Medtronic 693152 Technical Manual

Page 1
SPRINT FIDELIS™ 6931
Steroid eluting, tripolar, screw-in, ventricular lead with RV defibrillation coil electrode
Technical manual
Caution: Federal Law (USA) restricts this device to sale by or on the order of a physician (or properly licensed practitioner).
Page 2
The following are trademarks of Medtronic: Medtronic, Sprint Fidelis, Sprint Quattro Secure, Quick Twist
Page 3
Table of contents
Device description 5
Contents of package 6 Accessory descriptions 6
Indications 6
Contraindications 7
Warnings and precautions 7
Potential adverse events 10
Clinical study 11
Model 6947 clinical study 11 Model 6932 (single-coil) clinical study 14 Additional performance measurements with active can ICDs 16
Directions for use 16
Opening the package 16 Verifying the mechanical functioning of the helix electrode 17 Inserting the lead 19 Positioning the lead 20 Securing the helix electrode into the endocardium 21 Taking electrical measurements and defibrillation efficacy measurements 23 Anchoring the lead 24 Connecting the lead 25 Placing the device and leads into the pocket 25 Post-implant evaluation 26
Detailed device description 27
Specifications (nominal) 27 Specifications drawing (nominal) 28
Medtronic warranty 29
Service 29
6931 Technical Manual 3
Page 4
Page 5

Device description

The Medtronic Sprint Fidelis Model 6931 lead is a steroid eluting, tripolar, screw-in, ventricular lead with a right ventricular (RV) defibrillation coil electrode. The lead is designed for pacing, sensing, cardioversion, and defibrillation therapies.
The lead features an extendable/retractable helix electrode, silicone insulation with overlay, parallel conductors, titanium nitride coated platinum iridium tip and ring electrodes, and a RV coil electrode. See “Specifications drawing (nominal)” on page 28 for a lead drawing.
The helix electrode is common to the connector pin of the
1
bipolar leg.
IS-1
The ring electrode is common to the connector ring of the IS-1 bipolar leg.
The RV coil electrode is common to the DF-12 leg of the bifurcation, labeled and marked with a red band.
The RV coil delivers cardioversion and defibrillation therapies. Pacing and sensing occur between the helix and ring electrodes.
The IS-1 bipolar leg of the bifurcation features a lumen for stylet passage. The DF-1 connector legs will not accept a stylet.
The helix electrode can be actively fixed into the endocardium. The helix electrode can be extended or retracted by rotating the IS-1 connector pin with either the Quick Twist tool assembled on the lead or the white fixation tool.
The distal tip contains a maximum of 1.0 mg of dexamethasone acetate. Exposure to body fluids elutes the steroid from the lead tip. The steroid is known to suppress the inflammatory response that is believed to cause threshold rises typically associated with implanted pacing electrodes.
1
IS-1 refers to the International Connector Standard (ISO 5841-3) whereby pulse generators and leads so designated are assured of a basic mechanical fit.
2
DF-1 refers to the International Connector Standard (ISO 11318) whereby pulse generators and leads so designated are assured of a basic mechanical fit.
6931 Technical Manual 5
Page 6
Contents of package
The lead and accessories are provided sterile. Each package contains the following:
1 lead with 1 radiopaque anchoring sleeve1, stylet, and Quick Twist tool
1 vein lifter
1 slit anchoring sleeve
1 white fixation tool
2 pin caps
extra stylets
product literature
Accessory descriptions
Stylet – A stylet provides additional stiffness and controlled flexibility
for maneuvering the lead into position. Each stylet knob is labeled with the stylet diameter and length.
Anchoring sleeve – An anchoring sleeve secures the lead from moving and protects the lead insulation and conductors from damage caused by tight sutures.
Slit anchoring sleeve – A slit anchoring sleeve secures excess lead length in the device pocket.
Pin cap – A pin cap covers and insulates unused connector pins.
Vein lifter – A vein lifter facilitates lead insertion into a vessel.
White fixation tool – The white fixation tool facilitates connector
pin rotation.
Quick Twist tool – The Quick Twist tool facilitates both connector pin rotation and stylet insertion into the lead. This tool comes assembled on the lead.

Indications

The lead is intended for single, long-term use in the right ventricle.
This lead has application for patients for whom implantable cardioverter defibrillators are indicated.
1
Two radiopaque anchoring sleeves are provided with leads 85 cm or longer.
6 6931 Technical Manual
Page 7

Contraindications

Atrial use – The lead is contraindicated for the sole use of detection
and treatment of atrial arrhythmias.
Ventricular use – The lead is contraindicated for ventricular use in patients with tricuspid valvular disease or a tricuspid mechanical heart valve.
Transient ventricular tachyarrhythmias – The lead is contraindicated for patients with transient ventricular tachyarrhythmias due to reversible causes (drug intoxication, electrolyte imbalance, sepsis, hypoxia) or other factors (myocardial infarction, electric shock).
Steroid use – The lead is contraindicated in patients for whom a single dose of 1.0 mg of dexamethasone acetate may be contraindicated.

