LR Abnormal Control
LR unormal kontrol
LR-Abnormal-Gerinnungskontrolle
Μηφυσιολογικόςμάρτυρας LR
Control LR anormal
Epänormaali LR-kontrolli
Contrôle de coagulation anormale LR
Controllo anomalo LR
LR-controlevloeistof voor abnormalestollingstest
Abnorm kontroll for lave verdier
Controlo LR anormal
Контрольпатологическихзначений LR
LR-kontroll för onormala värden
Instructions for Use ■ Brugsanvisning ■ Gebrauchsanweisung ■ Οδηγίεςχρήσης ■
Instrucciones de uso ■ Käyttöohjeet ■ Mode d’emploi ■ Istruzioni per l'uso ■
Gebruiksinstructies ■ Bruksanvisning ■ Instruções de utilização ■ Инструкцияпоэксплуатации■ Bruksanvisning
Caution: Federal Law (USA) restricts this device to sale by
or on the order of a physician.
Trademarks may be registered and are the property of their respective owners.
Varemærker kan være registrerede og tilhører deres respektive ejere.
Marken sind eventuell eingetragen und sind das Eigentum ihrer jeweiligen Inhaber.
Τα εμπορικά σήματα μπορεί να είναι καταχωρημένα και είναι ιδιοκτησία των αντίστοιχων
κατόχων τους.
Las marcas comerciales pueden estar registradas y pertenecen a sus respectivos
propietarios.
Tavaramerkit voivat olla rekisteröityjä, ja ne ovat omistajiensa omaisuutta.
Les marques commerciales mentionnées peuvent être déposées et appartiennent à leurs
propriétaires respectifs.
I marchi di fabbrica possono essere registrati e sono di proprietà dei rispettivi titolari.
Handelsmerken kunnen zijn geregistreerd en zijn het eigendom van de desbetreffende
eigenaars.
Varemerkene kan være registrerte og tilhører de respektive eierne.
As marcas comerciais podem ser registadas e são propriedade dos respectivos detentores.
Тов арн ы е знаки могут быть зарегистрированными и являются собственностью их
владельцев.
Varumärken kan vara registrerade och tillhör respektive ägare.
Explanation of symbols on package labeling / Forklaring til symbolerne på
emballagens mærkater / Erläuterung der Symbole auf dem
Verpackungsetikett
συσκευασίας
del envase / Pakkauksen merkintöjen selitykset / Explication des
symboles des étiquettes sur l’emballage
etichette della confezione
productlabels / Forklaring av symboler på produktet og pakningen /
Explicação dos símbolos na documentação da embalagem / Объяснениесимволов, приведенныхнаупаковке
förpackningsetiketten
/ Επεξήγησητωνσυμβόλωνστησήμανσητης
/ Explicación de los símbolos que aparecen en el etiquetado
/ Verklaring van de symbolen op de
/ Spiegazione dei simboli sulle
/ Förklaring av symboler på
1
Conformité Européenne (European Conformity). This
symbol means that the device fully complies with European
Council Directive IVDD: 98/79/EC. / Conformité
Européenne (Europæisk Standard). Dette symbol betyder,
at enheden fuldt ud overholder Det Europæiske Råds
direktiv IVDD: 98/79/EF. / Conformité Européenne
(Europäische Konformität). Dieses Symbol besagt, dass
das Gerät die Anforderungen der Richtlinie 98/79/EG des
Europäischen Rats über Medizinprodukte für die In-VitroDiagnostik (IVDD) vollständig erfüllt. / Conformité
Européenne (Ευρωπαϊκή Συμμόρφωση). Το σύμβολο αυτό
υποδηλώνει ότι το προϊόν συμμορφώνεται πλήρως με την
Οδηγία του Ευρωπαϊκού Συμβουλίου περί
ιατροτεχνολογικών βοηθημάτων που χρησιμοποιούνται στη
διάγνωση in vitro: 98/79/EK. / Conformité Européene
(Conformidad europea). Este símbolo indica que el
dispositivo cumple totalmente la Directiva del Consejo
Europeo IVDD: 98/79/CE. / Conformité Européenne
(eurooppalainen vaatimustenmukaisuus). Tämä symboli
tarkoittaa, että laite noudattaa kaikilta osin Euroopan
unionin neuvoston in vitro -diagnostiikkaan tarkoitetuista
lääkinnällisistä laitteista annettua direktiiviä 98/79/EY. /
Conformité Européenne. Ce symbole signifie que
l'appareil est entièrement conforme à la Directive
européenne IVDD: 98/79/CE. /
2
Conformité Européenne (Conformità europea). Questo
simbolo significa che il dispositivo è pienamente conforme
alla Direttiva del Consiglio europeo IVDD 98/79/CE. /
Conformité Européenne (Europese Conformiteit). Dit
symbool betekent dat het product volledig voldoet aan
richtlijn van de Europese Raad IVDD: 98/79/EC. /
Conformité Européenne (samsvar med europeisk
standard). Dette symbolet betyr at enheten er fullstendig
i samsvar med EU-direktiv IVDD: 98/79/EF. / Conformité
Européenne (Conformidade Europeia). Este símbolo
significa que o dispositivo está em total conformidade com
a Directiva do Conselho Europeu IVDD: 98/79/CE. /
Conformité Européenne (Европейское Соответствие).
