Medtronic 53401 Quick Start Manual

MODEL 53401
TEMPORARY
EXTERNAL
TIP
PACEMAKER
Model 53401
Single Chamber Temporary Pacemaker
Replace the battery for each new patient. Do NOT reuse battery.
PRE-USE What to do ... What NOT to do ...
Battery
Do NOT ignore visible physical damage; the device may appear to work properly
immediately after being dropped or mishandled, but operational damage may
have occurred.
Check case for cracks/damage.
Check battery drawer for closure.
Check display for cracks/damage.
Physical
Condition
Do NOT use damaged leads or cables. Improper connection, displacement, or
fracture may result in pacemaker failure.
Do NOT reuse single-use cables.
Inspect cables and leads for possible defects, and secure
connection before each use.
Cables
Do NOT place the device in any area where the patient may interact with it.
Do not pla ce the tempora ry pacemaker i n any area outside
IN USE
Placement
Do NOT attempt to clean any internal surfaces, including battery compartment.
Do NOT immerse the device in water or cleaning agents; do NOT expose the unit
to ethers, acetone, or chlorinated solvents.
being dropped or mishandled, but operational damage may have occurred.
(see-throu gh plastic po cket) with
of the di rect observation by me dical staf f. If necessar y, insert
into a Disp osable Pouch
an att achment pane l (to ha ng from IV pole) th at protects
and holds the temporary pacemaker. Place the temporary
pacema ker in an area that mi nimizes access to th e controls by
unaut horized person nel, such as pat ients and visitors.
External surfaces of unit can be cleaned using a sponge or cloth
moistened with water or 70% isopropyl alcohol. For internal
surfaces, send to Medtronic for cleaning, safety, and
technical check.
Send to Medtronic for safety and technical check. Do NOT ignore; the device may appear to work appropriately immediately after
POST-USE
Clean
If dropped/
Visible exterior
fluid incursion may still occur.
Send to Medtronic for safety and technical check. Do NOT ignore; the temporary pacemaker was designed to minimize leakage, but
damage
If spilled on
Disposable Pouch available for the Model 53401 Temporary Pacemaker; Model #5409.
* Should service or repair be necessary, contact your local Medtronic representative.
2
MEDTRONIC MODEL 53401
CONTROLS AND INDICATORS
A
I
R
L
T
P
A
A
D
I
C
P
I
N
A
ATRIAL USE ONLY
RATE
-1
min
mA
SENSE
mV
G
ENABLE DISABLE
0.4 mV
R
HOLD TO DELIVER
PACE SENSE
OUTPUT
Async
1. Screen
2. Pacing and sensing status indicator LEDs
3. RAP indicator LED
4. ASYNC indicator LED
5. Battery indicator
 6. On/Obutton
7. Lock indicator
8. RATE dial
9. OUTPUT dial
10. SENSITIVITY dial
11. Lock/Unlock button
3
MEDTRONIC MODEL 53401
BASIC OPERATION
Connector Setup
1. Verify that the temporary pacemaker is turned off.
2. Plug the patient cable (5433, 5846, 5487) or surgical cable (5832S, 5833S) into the socket on the connector block on top of the temporary pacemaker.
3. Verify that the cable clicks when it is inserted into the temporary pacemaker connector receptacle. The audible click verifies that the plug is completely inserted into the receptacle.
4. Pull gently on the cable after insertion to ensure a good connection.
5. Connect the lead to the cable. Match positive (+) and negative (-) leads to positive (+) and negative (-) sockets or clips for the atrium or ventricle.
Note: To disconnec t the cable from the temporary pacemaker, do the following:
1. Press the connector release button on the patient cable plug.
2. Gently pull the plug from the receptacle.
4
MEDTRONIC MODEL 53401
BASIC OPERATION
To Turn On
§
Press and hold the On/Off button momentarily to turn on the temporary pacemaker.
The screen and the backlight illuminate, a self-test is initiated and the temporary pacemaker first searches for cardiac activity and then begins sensing and pacing (AAI/VVI mode).
§
When the temporary pacemaker is turned on, the values are:
RATE: 80 ppm (range 30-200 ppm) OUTPUT: 10 mA (range 0.1-25 mA) SENSITIVITY: 2 mV (range 0.1-20mV)
Note: If the batteries are near ing depletion, a red back light begins flashing behind the battery status indicator.
Warning: To prevent pacing i nto the vulnerable per iod of the T-wave, turn on the temporary pa cemaker and turn down OUTPU T to the minimum amplitude. Determine sensing thresholds before turning up OUTPUT to the threshold values.
To Turn Off
1. Unlock the temporary pacemaker, if it is locked (see following section).
2. Press the On/Off button for two seconds.
Note: If the On/Of f button is not held for two seconds, the tempora ry pacemaker remains on and continues to pace at the currently selected values.
5