Medtronic 6947M49 Technical Manual

Page 1
SPRINT QUATTRO SECURE® 6947M DXAC/DSP
Dexamethasone acetate and dexamethasone sodium phosphate steroid eluting, quadripolar, screw-in, ventricular lead with RV/SVC defibrillation coil electrodes
Technical Manual
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
Page 2
The following list includes trademarks or registered trademarks of Medtronic in the United States and possibly in other countries. All other trademarks are the property of their respective owners.
AccuRead, Medtronic, Sprint Quattro, Sprint Quattro Secure, Tensi-Lock
Page 3

Contents

1 Device description 3 2 Drug component description 3 3 Indications 4 4 Contraindications 4 5 Warnings and precautions 4 6 Adverse events 5 7 Drug information 6 8 Directions for use 7
9 Specifications (nominal) 12 10 Medtronic warranty 14 11 Service 14 12 Explanation of symbols on package labeling 14

1 Device description

The Medtronic Sprint Quattro Secure Model 6947M lead is a steroid-eluting, quadripolar, screw-in, ventricular lead with right ventricular (RV) and superior vena cava (SVC) defibrillation coil electrodes. This lead is designed for pacing, sensing, cardioversion, and defibrillation therapies.
The lead features an extendable and retractable helix electrode, silicone insulation, and parallel conductors. The four electrodes of the lead are the helix, ring, RV coil, and SVC coil. The lead also features Tensi-Lock1 and silicone-backfilled defibrillation coils.
The DF4-LLHH2 four-pole HV inline connector on the lead facilitates device connection during implant. The DF4 connector pin has a color band indicator that may be used to visually confirm proper connection to the device.
The RV and SVC coils deliver cardioversion and defibrillation therapies. Pacing and sensing occur between the helix and ring electrodes. An AccuRead analyzer cable interface tool (ACI tool) is attached to the lead to facilitate accurate electrical measurements during implant.
The helix electrode can be actively fixed into the endocardium. The helix electrode can be extended or retracted by rotating the DF4 connector pin with one of the purple fixation tools included in the package.
The distal tip contains a nominal dosage of 685 µg of dexamethasone acetate and 59 µg of dexamethasone sodium phosphate. Upon exposure to body fluids, the steroids elute from the lead tip. The steroids are known to suppress the inflammatory response that is believed to cause threshold rises typically associated with implanted pacing electrodes.

1.1 Package contents

Leads and accessories are supplied sterile. Each package contains the following items:
1 lead with a radiopaque anchoring sleeve3, stylet, and AccuRead tool
2 purple fixation tools
1 purple stylet guide
1 slit anchoring sleeve
1 vein lifter
extra stylets
product literature

1.2 Accessory descriptions

AccuRead analyzer cable interface tool – The AccuRead tool
facilitates accurate electrical measurements during implant and prevents possible connector damage.
Anchoring sleeve – An anchoring sleeve secures the lead to prevent it from moving and protects the lead insulation and conductors from damage caused by tight sutures.
Purple fixation tool – The purple fixation tool facilitates connector pin rotation.
Purple stylet guide – A stylet guide facilitates stylet insertion into the lead.
Slit anchoring sleeve – A slit anchoring sleeve secures excess lead length in the device pocket.
Stylet – A stylet provides additional stiffness and controlled flexibility for maneuvering the lead into position. Each stylet knob is labeled with the stylet diameter and corresponding lead length.
Vein lifter – A vein lifter facilitates lead insertion into a vein.

2 Drug component description

The active ingredients in the Model 6947M lead are dexamethasone acetate [21-(acetyloxy)-9-fluoro-11β, 17-trihydroxy-16α-methylpregna-1,4-diene-3,20-dione] and dexamethasone sodium phosphate [9-fluoro-11β, 17-dihydroxy-16α-methyl-21-(phosphonooxy) pregna-1,4-diene-3,20-diene-3,20-dione disodium salt]. The structural formula for these steroids is shown in the following figures.
Dexamethasone acetate is a white to practically white, odorless powder. It is a practically insoluble ester of dexamethasone, a synthetic adrenocortical steroid.
1
Tensi-Lock is an exclusive Medtronic design feature that utilizes lead body cables to act like a built-in locking stylet and add tensile strength to the lead.
2
DF4-LLHH refers to the International Connector Standard ISO 27186:2010, where the lead connector contacts are defined as low voltage (L) or high voltage (H).
3
Two radiopaque anchoring sleeves are provided on leads that are 85 cm or longer.
3
Page 4
Figure 1. Structural formula for dexamethasone acetate (DXAC) ­C24H31FO
6
Dexamethasone sodium phosphate (DSP) is an inorganic ester of dexamethasone, a synthetic adrenocortical steroid. Dexamethosone sodium phosphate is a white or slightly yellow crystalline powder. It is freely soluble in water and is very hygroscopic.
Figure 2. Structural formula for dexamethasone sodium phosphate (DSP) - C22H28FNa2O8P
The maximum dosage of dexamethasone acetate and dexamethasone sodium phosphate is less than 1.0 mg per lead.