Этот символ означает, что устройство полностью
соответствует требованиям директивы Европейского
Совета IVDD: 98/79/EC. / Conformité Européenne
(Europeisk standard). Denna symbol betyder att
utrustningen helt följer Europeiska rådets IVD-direktiv,
98/79/EG.
3
In Vitro Diagnostic Medical Device / Medicinsk udstyr til in
vitro-diagnostik / Medizinprodukt für die In-vitro-Diagnostik /
Ιατροτεχνολογικό βοήθημα που χρησιμοποιείται στη
διάγνωση in vitro / Dispositivo médico para diagnóstico in
vitro / In vitro -diagnostinen lääkinnällinen laite / Dispositif
médical de diagnostic in vitro / Dispositivo medico
diagnostico in vitro / In-vitro diagnostisch medisch
apparaat / In vitro-diagnostisk medisinsk utstyr / Dispositivo
médico de diagnóstico in vitro / Медицинское устройство, предназначенноедлядиагностики in vitro /
Medicinteknisk produkt för in vitro-diagnostik
Lot Number / Partinummer / Chargennummer / Αριθμόςπαρτίδας / Número de lote / Eränumero / Numéro de lot /
Codice del lotto / Partijnummer / Lotnummer / Número de
lote / Номерпартии / Lotnummer
Quantity / Antal / Anzahl / Ποσότητα / Cantidad / Määrä /
Quantité / Quantità / Aantal / Antall / Quantidade /
Количество / Mängd
Temperature Limit / Temperaturgrænse /
Temperaturbereich / Όριοθερμοκρασίας / Límite de
temperatura / Lämpötilaraja / Limite de température / Limiti
di temperatura / Temperatuurbereik / Temperaturgrense /
Limites de temperatura / Пределтемпературы /
Temperaturgräns
4
Use-By Date / Sidste anvendelsesdato / Zu verwenden bis
einschließlich / Ημερομηνία «Χρήσηέως» / No utilizar
después de / Viimeinen käyttöpäivämäärä / À utiliser
jusqu'au / Utilizzare entro / Uiterste gebruiksdatum / Siste
forbruksdag / Utilizar antes da data / Датаистечения срока годност и / Används före
Date of Manufacture / Fremstillingsdato /
Herstellungsdatum / Ημερομηνίακατασ κευής / Fecha de
fabricación / Valmistuspäivämäärä / Date de fabrication /
Data di fabbricazione / Productiedatum / Produksjonsdato /
Data de fabrico / Датаизготовления / Tillverkningsdatum
Manufactured In / Fremstillet i / Hergestellt in /
Κατασκευάσ τηκεσε / Fabricado en / Valmistuspaikka / Lieu
de fabrication / Fabbricato in / Vervaardigd in / Produsert i /
Fabricado em / Произведенов / Tillverkad i
5
Consult Instructions for Use / Se brugsanvisningen /
Gebrauchsanweisung beachten / Συμβουλευτείτε τις οδηγίεςχρήσης / Consultar las instrucciones de uso / Katso
käyttöohjeet / Consulter le mode d'emploi / Consultare le
istruzioni per l'uso / Raadpleeg gebruiksaanwijzing / Se i
bruksanvisningen / Consultar instruções de utilização /
Обратитеськинструкциипоэксплуатации / Se
bruksanvisningen
Authorized Representative in the European Community /
Autoriseret repræsentant i EF / Autorisierte Vertretung für
die Europäische Gemeinschaft / Εξουσιοδοτημένος αντιπρόσωποςστηνΕυρωπαϊκήΚοινότητα / Representante
autorizado en la Comunidad Europea / Valtuutettu edustaja
Euroopan yhteisön alueella / Représentant agréé dans la
Communauté européenne / Rappresentante autorizzato
nella Comunità europea / Geautoriseerd vertegenwoordiger
in de Europese Gemeenschap / Autorisert representant i
Det europeiske fellesskap / Representante Autorizado na
Comunidade Europeia / Уполномоченный представитель вЕС / Auktoriserad representant inom EG
Catalog Number / Katalognummer / Katalognummer /
Αριθμόςκαταλόγου / Número de catálogo /
Luettelonumero / Numéro de référence / Numero di
catalogo / Catalogusnummer / Katalognummer / Número de
catálogo / Номерпокаталогу / Katalognummer
6
For US Audiences Only / Gælder kun i USA / Gilt nur für
Leser in den USA / ΜόνογιαπελάτεςεντόςτωνΗΠΑ / Sólo
aplicable en EE.UU. / Koskee vain Yhdysvaltoja / Ne
s'applique qu'aux États-Unis / Esclusivamente per il
mercato statunitense / Alleen van toepassing voor de VS /
Gjelder bare i USA / Apenas aplicável aos E.U.A. / Тол ьк о дляСША / Gäller endast i USA
Control / Kontrol / Kontrolle / Μάρτυρας / Control / Kontrolli /
Contrôle / Controllo / Controlevloeistof / Kontroll / Controlo /
Контроль / Kontroll
Reconstitute with X mL of Deionized Water / Opblødning i X
ml afioniseret vand / Mit X ml deionisiertem Wasser
rekonstituieren / Ναανασυσταθείμε X mL απιονισμένουνερού / Reconstituir con X ml de agua desionizada /
Valmista käyttövalmiiksi lisäämällä X ml deionisoitua vettä /
À reconstituer avec X ml d'eau déionisée / Ricostituire con
X ml di acqua deionizzata / Reconstitueren in X ml
