Medtronic 6935 Technical Manual

Page 1
SPRINT QUATTRO SECURE S™ 6935
Steroid-eluting, tripolar, screw-in, ventricular lead with RV defibrillation coil electrode
Technical Manual
Caution: Federal Law (USA) restricts this device to sale by or on the order of a physician.
Page 2
The following are trademarks or registered trademarks of Medtronic in the United States and possibly in other countries:
Medtronic, Sprint Quattro, Sprint Quattro Secure, Sprint Quattro Secure S
Page 3
1 Description 3 2 Drug component description 3 3 Indications 4 4 Contraindications 4 5 Warnings and precautions 4 6 Potential complications 5 7 Drug Information 6 8 Adverse events and clinical trial data 6
9 Implant procedure 6 10 Specifications (nominal) 11 11 Medtronic warranty 12 12 Service 12

1 Description

The Medtronic Sprint Quattro Secure S Model 6935 lead is a steroid-eluting, tripolar, screw-in, ventricular lead with a right ventricular (RV) defibrillation coil electrode. The lead is designed for pacing, sensing, cardioversion, and defibrillation therapies.
The lead features an extendable/retractable helix electrode, silicone insulation with overlay, and parallel conductors. The 3 electrodes of the lead are the helix, ring, and RV coil. The ring and helix electrodes are platinized. See Figure 13 for an illustration of the lead.
The helix electrode is common to the connector pin of the IS-11 BI pin connector.
The ring electrode is common to the connector ring of the IS-1 BI connector.
The RV coil electrode is common to the connector pin of the RV DF-12 connector (red band).
The RV coil delivers cardioversion and defibrillation therapies. Pacing and sensing occur between the helix and ring electrodes.
The helix electrode is made of platinized platinum alloy that can be actively fixed into the endocardium. The helix electrode can be extended or retracted by rotating the IS-1 connector pin with the fixation tool.
The IS-1 bipolar leg of the bifurcation features a lumen for stylet passage. The DF-1 connector legs will not accept a stylet.
The distal tip of the lead contains a maximum of 1.0 mg dexamethasone sodium phosphate. Upon exposure to body fluids, the steroid elutes from the lead tip. The steroid is known to
suppress the inflammatory response that is believed to cause threshold rises typically associated with implanted pacing electrodes.

1.1 Package contents

Leads and accessories are supplied sterile. Each package contains the following items:
1 lead with 1 radiopaque anchoring sleeve3, stylet, and stylet guide
2 fixation tools
1 slit anchoring sleeve
1 vein lifter
2 pin caps
extra stylets
product documentation

1.2 Accessory descriptions

Anchoring sleeve – An anchoring sleeve secures the lead to
prevent it from moving and protects the lead insulation and conductors from damage caused by tight sutures.
Fixation tool – The fixation tool facilitates connector pin rotation.
Pin cap – A pin cap covers and insulates unused connector pins.
Slit anchoring sleeve – A slit anchoring sleeve secures excess
lead length in the device pocket.
Stylet – A stylet provides additional stiffness and controlled flexibility for maneuvering the lead into position. Each stylet knob is labeled with the stylet diameter and length.
Stylet guide – A stylet guide facilitates stylet insertion into the lead.
Vein lifter – A vein lifter facilitates lead insertion into a vein.

2 Drug component description

The active ingredient in the Model 6935 lead is dexamethasone sodium phosphate [9-fluoro-11β, 17-dihydroxy-16α-methyl-21-(phosphonooxy) pregna-1,4-diene-3,20-dione disodium salt]. Dexamethasone sodium phosphate (DSP) is an inorganic ester of dexamethasone, a synthetic adrenocortical steroid. DSP is a white or slightly yellow crystalline powder. It is freely soluble in water and is very hygroscopic.
The structural formula for this steroid is shown below.
1
IS-1 refers to the International Connector Standard (ISO 5841–3) whereby pulse generators and leads so designated are assured of a basic mechanical fit.
2
DF-1 refers to the International Connector Standard (ISO 11318) whereby pulse generators and leads so designated are assured of a basic mechanical fit.
3
Two radiopaque anchoring sleeves are provided with leads 85 cm or longer.
3
Page 4
Figure 1. Structural formula for dexamethasone sodium phosphate C22H28FNa2O8P
The maximum dosage of dexamethasone sodium phosphate is less than 1.0 mg per lead.