3 Indications

The lead is intended for single, long-term use in the right ventricle.
This lead has application for patients for whom implantable cardioverter defibrillators (ICDs) are indicated.

4 Contraindications

Atrial use – The lead is contraindicated for the sole use of
detection and treatment of atrial arrhythmias.
Ventricular use – The lead is contraindicated for ventricular use in patients with tricuspid valvular disease or a tricuspid mechanical heart valve.
Transient ventricular tachyarrhythmias – The lead is contraindicated for patients with transient ventricular tachyarrhythmias due to reversible causes (drug intoxication, electrolyte imbalance, sepsis, hypoxia) or other factors (myocardial infarction, electric shock).
Steroid use – The lead is contraindicated in patients for whom a single dose of 1.0 mg of dexamethasone acetate and dexamethasone sodium phosphate may be contraindicated.

5 Warnings and precautions

Inspecting the sterile package – Inspect the sterile package
before opening it.
If the seal of the package is damaged, contact a Medtronic representative.
Do not store this product above 40 °C (104 °F).
Do not use the product after its expiration date.
Single use – The lead is for single use only.
Sterilization – Medtronic has sterilized the package contents
with ethylene oxide before shipment. This lead is for single use only and is not intended to be resterilized.
Connector compatibility – The DF4-LLHH quadripolar connector port is not compatible with DF-1 or IS-1 leads. A DF4-LLHH quadripolar lead must be used with this port.
Electrophysiologic testing – Prior to lead implant, it is strongly recommended that patients undergo a complete cardiac evaluation, which should include electrophysiologic testing. Also, electrophysiologic evaluation and testing of the safety and efficacy of the proposed pacing, cardioversion, or defibrillation therapies are recommended during and after the implant of the system.
Steroid use – It has not been determined whether the warnings, precautions, or complications usually associated with injectable dexamethasone sodium phosphate or dexamethasone acetate apply to the use of this highly localized, controlled-release device. For a list of potential adverse effects, refer to the Physicians’ Desk Reference.
Handling the steroid tip – Avoid reducing the amount of steroid available before implanting the lead. Reducing the available amount of steroid may adversely affect low-threshold performance.
Do not allow the electrode surface to come in contact with surface contaminants.
Do not wipe or immerse the electrode in fluid, except blood, at the time of implant.
Handling the lead – Handle the lead with care at all times.
Protect the lead from materials that shed particles such as lint and dust. Lead insulators attract these particles.
Handle the lead with sterile surgical gloves that have been rinsed in sterile water or a comparable substance.
Do not severely bend, kink, or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pins.
Do not immerse the lead in mineral oil, silicone oil, or any other liquid, except blood, at the time of implant.
4
Page 5
Inserting the lead using a lead introducer that has a hemostasis valve may require a larger introducer than the size recommended. To avoid distortion of the coil electrode, do not withdraw the lead through a hemostasis valve.
Do not implant the lead without first verifying the mechanical functioning of the helix electrode. Refer to Section 8.2, “Verifying the mechanical functioning of the helix electrode”, page 7, for complete instructions.
Do not rotate the helix electrode after it is fully extended or fully retracted. Do not exceed the recommended maximum number of rotations to extend or retract the helix electrode. Exceeding the maximum number may result in fracture or distortion of the inner conductor or helix electrode. The number of rotations required to fully extend or retract the helix electrode is variable; refer to Section 9 for the recommended maximum number of rotations.
Handling the stylet – Handle the stylet with care at all times.
Do not use a sharp object to impart a curve to the distal end of the stylet.
Do not use excessive force or surgical instruments when inserting the stylet into the lead.
Avoid overbending or kinking the stylet.
Use a new stylet when blood or other fluids accumulate on the stylet. Accumulated blood or other fluids may damage the lead or cause difficulty in passing the stylet into the lead.
Necessary hospital equipment – Keep external defibrillation equipment nearby for immediate use during acute lead system testing, the implant procedure, or whenever arrhythmias are possible or intentionally induced during post-implant testing.
Line-powered and battery-powered equipment – An implanted lead forms a direct current path to the myocardium. During lead implant and testing, use only battery-powered equipment or line-powered equipment specifically designed for this purpose to protect against fibrillation that may be caused by alternating currents. Line-powered equipment used in the vicinity of the patient must be properly grounded. Lead connector pins must be insulated from any leakage currents that may arise from line-powered equipment.
Second anchoring sleeve – Leads that are 85 cm or longer include 2 anchoring sleeves. Use both anchoring sleeves to assure adequate fixation.
Concurrent devices – Output pulses, especially from unipolar devices, may adversely affect device sensing capabilities. If a patient requires a separate stimulation device, either permanent or temporary, allow enough space between the leads of the separate systems to avoid interference in the sensing capabilities of the devices. Previously implanted pulse generators and implantable cardioverter defibrillators should generally be explanted.
Magnetic resonance imaging (MRI) – Do not use magnetic resonance imaging (MRI) on patients who have this device implanted. MRI can induce currents on implanted leads,
potentially causing tissue damage and the induction of tachyarrhythmias.
Diathermy – People with metal implants such as pacemakers, implantable cardioverter defibrillators (ICDs), and accompanying leads should not receive diathermy treatment. The interaction between the implant and diathermy can cause tissue damage, fibrillation, or damage to the device components, which could result in serious injury, loss of therapy, and/or the need to reprogram or replace the device.
Chronic repositioning or removal of a screw-in lead –
Proceed with extreme caution if a lead must be removed or repositioned. Chronic repositioning or removal of screw-in transvenous leads may not be possible because of blood or fibrotic tissue development into the helix mechanism on the lead. In most clinical situations, it is preferable to abandon unused leads in place. Return all removed leads, unused leads, or lead sections to Medtronic for analysis.
Note: If a helix does not disengage from the endocardium by rotating the connector pin, rotating the lead body counterclockwise may withdraw the helix and decrease the possibility of damage to cardiovascular structures during removal.
Lead removal may result in avulsion of the endocardium, valve, or vein.
Lead junctions may separate, leaving the lead tip and bare wire in the heart or vein.
Chronic repositioning of a lead may adversely affect the low-threshold performance of a steroid lead.
An abandoned lead should be capped so that the lead does not transmit electrical signals.
Severed leads should have the remaining lead end sealed and the lead body sutured to adjacent tissue.
AccuRead tool – The AccuRead tool reduces the risk of connector damage, and reduces the risk of bridging and shorting that may occur while taking electrical measurements during implant. The potential for connector damage, bridging, and shorting is due to variations in analyzer cable terminals, as well as to the connector ring width and the proximity of the rings on the DF4 connector.