gedemineraliseerd water / Rekonstitueres med X ml
avionisert vann / Reconstituir com X ml de água
desionizada / Восстановить X млдеионизированнойводы / Rekonstitueras med X ml avjoniserat vatten
Wate r / Vand / Wasser / Νερό / Agua / Vesi / Eau / Acqua /
Wate r / Vann / Água / Вода / Vatten
Nonsterile / Ikke-steril / Unsteril / Μηστείρο / No estéril /
Epästeriili / Non stérile / Non sterile / Niet-steriel / Usteril /
Não estéril / Нестерильно / Icke-steril
Do Not Reuse / Må ikke genanvendes / Nicht
wiederverwenden / Μηνεπαναχρησιμοποιείτε / No
reutilizar / Kertakäyttöinen / Ne pas réutiliser / Non
riutilizzare / Niet opnieuw gebruiken / Skal ikke brukes flere
ganger / Não reutilizar / Неиспользовать повторно / Får ej
återanvändas
8
1 Intended Use
The CLOTtrac™ Low Range (LR) Abnormal Coagulation Control is used to confirm the
performance of the ACT instrument and LR-ACT cartridge.
For in vitro diagnostic use.
2 Summary
The Activated Clotting Time (ACT) test was originally described by Hattersley in 1966.1 The
test and its variations have been widely used to monitor the effect of heparin.
Controls have been developed by Medtronic to monitor the performance of this test. The
control consists of lyophilized citrated sheep plasma and erythrocytes. Nonsterile deionized
water is used to reconstitute the control (refer to Section 6 on page 10 for instructions on
reconstitution). Once reconstituted, the abnormal control behaves like a patient’s freshly
drawn, whole blood sample.
3 Principles
The control provides an extended clotting time, which can be used to satisfy quality
assurance programs requiring at least two levels of control testing. This control provides one
of the two levels.2 The reconstituted control mimics a freshly drawn, whole blood sample,
which has an extended clotting time due to heparinization, dilution, coagulopathy, etc.
4 Reagents
The control is made of lyophilized citrated sheep plasma and erythrocytes. Nonsterile
deionized water is used to reconstitute the controls.
English Instructions For Use 9
5 Storage and Stability
CLOTtrac LR Abnormal Coagulation Controls and deionized water require no special
handling during shipping and transport. Upon receipt, the control should be stored
refrigerated, from 2°C to 10°C (35.6°F to 50°F), until use. The expiration date of the controls
and water is indicated on the vial label. Do not use controls past expiration date. Once
reconstituted, the control is stable for 1 hour at room temperature, 15°C to 25°C (59°F
to 77°F).
6 Procedure
This product is intended for use only by trained individuals.
Materials Supplied:
• Deionized Water, 15 vials per box
• CLOTtrac LR Abnormal Coagulation Control, 15 vials per box
Materials Required, not supplied:
• 3-cc plastic syringe with needle
• Medtronic ACT Instrument
• Medtronic LR-ACT cartridges
•0.1 M CaCl
1. Remove the required number of CLOTtrac LR Abnormal Coagulation Controls and
2. For each vial of Iyophilized LR abnormal control, add 1.5 mL of deionized water using a
3. Allow at least 10 minutes for adequate rehydration. Failure to do so may lead to an
10 English Instructions For Use
2
deionized water vials.
syringe.
Note: Do not agitate the control until completely rehydrated.
improperly reconstituted control, which will give erratic results.
4. Once rehydrated, shake the control vigorously until the suspension is uniformly
dispersed and the control is completely reconstituted.
5. Tap the cartridge to resuspend the cartridge reagents.
6. Instructions specific to Medtronic's ACT Plus™ Instrument when using LR-ACT
cartridges:
a. From the “Main Menu”, select or confirm that the cartridge type is set to “LR-ACT”.
b. From the “QC Menu”, select “LR-AB” (LR Abnormal Control).
c. Transfer 0.2 mL of reconstituted control into each channel of the cartridge. Fill to
between the fill lines etched on each channel body.
d. Insert the cartridge into the ACT Plus™ instrument and allow to warm for 5 minutes.
e. Remove the cartridge from the instrument and immediately add one drop (40 µl) of
0.1 M CaCl2 to each channel of the cartridge.
Note: When adding CaCl2, hold the vial in a vertical position to ensure that a full drop of
CaCl2 is delivered.
f. Place the cartridge into the instrument and close the actuator heat block to start the
test.
g. Clot formation is signaled by an audible tone, the automatic opening of the actuator
heat block, and the displaying of the results.