3 Indications

The lead is intended for single, long-term use in the right ventricle.
This lead has application for patients for whom implantable cardioverter defibrillators are indicated.

4 Contraindications

Atrial use – The lead is contraindicated for the sole use of
detection and treatment of atrial arrhythmias.
Ventricular use – The lead is contraindicated for ventricular use in patients with tricuspid valvular disease or a tricuspid mechanical heart valve.
Transient ventricular tachyarrhythmias – The lead is contraindicated for patients with transient ventricular tachyarrhythmias due to reversible causes (drug intoxication, electrolyte imbalance, sepsis, hypoxia) or other factors (myocardial infarction, electric shock).
Steroid use – Use of steroid eluting transvenous leads is contraindicated in patients for whom a single dose of 1.0 mg dexamethasone sodium phosphate may be contraindicated.

5 Warnings and precautions

Line-powered and battery-powered equipment – An
implanted lead forms a direct current path to the myocardium. During lead implant and testing, use only battery-powered equipment or line-powered equipment specifically designed for this purpose to protect against fibrillation that may be caused by alternating currents. Line-powered equipment used in the vicinity of the patient must be properly grounded. Lead connector pins must be insulated from any leakage currents that may arise from line-powered equipment.
Diathermy – People with metal implants such as pacemakers, implantable cardioverter defibrillators (ICDs), and accompanying leads should not receive diathermy treatment. The interaction
between the implant and diathermy can cause tissue damage, fibrillation, or damage to the device components, which could result in serious injury, loss of therapy, and/or the need to reprogram or replace the device.
Electrophysiologic testing – Prior to lead implant, it is strongly recommended that patients undergo a complete cardiac evaluation, which should include electrophysiologic testing. Also, electrophysiologic evaluation and testing of the safety and efficacy of the proposed pacing, cardioversion, or defibrillation therapies are recommended during and after the implant of the system.
Single use – The lead is for single use only.
Inspecting the sterile package – Inspect the sterile package
with care before opening it.
Contact a Medtronic representative if the seal or package is damaged.
Do not store this product above 40 °C (104 °F).
Do not use the product after its expiration date.
Sterilization – Medtronic has sterilized the package contents with ethylene oxide before shipment. This lead is for single use only and is not intended to be resterilized.
Steroid use – It has not been determined whether the warnings, precautions, or complications usually associated with injectable dexamethasone sodium phosphate apply to the use of this highly localized, controlled-release device. For a list of potential adverse effects, refer to the Physicians’ Desk Reference.
Handling the steroid tip – Avoid reducing the amount of steroid available before implanting the lead. Reducing the available amount of steroid may adversely affect low-threshold performance.
Do not allow the electrode surface to come in contact with surface contaminants.
Do not wipe or immerse the electrode in fluid, except blood, at the time of implant.
Drug interactions – Drug interaction of dexamethasone sodium phosphate with this lead has not been studied.
Use of multiple leads – Prior to implanting the lead, total patient exposure to dexamethasone sodium phosphate should be considered.
Handling the lead – Handle the lead with care at all times.
Protect the lead from materials that shed particles such as lint and dust. Lead insulators attract these particles.
Handle the lead with sterile surgical gloves that have been rinsed in sterile water or a comparable substance.
Do not severely bend, kink, or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pins.
Do not immerse the lead in mineral oil, silicone oil, or any other liquid, except blood, at the time of implant.
4
Page 5
Inserting the lead using a lead introducer that features a hemostasis valve may require a larger introducer than the size recommended. To avoid distortion of the coil electrode, do not withdraw the lead through a hemostasis valve.
Do not implant the lead without first verifying the mechanical functioning of the helix electrode. Refer to Section 9.2, “Verifying the mechanical functioning of the helix electrode”, page 7, for complete instructions.
Do not rotate the helix electrode after it is fully extended or fully retracted. The number of rotations required to fully extend or retract the helix electrode is variable.
Furthermore, do not exceed the recommended maximum number of rotations to extend or retract the helix electrode. Exceeding the maximum number may result in fracture or distortion of the inner conductor or helix electrode. Refer to Section 10 for the recommended maximum number of rotations.
Handling the stylet – Handle the stylet with care at all times.
Do not use a sharp object to impart a curve to the distal end of the stylet.
Do not use excessive force or surgical instruments when inserting the stylet into the lead.
Avoid overbending or kinking the stylet.
Use a new stylet when blood or other fluids accumulate on the stylet. Accumulated blood or other fluids may damage the lead or cause difficulty in passing the stylet into the lead.
Necessary hospital equipment – Keep external defibrillation equipment nearby for immediate use during acute lead system testing, the implant procedure, or whenever arrhythmias are possible or intentionally induced during post-implant testing.
Concurrent devices – Output pulses, especially from unipolar devices, may adversely affect device sensing capabilities. If a patient requires a separate stimulation device, either permanent or temporary, allow enough space between the leads of the separate systems to avoid interference in the sensing capabilities of the devices. Previously implanted pulse generators and implantable cardioverter defibrillators should generally be explanted.
Chronic repositioning or removal of a screw-in lead –
Proceed with extreme caution if a lead must be removed or repositioned. Chronic repositioning or removal of screw-in transvenous leads may not be possible because of blood or fibrotic tissue development into the helix mechanism on the lead. In most clinical situations, it is preferable to abandon unused leads in place. Return all removed leads, unused leads, or lead sections to Medtronic for analysis.
Note: If a helix does not disengage from the endocardium by rotating the connector pin, rotating the lead body counterclockwise may withdraw the helix and decrease the
possibility of damage to cardiovascular structures during removal.
Lead removal may result in avulsion of the endocardium, valve, or vein.
Lead junctions may separate, leaving the lead tip and bare wire in the heart or vein.
Chronic repositioning of a lead may adversely affect the low-threshold performance of a steroid lead.
An abandoned lead should be capped so that the lead does not transmit electrical signals.
Severed leads should have the remaining lead end sealed and the lead body sutured to adjacent tissue.
Second anchoring sleeve – Leads 85 cm or longer feature 2 anchoring sleeves. Use both anchoring sleeves to assure adequate fixation.
Connector compatibility – Although the lead conforms to the International Connector Standards IS-1 and DF-1, do not attempt to use the lead with any device other than a commercially available implantable defibrillator system with which it has been tested and demonstrated to be safe and effective. The potential adverse consequences of using such a combination may include, but are not limited to, undersensing cardiac activity and failure to deliver necessary therapy.

