Medtronic 68000 Instructions for Use

CARDIOBLATE® 68000
SURGICAL ABLATION SYSTEM
Technical Manual
Explanation of symbols
European Conformity. This symbol means that the device fully complies with European Council Directive 93/42/EEC.
Attention, See Instructions for Use
Dangerous Voltage
Reorder Number
Serial Number
Manufacturer
Temperature Limitation
Nonsterile
Content
Atmospheric Pressure Limitation
Humidity Limitation
Date of Manufacture
Type CF Applied Part
Neutral Electrode Isolated from Earth High Frequency
Dispersive Electrode Pad
Electrogram
Monopolar
Bipolar
Danger: Risk of explosion if used in the presence of flammable anesthetics
With Respect To Electric Shock Fire And Mechanical Hazards Only In Accordance With UL60601-1/CAN/CSA C22.2 No. 601.1
Cardioblate® 68000 Technical Manual i
ETL Listing (US/CAN) This device meets the requirement of UL60601, CAN/CSA C22.2 No. 601-1M90 and IEC60601-2-2
Non-ionizing Radiation
Fuse Marking: Replace Fuses as Marked
Do not dispose of this product in the unsorted municipal waste stream. Dispose of this product according to local regulations. See http://recycling.medtronic.com for instructions on proper disposal of this product.
Bipolar Device
Monopolar Device
Overall RF Energy Delivery Time
Transmural
Remote Control Foot Switch
Graph Mode
Volume Level
Impedance
Wattage
Ohms, Unit of Impedance Measurement
RF Energy Delivery Button
Active Ablation Time
Universal Serial Bus (USB) Port Connector
Error Message Explanation
Watts, Unit of Energy Measurement
Maximum Limit
Minimum Limit
ii Cardioblate® 68000 Technical Manual
OFF (AC Power)
ON (AC Power)
Warning or Error Message
For US Audiences Only
China RoHS Standard (SJ/T11364-2006) Electronic Information Products Pollution Control Symbol. The number represents the years the device can be used before it must be recycled (environmental protection use period).
Authorized Representative in the European Community
Cardioblate® 68000 Technical Manual iii
iv Cardioblate® 68000 Technical Manual
Table of Contents
How to Use This Manual vii
Changing the Language Option vii

1 Overview 1-1

System Description 1-2
Indications for Use 1-4
Contraindications 1-5
Warnings and Precautions 1-5
Adverse Events 1-10

2 Generator Description 2-1

Components 2-2
Displays, Connectors, and Controls 2-3
Audible Tones 2-9
Specifications 2-9
Output Power Diagrams 2-12
Electromagnetic Emissions and Immunity Declaration (EN60601-1-2) 2-14
Applicable Standards 2-18

3 Generator Operation 3-1

Quick Reference Guide 3-2
Selecting and Attaching a Dispersive Electrode 3-3
Preparing and Powering Up the Generator 3-4
Connecting a Cardioblate
Verifying Generator Recognition of Attached Devices 3-8
Disconnecting the Cardioblate
Setting the RF Energy Mode and Parameters 3-9
Monopolar Mode Selection and Use 3-9
Bipolar Mode Selection and Use 3-14
Options 3-17
Powering Down the Generator 3-23
®
Ablation Device 3-6
®
Ablation Devices 3-9

4 Safety Shutdown and Troubleshooting 4-1

Safety Shutdown Conditions 4-2
Safety Shutdown Messages 4-3
Cardioblate® 68000 Technical Manual v
Other Messages 4-7
System Shutdown 4-7

5 Device Maintenance and Service 5-1

Replacing the Generator Fuse 5-2
Cleaning the Generator 5-2
Periodic Technical and Safety Inspection 5-3
Service 5-4

6Warranty6-1

Medtronic Warranty Information 6-2

A Index A-1

vi Cardioblate® 68000 Technical Manual
How to Use This Manual
This guide presents information for users of the Model 68000 Cardioblate
®
Generator. It contains information about the
following:
Overview
®
This chapter is a brief overview of the Cardioblate and the accessories that compose the Cardioblate
Generator
®
Surgical
Ablation System.
Generator Description
This chapter identifies and describes the generator case, controls, connectors, touch-screen information fields, and touch-screen interactive fields.
Generator Operation
This chapter provides instructions for preparing the patient, setting up the generator system, and treating the patient.
Safety Shutdown and Troubleshooting
This chapter provides information on safety shutdown conditions and messages and trouble-shooting.
Device Maintenance and Service
This chapter identifies how to replace the fuse and clean the generator and provides information on technical and safety inspections.
Warranty
This chapter describes the device and accessories warranties.
Changing the Language Option
This information is also provided in Chapter 3.
Note: English is the default language for the Medtronic Cardioblate
To change the language option:
1. Access the Options Menu
To access the Options Menu, press the Options button located in the lower-right side of the RF panel screen.
®
68000 Generator.
Cardioblate® 68000 Technical Manual vii
2. Select the Advanced screen
Select the Advanced option, which will bring up the Advanced Options screen.
3. Select the Language button
Press the Language button located in the lower left-side of the Advanced Options screen. Select the desired language. The following languages are available for use:
English
German
viii Cardioblate® 68000 Technical Manual
Spanish
French
Italian
Greek
Note: At any time, the backwards arrow button can be used to return to a previous screen.
Cardioblate® 68000 Technical Manual ix
x Cardioblate® 68000 Technical Manual
System Description 1-2
Indications for Use 1-4
Contraindications 1-5
Warnings and Precautions 1-5
Adverse Events 1-10
Overview1
1
Cardioblate® 68000 Technical Manual 1-1
Chapter 1

System Description

System Description
The Medtronic® Cardioblate® Surgical Ablation System Model 68000 consists of a generator and included accessories. The surgical ablation system is designed to deliver radiofrequency (RF) energy to selected sites via a Cardioblate device (hereafter referred to only as the ablation device). The generator is line-powered and is designed to be placed on a table.
The high-impact plastic generator case has a convenient carrying handle and a display screen. An easy-to-use latch mechanism secures and releases the display screen. The display screen provides users with real-time treatment information in graphical and digital formats. Message fields on the display screen also provide critical information during the treatment procedure.
The user interface includes a touch-screen that responds to gloved and ungloved touch. When the user touches a control-button icon on the screen, the system performs the function selected if no errors exist.
The user interface also provides unique audible tones to automatically notify the user of the treatment status or alert condition.
®
surgical ablation
Key features include the following:
Touch screen operation – The generator parameters can be easily set and adjusted using the touch screen panel.
RF energy modes – The generator controls the amount of energy being delivered through the ablation device in one of the following energy modes:
In monopolar mode, the generator monitors and controls the amount of RF energy delivered to the tissue between the ablation device and the dispersive electrode at a preselected power setpoint throughout the ablation.
In bipolar and enhanced bipolar modes, the generator monitors tissue impedance and adjusts the amount of power delivered to the tissue between the active electrode jaws to create transmural lesions.
EGM throughput is available in monopolar mode only.
1-2 Cardioblate® 68000 Technical Manual

Package Contents

1
3
2
Overview
System Description
Safety shutdowns – RF energy delivery stops if the operator-selected or manufacturer-programmed impedance, power, or temperature limits are exceeded. If a warning message appears, see Chapter 4, “Safety Shutdown and Troubleshooting,” for detailed information.
The Cardioblate® Surgical Ablation System (Model 68000) package may contain the following items (Figure 1-1):
One Model 68000 generator
One Model 60883 remote control foot switch
One power cord: Model 4807
Product literature
Figure 1-1. 68000 Generator
Cardioblate® 68000 Technical Manual 1-3
Chapter 1

Indications for Use

1. Model 68000 generator
2. Power cord
3. Remote control foot switch

Other Applicable Accessories

The following accessories are applicable to the Cardioblate® Surgical Ablation System but are not included in the Model 68000 package:
Medtronic® Cardioblate® surgical ablation devices
Model 60813 - Cardioblate® Surgical Ablation Pen
Model 60814 - Cardioblate® Surgical Ablation XL Pen
Model 60831 - Cardioblate® Surgical Ablation BP2 Device
Model 60841 - Cardioblate® Surgical Ablation LP Device
Models 49260/49261 - Cardioblate® Gemini™ Surgical Ablation Devices
Medtronic released products labeled for use with the Cardioblate
Caution: Cardioblate be used with the Cardioblate
Caution: Only use Medtronic Medtronic recommended accessories. The safety of use with other surgical ablation devices, catheters, or accessories has not been assessed.
Dispersive electrode with a conductive adhesive and a minimum surface area of 100 cm IEC 60601-2-2).
Model 60889 - Medtronic® Cardioblate® Diagnostic Kit
®
68000 Generator as an external device
®
bipolar device, Model 60821, cannot
®
68000 Generator.
®
ablation devices and
2
(must be compliant with
Indications for Use
The Cardioblate® Surgical Ablation System is intended to ablate cardiac tissue during cardiac surgery using radiofrequency energy.
1-4 Cardioblate® 68000 Technical Manual

Contraindications

The Cardioblate® Surgical Ablation System is contraindicated for patients with active endocarditis at the time of surgery.

