Medtronic 5086MRI45 Technical Manual

Page 1
CAPSUREFIX MRI™ SURESCAN™ 5086MRI
Steroid-eluting, bipolar, implantable, screw-in, ventricular/atrial, transvenous lead
Technical Manual
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
Page 2
The following list includes trademarks or registered trademarks of Medtronic in the United States and possibly in other countries. All other trademarks are the property of their respective owners.
CapSureFix, CapSureFix MRI, Medtronic, Quick Twist, SureScan
Page 3

Contents

9 Device description 12 10 Medtronic warranty 13 11 Service 13

1 Description

The Medtronic CapSureFix MRI SureScan Model 5086MRI steroid-eluting, bipolar, implantable, screw-in, ventricular/atrial, transvenous lead is designed for pacing and sensing applications in either the atrium or ventricle. The lead has been designed for use in the MRI environment when used with a Medtronic MRI SureScan IPG. The platinum alloy tip and ring electrodes feature a high-active surface area of titanium nitride microstructure. This electrode configuration contributes to low polarization.
The lead has a helical tip electrode made of platinum alloy that can be actively fixed in the endocardium. The helix electrode can be extended or retracted by rotating the lead connector pin with either the Quick Twist tool attached to the lead or with the white fixation tool. An active fixation lead is particularly beneficial for patients who have smooth or hypertrophic hearts where lead dislodgement may be a potential problem. The lead also has a second, larger electrode proximal to the helical tip electrode and an IS-1 Bipolar (BI) connector1 with one terminal pin. It features MP35N nickel alloy conductors and silicone rubber insulation. The outer insulation of the lead has been treated to facilitate ease of implant.
The distal tip contains a nominal dosage of 680 µg of dexamethasone acetate. Upon exposure to body fluids, the steroid elutes from the lead tip. The steroid is known to suppress the inflammatory response that is believed to cause threshold rises typically associated with implanted pacing electrodes.
The Medtronic SureScan pacing system includes a Medtronic SureScan device connected to Medtronic SureScan leads. Labeling for SureScan pacing system components displays the SureScan symbol and the MR Conditional symbol.
SureScan logo
MR Conditional symbol. The Medtronic MRI SureScan pacing system is MR Conditional and, as such, is designed to allow implanted patients the ability to undergo an MRI scan under the specified MRI conditions for use.
The MRI SureScan feature permits a mode of operation that allows a patient with a SureScan device to be safely scanned by an MRI machine while the device continues to provide appropriate pacing. When programmed to On, MRI SureScan operation disables arrhythmia detection, magnet mode, and all user-defined diagnostics. Before performing an MRI scan,
refer to the SureScan pacing system technical manual for important information about procedures and MRI-specific warnings and precautions.

1.1 Package contents

Leads and accessories are supplied sterile. Each package contains the following items:
1 lead with radiopaque anchoring sleeve, stylet, and Quick Twist tool
1 white fixation tool
1 vein lifter
extra stylets
product documentation

1.2 Accessory descriptions

Anchoring sleeve – An anchoring sleeve secures the lead to
prevent it from moving and protects the lead insulation and conductors from damage caused by tight sutures.
White fixation tool – The white fixation tool facilitates connector pin rotation.
Quick Twist tool – The Quick Twist tool facilitates both connector pin rotation and stylet insertion into the lead. This tool comes attached to the lead.
Stylet – A stylet provides additional stiffness and controlled flexibility for maneuvering the lead into position. Each stylet knob is labeled with the stylet diameter and length.
Vein lifter – A vein lifter facilitates lead insertion into a vein.

2 Drug component description

The active ingredient in the Model 5086MRI lead is dexamethasone acetate. Dexamethasone acetate is 9-Fluoro-11β, 17,21-trihydroxy-16α-methylpregna-1,4-diene-3,20-dione 21-acetate. Dexamethasone acetate has a molecular formula of C24H31FO6 and a molecular weight of 434.50. The MCRD (Monolithic controlled release device) excipient is silicone. See Figure 1 for the structural formula.
1
IS-1 BI refers to an International Connector Standard (ISO 5841-3) whereby pulse generators and leads so designated are assured of a basic mechanical fit.
3
Page 4
Figure 1.
The target dosage of dexamethasone acetate is 680 µg per lead.

3 Indications

The Medtronic MRI SureScan lead is indicated for use as a system consisting of a Medtronic MRI SureScan IPG implanted with SureScan leads. A complete system is required for use in the MRI environmentThis lead has
application where implantable dual chamber MR Conditional pacing systems are indicated.

4 Contraindications

Use of ventricular transvenous leads is contraindicated in patients with tricuspid valvular disease.
Use of ventricular transvenous leads is contraindicated in patients with mechanical tricuspid heart valves.
Use of steroid-eluting transvenous leads is contraindicated in patients for whom a single dose of 680 µg dexamethasone acetate may be contraindicated.

