Medtronic 6947-58 Technical Manual

SPRINT QUATTRO SECURE MRI™ SURESCAN™ 6947
MR Conditional, steroid-eluting, quadripolar, screw-in, ventricular lead with RV/SVC defibrillation coil electrodes
Technical Manual Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
The following list includes trademarks or registered trademarks of Medtronic in the United States and possibly in other countries. All other trademarks are the property of their respective owners.
Evera, Evera MRI, Medtronic, Sprint Quattro, Sprint Quattro Secure, Sprint Quattro Secure MRI, SureScan, Tensi-Lock

Contents

9 Implant procedure 8 10 Specifications (nominal) 13 11 Medtronic warranty 14 12 Service 14

1 Description

The Medtronic Sprint Quattro Secure MRI SureScan 6947 lead is a steroid-eluting, quadripolar, screw-in, ventricular lead with right ventricular (RV) and superior vena cava (SVC) defibrillation coil electrodes. This lead is designed for pacing, sensing, cardioversion, and defibrillation therapies. The lead has been tested for use in the Magnetic Resonance Imaging (MRI) environment. The following lead lengths are MR Conditional: 58 cm and 65 cm. Other lead lengths are not MR conditional. The lead also features Tensi-Lock and silicone-backed defibrillation coils1.
The lead features an extendable and retractable helix electrode, silicone insulation, and parallel conductors. The 4 electrodes of the lead are the helix, ring, RV coil, and SVC coil.
The tip electrode is common to the connector pin of the IS-1 BI connector.
The ring electrode is common to the connector ring of the IS-1 BI connector.
The RV coil electrode is common to the connector pin of the RV DF-13 connector (red band).
The SVC coil electrode is common to the connector pin of the SVC DF-1 connector (blue band).
The RV and SVC coils deliver cardioversion and defibrillation therapies. Pacing and sensing occur between the helix and ring electrodes.
The helix electrode is made of platinized platinum alloy that can be actively fixed into the endocardium. The helix electrode can be extended or retracted by rotating the IS-1 connector pin with the fixation tool.
The IS-1 bipolar leg of the trifurcation features a lumen for stylet passage. The DF-1 connectors do not accept stylets.
The distal tip contains a nominal dosage of 685 µg of dexamethasone acetate and 59 µg of dexamethasone sodium phosphate. Upon exposure to body fluids, the steroids elute from the lead tip. The steroids are known to suppress the inflammatory response that is believed to cause threshold rises typically associated with implanted pacing electrodes.

1.1 Medtronic SureScan system

A complete SureScan defibrillation system is required for use in the MR environment. A complete SureScan system includes a Medtronic SureScan device with the appropriate number of Medtronic SureScan leads.The Model 6947 lead is
part of the Medtronic SureScan defibrillation system. Labeling for SureScan system components displays the SureScan logo and the MR Conditional symbol. To verify that components are part of a SureScan system, visit http://www.mrisurescan.com. Any other combination may result in a hazard to the patient during an MRI scan.
SureScan logo
MR Conditional symbol. The Medtronic SureScan sys­tem is MR Conditional and is designed to allow implan­ted patients to undergo an MRI scan under the speci­fied MRI conditions for use.
The MRI SureScan feature permits a mode of operation that allows a patient with a SureScan device to be safely scanned by an MRI machine while the device continues to provide appropriate pacing. When programmed to On, MRI SureScan operation disables arrhythmia detection, magnet mode, and all user-defined diagnostics. Before performing an MRI scan,
2
refer to the SureScan system MRI technical manual for important information about procedures and MRI-specific warnings and precautions.

1.2 Package contents

Leads and accessories are supplied sterile. Each package contains the following items:
1 lead with 1 radiopaque anchoring sleeve, stylet, and stylet guide
1 slit anchoring sleeve
1 vein lifter
2 fixation tools
2 pin caps
extra stylets
product literature
1
Tensi-Lock is an exclusive Medtronic design feature that utilizes lead body cables to act like a built-in locking stylet and add tensile strength to the lead.
2
IS-1 refers to the International Connector Standard ISO 5841-3, whereby pulse generators and leads so designated are assured of a basic mechanical fit.
3
DF-1 refers to the International Connector Standard ISO 11318, whereby pulse generators and leads so designated are assured of a basic mechanical fit.
3

1.3 Accessory descriptions

Dispose of all single-use accessories according to local environmental requirements.
Anchoring sleeve – An anchoring sleeve secures the lead to prevent it from moving and protects the lead insulation and conductors from damage caused by tight sutures.
Fixation tool – A fixation tool facilitates connector pin rotation.
Pin cap – A pin cap covers and insulates unused connector pins.
Slit anchoring sleeve – A slit anchoring sleeve secures excess
lead length in the device pocket.
Stylet – A stylet provides additional stiffness and controlled flexibility for maneuvering the lead into position. Each stylet knob is labeled with the stylet diameter and corresponding lead length.
Stylet guide – A stylet guide facilitates stylet insertion into the lead.
Vein lifter – A vein lifter facilitates lead insertion into a vein.