6 Adverse events

Observed adverse events – The Model 6947 lead was utilized
in a prospective, nonrandomized, multicenter trial to assess the handling and performance of the lead.
Based upon its similarity to the Model 6947M, the clinical data from the Model 6947 lead supports the safety and efficacy of the Model 6947M lead. The Model 6947M lead is equivalent to the Model 6947 lead except for the connector system. Model 6947 uses the IS-1/DF-14 connector system; Model 6947M uses the Medtronic DF4-LLHH connector system.
From the 6947 clinical trial, a total of 22 cardiovascular-related adverse events were reported. Two of the 22 events were
4
DF-1 refers to the International Connector Standard ISO 11318:1993, whereby pulse generators and leads so designated are assured of a basic mechanical fit.
5
Page 6
ventricular lead-related and both occurred at implant. One event was a microdislodgement and one was induced VT as a result of lead manipulation.
Two deaths occurred in this patient group during the follow-up period. Both deaths were classified as non-sudden cardiac and were judged to be non-system related by an independent advisory committee.
Potential adverse events – The potential adverse events associated with the use of transvenous leads and pacing systems include, but are not limited to, the following events:
acceleration of tachyarrhythmias (caused by device)
air embolism
bleeding
body rejection phenomena, including local tissue reaction
cardiac dissection
cardiac perforation
cardiac tamponade
chronic nerve damage
constrictive pericarditis
death
device migration
endocarditis
erosion
excessive fibrotic tissue growth
extrusion
fibrillation or other arrhythmias
fluid accumulation
formation of hematomas/seromas or cysts
heart block
heart wall or vein wall rupture
hemothorax
infection
keloid formation
lead abrasion and discontinuity
lead migration/dislodgement
mortality due to inability to deliver therapy
muscle and/or nerve stimulation
myocardial damage
myocardial irritability
myopotential sensing
pericardial effusion
pericardial rub
pneumothorax
poor connection of the lead to the device, which may lead to oversensing, undersensing, or a loss of therapy
threshold elevation
thrombosis
thrombotic embolism
tissue necrosis
valve damage (particularly in fragile hearts)
venous occlusion
venous perforation
Other potential adverse events related to the lead include, but are not limited to, the following conditions:
insulation failure
lead conductor or electrode fracture

7 Drug information

7.1 Steroid mechanism of action

Steroid suppresses the inflammatory response that is believed to cause threshold rises typically associated with implanted pacing electrodes. Dexamethasone sodium phosphate and dexamethasone acetate are synthetic steroids of the glucocorticoid family. Glucocorticoids have potent anti-inflammatory actions via direct and indirect effects on major inflammatory cells. Glucocorticosteroids bind to a cytoplasmic glucocorticoid receptor as well as a membrane-bound receptor. Binding to the cytoplasmic receptor leads to receptor activation and translocation to the nucleus. The receptor interacts with specific DNA sequences within the regulatory regions of affected genes. Thus, glucocorticoids inhibit the production of multiple cell factors that are critical in generating the inflammatory response.