6 Potential complications

The potential complications related to the use of transvenous leads include, but are not limited to, the following patient-related conditions that can occur when the lead is being inserted or repositioned:
cardiac perforation
cardiac tamponade
constrictive pericarditis
embolism
endocarditis
fibrillation or other arrhythmias
heart wall rupture
hemothorax
infection
pneumothorax
thrombosis
tissue necrosis
Other potential complications related to the lead include, but are not limited to, the following complications:
insulation failure
lead conductor or electrode fracture
lead dislodgment
poor connection to the device, which may lead to oversensing, undersensing, or a loss of therapy
5
Page 6

7 Drug Information

7.1 Steroid mechanism of action

Steroid suppresses the inflammatory response that is believed to cause threshold rises typically associated with implanted pacing electrodes. Dexamethasone sodium phosphate is a synthetic steroid of the glucocorticoid family. Glucocorticoids have potent anti-inflammatory actions via direct and indirect effects on major inflammatory cells. Glucocorticosteroids bind to a cytoplasmic glucocorticoid receptor as well as a membrane-bound receptor. Binding to the cytoplasmic receptor leads to receptor activation and translocation to the nucleus. The receptor interacts with specific DNA sequences within the regulatory regions of affected genes. Thus, glucocorticoids inhibit the production of multiple cell factors that are critical in generating the inflammatory response.

