Medtronic 6935M62 Technical Manual

SPRINT QUATTRO SECURE S MRI™ SURESCAN™ 6935M
MR Conditional, steroid-eluting, tripolar, screw-in, ventricular lead with RV defibrillation coil electrode
Technical Manual
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
The following list includes trademarks or registered trademarks of Medtronic in the United States and possibly in other countries. All other trademarks are the property of their respective owners.
AccuRead, Evera, Evera MRI, Medtronic, Sprint Quattro, Sprint Quattro Secure, Sprint Quattro Secure S, Sprint Quattro Secure S MRI, SureScan, Tensi-Lock
1 Description 3 2 Drug component description 4 3 Indications 4 4 Contraindications 4 5 Warnings and precautions 4 6 Adverse events and clinical trial data 6 7 Drug information 6 8 Directions for use 7
9 Specifications (nominal) 13 10 Medtronic disclaimer of warranty 15 11 Service 15 12 Explanation of symbols on package labeling 15

1 Description

The Medtronic Sprint Quattro Secure S MRI SureScan Model 6935M lead is a steroid-eluting, tripolar, screw-in, ventricular lead with a right ventricular (RV) defibrillation coil electrode. The lead is designed for pacing, sensing, cardioversion, and defibrillation therapies. The following lead lengths are MR conditional: 55 cm and 62 cm. Other lead lengths are not MR conditional.
The lead features an extendable and retractable helix electrode, silicone insulation, and parallel conductors. The 3 electrodes of the lead are the helix, ring, and RV coil. The lead also features Tensi-Lock1 and silicone-backfilled defibrillation coils.
The Medtronic DF4-LLHO2 four-pole HV inline connector on the lead facilitates device connection during implant. The DF4 connector pin has a color band indicator that may be used to visually confirm proper connection to the device.
The RV coil delivers cardioversion and defibrillation therapies. Pacing and sensing occur between the helix and either the ring or RV coil electrode. An AccuRead analyzer cable interface tool (ACI tool) is attached to the lead to facilitate accurate electrical measurements during implant.
The helix electrode can be actively fixed into the endocardium. The helix electrode can be extended or retracted by rotating the DF4 connector pin with the purple fixation tool included in the package.
The distal tip contains a nominal dosage of 685 µg of dexamethasone acetate and 59 µg of dexamethasone sodium phosphate. Upon exposure to body fluids, the steroids elute from
the lead tip. The steroids are known to suppress the inflammatory response that is believed to cause threshold rises typically associated with implanted pacing electrodes.

1.1 Medtronic SureScan defibrillation system

A complete SureScan defibrillation system is required for use in the MR environment. A complete SureScan defibrillation system includes a Medtronic SureScan device with the appropriate number of Medtronic SureScan leads.
The Model 6935M lead is part of the Medtronic SureScan defibrillation system. Labeling for SureScan defibrillation system components displays the MR Conditional symbol. To verify that components are part of a SureScan system, visit http://www.mrisurescan.com. Any other combination may result in a hazard to the patient during an MRI scan.
The MRI SureScan feature permits a mode of operation that allows a patient with a SureScan system to be safely scanned by an MRI machine while the device continues to provide appropriate pacing. When programmed to On, MRI SureScan operation disables arrhythmia detection and all user-defined diagnostics. Before performing an MRI scan, refer to the MRI Technical Manual.

1.2 Package contents

Leads and accessories are supplied sterile. Each package contains the following items:
1 lead with a radiopaque anchoring sleeve, stylet, and ACI tool
2 purple fixation tools
1 purple stylet guide
1 slit anchoring sleeve
1 vein lifter
extra stylets
product literature

1.3 Accessory descriptions

Dispose of all single-use accessories according to local environmental requirements.
AccuRead analyzer cable interface (ACI) tool – The ACI tool facilitates accurate electrical measurements during implant and prevents possible connector damage.
Anchoring sleeve – An anchoring sleeve secures the lead to prevent it from moving and protects the lead insulation and conductors from damage caused by tight sutures.
Purple fixation tool – The purple fixation tool facilitates connector pin rotation.
Purple stylet guide – A stylet guide facilitates stylet insertion into the lead.
Slit anchoring sleeve – A slit anchoring sleeve secures excess lead length in the device pocket.
1
Tensi-Lock is an exclusive Medtronic design feature that utilizes lead body cables to act like a built-in locking stylet and add tensile strength to the lead.
2
DF4-LLHO refers to the international standard ISO 27186:2010, where the lead connector contacts are defined as low voltage (L), high voltage (H), or open (O).
3
Stylet – A stylet provides additional stiffness and controlled flexibility for maneuvering the lead into position. Each stylet knob is labeled with the stylet diameter and corresponding lead length.
Vein lifter – A vein lifter facilitates lead insertion into a vein.

