Medtronic 60813 Instructions for Use

CARDIOBLATE
Surgical Ablation Pen
®
Instructions for Use
Caution: Federal law (USA) restricts
this device to sale by or on the order of a physician.
Explanation of symbols on package labeling
Refer to the device labeling to see which symbols apply to this product
Nonpyrogenic
Sterilized Using Ethylene Oxide
Do Not Reuse
Do Not Resterilize
Use By
Lot Number
Quantity
Open Here
Consult Instructions for Use
For US Audiences Only
Upper Limit of Temperature
Humidity Limitation
Date of Manufacture
Catalog Number
Manufacturer
Contains di(2-ethylhexyl)phthalate (DEHP)
Authorized Representative in the European Community
Do not use if package is damaged
Type CF Applied Part
1
Cardioblate® Surgical Ablation Pen
7
Figure 1
1.
Molded acrylic handle with a silicone finger grip
2. Malleable stainless steel shaft with a silicone/fluoropolymer sheath
8.0 cm (3.15 in) long
3. Rounded distal tip features multiple fluid openings
4. Antidrip transition piece prevents fluid from running back onto the handle
5. Handle terminates into an integral saline fluid delivery tubing
304.8 cm (10 ft) in length that is molded to the electrical cable and detachable to the desired free length, with a female luer for connection to standard IV tubing systems
6. Handle also features an integral electrical cable 304.8 cm (10 ft) in length that terminates into a multi-pin connector
7. Pinch clamp
2
CARDIOBLATE
®

Surgical Ablation Pen

Model
60813

Description

The Cardioblate® Surgical Ablation Pen is a hand-held, single-use, monopolar, radiofrequency ablation device for use in cardiac surgery. It
a saline irrigation system to deliver fluid at the contact point between
has the tissue and electrode tip to cool the tissue during radiofrequency energy delivery. The device is intended for intermittent operation.
Sterile, nonpyrogenic, disposable, single use only.

Indications for Use

The Cardioblate® Surgical Ablation Pen is intended to ablate cardiac tissue during cardiac surgery using radiofrequency energy.

Contraindications

The Cardioblate® Surgical Ablation Pen should not be used for patients that have active endocarditis at time of surgery.

Adverse Effects

Possible adverse effects related to the creation of spot or linear lesions in cardiac tissue in combination with open heart surgery are:
tissue perforation
extension of extracorporeal bypass
perioperative heart rhythm disturbances (atrial and/or ventricular)
postoperative embolic complications
pericardial effusion or tamponade
injury to the great vessels
valve leaflet damage
conduction disturbances (SA/AV node)
acute ischemic myocardial event
thrombus formation
nerve damage
unintentional burns
pericarditis
pleural effusion
esophageal perforation
death
coronary sinus perforation
coronary artery spasm
atrial lead dislodgement
hypotension
cerebrovascular accident
transient ischemic attack
blood loss

Warnings

Do not activate ablation if saline is not flowing freely. Cooling of the outer tissue prevents the charring and tissue desiccation that may occur when applying radiofrequency energy without irrigation.
The device may be damaged before or during the procedure by improper handling or other intervening acts. If damaged, the device may fail to function properly, and may result in the following medical complications including, but not limited to:
tissue “popping” and perforation
Instructions for Use English 3
damage to the physiological conduction system
peripheral tissue ablation due to breach of insulation covering
The device shall be removed from the patient before defibrillation.
representation or warranty is made that failure or cessation of function
No of the device will not result in an adverse event, or that medical complications (including perforation of cardiac tissue) will not follow the procedure, or that the use of the device will in all cases restore adequate cardiac function.

Precautions

Caution: Federal law (USA) restricts this device to sale by or on the order
of a physician. Proper surgical procedures and techniques are the responsibility of the
medical professional. Each surgeon must evaluate the appropriateness of any procedure based on their own medical training and experience, and the type of surgical procedure.
Patient and procedure selection is solely a medical responsibility and the outcome is dependent on many variables including patient pathology, and surgical and perfusion procedures.
Use caution to avoid trauma to tissues not within the target area of ablation.
Make certain that saline fluid flows from the distal tip ports during use. Follow recommended power settings (step 8) for optimal ablation product
performance. Separation of tip from tissue contact during activation of RF ablation will
result in high impedance shutdown switching the generator power condition to STANDBY. Three short audible alarm beeps from the generator indicate a high impedance occurrence. Refer to the Cardioblate generator technical manual for reset instructions.
Maximum storage temperature: 40°C (104°F) Storage relative humidity: 5% to 85%, noncondensing This device was designed for single patient use only. Do not reuse,
reprocess, or resterilize this product. Reuse, reprocessing, or resterilization may compromise the structural integrity of the device and/or create a risk of contamination of the device, which could result in patient injury, illness, or death. Multiple uses can result in the occlusion of the irrigation openings and affect the device’s performance.
It is the responsibility of the user to dispose of the devices in accordance with local regulations and hospital procedures.
Due to the presence of phthalates in the product, the clinician must weigh the medical benefits of product use against the drawbacks of phthalate exposure for male children and pregnant or nursing women.