5 Warnings and precautions

Before performing an MRI scan, refer to the SureScan pacing system technical manual for MRI-specific warnings and precautions. Note: Medical procedure warnings and precautions that pertain to
the Medtronic implanted system are provided in the manual that is packaged with the device or on the Medtronic Manual Library website (www.medtronic.com/manuals).
Line-powered and battery-powered equipment – An implanted lead forms a direct current path to the myocardium. During lead implant and testing, use only battery-powered equipment or line-powered equipment specifically designed for this purpose to protect against fibrillation that may be caused by alternating currents. Line-powered equipment used in the vicinity of the patient must be properly grounded. Lead connector pins must be insulated from any leakage currents that may arise from line-powered equipment.
Diathermy treatment (including therapeutic ultrasound) –
Diathermy is a treatment that involves the therapeutic heating of body tissues. Diathermy treatments include high frequency, short wave, microwave, and therapeutic ultrasound. Except for
therapeutic ultrasound, do not use diathermy treatments on cardiac device patients. Diathermy treatments may result in serious injury or damage to an implanted device and lead system. Therapeutic ultrasound (including physiotherapy, high intensity therapeutic ultrasound, and high intensity focused ultrasound), is the use of ultrasound at higher energies than diagnostic ultrasound to bring heat or agitation into the body. Therapeutic ultrasound is acceptable if treatment is performed with a minimum separation distance of 15 cm (6 in) between the applicator and the implanted device and lead system, as long as the ultrasonic beam is pointing away from the device and lead system.
Magnetic resonance imaging (MRI) – An MRI is a type of medical imaging that uses magnetic fields to create an internal view of the body. If certain criteria are met and the warnings and precautions provided by Medtronic are followed, patients with an MR Conditional device and lead system are able to undergo an MRI scan; for details, refer to the MRI Technical Manual that Medtronic provides for the MR Conditional device.
Vessel and tissue damage – Use care when positioning the lead. Avoid known infarcted or thin ventricular wall areas to minimize the occurrence of perforation and dissection.
Single use – The lead and accessories are for single use only.
Inspecting the sterile package – Inspect the sterile package
with care before opening it.
If the seal or package is damaged, contact a Medtronic representative.
Store at 25 °C (77 °F). Excursions from this storage temperature are permitted in the range of 15 to 30 °C (59 to 86 °F). (See USP Controlled Room Temperature.) According to USP excursion conditions, transient spikes up to 40 °C (104 °F) are permitted as long as they do not exceed 24 hours.
Do not use the product after its expiration date.
Sterilization – Medtronic has sterilized the package contents with ethylene oxide before shipment. This lead is for single use only and is not intended to be resterilized.
Steroid use – It has not been determined whether the warnings, precautions, or complications usually associated with injectable dexamethasone acetate apply to the use of this highly localized, controlled-release lead. For a list of potential adverse effects, refer to the Physicians’ Desk Reference.
Handling the steroid tip – Avoid reducing the amount of steroid available before implanting the lead. Reducing the available amount of steroid may adversely affect low-threshold performance.
Do not allow the electrode surface to come in contact with surface contaminants.
Do not wipe or immerse the electrode in fluid, except blood, at the time of implant.
Handling a screw-in lead – Handle the lead with care at all times.
Do not implant the lead if it is damaged. Return the lead to a Medtronic representative.
Protect the lead from materials that shed small particles such as lint and dust. Lead insulators attract these particles.
4
Page 5
Handle the lead with sterile surgical gloves that have been rinsed in sterile water or a comparable substance.
Do not severely bend, kink, or stretch the lead.
Do not immerse the lead in mineral oil, silicone oil, or any other liquid, except blood, at the time of implant.
Do not use surgical instruments to grasp the lead or connector pin.
Do not force the lead if resistance is encountered during lead passage.
Exercise the helix electrode before implanting the lead. On initial extension, more rotations may be required to extend and retract the helix electrode, or the helix electrode may extend suddenly when torque is built up. Note: The estimated number of rotations (using the fixation tool) needed to initially extend or retract the helix electrode is stated in Table 3.
Ensure helix is retracted prior to implant.
The number of turns required to extend and subsequently retract the helix may be different. Verify helix electrode extension and retraction using fluoroscopy during implant (Figure 8). Overrotation of the connector pin may result in fracture or distortion of the inner conductor or retraction of the helix electrode out of its channel.
Handling the stylet – Handle the stylet with care at all times.
Curve the stylet before inserting it into the lead to achieve a curvature at the lead’s distal end. Do not use a sharp object to impart a curve to the distal end of the stylet.
Do not use excessive force or surgical instruments when inserting the stylet into the lead.
Avoid overbending or kinking the stylet.
Use a new stylet when blood or other fluids accumulate on the stylet. Accumulated blood or other fluids may damage the lead or cause difficulty in passing the stylet into the lead.
Necessary hospital equipment – Keep external defibrillation equipment nearby for immediate use during acute lead system testing, the implant procedure, or whenever arrhythmias are possible or intentionally induced during post-implant testing.
Concurrent devices – Output pulses, especially from unipolar devices, may adversely affect device sensing capabilities. If a patient requires a separate stimulation device, either permanent or temporary, allow enough space between the leads of the separate systems to avoid interference in the sensing capabilities of the devices. Previously implanted pulse generators and implantable cardioverter defibrillators should generally be explanted.
Chronic lead removal and the SureScan pacing system –
When implanting a SureScan pacing system, consider the risks associated with removing previously implanted leads before doing so. Abandoned leads or previously implanted non-SureScan labeled leads compromise the ability to safely scan the SureScan pacing system during MRI scans.
Chronic repositioning or removal of a screw-in lead –
Proceed with extreme caution if a lead must be removed or repositioned. Chronic repositioning or removal of screw-in transvenous leads may not be possible because of blood or
fibrotic tissue development into the helix mechanism on the lead. In most clinical situations, it is preferable to abandon unused leads in place. Return all removed leads, unused leads, or lead sections to Medtronic for analysis.
Note: If a helix electrode does not disengage from the endocardium by rotating the connector pin, rotating the lead body counterclockwise may withdraw the helix electrode and decrease the possibility of damage to cardiovascular structures during removal.
Lead removal may result in avulsion of the endocardium, valve, or vein.
Lead junctions may separate, leaving the lead tip and bare wire in the heart or vein.
Chronic repositioning of a lead may adversely affect the low-threshold performance of a steroid lead.
An abandoned lead should be capped so that the lead does not transmit electrical signals.
Severed leads should have the remaining lead end sealed and the lead body sutured to adjacent tissue.
Connector compatibility – Although the lead conforms to the IS-1 International Connector Standard, do not attempt to use the lead with any device other than a commercially available implantable pacing system with which it has been tested and demonstrated to be safe and effective. The potential adverse consequences of using such a combination may include, but are not limited to, undersensing cardiac activity and failure to deliver necessary therapy.