7.2 Pharmacodynamics of the Model 6947M lead

Pharmacokinetics – The pharmacokinetics (local drug levels
and systemic levels) of dexamethasone acetate (DXAC) and dexamethasone sodium phosphate (DSP) and their metabolites following lead implantation were not evaluated in human clinical trials.
The in-vivo elution profile of a tined pacemaker lead with a DSP monolithic controlled release device, based upon an assay of explanted leads, is shown in Mond and Stokes5.
Metabolism – The conversion of DSP to dexamethasone occurs within minutes; the conversion of DXAC to dexamethasone occurs within hours. The dexamethasone alcohol (dexamethasone) is the active glucocorticoid used in Medtronic leads. Steroid is applied to the tip and eluted through the electrode tip to the tissue interface where it will be used. Dexamethasone acetate and Dexamethasone sodium phosphate are hydrolyzed into dexamethasone, which is readily absorbed by the surrounding tissue and body fluids. Glucocorticoids, when given systemically, are eliminated primarily by renal excretion of inactive metabolites.
Mutagenesis, carcinogenicity and reproductive toxicology – The mutagenesis, carcinogenicity, and
reproductive toxicity of the Model 6947M lead have not been evaluated. However, the mutagenesis, carcinogenicity, and reproductive toxicity of dexamethasone acetate and dexamethasone sodium phosphate have been evaluated previously.
Carcinogenesis, mutagenesis, impairment of fertility – No adequate studies have been conducted in animals to determine whether corticosteroid have a potential for carcinogenesis (tumor
5
Mond, H. and Stokes, K.B., The Electrode - Tissue Interface: The Revolutionary Role of Steroid Elution, Pacing and Clinical Electrophysiology, Vol. 15, No. 1, pp 95-107
6
Page 7
initiation or promotion). Dexamethasone was genotoxic in assays for clastogenicity (including sister chromatid exchange in human lymphocytes) but not in an assay for mutagenicity in salmonella (Ames test).
Adrenocorticoids have been reported to increase or decrease the number and mobility of spermatozoa in some patients.
Pregnancy – Pregnancy category C. Dexamethasone acetate and Dexamethasone sodium phosphate have been shown to be teratogenic in many species when given in doses equivalent to the human dose. There are no adequate and well-controlled studies in pregnant women. Dexamethasone acetate and Dexamethasone sodium phosphate should be used during pregnancy only if the potential benefit justifies the potential risk to the fetus. Studies in mice, rats, and rabbits have shown that adrenocorticoids increase the incidence of cleft palate, placental insufficiency, and spontaneous abortions, and can decrease the intrauterine growth rate.
Nursing mothers – Systemically administered corticosteroids appear in human milk and could suppress growth, interfere with endogenous corticosteroid production, or cause other untoward effects in nursing infants. Because of the potential for serious adverse reactions in nursing infants from corticosteroids, a decision should be made whether to discontinue nursing or to use a non-steroidal lead, taking into account the importance of the lead and the drug to the mother.

8 Directions for use

8.1 Opening the package

Use the following steps to open the sterile package and inspect the lead:
1. Within the sterile field, open the sterile package and remove the lead and accessories.
2. Inspect the lead. Leads that are shorter than 85 cm should have 1 anchoring sleeve on the lead body. Leads that are 85 cm or longer should have 2 anchoring sleeves on the lead body.

