Opening the sterile package 7
Inserting the lead 7
Positioning the lead in the superior vena cava 9
Positioning the lead in the coronary sinus 9
Taking defibrillation efficacy measurements 10
Anchoring the lead 11
Connecting the lead 12
Placing the device and lead into the pocket 12
Post-implant evaluation 13
The Medtronic Transvene-CS/SVC Model 6937A Unipolar,
endocardial lead is designed for delivering cardioversion and
defibrillation therapies in either the coronary sinus or superior vena
cava. Silicone insulation with a polyurethane overlay surrounds the
low resistance conductor coil and the coil electrode is a platinum alloy.
The lead has a DF-1
section "Detailed device description."
Contents of package
The lead and accessories are provided sterile. Each package
contains the following:
■
One lead with one2 radiopaque anchoring sleeve, stylet, and
stylet guide
■
One vein lifter
■
Extra stylets
■
Product literature
Accessory descriptions
Stylets – A stylet provides additional stiffness and controlled
flexibility for maneuvering the lead into position. Each stylet knob is
labeled with the stylet diameter and length. Curving the stylet prior to
insertion into the lead will achieve a curvature at the lead’s distal end.
Stylet guide – A stylet guide facilitates stylet insertion into the lead.
A notch in the stylet guide allows connection of a surgical cable for
electrical measurements.
Anchoring sleeves – An anchoring sleeve secures the lead from
moving and protects the lead insulation and conductors from damage
caused by tight ligatures. An anchoring sleeve is radiopaque for
visualization on standard X-ray equipment.
Vein lifter – A vein lifter facilitates lead insertion into a vessel.
1
connector. For further information, refer to the
Indications for use
The lead is intended for single long-term use in the coronary sinus
or superior vena cava for delivering atrial cardioversion and
defibrillation therapies.
The lead has application for patients in which implantable
cardioverter defibrillators are indicated.
1
DF-1 refers to the International Connector Standard (ISO 11318) whereby
pulse generators and leads so designated are assured of a basic
mechanical fit.
2
Two anchoring sleeves are provided with leads of 85 cm or longer.
6937A Technical manual1
Note: Prior to lead implant, it is strongly recommended that patients
undergo a complete cardiac evaluation, which should include
extensive electrophysiologic testing. Also, extensive
electrophysiologic evaluation and testing of the safety and efficacy of
the proposed pacing, cardioversion, or defibrillation therapies are
recommended during and after the implantation of the system.
Contraindications
Arrhythmia types – The lead is contraindicated for patients with any
of the following conditions:
■
Transient ventricular tachyarrhythmias due to reversible
causes, such as drug intoxication, electrolyte imbalance,
sepsis, or hypoxia.
■
Transient ventricular tachyarrhythmias due to other factors,
such as myocardial infarction and electrocution.
Warnings and precautions
Inspecting the sterile package – Carefully inspect the package
prior to opening:
■
If the seal or package is damaged, contact your local
Medtronic representative.
■
Do not use the product after its expiration date.
■
The lead has been sterilized with ethylene oxide prior to
shipment. If the integrity of the sterile package has been
compromised prior to the expiration date, resterilize using
ethylene oxide.
Ethylene oxide resterilization – If the sterile package seal is
broken, resterilize the device using a validated ethylene oxide
process. Avoid resterilization techniques that could damage the lead:
■
Refer to sterilizer instructions for operating instructions.
■
Use an acceptable method for determining sterilizer
effectiveness, such as biological indicators.
■
Before resterilization, place the device in an ethylene oxide
permeable package.
■
Do not exceed temperatures of 55°C (130°F).
■
Do not resterilize more than one time.
■
After resterilization, allow the device to aerate ethylene
oxide residues.
26937A Technical manual
Handling the lead – Leads should be handled with great care at
all times:
■
Protect the lead from materials shedding particles such as lint
and dust. Lead insulators attract these particles.