6 Potential adverse events

The potential complications (listed in alphabetical order) related to the use of transvenous leads include, but are not limited to, the following patient-related conditions that can occur when the lead is being inserted or repositioned:
cardiac perforation
cardiac tamponade
fibrillation and other arrhythmias
heart wall rupture
infection
muscle or nerve stimulation
myocardial irritability
pericardial rub
pericarditis
pneumothorax
thrombotic and air embolism
thrombosis
valve damage (particularly in fragile hearts)
5
Page 6
Other potential complications related to the screw-in lead and the programmed parameters include, but are not limited to, the complications listed in the following table. Symptoms of the following potential complications include loss of capture or intermittent or continuous loss of capture or sensing2:
Complication
Lead dislodgement Reposition the lead.
Lead conductor or helix electrode fracture or insulation failure
Threshold elevation or exit blockaAdjust the implantable device out-
a
Evidence indicates that there is a higher frequency of exit block in the ventricle when using a screw-in lead. This should be considered when selecting a screw-in lead for use in the ventricle.
Corrective action to be consid­ered
Replace the lead. In some cases with a bipolar lead, the implantable device may be programmed to a unipolar configuration or the lead may be unipolarized.
put. Replace or reposition the lead.
Potential acute or chronic complications associated with screw-in lead placement that may require lead replacement to correct include, but are not limited to, the following:
Implant technique Potential complication
Forcing the lead through the intro­ducer
Use of too medial of an approach with venous introducer resulting in clavicle and first rib binding
Puncturing the periosteum and/or tendon when using subclavian introducer approach
Advancing the lead into the venous insertion site and/or through the veins without the stylet fully inser­ted
Bending the lead or manually pinching the lead body during extension or retraction of the helix electrode
Screw electrode damage, insula­tion damage
Conductor coil fracture, insulation damage, helix engagement issues
Conductor coil fracture, insulation damage
Tip distortion, insulation perfora­tion
Delayed torque transfer
In addition, prolonged implant procedures or repositioning the lead multiple times may allow blood or body fluids to build up on the helix mechanism. This may result in an increased number of rotations needed to extend or retract the helix electrode, which may damage the lead.

7 Drug information

7.1 Mechanism of action

Steroid suppresses the inflammatory response that is believed to cause threshold rises typically associated with implanted pacing electrodes. Dexamethasone acetate is a synthetic steroid of the
glucocorticoid family. Glucocorticoids have potent anti-inflammatory actions via direct and indirect effects on major inflammatory cells. Glucocorticosteroids bind to a cytoplasmic glucocorticoid receptor as well as a membrane-bound receptor. Binding to the cytoplasmic receptor leads to receptor activation and translocation to the nucleus. The receptor interacts with specific DNA sequences within the regulatory regions of affected genes. Thus, glucocorticoids inhibit the production of multiple cell factors that are critical in generating the inflammatory response.