7.2 Pharmacodynamics of the Model 6935 lead

Pharmacokinetics – The pharmacokinetics (local drug levels
and systemic levels) of dexamethasone sodium phosphate and its metabolites following lead implantation were not evaluated in human clinical trials.
The in-vivo elution profile of a tined pacemaker lead with a DSP monolithic controlled release device, based upon an assay of explanted leads, is shown in Mond and Stokes4.
Metabolism – The conversion of DSP to dexamethasone occurs within minutes. The dexamethasone alcohol (dexamethasone) is the active glucocorticoid used in Medtronic leads. Steroid is applied to the tip and eluted through the electrode tip to the tissue interface where it will be used. Dexamethasone sodium phosphate is hydrolyzed into dexamethasone, which is readily absorbed by the surrounding tissue and body fluids. Glucocorticoids, when given systemically, are eliminated primarily by renal excretion of inactive metabolites.
Mutagenesis, carcinogenicity and reproductive toxicology – The mutagenesis, carcinogenicity, and
reproductive toxicity of the Model 6935 lead have not been evaluated. However, the mutagenesis, carcinogenicity, and reproductive toxicity of dexamethasone sodium phosphate have previously been evaluated.
Carcinogenesis, mutagenesis, impairment of fertility – No adequate studies have been conducted in animals to determine whether corticosteroid have a potential for carcinogenesis (tumor initiation or promotion). Dexamethasone was genotoxic in assays for clastogenicity (including sister chromatid exchange in human lymphocytes) but not in an assay for mutagenicity in salmonella (Ames test).
Adrenocorticoids have been reported to increase or decrease the number and mobility of spermatozoa in some patients.
Pregnancy – Pregnancy category C. Dexamethasone sodium phosphate has been shown to be teratogenic in many species when given in doses equivalent to the human dose. There are no adequate and well-controlled studies in pregnant women.
Dexamethasone sodium phosphate should be used during pregnancy only if the potential benefit justifies the potential risk to the fetus. Studies in mice, rats, and rabbits have shown that adrenocorticoids increase the incidence of cleft palate, placental insufficiency, spontaneous abortions, and intrauterine growth rate.
Nursing mothers – Systemically administered corticosteroids appear in human milk and could suppress growth, interfere with endogenous corticosteroid production, or cause other untoward effects in nursing infants. Because of the potential for serious adverse reactions in nursing infants from corticosteroids, a decision should be made whether to discontinue nursing or to use a non-steroidal lead, taking into account the importance of the lead and the drug to the mother.

8 Adverse events and clinical trial data

Information regarding the Model 6935 lead clinical study and adverse events is available at www.medtronic.com/manuals. To view, download, print, or order the clinical study from the Medtronic website:
1. Navigate your web browser to www.medtronic.com/manuals.
2. Set the “Your Location” option to United States and click [OK].
3. Select the “Model # or Name” field on the left side of the screen and type “6935.”
4. Click [Search]. All technical literature for this lead will be listed.
The following clinical studies are related to this device:
Model 6932 RV Lead clinical study
Model 6947 RV Lead clinical study

9 Implant procedure

Proper surgical procedures and sterile techniques are the responsibility of the medical professional. The following procedures are provided for information only. Some implant techniques vary according to physician preference and the patient’s anatomy or physical condition.

9.1 Opening the package

Use the following steps to open the sterile package and inspect the lead:
1. Within the sterile field, open the sterile package and remove the lead and accessories.
2. Inspect the lead. Leads shorter than 85 cm should have 1 anchoring sleeve on the lead body. Leads 85 cm or longer should have 2 anchoring sleeves on the lead body.
4
Mond, H. and Stokes, K.B., The Electrode - Tissue Interface: The Revolutionary Role of Steroid Elution, Pacing and Clinical Electrophysiology, Vol. 15, No. 1, pp 95-107
6
Page 7