Warnings and Precautions

General

Product literature – Do not attempt to operate the Cardioblate
Surgical Ablation System or connect the ablation device to the generator prior to completely reading and understanding the product literature for the Cardioblate and the ablation device.
System compatibility – Only use Medtronic ablation devices and Medtronic-recommended accessories. The safety of use with other surgical ablation devices, catheters, or accessories has not been assessed.
Overview
Contraindications
®
Surgical Ablation System
®
Single Use Devices – Medtronic ablation devices designed for use with this generator are intended for single-time use. The Cardioblate Generator has functionality to prevent the reuse of the surgical ablation devices.
Accessories – Regularly inspect and test reusable cables and accessories. Do not use any cable or accessory that appears to be damaged, and take the following precautions:
Inspect for insulation damage, such as brittleness, cracking, thinning or bare spots.
Replace damaged accessories.
See Chapter 2, “Generator Description,” for detailed information regarding accessories.
Electrical shock and/or burns – To avoid the risk of electrical shock and/or burns to the patient, do not touch the patient while touching the outer housing or connections on the Cardioblate
®
generator. Do not allow the patient to come into contact with grounded metal surfaces during RF energy delivery.
Cardioblate® 68000 Technical Manual 1-5
Chapter 1
Warnings and Precautions
Electrostatic Discharge (ESD) is the sudden transfer or discharge of static electricity from one object to another. Human bodies can create ESD charges of up to 25,000 volts.
Connector pins identified with the ESD warning symbol should not be touched. NO connections should be made to these connectors unless ESD precautionary procedures are followed. ESD applied to one of these connectors may cause the generator to display an error message and interrupt RF energy delivery until the message is acknowledged and/or removed.
Precautionary procedures include the following:
– Prevent electrostatic charge build-up (eg, air conditioning,
humidification, conductive floor coverings, and nonsynthetic clothing);
– Discharge one’s body to the frame of the equipment or
system, to earth, or a large metal object prior to making a contact;
– Bond oneself by means of a wrist strap to the equipment or
system, or to earth prior to making a contact.

Ablation Therapy Hazards

Lesions – The long-term effects of lesions created by RF ablation
have not been established; note the following:
Any long-term effects of lesions in proximity to the conduction system or coronary vasculature are unknown.
The risks/benefits in asymptomatic patients have not been studied.
Artrioventricular (AV) node modification or septal accessory pathway ablation – Patients undergoing AV node modification or
septal accessory pathway ablation are at risk for complete AV blockage:
Closely monitor AV conduction after RF energy delivery
Using RF energy near implanted devices – RF energy may adversely affect implanted devices, such as pacemakers and implantable cardioverter-defibrillators (ICDs). To avoid this, do the following:
Keep external sources of pacing and defibrillation available during ablation.
Deactivate ICD detection during RF energy delivery.
1-6 Cardioblate® 68000 Technical Manual
Overview
Warnings and Precautions
Exercise extreme caution when delivering ablation energy in close proximity to implanted leads.
Perform complete implantable device testing before and after ablation.
Electrosurgery near flammable materials – The risk of igniting flammable gases or other materials is inherent in electrosurgery and cannot be eliminated by device design. Flammable agents used for cleaning or disinfecting, or as solvents of adhesives, should be allowed to evaporate before the application of high frequency surgery. There is a risk of pooling of flammable solutions under the patient or in body depressions such as the umbilicus, and in body cavities such as the vagina. Any fluid pooled in these areas should be mopped up before RF surgical equipment should be used. Attention should be called to the danger of ignition of endogenous gases. Take precautions to restrict flammable materials and substances from the electrosurgical site. Flammable materials may come in the following forms:
Anesthetics or skin preparation agents.
Natural chemicals produced within body cavities.
Chemicals originating in surgical drapes or other materials.
Unintentional patient burns – To avoid unintentional burns to the patient during RF energy delivery, do the following:
Minimize the distance between the dispersive electrode and the operating field.
Minimize skin-to-skin contact between parts of the patient's body by covering these areas with dry gauze.
Place monitoring electrodes as far as possible from surgical electrodes to minimize burns at the site of the monitoring electrodes.
When using multiple ablation devices, remove those devices not actively in use from patient contact. In all cases, monitoring systems incorporating high frequency current-limiting devices are recommended.
Handling RF devices during therapy – The generator is capable of delivering significant RF energy. Do not touch the ablation electrode of the ablation device or the dispersive electrode or both electrodes on a bipolar device while operating the generator.
Cardioblate® 68000 Technical Manual 1-7
Chapter 1
Warnings and Precautions
Equipment failure – The failure of the generator or hand piece could result in an unintended increase of output power. See Chapter 4, “Safety Shutdown and Troubleshooting,” for detailed information.

Generator Operation

High impedance – Monitor the device impedance measurement
display during RF energy delivery. If abnormally high impedance is observed, examine the ablation electrode for coagulum. If coagulum is present or any irrigation opening is found to be blocked, remove coagulum or replace the ablation device prior to continuing the procedure.
If high impedance is preventing ablation from occurring, do the following:
Check all connections.
Check for obvious defects or misapplication of the dispersive electrode (if the monopolar device is in use).
If a patient must be repositioned during a procedure, verify that there is proper contact between the patient and the dispersive electrode after the patient has been moved.
Low impedance – Some patients may have an abnormally low impedance between the ablation device and dispersive electrode, causing a safety shutdown. Moving the dispersive electrode to a location on the body that is farther from the device may correct this situation.
Fan – Do not obstruct the cooling fan located on the rear of the generator (Figure 2-5). Obstruction of the cooling fan could result in over-heating and device shutdown.

Cardioblate® Ablation Device Use

See Directions for Use for Cardioblate® ablation devices being used during the procedure.

Medical Environment

Line-powered equipment – Properly ground all line-powered
equipment used on or in the vicinity of the patient. An ablation device connected to an RF generator constitutes a direct, low-resistance current pathway to tissue, including the myocardium.
1-8 Cardioblate® 68000 Technical Manual
Overview
Warnings and Precautions
Necessary environment – Ablation procedures should be performed only by trained personnel in a fully equipped operating room as follows:
An emergency cardiovascular team should be available during the ablation procedure in case of a cardiovascular emergency.
The ablation device should be used only by or under the supervision of physicians trained in cardiac surgery and experienced in surgical ablation procedures.
Generator storage temperature – If the generator has been stored at temperatures greater than 30°C or less than 15°C, allow the generator to reach room temperature or wait at least 60 minutes before use, whichever is greater.
Fluid ingress – The generator may not function correctly if the electronic circuitry or the connectors are wet; note the following:
Do not allow any fluid or moisture into the generator or any connector or cables.
Do not hang fluids above the generator.
Do not immerse the cables into fluids.
Electromagnetic interference (EMI) – EMI produced by the generator during normal operation may adversely affect the performance of other equipment.
Please refer to applicable standards for additional information.
Cardioblate® 68000 Technical Manual 1-9
Chapter 1

Adverse Events

Adverse Events
The potential adverse events related to the use of ablation systems include, but are not limited to, the following:
tissue perforation
extension of extracorporeal bypass
perioperative heart rhythm disturbances (atrial and/or ventricular)
postoperative embolic complications
pericardial effusion or tamponade
injury to the great vessels
valve leaflet damage
conduction disturbances (SA/AV node)
acute ischemic myocardial event
thrombus formation
nerve damage
unintentional burns
pericarditis
pleural effusion
esophageal perforation
death
coronary sinus perforation
coronary artery spasm
atrial lead dislodgement
hypotension
pulmonary vein stenosis
cerebrovascular accident
transient ischemic attack
blood loss
1-10 Cardioblate® 68000 Technical Manual
Generator Description2
Components 2-2
Displays, Connectors, and Controls 2-3
Audible Tones 2-9
Specifications 2-9
Output Power Diagrams 2-12
Electromagnetic Emissions and Immunity Declaration (EN60601-1-2) 2-14
Applicable Standards 2-18
2
Cardioblate® 68000 Technical Manual 2-1
Chapter 2
3
4
5
2
1
6
7
8
9

Components

Components

Introduction

The Model 68000 Cardioblate® Generator (Figure 2-1) is lightweight and portable with state of the art electronics and design features that provide the user with a safe and effective Cardioblate
®
Surgical Ablation System.
Front View Back View
Figure 2-1. Model 68000 Cardioblate® Generator Case
The following are the basic user components:
1. Bezel Light Panel
2. USB Port
3. Transport Handle
4. Display Release Button
5. Display Screen
6. Therapy Connector Panel
7. Cooling Fan
8. Optional Device Connector
9. AC Power Panel
At the top of the cover assembly there is a bezel light panel with three LEDs: two blue and one green. When the center green LED is lit, it indicates that the AC power is on. When the two blue LEDs are flashing, the generator software is performing its self-test. When the blue LEDs are solid, RF energy is being applied and a lesion is in progress.
2-2 Cardioblate® 68000 Technical Manual

Displays, Connectors, and Controls

1
3
4
2
Displays, Connectors, and Controls

Right Side Panel

The applicable Medtronic® Cardioblate® surgical ablation devices and the dispersive electrode cable can be connected to the Cardioblate on the right side of the generator.
1. Cardioblate
This multipin connector provides the interface to a bipolar ablation device.
2. Cardioblate
This multipin connector provides the interface to a monopolar ablation device.
®
Generator via the therapy connector panel located
Figure 2-2. Right Side Therapy Connector Panel
®
Bipolar ablation device connector
®
Monopolar ablation device connector
Generator Description
3. Dispersive electrode connector
This connector provides the interface to the dispersive electrode via the dispersive electrode cable.
4. EGM Connection
This multipin connector provides the interface to EGM monitoring equipment.
Cardioblate® 68000 Technical Manual 2-3
Chapter 2
USB Port
Displays, Connectors, and Controls

Left Side Panel

On the left side panel there is a Universal Serial Bus (USB) port, which is intended for Medtronic field service personnel. It allows the field service personnel to connect a USB flash memory device to the generator. The flash memory device can be used for uploading application updates and for downloading stored data from the generator.
The Cardioblate loaded via this USB port. Refer to the product literature for detailed instructions.
®
Diagnostic Kit (Model 60889) application is also
Figure 2-3. USB Port location
2-4 Cardioblate® 68000 Technical Manual

