Dexamethasone acetate and dexamethasone sodium phosphate steroid eluting, quadripolar, screw-in,
ventricular lead with RV/SVC defibrillation coil electrodes
Technical Manual
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
The following list includes trademarks or registered trademarks of Medtronic in the United States and possibly in other countries.
All other trademarks are the property of their respective owners.
AccuRead, Medtronic, Sprint Quattro, Sprint Quattro Secure, Tensi-Lock
Contents
1 Device description3
2 Drug component description3
3 Indications4
4 Contraindications4
5 Warnings and precautions4
6 Adverse events5
7 Drug information6
8 Directions for use7
9 Specifications (nominal)12
10 Medtronic warranty14
11 Service14
12 Explanation of symbols on package labeling14
1 Device description
The Medtronic Sprint Quattro Secure Model 6947M lead is a
steroid-eluting, quadripolar, screw-in, ventricular lead with right
ventricular (RV) and superior vena cava (SVC) defibrillation coil
electrodes. This lead is designed for pacing, sensing,
cardioversion, and defibrillation therapies.
The lead features an extendable and retractable helix electrode,
silicone insulation, and parallel conductors. The four electrodes
of the lead are the helix, ring, RV coil, and SVC coil. The lead also
features Tensi-Lock1 and silicone-backfilled defibrillation coils.
The DF4-LLHH2 four-pole HV inline connector on the lead
facilitates device connection during implant. The DF4 connector
pin has a color band indicator that may be used to visually confirm
proper connection to the device.
The RV and SVC coils deliver cardioversion and defibrillation
therapies. Pacing and sensing occur between the helix and
ring electrodes. An AccuRead analyzer cable interface tool (ACI
tool) is attached to the lead to facilitate accurate electrical
measurements during implant.
The helix electrode can be actively fixed into the endocardium.
The helix electrode can be extended or retracted by rotating the
DF4 connector pin with one of the purple fixation tools included
in the package.
The distal tip contains a nominal dosage of 685 µg of
dexamethasone acetate and 59 µg of dexamethasone sodium
phosphate. Upon exposure to body fluids, the steroids elute from
the lead tip. The steroids are known to suppress the inflammatory
response that is believed to cause threshold rises typically
associated with implanted pacing electrodes.
1.1 Package contents
Leads and accessories are supplied sterile. Each package
contains the following items:
●
1 lead with a radiopaque anchoring sleeve3, stylet, and
AccuRead tool
●
2 purple fixation tools
●
1 purple stylet guide
●
1 slit anchoring sleeve
●
1 vein lifter
●
extra stylets
●
product literature
1.2 Accessory descriptions
AccuRead analyzer cable interface tool – The AccuRead tool
facilitates accurate electrical measurements during implant and
prevents possible connector damage.
Anchoring sleeve – An anchoring sleeve secures the lead to
prevent it from moving and protects the lead insulation and
conductors from damage caused by tight sutures.
Purple stylet guide – A stylet guide facilitates stylet insertion into
the lead.
Slit anchoring sleeve – A slit anchoring sleeve secures excess
lead length in the device pocket.
Stylet – A stylet provides additional stiffness and controlled
flexibility for maneuvering the lead into position. Each stylet knob
is labeled with the stylet diameter and corresponding lead length.
Vein lifter – A vein lifter facilitates lead insertion into a vein.
2 Drug component description
The active ingredients in the Model 6947M lead are
dexamethasone acetate [21-(acetyloxy)-9-fluoro-11β,
17-trihydroxy-16α-methylpregna-1,4-diene-3,20-dione] and
dexamethasone sodium phosphate [9-fluoro-11β,
17-dihydroxy-16α-methyl-21-(phosphonooxy)
pregna-1,4-diene-3,20-diene-3,20-dione disodium salt]. The
structural formula for these steroids is shown in the following
figures.
Dexamethasone acetate is a white to practically white, odorless
powder. It is a practically insoluble ester of dexamethasone, a
synthetic adrenocortical steroid.
1
Tensi-Lock is an exclusive Medtronic design feature that utilizes lead body cables to act like a built-in locking stylet and add tensile strength to the lead.
2
DF4-LLHH refers to the International Connector Standard ISO 27186:2010, where the lead connector contacts are defined as low voltage (L) or high voltage (H).