Warnings and precautions

For single use only – Do not resterilize and reimplant an
explanted lead.
Inspecting the sterile package – Inspect the package prior to opening.
If the seal or package is damaged, contact your local Medtronic representative.
Do not use the product after its expiration date.
Ethylene oxide resterilization – The lead has been sterilized with ethylene oxide prior to shipment. If the integrity of the sterile package has been compromised prior to the expiration date, resterilize using ethylene oxide. Avoid resterilization techniques that could damage the lead.
Refer to sterilizer instructions for operating instructions.
Use an acceptable method for determining sterilizer effectiveness, such as biological indicators.
Do not exceed temperatures of 55°C (131°F).
Do not resterilize more than 1 time.
After resterilization, allow the device to aerate ethylene oxide residues.
Electrophysiologic testing – Prior to lead implant, it is strongly recommended that patients undergo a complete cardiac evaluation, which should include electrophysiologic testing. Also, electrophysiologic evaluation and testing of the safety and efficacy of the proposed pacing, cardioversion, or defibrillation therapies are recommended during and after the implant of the system.
6931 Technical Manual 7
Page 8
Steroid use – It has not been determined whether the warnings, precautions, or complications usually associated with injectable dexamethasone acetate apply to the use of this highly localized, controlled-release device. For listing of potentially adverse effects, refer to the Physician’s Desk Reference.
Steroid tip – Reducing the available amount of steroid may adversely affect low-threshold performance. Avoid reducing the amount of steroid available prior to lead implant.
Do not allow the electrode surface to come in contact with surface contaminants.
Do not wipe or immerse the electrode in fluid, except blood, at the time of implant.
Handling the lead – Handle the lead with great care at all times.
Protect the lead from materials shedding particles such as lint and dust. Lead insulators attract these particles.
Handle the lead with sterile surgical gloves that have been rinsed in sterile water or a comparable substance.
Do not severely bend, kink, or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pins.
Do not immerse leads in mineral oil, silicone oil, or any other liquid, except blood, at the time of implant.
Do not implant the lead without first verifying the mechanical functioning of the helix electrode. Refer to the section “Verifying the mechanical functioning of the helix electrode” on page 17, for complete instructions.
Do not exceed the recommended maximum number of rotations to extend or retract the helix electrode. Exceeding the maximum number may result in fracture or distortion of the inner conductor or helix electrode. The number of rotations required to fully extend or retract the helix electrode is variable; refer to the section “Specifications (nominal)” on page 27, for the recommended maximum number of rotations.
Inserting the lead using a lead introducer that features a hemostasis valve may require a larger introducer than the size recommended. To avoid distortion of the coil electrode, do not withdraw the lead through a hemostasis valve.
Handling the stylets – Use care when handling stylets.
Do not use excessive force or surgical instruments when inserting a stylet.
Avoid overbending and kinking.
Use a new stylet when blood or other fluids accumulate on the stylet. Accumulated fluids may cause lead damage or difficulty in passing the stylet through the lead.
Do not use a sharp object to impart a curve to the distal end of the stylet.
8 6931 Technical Manual
Page 9
Necessary hospital equipment – Keep external defibrillation equipment nearby for immediate use during the acute lead system testing, implant procedure, or whenever arrhythmias are possible or intentionally induced during post-implant testing.
Line-powered equipment – An implanted lead forms a direct current path to the myocardium. During lead implant and testing, use only battery-powered equipment or line-powered equipment specifically designed for this purpose, to protect against fibrillation that may be caused by alternating currents. Line-powered equipment used in the vicinity of the patient must be properly grounded. Lead connector pins must be insulated from any leakage currents that may arise from line-powered equipment.
Second anchoring sleeve – Leads 85 cm or longer feature 2 anchoring sleeves. Use both anchoring sleeves to assure adequate fixation, see the section “Anchoring the lead” on page 24.
Concurrent devices – Output pulses, especially from unipolar devices, may adversely affect device sensing capabilities. If a patient requires a separate stimulation device, either permanent or temporary, allow enough space between the leads of the separate systems to avoid interference in the sensing capabilities of the devices. Previously implanted pulse generators, implantable cardioverter defibrillators, and leads should generally be explanted. Refer to “Chronic repositioning or removal” on page 9, for further information on explanting leads.
Diathermy – People with metal implants such as pacemakers, implantable cardioverter defibrillators (ICDs), and accompanying leads should not receive diathermy treatment. The interaction between the implant and diathermy can cause tissue damage, fibrillation, or damage to the device components, which could result in serious injury, loss of therapy, or the need to reprogram or replace the device.
Chronic repositioning or removal – Chronic repositioning or removal of leads may be difficult because of fibrotic tissue development. Return all removed leads, or lead segments, to Medtronic. If a lead must be removed or repositioned, proceed with extreme caution.
Lead removal may result in avulsion of the endocardium, valve, or vein.
Lead junctions may separate, leaving the lead tip and bare wire in the heart or vein.
Chronic repositioning may adversely affect the low-threshold performance of a steroid eluting lead.
Cap abandoned leads to avoid transmitting electrical signals.
For leads that have been severed, seal the remaining lead end and suture the lead to adjacent tissue.
6931 Technical Manual 9
Page 10
If a helix electrode does not disengage from the endocardium by rotating the connector pin, rotating the lead body counterclockwise may withdraw the helix electrode and decrease the possibility of damage to the cardiovascular structures during removal.
Connector compatibility – Although the lead conforms to the International Connector Standards IS-1 and DF-1, do not attempt to use the lead with any device other than a commercially available implantable defibrillator system with which it has been tested and demonstrated to be safe and effective. The potential adverse consequences of using such a combination may include, but are not limited to, undersensing cardiac activity and failure to deliver necessary therapy.

