Medtronic 5038-65 Technical Manual

Page 1
CAPSURE® VDD-2 5038
Steroid eluting, implantable, tined, bipolar atrial sensing, bipolar ventricular sensing, and pacing transvenous lead
Technical Manual
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
Page 2
The following list includes trademarks or registered trademarks of Medtronic in the United States and possibly in other countries. All other trademarks are the property of their respective owners.
CapSure, Medtronic, Thera
Page 3

Contents

1 Description 3 2 Indications 4 3 Contraindications 4 4 Warnings and precautions 4 5 Potential complications 5 6 Adverse events 5 7 Clinical trials summary 6 8 Directions for use 7
9 Detailed device description 10 10 Service 12 11 Medtronic warranty 12

1 Description

The Medtronic CapSure VDD-2 Model 5038 steroid eluting, implantable, tined, bipolar atrial sensing, bipolar ventricular sensing and pacing, transvenous lead is designed for bipolar sensing in the atrium, and bipolar sensing and pacing in the ventricle. The CapSure VDD-2 Model 5038 lead is intended for use with current Medtronic VDD pulse generators.
The lead has two ring electrodes for bipolar sensing in the atrium and a ring electrode and a tip electrode for bipolar sensing and pacing in the ventricle. All electrodes are composed of platinum alloy and the tip electrode features a hemispherical microporous surface composed of platinum that has been coated with steroid dexamethasone sodium phosphate.
The tip electrode contains a maximum of 1.0 mg of dexamethasone sodium phosphate, a portion of which is in a silicone rubber binder. Upon exposure to body fluids, the steroid elutes from the electrode. The lead is designed to provide low chronic pacing thresholds via steroid treatment of cardiac tissue near the lead tip. Steroid suppresses the inflammatory response that is believed to cause threshold rises typically associated with implanted pacing electrodes. The lower threshold provided by the steroid may permit reduced amplitude and pulse width settings in programmable pulse generators thereby increasing pacemaker longevity.
The lead also features four silicone tines near the electrode tip, nickel-alloy conductor coils and treated silicone outer insulation.
The lead comes with a bipolar ventricular connector, indicated by an embedded IS-1 BI VEN label, and a bipolar atrial connector, indicated with an ATRIAL SENSE label. The bipolar ventricular connector is an IS-11 bipolar (BI) connector. The ATRIAL SENSE
connector is a bipolar connector which meets the mechanical and functional requirements of the IS-1 standard, but is only designed for sensing functions.
The connector pin of the ventricular connector is electrically connected to the ventricular electrode tip, and the connector ring of the ventricular connector is electrically connected to the ventricular electrode ring. The connector pin of the atrial connector is electrically connected to the distal atrial electrode ring, and the connector ring of the atrial connector is electrically connected to the proximal atrial electrode ring.
Only the ventricular connector allows the use of a stylet wire.
The following standard combinations of lead length and A-V spacing are available:
Spacing between ventricular tip and
Lead model Lead length
5038S-52 52 cm 115 mm
5038-58 58 cm 135 mm
5038S-58 58 cm 115 mm
5038L-65 65 cm 155 mm
5038-65 65 cm 135 mm
center of atrial ring pair
The Model 5038-58, lead with 58 cm lead length and 135 mm spacing; and the Model 5038-65 lead, with 65 cm lead length and 135 mm spacing, should be appropriate for the majority of patients with average heart sizes. An X-ray of the patient’s heart may be useful in estimating the heart size and determining whether the length for the tip electrode to the midpoint of the atrial rings is appropriate. The Model 5038S-58 with 58 cm lead and 115 mm spacing and the Model 5038L-65, with 65 cm lead length and 155 mm spacing, may be useful for patients with unusually small or large hearts.
Note: Special care must be taken when prescribing this system for pediatric patients by assuring that the A-V spacing is appropriate for the heart size, that the atrial rings are placed within the atrium, and that adequate P-wave values are obtained.

1.1 Package contents

The lead and accessories are supplied sterile. Each package contains the following items:
1 lead with anchoring sleeve, stylet, and stylet guide
1 vein lifter
extra stylets
product documentation
1
IS-1 refers to an International Connector Standard (ISO 5841-3) whereby pulse generators and leads so designated are assured of a basic mechanical fit.
3
Page 4

2 Indications

The Model 5038 lead is designed to be used with Medtronic VDD pulse generators as part of a cardiac pacing system. The lead has application where implantable atrial sensing and ventricular pacing and sensing dual chamber pacing systems are indicated.

3 Contraindications

When tricuspid valvular disease is present, use of a ventricular transvenous lead is contraindicated.
The use of endocardial ventricular leads are contraindicated in patients with mechanical tricuspid heart valves.
Do not use steroid eluting leads in patients for whom a single dose of 1.0 mg dexamethasone sodium phosphate may be contraindicated.