7 Results
Results of control testing are reported in seconds. The control results should not be used in
interpreting patient results. Abnormal control values obtained should result in less than 12%
variation between channel clotting times.
Sample Calculation:
Listed Control Range=370 to 490 secs
English Instructions For Use 11
Channel 1 Result=400 secs
Channel 2 Result=420 secs
Mean Result=410 secs
12% of the Mean=49 secs
Difference=20 secs
The data are acceptable because 1) the mean clotting time, as well as both Channel clot
times, falls within the listed range and 2) the difference between Channel 1 and Channel 2
(20 secs) is less than 12% of the mean (49 secs).
8 Limitations
The CLOTtrac Abnormal LR Coagulation Control is intended for in vitro diagnostic use only.
The results are dependent on good technique and proper adherence to protocol. Strict
adherence to rehydration requirements is recommended.
The clotting time ranges listed are valid only when the control and cartridge have been stored
as indicated.
Failure to obtain clotting times within the established range invalidates the use of both the
instrument and the test until the problem is resolved.
Even though controls are of nonhuman origin, all biological samples should be considered
potentially hazardous and handled appropriately.
The deionized water is nonsterile. Each vial of water is intended for single use only (to prevent
contamination).
To ensure that the control performs properly, reconstitution should be achieved using only the
enclosed deionized water. It is recommended that the contents of the box be kept intact to
ensure availability of all components at all times.
12 English Instructions For Use
Allow at least 30 minutes to perform a complete quality control procedure from reconstitution
to final results.
If the results obtained do not agree with the range given in Specific Performance
Characteristics of the package insert (Section 9):
1. Ensure that the instrument is functioning properly according to the Maintenance and
Quality Control section in the instrument's Operator’s Manual.
2. Incubate or time for 5 minutes to ensure that the cartridge, instrument, and control are
at 37°C ± 0.5°C (98.6°F ± 0.9°F) before initiating the test.
3. Ensure that the cartridges and controls used have not exceeded their respective
expiration dates and have been properly stored.
4. Ensure that the control was properly reconstituted.
5. Inspect the control vial for visible clumping. If clumping is present in the vial, the
performance of the control may be affected.
6. Ensure that the test was initiated from the Quality Control screen, if applicable.
7. Repeat the test if the criteria in items 1 to 6 have been met.
8. Report any unresolved problems to Medtronic.
9 Specific Performance Characteristics
CLOTtrac LR Abnormal Coagulation Control is a known solution that is analyzed for quality
control purposes, providing the user with a range of values. This range is determined for each
specific lot of control. The user may use the provided range or, if desired, the user can
determine their own mean, standard deviation, and range for each lot of controls.
10 References
1. Hattersley, P. Activated Coagulation Time of Whole Blood. JAMA. 1966;136:436.
THE FOLLOWING DISCLAIMER OF WARRANTY APPLIES TO UNITED STATES
CUSTOMERS ONLY:
DISCLAIMER OF WARRANTY
THOUGH THE CLOTtrac™ LR ABNORMAL CONTROL, HEREAFTER REFERRED TO
AS “PRODUCT,” HAS BEEN MANUFACTURED UNDER CAREFULLY CONTROLLED
CONDITIONS, MEDTRONIC HAS NO CONTROL OVER THE CONDITIONS UNDER
WHICH THIS PRODUCT IS USED. MEDTRONIC, THEREFORE DISCLAIMS ALL
WARRANTIES, BOTH EXPRESS AND IMPLIED, WITH RESPECT TO THE PRODUCT,
INCLUDING, BUT NOT LIMITED TO, ANY IMPLIED WARRANTY OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. MEDTRONIC
SHALL NOT BE LIABLE TO ANY PERSON OR ENTITY FOR ANY MEDICAL EXPENSES
OR ANY DIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES CAUSED BY ANY
USE, DEFECT, FAILURE OR MALFUNCTION OF THE PRODUCT, WHETHER A CLAIM
FOR SUCH DAMAGES IS BASED UPON WARRANTY, CONTRACT, TORT OR
OTHERWISE. NO PERSON HAS ANY AUTHORITY TO BIND MEDTRONIC TO ANY
REPRESENTATION OR WARRANTY WITH RESPECT TO THE PRODUCT.
The exclusions and limitations set out ab ove are not intended to, and should not be construed
so as to, contravene mandatory provisions of applicable law. If any part or term of this
Disclaimer of Warranty is held to be illegal, unenforceable or in conflict with applicable law by
a court of competent jurisdiction, the validity of the remaining portions of this Disclaimer of
Warranty shall not be affected, and all rights and obligations shall be construed and enforced
as if this Disclaimer of Warranty did not contain the particular part or term held to be invalid.
