defibrillator if the implanted lead system fails to terminate a VF episode. At least 5 minutes should elapse between VF inductions.
For more information on obtaining electrical measurements, consult the technical manual supplied with the testing device.
Anchoring the lead
Caution: Use care when anchoring the lead.
Use only nonabsorbable sutures to anchor the lead.
Do not attempt to remove or cut the anchoring sleeve from the lead body.
During lead anchoring, take care to avoid dislodging the lead tip.
Do not secure sutures so tightly that they damage the vein, lead, or anchoring sleeve (Figure 10).
Do not tie a suture directly to the lead body (Figure 10).
24 English 6948 Technical manual
Figure 10.
Use the following steps to anchor the lead using all 3 grooves:
Note: The anchoring sleeves contain a radiopaque substance, which allows visualization of the anchoring sleeve on a standard x-ray and may aid in follow-up examinations.
1. Position the distal anchoring sleeve against or near the vein.
2. Secure the anchoring sleeve to the lead body by tying a suture firmly in each of the 3 grooves (Figure 11).
Figure 11.
3. Use at least 1 additional suture in 1 of the grooves to secure the anchoring sleeve and lead body to the fascia.
4. A second anchoring sleeve is provided with leads 85 cm or longer. For abdominal implants, redundant lead body (for example, a curve for strain relief) should be placed just proximal to the first anchoring sleeve. Then, the second anchoring sleeve may be lightly sutured to the lead body and fascia to hold the curve in place. This procedure helps isolate the vein entry site from tension on the proximal end of the lead body.
5. A slit anchoring sleeve may be used in the device pocket to secure excess lead length. First, secure the anchoring sleeve to the lead body. Then, orient the slit toward the fascia and secure the anchoring sleeve to the fascia with sutures.
Connecting the lead
Use the following steps to connect the lead to an implantable device:
1. Carefully remove the stylet and stylet guide. When removing the stylet and stylet guide, grip the lead firmly just below the connector pin, to prevent dislodgment.
2. Insert the lead connectors into the connector block. Consult the product literature packaged with the implantable device for instructions on proper lead connections.
6948 Technical manual English 25
Placing the device and leads into the pocket
Caution: Use care when placing the device and leads into
the pocket.
Ensure that the leads do not leave the device at an acute angle.
Do not grip the lead or device with surgical instruments.
Do not coil the lead. Coiling the lead can twist the lead body and may result in lead dislodgment (Figure 12).
Figure 12.
Use the following steps to place the device and leads into the pocket:
1. To prevent undesirable twisting of the lead body, rotate the device to loosely wrap the excess lead length (Figure 13).
Figure 13.
2. Insert the device and leads into the pocket.
3. Before closing the pocket, verify sensing, pacing, cardioversion, and defibrillation efficacy.
Post-implant evaluation
After implant, monitor the patient’s electrocardiogram until the patient is discharged. If a lead dislodges, it usually occurs during the immediate postoperative period.
Recommendations for verifying proper lead positioning include x-rays and pacing/sensing thresholds taken at pre-hospital discharge, 3 months after implant, and every 6 months thereafter.
In the event of a patient death, explant all implanted leads and devices and return them to Medtronic with a completed Product Information Report form. Call the appropriate phone number on the back cover if there are any questions on product handling procedures.
26 English 6948 Technical manual

Detailed device description

Specifications (nominal)
Parameter Model 6948
Type Quadripolar
Position Right ventricle
Fixation Tined
Length 40-110 cm
Connectors Unipolar (2): DF-1
Bipolar: IS-1
Materials Conductors: MP35N coil
Insulator: Silicone/ETFE
Electrodes: Titanium nitride coated platinum
RV/SVC coils: Platinum-clad tantalum
DF-1 connector pins: Stainless steel
IS-1 connector pin and ring: Stainless steel
Steroid Type: Dexamethasone acetate
Amount: 1.0 mg maximum (combined)
Steroid binder: Silicone
Conductor resistances
Pacing (unipolar): 44.7 at 65 cm
Pacing (bipolar): 91.7 at 65 cm
Defibrillation: < 2.4 at 65 cm
Diameters Lead body: 2.2 mm
Lead introducer (recommended size)
Without guide wire: 7.0 French
With guide wire: 9.0 French
MP35N composite cables
Overlay: Polyurethane
iridium
Dexamethasone sodium phosphate
Tip: 2.2 mm
6948 Technical manual English 27
Specifications drawing (nominal)
180 mm
12 mm
8 mm
Tip electrode
Surface area: 2.5 mm
Ring electrode
Surface area: 20.2 mm
RV coil electrode
Length: 62 mm Surface area: 513 mm
2
2
2
Electrical shadow area: 430 mm
SVC coil electrode
Length: 80 mm Surface area: 663 mm
2
Electrical shadow area: 556 mm
Anchoring sleeve
Note: Leads 85 cm or
longer have 2 anchoring sleeves
2
2
DF-1 connector (red band) Note: Connector
pin is common to RV coil electrode
IS-1 BI connector Note: Connector pin is common
DF-1 connector (blue band) Note: Connector
pin is common to SVC coil electrode
to tip electrode; connector ring is common to ring electrode
28 English 6948 Technical manual

Medtronic warranty

For complete warranty information, see the accompanying warranty document.

Service

Medtronic employs highly trained representatives and engineers located throughout the world to serve you and, upon request, to provide training to qualified hospital personnel in the use of Medtronic products. Medtronic also maintains a professional staff to provide technical consultation to product users. For medical consultation, Medtronic can often refer product users to outside medical consultants with appropriate expertise. For more information, contact your local Medtronic representative, or call or write Medtronic at the appropriate address or telephone number listed on the back cover.
6948 Technical manual English 29
World Headquarters
Medtronic, Inc. 710 Medtronic Parkway Minneapolis, MN 55432-5604 USA Internet: www.medtronic.com Tel. 763-514-4000 Fax 763-514-4879
Medtronic USA, Inc.
Toll-free in the USA (24-hour technical consultation for physicians and medical professionals) Bradycardia: 1-800-505-4636 Tachycardia: 1-800-723-4636 Heart Failure: 1-800-505-4636
Europe/Africa/Middle East Headquarters
Medtronic Europe Sàrl Route du Molliau 31 Case Postale CH-1131 Tolochenaz Switzerland Internet: www.medtronic.co.uk Tel. 41-21-802-7000 Fax 41-21-802-7900
Medtronic E.C. Authorized Representative/Distributed by
Medtronic B.V. Earl Bakkenstraat 10 6422 PJ Heerlen The Netherlands Tel. 31-45-566-8000 Fax 31-45-566-8668
Asia
Medtronic International Ltd. Suite 1602 16/F, Manulife Plaza The Lee Gardens, 33 Hysan Avenue Causeway Bay Hong Kong Tel. 852-2891-4068 Fax 852-2591-0313
Aust ralia
Medtronic Australasia Pty. Ltd. Unit 4/446 Victoria Road Gladesville NSW 2111 Australia Tel. 61-2-9879-5999 Fax 61-2-9879-5100
Canada
Medtronic of Canada Ltd. 6733 Kitimat Road Mississauga, Ontario L5N 1W3 Tel. 905-826-6020 Fax 905-826-6620 Toll-free in Canada: 1-800-268-5346
Japan
Medtronic Japan Solid Square West Tower 6F, 580 Horikawa-cho, Saiwai-ku, Kawasaki, Kanagawa 210-0913 Japan Tel. 81-44-540-6112 Fax 81-44-540-6200
*A02224001*
© Medtronic, Inc. 2004 All Rights Reserved
A02224001 2004-09-14
Loading...