2 Drug component description

The active ingredients in the Model 6935M lead are dexamethasone acetate [21-(acetyloxy)-9-fluoro-11β, 17-trihydroxy-16α-methylpregna-1,4-diene-3,20-dione] and dexamethasone sodium phosphate [9-fluoro-11β, 17-dihydroxy-16α-methyl-21-(phosphonooxy) pregna-1,4-diene-3,20-diene-3,20-dione disodium salt]. The structural formula for these steroids is shown in the following figures.
Dexamethasone acetate is a white to practically white, odorless powder. It is a practically insoluble ester of dexamethasone, a synthetic adrenocortical steroid.
Figure 1. Structural formula for dexamethasone acetate (DXAC) C24H31FO
Dexamethasone sodium phosphate is an inorganic ester of dexamethasone, a synthetic adrenocortical steroid. Dexamethasone sodium phosphate is a white or slightly yellow crystalline powder. It is freely soluble in water and is very hygroscopic.
Figure 2. Structural formula for dexamethasone sodium phosphate (DSP) C22H28FNa2O8P
The maximum dosage of dexamethasone acetate and dexamethasone sodium phosphate is less than 1.0 mg per lead.
6

3 Indications

The lead is intended for single, long-term use in the right ventricle.
This lead has application for patients for whom implantable cardioverter defibrillators (ICDs) are indicated.

4 Contraindications

Atrial use – The lead is contraindicated for the sole use of
detection and treatment of atrial arrhythmias.
Ventricular use – The lead is contraindicated for ventricular use in patients with tricuspid valvular disease or a tricuspid mechanical heart valve.
Transient ventricular tachyarrhythmias – The lead is contraindicated for patients with transient ventricular tachyarrhythmias due to reversible causes (drug intoxication, electrolyte imbalance, sepsis, hypoxia) or other factors (myocardial infarction, electric shock).
Steroid use – The lead is contraindicated in patients for whom a single dose of 1.0 mg of dexamethasone acetate and dexamethasone sodium phosphate may be contraindicated.