14 English Instructions For Use
THE FOLLOWING DISCLAIMER OF WARRANTY APPLIES TO CUSTOMERS OUTSIDE
THE UNITED STATES:
DISCLAIMER OF WARRANTY
ALTHOUGH THE CLOTtrac™ LR ABNORMAL CONTROL, HEREAFTER REFERRED TO
AS “PRODUCT,” HAS BEEN CAREFULLY DESIGNED, MANUFACTURED AND TESTED
PRIOR TO SALE, THE PRODUCT MAY FAIL TO PERFORM ITS INTENDED FUNCTION
SATISFACTORILY FOR A VARIETY OF REASONS. THE WARNINGS CONTAINED IN
THE PRODUCT LABELING PROVIDE MORE DETAILED INFORMATION AND ARE
CONSIDERED AN INTEGRAL PART OF THIS DISCLAIMER OF WARRANTY.
MEDTRONIC, THEREFORE, DISCLAIMS ALL WARRANTIES, BOTH EXPRESS AND
IMPLIED, WITH RESPECT TO THE PRODUCT. MEDTRONIC SHALL NOT BE LIABLE
FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES CAUSED BY ANY USE,
DEFECT OR FAILURE OF THE PRODUCT, WHETHER THE CLAIM IS BASED ON
WARRANTY, CONTRACT, TORT OR OTHERWISE.
The exclusions and limitations set out ab ove are not intended to, and should not be construed
so as to, contravene mandatory provisions of applicable law. If any part or term of this
Disclaimer of Warranty is held by any court of competent jurisdiction to be illegal,
unenforceable or in conflict with applicable law, the validity of the remaining portion of the
Disclaimer of Warranty shall not be affected, and all rights and obligations shall be construed
and enforced as if this Disclaimer of Warranty did not contain the particular part or term held
to be invalid.
English Instructions For Use 15
16 English Instructions For Use
1 Tilsigtet brug
CLOTtrac™ Low Range (LR) unormal koagulationskontrol bruges til at verificere præstationen for
ACT-instrumentet samt LR-ACT-kassetten.
Til brug i forbindelse med in vitro-diagnostik.
2 Resumé
Testen for aktiveret koaguleringstid (Activated Clotting Time; ACT) blev oprindeligt beskrevet
af Hattersley i 1966.1 Testen og dens varianter er blevet anvendt i stort omfang for at monitorere
heparins antikoagulerende virkning.
Medtronic har udviklet kontroller til monitorering af præstationen for denne test. Kontrollen
består af frysetørret, citreret fåreplasma og røde blodlegemer fra får. Der anvendes ikkesterilt, afioniseret vand til opblødning af kontrollen (se Afsnit 6 på side 18 for at få vejledning
i opblødning). Når den er opblødt, bruges den unormale kontrol på samme måde som en frisk
fuldblodsprøve fra en patient.
3 Principper
Kontrollen giver forlænget koagulationstid, hvilket kan benyttes til at imødekomme
kvalitetssikringsprogrammer, som kræver kontroltestning i mindst to niveauer. Denne
kontrol omfatter det ene af de to niveauer.2 Den genopblødte kontrol simulerer en frisk
fuldblodsprøve, der har en længere koagulationstid som følge af heparinisering, fortynding,
koagulopati osv.
4 Reagenser
Kontrollen er fremstillet af frysetørret, citreret fåreplasma og røde blodlegemer fra får.
Der anvendes ikke-sterilt, afioniseret vand til opblødning af kontrollerne.
Dansk Brugsanvisning 17
5 Opbevaring og stabilitet
CLOTtrac LR unormale koagulationskontroller og afioniseret vand kræver ingen særlig
håndtering under forsendelse og transport. Ved modtagelsen skal kontrollen være opbevaret
afkølet ved 2°C til 10°C (35,6°F til 50°F), indtil den tages i brug. Kontrollernes og vandets
udløbsdato står anført på flaskens mærkat. Kontrollerne må ikke anvendes efter udløbsdatoen.
Når kontrollen er opblødt, forbliver den stabil i 1 time ved stuetemperatur, 15°C til 25°C
(59°F til 77°F).
6 Procedure
Dette produkt er kun beregnet til at blive anvendt af dertil uddannet personale.
Leverede materialer:
• Afioniseret vand, 15 flasker pr. æske
• CLOTtrac LR unormal koagulationskontrol, 15 flasker pr. æske
Nødvendige materialer, som ikke medfølger:
• 3 ml plastsprøjte med kanyle
• Medtronic ACT-instrument
• Medtronic LR-ACT-kassetter
• 0,1 M CaCl
1. Udtag det nødvendige antal CLOTtrac LR unormale koagulationskontroller og flasker
2. For hver flaske frysetørret LR unormal kontrol skal der tilføres 1,5 ml afioniseret vand
3. Vent mindst 10 minutter, for at opnå tilstrækkelig rehydrering. Undladelse heraf kan
18 Dansk Brugsanvisning
2
med afioniseret vand.
ved hjælp af en sprøjte.
Bemærk: Kontrollen må ikke rystes, før den er fuldstændigt rehydreret.
medføre, at kontrollen ikke er opblødt korrekt og dermed viser fejlagtige resultater.