9.2 Verifying the mechanical functioning of the helix electrode

Before implant, verify the mechanical functioning of the helix electrode using the following steps:
1. With a stylet inserted into the lead, press both legs of the fixation tool together and place the most distal hole on the IS-1 connector pin (Figure 2). Note: The stylet guide may be removed by gently pulling it off. To reattach a stylet guide, gently push it onto the connector pin.
Figure 2.
2. Hold the IS-1 connector leg of the lead with the thumb on one side and 4 fingers on the other side. Keep the lead body and the IS-1 connector leg as straight as possible. Ensure that the stylet is fully inserted, then rotate the fixation tool clockwise until the helix electrode is fully extended (Figure 3). When the helix electrode is fully extended, approximately 1.5 to 2 helix coils are exposed.
Figure 3.
the lead is bent on either side of the lead bifurcation during helix electrode extension or retraction, the lead may be damaged.
Figure 4.
Caution: Over-rotating the connector pin after the helix
electrode is fully extended or fully retracted may damage the lead. The number of rotations required to extend or retract the helix electrode increases proportionately with the length of the lead. Additional curvatures made to the stylet may increase the number of rotations needed to extend or retract the helix electrode. Rotation should be stopped once full helix retraction is visually verified. Over-retraction of the helix may result in the inability to extend the helix. If the helix is unable to extend, use a new lead. During the initial helix electrode extension, the helix electrode may extend suddenly due to accumulated torque in the lead, or the helix electrode may require additional turns for extension.
3. Disconnect the fixation tool from the connector pin and release the proximal end of the lead body. Allow several seconds for relief of the residual torque in the lead.
4. After allowing for relief of the residual torque, reattach the fixation tool and rotate it counterclockwise until the helix electrode tip is retracted into the sheath.

9.3 Inserting the lead

Caution: Do not severely bend the IS-1 connector leg or the
lead body while extending the helix electrode (Figure 4). If
Caution: Use care when handling the lead during insertion.
Do not severely bend, kink, or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pins.
Insert the lead using the following techniques:
1. Select a site for lead insertion. The lead may be inserted by venotomy through several different venous routes, including the right or left cephalic vein or the external or internal jugular vein. Use the cephalic vein whenever possible to avoid lead damage in the first rib or clavicular (thoracic inlet) space.
7
Page 8
Cautions:
1
Certain anatomical abnormalities, such as thoracic outlet syndrome, may also precipitate pinching and subsequent fracture of the lead.
When using a subclavian approach, avoid techniques that may damage the lead.
Place the insertion site as far lateral as possible to avoid clamping the lead body between the clavicle and the first rib (Figure 5).
Figure 5.
1 Suggested entry site
Do not force the lead if significant resistance is encountered during lead passage.
Do not use techniques such as adjusting the patient’s posture to facilitate lead passage. If resistance is encountered, it is recommended that an alternate venous entry site be used.
2. Insert the tapered end of a vein lifter into the incised vein and gently push the lead tip underneath and into the vein (Figure 6). Note: A percutaneous lead introducer (PLI) kit may be used to facilitate insertion. Refer to the technical manual packaged with an appropriate percutaneous lead introducer for further instructions.
Figure 6.
3. Advance the lead into the right atrium using a straight stylet to facilitate movement through the veins.

9.4 Positioning the lead

Caution: Use care when handling the lead during positioning.
Do not severely bend, kink, or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pins.
Use the following steps to position the lead:
1. After the lead tip is passed into the atrium, advance the lead through the tricuspid valve. Replace the straight stylet with a gently curved stylet to add control when maneuvering the lead through the tricuspid valve. Caution: Do not use a sharp object to impart a curve to the distal end of the stylet. Imparting a curve to the stylet can be accomplished with a smooth-surface, sterile instrument (Figure 7).
Figure 7.
Note: When you pass the lead tip through the tricuspid valve
or chordae tendineae it may be difficult due to the flexible nature of the lead body. Rotate the lead body as the tip passes through the valve to facilitate passage.
2. After the lead tip is in the ventricle, the curved stylet may be replaced with a straight stylet. Withdraw the stylet slightly, to avoid using excessive tip force while achieving final electrode position. Avoid known infarcted or thin wall areas to minimize the occurrence of perforation.
3. Proper positioning of the helix electrode is essential for stable endocardial pacing. A satisfactory position usually is achieved when the lead tip points straight toward the apex, or when the distal end dips or bends slightly. Use fluoroscopy (lateral position) to ensure that the tip is not in a retrograde position or lodged in the coronary sinus. Note: With the helix electrode retracted, the distal end of the lead may be used to map a desirable site for electrode fixation. Mapping may reduce the need to repeatedly extend and fixate the helix electrode.
4. After placing the lead in a satisfactory position, extend the helix electrode by following the procedure in Section 9.5.