Potential adverse events

The potential adverse events related to the use of transvenous leads include, but are not limited to, the following patient-related conditions:
cardiac perforation
cardiac tamponade
constrictive pericarditis
embolism
endocarditis
fibrillation or other arrhythmias
heart wall rupture
hemothorax
infection
pneumothorax
thrombosis
tissue necrosis
Other potential adverse events related to the lead include, but are not limited to, the following:
Insulation failure
Lead conductor or electrode fracture
Lead dislodgment
Poor connection to the device, which may lead to oversensing, undersensing, or a loss of therapy
10 6931 Technical Manual
Page 11

Clinical study

Clinical data was not collected in the approval process for this lead. Clinical data from the Models 6947 and 6932 leads support the safety and efficacy of the Model 6931 lead.
The Model 6931 lead is a downsized version of the Medtronic Sprint Quattro Secure Model 6947 lead without an SVC defibrillation coil (same as the Sprint Model 6932 lead). The Model 6931 lead includes a combination of components used in currently marketed Medtronic leads with some enhancements. All functional features of the Model 6931 lead have been approved in currently marketed Medtronic leads. Previous clinical studies have demonstrated titanium nitride coated electrodes (6931) do not significantly change clinical pacing thresholds or sensing amplitudes as compared to platinized electrodes (6947). The overall surface of the defibrillation electrode falls within the range of currently approved Sprint leads.
Based upon its similarity to the Model 6947 and 6932 leads, the clinical data from these lead models supports the safety and efficacy of the Model 6931 lead (Table 1).
Table 1.
Sprint Fidelis Model 6931 features:
Silicone insulation with polyurethane overlay
Ring electrode for true bipolar sensing
Steroid eluting Steroid eluting
Active fixation extendable/ retractable helix
a
Standard
Sprint Fidelis Model 6931 features:
Single RV defibrillation coil Single RV defibrillation coil
a
electrode spacing
8 mm tip to ring spacing, 12 mm tip to RV coil spacing
Sprint Quattro Secure Model 6947 clinical data supports:
Silicone insulation with polyurethane overlay
Ring electrode for true bipolar sensing
Active fixation extendable/ retractable helix
a
Standard
Sprint Model 6932 clinical data supports:
electrode spacing
Model 6947 clinical study
The Model 6947 clinical study was a prospective, nonrandomized, multicenter trial assessing the lead handling and performance of the Sprint Quattro
The Model 6947 lead was implanted in 80 patients at 15 investigative centers in the United States and at 2 investigative centers in Canada between April 27, 2001 and August 2, 2001.
Secure Model 6947 Lead.
6931 Technical Manual 11
Page 12
Patients were included in this study if they met the following criteria:
1) able to receive a pectoral implant and 2) survived at least one episode of cardiac arrest due to a ventricular tachyarrhythmia; or had poorly tolerated, sustained ventricular tachycardia that occurred spontaneously; or had poorly tolerated, sustained ventricular tachycardia that could be induced.
Patients studied
The Model 6947 study population consisted of 63 males and 17 females. The mean age was 64.4 years. The most frequently reported indication for implant was inducible ventricular tachycardia without sudden cardiac death (SCD) (47.5%). The mean ejection fraction was 35.5%. The most frequently reported NYHA classifications were Class I (35.0%) and Class II (36.3%).
Cardiovascular history included coronary artery disease with myocardial infarction (62.5%), hypertension (58.8%), cardiomyopathy (65.0%), congestive heart failure (47.5%), syncope/ presyncope (46.3%) and previous cardiac surgery (68.8%).
Objectives
The objectives of the study were to report the following:
Pacing thresholds
R-wave amplitudes
Pacing impedance
Lead handling
Adverse events
Methods
Pulse width thresholds, R-wave amplitudes, and pacing impedances were measured at implant and at one month post-implant. Adverse events were collected throughout the study. A lead handling questionnaire was completed by the implanting physician at each implant. The implanting physicians evaluated the performance of the Model 6947 lead with regard to:
Ease of lead insertion into the vein
Ease of helix extension
Visibility of helix extension
Steerability
Torqueability
Lead placement time
Ability to traverse the tricuspid annulus
Comfort level with handling the lead
12 6931 Technical Manual
Slipperiness of lead surfaces
Stiffness of the lead
Ease of obtaining adequate R-wave sensing
Ease of obtaining adequate VF sensing
Ease of obtaining adequate DFTs
Overall ease of lead placement
Page 13
Results
The mean follow-up duration was 0.96 months (range: 0.00 - 1.71 months) with a cumulative follow-up duration of 76.82 months. The pace/sense measurements are summarized in Table 2.
Tab l e 2 . Pace/Sense measurements
Implant One month
Pulse-width threshold at one volt
N8061
Median (ms) 0.20 0.20
th
- 75th Percentile (0.20 - 0.20) (0.20 - 0.30)
25
Range (0.03 - 0.60) (0.20 - 0.80)
R-wave amplitude (EGM)
N7864
Median (mV) 8.0 9.3
th
- 75th Percentile (7.0 - 11.0) (7.5 - 12.8)
25
Range (3.0 - 24.0) (3.5 - 20.0)
Pacing lead impedance
N8066
Median (ohms) 564.0 481.0
th
- 75th Percentile (481.0 - 611.0) (444.0 - 521.0)
25
Range (378.0 - 985.0) (323.0 - 985.0)