4 Warnings and precautions

Sterilization – Medtronic has sterilized the package contents
with ethylene oxide before shipment. This device is for single use only and is not intended to be resterilized.
Single use – The lead and accessories are for single use only.
Line-powered and battery-powered equipment – An
implanted lead forms a direct current path to the myocardium. During lead implant and testing, use only battery-powered equipment or line-powered equipment specifically designed for this purpose to protect against fibrillation that may be caused by alternating currents. Line-powered equipment used in the vicinity of the patient must be properly grounded. Lead connector pins must be insulated from any leakage currents that may arise from line-powered equipment.
Diathermy treatment (including therapeutic ultrasound) –
Diathermy is a treatment that involves the therapeutic heating of body tissues. Diathermy treatments include high frequency, short wave, microwave, and therapeutic ultrasound. Except for therapeutic ultrasound, do not use diathermy treatments on cardiac device patients. Diathermy treatments may result in serious injury or damage to an implanted device and leads. Therapeutic ultrasound is the use of ultrasound at higher energies than diagnostic ultrasound to bring heat or agitation into the body. Therapeutic ultrasound is acceptable if treatment is performed with a minimum separation distance of 15 cm (6 in) between the applicator and the implanted device and leads.
Necessary hospital equipment – Defibrillating equipment should be kept nearby for immediate use during the implantation procedure.
Inspecting the package – Inspect the lead sterile package prior to opening. If the seal or package is damaged, contact your local Medtronic representative.
If the container, the sterile package, or the lead appear to have been damaged do not use the lead. Contact your Medtronic representative and return the damaged lead.
Before inserting the lead – Use an anchoring sleeve with all leads. Ensure that the anchoring sleeve is positioned close to the lead’s bifurcation sleeve at the connector side. This will prevent inadvertent passage of the sleeve into the vein.
Handling the lead – Handle the lead with care at all times.
Leads should be handled with great care at all times. Any severe bending, kinking, stretching, handling with surgical instruments or excessive force when inserting a stylet may cause permanent damage to the lead. If the lead is damaged, do not implant. Return the lead to your Medtronic representative.
Lead insulators attract small particles such as lint and dust. Therefore, to minimize contamination, protect the lead from materials shedding these substances. Handle the lead with sterile surgical gloves that have been rinsed in sterile water or equivalent.
Do not immerse leads in mineral oil, silicone oil or any other liquid.
Steroid elution – It has not been determined whether the warnings, precautions, or complications usually associated with injectable dexamethasone sodium phosphate apply to the use of this highly localized, controlled-release device. For a list of potential adverse effects, refer to the Physicians’ Desk Reference.
Do not allow the electrode surface to come in contact with surface contaminants. Do not wipe or immerse the electrode in fluid. Such treatment of a steroid eluting lead will reduce the amount of steroid available when the lead is implanted and may adversely affect low-threshold performance.
Chronic repositioning – Chronic repositioning or removal of tined leads may be difficult because of fibrotic tissue development. Chronic removal can result in avulsion of the endocardium, valve or vein. In addition, the lead conjunctions may separate, leaving the lead tip and bare wire in the heart or vein. In most clinical situations, it is preferable to abandon unused leads in place.
However, if a lead must be removed or repositioned, proceed with extreme caution. If a lead is removed, inspect it carefully for insulator or conductor coil damage. (Medtronic requests all removed or unused leads or portions thereof be returned for analysis.) If a lead is abandoned, it should be capped to avoid transmitting electrical signals from the pin to the heart. A lead that has been cut off should have the remaining lead end sealed and it should be sutured to adjacent tissue to avoid migration into the heart.
Chronic repositioning may adversely affect a steroid lead’s low-threshold performance.
4
Page 5
See additional cautions in the section titled “Directions for use.”

5 Potential complications

The potential complications (listed in alphabetical order) related to the use of transvenous leads include, but are not limited to, the following patient-related conditions that can occur when the lead is being inserted or repositioned:
air embolism
cardiac perforation
cardiac tamponade
fibrillation and other arrhythmias
heart wall rupture
infection
muscle or nerve stimulation
pericardial rub
pneumothorax
valve damage (particularly in fragile hearts, e.g. infants)
venous thrombosis leading to stenosis or emboli
Other potential complications related to the lead and the programmed parameters include, but are not limited to, the following:
Complication Symptom
Lead dislodgement Intermittent or contin-
uous loss of ventricu­lar capture or sens­inga, atrial sensing, false A-V triggering
Lead conductor frac­ture or insulation fail­ure
intermittent or contin­uous loss of ventricu­lar capture or sens­inga, or atrial sensing, or atrial and ventricu­lar oversensing
Threshold elevation or exit block
a
Transient loss of capture or sensing may occur for a short time following
Loss of ventricular
a
capture
surgery until lead stabilization takes place. If stabilization does not occur, lead dislodgement may be suspected.
Potential procedural complications during lead placement include, but are not limited to, the following:
Corrective action to be considered
Reposition the lead.
Replace the lead. In some cases with a bipolar lead, the implantable device may be programmed to a unipolar configu­ration or the lead may be unipolarized.
Adjust the pulse gen­erator output. Replace or reposition the lead.
Procedural compli­cation Problem
Forcing the lead through the introducer
Electrode, tines fixa­tion, or insulation
Corrective action to be considered
Replace the lead
damage
Binding of the lead between the clavicle and first rib
Potential conductor coil fracture, insula­tion compression
Replace the lead
fracture, polyurethane degradation
Periosteum and/or tendon puncture
Advancing withdraw stylet into lead simul­taneously with lead advancement
Conductor coil frac­ture
Tip distortion and per­foration
Replace the lead
Keep the stylet fully inserted until the lead is advanced into the heart