9.5 Securing the helix electrode into the endocardium

Secure the helix electrode using the following techniques:
1. Press both legs of the fixation tool together and place the most distal hole on the IS-1 connector pin (Step 1).
8
Page 9
2. Press the lead tip against the endocardium by gently pushing
A B
AB
the stylet and lead at the vein entry site.
3. Rotate the fixation tool clockwise until the helix electrode is fully extended. Caution: Do not severely bend the IS-1 connector leg or the lead body while extending the helix electrode. If the lead is bent on either side of the lead bifurcation during helix electrode extension or retraction, the lead may be damaged. Use fluoroscopy to verify helix electrode extension. The fluoroscope head may need to be rotated to obtain an adequate view. Both a visual and fluoroscopic view of a fully retracted and extended helix electrode is shown in Figure 8. Closing of the space between the indicator stop (A) and the indicator ring (B) implies complete extension of the helix electrode.
Figure 8.
1 Fully retracted visual 2 Fully retracted fluoroscopic 3 Fully extended visual 4 Fully extended fluoroscopic
Cautions:
The number of rotations required to fully extend or retract the helix electrode is variable. Rotation should be stopped once full helix extension or retraction is verified with fluoroscopy as shown in Figure 8. Over-retraction of the helix, during initial implant or subsequent repositioning, may result in the inability to extend the helix. If the helix is unable to extend, replace with a new lead.
Do not exceed the recommended maximum number of rotations to extend or retract the helix electrode. Exceeding the maximum number may result in fracture or distortion of the inner conductor or helix electrode. Refer to Section 10 for the recommended maximum number of rotations.
Prolonged implant procedures or multiple repositionings may allow blood or body fluids to build up on the helix electrode mechanism. This may result in an increased number of rotations required to extend or retract the helix electrode.
4. Remove the fixation tool from the IS-1 connector pin, and release the proximal end of the lead body. Allow several seconds for relief of the residual torque in the lead.
5. To assure helix electrode fixation, leave the stylet in place, hold the lead by the connector, and carefully rotate the lead body in 2 clockwise rotations.
6. Partially withdraw the stylet.
7. Obtain electrical measurements to verify satisfactory placement and electrode fixation. Refer to Section 9.6, “Taking electrical measurements and defibrillation efficacy measurements”, page 9.
8. Verify that the lead is affixed. Gently pull back on the lead, and check for resistance to verify fixation. A properly affixed helix electrode will remain in position. If the helix electrode is not properly affixed, the lead tip may become loose in the right ventricle.
9. If repositioning is required, reattach the fixation tool, and rotate counterclockwise until the helix electrode is retracted. Use fluoroscopy to verify withdrawal of the helix electrode before attempting to reposition.
10. After final positioning, make sure that the stylet and the stylet guide have been completely removed. When removing the stylet guide, grip the lead firmly just below the connector pin to help prevent lead dislodgment.
11. Obtain final electrical measurements. Refer to Section 9.6.

9.6 Taking electrical measurements and defibrillation efficacy measurements

Caution: Prior to taking electrical or defibrillation efficacy
measurements, move objects made from conductive materials, such as guide wires, away from all electrodes. Metal objects, such as guide wires, can short a lead and an active implantable device, causing electrical current to bypass the heart and possibly damage the implantable device and lead.
Use the following steps to take electrical measurements:
1. Attach a surgical cable to the lead connector pin. A notch in the style guide allows connection of a surgical cable for obtaining electrical measurements.
2. Use a testing device, such as a pacing system analyzer, for obtaining electrical measurements. For information on the use of the testing device, consult the product documentation for that device.
In order to demonstrate reliable defibrillation efficacy, obtain final defibrillation measurements for the lead system.