On the lead handling survey, for all items rated, the adjusted rating of the Model 6947 lead fell between 1 (very good) and 2 (excellent). The adjusted rating for each item is an average across physicians, accounting for multiple responses per physician.
The overall ease of lead placement was considered good, very good, or excellent by all implanting physicians, with an adjusted rating of 1.7 (the minimum rating on the questionnaire was - 2.0 and the maximum rating was 2.0).
Observed adverse events
The Sprint Quattro Secure Model 6947 Lead was utilized in a prospective, nonrandomized, multicenter trial to assess the handling and performance of the Model 6947 lead.
A total of 22 cardiovascular related adverse events were reported. Two of the 22 events were ventricular lead-related and both occurred at implant. One event was a microdislodgement and one was induced VT as a result of lead manipulation.
Two deaths occurred in this patient group during the follow-up period. Both deaths were classified as non-sudden cardiac and were judged to be non-system related by an independent advisory committee.
6931 Technical Manual 13
Page 14
Model 6947 conclusion
In this clinical study, the Model 6947 lead demonstrated acceptable clinical performance. Through questionnaire responses, implanting physicians verified acceptable overall lead performance and handling during the implant procedure.
Model 6932 (single-coil) clinical study
The following is a summary of the clinical results from the 6932 RV lead study with respect to the primary study objectives and additional performance measurements with Active Can ICDs. All of the results presented are based on an “as randomized” analysis.
Patient characteristics
The results summarize data for 336 patients, who were randomized (1:1) to either the Model 6932 lead (n=168) or the Model 6936 lead (n=168) between 6/9/95 and 3/22/96. Commercially available Medtronic ICDs were the devices used in the study. The mean follow-up was 2.39 ± 1.96 months with a cumulative follow-up of
803.4 months.
Primary endpoints
Patient characteristics by randomized lead
6932
N=168
Gender
% Male 133 (79.2%) 140 (83.3%)
Age at implant
Mean (years) Standard deviation (years)
Primary indication
SCD only VT only SCD & VT Other
Primary cardiovascular disease
CAD or MI Cardiomyopathy only Primary electrical disease Other
NYHA classification
Class I/II 142 (84.5%) 137 (81.5%)
Ejection fraction
Mean Standard deviation
59.4
±12.9
49 (29.2%) 88 (52.4%) 20 (11.9%)
11 (6.6%)
114 (67.9%)
30 (17.9%)
7 (4.2%)
17 (10.1%)
34.8&
±13.0%
6936
N=168
62.1
±11.4
45 (26.8%) 86 (51.2%) 28 (16.7%)
9 (5.4%)
117 (69.6%)
33 (19.6%)
7 (4.2%)
11 (6.6%)
34.9%
±13.4%
14 6931 Technical Manual
Page 15
Meeting defibrillation implant criterion
Meeting defibrillation implant criterion with the initial lead configuration was defined as having at least one successful defibrillation at 22 joules in the first three VF inductions (following a binary search protocol), or failing that, two successful defibrillations in the next two VF inductions at 24 joules. The initial lead configuration was defined as the initial system tested, allowing no polarity or position changes. The rate of meeting defibrillation implant criterion with the initial configuration for the 6932 lead was equivalent to the 6936 lead.
Meeting defibrillation implant criterion with
the initial lead configuration by randomized lead
6932 6936
Observed success rate
Adjusted success rate
b
p-value
a
Including only those patients in whom the implant testing protocol was fully completed
b
Adjusted for age, sex, ejection fraction, NYHA classification and primary indication for implant using logistic regression
a
b
99% (143/145) 96% (137/142)
100% 99%
0.077
Pacing thresholds through three months
A pulse-width threshold was measured at 2.8 V at implant, pre-hospital discharge evaluation (PDE), one month and three months. The pulse-width threshold for the 6932 lead was significantly lower than for the 6936 lead at each time point (p=0.0001).
Complication-free survival through three months
Pacing thresholds through 3 months by randomized lead
Pulse width (in msec) @ 2.8 V
Lead Implant PDE 1 month 3 month
6932 N
Adjusted mean
6936 N
Adjusted mean
a
p-value
a
Adjusted for age, sex, ejection fraction and NYHA classification using mixed effects regression
158
a
0.043
150
a
0.069
0.0001 0.0001 0.0001 0.0001
152
0.049
142
0.109
116
0.057
117
0.205
77
0.063
80
0.198
A complication was defined as an adverse event which required invasive intervention. The complication-free survival through three months for the 6932 lead was equivalent to the 6936 lead (p=0.87).
6931 Technical Manual 15
Page 16
Additional performance measurements with active can ICDs
Implant success
Implant success was achieved if the initial system tested was the system that was implanted (including polarity and position changes). For the 6932 lead, implant success with a single lead-Active Can system was 98% (114/116 patients).
Defibrillation threshold (DFT)
The DFT with the initial lead system was determined following a binary search protocol (starting with a 12 J shock). For the 6932 lead, the mean DFT with a single lead-Active Can system was 8.7 J.
Model 6932 conclusions
All of the stated objectives were met in this study. These findings demonstrate that the Model 6932 lead performance is equivalent to, or better than the Model 6936 lead performance, and that the 6932 lead is safe and effective for human use.

