Medtronic 60831 Instructions for Use

CARDIOBLATE® BP2
60831
Instructions for Use
Caution: Federal law (USA) restricts
this device to sale by or on the order of a physician.
Explanation of symbols on package labeling
Refer to the device labeling to see which symbols apply to this product
Nonpyrogenic
Sterilized Using Ethylene Oxide
Do Not Reuse
Do Not Resterilize
Do not use if package is damaged
Use By
Lot Number
Quantity
Open Here
Consult Instructions for Use
For US Audiences Only
Upper Limit of Temperature
Humidity Limitation
Date of Manufacture
Catalog Number
Manufacturer
Type CF Applied Part
Contains di(2-ethylhexyl)phthalate (DEHP)
Authorized Representative in the European Community
1
1
3
4
5
6
7
9
8
2
Figure 1
Adjustable parallel jaws, 7 cm (2.76 in) in length
1.
2. Seven cm (2.76 in) electrodes with embedded weeping polymer, 5 cm (1.97 in) of which is malleable
3. 300° rotation of jaw assembly
4. Flexible neck
5. Trigger
6. Trigger lock with raised thumb slide
7. Molded handle
8. Saline fluid delivery tubing 304.8 cm (10 ft) in length with a female luer for connection to standard IV tubing systems
9. Electrical cable 304.8 cm (10 ft) in length that terminates into a multi­pin connector
2
CARDIOBLATE® BP2

Surgical Ablation Device

Model
60831

Description

The Cardioblate® BP2 Surgical Ablation Device is a hand-held, single-
bipolar, radiofrequency ablation device for use in cardiac surgery. It
use, has a saline irrigation system to deliver fluid at the contact point between the tissue and electrode to cool the tissue during radiofrequency energy delivery. The device is intended for intermittent operation.
Sterile, nonpyrogenic, disposable, single use only.

Indications for Use

The Cardioblate® BP2 Surgical Ablation Device is intended to ablate cardiac tissue during cardiac surgery using radiofrequency energy.

Contraindications

The Cardioblate® BP2 Surgical Ablation Device should not be used for:
patients that have active endocarditis at time of surgery
ablation in a pool of blood (eg, through a purse string suture on a beating heart) (Effects of this type of ablation are unknown.)

Adverse Effects

Possible adverse effects related to the ablation of cardiac tissue in combination with open heart surgery are:
tissue perforation
extension of extracorporeal bypass
perioperative heart rhythm disturbances (atrial and/or ventricular)
postoperative embolic complications
pericardial effusion or tamponade
injury to the great vessels
valve leaflet damage
conduction disturbances (SA/AV node)
acute ischemic myocardial event
thrombus formation
nerve damage
unintentional burns
pericarditis
pleural effusion
esophageal perforation
death
coronary sinus perforation
coronary artery spasm
atrial lead dislodgement
hypotension
pulmonary vein stenosis
cerebrovascular accident
transient ischemic attack
blood loss

Warnings

Do not activate device if saline is not flowing freely. The cooling of the outer tissue prevents the charring and tissue desiccation that may occur when applying radiofrequency energy without irrigation.
Instructions for Use English 3
The device may be damaged before or during the procedure by improper handling or other intervening acts. If damaged, the device may fail to function properly, and may result in the following medical complications including, but not limited to:
damage to the physiological conduction system
peripheral tissue ablation due to breach of insulation covering
Device shall be removed from the patient before defibrillation.
representation or warranty is made that failure or cessation of function
No of the device will not result in an adverse event, or that medical complications (including perforation of cardiac tissue) will not follow the procedure, or that the use of the device will in all cases restore adequate cardiac function.

Precautions

Caution: Federal law (USA) restricts this device to sale by or on the order
of a physician. Proper surgical procedures and techniques are necessarily the
responsibility of the medical professional. Each surgeon must evaluate the appropriateness of any procedure based on his or her own medical training and experience, and the type of surgical procedure.
Patient and procedure selection is solely a medical responsibility and the outcome is dependent on many variables including patient pathology, surgical and perfusion procedures.
Use caution to avoid trauma to tissues not within the target area of ablation.
Purge air from system to ensure saline fluid flows from the electrodes during use.
Separation of electrodes from tissue contact during activation will result in a high impedance shutdown placing generator power condition to OFF. Placing electrodes in direct contact during radiofrequency energy activation results in a low impedance shut down and places the generator power condition to OFF. In both situations, three short audible alarm beeps will be heard. Refer to the Cardioblate generator technical manual for reset instructions.
Maximum storage temperature: 40°C (104°F) Storage relative humidity: 5% to 85%, noncondensing This device was designed for single patient use only. Do not reuse,
reprocess, or resterilize this product. Reuse, reprocessing, or resterilization may compromise the structural integrity of the device and/or create a risk of contamination of the device, which could result in patient injury, illness, or death. Multiple uses can result in the occlusion of the irrigation openings and affect the device’s performance. To prevent this from occurring, the generator has the capability to detect previously used ablation devices and will not allow their reuse.
It is the responsibility of the user to dispose of the devices in accordance with local regulations and hospital procedures.
Due to the presence of phthalates in the product, the clinician must weigh the medical benefits of product use against the drawbacks of phthalate exposure for male children and pregnant or nursing women.