7.2 Pharmacokinetics and metabolism

Pharmacokinetics – The pharmacokinetics (local drug levels
and systemic levels) of dexamethasone acetate and its metabolites following implant were not evaluated in human clinical trials. When delivered intra-muscularly, the lipid-soluble dexamethasone acetate is slowly absorbed throughout the tissue.
Metabolism – The conversion of dexamethasone acetate to dexamethasone occurs within hours. The dexamethasone alcohol (dexamethasone) is the active glucocorticoid used in this Medtronic lead. Steroid is applied via MCRD (Monolithic controlled release device) and eluted to the tissue interface where it will be used. The form of the steroid, whether it is a prodrug or the pharmacologically active dexamethasone, is irrelevant, as the steroid is directly present at the injury site to treat the inflammation. Dexamethasone acetate is hydrolyzed into dexamethasone, which is readily absorbed by the surrounding tissue and body fluids. Glucocorticoids, when given systemically, are eliminated primarily by renal excretion of inactive metabolites.

7.3 Mutagenesis, carcinogenicity, and reproductive toxicity

Mutagenesis, carcinogenicity and reproductive toxicity –
The mutagenesis, carcinogenicity, and reproductive toxicity or the Model 5086MRI lead have not been evaluated. However, the mutagenesis, carcinogenicity, and reproductive toxicity of dexamethasone acetate has been evaluated previously.
Carcinogenesis, mutagenesis, and impairment of fertility –
No adequate studies have been conducted in animals to determine whether corticosteroids have a potential for carcinogenesis (tumor initiation or promotion). Dexamethasone was genotoxic in assays for clastogenicity (including sister chromatid exchange in human lymphocytes) but not in an assay for mutagenicity in salmonella (Ames test).
Adrenocorticoids have been reported to increase or decrease the number and mobility of spermatozoa in some patients.

7.4 Pregnancy

Pregnancy – Dexamethasone acetate has been shown to be
teratogenic in many species when given in doses equivalent to the human dose. There are no adequate and well-controlled studies in pregnant women. Dexamethasone acetate should be used during pregnancy only if the potential benefit justifies the potential risk to the fetus. Studies in mice, rats, and rabbits have shown that
2
Transient loss of capture or sensing may occur for a short time following implant until lead stabilization takes place. If stabilization does not occur, lead dislodgement may be suspected.
6
Page 7
adrenocorticoids increase the incidence of cleft palate, placental
1
insufficiency, and spontaneous abortions, and can decrease the intrauterine growth rate.
Nursing mothers – Systemically administered corticosteroids appear in human milk and could suppress growth, interfere with endogenous corticosteroid production, or cause other untoward effects in nursing infants. Because of the potential for serious adverse reactions in nursing infants from corticosteroids, a decision should be made whether to discontinue nursing or to use a non-steroidal lead, taking into account the importance of the lead and the drug to the mother.

7.5 Lactation

Corticosteroids are secreted into human milk and there is a potential for serious adverse reactions. A decision should be made whether to nurse or to discontinue the drug, taking into account the importance of the drug to the mother. These potential risks of corticosteroids should also be considered along with any other steroidal therapy being received by the patient.

8 Implant procedure

Warning: When implanting a SureScan pacing system, consider
the risks associated with removing previously implanted leads before doing so. Abandoned leads or previously implanted non-SureScan labeled leads compromise the ability to safely scan the SureScan pacing system during MRI scans.
Proper surgical procedures and sterile techniques are the responsibility of the medical professional. Some implant techniques vary according to physician preference and the patient’s anatomy or physical condition.