Touch Screen Display

1
2
3
4
5
6
7
8 9
10 11 12
Display Screen
The color (640x480 pixels), liquid-crystal-display (LCD) screen closes into the case for storage and flips open on a hinge for use. The display screen provides users with real-time treatment information in graphical and digital formats. The screen’s hinge allows adjustment for different viewing angles. When the generator is on, the screen also remains on at all times and in all positions; there is no “screen saver” mode. Figure 2-4 details the touch screen display present upon start-up.
Generator Description
Displays, Connectors, and Controls
Figure 2-4. Touch screen display
1.
Total RF time
The cumulative amount of RF energy delivery time, since the generator was powered ON, is displayed. The counter resets to 0:00 when the generator is powered OFF and then ON.
2.
RF timer measurement display
This displays the duration (minutes:seconds) of each RF delivery cycle. The display automatically resets to 0:00 when the RF energy delivery begins and measures until RF energy delivery stops. The elapsed time of the delivered RF cycle is displayed until the next RF cycle begins.
Cardioblate® 68000 Technical Manual 2-5
Chapter 2
Displays, Connectors, and Controls
3.
Impedance measurement display
This displays the measured impedance (). For the monopolar device, this is between the ablation electrode of the ablation device and the dispersive electrode during the delivery of RF energy. For the bipolar device, this is between the jaws of the bipolar device during delivery of RF energy.
4.
RF power measurement display
This displays the amount of RF power (in watts) being delivered.
5.
Message Box
This displays text messages regarding generator status, including system warnings.
6.
RF energy button
The RF button both activates and deactivates the delivery of RF energy to the selected device.
7.
Bipolar button
The bipolar button selects the bipolar mode. This button can only be selected when a bipolar device is attached.The selected button remains active while the bipolar mode is in effect.
8.
Monopolar button
The monopolar button selects the monopolar mode. This button can only be selected when a monopolar device is attached. The selected button remains active while the monopolar mode is in effect.
9.
Options
The options button accesses the options menu. For further information on options, refer to “Options” on page 3-17.
10.
Speaker Volume Level
The speaker volume level icon is the readout for the audio control buttons. The number of curved lines increase and decrease with the volume settings. There are five volume levels; the lowest level is still audible and there is no OFF setting.
11.
Audio control
The audio control buttons adjust speaker volume.
12.
Information button
The information button provides details to the user regarding possible causes and solutions for device errors.
2-6 Cardioblate® 68000 Technical Manual

Rear Panel

1 2 3
4
5
6
7
Generator Description
Displays, Connectors, and Controls
The rear panel (Figure 2-5) contains the AC power, equipotential ground, and optional device connections. The AC power connection includes a power-entry plug assembly with an ON/OFF power switch and two power fuses.
A recessed, equipotential grounding terminal provides convenient access for grounding the generator with another electronic devices. There is also a connector for the cable to the optional Medtronic foot switch Model 60883 that activates the radio frequency delivery system.
The rear panel USB communications port is to be used only by Medtronic field service personnel. The port is secured against unauthorized use.
1.
Fan
Provides constant air flow to cool electronics when power is on. Do not obstruct the cooling fan. Obstruction of the fan could result
in overheating and device shutdown.
2.
Equipotential ground post
Provides common electrical ground for other electronic devices being used during the procedure.
3.
USB Port
Connection to USB cable for external digital device communication. This security-protected connector is for Medtronic use only.
Figure 2-5. Rear panel
Cardioblate® 68000 Technical Manual 2-7
Chapter 2
Displays, Connectors, and Controls
4.
Remote control foot switch connector
This connector provides the interface to the remote control foot switch.
5.
Power entry connector
This connector provides the interface to AC line power via the power cord.
6.
Fuse
This housing contains two field-replaceable 3.15 Amp. slow-blow fuses. Refer to “Replacing the Generator Fuse” on page 5-2.
7.
ON/OFF Switch

Remote Foot Switch

The Model 60883 Remote Foot Switch (Figure 2-6) is an optional remote control device that connects to the rear panel of the Model 68000 Cardioblate switch is properly connected to the generator, the user can press the foot switch to start RF energy delivery. To stop the RF energy delivery, the user releases the remote foot switch.
®
Generator. When the remote foot
The connector end of the remote foot switch has an easy-to-use connector release mechanism. To disconnect the cable from the generator, pull back on the textured grip ring to release the lock before removing the connector.
Figure 2-6. Model 60883 Remote Foot Switch
2-8 Cardioblate® 68000 Technical Manual

Audible Tones

Generator Description
Audible Tones
Tab le 2-1. Model 68000 audible tones
Generator tone Meaning
Two tones at regularly spaced intervals
Regularly spaced single medium-pitch beeps
Regularly spaced dual medium-pitch beeps
Three low-pitched beeps An error condition has caused a
Continuous medium-pitch tone Transmurality algorithm is complete

Specifications

Model 68000 Generator

Physical characteristics
Dimensions 55 x 35 x 14 cm (21.8 x 13.8 x 5.5 in)
Weight 9.5 kg (21 lbs)
Power requirement 100 - 240 VAC, 270VA, 50 - 60 Hz
Fuse protection Two 250 V, 3.15 Amp. slow-blow fuses
System error that may require a re-boot
RF energy is being delivered to the monopolar device
RF energy is being delivered to the bipolar device
safety shutdown
Table 2 -2. Model 68000 generator specifications
(user accessible)
Cardioblate® 68000 Technical Manual 2-9
Chapter 2
Specifications
Tabl e 2 -2. Model 68000 generator specifications (continued)
Functional characteristics
RF power output setpoint/limit (Monopolar Mode)
RF power output setpoint/limit (Bipolar Mode)
RF power output setpoint/limit (Enhanced Bipolar Mode)
RF power output frequency
Maximum RF voltage, between electrodes
Maximum RF Current
Impedance limits
RF timer limit (Monopolar Mode)
RF timer limit (Bipolar Mode)
Range: 1 to 50 W Increment: 1 W Accuracy: ± 10% or 3 W, whichever is
greater (Impedance ≥ 30 Ω and Impedance 400 Ω)
Accuracy: ± 20% or 3 W, whichever is greater (Impedance < 30 or Impedance >400Ω)
Range: 15 to 40 W (automatically varying) Increment: 5 W Accuracy: ± 10% or 3 W, whichever is
greater (Impedance ≥ 30 Ω and Impedance 400 Ω)
Accuracy: ± 20% or 2 W, whichever is greater (Impedance < 30 or Impedance >400Ω)
Range: 15 to 55 W (automatically varying)
Increment: 5 to 10 W, varying Accuracy: ± 10% or 3 W, whichever is
greater (Impedance ≥ 30 Ω and Impedance 400 Ω)
Accuracy: ± 20% or 2 W, whichever is greater (Impedance < 30 or Impedance >400Ω)
473 ± 5 kHz
180 Vrms ± 10%
2A ± 150 mA Range: 12 to 600
Increment: 5 Accuracy: ± 15% for 20 to 500 Accuracy: ± 20% or 5 (whichever is
greater) for 12 to 20 and 500 to 600
Range: 1 second to 2 minutes, inclusive Increment: 1 second Accuracy: ± 1% or 1 second, whichever is
greater
Range: 2 to 40 seconds (automatically controlled)
2-10 Cardioblate® 68000 Technical Manual
Generator Description
Tab le 2-2 . Model 68000 generator specifications (continued)
RF timer limit (Enhanced Bipolar Mode)
Default Settings
RF power
Impedance (Monopolar Mode)
Impedance (Bipolar Mode)
RF timer
Temperature and humidity information
Operating temperature and humidity
Storage/shipping temperature and humidity
Rear panel connectors
Fused power connector
Ground post
Remote control foot switch connector
Side panel connectors
Dispersive electrode
®
Cardioblate
monopolar
ablation device Cardioblate® bipolar
ablation device EGM connector
Safety information
a
Lemo® is a registered trademark of Lemo USA, Inc.
Range: 2 to 45 seconds (automatically controlled)
Monopolar mode: 25 W Bipolar mode: NA
Maximum: 500 Minimum: 12
Maximum: 500 Minimum: 12
Monopolar mode: 2 minutes
10 to 40°C, 80% relative humidity
-40 to 66°C, 95% relative humidity
3-prong fused mains (AC) power receptacle
Equipotential ground post
®a
3-pin Lemo
receptacle
2-pin return electrode
10-pin Lemo
16-pin Lemo
4-pin Lemo
®
®
®
receptacle
IEC 60601-1, Class I, Type CF, IPX0, no sterilization, equipment not suitable for use in the presence of a flammable anaesthetic mixture with air, oxygen, or nitrous oxide, continuous operation
UL 60601-1: With respect to electrical shock, fire, and mechanical hazards, only in accordance with UL 60601-1
Specifications
Cardioblate® 68000 Technical Manual 2-11
Chapter 2
0
20
40
60
80
100
120
0
25
50
100
150
200
250
300
350
400
450
500
550
600
650
700
750
800
850
900
950
1000
100 W
50 W
25 W
Power (W)
Load (OHMS)

Output Power Diagrams

Accessories

®
Medtronic
Model 60883 remote
control foot switch
Dimensions
Cable length
Weight
Connector
Environmental conditions
®
Medtronic
Model 4807, 4808, or
4809 power cord
Length
Connectors
Output Power Diagrams
Table 2 -3. Accessory specifications
13x15x5cm (5x6x2in)
472 cm (15.4 ft)
645g (1.42lbs)
3-pin Lemo
®
IP 68, water resistant
3 m (nominal) (10 ft)
3-pin hospital-grade safety plug
The following diagrams (Figure 2-7 through Figure 2-9) show the output power versus load impedance at three power setpoints: 25 W, 50 W, and 100 W at cold and warm operating temperatures. The generator was tested at fixed impedances using noninductive resistors.
2-12 Cardioblate® 68000 Technical Manual
Figure 2-7. Monopolar Load Curves
Generator Description
0
20
40
60
80
100
120
0
25
50
100
150
200
250
300
350
400
450
500
550
600
650
700
750
800
850
900
950
1000
100 W
50 W
25 W
Power (W)
Load (OHMS)
0.0
20.0
40.0
60.0
80.0
100.0
120.0
140.0
160.0
180.0
200.0
0 5 10 15 20 25 30 35 40 45 50
25 Ohms
50 Ohms
100 Ohms
200 Ohms
300 Ohms
400 Ohms
500 Ohms
600 Ohms
700 Ohms
0.0
20.0
40.0
60.0
80.0
100.0
120.0
140.0
160.0
180.0
200.0
0 5 10 15 20 25 30 35 40 45 50
25 Ohms
50 Ohms
100 Ohms
200 Ohms
300 Ohms
400 Ohms
500 Ohms
600 Ohms
700 Ohms
0.0
20.0
40.0
60.0
80.0
100.0
120.0
140.0
160.0
180.0
200.0
0 5 10 15 20 25 30 35 40 45 50
25 Ohms
50 Ohms
100 Ohms
200 Ohms
300 Ohms
400 Ohms
500 Ohms
600 Ohms
700 Ohms
Output Voltage (Vrms)
Power Setting (W)
Output Power Diagrams
Figure 2-8. Bipolar Load Curves
Figure 2-9. Voltage Curve
Cardioblate® 68000 Technical Manual 2-13
Chapter 2

