ACCELERATE
GRAFT DELIVERY
SYSTEM WITH
GRAFTON™ DBF
USER GUIDE
Graft material inside of Cannula shown for visual ization
Page 2
TABLE OF
CONTENTS
3Introduction
4Instrument and Implant Information
4 Instruments and Implants
4 Technical Specifications
5 Procedure
5 Tips
6 Graft Preparation
™
7 Filling the Accelerate
8 Loading the Accelerate
9 Attaching the Accelerate
10 Delivery of the Bone Graft Mixture
11 Reloading and Disassembly
Cannula with the Bone Graft Mixture
™
Cannula into the Accelerate™ Funnel
™
Funnel to the Accelerate™ Body
12 Graft Recovery
13Product Ordering Information
14Instructions for Use (IFU)
2
Page 3
INTRODUCTION
The Accelerate™ Graft Delivery System was developed and designed to empower the healthcare professional to leverage
the delivery of autograft combined with Grafton™ DBF. The seamless, simplified, and integrated delivery of bone graft with
the Accelerate™ Graft Delivery System enables controlled delivery in an efficient workflow. This User Guide is designed to
familiarize healthcare professionals with the use of Accelerate™ Instrument Set. Please carefully read this User Guide and
its appendix prior to use of the instruments and implants.
The development and design of the Accelerate™ Graft Delivery System revolved around a comprehensive solution to bone
grafting that focused on – the instrument, the implant, and the workflow.
Accelerate™ Design Pillars
LEVERAGE
AUTOGRAFT
EFFICIENT
WORKFLOW
Risks
See IFU for a full list of risks, warnings, and contraindications.
CONTROLLED
DELIVERY
3
Page 4
INSTRUMENT AND IMPLANT
INFORMATION
Instruments and Implants
Distal endProximal end
Accelerate
Cannula
Accelerate
™
Funnel
™
Graft Delivery Loading Platform
Pin
open
closed
Socket
Standard Bone Funnel
Accelerate
Accelerate
™
Graft Recovery Tool
™
Graft Recovery Platform
Clasp
Plunger
Release mechanism
TriggerBack handle
Accelerate
™
Funnel Pin
Technical Specications
Cannula
Cannula Bone Graft Capacity6.0cc
Accelerate
Accelerate™ Funnel Outer Diameter7.9mm
Accelerate
™
Funnel
™
Funnel Shaft Length7.1"
Accelerate
4
™
Body
Standard Bone Funnel
Standard Bone Funnel Inner Diameter6.85mm
Standard Bone Funnel Outer Diameter8.5mm
Standard Bone Funnel Shaft Length7.3"
Page 5
PROCEDURE
Tips
1 The Accelerate™ Instrument will clog if the bone chip
size exceeds 3.2mm in diameter.
2 Use Accelerate™ Graft Delivery Instrument as intended
only with Grafton™ DBF. Use with other products
(mineralized allograft, synthetics, dierent bone
graft formulations, etc.) will likely result in decreased
performance, clogging, and possible permanent
malfunction.
™
3 Only use Accelerate
BG Set T502XX Grafton™
DBF. The original T501XX Grafton™ DBF will not be
compatible as it does not contain the Cannula.
™
4 The Accelerate
Graft Delivery Instrument will not
work eciently and clog if used without the Cannula.
™
5 The Accelerate
Graft Delivery Instrument may
not function as intended (and is likely to clog) if the
autograft content of the nal bone grafting mixture
exceeds 50% of the total volume.
6 Close and open the Cannula once or twice to make
the closing of the Cannula easier once it has been
loaded with bone graft mixture.
7 If the Cannula won’t completely close, redistribute or
remove small amounts of bone grafting mixture until
the Cannula completely closes. Overlling of Cannula
increases the likelihood of clogging.
™
8 Do not force the Cannula into the Accelerate
Funnel.
If you experience resistance loading the Cannula into
the Accelerate™ Funnel, redistribute or remove small
amounts of bone grafting mixture.
™
9 The Accelerate
Instrument Set contains two
Accelerate™ Funnels for simultaneous preloading
of Cannulas with bone graft mixture.
™
10 If the Cannula (inside of the Accelerate
Funnel) is
hitting the Body and preventing the Accelerate™
Funnel from properly engaging, ensure the Cannula
is fully seated.
11 The proximal tip of the Cannula should be aligned
with the Plunger.