4. Efter rehydrering skal kontrollen rystes kraftigt, indtil opløsningen er ensartet fordelt
og kontrollen er helt opblødt.
5. Bank let på kassetten for at resuspendere reagenserne i kassetten.
6. Specific vejledning til Medtronic ACT Plus™-instrumentet ved brug af LR-ACT-kassetter:
a. Vælg eller bekræft i "Hovedmenu", at kassettetypen er indstillet til "LR-ACT".
b. Vælg "LR-AB" (LR unormal kontrol) i "QC Menu".
c. Overfør 0,2 ml opblødt kontrol til hver af kassettens kanaler. Fyld op til mellem de
angivne fyldemærker i hver kanal.
d. Sæt den fyldte kassette i ACT Plus™-instrumentet, og lad det varme op i 5 minutter.
e. Fjern kassetten fra instrumentet, og tilsæt straks én dråbe (40 µl) med 0,1 M CaCl2
i hver af kassettens kanaler.
Bemærk: Ved tilsætningen af CaCl2 skal flasken holdes lodret for at sikre, at der tilføres
en hel dråbe CaCl2.
f. Sæt kassetten i instrumentet, og luk aktuatorens varmeblok for at påbegynde testen.
g. Ved koageldannelse lyder der en tone, aktuatorens varmeblok åbner automatisk og
resultaterne vises.
Dansk Brugsanvisning 19
7 Resultater
Resultaterne af kontroltest er angivet i sekunder. Kontrolresultaterne bør ikke anvendes til
fortolkning af patientresultater. Unormale kontrolværdier, der indsamles, bør vise under 12%
afvigelse mellem kanalernes koagulationstider.
Prøveberegning:
Angivet kontrolområde=370 til 490 sekunder
Kanal 1 resultat=400 sekunder
Kanal 2 resultat=420 sekunder
Middelværdi for resultat=410 sekunder
12% af middelværdien=49 sekunder
Difference=20 sekunder
Dataene er acceptable, fordi 1) middelværdien for koagulationstiden samt begge kanalers
koagulationstider ligger inden for det angivne område, og 2) differencen mellem Kanal 1
og Kanal 2 (20 sekunder) er mindre end 12% af middelværdien (49 sekunder).
8 Begrænsninger
CLOTtrac unormal LR-koagulationskontrol er udelukkende beregnet til in vitrodiagnosticering.
Resultaterne er afhængige af god teknik og korrekt efterlevelse af protokollen. Det anbefales,
at rehydreringskravene overholdes fuldt ud.
De angivne koaguleringsstidsområder er kun gældende, når kontrollen og kassetten er blevet
opbevaret som anvist.
20 Dansk Brugsanvisning
Hvis der ikke opnås koagulationstider inden for det vedtagne område, er brugen af både
instrumentet og testen ugyldig, indtil problemet er blevet afhjulpet.
Selv om kontrollerne ikke stammer fra mennesker, skal alle biologiske prøver anses for
at være potentielt farlige, og de skal derfor håndteres korrekt.
Det afioniserede vand er ikke sterilt. Hver vandflaske er kun beregnet til engangsbrug
(for at forhindre kontaminering).
For at sikre at kontrollens præstationer er korrekte, må opblødning kun foretages med
det medfølgende afioniserede vand. Det anbefales, at æskens indhold bevares intakt,
for at sikre, at alle komponenter altid er tilgængelige.
Sørg for at have mindst 30 minutter til rådighed til at udføre en fuld kvalitetskontrolprocedure
fra opblødning til de endelige resultater.
Hvis de opnåede resultater ikke stemmer overens med det område, der er angivet i de
specifikke præstationsegenskaber på indlægssedlen (Afsnit 9):
1. Kontrollér, at instrumentet fungerer korrekt i henhold til afsnittet Vedligeholdelse
og kvalitetskontrol i betjeningshåndbogen til instrumentet.
2. Udfør inkubation, eller afvent i 5 minutter for at sikre, at kassetten, instrumentet
og kontrollen er på 37°C ± 0,5°C (98,6°F ± 0,9°F), inden testen påbegyndes.
3. Kontrollér, at de anvendte kassetter og kontroller ikke har overskredet sidste
anvendelsesdato, og at de er blevet opbevaret korrekt.
4. Kontrollér, om kontrollen var korrekt opblødt.
5. Undersøg kontrolflasken for synlige tegn på klumpning. Hvis der findes klumpning
i flasken, kan det påvirke kontrollens præstation.
6. Kontrollér, at testen blev påbegyndt via skærmen Kvalitetskontrol, hvis det er nødvendigt.
7. Gentag testen, hvis kriterierne i punkt 1 til 6 er blevet opfyldt.
8. Rapportér eventuelle uløste problemer til Medtronic.
Dansk Brugsanvisning 21
9 Specifikke præstationsegenskaber
CLOTtrac LR unormal koagulationskontrol er en kendt opløsning, som bliver analyseret
med henblik på kvalitetskontrol, og som giver brugeren en række værdiområder. Området
bestemmes for hvert enkelt kontrolparti. Brugeren kan anvende det angivne område, eller
hvis ønsket, bestemme sine egne middelværdier, standardafvigelser og område for hvert
kontrolparti.