2 Indications

The lead is intended for single, long-term use in the right ventricle.
This lead has application for patients for whom implantable cardioverter defibrillators (ICDs) are indicated.

3 Drug component description

The active ingredients in the 6947 lead are dexamethasone acetate [21-(acetyloxy)-9-fluoro-11β, 17-trihydroxy-16α-methylpregna-1,4-diene-3,20-dione] and dexamethasone sodium phosphate [9-fluoro-11β, 17-dihydroxy-16α-methyl-21-(phosphonooxy) pregna-1,4-diene-3,20-diene-3,20-dione disodium salt].
The structural formula for each of these steroids is shown below:

Figure 1. Structural formula for dexamethasone sodium phosphate (DSP)

Dexamethasone sodium phosphate is an inorganic ester of dexamethasone, a synthetic adrenocortical steroid. Dexamethasone sodium phosphate is a white or slightly yellow crystalline powder. It is freely soluble in water and is very hygroscopic.

Figure 2. Structural formula for dexamethaxone acetate (DXAC)

Dexamethasone acetate is a white to practically white, odorless powder. Dexamethasone acetate is a practically insoluble ester of dexamethasone, a synthetic adrenocortical steroid.
The maximum dosage of dexamethasone sodium phosphate and dexamethasone acetate is less than 1.0 mg per lead.

4 Contraindications

Before performing an MRI scan, refer to the SureScan system MRI Technical Manual for MRI-specific contraindications.
Atrial use – The lead is contraindicated for the sole use of
detection and treatment of atrial arrhythmias.
Ventricular use – The lead is contraindicated for ventricular use in patients with tricuspid valvular disease or a tricuspid mechanical heart valve.
Transient ventricular tachyarrhythmias – The lead is contraindicated for patients with transient ventricular tachyarrhythmias due to reversible causes (drug intoxication, electrolyte imbalance, sepsis, hypoxia) or other factors (myocardial infarction, electric shock).
Steroid use – The lead is contraindicated in patients for whom a combined dose of 1.0 mg of dexamethasone sodium phosphate and dexamethasone acetate may be contraindicated.
4