12 It may take several Trigger squeezes prior to
delivery of the bone graft mixture. Typically, the
bone graft mixture in the Cannula rst compresses,
and then begins to ow through distal end of the
™
Accelerate
Funnel.
13 If there is increasing resistance/diculty in squeezing
™
the Trigger, remove the Accelerate
Graft Delivery
Instrument from the surgical site and disassemble in
accordance with the instructions in the Reloading and
Disassembly section of this User Guide.
14 If the Plunger is hard to pull back, conrm that the
Trigger has been fully reset, and sits ush with the
™
Accelerate
Body.
15 The Cannula is intended for SINGLE USE ONLY. If
additional bone graft is needed, repeat the loading
process with a new Cannula until the desired amount
of bone graft mixture is delivered.
16 If the Cannula is dicult to remove, simultaneously
™
twist the Accelerate
Funnel while pulling out the
Cannula.
17 For rapid bone graft delivery, it is recommended that
™
the spare Accelerate
Funnel is prepped and loaded
with a Cannula to allow for rapid Accelerate™ Funnel
exchange.
18 Be sure to remove the Cannula after use (and prior
to sterilization).
Page 6
ACCELERATE
PROCEDURE
™
BONE GRAFT DELIVERY SYSTEM
Graft Preparation
Autograft Preparation
Mill (or mince) the autologous bone until the graft pieces
are less than 3.2mm in diameter. The specied sizes can
be obtained using Medtronic Bone Mill (Midas Rex™ Bone
Mill, Figure 1). If using a Stryker Bone Mill, only the Fine
Blade can mill to the required bone chip sizes.
Note
The instrument will clog if the bone chip size exceeds
3.2mm in diameter.
Note
Use Accelerate™ Graft Delivery Instrument as intended
only with Grafton™ DBF. Use with other products
(mineralized allograft, synthetics, dierent bone
graft formulations, etc.) will likely result in decreased
performance, clogging, and possible permanent
malfunction.
Grafton™ DBF Preparation
Hydrate the Grafton™ DBF with blood, bone marrow
aspirate, sterile saline, or sterile water in accordance
with the Instructions For Use (IFU).
Note
Only use Accelerate™ BG Set T502XX Grafton™ DBF. The
original T501XX Grafton™ DBF will not be compatible as it
does not contain the Cannula.
Note
The Accelerate™ Graft Delivery Instrument will not work
eciently and clog if used without the Cannula.
“Bone Graft Mixture” Preparation
Combine the hydrated Grafton™ DBF with up to an equal
volume of autograft (milled to less than 3.2mm in diameter)
(Figure 2). Add additional uid to the bone graft mixture to
obtain the desired consistency.
Note
The Accelerate™ Graft Delivery Device may not function as
intended (and is likely to clog) if the autograft content of the
nal bone grafting mixture exceeds 50% of the total volume.
Figure 2Figure 1
6
Page 7
ACCELERATE™ BONE GRAFT DELIVERY SYSTEM
PROCEDURE
Filling the Accelerate™ Cannula
with the Bone Graft Mixture
Remove the Cannula from Accelerate™ BG Set T502XX
Grafton™ DBF box using standard aseptic techniques.
Note
Close and open the Cannula once or twice to make the
closing of the Cannula easier once it has been loaded with
bone graft mixture.
Place the Cannula(s) into the grooves on the Accelerate
™
Graft Delivery Loading Platform concave up (Figure 3).
Up to two Cannulas may be placed simultaneously on
Figure 3
the Loading Platform. Spread the bone grafting mixture
uniformly in the Cannula ensuring the bone graft mixture
stays within the top of the Cannula curvature (Figure 4). After
the Cannula is lled, close the Cannula (Figure 5).
Note
If the Cannula won’t completely close, redistribute or
remove small amounts of bone grafting mixture until the
Cannula completely closes. Overlling of Cannula increases
the likelihood of clogging.
Figure 4
Figure 5
7
Page 8
ACCELERATE™ BONE GRAFT DELIVERY SYSTEM
PROCEDURE
Loading the Accelerate™ Cannula
™
into the Accelerate
Funnel
Insert the lled Cannula into the proximal end of
Accelerate™ Funnel (Figure 6).
Note
Do not force the Cannula into the Accelerate™ Funnel. If
you experience resistance loading the Cannula into the
Accelerate™ Funnel, redistribute or remove small amounts
of bone grafting mixture.