8.2 Verifying the mechanical functioning of the helix electrode

Before implant, verify the mechanical functioning of the helix electrode using the following steps:
1. If needed, slide the stylet guide away from the connector pin, and then press both legs of the fixation tool together and place the most distal hole on the DF4 connector pin (Figure 3).
Figure 3.
Proper surgical procedures and sterile techniques are the responsibility of the medical professional. The following procedures are provided for information only. Some implant techniques vary according to physician preference and the patient’s anatomy or physical condition. Each physician must apply the information in these instructions according to professional medical training and experience.
The implant procedure generally includes the following steps:
Opening the package
Verifying the mechanical functioning of the helix electrode
Inserting the lead
Positioning a screw-in ventricular lead
Securing the helix electrode into the endocardium
Taking electrical measurements and defibrillation efficacy measurements
Anchoring the lead
Connecting the lead
Placing the device and lead into the pocket
2. Keep the lead body and the DF4 connector sleeve as straight as possible. Ensure that the stylet is fully inserted, then rotate the fixation tool clockwise until the helix electrode is fully extended (Figure 4). When the helix electrode is fully extended, approximately 1.5 to 2 helix coils are exposed.
Figure 4.
Caution: Do not severely bend the DF4 connector sleeve or
the lead body while extending the helix electrode. Caution: Overrotating the connector pin after the helix electrode is fully extended or fully retracted may damage the lead.
7
Page 8
The number of rotations required to extend or retract the helix
1
electrode increases proportionately with the length of the lead. Additional curvatures made to the stylet may increase the number of rotations needed to extend or retract the helix electrode. Rotation of the fixation tool should be stopped once full helix retraction is visually verified. Overretraction of the helix may result in the inability to extend the helix. If the helix is unable to extend, use a new lead. Note: To determine the number of rotations applied to the lead, count the number of rotations of the fixation tool. See Section 9 for the maximum number of rotations to extend or retract the helix electrode. During the initial helix electrode extension, the helix electrode may extend suddenly due to accumulated torque in the lead, or the helix electrode may require additional turns for extension.
3. Disconnect the fixation tool from the connector pin and release the proximal end of the lead body. Allow several seconds for relief of the residual torque in the lead.
4. After allowing for relief of the residual torque, reattach the fixation tool and rotate it counterclockwise until the helix electrode tip is retracted into the sheath.

8.3 Inserting the lead

Caution: Use care when handling the lead during insertion.
Do not severely bend, kink, or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pins.
Insert the lead using the following techniques:
1. Select a site for lead insertion. The lead may be inserted by venotomy through several different venous routes, including the right or left cephalic vein, the subclavian vein, or the external or internal jugular vein. Use the cephalic vein whenever possible to avoid lead damage in the first rib or clavicular (thoracic inlet) space.
Cautions:
Certain anatomical abnormalities, such as thoracic outlet syndrome, may also precipitate pinching and subsequent fracture of the lead.
When using a subclavian approach, avoid techniques that may damage the lead.
Place the insertion site as far lateral as possible to avoid clamping the lead body between the clavicle and the first rib (Figure 5).
Figure 5.
Do not force the lead if significant resistance is encountered during lead passage.
Do not use techniques such as adjusting the patient’s posture to facilitate lead passage. If resistance is encountered, it is recommended that an alternate venous entry site be used.
2. Insert the tapered end of a vein lifter into the incised vein and gently push the lead tip underneath and into the vein (Figure 6). Note: A percutaneous lead introducer (PLI) kit may be used to facilitate insertion. If an introducer is used, it should be at least 3.0 mm (9 French). Refer to the technical manual packaged with an appropriate percutaneous lead introducer for further instructions.
Figure 6.
3. Advance the lead into the right atrium using a straight stylet to facilitate movement through the veins.

8.4 Positioning a screw-in ventricular lead

Caution: Use care when handling the lead during positioning.
Do not severely bend, kink, or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pins.
Use the following steps to position the lead:
1. After the lead tip is passed into the atrium, advance the lead through the tricuspid valve. Replace the straight stylet with a gently curved stylet to add control when maneuvering the lead through the tricuspid valve. Caution: Do not use a sharp object to impart a curve to the distal end of the stylet. Imparting a curve to the stylet can be accomplished with a smooth-surface, sterile instrument (Figure 7).
Figure 7.
1 Suggested entry site
8
Note: When you pass the lead tip through the tricuspid valve or chordae tendineae, it may be difficult due to the flexible nature of the lead body. Rotate the lead body as the tip passes through the valve to facilitate passage.
Page 9
2. After the lead tip is in the ventricle, the curved stylet may be
1
2
1 2
replaced with a straight stylet. Withdraw the stylet slightly, to avoid using excessive tip force while achieving final electrode position. Avoid known infarcted or thin wall areas to minimize the occurrence of perforation.
3. Proper positioning of the helix electrode is essential for stable endocardial pacing. A satisfactory position usually is achieved when the lead tip points straight toward the apex, or when the distal end dips or bends slightly. Use fluoroscopy (lateral position) to ensure that the tip is not in a retrograde position or lodged in the coronary sinus. Note: With the helix electrode retracted, the distal end of the lead may be used to map a desirable site for electrode fixation. Mapping may reduce the need to repeatedly extend and fixate the helix electrode.
4. After placing the lead in a satisfactory position, extend the helix electrode by following the procedure in Section 8.5.