5 Warnings and precautions

A complete SureScan defibrillation system is required for use in the MR environment. A complete SureScan defibrillation system includes a SureScan device with the appropriate number of SureScan leads. To verify that
components are part of a SureScan system, visit http://mrisurescan.com. Any other combination may result in a hazard to the patient during an MRI scan.
Single use – The lead and accessories are for single use only.
Inspecting the sterile package – Inspect the sterile package
with care before opening it.
If the seal or package is damaged, contact a Medtronic representative.
Store at 25 °C (77 °F). Excursions from this storage temperature are permitted in the range of 15 to 30 °C (59 to 86 °F). (See USP Controlled Room Temperature.) According to USP excursion conditions, transient spikes up to 40 °C (104 °F) are permitted as long as they do not exceed 24 hours.
Do not use the product after its expiration date.
Sterilization – Medtronic has sterilized the package contents with ethylene oxide before shipment. This lead is for single use only and is not intended to be resterilized.
Electrophysiologic testing – Prior to lead implant, it is strongly recommended that patients undergo a complete cardiac evaluation, which should include electrophysiologic testing. Also, electrophysiologic evaluation and testing of the safety and efficacy of the proposed pacing, cardioversion, or defibrillation therapies are recommended during and after the implant of the system.
Steroid use – It has not been determined whether the warnings, precautions, or complications usually associated with injectable dexamethasone sodium phosphate or dexamethasone acetate apply to the use of this highly localized, controlled-release lead. For a list of potentially adverse effects, refer to the Physicians’ Desk Reference.
Handling the steroid tip – Avoid reducing the amount of steroid available before implanting the lead. Reducing the available amount of steroid may adversely affect low-threshold performance.
Do not allow the electrode surface to come in contact with surface contaminants.
Do not wipe or immerse the electrode in fluid, except blood, at the time of implant.
Handling the lead – Handle the lead with care at all times.
Protect the lead from materials shedding particles such as lint and dust. Lead insulators attract these particles.
Handle the lead with sterile surgical gloves that have been rinsed in sterile water or a comparable substance.
Do not severely bend, kink, or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pins.
Do not immerse the leads in mineral oil, silicone oil, or any other liquid, except blood, at the time of implantation.
Do not implant the lead without first verifying the mechanical functioning of the helix electrode. Refer to Section 9.2, “Verifying the mechanical functioning of the helix electrode”, page 9 for complete instructions.
Do not rotate the helix electrode after it is fully extended or fully retracted. The number of rotations required to fully extend or retract the helix electrode is variable. Furthermore, do not exceed the recommended maximum number of rotations to extend or retract the helix electrode. Exceeding the maximum number may result in fracture or distortion of the inner
conductor or helix electrode. Refer to Section 10.1 for the recommended maximum number of rotations.
Inserting the lead using a lead introducer that has a hemostasis valve may require a larger introducer than the size recommended. To avoid distortion of the coil electrode, do not withdraw the lead through a hemostasis valve.
Handling the stylet – Handle the stylet with care at all times.
Do not use a sharp object to impart a curve to the distal end of the stylet.
Do not use excessive force or surgical instruments when inserting the stylet into the lead.
Avoid overbending or kinking the stylet.
Use a new stylet when blood or other fluids accumulate on the stylet. Accumulated blood or other fluids may damage the lead or cause difficulty in passing the stylet into the lead.
Necessary hospital equipment – Keep external defibrillation equipment nearby for immediate use during acute lead system testing, the implant procedure, or whenever arrhythmias are possible or intentionally induced during post-implant testing.
Line-powered and battery-powered equipment – An implanted lead forms a direct current path to the myocardium. During lead implant and testing, use only battery-powered equipment or line-powered equipment specifically designed for this purpose to protect against fibrillation that may be caused by alternating currents. Line-powered equipment used in the vicinity of the patient must be properly grounded. Lead connector pins must be insulated from any leakage currents that may arise from line-powered equipment.
Concurrent devices – Output pulses, especially from unipolar devices, may adversely affect device sensing capabilities. If a patient requires a separate stimulation device, either permanent or temporary, allow enough space between the leads of the separate systems to avoid interference in the sensing capabilities of the devices. Previously implanted pulse generators and implantable cardioverter defibrillators should generally be explanted.
Magnetic resonance imaging (MRI) – An MRI is a type of medical imaging that uses magnetic fields to create an internal view of the body. If certain criteria are met and the warnings and precautions provided by Medtronic are followed, patients with an MR Conditional device and lead system are able to undergo an MRI scan; for details, refer to the SureScan MRI technical manual that Medtronic provides for an MR Conditional device.
Diathermy treatment (including therapeutic ultrasound) –
Diathermy is a treatment that involves the therapeutic heating of body tissues. Diathermy treatments include high frequency, short wave, microwave, and therapeutic ultrasound. Except for therapeutic ultrasound, do not use diathermy treatments on cardiac device patients. Diathermy treatments may result in serious injury or damage to an implanted device and lead system. Therapeutic ultrasound (including physiotherapy, high intensity therapeutic ultrasound, and high intensity focused ultrasound), is the use of ultrasound at higher energies than diagnostic ultrasound to bring heat or agitation into the body. Therapeutic ultrasound is acceptable if treatment is performed with a minimum
5
separation distance of 15 cm (6 in) between the applicator and the implanted device and lead system, as long as the ultrasonic beam is pointing away from the device and lead system.
Chronic lead removal and the SureScan defibrillation system – When implanting a SureScan defibrillation system,
consider the risks associated with removing previously implanted leads before doing so. Abandoned leads or previously implanted non-SureScan labeled leads compromise the ability to safely scan the SureScan defibrillation system during MRI scans.
Chronic repositioning or removal of a screw-in lead –
Proceed with extreme caution if a lead must be removed or repositioned. Chronic repositioning or removal of screw-in transvenous leads may not be possible because of blood or fibrotic tissue development into the helix mechanism on the lead. In most clinical situations, it is preferable to abandon unused leads in place. Return all removed leads, unused leads, or lead sections to Medtronic for analysis.
Note: If a helix does not disengage from the endocardium by rotating the connector pin, rotating the lead body counterclockwise may withdraw the helix and decrease the possibility of damage to cardiovascular structures during removal.
Lead removal may result in avulsion of the endocardium, valve, or vein.
Lead junctions may separate, leaving the lead tip and bare wire in the heart or vein.
Chronic repositioning of a lead may adversely affect the low-threshold performance of a steroid lead.
An abandoned lead should be capped so that the lead does not transmit electrical signals.
Severed leads should have the remaining lead end sealed and the lead body sutured to adjacent tissue.
External defibrillation and cardioversion – External defibrillation and cardioversion are therapies that deliver an electrical shock to the heart to convert an abnormal heart rhythm to a normal rhythm.
Medtronic cardiac devices are designed to withstand exposure to external defibrillation and cardioversion. While damage to an implanted system from an external shock is rare, the probability increases with increased energy levels. These procedures may also temporarily or permanently elevate pacing thresholds or temporarily or permanently damage the myocardium. If external defibrillation or cardioversion are required, consider the following precautions:
Use the lowest clinically appropriate energy.
Position the patches or paddles no closer than 15 cm (6 in) to the device.
Position the patches or paddles perpendicular to the device and lead system.
If an external defibrillation or cardioversion is delivered within 15 cm (6 in) of the device, use a Medtronic programmer to evaluate the device and lead system.

6 Drug information

Steroid mechanism of action – Steroid suppresses the
inflammatory response that is believed to cause threshold rises typically associated with implanted pacing electrodes. Dexamethasone sodium phosphate and dexamethasone acetate are synthetic steroids of the glucocorticoid family. Glucocorticoids have potent anti-inflammatory actions via direct and indirect effects on major inflammatory cells. Glucocorticosteroids bind to a cytoplasmic glucocorticoid receptor as well as a membrane-bound receptor. Binding to the cytoplasmic receptor leads to receptor activation and translocation to the nucleus. The receptor interacts with specific DNA sequences within the regulatory regions of affected genes. Thus, glucocorticoids inhibit the production of multiple cell factors that are critical in generating the inflammatory response.

