Medtronic 6937A-35 Technical Manual

TRANSVENE® -CS/SVC 6937A
Unipolar, endocardial, CS/SVC lead
Technical manual
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
Table of contents
Device description 1
Contents of package 1 Accessory descriptions 1
Indications for use 1
Contraindications 2
Warnings and precautions 2
Adverse events 4
Observed adverse events 4 Potential adverse events 5
Clinical studies 5
Directions for use 7
Opening the sterile package 7 Inserting the lead 7 Positioning the lead in the superior vena cava 9 Positioning the lead in the coronary sinus 9 Taking defibrillation efficacy measurements 10 Anchoring the lead 11 Connecting the lead 12 Placing the device and lead into the pocket 12 Post-implant evaluation 13
Detailed device description 14
Specifications (nominal) 14 Specifications drawings (nominal) 15
Special notice 16
Service 16

Device description

The Medtronic Transvene-CS/SVC Model 6937A Unipolar, endocardial lead is designed for delivering cardioversion and defibrillation therapies in either the coronary sinus or superior vena cava. Silicone insulation with a polyurethane overlay surrounds the low resistance conductor coil and the coil electrode is a platinum alloy. The lead has a DF-1 section "Detailed device description."

Contents of package

The lead and accessories are provided sterile. Each package contains the following:
One lead with one2 radiopaque anchoring sleeve, stylet, and stylet guide
One vein lifter
Extra stylets
Product literature

Accessory descriptions

Stylets – A stylet provides additional stiffness and controlled
flexibility for maneuvering the lead into position. Each stylet knob is labeled with the stylet diameter and length. Curving the stylet prior to insertion into the lead will achieve a curvature at the lead’s distal end.
Stylet guide – A stylet guide facilitates stylet insertion into the lead. A notch in the stylet guide allows connection of a surgical cable for electrical measurements.
Anchoring sleeves – An anchoring sleeve secures the lead from moving and protects the lead insulation and conductors from damage caused by tight ligatures. An anchoring sleeve is radiopaque for visualization on standard X-ray equipment.
Vein lifter – A vein lifter facilitates lead insertion into a vessel.
1
connector. For further information, refer to the

Indications for use

The lead is intended for single long-term use in the coronary sinus or superior vena cava for delivering atrial cardioversion and defibrillation therapies.
The lead has application for patients in which implantable cardioverter defibrillators are indicated.
1
DF-1 refers to the International Connector Standard (ISO 11318) whereby pulse generators and leads so designated are assured of a basic mechanical fit.
2
Two anchoring sleeves are provided with leads of 85 cm or longer.
6937A Technical manual 1
Note: Prior to lead implant, it is strongly recommended that patients undergo a complete cardiac evaluation, which should include extensive electrophysiologic testing. Also, extensive electrophysiologic evaluation and testing of the safety and efficacy of the proposed pacing, cardioversion, or defibrillation therapies are recommended during and after the implantation of the system.

Contraindications

Arrhythmia types – The lead is contraindicated for patients with any
of the following conditions:
Transient ventricular tachyarrhythmias due to reversible causes, such as drug intoxication, electrolyte imbalance, sepsis, or hypoxia.
Transient ventricular tachyarrhythmias due to other factors, such as myocardial infarction and electrocution.