Electromagnetic Emissions and Immunity Declaration (EN60601-1-2)

Electromagnetic Emissions and Immunity Declaration (EN60601-1-2)
Table 2 -4. Electromagnetic Emissions
The Model 68000 Generator is intended for use in the electromagnetic environment specified below. The customer or the user of the generator should ensure that it is used in such an environment.
Emissions test Compliance Electromagnetic Environment –
Radio-frequency (RF) emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC/EN 61000-3-2
Voltage fluctuations/ flicker emissions IEC/EN 61000-3-3
The Model 68000 Generator is intended for use in the electromagnetic environment specified below. The customer or the user of the generator should ensure that it is used in such an environment.
Immunity Test EN 60601
Test Level
Electrostatic discharge (ESD):
IEC/EN 61000-4-2
±6 kV contact ±8 kV air
Group 2 The Model 68000 must emit
electromagnetic energy in order to perform its intended function. Nearby electronic equipment may be affected.
Class A The Model 68000 is suitable for use
Class A
Complies
Tab le 2-5. Electromagnetic Immunity
Compliance Level Electromagnetic
±6 kV contact ±8 kV air
in all establishments other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Guidance
Environment – Guidance
Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
2-14 Cardioblate® 68000 Technical Manual
Electrical fast transient/burst:
IEC/EN 61000-4-4
Surge: IEC/EN 61000-4-5
Voltage dips, short interruptions and voltage variations on power supply input lines:
IEC/EN 61000-4-11
Generator Description
Electromagnetic Emissions and Immunity Declaration (EN60601-1-2)
Table 2 -5. Electromagnetic Immunity
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV line(s) to line(s) ±2 kV line(s) to Earth
<5% U
T
(>95% dip in UT) for 0.5 cycle
40% U
T
(60% dip in UT) for 5 cycles
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV line(s) to line(s) ±2 kV line(s) to Earth
<5% U
T
(>95% dip in UT) for 0.5 cycle
40% U
T
(60% dip in UT) for 5 cycles
Main power quality should be that of a typical commercial or hospital environment.
If the user of the device requires continued operation during power main interruptions, it is recommended that the device be powered from an uninterruptible power supply or a battery.
Power frequency (50/60 Hz) magnetic field:
IEC/EN 61000-4-8
Note: U
is the A.C. mains voltage prior to application of the test level.
T
70% U
T
(30% dip in UT) for 25 cycles
<5% U
T
70% U
T
(30% dip in UT) for 25 cycles
(>95% dip in UT) for 25 seconds
3 A/m 3 A/m Power frequency magnetic
fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
Cardioblate® 68000 Technical Manual 2-15
Chapter 2
Electromagnetic Emissions and Immunity Declaration (EN60601-1-2)
Tabl e 2 -6. Electromagnetic Immunity, IEC 60601-1-2 Table 204
Immunity Test EN 60601
Test Level
Portable and mobile RF communications equipment should be used no closer to any part of the Model 68000, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Compliance Level Electromagnetic
Environment — Guidance
Recommended separation
distance
Conducted RF
IEC/EN 61000-4-6
Radiated RF
IEC/EN 61000-4-3
3Vrms
150 kHz to 80 MHz
3V/m
80 MHz to 2.5 GHz
3Vrms d=1.2√ P
3V/m d=1.2√ P 80 MHz to 800 MHz
d=2.3√ P 800 MHz to 2.5 GHz
Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than the compliance level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with the following symbol:
Notes:
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Model 68000 is used exceeds the applicable RF compliance level above, the Model 68000 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the device.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
a
2-16 Cardioblate® 68000 Technical Manual
Generator Description
Electromagnetic Emissions and Immunity Declaration (EN60601-1-2)
Table 2 -7. Recommended separation distances between portable and mobile radio frequency
(RF) Communication equipment and the Model 68000 Generator. IEC/EN 60601-1-2 Table 206
The Model 68000 Generator is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the generator can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the generator as recommended below, according to the maximum output power or the communications equipment.
Rated maximum
output power of
Separation distance according to frequency of transmitter
(m)
transmitter
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
(W) d = 1.2Pd=1.2√ Pd=2.3√P
0.01 0.35 0.12 0.23
0.1 1.1 0.38 0.73
1 3.5 1.2 2.3
10 11 3.8 7.3
100 35 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Notes:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
Cautions:
Do not use non-Medtronic components with Medtronic in-line-powered external devices. The use of non-Medtronic components may result in damage to Medtronic components, increased emissions, or decreased electromagnetic immunity of the Medtronic devices or systems.
Do not use Medtronic in-line-powered external devices adjacent to, or stacked with, other electronic devices. Using Medtronic devices in these configurations may result in decreased electromagnetic immunity of the Medtronic devices or systems.
®
RF Cable: The Medtronic
Cardioblate® surgical ablation devices’ cable length is approximately
3 m in length.
Cardioblate® 68000 Technical Manual 2-17
Chapter 2

Applicable Standards

Applicable Standards
IEC 60601-1 IEC 60601-1-AM1 IEC 60601-1-AM2
IEC 60601-1-2 Electromagnetic compatibility
IEC 60601-2-2 Medical electrical equipment - Part 2: Particular
EN 60601-1-4 Medical electrical equipment - Part 1: General
UL 60601-1 Grounded construction, ordinary connection
FCC Part 18 FCC standard for EMI/EMC requirements
Class I, Type CF, ordinary equipment, no sterilization, benign environment, continuous mode of operation, applied part specifically designed for contact directly with the heart
requirements for the safety of high-frequency surgical equipment
requirements for safety - 4. collateral standard: programmable electrical medical systems
This equipment has been tested and found to comply with the limits for medical devices to IEC 60601-1-2 [or EN 60601-1-2 or Medical Device Directive 93/42/EEC]. This testing shows the device provides reasonable protection against harmful interference in a typical medical installation. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to other devices or is negatively impacted by other devices, the user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the devices.
Increase the separation between the devices.
Connect the equipment to an outlet on a different circuit.
Consult the manufacturer or field service technician for help.
2-18 Cardioblate® 68000 Technical Manual
Generator Operation
Quick Reference Guide 3-2
Selecting and Attaching a Dispersive Electrode 3-3
Preparing and Powering Up the Generator 3-4
Connecting a Cardioblate® Ablation Device 3-6
Verifying Generator Recognition of Attached Devices 3-8
Disconnecting the Cardioblate® Ablation Devices 3-9
Setting the RF Energy Mode and Parameters 3-9
Monopolar Mode Selection and Use 3-9
Bipolar Mode Selection and Use 3-14
3
3
Options 3-17
Powering Down the Generator 3-23
Cardioblate® 68000 Technical Manual 3-1
Chapter 3

Quick Reference Guide

Quick Reference Guide
A quick reference guide is located on the surface of the generator, under the display screen as shown in Figure 3-1. Detailed steps regarding setup and operation are on the following pages. Refer to the ablation device Instructions for Use for specific information on device use.
1. Attach the dispersive electrode.
2. Plug in the power cord and foot switch.
3. Set up saline irrigation.
4. Attach devices to the generator after the self-test is complete.
3-2 Cardioblate® 68000 Technical Manual
Figure 3-1. Quick Reference Guide
Generator Operation
Back Mid-back Buttocks

Selecting and Attaching a Dispersive Electrode

Selecting and Attaching a Dispersive Electrode

Selecting a Dispersive Electrode and Application Site

Note: A dispersive electrode is used only with a monopolar
device.
1. Select a dispersive electrode with the following characteristics:
The dispersive electrode must have a conductive adhesive backing and a minimum surface area of 100 cm
The dispersive electrode must meet or exceed American National Standards Institute (ANSI) and Association for the Advancement of Medical Instrumentation (AAMI) requirements.
Refer to the product literature packaged with the dispersive electrode for additional information on the use of the dispersive electrode.
2. Select an application site for the dispersive electrode that allows the entire area of the dispersive electrode to adhere to the patient's body and is as close to the heart as possible. The recommended sites are the patient's back, mid-back, or buttocks (Figure 3-2).
2
.
Caution: Place electrocardiogram (ECG) electrodes and other monitoring probes as far away as possible from the ablation site and the attached dispersive electrode to reduce the risk of burns.
Figure 3-2. Dispersive electrode application site
Cardioblate® 68000 Technical Manual 3-3
Chapter 3
1

Preparing and Powering Up the Generator

Attaching the Dispersive Electrode

1. Shave, clean, and dry the application site.
2. Apply the dispersive electrode to the application site according to the instructions in the dispersive electrode manufacturer’s product literature.
3. Use antistatic (nonconductive) sheeting to keep the patient from coming into contact with metal parts that are grounded or can act as a ground (eg, the operating table, supports).
4. Place dry gauze on the patient in areas of potential skin-to-skin contact (eg, between arms and torso).
Caution: Ensure the entire dispersive electrode is attached to the patient’s body.
Preparing and Powering Up the Generator

Setting Up the Generator

1. Place the RF generator on a suitable table that does not obstruct the rear-panel cooling fan air flow. Then, press the display release button and position the display screen.
2. Verify that all of the components are present. See “Package Contents” on page 1-3.
3. Inspect the components. If any component is damaged, contact your local Medtronic representative.
3-4 Cardioblate® 68000 Technical Manual
Figure 3-3. Open Display Screen