7 Pharmacodynamics of the Model 6947 lead

Pharmacokinetics – The pharmacokinetics (local drug levels
and systemic levels) of dexamethasone acetate (DXAC) and dexamethasone sodium phosphate (DSP) and their metabolites following lead implantation were not evaluated in human clinical trials.
The in-vivo elution profile of a tined pacemaker lead with a DSP monolithic controlled release device, based upon an assay of explanted leads, is shown in Mond and Stokes4.
Metabolism – The conversion of DSP to dexamethasone occurs within minutes; the conversion of DXAC to dexamethasone occurs within hours. The dexamethasone alcohol (dexamethasone) is the active glucocorticoid used in Medtronic leads. Steroid is applied to the tip and eluted through the electrode tip to the tissue interface where it will be used. Dexamethasone acetate and Dexamethasone sodium phosphate are hydrolyzed into dexamethasone, which is readily absorbed by the surrounding tissue and body fluids. Glucocorticoids, when given systemically, are eliminated primarily by renal excretion of inactive metabolites.
Mutagenesis, carcinogenicity and reproductive toxicology –
The mutagenesis, carcinogenicity, and reproductive toxicity of the Model 6947 lead have not been evaluated. However, the mutagenesis, carcinogenicity, and reproductive toxicity of dexamethasone acetate and dexamethasone sodium phosphate have been evaluated previously.
Carcinogenesis, mutagenesis, impairment of fertility – No adequate studies have been conducted in animals to determine whether corticosteroid have a potential for carcinogenesis (tumor initiation or promotion). Dexamethasone was genotoxic in assays for clastogenicity (including sister chromatid exchange in human lymphocytes) but not in an assay for mutagenicity in salmonella (Ames test).
Adrenocorticoids have been reported to increase or decrease the number and mobility of spermatozoa in some patients.
4
Mond, H. and Stokes, K.B., The Electrode - Tissue Interface: The Revolutionary Role of Steroid Elution, Pacing and Clinical Electrophysiology, Vol. 15, No. 1, pp 95-107
6
Pregnancy – Pregnancy category C. Dexamethasone acetate and Dexamethasone sodium phosphate have been shown to be teratogenic in many species when given in doses equivalent to the human dose. There are no adequate and well-controlled studies in pregnant women. Dexamethasone acetate and Dexamethasone sodium phosphate should be used during pregnancy only if the potential benefit justifies the potential risk to the fetus. Studies in mice, rats, and rabbits have shown that adrenocorticoids increase the incidence of cleft palate, placental insufficiency, and spontaneous abortions, and can decrease the intrauterine growth rate.
Nursing mothers – Systemically administered corticosteroids appear in human milk and could suppress growth, interfere with endogenous corticosteroid production, or cause other untoward effects in nursing infants. Because of the potential for serious adverse reactions in nursing infants from corticosteroids, a decision should be made whether to discontinue nursing or to use a non-steroidal lead, taking into account the importance of the lead and the drug to the mother.