5 Warnings and precautions

A complete SureScan defibrillation system is required for use in the MR environment. Before performing an MRI scan, refer to the MRI Technical Manual for MRI-specific warnings and precautions.
Inspecting the sterile package – Inspect the sterile package
before opening it.
If the seal of the package is damaged, contact a Medtronic representative.
Do not store this product above 40 °C (104 °F).
Do not use the product after its expiration date.
Single use – The lead and accessories are for single use only.
Sterilization – Medtronic has sterilized the package contents
with ethylene oxide before shipment. This lead is for single use only and is not intended to be resterilized.
Connector compatibility – Although the lead conforms to the International Connector Standard for DF4, do not attempt to use the lead with any device other than a commercially available implantable defibrillator system with which it has been tested and demonstrated to be safe and effective. The potential adverse consequences of using such a combination may include, but are not limited to, undersensing cardiac activity and failure to deliver necessary therapy.
Electrophysiologic testing – Prior to lead implant, it is strongly recommended that patients undergo a complete cardiac evaluation, which should include electrophysiologic testing. Also, electrophysiologic evaluation and testing of the safety and efficacy of the proposed pacing, cardioversion, or defibrillation
4
therapies are recommended during and after the implant of the system.
Steroid use – It has not been determined whether the warnings, precautions, or complications usually associated with injectable dexamethasone sodium phosphate or dexamethasone acetate apply to the use of this highly localized, controlled-release device. For a list of potential adverse effects, refer to the Physicians’ Desk Reference.
Handling the steroid tip – Avoid reducing the amount of steroid available before implanting the lead. Reducing the available amount of steroid may adversely affect low-threshold performance.
Do not allow the electrode surface to come in contact with surface contaminants.
Do not wipe or immerse the electrode in fluid, except blood, at the time of implant.
Handling the lead – Handle the lead with care at all times.
Protect the lead from materials that shed particles such as lint and dust. Lead insulators attract these particles.
Handle the lead with sterile surgical gloves that have been rinsed in sterile water or a comparable substance.
Do not severely bend, kink, or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pins.
Do not immerse the lead in mineral oil, silicone oil, or any other liquid, except blood, at the time of implant.
Inserting the lead using a lead introducer that has a hemostasis valve may require a larger introducer than the size recommended. To avoid distortion of the coil electrode, do not withdraw the lead through a hemostasis valve.
Do not implant the lead without first verifying the mechanical functioning of the helix electrode. Refer to Section 8.2, for complete instructions.
Do not rotate the helix electrode after it is fully extended or fully retracted. Do not exceed the recommended maximum number of rotations to extend or retract the helix electrode. Exceeding the maximum number may result in fracture or distortion of the inner conductor or helix electrode. The number of rotations required to fully extend or retract the helix electrode is variable; refer to Chapter 9 for the recommended maximum number of rotations.
Handling the stylet – Handle the stylet with care at all times.
Do not use a sharp object to impart a curve to the distal end of the stylet.
Do not use excessive force or surgical instruments when inserting the stylet into the lead.
Avoid overbending or kinking the stylet.
Use a new stylet when blood or other fluids accumulate on the stylet. Accumulated blood or other fluids may damage the lead or cause difficulty in passing the stylet into the lead.
Necessary hospital equipment – Keep external defibrillation equipment nearby for immediate use during acute lead system testing, the implant procedure, or whenever arrhythmias are possible or intentionally induced during post-implant testing.
Line-powered and battery-powered equipment – An implanted lead forms a direct current path to the myocardium. During lead implant and testing, use only battery-powered equipment or line-powered equipment specifically designed for this purpose to protect against fibrillation that may be caused by alternating currents. Line-powered equipment used in the vicinity of the patient must be properly grounded. Lead connector pins must be insulated from any leakage currents that may arise from line-powered equipment.
Concurrent devices – Output pulses, especially from unipolar devices, may adversely affect device sensing capabilities. If a patient requires a separate stimulation device, either permanent or temporary, allow enough space between the leads of the separate systems to avoid interference in the sensing capabilities of the devices. Previously implanted pulse generators and implantable cardioverter defibrillators should generally be explanted.
Magnetic resonance imaging (MRI) – An MRI is a type of medical imaging that uses magnetic fields to create an internal view of the body. If certain criteria are met and the warnings and precautions provided by Medtronic are followed, patients with an MR Conditional device and lead system are able to undergo an MRI scan; for details, refer to the MRI Technical Manual that Medtronic provides for an MR Conditional device.
Diathermy treatment (including therapeutic ultrasound) –
Diathermy is a treatment that involves the therapeutic heating of body tissues. Diathermy treatments include high frequency, short wave, microwave, and therapeutic ultrasound. Except for therapeutic ultrasound, do not use diathermy treatments on cardiac device patients. Diathermy treatments may result in serious injury or damage to an implanted device and lead system. Therapeutic ultrasound (including physiotherapy, high intensity therapeutic ultrasound, and high intensity focused ultrasound), is the use of ultrasound at higher energies than diagnostic ultrasound to bring heat or agitation into the body. Therapeutic ultrasound is acceptable if treatment is performed with a minimum separation distance of 15 cm (6 in) between the applicator and the implanted device and lead system, as long as the ultrasonic beam is pointing away from the device and lead system.
Chronic lead removal and the SureScan defibrillation system – When implanting a SureScan defibrillation system,
consider the risks associated with removing previously implanted leads before doing so. Abandoned leads or previously implanted non-SureScan labeled leads compromise the ability to safely scan the SureScan defibrillation system during MRI scans.
Chronic repositioning or removal of a screw-in lead –
Proceed with extreme caution if a lead must be removed or repositioned. Chronic repositioning or removal of screw-in transvenous leads may not be possible because of blood or fibrotic tissue development into the helix mechanism on the lead. In most clinical situations, it is preferable to abandon unused leads in place. Return all removed leads, unused leads, or lead sections to Medtronic for analysis.
5
Note: If a helix does not disengage from the endocardium by rotating the connector pin, rotating the lead body counterclockwise may withdraw the helix and decrease the possibility of damage to cardiovascular structures during removal.
Lead removal may result in avulsion of the endocardium, valve, or vein.
Lead junctions may separate, leaving the lead tip and bare wire in the heart or vein.
Chronic repositioning of a lead may adversely affect the low-threshold performance of a steroid lead.
An abandoned lead should be capped so that the lead does not transmit electrical signals.
Severed leads should have the remaining lead end sealed and the lead body sutured to adjacent tissue.
AccuRead tool – The AccuRead tool reduces the risk of connector damage, and reduces the risk of bridging and shorting that may occur while taking electrical measurements during implant. The potential for connector damage, bridging, and shorting is due to variations in analyzer cable terminals, as well as to the connector ring width and the proximity of the rings on the DF4 connector.

6 Adverse events and clinical trial data

Information regarding clinical studies and adverse events related to this lead is available at www.medtronic.com/manuals. The following clinical studies are related to this lead:
Model 6932 RV Lead clinical study
Model 6947 RV Lead clinical study
Evera MRI System study – This clinical study was executed to confirm safety and efficacy of the Evera MRI system in the clinical MRI environment when subjects receive MRI scans up to 2 W/kg SAR without positioning restrictions (MRI scans may occur anywhere on the body), providing support for the Sprint Quattro Secure S MRI SureScan Model 6935M device.
If you do not have web access, a printed copy of the related clinical study summary can be obtained from your Medtronic representative, or you can call the toll-free number located on the back cover.
Potential adverse events – The potential adverse events associated with the use of transvenous leads and pacing systems include, but are not limited to, the following events:
acceleration of tachyarrhythmias (caused by device)
air embolism
bleeding
body rejection phenomena, including local tissue reaction
cardiac dissection
cardiac perforation
cardiac tamponade
chronic nerve damage
constrictive pericarditis
death
device migration
endocarditis
erosion
excessive fibrotic tissue growth
extrusion
fibrillation or other arrhythmias
fluid accumulation
formation of hematomas/seromas or cysts
heart block
heart wall or vein wall rupture
hemothorax
infection
keloid formation
lead abrasion and discontinuity
lead migration/dislodgement
mortality due to inability to deliver therapy
muscle and/or nerve stimulation
myocardial damage
myocardial irritability
myopotential sensing
pericardial effusion
pericardial rub
pneumothorax
poor connection of the lead to the device, which may lead to oversensing, undersensing, or a loss of therapy
threshold elevation
thrombosis
thrombotic embolism
tissue necrosis
valve damage (particularly in fragile hearts)
venous occlusion
venous perforation
Other potential adverse events related to the lead include, but are not limited to, the following conditions:
insulation failure
lead conductor or electrode fracture
Additional potential adverse events associated with the use of ICD systems include, but are not limited to, the following events:
inappropriate shocks
potential mortality due to inability to defibrillate
shunting current or insulating myocardium during defibrillation