Connecting Cables to the Generator

2
1
3
Note: Refer to “Displays, Connectors, and Controls” on page 2-3 for connector locations on the generator.
Figure 3-4. AC Power Panel and Optional Device Connectors
1. Connect the power cord:
a. Plug the power cord into the power entry connector on the back of
the generator (see Figure 3-4).
b. Plug the power cord into an AC outlet.
2. Connect the remote control foot switch to the generator:
a. Align the foot switch cable connector to the foot switch connector
on the back of the generator (see Figure 3-4).
b. Insert the foot switch connector until the metal locking ring clicks.
Note: To disconnect the foot switch from the generator, pull back on the textured grip ring to release the lock before removing the connector.
3. Monopolar Devices Only: Dispersive electrodes are required for operation with monopolar devices. Dispersive electrodes have two styles — with or without cables pre-attached. For dispersive electrodes with cables attached, simply connect the cable end to the generator. If the dispersive electrode does not have a pre-attached cable, use a separate cable and connect one end to the dispersive electrode and the other end to the generator.
Generator Operation
Preparing and Powering Up the Generator
Figure 3-5. Therapy Connector Panel
Cardioblate® 68000 Technical Manual 3-5
Chapter 3
1
Connecting a Cardioblate
®
Ablation Device

Powering Up the Generator

1. Press the AC power button up on the rear panel once to
power up the generator.
2. Once the generator completes its self-test, the RF panel screen will appear.
Figure 3-6. AC Power Panel
Figure 3-7. RF Panel Screen

Connecting a Cardioblate® Ablation Device

Note: Refer to the product literature supplied with the ablation
device for further information on its use.
Note: Ensure the generator is turned on and the self-test completed prior to attaching ablation devices.
3-6 Cardioblate® 68000 Technical Manual
Generator Operation
1
Connecting a Cardioblate
®
Ablation Device
Note: Do not attempt to reuse an ablation device with this generator. If the ablation device has been previously used, the error message, “Device Error. Replace Device.” will appear. Replace the ablation device before proceeding.
Note: If the ablation device is not a Medtronic ablation device designed for use with this system, the error message “Device Error. Invalid device attached.” will appear until the device is replaced with a Medtronic ablation device.
Connecting the Monopolar Ablation Device
1. Orient the monopolar ablation device plug with the arrow pointing up.
2. Insert the multipin connector of the monopolar ablation device into the monopolar device port on the right side of the generator until it locks into place (Figure 3-8). Do not force the connection.
Figure 3-8. Monopolar Device Port
Note: After the device is connected and recognized, the
Monopolar Button will become active (dark grey).
Note: An error message will appear if the dispersive electrode is not attached and recognized.
Connecting the Bipolar Ablation Device
1. Orient the bipolar ablation device plug with the arrow pointing up.
2. Insert the multipin connector of the bipolar ablation device into the bipolar device port on the right side of the generator until it locks into place (Figure 3-9). Do not force the connection.
Cardioblate® 68000 Technical Manual 3-7
Chapter 3
2

Verifying Generator Recognition of Attached Devices

Figure 3-9. Bipolar Device Port
Note: After the device is connected and recognized, the
Bipolar button will become active (dark grey).
Verifying Generator Recognition of Attached Devices
1. Press the Options button on the Generator RF Panel.
2. Press the Setup button. The Setup and Connection Status screen appears (Figure 3-10).
3. Verify that the components attached to the generator are recognized. If the component is recognized, a checkmark will appear on the screen in beside the device icon. If the components are not recognized, a red arrow will appear on the screen pointing to the generator icon.
4. Press the backwards arrow button RF Panel screen.
to return to the main
Figure 3-10. Setup and Connection Status screen, devices not attached
3-8 Cardioblate® 68000 Technical Manual
or recognized
Generator Operation
Disconnecting the Cardioblate
®
Ablation Devices

Disconnecting the Cardioblate® Ablation Devices

To disconnect the devices from the generator, pull back on the grip ring to release the lock before removing the device.

Setting the RF Energy Mode and Parameters

The generator can control the amount of energy being delivered through an ablation device in two modes.
In monopolar mode, the generator monitors and controls the amount of RF energy delivered to the tissue between the ablation device and the dispersive electrode at a preselected power setpoint throughout the ablation.
In bipolar mode, the generator monitors tissue impedance and adjusts the amount of power delivered to the tissue between the active electrode jaws to create transmural lesions.

Monopolar Mode Selection and Use

Selecting the Monopolar Mode

Select the Monopolar Button
Using the touch screen buttons located at the lower-left corner of the RF Panel display, select the Monopolar button . When the monopolar mode is selected, the Monopolar button will appear depressed and grey (Figure 3-11).
1. The display screen will show the limit and setpoint values from the previous use of the generator.
2. Review the displayed values for RF power setpoint and RF timer limit. Adjust the values as necessary.
Cardioblate® 68000 Technical Manual 3-9
Chapter 3
Monopolar Mode Selection and Use
Figure 3-11. RF Panel in Monopolar Mode

Adjusting the Setpoint Values and Limits in the Monopolar Mode

Refer to Table 3-1 for a more detailed description of adjustable parameter functions in monopolar mode.
1. Adjust the RF power setpoint in 1 W increments by pressing the button to increase or button to decrease the value (Figure 3-11).
Caution: The selected output power should be as low as possible for the intended purpose. Refer to the device product literature for detailed instructions. A low output power may be useful for the following situations:
Ablation of targets where there is risk of collateral damage to structures in close proximity, eg, ablation of a septal pathway very close to the AV node or esophageal damage.
Ablation in locations where blood flow may be limited postablation.
Ablation in patients who may have unusually thin cardiac walls.
3-10 Cardioblate® 68000 Technical Manual
Generator Operation
Monopolar Mode Selection and Use
Tab le 3-1 . Operational modes and adjustable parameter functions
Mode Parameter and function Safety shutdown
condition
Monopolar mode
RF power setpoint is the
wattage of RF energy delivery.
Impedance limit is
automatically set.
RF timer limit is the maximum amount of time in an RF energy delivery cycle.

Adjusting the RF Timer Limits in the Monopolar Mode

1. Access the Limits menu by pressing the Options button on the RF Panel screen and then the Limits button.
2. On the Limits screen, adjust the RF timer in 1-second increments by pressing value.
to increase or to decrease the
If the measured power varies from the setpoint, the generator will stop RF energy delivery.
If the measured impedance is outside the limits, the generator will stop RF energy delivery.
If set time is exceeded, the generator will stop RF energy delivery.
Figure 3-12. Limits screen
3. Verify the parameters are set properly before depressing the remote control foot switch or pressing the RF button
to
deliver energy.
Cardioblate® 68000 Technical Manual 3-11
Chapter 3
Monopolar Mode Selection and Use
Note: The RF Timer Limits can be changed at any point during the procedure as long as RF energy is not active.

Delivering Radiofrequency (RF) Energy in the Monopolar Mode

Caution: If there are any concerns regarding the safety or efficacy
or the equipment, do not deliver RF energy. Contact your local Medtronic representative.
Note: Refer to the product literature for ablation device instructions prior to use.
1. Select the monopolar mode by pressing the appropriate button on the touch screen, and verify that the appropriate device is active.
Note: Monopolar mode can only be selected if the monopolar device is attached and recognized by the generator.
Note: RF energy can only be delivered to one device at a time.
2. To deliver RF energy, either press and hold the remote control foot switch or press the RF button screen. RF energy is delivered through the monopolar ablation device.
on the touch
Delivery of RF energy is indicated by:
The light panel illuminating blue.
The message box displays “Ablating.”
The generator beeping. While in monopolar mode, the generator will beep once per second.
The touch screen display shows the ablation time, total accumulated RF time, impedance, and power during ablation.
Caution: Accidental or unintentional activation of the foot switch may cause RF energy to be delivered.
3. To change the power, timer setpoint values, and limits, see “Adjusting the Setpoint Values and Limits in the Monopolar Mode” on page 3-10, with the exception noted below.
Note: During RF energy delivery in monopolar mode, the or buttons must be pressed repeatedly to adjust values.
Note: Device selection cannot be changed during the delivery of RF energy.
3-12 Cardioblate® 68000 Technical Manual
Generator Operation
Monopolar Mode Selection and Use
4. RF energy delivery will stop when the RF timer limit is reached or the user chooses to end RF energy delivery. To stop RF energy delivery before the timer reaches the limit, release the foot switch or press the RF button .
Note: To deliver RF energy, either press and hold the remote control foot switch or press the RF button
on the touch
screen. If the remote control foot switch is used to deliver RF power and the RF button is pressed to cease the delivery of RF power, the foot switch must be released before RF power can be delivered again.
Note: RF energy delivery will stop if a setpoint or limit is exceeded, causing a safety shutdown. An error message will appear in the message box. Press the Information Button more information. For information about safety shutdown messages, see “Safety Shutdown Messages” in Chapter 4.
5. When RF energy delivery ends, the generator will return to standby as shown in Figure 3-13.
for
Figure 3-13. RF Panel standby mode
Cardioblate® 68000 Technical Manual 3-13
Chapter 3

Bipolar Mode Selection and Use

Bipolar Mode Selection and Use

Selecting the Bipolar Mode

Select the Bipolar Button
Using the touch screen buttons located at the lower-left corner of the RF Panel display, select the Bipolar button . When the bipolar mode is selected, the Bipolar button will appear depressed and dark grey.
1. The values for RF power setpoint and limits, impedance limit, and RF timer limit will be adjusted automatically.
Refer to Table 3-2 for a more detailed description of adjustable parameter functions in bipolar mode.
Tabl e 3 -2. Operational modes and adjustable parameter functions
Mode Parameter and function Safety shutdown
Bipolar mode RF power setpoint and
limit are automatically set.
Impedance limit is
automatically set.
condition
If the measured power varies from the algorithm setpoint, the generator will stop RF energy delivery.
If the measured impedance is outside the limits, the generator will stop RF energy delivery.
RF timer limit is automatically set.
If maximum time is exceeded, generator will stop RF energy delivery.