Warnings and precautions

Inspecting the sterile package – Carefully inspect the package
prior to opening:
If the seal or package is damaged, contact your local Medtronic representative.
Do not use the product after its expiration date.
The lead has been sterilized with ethylene oxide prior to shipment. If the integrity of the sterile package has been compromised prior to the expiration date, resterilize using ethylene oxide.
Ethylene oxide resterilization – If the sterile package seal is broken, resterilize the device using a validated ethylene oxide process. Avoid resterilization techniques that could damage the lead:
Refer to sterilizer instructions for operating instructions.
Use an acceptable method for determining sterilizer effectiveness, such as biological indicators.
Before resterilization, place the device in an ethylene oxide permeable package.
Do not exceed temperatures of 55°C (130°F).
Do not resterilize more than one time.
After resterilization, allow the device to aerate ethylene oxide residues.
2 6937A Technical manual
Handling the lead – Leads should be handled with great care at all times:
Protect the lead from materials shedding particles such as lint and dust. Lead insulators attract these particles.
Handle the lead with sterile surgical gloves that have been rinsed in sterile water or a comparable substance.
Do not severely bend, kink, or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pins.
Do not immerse leads in mineral oil, silicone oil, or any other liquid, except blood at the time of implantation.
Inserting the lead using a lead introducer that features a hemostasis valve may require a larger introducer than the size recommended. Do not withdraw the lead through a hemostasis valve, to avoid distortion of the coil electrode.
Handling the stylets – Use care when handling stylets:
Do not use excessive force or surgical instruments when inserting a stylet.
Avoid overbending, kinking, or blood contact.
Use a new stylet when blood or other fluids accumulate on the stylet. Repeated insertions of a stylet increases the probability that blood or other fluids may accumulate on the stylet, which may cause difficulty in passing the stylet through the lead or damage to the lead.
Do not use a sharp object to impart a curve to the distal end of the stylet, to avoid damage to the stylet.
Necessary hospital equipment – Keep external defibrillation equipment nearby for immediate use during the acute lead system testing, implantation procedure, or whenever arrhythmias are possible or intentionally induced during post-implant testing.
Line-powered equipment – An implanted lead forms a direct current path to the myocardium. During lead implantation and testing, use only battery-powered equipment or line-powered equipment specifically designed for this purpose, to protect against fibrillation that may be caused by alternating currents. Line-powered equipment used in the vicinity of the patient must be properly grounded. Lead connector pins must be insulated from any leakage currents that may arise from line-powered equipment.
Second anchoring sleeve – For leads 85 cm or longer, which feature two anchoring sleeves, use both anchoring sleeves as described in the “Anchoring the lead” section, to assure adequate fixation.
Chronic repositioning or removal – Chronic repositioning or removal of leads may be difficult because of fibrotic tissue development. Return all removed or unused leads to Medtronic. If a lead must be removed or repositioned, proceed with extreme caution:
6937A Technical manual 3
Removal of the lead may result in avulsion of the endocardium, valve, or vein.
If a lead is abandoned, it should be capped to avoid transmitting electrical signals.
A lead that has been cut off should have the remaining lead end sealed, and the lead should be sutured to adjacent tissue to avoid migration into the heart.
Connector compatibility – Although the lead conforms to the International Connector Standard DF-1, do not attempt to use the Model 6937A lead with any device other than a commercially available implantable defibrillator system with which it has been tested and demonstrated to be safe and effective. The potential adverse consequences of using such a combination may include, but are not limited to, undersensing cardiac activity and failure to deliver necessary therapy.

Adverse events

Observed adverse events

The Transvene-CS/SVC Model 6937A Unipolar, endocardial, CS/ SVC lead was utilized in the SVC and coronary sinus during the Model 7250 Jewel patients who received a Model 7250 AF Implantable Cardioverter Defibrillator received a Model 6937A lead in either the coronary sinus or the SVC. Individual patient implant time averaged 12.8 months, with a range of 0.1 to 24.7 months.
Eight patient deaths occurred in this group of patients. All deaths were reviewed and judged to be non-system related by an independent advisory committee. The deaths were attributed to: congestive heart failure (3), ischemic cardiomyopathy (1), cardiomyopathy (1), hypoxic encephalopathy (1), lung cancer (1) and cardiogenic shock/respiratory failure (1).
Lead-related complications and lead-related observations are summarized in Table 1:
Tab le 1 . Lead-related complications and observationsa all patients (N = 114)
Complications
Lead dislodgment 3 3
To t al 3 3
Observations
Subclavian vein thrombosis 1 1
To t a l 1 1
a
database closure: 5/31/2000
AF clinical study. One hundred fourteen of 676
# of events # of patients
4 6937A Technical manual

Potential adverse events

The potential adverse events related to the use of endocardial leads include, but are not limited to, the following patient-related conditions:
Cardiac perforation
Cardiac tamponade
Constrictive pericarditis
Coronary sinus perforation
Embolism
Endocarditis
Fibrillation or other arrhythmias
Heart wall rupture
Hemothorax
Infection
Pneumothorax
Thrombosis
Tissue necrosis
Other potential adverse events related to the lead include, but are not limited to, the following:
Insulation failure
Lead conductor or electrode fracture
Lead dislodgement
Poor connection to the device, which may lead to oversensing, undersensing, or a loss of therapy.