7 Drug information

7.1 Steroid mechanism of action

Steroid suppresses the inflammatory response that is believed to cause threshold rises typically associated with implanted pacing electrodes. Dexamethasone sodium phosphate and dexamethasone acetate are synthetic steroids of the glucocorticoid family. Glucocorticoids have potent anti-inflammatory actions via direct and indirect effects on major inflammatory cells. Glucocorticosteroids bind to a cytoplasmic glucocorticoid receptor as well as a membrane-bound receptor. Binding to the cytoplasmic receptor leads to receptor activation and translocation to the nucleus. The receptor interacts with
6
specific DNA sequences within the regulatory regions of affected genes. Thus, glucocorticoids inhibit the production of multiple cell factors that are critical in generating the inflammatory response.

7.2 Pharmacodynamics of the Model 6935M lead

Pharmacokinetics – The pharmacokinetics (local drug levels
and systemic levels) of dexamethasone acetate (DXAC) and dexamethasone sodium phosphate (DSP) and their metabolites following lead implantation were not evaluated in human clinical trials.
The in-vivo elution profile of a tined pacemaker lead with a DSP monolithic controlled release device, based upon an assay of explanted leads, is shown in Mond and Stokes3.
Metabolism – The conversion of DSP to dexamethasone occurs within minutes; the conversion of DXAC to dexamethasone occurs within hours. The dexamethasone alcohol (dexamethasone) is the active glucocorticoid used in Medtronic leads. Steroid is applied to the tip and eluted through the electrode tip to the tissue interface where it will be used. Dexamethasone acetate and Dexamethasone sodium phosphate are hydrolyzed into dexamethasone, which is readily absorbed by the surrounding tissue and body fluids. Glucocorticoids, when given systemically, are eliminated primarily by renal excretion of inactive metabolites.
Mutagenesis, carcinogenicity and reproductive toxicology –
The mutagenesis, carcinogenicity, and reproductive toxicity of the Model 6935M lead have not been evaluated. However, the mutagenesis, carcinogenicity, and reproductive toxicity of dexamethasone acetate and dexamethasone sodium phosphate have been evaluated previously.
Carcinogenesis, mutagenesis, impairment of fertility – No adequate studies have been conducted in animals to determine whether corticosteroid have a potential for carcinogenesis (tumor initiation or promotion). Dexamethasone was genotoxic in assays for clastogenicity (including sister chromatid exchange in human lymphocytes) but not in an assay for mutagenicity in salmonella (Ames test).
Adrenocorticoids have been reported to increase or decrease the number and mobility of spermatozoa in some patients.
Pregnancy – Pregnancy category C. Dexamethasone acetate and Dexamethasone sodium phosphate have been shown to be teratogenic in many species when given in doses equivalent to the human dose. There are no adequate and well-controlled studies in pregnant women. Dexamethasone acetate and Dexamethasone sodium phosphate should be used during pregnancy only if the potential benefit justifies the potential risk to the fetus. Studies in mice, rats, and rabbits have shown that adrenocorticoids increase the incidence of cleft palate, placental insufficiency, and spontaneous abortions, and can decrease the intrauterine growth rate.
Nursing mothers – Systemically administered corticosteroids appear in human milk and could suppress growth, interfere with
endogenous corticosteroid production, or cause other untoward effects in nursing infants. Because of the potential for serious adverse reactions in nursing infants from corticosteroids, a decision should be made whether to discontinue nursing or to use a non-steroidal lead, taking into account the importance of the lead and the drug to the mother.

8 Directions for use

Warning: Before implanting a SureScan defibrillation system,
consider the risks associated with removing previously implanted leads. Abandoned leads or previously implanted leads not tested for MRI compatibility compromise the ability to safely scan the SureScan defibrillation system during MRI scans.
Proper surgical procedures and sterile techniques are the responsibility of the medical professional. The following procedures are provided for information only. Some implant techniques vary according to physician preference and the patient’s anatomy or physical condition. Each physician must apply the information in these instructions according to professional medical training and experience.