8.1 Verifying the mechanical functioning of the helix electrode

Note: The package includes 2 tools, the Quick Twist tool attached
to the lead and the white fixation tool. Either tool may be used to verify the mechanical functioning of the helix electrode. The choice of tool is left to the discretion of the physician.
Before implant, verify the mechanical functioning of the helix electrode using the following steps:
1. Attach either the Quick Twist tool or the white fixation tool to the lead. Ensure that the stylet is inserted into the lead and proceed as indicated, according to the tool being used.
a. Quick Twist tool: Push the Quick Twist tool onto the
connector pin (Figure 2).
Figure 2.
b. White fixation tool: Press both legs of the white fixation
tool together and place the most distal hole on the connector pin (Figure 3).
Figure 3.
1 The most distal hole of the white fixation tool.
2. Keep the lead body and the IS-1 connector sleeve as straight as possible (Figure 3). Ensure that the stylet is fully inserted, then rotate the selected fixation tool clockwise until the helix electrode is fully extended (Figure 4 or Figure 5). When the helix electrode is fully extended, approximately 1.5 to 2 helix coils are exposed.
Figure 4.
7
Page 8
Figure 5.
1
Caution: Do not severely bend the IS–1 connector sleeve or
the lead body while extending the helix electrode. Caution: Overrotating the connector pin after the helix electrode is fully extended may damage the lead. The number of rotations required to extend the helix electrode increases proportionately with the length of the lead. Additional curvatures made to the stylet may increase the number of rotations needed to extend or retract the helix electrode. Refer to Table 3 for the estimated number of rotations required to extend or retract the helix electrode. Note: The number of rotations required to extend the helix electrode is variable depending on the lead path. During the initial helix electrode extension, the helix electrode may extend suddenly due to accumulated torque in the lead, or the helix electrode may require additional rotations for extension.
3. Disconnect the selected fixation tool from the connector pin and release the proximal end of the lead body. Allow several seconds for relief of the residual torque in the lead.
4. After allowing for relief of the residual torque, reattach the selected fixation tool and rotate it counterclockwise until the helix electrode tip is retracted into the sheath.

8.2 Inserting the lead

Caution: Use care when handling the lead during insertion.
Do not severely bend, kink, or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pins.
Insert the lead using the following techniques:
1. Select a site for lead insertion. The lead may be inserted by venotomy through several different venous routes, including the right or left cephalic vein or the external or internal jugular vein. Use the cephalic vein whenever possible to avoid lead damage in the first rib or clavicular (thoracic inlet) space.
Cautions:
Certain anatomical abnormalities, such as thoracic outlet syndrome, may also precipitate pinching and subsequent fracture of the lead.
When using a subclavian approach, avoid techniques that may damage the lead.
Place the insertion site as far lateral as possible to avoid clamping the lead body between the clavicle and the first rib (Figure 6).
Figure 6.
1 Suggested entry site
Do not force the lead if significant resistance is encountered during lead passage.
Do not use techniques such as adjusting the patient’s posture to facilitate lead passage. If resistance is encountered, it is recommended that an alternate venous entry site be used.
2. Insert the tapered end of a vein lifter into the incised vein and gently push the lead tip underneath and into the vein (Figure 7). Note: A percutaneous lead introducer (PLI) kit may be used to facilitate insertion. If a slittable introducer is used, it should be at least 2.6 mm (8 French). Refer to the technical manual packaged with an appropriate percutaneous lead introducer for further instructions.
Figure 7.
3. Advance the lead into the right atrium using a straight stylet to facilitate movement through the veins.

8.3 Positioning a screw-in ventricular lead

Caution: Use care when handling the lead during positioning:
Do not severely bend, kink, or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pin.
1. Insert the tapered end of a vein lifter into the incised vein and gently push the lead tip underneath and into the vein. A vein lifter facilitates lead insertion.
2. Advance the lead into the right atrium using a straight stylet to facilitate movement through the veins.
8
Page 9
3. Advance the lead through the tricuspid valve. Replacing the straight stylet with a gently curved stylet may add control in maneuvering the lead through the tricuspid valve. Then, advance the lead directly through the tricuspid valve, or project the lead tip against the lateral atrial wall and retract the curved portion of the lead body through the tricuspid valve until the lead tip enters the ventricle.
4. Position the lead in the ventricle using the following techniques. Accurate positioning of the helix electrode is essential for stable pacing. Caution: If there is reason to believe the patient has an unusually thin wall at the apex of the right ventricle, you may want to consider another site for placement of the lead. Caution: If placing the lead in or near the right ventricular apex, use caution if passing the distal end of the lead directly from the valve to the apex. This technique may result in excessive tip pressure. Caution: If an awake patient feels a twinge of pain, this may be an early sign of perforation. Using one of the following techniques may help minimize transmission of pressure directly toward the tip of the lead:
Partially withdraw the stylet so that the stylet tip is proximal to the electrode ring while positioning the lead, to minimize tip stiffness. The stylet can then be gently advanced to the tip of the lead before securing the electrode in the endocardium.
A curved stylet may be used during positioning to minimize direct pressure on the apex.
Using a curved stylet, or partially withdrawing the stylet to allow the lead to be carried by blood flow, the lead may be curved up toward the outflow tract and then allowed to fall gently into position near the apex by pulling back on
the lead body. Use fluoroscopy (lateral position) to ensure that the tip is not in a retrograde position or is not lodged in the coronary sinus.
5. After placing the lead in a satisfactory position, extend the helix electrode by following the procedure in Section 8.5.