Delivering Radiofrequency (RF) Energy in the Bipolar Mode

Caution: If there are any concerns regarding the safety or efficacy
or the equipment, do not deliver RF energy. Contact your local Medtronic representative.
Note: Refer to the product literature for ablation device instructions prior to use.
1. Select the bipolar mode by pressing the appropriate button on the touch screen, and verify that the appropriate device is active.
3-14 Cardioblate® 68000 Technical Manual
Generator Operation
Bipolar Mode Selection and Use
Note: Bipolar mode can only be selected if a bipolar device is attached and recognized by the generator.
Note: RF energy can only be delivered to one device at a time.
2. To deliver RF energy, either press and hold the remote control foot switch or press the RF button on the touch screen. RF energy is delivered through the bipolar ablation device.
Delivery of RF energy is indicated by:
The light panel illuminating blue.
The message box displays “Ablating.”
The generator beeping. While in bipolar mode, the generator will beep twice per second.
The screen shows the ablation time, total accumulated RF time, impedance, and power during ablation.
Caution: Accidental or unintentional activation of the foot switch may cause RF energy to be delivered.
Note: Device selection cannot be changed during the delivery of RF energy.
3. RF energy delivery will stop when the RF timer limit is reached or the user chooses to end RF energy delivery. To stop RF energy delivery before the timer reaches the limit, release the foot switch or press the RF button .
Note: To deliver RF energy with a BP2 or LP device, either press and hold the remote control foot switch or press the RF button on the touch screen. If the remote control foot switch is used to deliver RF power and the RF button is pressed to cease the delivery of RF power, the foot switch must be released before RF power can be delivered again.
Note: Achievement of a transmural lesion in the bipolar mode is indicated by the generator tone changing from a beeping to a steady tone and “Transmurality reached. Power is ON.” appearing on the screen (Figure 3-14). The green transmurality symbol will also appear next to the power output. RF energy will continue until the surgeon stops ablating. To stop RF energy, release the foot switch or press the RF button .
Cardioblate® 68000 Technical Manual 3-15
Chapter 3
Bipolar Mode Selection and Use
RF energy delivery will also stop if a setpoint or limit is exceeded, causing a safety shutdown. For information about safety shutdown messages, see “Safety Shutdown Messages” in Chapter 4.
4. When RF energy delivery ends, the generator will return to standby.
Figure 3-14. Transmurality Reached
3-16 Cardioblate® 68000 Technical Manual
Figure 3-15. RF Panel standby mode

Options

Options Menu

Generator Operation
Options
The Options Menu provides user-friendly instructions for setting up the device and defining limits, and includes advanced features such as the Graph Mode. Software version information can also be found in the Options Menu.
To access the options menu, press the Options button located in the lower-right side of the RF panel screen to access the Set Up, Limits, and Advanced screens.
Figure 3-16. Options Menu
Note: At any time the backwards arrow button can be used
to return to a previous screen.
Set Up Screen
Selecting the Set Up option will display the device connections screen (Figure 3-17).
Cardioblate® 68000 Technical Manual 3-17
Chapter 3
Options
Figure 3-17. Setup and Connections Status screen
The Setup and Connections Status screen indicates which devices the generator recognizes as being currently attached. Next to each device is an Information button
for selecting
the information screen.
The information screen provides detailed instructions for setting up and attaching the devices to the generator.
Additional information is provided for the monopolar and bipolar devices by selecting the Setup button displayed on the information screen.
Limits Screen
Selecting the Limits option will bring up the limits selection screen (Figure 3-18).
3-18 Cardioblate® 68000 Technical Manual
Generator Operation
Options
Figure 3-18. Limits screen
The time and impedance limits are listed for the bipolar and monopolar devices. Only the RF Timer Limit for monopolar mode can be adjusted.
1. Adjust the RF Timer Limit in 1-second increments by pressing
to increase or to decrease the value.
2. Verify the parameters are set properly before depressing the remote control foot switch or pressing the RF button to deliver energy.
Note: The RF Timer Limits can be changed at any point during the procedure as long as RF energy is not active.
Note: Maximum and minimum limits cannot be altered.
Advanced Screen
Selecting the Advanced option will bring up the advanced options screen (Figure 3-19).
Cardioblate® 68000 Technical Manual 3-19
Chapter 3
Options
Figure 3-19. Advanced screen
This screen includes the following parameters:
Graph selection: This allows the user to turn on or off the display of the impedance/power graph on the RF Panel screen. Touch the button to enable the graph. Figure 3-20 shows the RF Panel when graph mode is selected.
3-20 Cardioblate® 68000 Technical Manual
Figure 3-20. RF Panel in graph mode
Generator Operation
Options
The Advanced screen also displays the following software information:
Application Version
User Interface Version
Core Version
FPGA Version
Changing the Language Option
®
Note: English is the default language for the Cardioblate Generator.
Press the Language button located in the lower left-side of the Advanced Options screen. Select the desired language. The following languages are available for use:
English
German
Spanish
French
Italian
Greek
Figure 3-21. Changing Language Option Screen
Cardioblate® 68000 Technical Manual 3-21
Chapter 3
Options

Optional Connections

Caution: Peripheral equipment connected to the Cardioblate®
Generator and any system formed by the generator and the connected peripherals shall be certified according to the respective IEC standards (eg, IEC 950 for data processing equipment and IEC 60601-1 for medical equipment). It is the responsibility of the person who connects the peripheral equipment to comply with the required standards. Consult the Technical Services Department or your local representative if additional information is required.
The following connection is optional:

Electrogram (EGM) Monitoring System

Note: For this option, a Medtronic Model 05328 EGM cable is
required.
To connect an EGM monitoring system to the generator using the EGM cable, do the following:
1. Position the metal EGM cable connector so that the red dot is on top.
2. Insert the connector into the EGM connector on the generator side panel until the connector locks into place. Do not force the connection.
3. Connect the opposite end of the cable to the EGM monitor.
Note: To disconnect the EGM cable from the generator, pull back on the textured grip ring to release the lock before removing the connector.
Caution: If noise is noted on the EGM during RF power delivery, try the following to diminish noise:
Remove surface ECG skin electrodes and the dispersive electrode. Prepare these areas and apply new skin electrodes and the dispersive electrode according to instructions in the manufacturer's product literature.
Turn off other equipment in the room if possible.
3-22 Cardioblate® 68000 Technical Manual

Powering Down the Generator

Turn Off the Generator

1. Turn OFF the generator by pressing the AC power button located on the rear panel.
2. Disconnect all of the devices.
3. Close the display.
4. Discard single-use ablation devices per standard procedure.
Generator Operation
Powering Down the Generator
Cardioblate® 68000 Technical Manual 3-23
Chapter 3
Powering Down the Generator
3-24 Cardioblate® 68000 Technical Manual
Safety Shutdown and
Safety Shutdown Conditions 4-2
Safety Shutdown Messages 4-3
Other Messages 4-7
System Shutdown 4-7
Troubleshooting
4
4
Cardioblate® 68000 Technical Manual 4-1
Chapter 4

Safety Shutdown Conditions

Safety Shutdown Conditions
The generator will perform a safety shutdown to prevent an unsafe operating condition. During a safety shutdown, RF energy delivery will stop and will not be initiated again until the condition that caused the safety shutdown is resolved.

Alert Messages

An alert message occurs when the generator detects a problem. An alert message can appear on the screen either during a lesion, before a lesion, or after a lesion. Alert messages also include the probable cause and the recommended corrective action.
If an alert condition occurs with the RF energy ON, the alert automatically stops the lesion. It turns OFF the RF energy
output to both channels. The simultaneous lesion abort and alert message comprises the system's safety shutdown feature.
All system alert messages require some corrective action by the user.
Some alert messages stay on the LCD screen until the user clears the causing condition.
There is also an audible alarm for each message.
The following will occur during a safety shutdown:
RF energy is terminated.
The generator beeps three times.
All controls become inactive until the safety shutdown condition is resolved.
Blue LEDs turn off.
An “alert” message displays on the LCD screen and remains on until the cause is corrected.
The following sections present possible causes of and possible solutions for several safety shutdown messages.
If the surgical ablation system does not operate properly after taking the steps listed in the tables or listed in the message itself, contact technical support as listed on the rear panel of the device or your local Medtronic representative.
4-2 Cardioblate® 68000 Technical Manual