Clinical studies

Model 6937A data was derived from two clinical investigations of the Model 7250 Jewel AF device. One of the clinical investigations focused on patients with VT/AT indications and the other investigation focused on patients with AF indications. The clinical investigations were conducted in the U.S., Europe, and Canada. These studies were conducted primarily to verify that the atrial prevention and treatment therapies of the Model 7250 Jewel AF were safe and effective as evaluated in their respective patient populations.
In the VT/AT clinical investigation, the Model 6937A lead was utilized in 35 of 530 patients receiving a Model 7250 Jewel AF device. VT/AT patients had to meet the following eligibility criteria: 1) at least one episode of cardiac arrest due to ventricular tachyarrhythmia not caused by an acute myocardial infarction; or recurrent, sustained VT (spontaneous or inducible), and 2) at least two documented episodes of some form of atrial tachyarrhythmia occurring in the last year, with at least one episode documented on ECG.
In the AF-only clinical investigation, the Model 6937A lead was utilized in 79 of 146 patients receiving a Model 7250 Jewel AF device. AF-only patients had to meet the following eligibility criteria: 1) at least two episodes of symptomatic, drug refractory atrial fibrillation or atrial flutter occurring in the last three months, with at least one episode documented on ECG.
6937A Technical manual 5
Patients studied – Patient demographics for both studies are listed in Table 2.
Table 2. Patient demographics
Patient characteristic VT/AT (N=35) AF-only (N=79) P-Value
Gender (N,%)
Male 30 (85.7%) 58 (73.4%) 0.230
Female 5 (14.3%) 21 (26.6%)
Age (years)
Mean 67.4 64.3 0.157
Range 31.2 - 84.5 38.1 - 82.0
Standard deviation 11.4 10.7
Spontaneous ventricular arrhythmia history (non-exclusive, N,%)
Sustained VT 21 (60.0%) 1 (1.3%) <0.001
Nonsustained VT 11 (31.4%) 5 (6.3%) 0.001
Ventricular flutter 1 (2.9%) 0 0.674
Ventricular fibrillation 8 (22.9%) 2 (2.5%) 0.001
Spontaneous atrial arrhythmia history (non-exclusive, N,%)
Atrial fibrillation 31 (88.6%) 77 (97.5%) 0.132
Atrial flutter 5 (14.3%) 19 (24.1%) 0.352
Methods - For both studies, clinical data was collected at pre-implant assessment/enrollment, implant, scheduled and unscheduled follow­up visits; and for system modifications and patient deaths. Scheduled follow-up visits were required at one, three, and six months post­implant, with continued follow-ups every six months afterwards.
Atrial defibrillation thresholds (A-DFTs) were determined by using a two-tiered step-up protocol.
Results – The A-DFT results for both studies are provided in Table 3.
.
A-DFT at implant VT/AT
Mean A-DFT (J) 3.8 6.6 6.2
Range 2 - 6 2 -18 2 - 18
Standard deviation 1.8 4.8 4.6
Tab le 3 . A-DFTs at implant
(N=10)
AF-only
(N=53)
Combined
(N=63)
Conclusion – The clinical experience with the Model 6937A CS/SVC
lead demonstrates that the lead is safe and effective for human use.
6 6937A Technical manual

Directions for use

Proper surgical procedures and sterile techniques are the responsibility of the medical professional. The implant procedures described in this manual are furnished for information only. Each physician must apply the information in these instructions according to professional medical training and experience.
The implantation procedure generally includes the following steps:

Opening the sterile package

Inserting the lead

Positioning the lead in the superior vena cava or coronary sinus
Taking defibrillation efficacy measurements
Anchoring the lead
Connecting the lead
Placing the device and lead into the pocket
Opening the sterile package
Open the sterile package and inspect the lead:
1. Within the sterile field, open the sterile package and remove the lead and accessories.
2. Inspect the lead. Leads shorter than 85 cm should have one anchoring sleeve. Leads 85 cm or longer should have two anchoring sleeves.
Inserting the lead
Caution: Use care when handling the lead during insertion:
Do not severely bend, kink, or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pins.
Insert the lead using the techniques described below:
1. Select a site for lead insertion. The lead may be inserted by venotomy through several different venous routes, including the right or left cephalic vein or the external or internal jugular vein. The lead may also be inserted through the subclavian vein by using a percutaneous lead introducer kit. Use the cephalic vein whenever possible to avoid lead damage in the first rib/ clavicular (thoracic inlet) space.
Certain anatomical abnormalities, such as thoracic outlet syndrome, may also precipitate pinching and subsequent fracture of the lead.
6937A Technical manual 7
2. If using a cephalic approach, insert the tapered end of a vein lifter into the incised vein and gently push the lead tip underneath and into the vein (Figure 1).
Figure 1.
Caution: If the subclavian approach is required, avoid techniques that may damage the lead:
Insertion via a subclavian approach should be done as far lateral as possible to avoid clamping the lead body between the clavicle and the first rib (Figure 2). Clamping the lead may cause lead damage which may result in the loss of therapy.
Figure 2.
Do not force the lead if significant resistance is encountered during lead passage.
Do not use techniques such as adjusting the patient’s posture (e.g., raising the arm or putting a towel behind the back) to facilitate lead passage. If resistance is encountered, it is recommended that an alternate venous entry site be used, such as the cephalic vein.
3. Advance the lead into the right atrium or the superior vena cava using a straight stylet to facilitate movement through the veins.
8 6937A Technical manual

Positioning the lead in the superior vena cava

Caution: Use care when handling the lead during positioning.
Do not severely bend, kink, or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pins.
Position the lead in the superior vena cava using the technique described below.
1. With a straight stylet inserted in the lead, advance the lead into the superior vena cava.
Note: If the distal end of the lead tip enters the right atrium, the lead has been advanced too far. Reposition the lead.
Use fluoroscopy to ensure that the distal end of the coil electrode is properly positioned in the superior vena cava. Proper positioning of the coil electrode is essential for defibrillation efficacy.

Positioning the lead in the coronary sinus

Caution: Use care when handling the lead during positioning.
Do not severely bend, kink, or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pins.
Position the lead in the coronary sinus using the technique described below:
1. Bend the stylet into a gentle curve. Imparting a curve to the stylet can be accomplished with a smooth-surface, sterile instrument as shown (Figure 3).
Caution: Do not use a sharp object to impart a curve to the distal end of the stylet, to avoid damage to the stylet.
Figure 3.
2. Advance the lead into the right atrium and rotate the curved stylet to manipulate the lead tip over the tricuspid valve. Move the lead tip medially and posteriorly to the coronary sinus ostium. Keep the stylet tip inserted just inside the coronary sinus ostium and advance the lead off of the stylet into the coronary sinus.
6937A Technical manual 9
Note: If the lead goes through the tricuspid valve opening, retract the lead with a gentle posterior rotation and then advance it toward the coronary ostium.
3. Position the coil electrode so that it is entirely within the coronary sinus. Push the lead to a position as distal as possible in the coronary sinus.
Use fluoroscopy (P-A and lateral position) to ensure the coil electrode is properly positioned in the coronary sinus. Proper positioning of the coil electrode is essential for defibrillation efficacy.