Directions for use

Proper surgical procedures and sterile techniques are the responsibility of the medical professional. The following procedures are provided for information only. Each physician must apply the information in these instructions according to professional medical training and experience.
The implant procedure generally includes the following steps:
Opening the package
Verifying the mechanical functioning of the helix electrode
Inserting the lead
Positioning the lead
Securing the helix electrode into the endocardium
Taking electrical measurements and defibrillation efficacy measurements
Anchoring the lead
Connecting the lead
Placing the device and leads into the pocket
Post-implant evaluation
Opening the package
Use the following steps to open the sterile package and inspect the lead:
1. Within the sterile field, open the sterile package and remove the lead and accessories.
2. Inspect the lead. Leads shorter than 85 cm should have 1 anchoring sleeve on the lead body. Leads 85 cm or longer should have 2 anchoring sleeves on the lead body.
16 6931 Technical Manual
Page 17
Verifying the mechanical functioning of the helix electrode
Note: The package includes 2 tools, the Quick Twist tool assembled
on the lead and the white fixation tool. Either tool may be used to verify the mechanical functioning of the helix electrode. The choice of tool is left to the discretion of the physician.
Before implant, verify the mechanical functioning of the helix electrode using the following steps:
1. Attach either the Quick Twist tool or the white fixation tool to the lead. Ensure that the stylet is inserted into the lead and proceed as indicated, according to the tool being used.
a. Quick Twist tool: Push the Quick Twist tool onto the IS-1
connector pin (Figure 1).
Figure 1.
b. White fixation tool: Press both legs of the white fixation
tool together and place the most distal hole on the IS-1 connector pin (Figure 2).
Figure 2.
2. Hold the IS-1 connector leg of the lead with the thumb on one side and 4 fingers on the other side. Keep the lead body and the IS-1 connector leg as straight as possible (Figure 2). Ensure that the stylet is fully inserted, then rotate the selected fixation tool clockwise until the helix electrode is fully extended (Figure 3a or Figure 3b). When the helix electrode is fully extended, approximately 1.5 to 2 helix coils are exposed.
6931 Technical Manual 17
Page 18
a
b
Figure 3.
Caution: Do not severely bend the IS-1 connector leg or the lead body while extending the helix electrode (Figure 4). If the lead is bent on either side of the lead bifurcation during helix electrode extension or retraction, the lead may be damaged.
Figure 4.
Caution: Overrotating the connector pin after the helix electrode is fully extended may damage the lead.
Note: To determine the number of rotations applied to the lead, count the number of rotations of the white flap on the Quick Twist tool. The number of necessary rotations is equivalent to the number required for the white fixation tool. See “Specifications (nominal)” on page 27 for the maximum number of rotations to extend or retract the helix electrode.
18 6931 Technical Manual
Page 19
The number of rotations required to extend the helix electrode increases proportionately with the length of the lead. Additional curvatures made to the stylet may increase the number of rotations needed to extend or retract the helix electrode. Refer to the section “Specifications (nominal)” on page 27, for the maximum number of rotations to extend or retract the helix electrode.
During the initial helix electrode extension, the helix electrode may extend suddenly due to accumulated torque in the lead, or the helix electrode may require additional turns for extension.
3. Disconnect the selected fixation tool from the connector pin and release the proximal end of the lead body. Allow several seconds for relief of the residual torque in the lead.
4. After allowing for relief of the residual torque, reattach the selected fixation tool and rotate it counterclockwise until the helix electrode tip is retracted into the sheath.
Inserting the lead
Caution: Use care when handling the lead during insertion.
Do not severely bend, kink, or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pins.
Insert the lead using the following techniques:
1. Select a site for lead insertion. The lead may be inserted by venotomy through several different venous routes, including the right or left cephalic vein or the external or internal jugular vein. Use the cephalic vein whenever possible to avoid lead damage in the first rib or clavicular (thoracic inlet) space.
Caution: Certain anatomical abnormalities, such as thoracic outlet syndrome, may also precipitate pinching and subsequent fracture of the lead.
Caution: When using a subclavian approach, avoid techniques that may damage the lead.
Place the insertion site as far lateral as possible to avoid clamping the lead body between the clavicle and the first rib (Figure 5).
Figure 5.
Do not force the lead if significant resistance is encountered during lead passage.
6931 Technical Manual 19
Page 20
Do not use techniques such as adjusting the patient’s posture to facilitate lead passage. If resistance is encountered, it is recommended that an alternate venous entry site be used.
2. Insert the tapered end of a vein lifter into the incised vein and gently push the lead tip underneath and into the vein (Figure 6).
Note: A percutaneous lead introducer (PLI) kit may be used to facilitate insertion. Refer to the technical manual packaged with an appropriate percutaneous introducer for further instructions.
Figure 6.
3. Advance the lead into the right atrium using a straight stylet to facilitate movement through the veins.