3
Two radiopaque anchoring sleeves are provided on leads that are 85 cm or longer.
3
Figure 1. Structural formula for dexamethasone acetate (DXAC) C24H31FO
6
Dexamethasone sodium phosphate (DSP) is an inorganic ester
of dexamethasone, a synthetic adrenocortical steroid.
Dexamethosone sodium phosphate is a white or slightly yellow
crystalline powder. It is freely soluble in water and is very
hygroscopic.
Figure 2. Structural formula for dexamethasone sodium
phosphate (DSP) - C22H28FNa2O8P
The maximum dosage of dexamethasone acetate and
dexamethasone sodium phosphate is less than 1.0 mg per lead.
3 Indications
The lead is intended for single, long-term use in the right ventricle.
This lead has application for patients for whom implantable
cardioverter defibrillators (ICDs) are indicated.
4 Contraindications
Atrial use – The lead is contraindicated for the sole use of
detection and treatment of atrial arrhythmias.
Ventricular use – The lead is contraindicated for ventricular use
in patients with tricuspid valvular disease or a tricuspid
mechanical heart valve.
Transient ventricular tachyarrhythmias – The lead is
contraindicated for patients with transient ventricular
tachyarrhythmias due to reversible causes (drug intoxication,
electrolyte imbalance, sepsis, hypoxia) or other factors
(myocardial infarction, electric shock).
Steroid use – The lead is contraindicated in patients for whom a
single dose of 1.0 mg of dexamethasone acetate and
dexamethasone sodium phosphate may be contraindicated.
5 Warnings and precautions
Inspecting the sterile package – Inspect the sterile package
before opening it.
●
If the seal of the package is damaged, contact a Medtronic
representative.
●
Do not store this product above 40 °C (104 °F).
●
Do not use the product after its expiration date.
Single use – The lead is for single use only.
Sterilization – Medtronic has sterilized the package contents
with ethylene oxide before shipment. This lead is for single use
only and is not intended to be resterilized.
Connector compatibility – The DF4-LLHH quadripolar
connector port is not compatible with DF-1 or IS-1 leads. A
DF4-LLHH quadripolar lead must be used with this port.
Electrophysiologic testing – Prior to lead implant, it is strongly
recommended that patients undergo a complete cardiac
evaluation, which should include electrophysiologic testing. Also,
electrophysiologic evaluation and testing of the safety and
efficacy of the proposed pacing, cardioversion, or defibrillation
therapies are recommended during and after the implant of the
system.
Steroid use – It has not been determined whether the warnings,
precautions, or complications usually associated with injectable
dexamethasone sodium phosphate or dexamethasone acetate
apply to the use of this highly localized, controlled-release device.
For a list of potential adverse effects, refer to the Physicians’ DeskReference.
Handling the steroid tip – Avoid reducing the amount of steroid
available before implanting the lead. Reducing the available
amount of steroid may adversely affect low-threshold
performance.
●
Do not allow the electrode surface to come in contact with
surface contaminants.
●
Do not wipe or immerse the electrode in fluid, except blood,
at the time of implant.
Handling the lead – Handle the lead with care at all times.
●
Protect the lead from materials that shed particles such as lint
and dust. Lead insulators attract these particles.
●
Handle the lead with sterile surgical gloves that have been
rinsed in sterile water or a comparable substance.
●
Do not severely bend, kink, or stretch the lead.
●
Do not use surgical instruments to grasp the lead or
connector pins.
●
Do not immerse the lead in mineral oil, silicone oil, or any other
liquid, except blood, at the time of implant.
4
●
Inserting the lead using a lead introducer that has a
hemostasis valve may require a larger introducer than the size
recommended. To avoid distortion of the coil electrode, do
not withdraw the lead through a hemostasis valve.
●
Do not implant the lead without first verifying the mechanical
functioning of the helix electrode. Refer to Section 8.2,
“Verifying the mechanical functioning of the helix electrode”,
page 7, for complete instructions.
●
Do not rotate the helix electrode after it is fully extended or
fully retracted. Do not exceed the recommended maximum
number of rotations to extend or retract the helix electrode.
Exceeding the maximum number may result in fracture or
distortion of the inner conductor or helix electrode. The
number of rotations required to fully extend or retract the helix
electrode is variable; refer to Section 9 for the recommended
maximum number of rotations.