8.4 Positioning a screw-in atrial lead

Caution: Use care when handling the lead during positioning:
Do not severely bend, kink, or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pin.
The following procedure is suggested for atrial placement of the lead:
1. Insert the tapered end of a vein lifter into the incised vein and gently push the lead tip underneath and into the vein. A vein lifter facilitates lead insertion.
2. Advance the lead into the right atrium using a straight stylet to facilitate movement through the veins. After the lead tip is passed into the atrium, replace the straight stylet with a gently curved stylet or one of the J-shaped stylets supplied with the lead.
3. Direct the lead tip into an appropriate position. Accurate positioning of the helix electrode is essential for stable pacing and sensing. Generally, a satisfactory position has the lead tip situated against the atrial endocardium in or near the apex of the appendage. As viewed on the fluoroscope (A-P view), the lead tip points medially and forward toward the left atrium. A successful position is usually achieved with an anterior, medial, or lateral tip location.
Caution: If an awake patient feels a twinge of pain, this may be an early sign of perforation.
After placing the lead tip in a satisfactory position, extend the helix electrode by following the procedure in Section 8.5.

8.5 Securing the helix electrode into the endocardium

Note: The package includes 2 tools, the Quick Twist tool attached
to the lead and the white fixation tool. Either tool may be used to secure the helix electrode into the endocardium. The choice of tool is left to the discretion of the physician.
Secure the helix electrode using the following techniques:
1. Attach either the Quick Twist tool or the white fixation tool to the lead. Ensure that the stylet is inserted into the lead and proceed as indicated, according to the tool being used.
a. Quick Twist tool: Push the Quick Twist tool onto the
connector pin (Figure 2).
b. White fixation tool: Press both legs of the white fixation
tool together and place the most distal hole on the connector pin (Figure 3).
2. Press the lead tip against the endocardium using the appropriate technique:
a. Ventricular placement: Press the lead tip against the
endocardium by gently pushing the stylet and lead at the vein entry site.
b. Atrial placement: With the lead tip advanced into the
atrium and a J-shaped or gently curved stylet in the lead, press the lead tip against the endocardium by gently pulling the stylet and the lead at the vein entry site.
3. Rotate the selected fixation tool clockwise until the helix electrode is fully extended. Caution: Do not severely bend the IS-1 connector sleeve or the lead body while extending the helix electrode.
4. Use fluoroscopy to verify helix electrode extension. This is the only reliable method to confirm extension. Extension of the space between the indicator ring (A) and the drive mechanism (B) implies complete exposure of the helix electrode (Figure 8). The top view illustrates no gap between the indicator ring and the drive mechanism (retracted) and the bottom view illustrates a gap (extended).
9
Page 10
Cautions:
Retracted
Extended
The estimated number of rotations required to fully extend or retract the helix electrode is variable. Refer to Table 3, for the estimated number of rotations.
Prolonged implant procedures or repositioning the lead multiple times may allow blood or body fluids to build up on the helix mechanism. This may result in an increased number of rotations required to extend or retract the helix electrode, which may damage the lead.
Figure 8.
5. Remove the selected fixation tool from the IS-1 connector pin and release the proximal end of the lead body. Allow several seconds for relief of the residual torque in the lead.
6. Partially withdraw the stylet.
7. Verify that the lead is affixed.
a. For a lead placed in the ventricle: Gently pull back on
the lead and check for resistance to verify affixation. A properly affixed helix electrode will remain in position. If the helix electrode is not properly affixed, the lead tip may become loose in the right ventricle.
b. For a lead placed in the atrial appendage: After the
lead tip is fixed, allow lead slack to build up in the atrium. Lead slack helps prevent tip dislodgement. Enough slack is assumed present if, under fluoroscopy, the lead assumes an “L” shape during deep inspiration. Avoid excessive slack buildup that may cause the loop of the lead to drop near the tricuspid valve.
8. If repositioning is required, reattach the selected fixation tool and rotate counterclockwise until the helix electrode is retracted. Use fluoroscopy to verify withdrawal of the helix electrode before attempting to reposition (Figure 8). Caution: Do not rotate the selected fixation tool more than the number of rotations required to fully retract the helix electrode.
9. After final positioning, remove the stylet and the Quick Twist tool completely. When removing the Quick Twist tool, grip the lead firmly just below the connector pin to help prevent lead dislodgement.
10. Obtain final electrical measurements.

8.6 Taking electrical measurements

Take electrical measurements:
1. Attach the clip of a surgical cable to the notch on the stylet guide (Figure 9).
Figure 9.
Note: A unipolar lead requires the use of an indifferent
electrode.
2. Use an implant support instrument to obtain electrical measurements. Medtronic recommends using a pacing system analyzer. For information on the use of the implant support instrument, see the product literature for that device. Satisfactory lead placement is indicated by low stimulation thresholds and adequate sensing of intracardiac signal amplitudes. Refer to Table 1 for recommended stimulation threshold and sensing amplitude measurements at implant.
A low stimulation threshold provides for a desirable safety margin, allowing for a possible rise in thresholds that may occur within 2 months following implant.
Adequate sensing amplitudes ensure that the lead is properly sensing intrinsic cardiac signals. Minimum signal requirements depend on the device’s sensitivity capabilities. Acceptable acute signal amplitudes for the lead must be greater than the minimum device sensing capabilities, including an adequate safety margin to account for lead maturity.