Positioning the lead
Caution: Use care when handling the lead during positioning.
Do not severely bend, kink or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pins.
Use the following steps to position the lead:
1. After the lead tip is passed into the atrium, advance the lead through the tricuspid valve. Replacing the straight stylet with a gently curved stylet may add control in maneuvering the lead through the tricuspid valve.
Caution: Do not use a sharp object to impart a curve to the distal end of the stylet. Imparting a curve to the stylet can be accomplished with a smooth-surface, sterile instrument (Figure 7).
Figure 7.
20 6931 Technical Manual
Page 21
Note: Passing the lead tip through the tricuspid valve or chordae tendineae may be difficult due to the flexible nature of the lead body. Rotating the lead body as the tip passes through the valve may facilitate passage.
2. After the lead tip is in the ventricle, the curved stylet may be replaced with a straight stylet. Withdraw the stylet slightly, to avoid using excessive tip force while achieving final electrode position. Avoid known infarcted or thin wall areas, to minimize the occurrence of perforation.
3. Proper positioning of the helix electrode is essential for stable endocardial pacing. A satisfactory position usually is achieved when the lead tip points straight toward the apex, or when the distal end dips or bends slightly. Use fluoroscopy (lateral position) to ensure that the tip is not in a retrograde position or lodged in the coronary sinus.
Note: With the helix electrode retracted, the distal end of the lead may be used to map a desirable site for electrode fixation. Mapping may reduce the need to repeatedly extend and fixate the helix electrode.
4. After placing the lead in a satisfactory position, extend the helix electrode by following the procedure in the section “Securing the helix electrode into the endocardium” on page 21.
Securing the helix electrode into the endocardium
Note: The package includes 2 tools, the Quick Twist tool assembled
on the lead and the white fixation tool. Either tool may be used to secure the helix electrode into the endocardium. The choice of tool is left to the discretion of the physician.
Secure the helix electrode using the following techniques:
1. Attach either the Quick Twist tool or the white fixation tool to the lead. Ensure that the stylet is inserted into the lead and proceed as indicated, according to the tool being used.
a. Quick Twist tool: Push the Quick Twist tool onto the IS-1
connector pin. (Figure 1).
b. White fixation tool: Press both legs of the white fixation
tool together and place the most distal hole on the IS-1 connector pin (Figure 2).
2. Press the lead tip against the endocardium by gently pushing the stylet and lead at the vein entry site.
3. Rotate the selected fixation tool clockwise until the helix electrode is fully extended (see Figure 3a or Figure 3b).
Caution: Do not severely bend the IS-1 connector leg or the lead body while extending the helix electrode (Figure 4). If the lead is bent on either side of the lead bifurcation during helix electrode extension or retraction, the lead may be damaged.
Use fluoroscopy to verify electrode extension (Figure 8). The fluoroscope head may need to be rotated to obtain an adequate view.
6931 Technical Manual 21
Page 22
Closing of the space between the indicator stop (A) and the indicator ring (B) implies complete extension of the helix electrode.
VISUAL
VISUAL
Fully Retracted
FLUOROSCOPIC
Fully Extended
FLUOROSCOPIC
Figure 8.
A B
AB
Caution: Exceeding the maximum number of rotations may damage the lead. The maximum number of rotations required to fully extend or retract the helix electrode is variable; refer to the section, “Specifications (nominal)” on page 27, for the maximum number of rotations.
Caution: Prolonged implant procedures or multiple repositionings may allow blood or body fluids to build up on the helix electrode mechanism. This may result in an increased number of rotations required to extend or retract the helix electrode.
4. Remove the selected fixation tool from the IS-1 connector pin and release the proximal end of the lead body. Allow several seconds for relief of the residual torque in the lead.
5. To assure helix electrode fixation, leave the stylet in place, hold the lead by the connector, and carefully rotate the lead body in 2 clockwise rotations.
6. Partially withdraw the stylet.
7. Obtain electrical measurements to verify satisfactory placement and electrode fixation. Refer to the section “Taking electrical measurements and defibrillation efficacy measurements” on page 23.
8. Verify that the lead is affixed. Gently pull back on the lead and check for resistance to verify fixation. A properly affixed helix electrode will remain in position. If the helix electrode is not properly affixed, the lead tip may become loose in the right ventricle.
9. If repositioning is required, reattach the selected fixation tool and rotate counterclockwise until the helix electrode is retracted. Use fluoroscopy to verify withdrawal of the helix electrode before attempting to reposition.
10. After final positioning, remove the stylet and the Quick Twist tool completely. When removing the Quick Twist tool, grip the lead firmly just below the connector pin to help prevent lead dislodgment.
22 6931 Technical Manual
Page 23
11. Obtain final electrical measurements. Refer to the section “Taking electrical measurements and defibrillation efficacy measurements” on page 23.
Taking electrical measurements and defibrillation efficacy measurements
Caution: Prior to taking electrical or defibrillation efficacy