8 Adverse events and clinical trial data

Observed adverse events – The Model 6947 Lead was utilized
in a prospective, nonrandomized, multicenter trial to assess the handling and performance of the Sprint Quattro Secure Model 6947 Lead.
A total of 22 cardiovascular-related adverse events were reported. Two of the 22 events were ventricular lead-related and both occurred at implant. One event was a microdislodgement and one was induced VT as a result of lead manipulation.
Two deaths occurred in this patient group during the follow-up period. Both deaths were classified as non-sudden cardiac and were judged to be non-system related by an independent advisory committee.
Information regarding clinical studies and adverse events related to this lead is available at www.medtronic.com/manuals. The following clinical studies are related to this lead:
Model 6932 RV Lead clinical study
Model 6947 RV Lead clinical study
Evera MRI System study – This clinical study was executed to confirm safety and efficacy of the Evera MRI system in the clinical MRI environment when subjects receive MRI scans up to 2 W/kg SAR without positioning restrictions (MRI scans may occur anywhere on the body), providing support for the Sprint Quattro Secure Model 6947 device.
If you do not have web access, a printed copy of the related clinical study summary can be obtained from your Medtronic representative, or you can call the toll-free number located on the back cover.
Potential adverse events – The potential adverse events associated with the use of transvenous leads and pacing systems include, but are not limited to, the following events:
acceleration of tachyarrhythmias (caused by device)
air embolism
bleeding
body rejection phenomena, including local tissue reaction
cardiac dissection
cardiac perforation
cardiac tamponade
chronic nerve damage
constrictive pericarditis
death
device migration
endocarditis
erosion
excessive fibrotic tissue growth
extrusion
fibrillation or other arrhythmias
fluid accumulation
formation of hematomas/seromas or cysts
heart block
heart wall or vein wall rupture
hemothorax
infection
keloid formation
lead abrasion and discontinuity
lead migration/dislodgement
mortality due to inability to deliver therapy
muscle and/or nerve stimulation
myocardial damage
myocardial irritability
myopotential sensing
pericardial effusion
pericardial rub
pneumothorax
poor connection of the lead to the device, which may lead to oversensing, undersensing, or a loss of therapy
threshold elevation
thrombosis
thrombotic embolism
tissue necrosis
valve damage (particularly in fragile hearts)
venous occlusion
venous perforation
Other potential adverse events related to the lead include, but are not limited to, the following conditions:
insulation failure
lead conductor or electrode fracture
lead dislodgement
poor connection to the device, which may lead to oversensing, undersensing, or a loss of therapy
The Model 6947 clinical study was a prospective, nonrandomized, multicenter trial assessing the lead handling and performance of the Sprint Quattro Secure Model 6947 Lead.
The Model 6947 lead was implanted in 80 patients at 15 investigative centers in the United States and at 2 investigative centers in Canada between April 27, 2001 and August 2, 2001.
7
Patients were included in this study if they met the following criteria: 1) able to receive a pectoral implant and 2) survived at least one episode of cardiac arrest due to a ventricular tachyarrhythmia; or had poorly tolerated, sustained ventricular tachycardia that occurred spontaneously; or had poorly tolerated, sustained ventricular tachycardia that could be induced.
Patients studied – The Model 6947 study population consisted of 63 males and 17 females. The mean age was 64.4 years. The most frequently reported indication for implant was inducible ventricular tachycardia without sudden cardiac death (SCD) (47.5%). The mean ejection fraction was 35.5%. The most frequently reported NYHA classifications were Class I (35.0%) and Class II (36.3%).
Cardiovascular history included coronary artery disease with myocardial infarction (62.5%), hypertension (58.8%), cardiomyopathy (65.0%), congestive heart failure (47.5%), syncope/presyncope (46.3%) and previous cardiac surgery (68.8%).
Objectives – The objectives of the study were to report the following:
Pacing thresholds
R-wave amplitudes
Pacing impedance
Lead handling
Adverse events
Methods – Pulse width thresholds, R-wave amplitudes, and pacing impedances were measured at implant and at one month post-implant. Adverse events were collected throughout the study. A lead handling questionnaire was completed by the implanting physician at each implant. The implanting physicians evaluated the performance of the Model 6947 lead with regard to:
Ease of lead insertion into the vein
Ease of helix extension
Visibility of helix extension
Steerability
Torqueability
Lead placement time
Ability to traverse the tricuspid annulus
Comfort level with handling the lead
Slipperiness of lead surfaces
Stiffness of the lead
Ease of obtaining adequate R-wave sensing
Ease of obtaining adequate VF sensing
Ease of obtaining adequate DFTs
Overall ease of lead placement
Results – The mean follow-up duration was 0.96 months (range: 0.00 – 1.71 months) with a cumulative follow-up duration of 76.82 months. The pace/sense measurements are summarized in Table 1.

Table 1. Pace/Sense measurements

Implant One month
Pulse-width threshold at one volt
N 80 61
Median 0.20 ms 0.20 ms
25th – 75th Percentile (0.20 – 0.20) (0.20 – 0.30)
Range (0.03 – 0.60) (0.20 – 0.80)
R-wave amplitude (EGM)
N 78 64
Median 8.0 mV 9.3 mV
25th – 75th Percentile (7.0 – 11.0) (7.5 – 12.8)
Range (3.0 – 24.0) (3.5 – 20.0)
Pacing lead impedance
N 80 66
Median 564.0 Ω 481.0 Ω
25th – 75th Percentile (481.0 – 611.0) (444.0 – 521.0)
Range (378.0 – 985.0) (323.0 – 985.0)
On the lead handling survey, for all items rated, the adjusted rating of the Model 6947 lead fell between 1 (very good) and 2 (excellent). The adjusted rating for each item is an average across physicians, accounting for multiple responses per physician.
The overall ease of lead placement was considered good, very good, or excellent by all implanting physicians, with an adjusted rating of 1.7 (the minimum rating on the questionnaire was –2.0 and the maximum rating was 2.0).
Conclusion – In this clinical study, the Model 6947 lead demonstrated acceptable clinical performance. Through questionnaire responses, implanting physicians verified acceptable overall lead performance and handling during the implant procedure.

9 Implant procedure

Warning: Before implanting a SureScan system, consider the
risks associated with removing previously implanted leads. Abandoned leads or previously implanted leads not tested for MRI compatibility compromise the ability to safely scan the SureScan system during MRI scans.
Proper surgical procedures and sterile techniques are the responsibility of the medical professional. The following procedures are provided for information only. Some implant techniques vary according to physician preference and the patient’s anatomy or physical condition. Each physician must apply the information in these instructions according to professional medical training and experience.
8

9.1 Opening the package

Use the following steps to open the sterile package and inspect the lead:
1. Within the sterile field, open the sterile package and remove the lead and accessories.
2. Inspect the lead to verify that there is an anchoring sleeve on the lead body.