Handling the stylet – Handle the stylet with care at all times.
●
Do not use a sharp object to impart a curve to the distal end
of the stylet.
●
Do not use excessive force or surgical instruments when
inserting the stylet into the lead.
●
Avoid overbending or kinking the stylet.
●
Use a new stylet when blood or other fluids accumulate on
the stylet. Accumulated blood or other fluids may damage the
lead or cause difficulty in passing the stylet into the lead.
Necessary hospital equipment – Keep external defibrillation
equipment nearby for immediate use during acute lead system
testing, the implant procedure, or whenever arrhythmias are
possible or intentionally induced during post-implant testing.
Line-powered and battery-powered equipment – An
implanted lead forms a direct current path to the myocardium.
During lead implant and testing, use only battery-powered
equipment or line-powered equipment specifically designed for
this purpose to protect against fibrillation that may be caused by
alternating currents. Line-powered equipment used in the vicinity
of the patient must be properly grounded. Lead connector pins
must be insulated from any leakage currents that may arise from
line-powered equipment.
Second anchoring sleeve – Leads that are 85 cm or longer
include 2 anchoring sleeves. Use both anchoring sleeves to
assure adequate fixation.
Concurrent devices – Output pulses, especially from unipolar
devices, may adversely affect device sensing capabilities. If a
patient requires a separate stimulation device, either permanent
or temporary, allow enough space between the leads of the
separate systems to avoid interference in the sensing capabilities
of the devices. Previously implanted pulse generators and
implantable cardioverter defibrillators should generally be
explanted.
Magnetic resonance imaging (MRI) – Do not use magnetic
resonance imaging (MRI) on patients who have this device
implanted. MRI can induce currents on implanted leads,
potentially causing tissue damage and the induction of
tachyarrhythmias.
Diathermy – People with metal implants such as pacemakers,
implantable cardioverter defibrillators (ICDs), and accompanying
leads should not receive diathermy treatment. The interaction
between the implant and diathermy can cause tissue damage,
fibrillation, or damage to the device components, which could
result in serious injury, loss of therapy, and/or the need to
reprogram or replace the device.
Chronic repositioning or removal of a screw-in lead –
Proceed with extreme caution if a lead must be removed or
repositioned. Chronic repositioning or removal of screw-in
transvenous leads may not be possible because of blood or
fibrotic tissue development into the helix mechanism on the lead.
In most clinical situations, it is preferable to abandon unused
leads in place. Return all removed leads, unused leads, or lead
sections to Medtronic for analysis.
Note: If a helix does not disengage from the endocardium by
rotating the connector pin, rotating the lead body
counterclockwise may withdraw the helix and decrease the
possibility of damage to cardiovascular structures during
removal.
●
Lead removal may result in avulsion of the endocardium,
valve, or vein.
●
Lead junctions may separate, leaving the lead tip and bare
wire in the heart or vein.
●
Chronic repositioning of a lead may adversely affect the
low-threshold performance of a steroid lead.
●
An abandoned lead should be capped so that the lead does
not transmit electrical signals.
●
Severed leads should have the remaining lead end sealed
and the lead body sutured to adjacent tissue.
AccuRead tool – The AccuRead tool reduces the risk of
connector damage, and reduces the risk of bridging and shorting
that may occur while taking electrical measurements during
implant. The potential for connector damage, bridging, and
shorting is due to variations in analyzer cable terminals, as well
as to the connector ring width and the proximity of the rings on the
DF4 connector.
6 Adverse events
Observed adverse events – The Model 6947 lead was utilized
in a prospective, nonrandomized, multicenter trial to assess the
handling and performance of the lead.
Based upon its similarity to the Model 6947M, the clinical data
from the Model 6947 lead supports the safety and efficacy of the
Model 6947M lead. The Model 6947M lead is equivalent to the
Model 6947 lead except for the connector system. Model 6947
uses the IS-1/DF-14 connector system; Model 6947M uses the
Medtronic DF4-LLHH connector system.
From the 6947 clinical trial, a total of 22 cardiovascular-related
adverse events were reported. Two of the 22 events were
4
DF-1 refers to the International Connector Standard ISO 11318:1993, whereby pulse generators and leads so designated are assured of a basic mechanical fit.