When properly inserted, the Cannula will protrude
approximately 5mm from the proximal end of the
Accelerate™ Funnel (Figure 7). An internal ledge prevents
the Cannula from sliding out of the distal end of the
Accelerate™ Funnel.
Note
The Accelerate™ Instrument Set contains two
Accelerate™ Funnels for simultaneous preloading
of bone graft mixture.
Figure 6
Cannula protrudes 5mmFigure 7
8
Page 9
ACCELERATE™ BONE GRAFT DELIVERY SYSTEM
PROCEDURE
Attaching the Accelerate™ Funnel
™
to the Accelerate
Body
Slide the Accelerate™ Funnel Pin into the Body Socket at a
perpendicular angle (Figure 8). While depressing the Clasp
on the Body (Figure 9), rotate the Accelerate™ Funnel into
alignment with the Accelerate™ Body (Figure 10). Once the
Accelerate™ Funnel and Body are aligned, release the Clasp
and ensure the Accelerate™ Funnel is securely attached to
the Body.
Note
If the Cannula (inside of the Accelerate™ Funnel) is hitting the
Body and preventing the Accelerate™ Funnel from properly
engaging, ensure the Cannula is fully seated (Figure 11).
Note
The proximal tip of the Cannula should be aligned with
the Plunger.
Advance the Plunger into the Accelerate™ Funnel by
squeezing the Trigger (approximately 2-3 times). The
positive engagement of the Plunger into the Cannula will
improve the overall stability of the Instrument.
Figure 8
Figure 10
Funnel Pin
9
Figure 11
Figure 9
Cannula protrudes 5mm
Page 10
ACCELERATE
PROCEDURE
™
BONE GRAFT DELIVERY SYSTEM
Delivery of the Bone Graft Mixture
Squeeze the Trigger to begin delivery of the bone graft
mixture to the surgical site. The Plunger will advance the
bone graft mixture with each Trigger squeeze until the
Cannula is empty (Figure 12).
Note
It may take several Trigger squeezes prior to delivery of
the bone graft mixture. Typically, the bone graft mixture
in the Cannula rst compresses, and then begins to ow
through distal end of the Accelerate™ Funnel.
The bone graft mixture should ow easily with each
Trigger squeeze.
Note
If there is increasing resistance/diculty in squeezing the
Trigger, remove the Accelerate™ Graft Delivery Instrument
from the surgical site and disassemble it in accordance with
the instructions in the Reloading and Disassembly section of
this User Guide.
Figure 12
10
Page 11
ACCELERATE™ BONE GRAFT DELIVERY SYSTEM
PROCEDURE
Reloading and Disassembly
To remove the Plunger from the Accelerate™ Funnel,
simultaneously grasp the Back Handle (do not squeeze
the Trigger) and depress the grey Plunger Release
Tab with one hand. While depressing the Release Tab,
completely pull the Plunger out of the Accelerate™ Funnel
with the other hand (Figure 13).
Next, remove the Accelerate™ Funnel by depressing
the Clasp with one hand, and rotating it perpendicular
to the Body with the other hand (Figure 14). Disengage
the Accelerate™ Funnel from the Body, then remove the
Cannula from the Accelerate™ Funnel and discard the
Cannula (Figure 15).
Note
If the Plunger is hard to pull back, conrm that the
Trigger has been fully reset, and sits ush with the
Accelerate™ Body.
Note
The Cannula is intended for SINGLE USE ONLY. If additional
bone graft is needed, repeat the loading process with a new
Cannula until the desired amount of bone graft mixture is
delivered.
Note
If the Cannula is dicult to remove, simultaneously twist
the Accelerate™ Funnel while pulling out the Cannula.
Note
For rapid bone graft delivery, it is recommended that the
spare Accelerate™ Funnel is prepped and loaded with a
Cannula to allow for rapid Accelerate™ Funnel exchange.
Note
Be sure to remove the Cannula after use (and prior
to sterilization).
Figure 13Figure 14
Figure 15
11
Page 12
ACCELERATE™ BONE GRAFT DELIVERY SYSTEM
PROCEDURE
Graft Recovery
If a jam occurs, there are provisions to recover the clogged
bone graft. First, remove the Accelerate™ Graft Delivery
Instrument from the surgical site. Follow the disassembly
instructions as outlined in the previous Reloading and
Disassembly section to remove the Accelerate™ Funnel
from the Body.