Table 1. Recommended measurements at implant (when using a pacing system analyzer)

Measurements required Acutea lead system
Capture threshold (at 0.5 ms pulse width)
Pacing impedance 200–1000 Ω 200–1000 Ω
≤1.0 V ≤3.0 V
Chronicb lead sys-
tem
9
Page 10
Table 1. Recommended measurements at implant (when using a pacing system analyzer) (continued)
Measurements required Acutea lead system
Filtered R-wave amplitude (during sinus rhythm)
Slew rate ≥0.75 V/s ≥0.45 V/s
a
<30 days after implant.
b
>30 days after implant.
≥5 mV ≥3 mV
Chronicb lead sys-
tem
If initial electrical measurements deviate from the recommended values, it may be necessary to repeat the testing procedure 15 min after final positioning. Initial electrical measurements may deviate from the recommended values:
Initial impedance values may exceed the measuring capabilities of the testing device, resulting in an error message.
Values may vary depending upon lead type, implantable device settings, cardiac tissue condition, and drug interactions.
If electrical measurements do not stabilize to acceptable levels, it may be necessary to reposition the lead and repeat the testing procedure.
In order to keep patient morbidity and mortality to a minimum, patients should be rescued promptly with an external defibrillator if the implanted lead system fails to terminate a VF episode. At least 5 min should elapse between VF inductions.
For more information on obtaining electrical measurements, consult the product documentation supplied with the testing device.

9.7 Anchoring the lead

Caution: Use care when anchoring the lead.
Use only nonabsorbable sutures to anchor the lead.
Do not attempt to remove or cut the anchoring sleeve from the lead body.
During lead anchoring, take care to avoid dislodging the lead tip.
Do not secure sutures so tightly that they damage the vein, lead, or anchoring sleeve (Figure 9).
Do not tie a suture directly to the lead body (Figure 9).
Figure 9.
Use the following steps to anchor the lead using all 3 grooves:
1. Position the distal anchoring sleeve against or near the vein.
2. Secure the anchoring sleeve to the lead body by tying a suture firmly in each of the 3 grooves (Figure 10).
Figure 10.
3. Use at least 1 additional suture in 1 of the grooves to secure the anchoring sleeve and lead body to the fascia.
4. A second anchoring sleeve is provided with leads 85 cm or longer. For abdominal implants, redundant lead body (for example, a curve for strain relief) should be placed just proximal to the first anchoring sleeve. Then, the second anchoring sleeve may be lightly sutured to the lead body and fascia to hold the curve in place. This procedure helps isolate the vein entry site from tension on the proximal end of the lead body.
5. A slit anchoring sleeve may be used in the device pocket to secure excess lead length. First, secure the anchoring sleeve to the lead body. Then, orient the slit toward the fascia and secure the anchoring sleeve to the fascia with sutures.

9.8 Connecting the lead

Use the following steps to connect the lead to an implantable device:
1. Carefully remove the stylet and stylet guide. When removing the stylet and stylet guide, grip the lead firmly just below the connector pin to prevent dislodgment.
2. Insert the lead connectors into the lead connector block. Consult the product documentation packaged with the implantable device for instructions on proper lead connections.

9.9 Placing the device and lead into the pocket

Caution: Use care when placing the device and leads into
the pocket.
Ensure that the leads do not leave the device at an acute angle.
Do not grip the lead or device with surgical instruments.
Do not coil the lead. Coiling the lead can twist the lead body and may result in lead dislodgment (Figure 11).
10
Page 11
Figure 11.
Use the following steps to place the device and leads into the pocket:
1. To prevent undesirable twisting of the lead body, rotate the device to loosely wrap the excess lead length (Figure 12).
Figure 12.
2. Insert the device and leads into the pocket.
3. Before closing the pocket, verify sensing, pacing, cardioversion, and defibrillation efficacy.