5
ventricular lead-related and both occurred at implant. One event
was a microdislodgement and one was induced VT as a result of
lead manipulation.
Two deaths occurred in this patient group during the follow-up
period. Both deaths were classified as non-sudden cardiac and
were judged to be non-system related by an independent
advisory committee.
Potential adverse events – The potential adverse events
associated with the use of transvenous leads and pacing systems
include, but are not limited to, the following events:
●
acceleration of tachyarrhythmias (caused by device)
●
air embolism
●
bleeding
●
body rejection phenomena, including local tissue reaction
●
cardiac dissection
●
cardiac perforation
●
cardiac tamponade
●
chronic nerve damage
●
constrictive pericarditis
●
death
●
device migration
●
endocarditis
●
erosion
●
excessive fibrotic tissue growth
●
extrusion
●
fibrillation or other arrhythmias
●
fluid accumulation
●
formation of hematomas/seromas or cysts
●
heart block
●
heart wall or vein wall rupture
●
hemothorax
●
infection
●
keloid formation
●
lead abrasion and discontinuity
●
lead migration/dislodgement
●
mortality due to inability to deliver therapy
●
muscle and/or nerve stimulation
●
myocardial damage
●
myocardial irritability
●
myopotential sensing
●
pericardial effusion
●
pericardial rub
●
pneumothorax
●
poor connection of the lead to the device, which may lead to
oversensing, undersensing, or a loss of therapy
●
threshold elevation
●
thrombosis
●
thrombotic embolism
●
tissue necrosis
●
valve damage (particularly in fragile hearts)
●
venous occlusion
●
venous perforation
Other potential adverse events related to the lead include, but are
not limited to, the following conditions:
●
insulation failure
●
lead conductor or electrode fracture
7 Drug information
7.1 Steroid mechanism of action
Steroid suppresses the inflammatory response that is believed to
cause threshold rises typically associated with implanted pacing
electrodes. Dexamethasone sodium phosphate and
dexamethasone acetate are synthetic steroids of the
glucocorticoid family. Glucocorticoids have potent
anti-inflammatory actions via direct and indirect effects on major
inflammatory cells. Glucocorticosteroids bind to a cytoplasmic
glucocorticoid receptor as well as a membrane-bound receptor.
Binding to the cytoplasmic receptor leads to receptor activation
and translocation to the nucleus. The receptor interacts with
specific DNA sequences within the regulatory regions of affected
genes. Thus, glucocorticoids inhibit the production of multiple cell
factors that are critical in generating the inflammatory response.
7.2 Pharmacodynamics of the Model 6947M lead
Pharmacokinetics – The pharmacokinetics (local drug levels
and systemic levels) of dexamethasone acetate (DXAC) and
dexamethasone sodium phosphate (DSP) and their metabolites
following lead implantation were not evaluated in human clinical
trials.
The in-vivo elution profile of a tined pacemaker lead with a DSP
monolithic controlled release device, based upon an assay of
explanted leads, is shown in Mond and Stokes5.
Metabolism – The conversion of DSP to dexamethasone occurs
within minutes; the conversion of DXAC to dexamethasone
occurs within hours. The dexamethasone alcohol
(dexamethasone) is the active glucocorticoid used in Medtronic
leads. Steroid is applied to the tip and eluted through the electrode
tip to the tissue interface where it will be used. Dexamethasone
acetate and Dexamethasone sodium phosphate are hydrolyzed
into dexamethasone, which is readily absorbed by the
surrounding tissue and body fluids. Glucocorticoids, when given
systemically, are eliminated primarily by renal excretion of
inactive metabolites.
Mutagenesis, carcinogenicity and reproductive
toxicology – The mutagenesis, carcinogenicity, and
reproductive toxicity of the Model 6947M lead have not been
evaluated. However, the mutagenesis, carcinogenicity, and
reproductive toxicity of dexamethasone acetate and
dexamethasone sodium phosphate have been evaluated
previously.
Carcinogenesis, mutagenesis, impairment of fertility – No
adequate studies have been conducted in animals to determine
whether corticosteroid have a potential for carcinogenesis (tumor
5
Mond, H. and Stokes, K.B., The Electrode - Tissue Interface: The Revolutionary Role of Steroid Elution, Pacing and Clinical Electrophysiology, Vol. 15, No. 1, pp
95-107
6
Loading...
+ 12 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.