Taking defibrillation efficacy measurements

Caution: Prior to taking defibrillation efficacy measurements, move
objects made of conductive materials, such as guidewires, away from all electrodes. Metal objects, such as guidewires, can short a lead and active implantable device, causing electrical current to bypass the heart and possibly damage the implantable device and lead.
Ventricular defibrillation efficacy measurements –
Establish ventricular defibrillation efficacy using the technique described below:
1. Make sure the lead is in the desired position.
2. Remove the stylet from the lead.
3. Connect the lead to the device.
4. Put the device in the subcutaneous pocket.
5. Establish ventricular defibrillation efficacy with at least a 10 J safety margin.
In order to keep patient morbidity and mortality to a minimum, patients should be rescued promptly with an external defibrillator if the implanted lead system fails to terminate a VF episode. At least five minutes should elapse between VF inductions.
6. After establishing defibrillation efficacy, anchor the lead.
Atrial defibrillation efficacy measurements – After establishing ventricular defibrillation efficacy, atrial defibrillation efficacy measurements may be obtained using the technique described below:
1. Make sure the leads are in the desired position.
2. Remove the stylet from the lead.
3. Connect the lead to the device if this has not already been done while taking ventricular defibrillation efficacy measurements.
4. Put the device in the subcutaneous pocket if this has not already been done while taking ventricular defibrillation efficacy measurements.
5. Establish atrial defibrillation efficacy.
6. If the leads have not already been anchored, anchor the leads.
10 6937A Technical manual

Anchoring the lead

Caution: Use care when anchoring the lead:
Use only nonabsorbable sutures to anchor the lead.
Do not attempt to remove or cut the anchoring sleeve. This may cause damage to the lead insulation.
During anchoring, take care to avoid dislodging the lead.
Do not secure ligatures so tightly that they damage the vein, lead, or anchoring sleeve.
Do not tie a ligature directly to the lead body (Figure 4).
Figure 4.
Anchor the lead using all three grooves:
1. Position the anchoring sleeve against or near the vein.
2. Secure the anchoring sleeve to the lead body by tying a suture firmly in each of the three grooves (Figure 5). Suturing in the triple grooves is the only recommended location for the sutures.
Figure 5.
3. After securing the anchoring sleeve to the lead body, use at least one additional suture in one of the grooves to secure the anchoring sleeve and lead body to the fascia.
4. A second anchoring sleeve is provided with leads 85 cm or longer. For abdominal implants, redundant lead body (e.g., a curve for strain relief) should be placed just proximal to the first anchoring sleeve, then the second anchoring sleeve may be lightly sutured to the lead body and fascia to hold the curve in place. This procedure helps isolate the vein entry site from tension on the proximal end of the lead body.
6937A Technical manual 11

Connecting the lead

Note: If tunneling the lead, record each lead’s serial number along
with the function of the lead before tunneling.
Use the following techniques to connect the lead to an implantable device:
1. If tunneling the lead connectors to the implantable device is required, use a chest tube to prevent damage to the lead. Tunnel the lead subcutaneously to the device pocket.
Caution: Do not use surgical instruments to grasp the lead or connector pins when tunneling.
2. Insert the lead connectors into the connector block. Consult the manual packaged with the implantable device for instructions on proper lead connections.
Caution: Always remove the stylet before connecting the lead to the implantable device. Failure to remove the stylet may result in lead failure.
3. Before closing the pocket, verify cardioversion, and defibrillation efficacy.

Placing the device and lead into the pocket

Caution: Use care when placing the device and lead into the pocket:
Ensure that the lead does not leave the device at an acute angle. A sharp angle or excessive pressure places undue stress on the lead conductor and insulation.
Do not grip the lead or device with surgical instruments.
Do not coil the lead. Coiling the lead can twist the lead body and may result in lead dislodgment (Figure 6).
Figure 6.
12 6937A Technical manual
Use the following techniques to place the implantable device and lead into the pocket:
1. To prevent undesirable twisting of the lead body, rotate the device to loosely wrap the excess lead length (Figure 7).
Figure 7.
Place the device and lead into the pocket.

Post-implant evaluation

After implantation, monitor the patient’s electrocardiogram continuously. If a lead dislodges, it usually occurs during the immediate postoperative period.
The patient should have X-rays taken at pre-hospital discharge, three months after implant, and every six months thereafter to verify proper lead position and to check for conductor fractures.
In the event of a patient death, all implanted leads and devices should be explanted and returned to Medtronic with a completed Product Information Report. Any questions on product handling procedures can be addressed by calling the appropriate number on the back cover.
6937A Technical manual 13

Detailed device description

Specifications (nominal)