measurements, move objects made from conductive materials, such as guide wires, away from all electrodes. Metal objects, such as guide wires, can short a lead and an active implantable device, causing electrical current to bypass the heart and possibly damage the implantable device and lead.
Use the following steps to take electrical measurements:
1. Ensure that the Quick Twist tool is disconnected from the IS-1 connector pin.
2. Attach a surgical cable to the lead connector pin.
3. Use a testing device, such as a pacing system analyzer, for obtaining electrical measurements. For information on the use of the testing device, consult the product literature for that device.
In order to demonstrate reliable defibrillation efficacy, obtain final defibrillation measurements for the lead system.
Tab l e 3 . Recommended measurements at implant
(when using a pacing system analyzer)
Measurements required Acute
Capture threshold (at 0.5 ms pulse width) 1.0 V 3.0 V
Pacing impedance 200 - 1000 ohms 200 - 1000 ohms
Filtered R-wave amplitude (during sinus rhythm)
Slew rate
a
< 30 days after implant.
b
> 30 days after implant.
a
lead system
5mV ≥ 3mV
0.75 V/s 0.45 V/s
Chronicb lead system
If initial electrical measurements deviate from the recommended values, it may be necessary to repeat the testing procedure 15 minutes after final positioning. Initial electrical measurements may deviate from the recommended values:
Initial impedance values may exceed the measuring capabilities of the testing device, resulting in an error message.
Values may vary depending upon lead type, implantable device settings, cardiac tissue condition, and drug interactions.
If electrical measurements do not stabilize to acceptable levels, it may be necessary to reposition the lead and repeat the testing procedure.
6931 Technical Manual 23
Page 24
In order to keep patient morbidity and mortality to a minimum, patients should be rescued promptly with an external defibrillator if the implanted lead system fails to terminate a VF episode. At least 5 minutes should elapse between VF inductions.
For more information on obtaining electrical measurements, consult the technical manual supplied with the testing device.
Anchoring the lead
Caution: Use care when anchoring the lead.
Use only nonabsorbable sutures to anchor the lead.
Do not attempt to remove or cut the anchoring sleeve from the lead body.
During lead anchoring, take care to avoid dislodging the lead tip.
Do not secure sutures so tightly that they damage the vein, lead, or anchoring sleeve (Figure 9).
Do not tie a suture directly to the lead body (Figure 9).
Figure 9.
Use the following steps to anchor the lead using all 3 grooves:
Note: The anchoring sleeves contain a radiopaque substance, which allows visualization of the anchoring sleeve on a standard x-ray and may aid in follow-up examinations.
1. Position the distal anchoring sleeve against or near the vein.
2. Secure the anchoring sleeve to the lead body by tying a suture firmly in each of the 3 grooves (Figure 10).
Figure 10.
3. Use at least 1 additional suture in 1 of the grooves to secure the anchoring sleeve and lead body to the fascia.
24 6931 Technical Manual
Page 25
4. A second anchoring sleeve is provided with leads 85 cm or longer. For abdominal implants, redundant lead body (for example, a curve for strain relief) should be placed just proximal to the first anchoring sleeve. Then, the second anchoring sleeve may be lightly sutured to the lead body and fascia to hold the curve in place. This procedure helps isolate the vein entry site from tension on the proximal end of the lead body.
5. A slit anchoring sleeve may be used in the device pocket to secure excess lead length. First, secure the anchoring sleeve to the lead body. Then, orient the slit toward the fascia and secure the anchoring sleeve to the fascia with sutures.
Connecting the lead
Use the following steps to connect the lead to an implantable device:
1. Carefully remove the stylet and Quick Twist tool. When removing the stylet and Quick Twist tool, grip the lead firmly just below the connector pin, to prevent dislodgment.
2. Insert the lead connectors into the connector block. Consult the product literature packaged with the implantable device for instructions on proper lead connections.
Placing the device and leads into the pocket
Caution: Use care when placing the device and leads into
the pocket.
Ensure that the leads do not leave the device at an acute angle.
Do not grip the lead or device with surgical instruments.
Do not coil the lead. Coiling the lead can twist the lead body and may result in lead dislodgment (Figure 11).
Figure 11.
Use the following steps to place the device and leads into the pocket:
1. To prevent undesirable twisting of the lead body, rotate the device to loosely wrap the excess lead length (Figure 12).
6931 Technical Manual 25
Page 26
Figure 12.
2. Insert the device and leads into the pocket.
3. Before closing the pocket, verify sensing, pacing, cardioversion, and defibrillation efficacy.
Post-implant evaluation
After implant, monitor the patient’s electrocardiogram until the patient is discharged. If a lead dislodges, it usually occurs during the immediate postoperative period.
Recommendations for verifying proper lead positioning include x-rays and pacing/sensing thresholds taken at pre-hospital discharge, 3 months after implant, and every 6 months thereafter.
In the event of a patient death, explant all implanted leads and devices and return them to Medtronic with a completed Product Information Report form. Call the appropriate phone number on the back cover if there are any questions on product handling procedures.
26 6931 Technical Manual
Page 27