Table 1. Recommended measurements at implant

Ventricle Atrium
Maximum acute stimulation thresholds
Minimum acute sensing amplitudes 5.0 mV 2.0 mV
a
At pulse duration setting of 0.5 ms.
a
1.0 V
3.0 mA
1.5 V
4.5 mA
3. If electrical measurements do not stabilize to acceptable levels, repositioning the lead and repeating the testing procedure may be necessary. Note: Initial electrical measurements may deviate from the recommendations because of acute cellular trauma. If such a deviation occurs, wait 5 to 15 min and repeat the testing procedure. Values may vary depending upon lead type, device settings, cardiac tissue condition, and drug interactions.

8.6.1 Checking diaphragmatic stimulation for screw-in leads

Diaphragmatic stimulation should also be checked by pacing at 10 V and observing on fluoroscopy whether the diaphragm contracts with each paced stimulus. If diaphragmatic stimulation occurs, reposition the lead.
10
Page 11
8.6.2 Taking pacing impedance (or resistance)
1
measurements
Pacing impedance (or resistance) is used to assess device function and lead integrity during routine device patient follow-up sessions and to assist in troubleshooting suspected lead failures. Additional troubleshooting procedures include ECG analysis, visual inspection, measurement of thresholds, and electrogram characteristics.
Pacing impedance values are affected by many factors including lead position, electrode size, conductor design and integrity, insulation integrity, and the patient’s electrolyte balance. Apparent pacing impedance is also significantly affected by the measurement technique. Comparison of pacing impedance should be done using consistent methods of measurements and equipment.
An impedance higher or lower than the typical values is not necessarily a conclusive indication of a lead failure. Other causes must be considered as well. Before reaching a conclusive diagnosis, the full clinical picture must be considered. The full clinical picture includes pacing artifact size and morphology changes in 12-lead analog ECGs, muscle stimulation with bipolar leads, sensing and/or capture problems, patient symptoms, and device characteristics.
Recommendations for clinically monitoring and evaluating leads in terms of impedance characteristics are listed below.
Consider the following recommendations for devices with telemetry readout of impedance:
Routinely monitor and record impedance values at implant and follow-up sessions using consistent output settings. Note: Impedance values may be different at different programmable output settings (for example, pulse width or pulse amplitude) of the device or pacing system analyzer.
Establish a baseline chronic impedance value once the impedance has stabilized, generally within 6 to 12 months after implant.
Monitor for significant impedance changes and abnormal values.
Where impedance abnormalities occur, closely monitor the patient for indications of pacing and sensing problems. The output settings used for measuring impedance should be the same as those used for the original measurements.
For patients at high risk, such as implantable device-dependent patients, consider further action such as increased frequency of monitoring, provocative maneuvers, and ambulatory ECG monitoring.
Consider the following recommendations for devices without telemetry:
Record the impedance value at implant. Also record the measurement device, its output settings, and the procedure used.
At the time of device replacement, if pacing system analyzer-measured impedance is abnormal, carefully
evaluate lead integrity (including thresholds and physical appearance) and patient condition before electing to reuse the lead.
Impedances below 250 Ω may result in excessive battery current drain, which may seriously compromise device longevity, regardless of lead integrity.
For more information on obtaining electrical measurements, consult the product literature supplied with the testing device.

8.7 Anchoring the lead

Cautions:
Use care when anchoring the lead.
Use an anchoring sleeve with all leads.
Do not use absorbable sutures to anchor the lead.
Do not secure the sutures so tightly that they damage the vein, lead, or anchoring sleeve.
Do not use the anchoring sleeve tabs for suturing (Figure 10).
Do not tie a suture directly to the lead body (Figure 11).
Do not dislodge the lead tip.
Do not attempt to remove or cut the anchoring sleeve.
Do not remove the tabs on anchoring sleeves. Tabs are provided to minimize the possibility of the sleeve entering the vein.
If a large diameter percutaneous lead introducer (PLI) sheath is used, take extreme care to prevent passage of the anchoring sleeve into the PLI lumen or the venous system.
Figure 10.
1 Anchoring sleeve tab
Figure 11.
With a triple groove anchoring sleeve, generally 2 or 3 of the grooves may be used with the following procedure.
Note: The anchoring sleeves contain a radiopaque substance, which allows visualization of the anchoring sleeve on a standard x-ray and may aid in follow-up examinations.
Anchor the lead:
1. Position the anchoring sleeve close to the lead’s connector pin to prevent inadvertent passage of the sleeve into the vein.
2. Insert the anchoring sleeve partially into the vein.
3. Use the most distal suture groove to secure the anchoring sleeve to the vein.
11
Page 12
4. Use the middle groove to secure the anchoring sleeve to the fascia and lead (Figure 12):
a. Create a base by looping a suture through the fascia
underneath the middle groove and tying a knot.
b. Firmly wrap the suture around the middle groove and tie
a second knot.
Figure 12.
5. If anchoring with all 3 grooves, use the third and most proximal groove to secure the anchoring sleeve to the lead body (Figure 13).
Figure 13.