8.5 Securing the helix electrode into the endocardium

Secure the helix electrode using the following techniques:
1. If needed, slide the stylet guide away from the connector pin, and then press both legs of the fixation tool together and place the most distal hole on the DF4 connector pin (Figure 3).
2. Ensure that the stylet is inserted into the lead, and then press the lead tip against the endocardium by gently pushing the stylet and lead at the vein entry site.
3. Rotate the fixation tool clockwise until the helix electrode is fully extended (see Figure 4). Caution: Do not severely bend the DF4 connector sleeve or the lead body while extending the helix electrode. Use fluoroscopy to verify helix electrode exposure. Both a visual and fluoroscopic view of a fully retracted helix electrode is shown in Figure 8. Both a visual and fluoroscopic view of a fully extended helix electrode is shown in Figure 9. Closing of the space between the crimp sleeve and the indicator ring implies complete exposure of the helix electrode.
Figure 8.
1 Crimp sleeve 2 Indicator ring
Figure 9.
1 Crimp sleeve 2 Indicator ring
Cautions:
The number of rotations required to fully extend or retract the helix electrode is variable. Rotation should be stopped once full helix extension or retraction is verified with fluoroscopy as shown in Figure 8 and Figure 9. Overretraction of the helix, during initial implant or subsequent repositioning, may result in the inability to extend the helix. If the helix is unable to extend, replace with a new lead.
Do not exceed the recommended maximum number of rotations to extend or retract the helix electrode. Exceeding the maximum number may result in fracture or distortion of the inner conductor or helix electrode. Refer to Section 9 for the recommended maximum number of rotations.
Prolonged implant procedures or multiple repositionings may allow blood or body fluids to build up on the helix electrode mechanism. This may result in an increased number of rotations required to extend or retract the helix electrode.
4. Remove the fixation tool from the DF4 connector pin, and release the proximal end of the lead body. Allow several seconds for relief of the residual torque in the lead.
5. To assure helix electrode fixation, leave the stylet in place, hold the lead by the connector, and carefully rotate the lead body in 2 clockwise rotations.
6. Partially withdraw the stylet.
7. Obtain electrical measurements to verify satisfactory placement and electrode fixation. Refer to Section 8.6, “Taking electrical measurements and defibrillation efficacy measurements”, page 9.
8. Verify that the lead is affixed. Gently pull back on the lead, and check for resistance to verify fixation. A properly affixed helix electrode will remain in position. If the helix electrode is not properly affixed, the lead tip may become loose in the right ventricle.
9. If repositioning is required, reattach the fixation tool, and rotate counterclockwise until the helix electrode is retracted. Use fluoroscopy to verify withdrawal of the helix electrode before attempting to reposition.
10. After final positioning, make sure that the stylet and the fixation tool have been completely removed.
11. Obtain final electrical measurements. Refer to Section 8.6.

8.6 Taking electrical measurements and defibrillation efficacy measurements

Caution: Prior to taking electrical or defibrillation efficacy
measurements, move objects made from conductive materials, such as guide wires, away from all electrodes. Metal objects, such as guide wires, can short a lead and an active implantable device, causing electrical current to bypass the heart and possibly damage the implantable device and lead.
The ACI tool is used to facilitate accurate electrical measurements during implant. The lead package will contain one of two existing ACI tool designs as shown in Figure 10 and Figure 11. When attaching or removing the ACI tool:
9
Page 10
1. Be sure to grasp the grooves of the plastic housing (see Figure 10) and not the metallic contacts.
Figure 10.
2. Grasp the ACI tool in the most convenient location (see Figure 11).
Figure 11.
Figure 12.
Caution: The AccuRead tool reduces the risk of connector
damage, and reduces the risk of bridging and shorting that may occur while taking electrical measurements during implant. The potential for connector damage, bridging, and shorting is due to variations in analyzer cable terminals, as well as to the connector ring width and the proximity of the rings on the DF4 connector.
Note: The ACI tool may be removed or attached at any time during the procedure using the slit on the side of the tool (see Figure 12 or Figure 13).
1 Removing the ACI tool from the connector pin 2 Removing the ACI tool from the stylet using the slit on the side of the
tool (Do not let go of the ACI tool; the ACI tool can fall off the stylet.)
Figure 13.
1 Removing the ACI tool from the connector pin 2 Removing the ACI tool from the stylet using the slit on the side of the
tool
Use the following steps to take electrical measurements:
1. Ensure that the fixation tool is disconnected from the DF4 connector pin.
2. Ensure that the lead connector is completely inserted into the ACI tool. The connector pin will be completely accessible if the ACI tool is properly attached (see Figure 14 or Figure 15).
10
Page 11
Figure 14.
Figure 15.
1 When properly attached, all 3 contacts are visible through the ACI
tool openings.
3. Attach a surgical cable to the ACI tool. Line up the cable clips with the contacts on the ACI tool to ensure that accurate readings are obtained. (See Figure 22 for specific contacts.)
4. Use a testing device, such as a pacing system analyzer, for obtaining electrical measurements (see Table 1 for recommended measurements). For information on the use of the testing device, consult the product literature for that device.
5. After the electrical measurements are complete, remove the surgical cable from the ACI tool before removing the tool from the lead.
In order to demonstrate reliable defibrillation efficacy, obtain final defibrillation measurements for the lead system.