Safety Shutdown Messages

Tab le 4-1 . Safety shutdown messages
Error message Possible cause Possible solution
Safety Shutdown and Troubleshooting
Safety Shutdown Messages
“Ablation stopped: Device not attached.”
“Ablation stopped: Maximum ablation time exceeded.”
“Current limit exceeded.”
“Device error. Invalid device attached.”
The device was disconnected from the generator while an ablation was in process.
The maximum time limit has been reached.
The maximum current limit for the generator has been reached.
A malfunctioning device or a device that does not conform to the specifications of this generator has been detected.
1) Reattach device.
2) Restart the ablation.
In Monopolar Mode:
1) Check the maximum time and adjust time limit if necessary.
2) Restart the ablation.
In Bipolar Mode:
1) Halt the ablation after transmurality has been indicated.
2) Restart the ablation.
1) Turn the generator OFF then ON.
1) Replace the ablation device.
“Device error. Replace device.”
“Dispersive electrode not attached.”
The device has been previously used.
The dispersive electrode is not attached to the generator.
The dispersive electrode is faulty.
Cardioblate® 68000 Technical Manual 4-3
1) Replace the ablation device.
1) Reattach the dispersive electrode.
2) Restart the ablation.
1) Replace the dispersive electrode.
2) Restart the ablation.
Chapter 4
Safety Shutdown Messages
Error message Possible cause Possible solution
Tab le 4-1 . Safety shutdown messages (continued)
“Footswitch error. Cannot detect footswitch.”
“Generator overheated. Turn generator OFF and allow to cool.”
“Internal error. Turn generator OFF then ON.”
“Internal software error. Turn generator OFF then ON.”
“Limit exceeded. High dispersive electrode impedance.”
The foot switch connection is faulty.
The foot switch is faulty.
The internal temperature limit of the generator has been exceeded.
A problem has been detected with one of the generator’s hardware components.
An invalid parameter was specified by the software.
The dispersive electrode connection is faulty.
1) Verify the foot switch connection.
2) Turn the generator OFF then ON.
1) Remove or replace the foot switch.
1) Turn the generator OFF.
2) Allow the generator to cool.
3) Ensure that the fan located on the rear panel and the air intake vents on the bottom of the generator are not blocked.
4) Turn the generator ON.
1) Turn the generator OFF then ON.
1) Turn the generator OFF then ON.
1) Reattach the dispersive electrode.
2) Restart the ablation.
“Limit exceeded: High impedance. Restart ablation to continue.”
4-4 Cardioblate® 68000 Technical Manual
The dispersive electrode is faulty.
Lifting of electrode from tissue while electrode is active.
1) Replace the dispersive electrode.
2) Restart the ablation.
In Monopolar Mode:
1) Place the electrode on target tissue.
2) Reapply power.
Safety Shutdown and Troubleshooting
Safety Shutdown Messages
Tab le 4-1 . Safety shutdown messages (continued)
Error message Possible cause Possible solution
“Limit exceeded: Low impedance. Restart ablation to continue.”
Restricted irrigation flow
The dispersive electrode/ablation device connection is faulty.
Coagulum formed on the ablation e lectrode of the ablation device.
The ablation device is faulty.
Restricted irrigation flow
The dispersive electrode is improperly placed.
The ablation device is faulty.
1) Check irrigation set-up to ensure good flow to electrodes (pressure bag inflated between 150-300 mm Hg, 1000 mL bag of normal saline, all clamps on line open).
In Monopolar Mode:
1) Verify the connection of the dispersive electrode to the patient, the dispersive electrode cable to the electrode and to the generator.
2) Verify the connection of the ablation device to the generator.
3) Restart ablation.
In Bipolar Mode:
1) Replace ablation device.
2) Restart ablation.
1) Replace the ablation device.
2) Restart ablation.
1) Check irrigation set-up to ensure good flow to electrodes (pressure bag inflated between 150-300 mm Hg, 1000 mL bag of normal saline, all clamps on line open).
In Monopolar Mode:
1) Move the dispersive electrode farther from the heart.
2) Reapply power.
1) Replace the ablation device.
2) Reapply power.
Cardioblate® 68000 Technical Manual 4-5
Chapter 4
Safety Shutdown Messages
Error message Possible cause Possible solution
Tab le 4-1 . Safety shutdown messages (continued)
In Bipolar Mode:
The device electrodes are in direct contact.
1) Open the jaws of the bipolar electrodes.
2) Reposition on the tissue so that the electrodes are not in direct contact.
3) Reapply power.
“Power measurement error: P<>VI. Restart ablation to continue.”
“Power measurement error. Restart ablation to continue.”
“XXXX system error. Turn generator OFF then ON.”
“Voltage limit exceeded. Restart ablation to continue.”
The ablation device is faulty.
This occurs when the internal power calculation falls outside of the acceptable range during ablation.
This occurs when the wattage readings fall outside the acceptable range during ablation.
A problem with the generator has been detected.
The maximum voltage limit for the generator has been reached.
1) Replace the ablation device.
2) Reapply power.
1) Restart ablation to continue.
1) Restart the ablation to continue.
1) Turn the generator OFF then ON.
1) Turn the generator OFF then ON.
4-6 Cardioblate® 68000 Technical Manual

Other Messages

Message Explanation
“Ablating” RF energy is being delivered.
“Ablation stopped. Press RF button or footswitch to start ablation.”
Safety Shutdown and Troubleshooting
Other Messages
Tabl e 4 -2 . Other messages
RF energy has terminated and needs to be restarted to continue ablating.
“No devices are attached.”
“Press RF button or footswitch to start ablation.”
“Select ablation mode.”
“Transmurality reached. Power is ON”

System Shutdown

When a complete system shutdown occurs, a solid blue screen will appear on the display panel. If this occurs, please note the error message on the screen and contact technical support as listed on the rear panel or your local Medtronic representative.
The generator is ready for the device to be attached.
RF energy is not active and needs to be started to ablate.
Ablation mode, bipolar or monopolar, must be selected to activate a device. RF energy will only flow to one device at a time.
The generator algorithm notes that a transmural lesion has been obtained. A solid tone is emitted from the generator. Energy delivery will continue until the user releases the foot switch, presses the power button on the generator, releases the jaws and eliminates contact with the tissue, or the maximum time limit is reached.
Cardioblate® 68000 Technical Manual 4-7
Chapter 4
System Shutdown
4-8 Cardioblate® 68000 Technical Manual
Device Maintenance and Service
Replacing the Generator Fuse 5-2
Cleaning the Generator 5-2
Periodic Technical and Safety Inspection 5-3
Service 5-4
5
5
Cardioblate® 68000 Technical Manual 5-1
Chapter 5
Fuse

Replacing the Generator Fuse

Replacing the Generator Fuse
The Cardioblate® Generator has two 250 V 3.15 Amp. slow-blow fuses that can be replaced in the field.

Replacing the Fuse

Refer to Figure 5-1 along with the following instructions.
1. Pull out the door of the fuse housing.
2. Replace the fuses.
3. Push in the door of the fuse housing.
Figure 5-1. Replacing fuse

Cleaning the Generator

Caution: Do not clean the components of the generator with
aggressive solvents or the equipment may be damaged.
Caution: Do not immerse the components for cleaning.

Cleaning the Components of the Cardioblate® Surgical Ablation System

1. Clean the generator and its components by wiping them with a damp cloth. If necessary, use a mild detergent to remove all visible blood and body fluid contaminants.
2. Dry all of the components completely.
5-2 Cardioblate® 68000 Technical Manual
Device Maintenance and Service

Periodic Technical and Safety Inspection

Periodic Technical and Safety Inspection
Technical and safety inspections of the generator and its accessories must be performed at least once every 12 months by persons, who, based on their training, knowledge, and practical experience, are capable of adequately performing such inspections and who do not require instructions with regard to the technical and safety inspection.
Warning: If technical and safety inspection reveals a defect that could harm the patient, users, or third parties, the device should not be used until it has been properly repaired. The operator must immediately notify the appropriate authorities and Medtronic of these defects.

Visual Inspection

The following must be visually inspected:
The technical manual is available and complete.
Labels, cautions, or warnings are placed correctly and in all required locations.
The device, accessories, and all wiring show no signs of physical damage.

Operating Test

The following must be tested for functionality:
Foot switch
Touch screen: buttons and displays
Audio output
Cardioblate® 68000 Technical Manual 5-3
Chapter 5

Service

Electrical Safety Test: IEC 60601-1

Follow test instructions in the appropriate area of the standard.
Protective earth resistance: 0.1 ohm
Earth leakage current: 0.5 mA (500 µA)
Enclosure leakage current: 0.1 mA (100 µA)
Patient leakage current: 0.01 mA (10 µA)
Patient auxiliary current (dc): 0.01 mA (10 µA)
Patient auxiliary current (ac): 0.01 mA (10 µA)
Caution: USE EXTREME CAUTION WHEN WORKING AROUND TEST LEADS AND LOADS WHILE RF ENERGY IS BEING APPLIED. DO NOT TOUCH ANY EXPOSED LEADS. USE INSULATED LEADS IF POSSIBLE. RISK OF SHOCK OR TISSUE ABLATION EXISTS WITH THESE NONCLINICAL SET UPS.

Functional Test

Service
In order to perform critical functional tests of the Medtronic® Cardioblate
®
Generator Model 68000, specific software and hardware are required which is available for purchase from your local Medtronic representative. Specific instructions on how to perform these tests are included as part of the Medtronic Cardioblate
®
Diagnostic Kit (Model 60889).
®
Medtronic employs highly trained representatives and engineers located throughout the world to serve you and, upon request, to provide training to qualified hospital personnel in the use of Medtronic products. Medtronic also maintains a professional staff to provide technical consultation to product users. For medical consultation, Medtronic can often refer product users to outside medical consultants with appropriate expertise. For more information, contact your local Medtronic representative, or call or write Medtronic at the appropriate address or telephone number listed on the back cover.
5-4 Cardioblate® 68000 Technical Manual
Device Maintenance and Service
Service
The Cardioblate® Surgical Ablation System Model 68000 has been carefully engineered, manufactured and quality tested to provide long, trouble-free service. Should service or repair be necessary, contact your local Medtronic representative at the appropriate address or telephone number listed on the back cover.
Note: Ship the Cardioblate
®
Surgical Ablation System Model 68000 back to Medtronic in its original shipping container. If the original shipping container is not available, contact your Medtronic representative.
A serial number identifying each individual device is printed on the back of the device. This serial number should be referenced in any correspondence regarding this device.
Contact your local Medtronic representative for replacement parts, such as cables, footswitches.
At the end of its useful life, dispose of the Cardioblate
®
Surgical Ablation System Model 68000 in accordance with local environmental requirements.
Cardioblate® 68000 Technical Manual 5-5
Chapter 5
Service
5-6 Cardioblate® 68000 Technical Manual
Medtronic Warranty Information 6-2
Warranty6
6
Cardioblate® 68000 Technical Manual 6-1
Chapter 6