9.2 Verifying the mechanical functioning of the helix electrode

Before implantation, verify the mechanical functioning of the helix electrode:
1. With a stylet inserted into the lead, press both legs of the fixation tool together and place the most distal hole of the fixation tool on the connector pin (Figure 3). Note: The stylet guide may be removed by gently pulling it off. To reattach a stylet guide, gently push it onto the connector pin.
Figure 3.
2. Hold the IS-1 connector leg of the lead with the thumb on one side and four fingers on the other side. Keep the lead body and the IS-1 connector leg as straight as possible (Figure 3). Ensure that the stylet is fully inserted, then rotate the fixation tool clockwise until the helix electrode is fully extended. When the helix electrode is fully extended, approximately 1.5 to 2 helix coils are exposed. Caution: Do not severely bend the IS-1 connector leg or the lead body while extending the helix electrode (Figure 4). If the lead is bent on either side of the lead trifurcation during helix electrode extension or retraction, the lead may be damaged.
Figure 4.
Caution: Over-rotating the connector pin after the helix
electrode is fully extended or fully retracted may damage the lead. The number of rotations required to extend or retract the helix electrode increases proportionately with the length of the lead. Additional curvatures made to the stylet may increase the number of rotations needed to extend or retract the helix electrode. Rotation of the fixation tool should be stopped once full helix retraction is visually verified. Over-retraction of the helix may result in the inability to extend the helix. If the helix is unable to extend, use a new lead. During the initial helix electrode extension, the helix electrode may extend suddenly due to accumulated torque in the lead, or the helix electrode may require additional turns for extension.
3. Disconnect the fixation tool from the connector pin and release the proximal end of the lead body. Allow several seconds for relief of the residual torque in the lead.
4. After allowing for relief of the residual torque, reattach the fixation tool and rotate it counterclockwise until the helix electrode tip is retracted into the sheath.

9.3 Inserting the lead

Caution: Use care when handling the lead during insertion.
Do not severely bend, kink, or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pins.
Insert the lead using the following techniques:
1. Select a site for lead insertion. The lead may be inserted by venotomy through several different venous routes, including the right or left cephalic vein, the subclavian vein, or the external or internal jugular vein. Use the cephalic vein whenever possible to avoid lead damage in the first rib or clavicular (thoracic inlet) space.
9
Cautions:
1
Certain anatomical abnormalities, such as thoracic outlet syndrome, may also precipitate pinching and subsequent fracture of the lead.
When using a subclavian approach, avoid techniques that may damage the lead.
Place the insertion site as far lateral as possible to avoid clamping the lead body between the clavicle and the first rib (Figure 5).
Figure 5.
1 Suggested entry site
Do not force the lead if significant resistance is encountered during lead passage.
Do not use techniques such as adjusting the patient’s posture to facilitate lead passage. If resistance is encountered, it is recommended that an alternate venous entry site be used.
2. Insert the tapered end of a vein lifter into the incised vein and gently push the lead tip underneath and into the vein (Figure 6). Note: A percutaneous lead introducer (PLI) kit may be used to facilitate insertion. Refer to the technical manual packaged with an appropriate percutaneous lead introducer for further instructions.
Figure 6.
3. Advance the lead into the right atrium using a straight stylet to facilitate movement through the veins.

9.4 Positioning a screw-in ventricular lead

Caution: Use care when handling the lead during positioning.
Do not severely bend, kink, or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pins.
Use the following steps to position the lead:
1. After the lead tip is passed into the atrium, advance the lead through the tricuspid valve. Replace the straight stylet with a gently curved stylet to add control when maneuvering the lead through the tricuspid valve. Caution: Do not use a sharp object to impart a curve to the distal end of the stylet. Imparting a curve to the stylet can be accomplished with a smooth-surface, sterile instrument (Figure 7).
Figure 7.
Note: When you pass the lead tip through the tricuspid valve
or chordae tendineae, it may be difficult due to the flexible nature of the lead body. Rotate the lead body as the tip passes through the valve to facilitate passage.
2. After the lead tip is in the ventricle, the curved stylet may be replaced with a straight stylet. Withdraw the stylet slightly, to avoid using excessive tip force while achieving final electrode position. Avoid known infarcted or thin wall areas to minimize the occurrence of perforation.
3. Proper positioning of the helix electrode is essential for stable endocardial pacing. A satisfactory position usually is achieved when the lead tip points straight toward the apex, or when the distal end dips or bends slightly. Use fluoroscopy (lateral position) to ensure that the tip is not in a retrograde position or lodged in the coronary sinus. Note: With the helix electrode retracted, the distal end of the lead may be used to map a desirable site for electrode fixation. Mapping may reduce the need to repeatedly extend and fixate the helix electrode.
4. After placing the lead in a satisfactory position, extend the helix electrode by following the procedure in Section 9.5.