8.1 Opening the package

Use the following steps to open the sterile package and inspect the lead:
1. Within the sterile field, open the sterile package and remove the lead and accessories.
2. Inspect the lead. Leads that are shorter than 85 cm should have 1 anchoring sleeve on the lead body.

8.2 Verifying the mechanical functioning of the helix electrode

Before implant, verify the mechanical functioning of the helix electrode using the following steps:
1. If needed, slide the stylet guide away from the connector pin, and then press both legs of the fixation tool together and place the most distal hole on the DF4 connector pin (Figure 3).
Figure 3.
3
Mond, H. and Stokes, K.B., The Electrode - Tissue Interface: The Revolutionary Role of Steroid Elution, Pacing and Clinical Electrophysiology, Vol. 15, No. 1, pp 95-107
7
2. Keep the lead body and the DF4 connector sleeve as straight
1
as possible. Ensure that the stylet is fully inserted, then rotate the fixation tool clockwise until the helix electrode is fully extended (Figure 4). When the helix electrode is fully extended, approximately 1-1/2 to 2 helix coils are exposed.
Figure 4.
Caution: Do not severely bend the DF4 connector sleeve or
the lead body while extending the helix electrode. Caution: Overrotating the connector pin after the helix electrode is fully extended or fully retracted may damage the lead. The number of rotations required to extend or retract the helix electrode increases proportionately with the length of the lead. Additional curvatures made to the stylet may increase the number of rotations needed to extend or retract the helix electrode. Rotation of the fixation tool should be stopped once full helix retraction is visually verified. Overretraction of the helix may result in the inability to extend the helix. If the helix is unable to extend, use a new lead. Note: To determine the number of rotations applied to the lead, count the number of rotations of the fixation tool. See Chapter 9 for the maximum number of rotations to extend or retract the helix electrode. During the initial helix electrode extension, the helix electrode may extend suddenly due to accumulated torque in the lead, or the helix electrode may require additional turns for extension.
3. Disconnect the fixation tool from the connector pin and release the proximal end of the lead body. Allow several seconds for relief of the residual torque in the lead.
4. After allowing for relief of the residual torque, reattach the fixation tool and rotate it counterclockwise until the helix electrode tip is retracted into the sheath.

8.3 Inserting the lead

Cautions:
Certain anatomical abnormalities, such as thoracic outlet syndrome, may also precipitate pinching and subsequent fracture of the lead.
When using a subclavian approach, avoid techniques that may damage the lead.
Place the insertion site as far lateral as possible to avoid clamping the lead body between the clavicle and the first rib (Figure 5).
Figure 5.
1 Suggested entry site
Do not force the lead if significant resistance is encountered during lead passage.
Do not use techniques such as adjusting the patient’s posture to facilitate lead passage. If resistance is encountered, it is recommended that an alternate venous entry site be used.
2. Insert the tapered end of a vein lifter into the incised vein and gently push the lead tip underneath and into the vein (Figure 6). Note: A percutaneous lead introducer (PLI) kit may be used to facilitate insertion. If an introducer is used, it should be at least 3.0 mm (9 French). Refer to the technical manual packaged with an appropriate percutaneous lead introducer for further instructions.
Figure 6.
Caution: Use care when handling the lead during insertion.
Do not severely bend, kink, or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pins.
Insert the lead using the following techniques:
1. Select a site for lead insertion. The lead may be inserted by venotomy through several different venous routes, including the right or left cephalic vein, the subclavian vein, or the external or internal jugular vein. Use the cephalic vein whenever possible to avoid lead damage in the first rib or clavicular (thoracic inlet) space.
8
3. Advance the lead into the right atrium using a straight stylet to facilitate movement through the veins.

8.4 Positioning a screw-in ventricular lead

1
2
1 2
Caution: Use care when handling the lead during positioning.
Do not severely bend, kink, or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pins.
Use the following steps to position the lead:
1. After the lead tip is passed into the atrium, advance the lead through the tricuspid valve. Replace the straight stylet with a gently curved stylet to add control when maneuvering the lead through the tricuspid valve. Caution: Do not use a sharp object to impart a curve to the distal end of the stylet. Imparting a curve to the stylet can be accomplished with a smooth-surface, sterile instrument (Figure 7).
Figure 7.
Note: When you pass the lead tip through the tricuspid valve
or chordae tendineae, it may be difficult due to the flexible nature of the lead body. Rotate the lead body as the tip passes through the valve to facilitate passage.
2. After the lead tip is in the ventricle, the curved stylet may be replaced with a straight stylet. Withdraw the stylet slightly, to avoid using excessive tip force while achieving final electrode position. Avoid known infarcted or thin wall areas to minimize the occurrence of perforation.
3. Proper positioning of the helix electrode is essential for stable endocardial pacing. A satisfactory position usually is achieved when the lead tip points straight toward the apex, or when the distal end dips or bends slightly. Use fluoroscopy (lateral position) to ensure that the tip is not in a retrograde position or lodged in the coronary sinus. Note: With the helix electrode retracted, the distal end of the lead may be used to map a desirable site for electrode fixation. Mapping may reduce the need to repeatedly extend and fixate the helix electrode.
4. After placing the lead in a satisfactory position, extend the helix electrode by following the procedure in Section 8.5.