10 Henvisninger
1. Hattersley, P. Activated Coagulation Time of Whole Blood. JAMA. 1966;136:436.
FØLGENDE ANSVARSFRASKRIVELSE GÆLDER FOR KUNDER UDEN FOR USA:
ANSVARSFRASKRIVELSE
SELVOM CLOTtrac™ LR UNORMAL KONTROLLEN, I DET FØLGENDE KALDET
"PRODUKTET", ER BLEVET OMHYGGELIGT KONSTRUERET, FREMSTILLET OG
AFPRØVET INDEN SALG, KAN DER AF DIVERSE ÅRSAGER FINDES RISIKO FOR,
AT PRODUKTET IKKE FUNGERER TILFREDSSTILLENDE EFTER HENSIGTEN.
ADVARSLERNE, SOM FREMGÅR AF PRODUKTETS MÆRKATER, GIVER
NÆRMERE OPLYSNINGER OG BETRAGTES SOM EN INTEGRERET DEL AF
DENNE ANSVARSFRASKRIVELSE. MEDTRONIC FRASKRIVER SIG DERFOR
ALLE GARANTIER, UDTRYKKELIGE SÅVEL SOM STILTIENDE, VEDRØRENDE
PRODUKTET. MEDTRONIC KAN IKKE HOLDES ANSVARLIG FOR TILFÆLDIGE
SKADER ELLER FØLGESKADER SOM FØLGE AF BRUG AF PRODUKTET ELLER
DETS DEFEKT ELLER FEJLFUNKTION, UANSET OM KRAVET BYGGER PÅ
GARANTIRET, AFTALERET, ERSTATNINGSRET ELLER ANDET JURIDISK
GRUNDLAG.
Ovennævnte undtagelser og begrænsninger har ikke til hensigt at være, og må ikke fortolkes
som værende, i modstrid med ufravigelige bestemmelser i gældende lovgivning. Hvis noget
afnsit eller vilkår i denne ansvarsfraskrivelse af en retsinstans i en kompetent jurisdiktion
anses for ulovligt, uden retskraft eller i modstrid med gældende lovgivning, berøres de
resterende dele af ansvarsfraskrivelsen ikke, og alle rettigheder og forpligtelser skal tolkes
og håndhæves, som om denne ansvarsfraskrivelse ikke indeholdt det pågældende afsnit
eller vilkår, der anses for ugyldigt.
Dansk Brugsanvisning 23
24 Dansk Brugsanvisning
1 Verwendungszweck
Die CLOTtrac™ Low Range (LR) Abnormal-Gerinnungskontrolle dient zur Überprüfung der
einwandfreien Funktion des ACT-Geräts und der LR-ACT-Kartuschen.
Zur In-vitro-Diagnostik.
2 Kurzbeschreibung
Die Bestimmung der aktivierten Gerinnungszeit (Activated Clotting Time, ACT) wurde
erstmals 1966 von Hattersley beschrieben.1 Der Test und seine Varianten werden seitdem in
großem Umfang zur Überwachung der gerinnungshemmenden Wirkung von Heparin
eingesetzt.
Medtronic hat Gerinnungskontrollen entwickelt, um die Leistungsfähigkeit dieses Tests zu
überwachen. Die Gerinnungskontrolle besteht aus lyophilisiertem Zitratplasma und
Erythrozyten von Schafen. Die Rekonstitution der Gerinnungskontrolle erfolgt mit unsterilem
deionisiertem Wasser (Anweisungen zur Rekonstitution finden Sie in Kapitel 6 auf Seite 26).
Nach der Rekonstitution verhält sich die Abnormal-Gerinnungskontrolle wie eine frisch
entnommene Vollblutprobe des Patienten.
3 Anwendungsprinzip
Die Gerinnungskontrolle sorgt für eine verlängerte Gerinnungszeit, die genutzt werden kann,
um Qualitätssicherungsprogramme zu erfüllen, die Kontrolltests auf mindestens zwei
Ebenen verlangen. Diese Gerinnungskontrolle bietet eine der beiden Ebenen.2 Die
rekonstituierte Gerinnungskontrolle simuliert eine frisch entnommene Vollblutprobe, die
aufgrund von Heparinisierung, Verdünnung, Koagulopathie usw. eine verlängerte
Gerinnungszeit aufweist.
Deutsch Gebrauchsanweisung 25
4 Reagenzien
Die Gerinnungskontrolle besteht aus lyophilisiertem Zitratplasma und Erythrozyten von
Schafen. Die Rekonstitution der Gerinnungskontrollen erfolgt mit unsterilem deionisiertem
Wasser.