Attach the proximal end of the Accelerate™ Funnel to
the top of the Graft Recovery Platform (Figure 16). Then,
insert the Graft Recovery Tool into the distal tip of the
Accelerate™ Funnel (Figure 17). To push the Cannula out of
the Accelerate™ Funnel, gently mallet the Graft Recovery
Tool into the Accelerate™ Funnel (Figure 18). The Cannula
may bend and or collapse as it is malleted from the
Accelerate™ Funnel (Figure 19). Recover bone graft mixture
by gently opening the Cannula.
The recovered bone graft mixture may be delivered with the
Standard Bone Funnel included in the set. The Graft Recovery
Tool will act as the bone tamp for this backup graft delivery
method (Figure 20).
T50212Accelerate™ BG Set T50212 Grafton™ DBF 12cc4
Accessory Ordering Information
Midas Rex™ Bone Mill
Part Number SAP Description
BM110Midas Rex™ Bone Mill Base
BM120Midas Rex™ Bone Mill Console
BM130Midas Rex™ Bone Mill Instrument
BM210Midas Rex™ Dual Blade Disposable
EA600 Power Cord, US
13
Page 14
IMPORTANT PRODUCT
INFORMATION
Accelerate™ Bone Graft
Delivery System
Intended Use
These orthopedic manual surgical instruments are
intended for use in surgical procedures to manipulate
tissue, bone, or for use with other devices in orthopedic
surgery. An instrument may incorporate a measuring
function which has uses as described on the label and the
instrument.
DO NOT IMPLANT THE INSTRUMENTS.
MEDTRONIC does not and cannot warrant the use
of this instrument nor any of the component parts
upon which repairs have been made or attempted,
except as performed by MEDTRONIC or an authorized
MEDTRONIC repair representative. Implied warranties of
merchantability and tness for a particular purpose or use
are specically excluded.
Risks
Breakage, slippage, misuse, or mishandling of
instruments, such as on sharp edges, may cause injury
to the patient or operative personnel. It is important that
the surgeon exercise extreme caution when working in
close proximity to vital organs, nerves or vessels, and that
the forces applied while correcting the position of the
instrumentation is not excessive, such that it might cause
injury to the patient.
For a complete list of indications, safety, and warnings for
Accelerate™ Graft Delivery Instrument Set, please visit
For more information visit Medtronic.com or
call (800) 933-2635.
Grafton™ DBF
Indications
Grafton™ DBF can be used in orthopedic or reconstructive
bone grafting procedures. The product can also be used
in bone grafting procedures in combination with
autologous bone or other forms of allograft bone, or
alone as a bone graft.
Contraindications
The presence of infection at the transplantation site is a
contraindication for the use of this allograft.
Caution
This allograft may contain trace amounts of antibiotics
(gentamicin), antiseptic (povidone-iodine) and alcohol
solutions. Caution should be exercised if the patient is
allergic to these antibiotics or chemicals.
Precautions
Extensive donor blood serum testing, medical and social
history screening procedures, and tissue microbiological
testing were used in the qualication of tissue donors.
Despite the viral inactivation and extensive tissue donor
selection and qualication processes used in providing
this tissue graft (see DONOR SCREENING AND TESTING),
transmission of a communicable disease is still possible.
Bacterial infection at the graft site may also occur. Adverse
outcomes potentially attributable to
Grafton™ DBF must be reported promptly to Medtronic.
If injecting Grafton™ DBF into the defect site, precaution
should be taken not to:
over-pressurize the delivery device, as this may lead to
extrusion of the device beyond the site of its intended
application and damage to the surrounding tissues.
over-pressurize the defect site, as this may lead to fat
embolization or embolization of the device material into
the bloodstream.
For a complete list of indications, safety, and warnings
for Grafton™ DBF, please visit https://manuals.
For more information visit Medtronic.com or call (800) 933-2635.
15
Page 16
Medtronic
Spinal and Biologics Business
Worldwide Headquarters
2600 Sofamor Danek Drive
Memphis, TN 38132
The Use r Guide shown i s for illustr ative
purposes only. The technique(s) actually
empl oyed in each case w ill always
depend upon the medical judgment
of the su rgeon exercised b efore and
duri ng surgery as to t he best mod e of
treatm ent for eachpat ient.
Consul t instruc tions for use at t his
website www.medtronic.com/
manuals.
Medtronic Sofamor Danek USA, Inc.
1800 Pyramid Place
Memphis, TN 38132