9.5 Securing the helix electrode into the endocardium

Secure the helix electrode using the following techniques:
1. Press both legs of the fixation tool together and place the most distal hole on the IS-1 connector pin (Step 1).
10
2. Press the lead tip against the endocardium by gently pushing
A B
AB
the stylet and lead at the vein entry site.
3. Rotate the fixation tool clockwise until the helix electrode is fully extended. Caution: Do not severely bend the IS-1 connector leg or the lead body while extending the helix electrode. If the lead is bent on either side of the lead trifurcation during helix electrode extension or retraction, the lead may be damaged. Use fluoroscopy to verify helix electrode extension. The fluoroscope head may need to be rotated to obtain an adequate view. Both a visual and fluoroscopic view of a fully retracted and extended helix electrode is shown in Figure 8. Closing of the space between the indicator stop (A) and the indicator ring (B) implies complete extension of the helix electrode.
Figure 8.
1 Fully retracted visual 2 Fully retracted fluoroscopic 3 Fully extended visual 4 Fully extended fluoroscopic
Cautions:
The number of rotations required to fully extend or retract the helix electrode is variable. Rotation should be stopped once full helix extension or retraction is verified with fluoroscopy as shown in Figure 8. Over-retraction of the helix, during initial implant or subsequent repositioning, may result in the inability to extend the helix. If the helix is unable to extend, replace with a new lead.
Do not exceed the recommended maximum number of rotations to extend or retract the helix electrode. Exceeding the maximum number may result in fracture or distortion of the inner conductor or helix electrode. Refer to Chapter 10 for the recommended maximum number of rotations.
Prolonged implant procedures or multiple repositionings may allow blood or body fluids to build up on the helix electrode mechanism. This may result in an increased number of rotations required to extend or retract the helix electrode.
4. Remove the fixation tool from the IS-1 connector pin, and release the proximal end of the lead body. Allow several seconds for relief of the residual torque in the lead.
5. To assure helix electrode fixation, leave the stylet in place, hold the lead by the connector, and carefully rotate the lead body in 2 clockwise rotations.
6. Partially withdraw the stylet.
7. Obtain electrical measurements to verify satisfactory placement and electrode fixation. Refer to Section 9.6, “Taking electrical measurements and defibrillation efficacy measurements”, page 11.
8. Verify that the lead is affixed. Gently pull back on the lead, and check for resistance to verify fixation. A properly affixed helix electrode will remain in position. If the helix electrode is not properly affixed, the lead tip may become loose in the right ventricle.
9. If repositioning is required, reattach the fixation tool, and rotate counterclockwise until the helix electrode is retracted. Use fluoroscopy to verify withdrawal of the helix electrode before attempting to reposition.
10. After final positioning, make sure that the stylet, stylet guide, and fixation tool have been completely removed. When removing the stylet guide, grip the lead firmly just below the connector pin to help prevent lead dislodgement.
11. Obtain final electrical measurements. Refer to Section 9.6.

9.6 Taking electrical measurements and defibrillation efficacy measurements

Caution: Prior to taking electrical or defibrillation efficacy
measurements, move objects made from conductive materials, such as guide wires, away from all electrodes. Metal objects, such as guide wires, can short a lead and an active implantable device, causing electrical current to bypass the heart and possibly damage the implantable device and lead.
Use the following steps to take electrical measurements:
1. Attach a surgical cable to the lead connector pin. A notch in the stylet guide allows connection of a surgical cable for obtaining electrical measurements.
2. Use a testing device, such as a pacing system analyzer, for obtaining electrical measurements. For information about the use of the testing device, consult the product documentation for that device.
In order to demonstrate reliable defibrillation efficacy, obtain final defibrillation measurements for the lead system.

Table 2. Recommended measurements at implant (when using a pacing system analyzer)

Measurements required Acutea lead system
Capture threshold (at 0.5 ms pulse width)
Pacing impedance 200–1000 Ω 200–1000 Ω
≤1.0 V ≤3.0 V
Chronicb lead sys-
tem
11
Table 2. Recommended measurements at implant (when using a pacing system analyzer) (continued)
Measurements required Acutea lead system
Filtered R-wave amplitude (during sinus rhythm)
Slew rate ≥0.75 V/s ≥0.45 V/s
a
≤30 days after implant.
b
>30 days after implant.
≥5 mV ≥3 mV
Chronicb lead sys-
tem
If pacing thresholds and R-wave amplitudes do not stabilize to acceptable levels, it may be necessary to reposition the lead and repeat the testing procedure.
Values may vary depending upon implantable device settings, cardiac tissue condition, and drug interactions.
Warning: If the implanted lead system fails to terminate a VF episode, rescue the patient promptly with an external defibrillator . At least 5 min should elapse between VF inductions.
For more information about obtaining electrical measurements, consult the product documentation supplied with the testing device.

9.7 Anchoring the lead

Caution: Use care when anchoring the lead.
Use only nonabsorbable sutures to anchor the lead.
Do not attempt to remove or cut the anchoring sleeve from the lead body.
During lead anchoring, take care to avoid dislodging the lead tip.
Do not secure sutures so tightly that they damage the vein, lead, or anchoring sleeve (Figure 9).
Do not tie a suture directly to the lead body (Figure 9).
Figure 9.
Figure 10.
3. Use at least one additional suture in one of the grooves to secure the anchoring sleeve and lead body to the fascia.
4. A slit anchoring sleeve may be used in the device pocket to secure excess lead length. First, secure the anchoring sleeve to the lead body. Then, orient the slit toward the fascia and secure the anchoring sleeve to the fascia with sutures.

9.8 Connecting the lead

Use the following steps to connect the lead to an implantable device:
1. Make sure that the stylet and all accessories have been completely removed. When removing the accessories, grip the lead firmly just below the connector pin to prevent dislodgement.
2. Insert the lead connectors into the connector block. Consult the product documentation packaged with the implantable device for instructions on proper lead connections.