8.5 Securing the helix electrode into the endocardium

Secure the helix electrode using the following techniques:
1. If needed, slide the stylet guide away from the connector pin, and then press both legs of the fixation tool together and place the most distal hole on the DF4 connector pin (Figure 3).
2. Ensure that the stylet is inserted into the lead, and then press the lead tip against the endocardium by gently pushing the stylet and lead at the vein entry site.
3. Rotate the fixation tool clockwise until the helix electrode is fully extended (see Figure 4). Caution: Do not severely bend the DF4 connector sleeve or the lead body while extending the helix electrode. Use fluoroscopy to verify helix electrode exposure. Both a visual and fluoroscopic view of a fully retracted helix electrode is shown in Figure 8. Both a visual and fluoroscopic view of a fully extended helix electrode is shown in Figure 9. Closing of the space between the crimp sleeve and the indicator ring implies complete exposure of the helix electrode.
Figure 8.
1 Crimp sleeve 2 Indicator ring
Figure 9.
1 Crimp sleeve 2 Indicator ring
9
Cautions:
The number of rotations required to fully extend or retract the helix electrode is variable. Rotation should be stopped once full helix extension or retraction is verified with fluoroscopy as shown in Figure 8 and Figure 9. Overretraction of the helix, during initial implant or subsequent repositioning, may result in the inability to extend the helix. If the helix is unable to extend, replace with a new lead.
Do not exceed the recommended maximum number of rotations to extend or retract the helix electrode. Exceeding the maximum number may result in fracture or distortion of the inner conductor or helix electrode. Refer to Chapter 9 for the recommended maximum number of rotations.
Prolonged implant procedures or multiple repositionings may allow blood or body fluids to build up on the helix electrode mechanism. This may result in an increased number of rotations required to extend or retract the helix electrode.
4. Remove the fixation tool from the DF4 connector pin, and release the proximal end of the lead body. Allow several seconds for relief of the residual torque in the lead.
5. To assure helix electrode fixation, leave the stylet in place, hold the lead by the connector, and carefully rotate the lead body in 2 clockwise rotations.
6. Partially withdraw the stylet.
7. Obtain electrical measurements to verify satisfactory placement and electrode fixation. Refer to Section 8.6, “Taking electrical measurements and defibrillation efficacy measurements”, page 10.
8. Verify that the lead is affixed. Gently pull back on the lead, and check for resistance to verify fixation. A properly affixed helix electrode will remain in position. If the helix electrode is not properly affixed, the lead tip may become loose in the right ventricle.
9. If repositioning is required, reattach the fixation tool, and rotate counterclockwise until the helix electrode is retracted. Use fluoroscopy to verify withdrawal of the helix electrode before attempting to reposition.
10. After final positioning, make sure that the stylet and the fixation tool have been completely removed.
11. Obtain final electrical measurements. Refer to Section 8.6.

8.6 Taking electrical measurements and defibrillation efficacy measurements

1. Be sure to grasp the grooves of the plastic housing (see Figure 10) and not the metallic contacts.
Figure 10.
2. Grasp the ACI tool in the most convenient location (see Figure 11).
Figure 11.
Caution: The AccuRead tool reduces the risk of connector
damage, and reduces the risk of bridging and shorting that may occur while taking electrical measurements during implant. The potential for connector damage, bridging, and shorting is due to variations in analyzer cable terminals, as well as to the connector ring width and the proximity of the rings on the DF4 connector.
Note: The ACI tool may be removed or attached at any time during the procedure using the slit on the side of the tool (see Figure 12 or Figure 13).
Figure 12.
Caution: Prior to taking electrical or defibrillation efficacy
measurements, move objects made from conductive materials, such as guide wires, away from all electrodes. Metal objects, such as guide wires, can short a lead and an active implantable device, causing electrical current to bypass the heart and possibly damage the implantable device and lead.
The ACI tool is used to facilitate accurate electrical measurements during implant. The lead package will contain one of two existing ACI tool designs as shown in Figure 10 and Figure 11. When attaching or removing the ACI tool:
10
1 Removing the ACI tool from the connector pin 2 Removing the ACI tool from the stylet using the slit on the side of the
tool (Do not let go of the ACI tool; the ACI tool can fall off the stylet.)
Figure 13.
1 Removing the ACI tool from the connector pin 2 Removing the ACI tool from the stylet using the slit on the side of the
tool
Use the following steps to take electrical measurements:
1. Ensure that the fixation tool is disconnected from the DF4 connector pin.
2. Ensure that the lead connector is completely inserted into the ACI tool. The connector pin will be completely accessible if the ACI tool is properly attached (see Figure 14 or Figure 15).
Figure 14.
Figure 15.
In order to demonstrate reliable defibrillation efficacy, obtain final defibrillation measurements for the lead system.