Table 1. Recommended measurements at implant (when using a pacing system analyzer)

Measurements required Acutea lead system
Capture threshold (at 0.5 ms pulse width)
Pacing impedance 200–1000 Ω 200–1000 Ω
Filtered R-wave amplitude (during sinus rhythm)
Slew rate ≥0.75 V/s ≥0.45 V/s
a
<30 days after implant.
b
>30 days after implant.
≤1.0 V ≤3.0 V
≥5 mV ≥3 mV
Chronicb lead sys-
tem
Values may vary depending upon lead type, implantable device settings, cardiac tissue condition, and drug interactions.
If electrical measurements do not stabilize to acceptable levels, it may be necessary to reposition the lead and repeat the testing procedure.
In order to keep patient morbidity and mortality to a minimum, patients should be rescued promptly with an external defibrillator if the implanted lead system fails to terminate a VF episode. At least 5 min should elapse between VF inductions.
For more information about obtaining electrical measurements, consult the product documentation supplied with the testing device.

8.7 Anchoring the lead

Caution: Use care when anchoring the lead.
Use only nonabsorbable sutures to anchor the lead.
Do not attempt to remove or cut the anchoring sleeve from the lead body.
During lead anchoring, take care to avoid dislodging the lead tip.
Do not secure sutures so tightly that they damage the vein, lead, or anchoring sleeve (Figure 16).
Do not tie a suture directly to the lead body (Figure 16).
Figure 16.
Use the following steps to anchor the lead using all 3 grooves:
Note: The anchoring sleeves contain a radiopaque substance, which allows visualization of the anchoring sleeve on a standard x-ray and may aid in follow-up examinations.
1. Position the distal anchoring sleeve against or near the vein.
2. Secure the anchoring sleeve to the lead body by tying a suture firmly in each of the 3 grooves (Figure 17).
Figure 17.
If initial electrical measurements deviate from the recommended values, it may be necessary to repeat the testing procedure 15 min after final positioning. Initial electrical measurements may deviate from the recommended values:
Initial impedance values may exceed the measuring capabilities of the testing device, resulting in an error message.
3. Use at least one additional suture in one of the grooves to secure the anchoring sleeve and lead body to the fascia.
11
Page 12
4. A second anchoring sleeve is provided with leads that are 85 cm or longer. For abdominal implants, the redundant lead body (for example, a curve for strain relief) should be placed just proximal to the first anchoring sleeve. Then, the second anchoring sleeve may be lightly sutured to the lead body and fascia to hold the curve in place. This procedure helps isolate the vein entry site from tension on the proximal end of the lead body.
5. A slit anchoring sleeve may be used in the device pocket to secure excess lead length. First, secure the anchoring sleeve to the lead body. Then, orient the slit toward the fascia and secure the anchoring sleeve to the fascia with sutures.

8.8 Connecting the lead

Use the following steps to connect the lead to an implantable device:
1. Make sure that the stylet and all accessories have been completely removed. When removing the accessories, grip the lead firmly just below the AccuRead tool on the connector to prevent dislodgement.
2. Push the lead or plug into the header block until the color band on the tip of the lead connector pin is visible in the pin viewing area (see Figure 18). The color band will be visible when the lead is fully inserted. Consult the product literature packaged with the implantable device for instructions on proper lead connections.

Figure 18. Lead connector pin viewing area

Figure 19.
Use the following steps to place the device and leads into the pocket:
1. To prevent undesirable twisting of the lead body, rotate the device to loosely wrap the excess lead length (Figure 20).
Figure 20.
2. Insert the device and leads into the pocket.
3. Before closing the pocket, verify sensing, pacing, cardioversion, and defibrillation efficacy.

8.10 Post-implant evaluation

1 Lead tip extends past setscrew block; lead connector pin is visible
in pin viewing area (color band may be used to verify full lead
insertion) 2 Setscrew block, located behind grommet 3 Lead

8.9 Placing the device and lead into the pocket

Caution: Use care when placing the device and leads into
the pocket.
Ensure that the leads do not leave the device at an acute angle.
Do not grip the lead or device with surgical instruments.
Do not coil the lead. Coiling the lead can twist the lead body and may result in lead dislodgement (Figure 19).
12
After implant, monitor the patient’s electrocardiogram until the patient is discharged. If a lead dislodges, it usually occurs during the immediate postoperative period.
Recommendations for verifying proper lead positioning include x-rays and pacing and sensing thresholds taken at pre-hospital discharge, 3 months after implant, and every 6 months thereafter.
In the event of a patient death, explant all implanted leads and devices and return them to Medtronic with a completed Product Information Report form. Call the appropriate phone number on the back cover if there are any questions on product handling procedures.