Connections to External Devices

1. Connection to the Cardioblate generator is as follows: a. Position the multi-pin connector of the bipolar device so the
alignment indicator matches the generator port indicator.
b. Insert the multi-pin connector of the bipolar device into the device
port on the generator until it locks in place. Do not force the connection.
Note: Connect to RF generators (only IEC / EN 60601-1 Type CF) with the following specifications:
4 Instructions for Use English
RF power output (Transmurality
Mode) RF power output fre-
quency Maximum RF volt-
age, between elec­trodes
Note: If additional devices are to be used, refer to their specific instructions for connections to external devices.
Sterile normal saline, commercially available pressure cuff, and IV
2. tubing set are required for use with the bipolar device. These are not included with the Cardioblate® BP2 Surgical Ablation Device.
Model 60880/60890 Model 68000
25 - 40 W (automatically varying)
479 - 489 kHz 468 - 478 kHz
159 Vrms 180 Vrms ±10%
15 - 40 W (automatically vary-
ing)

Instructions for Use

1. Inspect the package and device to ensure the expiration date has not passed and damage has not occurred to the device during shipping and handling. Do not use the device if the packaging or device is damaged or if the expiration date has elapsed. Open the package and transfer the device onto the sterile field utilizing aseptic technique.
2. Uncoil both the saline delivery tubing and the electrical cable carefully to avoid kinking and tangling.
3. Attach the saline delivery tubing female luer connector to an IV tubing set and 1000 mL bag of sterile normal saline. (0.9% normal saline is recommended because it provides a low impedance pathway for the radiofrequency energy to the tissue.)
4. If both a bipolar and monopolar device are being used, remove the sterile caps from the “Y” connector (Model 10004D) provided and attach the female luer of both devices. Then insert the IV line into the “Y” connector.
Figure 2.
Insert a 1000 mL bag of 0.9% normal saline into commercially
5. available pressure cuff and inflate to between 150 and 300 mm Hg.
6. Open the IV line and squeeze the trigger to purge the system of air until fluid is seen coming from the electrodes on the bipolar device. Release the trigger to stop IV flow.
Instructions for Use English 5
1
2
2
1. 7 cm (2.76 in) active electrodes
2.
5 cm (1.97 in) malleable
Figure 3.
7. a. Attach the bipolar electrical cable to the generator per instructions outlined in “Connections to External Devices.”
b. Shape
the electrodes as required for the targeted tissue. Engage the lock and squeeze the trigger to bring the electrodes together. Grip the electrodes with both hands to shape them. Care should be taken to ensure the electrodes are as parallel as possible. If the electrodes are out of alignment greater than 4 mm, the ability of the device to create a transmural lesion across the entire electrode length may be affected.
c. Release the lock by moving it backward and open the electrodes.
Use the flexible neck and 300° roll to assist in positioning the electrodes on the targeted tissue. Once the electrodes are correctly positioned, engage the lock and squeeze the trigger to bring the electrodes in full contact with the tissue. The lock must be engaged to ensure accurate transmurality readings.
d. Confirm the generator selector is placed on the bipolar device.
8. Prior to tissue ablation: a. Squeeze the trigger and make certain that saline is flowing freely
out of the electrodes. Continuous flow of saline is necessary during ablation to cool the outer tissue. Cooling of the outer tissue prevents charring and tissue desiccation and allows radiofrequency energy to be delivered deeper into the tissue to more effectively create transmural lesions.
Note: While irrigation increases the effectiveness of radiofrequency ablation, ablating in a pool of fluid can have a detrimental effect. Radiofrequency energy can shunt between the two electrodes and the transmurality algorithm may be less effective. Suctioning of excessive fluid at the ablation site is recommended to prevent this from occurring.
b. When using a bipolar device, the radiofrequency ablation
generator will default to the TRANSMURALITY control mode.
c. In the TRANSMURALITY mode, all generator parameters will be
automatically adjusted.
9. The current is activated in the device using the RF energy button or the foot switch provided with the Cardioblate generator. The current flows between the electrodes.
Table 1. Anticipated Energy Delivery at Various Tissue Thicknesses
Tissue Thickness (7 cm length) Energy Delivered (W)
10. The lesion becomes visible by a white coloration on the tissue. In TRANSMURALITY once a transmural lesion is created. This is the indication to the surgeon to stop ablating by releasing the foot switch or pressing the RF energy power button on the generator.
1 mm 439±40 3 mm 505±25 5 mm 557±28
control mode, the generator will emit a solid tone
6 Instructions for Use English
0
2
0
3
04
0
5
06
07
08
0
9
0
01
0
2000
4000
6000
8000
10000
12000
14000
16000
18000
20000
0
1
5
1
02
5
2
03
53
04
54
05
1
3
2
3
1
1. Average Impedance (Ohms)
Time (msec)
2.
3. Power (Watts)
Figure 4. Typical Tissue Impedance Values with
Corresponding Power Levels.
(Approximate total energy
delivered 570 W)
11. Ablating beyond recommended ablation times may result in tissue perforation.
Note: To prevent the porous jaws from drying out and clogging, maintain the jaws immersed in saline or covered with a wet sponge between ablations.
12.
Upon completion of the surgical procedure, disconnect the device from the generator and discard.