Connections to External Devices

Note: Do not attempt to reuse the ablation device with the Cardioblate
generator. The generator has the capability to detect previously used ablation devices and will not allow their reuse.
1. Connection to the Cardioblate generator is as follows: a. Position the multi-pin connector of the pen so the alignment
indicator matches the monopolar generator port indicator.
b. Insert the multi-pin connector of the pen into the device port on
the generator until it locks in place. Do not force the connection. Note: Connect to RF generators (only IEC / EN 60601-1
Type CF) with the following specifications:
RF power output setpoint/limit
Model 60880/60890 Model 68000
25 - 50 W (manually variable)
25 - 50 W (manually varia-
ble)
4 Instructions for Use English
RF power output frequency
Maximum RF volt­age, between elec­trodes
Note: If additional devices are to be used, refer to their specific directions for connection information.
Sterile normal saline, commercially available pressure bag, and IV
2. tubing set are required.
Model 60880/60890 Model 68000
479 - 489 kHz 468 - 478 kHz
159 Vrms 180 Vrms ±10%

Instructions for Use

1. Inspect the package and device to ensure the expiration date has not passed and no damage occurred to the device during shipping and handling. Do not use the device if the packaging or device is damaged or if the expiration date has elapsed. Open the package and transfer the device onto the sterile field utilizing an aseptic technique.
2. Place a large indifferent electrode pad with electrical cable securely on the back of the patient’s body. (Refer to generator manual and indifferent electrode instructions for use for specific instructions on electrode placement and skin care.)
3. Attach the electrical cable from the indifferent electrode securely to the appropriate patient grounding connection site on the Cardioblate generator. (Refer to the Cardioblate generator technical manual.)
4. Uncoil both the saline delivery tubing and the electrical cable carefully to avoid kinking and tangling.
5. Attach the saline delivery tubing female luer connector to an IV tubing set and sterile normal saline bag. (0.9% normal saline is recommended because it provides a low impedance pathway for the radiofrequency energy to the tissue.)
If both a bipolar and monopolar device are being used, attach the female luer of both devices to the “Y” connector (Model 10004D) provided. Then insert the IV line into the “Y” connector.
6. Place the sterile normal saline bag into the pressure bag. Inflate the pressure bag to between 150 and 300 mm Hg. Purge the system of air until fluid is seen coming from the openings on the distal tip of the device (Figure 2). Clamp off tubing set until ready for use.
Figure 2.
7. Attach
8. Prior to tissue ablation:
the pen electrical cable to the Cardioblate generator connector
per instructions outlined in “Connections to External Devices.”
a. Open the IV line and make certain that saline is flowing freely out
of the openings in the distal tip of the device. Continuous flow of saline is necessary during ablation to cool the outer tissue. Cooling of the outer tissue prevents charring and tissue desiccation and allows radiofrequency energy to be delivered deeper into the tissue to more effectively create transmural lesions.
b. Set the radiofrequency ablation generator to the POWER mode
and set the power between 20-30 watts maximum output. Ideal power settings should be determined by the surgeon during tissue ablation.
Instructions for Use English 5
9. The distal tip of the device is dragged gently across the targeted intracardiac tissue under direct visualization by the surgeon. The current is activated in the device using the foot switch provided with
Cardioblate generator. The current flows between the ablation tip
the and the large indifferent electrode on the patient’s body.
10. The lesion becomes visible as a white coloration on the tissue. The tip is moved in an oscillating motion at a recommended speed of 1 cm/sec. The surgeon will determine depth, width, and location of the ablation.
11. Do not dwell in a target area too long; sudden tissue “popping” and/or perforation may occur.
12. Upon completion of the surgical procedure, disconnect the device and indifferent electrode from the Cardioblate generator and discard.

Clinical Data on Conduction Block

In vitro testing was conducted creating lesion sets on bovine atria. In vivo testing was performed on porcine cardiac tissue for the Cardioblate Surgical Ablation Pen. Transmural lesions were demonstrated with a 95/95 confidence level. The testing resulted in the following lesion dimensions:
Table 1. Anticipated Lesion Depth at Various Power Levels
95% Confidence
interval for 10 sec-
ond oscillation,
1 cm
2.3 – 3.5 mm 3.5 – 4.4 mm 20
2.9 – 3.7 mm 4.7 – 5.7 mm 30
These results are consistent with a small, retrospective study conducted
two investigational centers of 69 patients during 2006-2007. Thirty-five
at of those patients received cardiac tissue ablation using irrigated radiofrequency (RF) energy. Of the 35 patients, 29 (82.9%) underwent conduction block assessment confirming transmurality of cardiac tissue lesions. The remaining 6 patients were not assessed.
95% Confidence
interval for 20 sec-
ond oscillation,
1 cm
Power (W) @
150 - 300 mm Hg pres-
on saline irrigation
sure
6 Instructions for Use English
The following disclaimer of warranty applies to United States customers only:

Disclaimer of Warranty

ALTHOUGH THE CARDIOBLATE® SURGICAL ABLATION PEN, HEREAFTER REFERRED TO AS “PRODUCT” HAS BEEN MANUFACTURED UNDER CAREFULLY CONTROLLED CONDITIONS, MEDTRONIC HAS NO CONTROL OVER THE CONDITIONS UNDER WHICH THIS PRODUCT IS USED. MEDTRONIC, THEREFORE DISCLAIMS ALL WARRANTIES, BOTH EXPRESS AND IMPLIED, WITH RESPECT TO THE PRODUCT, INCLUDING, BUT NOT LIMITED TO, ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. MEDTRONIC FOR ANY MEDICAL EXPENSES OR ANY DIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES CAUSED BY ANY USE, DEFECT, FAILURE OR MALFUNCTION OF THE PRODUCT, WHETHER A CLAIM FOR SUCH DAMAGES IS BASED UPON WARRANTY, CONTRACT, TORT OR OTHERWISE. NO PERSON HAS ANY AUTHORITY TO BIND MEDTRONIC TO ANY REPRESENTATION OR WARRANTY WITH RESPECT TO THE PRODUCT.
The exclusions and limitations set out above are not intended to, and should not be construed so as to, contravene mandatory provisions of applicable law. If any part or term of this Disclaimer of Warranty is held to be illegal, unenforceable or in conflict with applicable law, by a court of competent jurisdiction, the validity of the remaining portions of this Disclaimer shall be construed and enforced as if this Disclaimer of Warranty did not contain the particular part or term held to be invalid.
SHALL NOT BE LIABLE TO ANY PERSON OR ENTITY
of Warranty shall not be affected, and all rights and obligations
The following disclaimer of warranty applies to customers outside the United States:
Disclaimer of Warranty
ALTHOUGH THE CARDIOBLATE® SURGICAL ABLATION PEN, HEREAFTER REFERRED TO AS “PRODUCT” HAS BEEN CAREFULLY DESIGNED, MANUFACTURED AND TESTED PRIOR TO SALE, THE PRODUCT MAY FAIL TO PERFORM ITS INTENDED FUNCTION SATISFACTORILY FOR A VARIETY OF REASONS. THE WARNINGS CONTAINED IN THE PRODUCT LABELING PROVIDE MORE DETAILED INFORMATION AND ARE CONSIDERED AN INTEGRAL PART OF THIS DISCLAIMER OF WARRANTY. MEDTRONIC, THEREFORE, DISCLAIMS ALL WARRANTIES, BOTH EXPRESS AND IMPLIED, WITH RESPECT TO THE PRODUCT. MEDTRONIC SHALL NOT BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES CAUSED BY ANY USE, DEFECT, OR FAILURE OF THE PRODUCT, WHETHER THE CLAIM IS BASED ON WARRANTY, CONTRACT, TORT OR OTHERWISE.
The exclusions and limitations set out above are not intended to, and should not be construed so as to, contravene mandatory provisions of applicable law. If any part or term of this Disclaimer of Warranty is held by any court of competent jurisdiction to be illegal, unenforceable or in conflict with applicable law, the validity of the remaining portion of the Disclaimer shall be construed and enforced as if this Disclaimer of Warranty did not contain the particular part or term held to be invalid.
of Warranty shall not be affected, and all rights and obligations
Instructions for Use English 7
Europe
*M940535A001*
Europe/Africa/Middle East Headquarters
Medtronic International Trading Sàrl Route du Molliau 31 Case Postale 84 CH - 1131 Tolochenaz Switzerland Internet: www.medtronic.co.uk Tel. 41-21-802-7000 Fax 41-21-802-7900
Authorized Representative in the European Community
Medtronic B.V. Earl Bakkenstraat 10 6422 PJ Heerlen The Netherlands Tel. 31-45-566-8000 Fax 31-45-566-8668
Asia-Pacific
Japan
Medtronic Japan Comodio Shiodome 5F 2-14-1 Higashi-Shimbashi, Minato-ku Tokyo 105-0021 Japan Tel. 81-3-6430-2011 Fax 81-3-6430-7140
Australia
Medtronic Australasia Pty. Ltd. 97 Waterloo Road North Ryde NSW 2113 Australia Tel. 61-2-9857-9000 Fax 61-2-9878-5100
Asia
Medtronic International Ltd.
1602 16/F, Manulife Plaza
Suite The Lee Gardens, 33 Hysan Avenue Causeway Bay Hong Kong Tel. 852-2891-4068 Fax 852-2591-0313
Americas
Latin America
Medtronic Latin America 3750 NW 87th Avenue Suite 700 Miami, FL 33178 USA Tel. 305-500-9328 Fax 786-709-4244
Canada
Medtronic of Canada Ltd. 99 Hereford Street Brampton, Ontario L6Y 0R3 Canada Tel. 905-460-3800 Fax 905-826-6620 Toll-free: 1-800-268-5346
United States
Manufacturer: Medtronic, Inc. 710 Medtronic Parkway Minneapolis, MN 55432 USA Internet: www.medtronic.com Tel. 763-514-4000 Fax 763-391-9100 Toll-free: 1-800-328-2518 (24-hour consultation service)
© 2004, 2008, 2011 Medtronic M940535A001 Rev. 1B
Loading...