9 Specifications (nominal)

9.1 Detailed device description

Table 2. Specifications (nominal)

Parameter Model 6947M
Type Quadripolar
Position Right ventricle
Page 13
Table 2. Specifications (nominal) (continued)
1
234
5
Parameter Model 6947M
Fixation Extendable/retractable helix
Length 45–110 cm
Connector Quadripolar/true
Materials Conductors: MP35N coil
Electrodes (pace, sense): Platinized platinum alloy
DF4 pin and rings: MP35N
Steroid Type: Dexamethasone acetate and
Conductor resistances
Helix length (extended) 1.8 mm
Diameters Lead body: 2.8 mm
Lead introducer (recommended size)
Pacing (unipolar): 27.9 Ω (62 cm)
without guide wire: 3.0 mm (9.0 French)
bipolar:
Insulation: Silicone, PTFE, ETFE
Overlay: Polyurethane
Seal Zone: PEEK
RV/SVC coils: Platinum-clad tantalum
DF4 pin: MP35N
Amount: 685 µg of dexamethasone ace-
Steroid binder: Silicone
Pacing (bipolar): 29.3 Ω (62 cm)
Defibrillation: 1.4 Ω (62 cm)
with guide wire: 3.7 mm (11.0 French)
Four-pole inline (DF4-LLHH)
MP35N composite cables
dexamethasone sodium phos­phate
tate 59 µg of dexamethasone sodium phosphate
Tip: 2.8 mm
Helix: 1.4 mm

Table 3. Maximum number of rotations to extend or retract the helix electrode

Lead length Number of rotations
49 cm 17
55 cm 18
62 cm 20
72 cm 22
97 cm 27

Table 4. Respective electrode distances

Helix electrode to ring electrode 8 mm
Helix electrode to RV coil electrode 12 mm
Helix electrode to SVC coil electrode 180 mm

Figure 21. Model 6947M distal lead components

1 Helix electrode; surface area: 5.7 mm 2 Ring electrode; surface area: 25.2 mm 3 RV coil electrode; length: 57 mm; surface area: 614 mm2; electrical
shadow area: 506 mm
4 SVC coil electrode; length: 80 mm; surface area: 860 mm2; electrical
shadow area: 709 mm
5 Anchoring sleeve; leads 85 cm or longer have 2 anchoring sleeves
2
2
2
2
13
Page 14

Figure 22. Model 6947M proximal lead components

EC REP

12 Explanation of symbols on package labeling

Refer to the package labels to see which symbols apply to this product.

Table 5. Explanation of symbols on package labeling

Symbol Explanation
Conformité Européenne (European Conformity). This symbol means that the device fully complies with AIMD Directive 90/385/EEC (0123).
Do not use if package is damaged
Do not reuse
Upper limit of temperature
Open here
Sterilized using ethylene oxide
Caution
1 AccuRead tool 2 Connector pin 3 Stylet 4 SVC contact 5 RV contact 6 Ring (+) contact 7 Tip (-)

10 Medtronic warranty

For complete warranty information, see the accompanying warranty document.

11 Service

Medtronic employs highly trained representatives and engineers located throughout the world to serve you and, upon request, to provide training to qualified hospital personnel in the use of Medtronic products. Medtronic also maintains a professional staff to provide technical consultation to product users. For more information, contact your local Medtronic representative, or call or write Medtronic at the appropriate telephone number or address listed on the back cover.
14
Consult instructions for use
Date of manufacture
Manufacturer
Authorized representative in the European com­munity
For US audiences only
Use by
Reorder number
Serial number
Lot number
Page 15
Table 5. Explanation of symbols on package labeling (continued)
Symbol Explanation
Package contents
Product documentation
Accessories
Inner diameter
Lead
Lead length
Transvenous ventricular lead
Transvenous lead with two defibrillation electrodes
Pace
Sense
Defibrillation
Extendable and retractable screw-in
Steroid eluting
Lead introducer
Lead introducer with guide wire
15
Page 16
Page 17
Page 18
World Headquarters
*M953938A001*
Medtronic, Inc. 710 Medtronic Parkway Minneapolis, MN 55432 USA www.medtronic.com Tel. +1 763 514 4000 Fax +1 763 514 4879
Medtronic USA, Inc.
Toll-free in the USA (24-hour technical consultation for physicians and medical professionals) Bradycardia: +1 800 505 4636 Tachycardia: +1 800 723 4636
Europe/Africa/Middle East Headquarters
Medtronic International Trading Sàrl Route du Molliau 31 Case Postale 84 CH-1131 Tolochenaz Switzerland www.medtronic.com Tel. +41 21 802 7000 Fax +41 21 802 7900
Medtronic E.C. Authorized Representative
Medtronic B.V. Earl Bakkenstraat 10 6422 PJ Heerlen The Netherlands Tel. +31 45 566 8000 Fax +31 45 566 8668
Technical manuals: www.medtronic.com/manuals
© Medtronic, Inc. 2012 M953938A001A 2012-10-05
Loading...