5 Lagerung und Stabilität
CLOTtrac LR-Abnormal-Gerinnungskontrollen und deionisiertes Wasser erfordern keine
spezielle Handhabung während des Versands und Transports. Die Gerinnungskontrolle
sollte nach dem Empfang bis zur Verwendung gekühlt bei 2 °C bis 10 °C (35,6 °F bis 50 °F)
aufbewahrt werden. Das Verwendbarkeitsdatum der Gerinnungskontrollen und des Wassers
ist auf dem Etikett der Ampulle angegeben. Verwenden Sie die Gerinnungskontrollen nicht,
wenn das Verwendbarkeitsdatum überschritten ist. Nach der Rekonstitution ist die
Gerinnungskontrolle bei Raumtemperatur, d. h. 15 °C bis 25 °C (59 °F bis 77 °F), 1 Stunde
lang stabil.
6 Vorgehensweise
Dieses Produkt ist nur für die Benutzung durch dafür ausgebildete Personen vorgesehen.
Im Lieferumfang enthaltene Materialien:
• Deionisiertes Wasser, 15 Ampullen pro Box
• CLOTtrac LR-Abnormal-Gerinnungskontrolle, 15 Ampullen pro Box
Erforderliche Materialien (nicht mitgeliefert):
• 3-ml-Kunststoffspritze mit Kanüle
• Medtronic ACT-Gerät
• Medtronic LR-ACT-Kartuschen
• 0,1 M CaCl
26 Deutsch Gebrauchsanweisung
2
1. Entnehmen Sie die erforderliche Anzahl an Ampullen mit CLOTtrac LR-AbnormalGerinnungskontrolle und mit deionisiertem Wasser.
2. Fügen Sie für jede Ampulle mit lyophilisierter LR-Abnormal-Gerinnungskontrolle mit
einer Spritze 1,5 ml deionisiertes Wasser hinzu.
Hinweis: Schütteln Sie die Gerinnungskontrolle erst nach vollständiger Rehydrierung.
3. Warten Sie mindestens 10 Minuten, um eine adäquate Rehydrierung zu gewährleisten.
Nichteinhaltung dieser Wartezeit kann zu einer nicht ordnungsgemäß rekonstituierten
Gerinnungskontrolle und in der Folge zu falschen Ergebnissen führen.
4. Schütteln Sie die Ampulle nach erfolgter Rehydrierung kräftig, bis sich der Aufschluss
gleichmäßig aufgelöst hat und die Gerinnungskontrolle vollständig rekonstituiert ist.
5. Klopfen Sie gegen die Kartusche, um die Kartuschenreagenzien zu resuspendieren.
6. Spezielle Anweisungen für das ACT Plus™ Gerät von Medtronic bei Verwendung von
LR-ACT-Kartuschen:
a. Wählen oder bestätigen S ie in der Anzeige "Hauptmenü", dass der Kart uschentyp auf
"LR-ACT" eingestellt ist.
b. Wählen Sie im Menü "Qualitätskontrolle (QK)" den Eintrag "LR-AB".
c. Geben Sie 0,2 ml der rekonstituierten Gerinnungskontrolle in jeden Kartuschenkanal.
Der Füllstand muss zwischen den Fülllinien der jeweiligen Kartuschenkanäle liegen.
d. Setzen Sie die Kartusche in das ACT Plus™ Gerät ein und lassen Sie sie 5 Minuten
lang erwärmen.
e. Nehmen Sie Kartusche aus dem Gerät heraus und geben Sie sofort einen Tropfen
(40 µl) der 0,1 M CaCl2-Lösung zu jedem Kartuschenkanal hinzu.
Hinweis: Halten Sie die Ampulle beim Hinzufügen der CaCl2-Lösung vertikal, um
sicherzustellen, dass ein vollständiger Tropfen CaCl2 abgegeben wird.
Deutsch Gebrauchsanweisung 27
f. Setzen Sie die Kartusche in das Gerät ein und schließen Sie den Heizblock des
Betätigungselements, um den Test zu starten.
g. Die Gerinnungsbildung wird durch einen Signalton gemeldet, der Heizblock des
Betätigungselements öffnet sich und die Ergebnisse werden angezeigt.
7 Ergebnisse
Die Ergebnisse der Kontrolltests werden in Sekunden angegeben. Diese Ergebnisse dürfen
nicht zur Interpretation von Patientenergebnissen herangezogen werden. Bei abnormalen
Werten der Gerinnungskontrolle sollte die Abweichung zwischen den Gerinnungszeiten der
Kanäle weniger als 12 % betragen.
Probenberechnung:
Aufgeführter Bereich der
Gerinnungskontrolle
Ergebnis Kanal 1=400 Sekunden
Ergebnis Kanal 2=420 Sekunden
Mittelwert=410 Sekunden
12 % des Mittelwerts=49 Sekunden
Differenz=20 Sekunden
Die Daten sind akzeptabel, weil zum einen die mittlere Gerinnungszeit sowie die Gerinnungszeiten beider Kanäle innerhalb des aufgeführten Bereichs liegen und zum anderen die
Differenz zwischen Kanal 1 und Kanal 2 (20 s) weniger als 12 % des Mittelwerts (49 s)
beträgt.
28 Deutsch Gebrauchsanweisung
=370 bis 490 Sekunden