9.10 Post-implant evaluation

After implant, monitor the patient’s electrocardiogram until the patient is discharged. If a lead dislodges, it usually occurs during the immediate postoperative period.
Recommendations for verifying proper lead positioning include x-rays and pacing/sensing thresholds taken at pre-hospital discharge, 3 months after implant, and every 6 months thereafter.
In the event of a patient death, explant all implanted leads and devices and return them to Medtronic with a completed Product Information Report form. Call the appropriate phone number on the back cover if there are any questions on product handling procedures.

10 Specifications (nominal)

10.1 Detailed device description

Table 2. Specifications (nominal)

Parameter Model 6935
Type Tripolar
Position Right ventricle
Fixation Extendable/retractable helix
Length 40–110 cm
Connectors Unipolar: DF-1
Bipolar: IS-1
Materials Conductors: MP35N coil
Insulation: Silicone, PTFE, ETFE
Overlay: Polyurethane
Electrodes (pace, sense): Platinized platinum alloy
RV coil: Platinum-clad tantalum
DF-1 pin: Stainless steel
IS-1 pin and rings: Stainless steel
Steroid Type: Dexamethasone sodium phos-
Amount: 1.0 mg maximum
Steroid binder: Polyurethane
Conductor resistances
Helix length (extended) 1.8 mm
Diameters Lead body: 2.8 mm
Lead introducer (recommended size)
Pacing (unipolar): 29 Ω (65 cm)
Pacing (bipolar): 32.3 Ω (65 cm)
Defibrillation: <1.2 Ω (65 cm)
without guide wire: 3.0 mm (9.0 French)
with guide wire: 3.7 mm (11.0 French)

Table 3. Maximum number of rotations to extend or retract the helix electrode

Lead length Number of rotations
52 cm 16
58 cm 18
65 cm 20
75 cm 22
100 cm 27
MP35N composite cables
phate
Tip: 2.8 mm
Helix: 1.4 mm
11
Page 12

Figure 13. Model 6935 distal lead components

7
8
7 DF-1 connector (red band): connector pin is common to RV coil
electrode
8 IS-1 BI connector: connector pin is common to tip electrode;
connector ring is common to ring electrode

11 Medtronic warranty

For complete warranty information, see the accompanying warranty document.

12 Service

Medtronic employs highly trained representatives and engineers located throughout the world to serve you and, upon request, to provide training to qualified hospital personnel in the use of Medtronic products. Medtronic also maintains a professional staff to provide technical consultation to product users. For more information, contact your local Medtronic representative, or call or write Medtronic at the appropriate address or telephone number listed on the back cover.
1 Helix electrode, surface area: 5.7 mm 2 Ring electrode, surface area: 25.2 mm 3 RV coil electrode, length: 57 mm; surface area: 614 mm2; electrical
shadow area: 506 mm 4 Anchoring sleeve: leads 85 cm or longer have 2 anchoring sleeves 5 Helix electrode to ring electrode, length 8 mm 6 Helix electrode to RV coil, length: 12 mm
12
2
2
2
Page 13
Page 14
World Headquarters
*M945870A001*
Medtronic, Inc. 710 Medtronic Parkway Minneapolis, MN 55432-5604 USA www.medtronic.com Tel. +1-763-514-4000 Fax +1-763-514-4879
Medtronic USA, Inc.
Toll-free in the USA (24-hour technical consultation for physicians and medical professionals) Bradycardia: 1-800-505-4636 Tachycardia: 1-800-723-4636
Europe/Africa/Middle East Headquarters
Medtronic International Trading Sàrl Route du Molliau 31 Case Postale 84 CH-1131 Tolochenaz Switzerland www.medtronic.com Tel. +41-21-802-7000 Fax +41-21-802-7900
Medtronic E.C. Authorized Representative/Distributed by
Medtronic B.V. Earl Bakkenstraat 10 6422 PJ Heerlen The Netherlands Tel. +31-45-566-8000 Fax +31-45-566-8668
Technical manuals: www.medtronic.com/manuals
© Medtronic, Inc. 2010 M945870A001A 2010-10-20
Loading...