9.9 Placing the device and lead into the pocket

Caution: Use care when placing the device and leads into
the pocket.
Ensure that the leads do not leave the device at an acute angle.
Do not grip the lead or device with surgical instruments.
Do not coil the lead. Coiling the lead can twist the lead body and may result in lead dislodgement (Figure 11).
Figure 11.
Use the following steps to anchor the lead using all 3 grooves:
Note: The anchoring sleeve contains a radiopaque substance, which allows visualization of the anchoring sleeve on a standard x-ray and may aid in follow-up examinations.
1. Position the distal anchoring sleeve against or near the vein.
2. Secure the anchoring sleeve to the lead body by tying a suture firmly in each of the 3 grooves (Figure 10).
12
Use the following steps to place the device and leads into the pocket:
1. To prevent undesirable twisting of the lead body, rotate the device to loosely wrap the excess lead length (Figure 12).
Figure 12.
2. Insert the device and leads into the pocket.
3. Before closing the pocket, verify sensing, pacing, cardioversion, and defibrillation efficacy.

9.10 Post-implant evaluation

After implant, monitor the patient’s electrocardiogram until the patient is discharged. If a lead dislodges, it typically occurs shortly after implant.
Recommendations for verifying proper lead positioning include x-rays and pacing and sensing threshold measurements.
In the event of a patient death, explant all implanted leads and devices and return them to Medtronic with a completed Product Information Report form. Call the appropriate phone number on the back cover if there are any questions on product handling procedures.

10 Specifications (nominal)

10.1 Detailed device description

Table 3. Specifications (nominal)

Parameter Model 6947
Type Quadripolar
Position Right ventricle
Fixation Extendable/retractable helix
Length 58 cm, 65 cm
Connector Unipolar (2) DF-1
Bipolar IS-1
Materials Conductors: MP35N coil
MP35N composite cables
Insulation: Silicone, PTFE, ETFE
Overlay: Polyurethane
Electrodes (pace, sense): Platinized platinum alloy
RV/SVC coils: Platinum-clad tantalum
DF-1 pins: Stainless steel
IS-1 pin and ring: Stainless steel
Steroid Type: Dexamethasone acetate
Dexamethasone sodium phos­phate
Amount: 685 µg dexamethasone acetate
59 µg dexamethasone sodium phosphate
Steroid binder: Silicone
Table 3. Specifications (nominal) (continued)
Parameter Model 6947
Conductor resistances
Helix length (extended) 1.8 mm
Diameters Lead body: 2.8 mm
Lead introducer (recommended size)
Pacing (unipolar): 29 Ω (65 cm)
Pacing (bipolar): 32.3 Ω (65 cm)
Defibrillation: <1.2 Ω (65 cm)
Tip: 2.8 mm
Helix: 1.4 mm
without guide wire: 3.0 mm (9.0 French)
with guide wire: 3.7 mm (11.0 French)

Table 4. Maximum number of rotations to extend or retract the helix electrode

Lead length Number of rotations
58 cm 18
65 cm 20

Table 5. Respective electrode distances

Helix electrode to ring electrode 8 mm
Helix electrode to RV coil electrode 12 mm
Helix electrode to SVC coil electrode 180 mm
13

Figure 13. Model 6947 distal lead components

1
234
5

Figure 14. Model 6947 proximal lead components

1 Helix electrode; surface area: 5.7 mm 2 Ring electrode; surface area: 25.2 mm 3 RV coil electrode; length: 57 mm; surface area: 614 mm2; electrical
shadow area: 506 mm
4 SVC coil electrode; length: 80 mm; surface area: 860 mm2; electrical
shadow area: 709 mm
5 Anchoring sleeve
14
2
2
2
2
1 DF-1 Connector (red band); Connector pin is common to RV coil
electrode
2 DF-1 Connector (blue band); Connector pin is common to SVC coil
electrode
3 IS-1 BI contact; Connector pin is common to tip electrode; connector
ring is common to ring electrode

11 Medtronic warranty

For complete warranty information, see the accompanying warranty document.

12 Service

Medtronic employs highly trained representatives and engineers located throughout the world to serve you and, upon request, to provide training to qualified hospital personnel in the use of Medtronic products. Medtronic also maintains a professional staff to provide technical consultation to product users. For more information, contact your local Medtronic representative, or call or write Medtronic at the appropriate telephone number or address listed on the back cover.
Medtronic, Inc.
*M965526A001*
710 Medtronic Parkway Minneapolis, MN 55432 USA www.medtronic.com +1 763 514 4000
Medtronic USA, Inc.
Toll-free in the USA (24-hour technical consultation for physicians and medical professionals) Bradycardia: +1 800 505 4636 Tachycardia: +1 800 723 4636
Europe/Middle East/Africa
Medtronic International Trading Sàrl Route du Molliau 31 Case Postale 84 CH-1131 Tolochenaz Switzerland +41 21 802 7000
Technical manuals
www.medtronic.com/manuals
© 2016 Medtronic M965526A001 B 2016-03-10
Loading...