Table 1. Recommended measurements at implant (when using a pacing system analyzer)

Measurements required Acutea lead system
Capture threshold (at 0.5 ms pulse width)
Pacing impedance 200–1000 Ω 200–1000 Ω
Filtered R-wave amplitude (during sinus rhythm)
Slew rate ≥0.75 V/s ≥0.45 V/s
a
<30 days after implant.
b
>30 days after implant.
≤1.0 V ≤3.0 V
≥5 mV ≥3 mV
Chronicb lead sys-
tem
If initial electrical measurements deviate from the recommended values, it may be necessary to repeat the testing procedure 15 min after final positioning. Initial electrical measurements may deviate from the recommended values:
Initial impedance values may exceed the measuring capabilities of the testing device, resulting in an error message.
Values may vary depending upon lead type, implantable device settings, cardiac tissue condition, and drug interactions.
If electrical measurements do not stabilize to acceptable levels, it may be necessary to reposition the lead and repeat the testing procedure.
Warning: If the implanted lead system fails to terminate a VF episode, rescue the patient promptly with an external defibrillator. At least 5 min should elapse between VF inductions.
For more information about obtaining electrical measurements, consult the product documentation supplied with the testing device.

8.7 Anchoring the lead

1 When properly attached, all 3 contacts are visible through the ACI tool
openings.
3. Attach a surgical cable to the ACI tool. Line up the cable clips with the contacts on the ACI tool to ensure that accurate readings are obtained.
4. Use a testing device, such as a pacing system analyzer, for obtaining electrical measurements (see Table 1 for recommended measurements). For information on the use of the testing device, consult the product literature for that device.
5. After the electrical measurements are complete, remove the surgical cable from the ACI tool before removing the tool from the lead.
Caution: Use care when anchoring the lead.
Use only nonabsorbable sutures to anchor the lead.
Do not attempt to remove or cut the anchoring sleeve from the lead body.
During lead anchoring, take care to avoid dislodging the lead tip.
Do not secure sutures so tightly that they damage the vein, lead, or anchoring sleeve (Figure 16).
Do not tie a suture directly to the lead body (Figure 16).
11
Figure 16.
Use the following steps to anchor the lead using all 3 grooves:
Note: The anchoring sleeves contain a radiopaque substance, which allows visualization of the anchoring sleeve on a standard x-ray and may aid in follow-up examinations.
1. Position the distal anchoring sleeve against or near the vein.
2. Secure the anchoring sleeve to the lead body by tying a suture firmly in each of the 3 grooves (Figure 17).
Figure 17.
3. Use at least one additional suture in one of the grooves to secure the anchoring sleeve and lead body to the fascia.
4. For abdominal implants, the redundant lead body (for example, a curve for strain relief) should be placed just proximal to the first anchoring sleeve. Then, the second anchoring sleeve may be lightly sutured to the lead body and fascia to hold the curve in place. This procedure helps isolate the vein entry site from tension on the proximal end of the lead body.
5. A slit anchoring sleeve may be used in the device pocket to secure excess lead length. First, secure the anchoring sleeve to the lead body. Then, orient the slit toward the fascia and secure the anchoring sleeve to the fascia with sutures.

8.8 Connecting the lead

Use the following steps to connect the lead to an implantable device:
1. Make sure that the stylet and all accessories have been completely removed. When removing the accessories, grip the lead firmly just below the ACI tool on the connector to prevent dislodgement.
2. Push the lead or plug into the header block until the color band on the tip of the lead connector pin is visible in the pin viewing area (see Figure 18). The color band will be visible when the lead is fully inserted. Consult the product literature packaged with the implantable device for instructions on proper lead connections.

Figure 18. Lead connector pin viewing area

1 Lead tip extends past setscrew block; lead connector pin is visible in
pin viewing area (color band may be used to verify full lead insertion) 2 Setscrew block, located behind grommet 3 Lead

8.9 Placing the device and lead into the pocket

Caution: Use care when placing the device and leads into
the pocket.
Ensure that the leads do not leave the device at an acute angle.
Do not grip the lead or device with surgical instruments.
Do not coil the lead. Coiling the lead can twist the lead body and may result in lead dislodgement (Figure 19).
Figure 19.
Use the following steps to place the device and leads into the pocket:
1. To prevent undesirable twisting of the lead body, rotate the device to loosely wrap the excess lead length (Figure 20).
Figure 20.
2. Insert the device and leads into the pocket.
12
3. Before closing the pocket, verify sensing, pacing, cardioversion, and defibrillation efficacy.