Medtronic Warranty Information

Medtronic Warranty Information

Limited Warranty for the Model 68000 Generator

A. This LIMITED WARRANTY provides the following assurance
to the purchaser of the Medtronic generator, hereafter referred to as the “Equipment,” that should the Equipment fail to function within normal tolerances due to a defect in materials or workmanship within a period of one year, commencing with the delivery of the Equipment to the purchaser, Medtronic will at its option: (a) repair or replace any defective part or parts of the Equipment; (b) issue to the purchaser a credit equal to the original Equipment purchase price (but not to exceed the value of the replacement Equipment), against the purchase of replacement Equipment, or (c) provide functionally comparable replacement Equipment at no charge.
B. To qualify for this repair, replacement or credit, the following
conditions must be met:
(1) The Equipment must be returned to Medtronic within
sixty (60) days after discovery of the defect (Medtronic may, at its option, repair the Equipment on site).
(2) The Equipment must not have been repaired or altered by
someone other than Medtronic in any way that, in the judgment of Medtronic, affects its stability and reliability.
(3) The Equipment must not have been subjected to misuse,
abuse or accident.
C. This LIMITED WARRANTY is limited to its express term. In
particular, Medtronic is not responsible for any incidental or consequential damages based on any use, defect or failure of the Equipment, whether the claim is based on warranty, contract, tort or otherwise.
®
Cardioblate® Model 68000
D. The exclusions and limitations set out above are not intended
to, and should not be construed so as to, contravene mandatory provisions of applicable law. If any part or term of this LIMITED WARRANTY is held by any court of competent jurisdiction to be illegal, unenforceable or in conflict with applicable law, the validity of the remaining portion of the LIMITED WARRANTY shall not be affected, and all rights and obligations shall be construed and enforced as if this LIMITED WARRANTY did not contain the particular part or term held to be invalid.
6-2 Cardioblate® 68000 Technical Manual

Disclaimer of Warranty for the Accessories

Although the Medtronic® Model 60883 remote control foot switch, Models 4807, 4808, and 4809 power cords, Model 05328 EGM cable, hereafter referred to as “Product,” have been carefully designed, manufactured and tested prior to sale, the Product may fail to perform its intended function satisfactorily for a variety of reasons. The warnings contained in the Product labeling provide more detailed information and are considered an integral part of the DISCLAIMER OF WARRANTY. Medtronic, therefore disclaims all warranties, both express and implied, with respect to the Product. Medtronic shall not be liable for any incidental or consequential damages caused by any use, defect, or failure of the Product, whether the claim is based on warranty, contract, tort or otherwise.
The exclusions and limitations set out above are not intended to, and should not be construed so as to contravene mandatory provisions of applicable law. If any part or term of the DISCLAIMER OF WARRANTY is held by any court of competent jurisdiction to be illegal, unenforceable or in conflict with applicable law, the validity of the remaining portion of the DISCLAIMER OF WARRANTY shall not be affected, and all rights and obligations shall be construed and enforced as if this DISCLAIMER OF WARRANTY did not contain the particular part or term held to be invalid.
Warranty
Medtronic Warranty Information
Cardioblate® 68000 Technical Manual 6-3
Chapter 6
Medtronic Warranty Information
6-4 Cardioblate® 68000 Technical Manual
A AC Power 2-7
Button 3-6, 3-23 Connection 2-7 Outlet 3-5
Accessories 1-4, 5-3
Specifications 2-12 Warnings and Precautions 1-6
Warranty 6-3 Adverse Events 1-10 Applicable Standards 2-18 Application Site 3-3–3-4 Audio 2-6
Control 2-6
Control Buttons 2-6
Output 5-3
Tones 2-9 B Bipolar Ablation Device 2-3, 2-11
Connecting 3-6–3-7
Connector 2-3
Impedance Measurement Display 2-6 Bipolar Button 2-6 Bipolar Mode 2-6
Default Settings 2-11
Delivering RF Energy 3-14–3-16
Description 1-2
Safety Shutdown Messages 4-3–4-6
IndexA
A
Cardioblate® 68000 Technical Manual A-1
Appendix A
Selecting 3-14 Selection and Use 3-14–3-16
Specifications 2-9–2-11 Burns 1-5, 1-7, 3-3 C Cardioblate Diagnostic Kit 1-4, 5-4 Cardioblate Surgical Ablation System 1-2, 1-4, 2-2, 5-2, 5-5
Components 2-2
Package Contents 1-3 Cleaning 1-7, 5-2
Components 5-2
Generator 5-2 D Dispersive Electrode 1-2, 1-4, 1-7, 1-8, 2-3, 2-6, 2-11, 3-5, 3-22
Connector 2-3
Safety Shutdown Messages 4-3–4-6
Selecting and Attaching 3-3–3-4 Display Screen 2-5–2-6, 3-2, 3-4
Description 1-2 E Electrogram (EGM)
Connection 2-3
Connector 2-3, 2-11, 3-22
Monitoring System 3-22
Warranty 6-3 Electromagnetic Compatibility Declaration 2-14–2-16 Electromagnetic Interference 1-9, 2-17 Electrostatic Discharge 1-6 Enhanced Bipolar Mode
Description 1-2
Specifications 2-10 Equipotential Ground Post 2-7, 2-11 F Failure 1-8 Fan 1-8, 2-7, 3-4 Flash Memory Device 2-4
A-2 Cardioblate® 68000 Technical Manual
Fluid Ingress 1-9 Foot Switch 1-4, 2-7, 2-8, 2-11, 3-5
Operating Test 5-3 Safety Shutdown Messages 4-4 Specifications 2-12 Warranty 6-3
Fuse 2-7, 2-8
Replacing 5-2
Specifications 2-9 G Generator 1-2, 1-3, 1-4, 1-9, 2-2, 2-8, 2-12
Cleaning 5-2
Connecting Cables 3-5
Description 1-2–1-3
Powering Down 3-23
Powering Up 3-6
Setting Up 3-4
Specifications 2-9–2-11
Warnings and Precautions 1-5–1-8 H High Impedance 1-8
Safety Shutdown Messages 4-4 I Impedance 1-2, 1-3, 2-10, 2-12
Default Settings 2-11
Limits 2-10
Measurement Display 2-6 Implanted Devices 1-6 Information Button 2-6 L Left Side Panel 2-4 Lesions 1-6 Low Impedance 1-8
Safety Shutdown Messages 4-5 M Message Box 2-6
Index
Cardioblate® 68000 Technical Manual A-3
Appendix A
Monopolar Ablation Device 2-3, 2-11, 3-5
Connecting 3-6–3-7 Connector 2-3
Impedance Measurement Display 2-6 Monopolar Button 2-6 Monopolar Mode 3-9
Adjusting the RF Timer Limits 3-11–3-12
Adjusting the Setpoint Values and Limits 3-10–3-11
Default Settings 2-11
Delivering RF Energy 3-12–3-13
Description 1-2
Selecting 3-9
Selection and Use 3-9–3-13
Specifications 2-9–2-11 O ON/OFF Switch 2-7, 2-8 Optional Connections 3-22 Options 2-6, 3-17–3-22 Options Menu 2-6, 3-17–3-21
Advanced 3-19–3-21 Output Power Diagrams 2-12–2-13 P Package Contents 1-3 Power Cord 1-4, 2-8, 3-5
Specifications 2-12
Warranty 6-3 Power Entry Connector 2-8, 3-5 Q Quick Reference Guide 3-2 R Rear Panel 2-7–2-8, 3-6, 4-2
Connectors 2-11 RF Energy 1-2, 1-3, 1-4, 1-6, 2-6, 4-2
Button 2-6
Delivering in Bipolar Mode 3-14–3-16
Delivering in Monopolar Mode 3-12–3-13
A-4 Cardioblate® 68000 Technical Manual
Modes 1-2
Warnings and Precautions 1-5–1-9 RF Panel Screen 3-6 RF Power Measurement Display 2-6 RF Power Output
Functional Characteristics 2-10 RF Timer Measurement Display 2-5 Right Side Panel 2-3 S Safety Shutdowns 1-3, 4-7
Conditions 4-2
Messages 4-3–4-6 Service 5-4–5-5 Speaker Volume Level 2-6 Specifications 2-9–2-12 Storage Temperature 1-9 T Therapy Connector Panel 2-3 Total RF time 2-5 U USB Port 2-7
Location 2-4 User Interface 1-2
Index
Cardioblate® 68000 Technical Manual A-5
Appendix A
A-6 Cardioblate® 68000 Technical Manual
Europe
Europe/Africa/Middle East Headquarters
Medtronic International Trading Sàrl Route du Molliau 31 Case Postale 84 CH - 1131 Tolochenaz Switzerland Internet: www.medtronic.co.uk Tel. 41-21-802-7000 Fax 41-21-802-7900
Authorized Representative in the European Community
Medtronic B.V. Earl Bakkenstraat 10 6422 PJ Heerlen The Netherlands Tel. 31-45-566-8000 Fax 31-45-566-8668
Asia-Pacific
Japan
Medtronic Japan Comodio Shiodome 5F 2-14-1 Higashi-Shimbashi, Minato-ku Tokyo 105-0021 Japan Tel. 81-3-6430-2011 Fax 81-3-6430-7140
Australia
Medtronic Australasia Pty. Ltd. 97 Waterloo Road North Ryde NSW 2113 Australia Tel. 61-2-9857-9000 Fax 61-2-9878-5100
Asia
Medtronic International Ltd. Suite 1602 16/F, Manulife Plaza The Lee Gardens, 33 Hysan Avenue Causeway Bay Hong Kong Tel. 852-2891-4068 Fax 852-2591-0313
Americas
Latin America
Medtronic Latin America 3750 NW 87th Avenue Suite 700 Miami, FL 33178 Tel. 305-500-9328 Fax 786-709-4244
Canada
Medtronic of Canada Ltd. 99 Hereford Street Brampton, Ontario L6Y 0R3 Canada Tel. 905-460-3800 Fax 905-826-6620 Toll-free: 1-800-268-5346
United States
Manufacturer:
Medtronic, Inc. 710 Medtronic Parkway Minneapolis, MN 55432 USA Internet: www.medtronic.com Tel. 763-514-4000 Fax 763-391-9100 Toll-free: 1-800-328-2518 (24-hour consultation service)
© 2006, 2007, 2011 Medtronic M948371A001 Rev. 1A
*M948371A001*
Loading...