Detailed device description

Specifications (nominal)
Parameter Model 6931
Ty pe Tr ip o la r
Position Right ventricle
Fixation Extendable/Retractable helix
Length 40-110 cm
Connectors Unipolar: DF-1
Bipolar: IS-1
Materials Conductors: MP35N coil
Insulation: Silicone, ETFE
Electrodes (pace, sense): Titanium nitride coated
DF-1 pin: Stainless steel
IS-1 pin and ring: Stainless steel
Steroid Type: Dexamethasone acetate
Steroid binder: Silicone
Conductor resistances Pacing (unipolar): 21.6 (65 cm)
Pacing (bipolar): 68.6 (65 cm)
Defibrillation: <2.4 (65 cm)
Helix length (extended) 1.8 mm
Diameters Lead body: 2.2 mm
Lead introducer (recommended size)
without guide wire: 7.0 French
with guide wire: 9.0 French
MP35N composite cables
Overlay: Polyurethane
platinum iridium
RV coil: Platinum-clad tantalum
Amount: 1.0 mg maximum
Tip: 2.2 mm
Helix: 1.2 mm
Maximum number of rotations to extend or retract the helix electrode
Lead length Number of rotations
52 cm 16
58 cm 18
65 cm 20
75 cm 22
100 cm 27
6931 Technical Manual 27
Page 28
Specifications drawing (nominal)
12 mm
8 mm
Helix electrode Surface area: 4.2 mm
Ring electrode
Surface area: 20.2 mm
2
2
RV coil electrode
Length: 62 mm Surface area: 513 mm
2
Electrical shadow area: 430 mm
Anchoring sleeve
Note: Leads 85 cm or
longer have 2 anchoring sleeves
2
DF-1 connector (red band) Note: Connector
pin is common to RV coil electrode
28 6931 Technical Manual
IS-1 BI connector Note: Connector pin is common
to tip electrode; connector ring is common to ring electrode
Page 29

Medtronic warranty

For complete warranty information, see the accompanying warranty document.

Service

Medtronic employs highly trained representatives and engineers located throughout the world to serve you and, upon request, to provide training to qualified hospital personnel in the use of Medtronic products. Medtronic also maintains a professional staff to provide technical consultation to product users. For medical consultation, Medtronic can often refer product users to outside medical consultants with appropriate expertise. For more information, contact your local Medtronic representative, or call or write Medtronic at the appropriate address or telephone number listed on the back cover.
6931 Technical Manual 29
Page 30
Page 31
Page 32
World Headquarters
Medtronic, Inc. 710 Medtronic Parkway Minneapolis, MN 55432-5604 USA Internet: www.medtronic.com Tel. 763-514-4000 Fax 763-514-4879
Medtronic USA, Inc.
Toll-free in the USA (24-hour technical consultation for physicians and medical professionals) Bradycardia: 1-800-505-4636 Tachycardia: 1-800-723-4636 Heart Failure: 1-800-505-4636
Europe/Africa/Middle East Headquarters
Medtronic Europe Sàrl Route du Molliau 31 Case Postale CH-1131 Tolochenaz Switzerland Internet: www.medtronic.co.uk Tel. 41-21-802-7000 Fax 41-21-802-7900
Medtronic E.C. Authorized Representative/Distributed by
Medtronic B.V. Earl Bakkenstraat 10 6422 PJ Heerlen The Netherlands Tel. 31-45-566-8000 Fax 31-45-566-8668
Asia
Medtronic International Ltd. Suite 1602 16/F, Manulife Plaza The Lee Gardens, 33 Hysan Avenue Causeway Bay Hong Kong Tel. 852-2891-4068 Fax 852-2591-0313
Aust ralia
Medtronic Australasia Pty. Ltd. Unit 4/446 Victoria Road Gladesville NSW 2111 Australia Tel. 61-2-9879-5999 Fax 61-2-9879-5100
Canada
Medtronic of Canada Ltd. 6733 Kitimat Road Mississauga, Ontario L5N 1W3 Tel. 905-826-6020 Fax 905-826-6620 Toll-free in Canada: 1-800-268-5346
Japan
Medtronic Japan Solid Square West Tower 6F, 580 Horikawa-cho, Saiwai-ku, Kawasaki, Kanagawa 210-0913 Japan Tel. 81-44-540-6112 Fax 81-44-540-6200
*A09204001*
© Medtronic, Inc. 2004 All Rights Reserved
A09204001 2004-09-14
Loading...