8.8 Connecting the lead

Caution: Always remove the stylet and stylet guide before
connecting the lead to the device. Failure to remove the stylet and stylet guide may result in lead failure.
Connect the lead to the device:
1. Carefully and completely remove the stylet and stylet guide. Note: When removing the stylet and stylet guide, firmly grip the lead just below the connector pin to help prevent possible lead dislodgement.
2. Obtain final electrical measurements.
3. Insert the lead connector into the connector block on the device. For instructions on proper lead connections, see the product documentation supplied with the device.

8.9 Placing the device and lead into the pocket

Cautions:
Use care when placing the device and lead into the pocket.
Ensure that the lead does not leave the device at an acute angle.
Do not grip the lead or device with surgical instruments.
Do not coil the lead (Figure 14). Coiling the lead can twist the lead body and may result in lead dislodgement.
Figure 14.
Caution: To prevent undesirable twisting of the lead body, wrap
the excess lead length loosely under the device and place both the device and the lead into the subcutaneous pocket.
Place the device and lead into the pocket:
1. Rotate the device to loosely wrap the excess lead length under the device (Figure 15).
Figure 15.
2. Insert the device and lead into the pocket.
3. Suture the pocket closed.
4. Monitor the patient’s electrocardiogram until the patient is discharged. If a lead dislodges, it usually occurs during the immediate postoperative period.

9 Device description

9.1 Specifications (nominal)

Table 2. Specifications (nominal)

Parameter Model 5086MRI
Type Bipolar
Chamber Atrium/Ventricle
Fixation Screw-in
Lengths 45, 52, 58 cm
Connector IS-1 BI
Materials Conductor: MP35N nickel alloy
Insulation: Treated silicone rubber
Connector pin: Stainless steel
Connector ring: Stainless steel
Electrode materi­als
Helix electrode: Titanium nitride coated plati-
num alloy
Ring electrode: Titanium nitride coated plati-
num alloy
12
Page 13
Table 2. Specifications (nominal) (continued)
Parameter Model 5086MRI
Electrode surface area
Tip to ring spacing 10 mm
Diameter Lead body: 2.3 mm
Lead introducer (recommended size)
Helix length (fully extended) 1.8 mm
Resistance Unipolar: 70.0 Ω (58 cm)
Steroid Type: Dexamethasone acetate
Nominal dosage of steroid 680 µg (target dose)
Steroid binder Silicone
Helix: 4.2 mm
Ring: 24 mm
without guide
wire:
with guide wire: 3.7 mm (11 French)
Bipolar: 105.0 Ω (58 cm)
2
2
2.7 mm (8 French)

Table 3. Estimated number of rotations required to extend or retract the helix electrode for initial placement

Lead length Straight stylet J-shaped stylet
45 cm 12 18
52 cm 13 20
58 cm 14 21
a
The number of rotations required to extend the helix electrode is variable depending on the lead path. In addition, the number of turns required to extend and subsequently retract the helix may be different. Verify extension and retraction with fluoroscopy.
a

10 Medtronic warranty

For complete warranty information, see the accompanying warranty document.

11 Service

Medtronic employs highly trained representatives and engineers located throughout the world to serve you and, upon request, to provide training to qualified hospital personnel in the use of Medtronic products. Medtronic also maintains a professional staff to provide technical consultation to product users. For more information, contact your local Medtronic representative, or call or write Medtronic at the appropriate telephone number or address listed on the back cover.
13
Page 14
Page 15
Page 16
Medtronic, Inc.
*M955299A001*
710 Medtronic Parkway Minneapolis, MN 55432 USA www.medtronic.com +1 763 514 4000
Medtronic USA, Inc.
Toll-free in the USA (24-hour technical consultation for physicians and medical professionals) Bradycardia: +1 800 505 4636 Tachycardia: +1 800 723 4636
Europe/Middle East/Africa
Medtronic International Trading Sàrl Route du Molliau 31 Case Postale 84 CH-1131 Tolochenaz Switzerland +41 21 802 7000
Technical manuals
www.medtronic.com/manuals
© 2015 Medtronic, Inc. M955299A001 A 2015-09-23
Loading...