■
Handle the lead with sterile surgical gloves that have been
rinsed in sterile water or a comparable substance.
■
Do not severely bend, kink, or stretch the lead.
■
Do not use surgical instruments to grasp the lead or
connector pins.
■
Do not immerse leads in mineral oil, silicone oil, or any other
liquid, except blood at the time of implantation.
■
Inserting the lead using a lead introducer that features a
hemostasis valve may require a larger introducer than the size
recommended. Do not withdraw the lead through a hemostasis
valve, to avoid distortion of the coil electrode.
Handling the stylets – Use care when handling stylets:
■
Do not use excessive force or surgical instruments when
inserting a stylet.
■
Avoid overbending, kinking, or blood contact.
■
Use a new stylet when blood or other fluids accumulate on the
stylet. Repeated insertions of a stylet increases the probability
that blood or other fluids may accumulate on the stylet, which
may cause difficulty in passing the stylet through the lead or
damage to the lead.
■
Do not use a sharp object to impart a curve to the distal end of
the stylet, to avoid damage to the stylet.
Necessary hospital equipment – Keep external defibrillation
equipment nearby for immediate use during the acute lead system
testing, implantation procedure, or whenever arrhythmias are
possible or intentionally induced during post-implant testing.
Line-powered equipment – An implanted lead forms a direct
current path to the myocardium. During lead implantation and testing,
use only battery-powered equipment or line-powered equipment
specifically designed for this purpose, to protect against fibrillation
that may be caused by alternating currents. Line-powered equipment
used in the vicinity of the patient must be properly grounded. Lead
connector pins must be insulated from any leakage currents that may
arise from line-powered equipment.
Second anchoring sleeve – For leads 85 cm or longer, which
feature two anchoring sleeves, use both anchoring sleeves as
described in the “Anchoring the lead” section, to assure
adequate fixation.
Chronic repositioning or removal – Chronic repositioning or
removal of leads may be difficult because of fibrotic tissue
development. Return all removed or unused leads to Medtronic. If a
lead must be removed or repositioned, proceed with extreme caution:
6937A Technical manual3
■
Removal of the lead may result in avulsion of the endocardium,
valve, or vein.
■
If a lead is abandoned, it should be capped to avoid transmitting
electrical signals.
■
A lead that has been cut off should have the remaining lead end
sealed, and the lead should be sutured to adjacent tissue to
avoid migration into the heart.
Connector compatibility – Although the lead conforms to the
International Connector Standard DF-1, do not attempt to use the
Model 6937A lead with any device other than a commercially
available implantable defibrillator system with which it has been
tested and demonstrated to be safe and effective. The potential
adverse consequences of using such a combination may include, but
are not limited to, undersensing cardiac activity and failure to deliver
necessary therapy.
Adverse events
Observed adverse events
The Transvene-CS/SVC Model 6937A Unipolar, endocardial, CS/
SVC lead was utilized in the SVC and coronary sinus during the
Model 7250 Jewel
patients who received a Model 7250 AF Implantable Cardioverter
Defibrillator received a Model 6937A lead in either the coronary sinus
or the SVC. Individual patient implant time averaged 12.8 months,
with a range of 0.1 to 24.7 months.
Eight patient deaths occurred in this group of patients. All deaths
were reviewed and judged to be non-system related by an
independent advisory committee. The deaths were attributed to:
congestive heart failure (3), ischemic cardiomyopathy (1),
cardiomyopathy (1), hypoxic encephalopathy (1), lung cancer (1) and
cardiogenic shock/respiratory failure (1).
Lead-related complications and lead-related observations are
summarized in Table 1:
Tab le 1 . Lead-related complications and observationsa all patients (N = 114)
Complications
Lead dislodgment33
To t al33
Observations
Subclavian vein thrombosis11
To t a l 11
a
database closure: 5/31/2000
AF clinical study. One hundred fourteen of676
# of events# of patients
46937A Technical manual
Loading...
+ 16 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.