Clinical Data on Conduction Block

In vitro testing was conducted creating lesion set on porcine atria. In vivo testing was performed on porcine cardiac tissue for the Cardioblate BP2 Surgical Ablation Device. Transmural lesions were demonstrated with a 95/95 confidence level. The testing resulted in the following lesion dimensions:
n=3
Max Transmurality Depth (mm) 2.4 ± 2.3 Lesion Length (cm) 3.2 ± 0.6 Lesion Width (mm) 2.8 ± 0.5
These results are consistent with a small, retrospective study conducted
two investigational centers of 69 patients during 2006-2007. Thirty-five
at of those patients received cardiac tissue ablation using irrigated radiofrequency (RF) energy. Of the 35 patients, 29 (82.9%) underwent conduction block assessment confirming transmurality of cardiac tissue lesions. The remaining 6 patients were not assessed.
Instructions for Use English 7
The following disclaimer of warranty applies to United States customers only:

Disclaimer of Warranty

ALTHOUGH THE CARDIOBLATE® BP2 SURGICAL ABLATION DEVICE, HEREAFTER REFERRED TO AS “PRODUCT” HAS BEEN MANUFACTURED UNDER CAREFULLY CONTROLLED CONDITIONS, MEDTRONIC HAS NO CONTROL OVER THE CONDITIONS UNDER WHICH THIS PRODUCT IS USED. MEDTRONIC, THEREFORE DISCLAIMS ALL WARRANTIES, BOTH EXPRESS AND IMPLIED, WITH RESPECT TO THE PRODUCT, INCLUDING, BUT NOT LIMITED TO, ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. MEDTRONIC FOR ANY MEDICAL EXPENSES OR ANY DIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES CAUSED BY ANY USE, DEFECT, FAILURE OR MALFUNCTION OF THE PRODUCT, WHETHER A CLAIM FOR SUCH DAMAGES IS BASED UPON WARRANTY, CONTRACT, TORT OR OTHERWISE. NO PERSON HAS ANY AUTHORITY TO BIND MEDTRONIC TO ANY REPRESENTATION OR WARRANTY WITH RESPECT TO THE PRODUCT.
The exclusions and limitations set out above are not intended to, and should not be construed so as to, contravene mandatory provisions of applicable law. If any part or term of this Disclaimer of Warranty is held to be illegal, unenforceable or in conflict with applicable law, by a court of competent jurisdiction, the validity of the remaining portions of this Disclaimer shall be construed and enforced as if this Disclaimer of Warranty did not contain the particular part or term held to be invalid.
SHALL NOT BE LIABLE TO ANY PERSON OR ENTITY
of Warranty shall not be affected, and all rights and obligations
8 Instructions for Use English
Europe
*M940537A001*
Europe/Africa/Middle East Headquarters
Medtronic International Trading Sàrl Route du Molliau 31 Case Postale 84 CH - 1131 Tolochenaz Switzerland Internet: www.medtronic.co.uk Tel. 41-21-802-7000 Fax 41-21-802-7900
Authorized Representative in the European Community
Medtronic B.V. Earl Bakkenstraat 10 6422 PJ Heerlen The Netherlands Tel. 31-45-566-8000 Fax 31-45-566-8668
Asia-Pacific
Japan
Medtronic Japan Comodio Shiodome 5F 2-14-1 Higashi-Shimbashi, Minato-ku Tokyo 105-0021 Japan Tel. 81-3-6430-2011 Fax 81-3-6430-7140
Australia
Medtronic Australasia Pty. Ltd. 97 Waterloo Road North Ryde NSW 2113 Australia Tel. 61-2-9857-9000 Fax 61-2-9878-5100
Asia
Medtronic International Ltd.
1602 16/F, Manulife Plaza
Suite The Lee Gardens, 33 Hysan Avenue Causeway Bay Hong Kong Tel. 852-2891-4068 Fax 852-2591-0313
Americas
Latin America
Medtronic Latin America 3750 NW 87th Avenue Suite 700 Miami, FL 33178 USA Tel. 305-500-9328 Fax 786-709-4244
Canada
Medtronic of Canada Ltd. 99 Hereford Street Brampton, Ontario L6Y 0R3 Canada Tel. 905-460-3800 Fax 905-826-6620 Toll-free: 1-800-268-5346
United States
Manufacturer: Medtronic, Inc. 710 Medtronic Parkway Minneapolis, MN 55432 USA Internet: www.medtronic.com Tel. 763-514-4000 Fax 763-391-9100 Toll-free: 1-800-328-2518 (24-hour consultation service)
© 2006, 2009, 2011 Medtronic M940537A001 Rev. 1B