Parameter Model 6937A
Type Unipolar
Position CS or SVC
Fixation None
Length 20-110 cm
Connector DF-1
Materials Conductor:
Diameters Lead body:
Conductor resistances Defibrillation: 1.5 (max) at 58 cm
Lead introducer (recommended size)
Insulator:
Outer layer:
Coil electrode:
Coil electrode:
without guide wire:
with guide wire:
Multifilar MP35N composite Silicone Polyurethane tubing Platinum alloy
2.5 mm
2.3 mm
9 French
10.5 French
14 6937A Technical manual

Specifications drawings (nominal)

Coil electrode
Surface area: 125 mm Electrical shadow area: 355 mm Length: 50 mm
Anchoring sleeve Note: Leads 85 cm or
longer feature a second anchoring sleeve
2
2
DF-1 connector Note: Connector pin is
common to coil electrode
6937A Technical manual 15

Special notice

Medtronic implantable leads are implanted in the extremely hostile environment of the human body. Leads are necessarily very small in diameter and must still be very flexible, which unavoidably reduces their potential performance or longevity. Leads may fail to function for a variety of causes, including, but not limited to: medical complications, body rejection phenomena, allergic reaction, fibrotic tissue, or failure of leads by breakage or by breach of their insulation covering. In addition, despite the exercise of all due care in design, component selection, manufacture, and testing prior to sale, leads may be easily damaged before, during, or after insertion by improper handling or other intervening acts. Consequently, no representation or warranty is made that failure or cessation of function of leads will not occur or that the body will not react adversely to the implantation of leads or that medical complications (including perforation of the heart) will not follow the implantation of leads or that the lead will, in all cases, restore adequate cardiac function.
For complete warranty information, see the accompanying card enclosed in the package.

Service

Medtronic employs highly trained representatives and engineers located throughout the world to serve you and, upon request, to provide training to qualified hospital personnel in the use of Medtronic products. Medtronic also maintains a professional staff to provide technical consultation to product users. For medical consultation, Medtronic can often refer product users to outside medical consultants with appropriate expertise. For more information, contact your local Medtronic representative, or call or write Medtronic at the appropriate address or telephone number listed on the back cover.
16 6937A Technical manual
Europe
Europe/Africa/Middle East Headquarters
Medtronic Europe S.A. Route du Molliau 1131 Tolochenaz Switzerland Internet: www.medtronic.co.uk Tel. 41-21-802-7000 Fax 41-21-802-7900
Medtronic E.C. Authorized Representative
Medtronic B.V. Wenckebachstraat 10 6466 NC Kerkrade The Netherlands Tel. 31-45-566-8000 Fax 31-45-566-8668
Asia-Pacific
Japan
Medtronic Japan Solid Square West Tower 6F, 580 Horikawa-cho, Saiwai-ku, Kawasaki, Kanagawa 210-0913 Japan Tel. 81-44-540-6112 Fax 81-44-540-6200
Australi a
Medtronic Australasia Pty. Ltd. Unit 4/446 Victoria Road Gladesville NSW 2111 Australia Tel. 61-2-9879-5999 Fax 61-2-9879-5100
Asia
Medtronic International Ltd. Suite 1602 1 6/F, Manulife Plaz a The Lee Gardens, 33 Hysan Avenue Causeway Bay Hong Kong Tel. 852-2891-4068 Fax 852-2591-0313
Americas
Latin America Headquarters
Medtronic, Inc. 710 Medtronic Parkway NE Minneapolis, MN 55432-5604 USA Tel. 763-514-4000 Fax 763-514-4879
Canada
Medtronic of Canada Ltd. 6733 Kitimat Road Mississauga, Ontario L5N 1W3 Tel. 905-826-6020 Fax 905-826-6620 Toll-free in Canada: 1-800-268-5346
United States
World Headquarters Medtronic, Inc. 710 Medtronic Parkway NE Minneapolis, MN 55432-5604 USA Internet: www.medtronic.com Tel. 763-514-4000 Fax 763-514-4879
Medtronic USA, Inc.
Toll-free in the USA: 1-800-723-4636 (24-hour consultation for physicians and medical professionals)
*197885001*
© Medtronic, Inc. 2001 All Rights Reserved
UCX197885001 197885001 April 2001
Loading...