8.10 Post-implant evaluation

After implant, monitor the patient’s electrocardiogram until the patient is discharged. If a lead dislodges, it usually occurs during the immediate postoperative period.
Recommendations for verifying proper lead positioning include x-rays and pacing and sensing thresholds taken at pre-hospital discharge, 3 months after implant, and every 6 months thereafter.
In the event of a patient death, explant all implanted leads and devices and return them to Medtronic with a completed Product Information Report form. Call the appropriate phone number on the back cover if there are any questions on product handling procedures.

9 Specifications (nominal)

9.1 Detailed device description

Table 2. Specifications (nominal)

Parameter Model 6935M
Type Tripolar
Position Right ventricle
Fixation Extendable/retractable helix
Length 55 cm, 62 cm
Connector Quadripolar/true
Materials Conductors: MP35N coil
Electrodes (pace, sense): Platinized platinum alloy
Steroid Type: Dexamethasone acetate and
Conductor resistances
Helix length (extended) 1.8 mm
Diameters Lead body: 2.8 mm
Pacing (unipolar): 27.9 Ω (62 cm)
bipolar:
Insulation: Silicone, PTFE, ETFE
Overlay: Polyurethane
Seal Zone: PEEK
RV coil: Platinum-clad tantalum
DF4 pin: MP35N
DF4 rings: MP35N
Amount: 685 µg of dexamethasone ace-
Steroid binder: Silicone
Pacing (bipolar): 29.3 Ω (62 cm)
Defibrillation: 1.4 Ω (62 cm)
Four-pole inline (DF4-LLHO)
MP35N composite cables
dexamethasone sodium phos­phate
tate 59 µg of dexamethasone sodium phosphate
Tip: 2.8 mm
Table 2. Specifications (nominal) (continued)
Parameter Model 6935M
Helix: 1.4 mm
Lead introducer (recommended size)
without guide wire: 3.0 mm (9.0 French)
with guide wire: 3.7 mm (11.0 French)

Table 3. Maximum number of rotations to extend or retract the helix electrode

Lead length Number of rotations
55 cm 18
62 cm 20

Table 4. Respective electrode distances

Helix electrode to ring electrode 8 mm
Helix electrode to RV coil electrode 12 mm
13

Figure 21. Model 6935M distal lead components

Inactive
Inactive

Figure 22. Model 6935M proximal lead components

1 Helix electrode; surface area: 5.7 mm 2 Ring electrode; surface area: 25.2 mm 3 RV coil electrode; length: 57 mm; surface area: 614 mm2; electrical
shadow area: 506 mm
2
2
2
4 Anchoring sleeve
14
1 AccuRead tool 2 Connector pin 3 Stylet 4 RV contact 5 Ring (+) contact 6 Tip (-)

10 Medtronic disclaimer of warranty

For complete warranty information, see the accompanying warranty document.
Table 5. Explanation of symbols on package labeling (continued)
Symbol Explanation
Authorized representative in the European com­munity

11 Service

Medtronic employs highly trained representatives and engineers located throughout the world to serve you and, upon request, to provide training to qualified hospital personnel in the use of Medtronic products. Medtronic also maintains a professional staff to provide technical consultation to product users. For more information, contact your local Medtronic representative, or call or write Medtronic at the appropriate telephone number or address listed on the back cover.

12 Explanation of symbols on package labeling

Refer to the package labels to see which symbols apply to this product.

Table 5. Explanation of symbols on package labeling

Symbol Explanation
Conformité Européenne (European Conformity). This symbol means that the device fully complies with AIMD Directive 90/385/EEC (0123).
Do not use if package is damaged
Do not reuse
Upper limit of temperature
For US audiences only
Use by
Reorder number
Serial number
Lot number
Package contents
Product documentation
Accessories
Inner diameter
Lead
Lead length
Open here
Sterilized using ethylene oxide
Caution
Consult instructions for use
Date of manufacture
Manufacturer
Transvenous ventricular lead
Transvenous lead with one defibrillation electrode
Pace
Sense
Defibrillation
Extendable and retractable screw-in
15
Table 5. Explanation of symbols on package labeling (continued)
Symbol Explanation
Steroid-eluting
Lead introducer
Lead introducer with guide wire
MR Conditional symbol. The Medtronic SureScan defibrillation system is MR Conditional and is designed to allow implanted patients to undergo an MRI scan under the specified MRI conditions for use.
SureScan symbol
16
Medtronic, Inc.
*M961371A001*
710 Medtronic Parkway Minneapolis, MN 55432 USA www.medtronic.com +1 763 514 4000
Medtronic USA, Inc.
Toll-free in the USA (24-hour technical consultation for physicians and medical professionals) Bradycardia: +1 800 505 4636 Tachycardia: +1 800 723 4636
Europe/Middle East/Africa
Medtronic International Trading Sàrl Route du Molliau 31 Case Postale 84 CH-1131 Tolochenaz Switzerland +41 21 802 7000
Technical manuals
www.medtronic.com/manuals
© 2015 Medtronic, Inc. M961371A001 B 2015-07-05
Loading...