6 Adverse events

6.1 Adverse events

The clinical investigation of the Medtronic Thera VDD/CapSure VDD2 system involved 178 systems implanted in 178 patients for a total of 2100 cumulative device months of experience (mean =
11.8 months). Thirteen patients died during the course of the clinical study. None of the deaths were judged to be related to the device. Adverse events (AEs), including 9 complications and 29 observations, were reported during the clinical investigation. The following table reports these data on a per patient and a per patient-month basis.
# of
Type of AE
patients
(n=178)
% of
patients
Observationsa(29)
Angina 1 0.56% 1 .0005 2100
Atrial Over-
1 0.56% 1 .0005 2100
sensing
Atrial Under-
16 8.99% 16 .0076 131
sensing
Dizziness 3 1.69% 3 .0014 700
Elevated
2 1.12% 2 .0010 1050 Pacing Thresholds
Fatigue 2 1.12% 2 .0010 1050
Palpitations 2 1.12% 2 .0010 1050
Pain/Swel-
2 1.12% 2 .0010 1050 ling at Pocket Site
Complicationsb (9)
AE
#
s
AE/Pt-M
O
(n=210
0)
Pt Mos
Between
AEs
2
This data was obtained with the Medtronic Model 5032 CapSure VDD lead in a clinical study which concluded in February 1995. Clinical studies were not performed on the Model 5038 CapSure VDD-2 lead due to its similarity in design and function to the Model 5032 lead. The clinical data collected for the Model 5032 lead therefore supports the safety and efficacy claims for the Model 5038 lead and is included here for reference purposes. The Model 5038 lead differs from the Model 5032 lead in the following minor ways: downsized bifurcation area, smaller introducer size (9 French without guidewire; with guidewire, 12 French), increased stylet socket depth.
5
Page 6
# of
Type of AE
Lead reposi­tioned
Pulse gener­ator reposi­tioned
Reprogram­med to VVI/R due to loss of atrial sensing
a
Observations are adverse events which are correctable by noninvasive measures, e.g., reprogramming.
b
Complications are adverse events requiring invasive measures to correct (e.g., surgical intervention, programming out of VDD mode) or the loss of a significant function of the device.
patients
(n=178)
6 3.40% 6 .0029 350
1 0.56% 1 .0005 2100
2 1.12% 2 .0010 1050
% of
patients
AE
#
s
AE/Pt-M
O
(n=210
0)
Pt Mos
Between
AEs
Other possible side effects include, but are not limited to, body rejection phenomena including local tissue reaction, fibrotic tissue formation, muscle and nerve stimulation, infection, erosion of pacemaker lead through skin, myopotential sensing, transvenous lead-related thrombosis, embolism, and cardiac tamponade. Also, see the section Chapter 5.

7 Clinical trials summary

7.1 Conclusion

The Medtronic Thera VDD/CapSure VDD pacing system3 was proven to be safe and effective. VDD mode retention rate was
98.8% (excluding patients who were reprogrammed to VVI/R due to changing sinus condition). With the enhanced 0.18 mV capability, VDD mode retention rate is 99.4%. These results suggest that the Thera VDD system provides stable and reliable chronic atrial sensing performance.

7.1.1 Primary objectives

Primary Objective One: Demonstrate the safety of the Thera VDD/CapSure VDD pacing system.
Primary Objective Two: Demonstrate effective atrial sensing and tracking provided by the Thera VDD/CapSure VDD pacing system.

7.1.2 Methods

Prior to study enrollment, all patients were screened for adequate sinus node competence by exercise testing, hall walk, ECG or Holter monitoring. Patients where randomly assigned to either the Thera VDD/CapSure VDD system or a two-lead control system in a 1:1 ratio.
Controls were implanted with new, standard atrial and ventricular leads. Active or passive fixation leads were allowed in the atrium. Bipolar, passive fixation, steroid eluting leads (e.g., Medtronic Model 5024 or Medtronic 4024) were required in the ventricle.

7.1.3 Results

Forty-three centers worldwide are participating in an ongoing clinical study to evaluate atrial sensing performance with the Medtronic single-lead VDD system (Thera VDD/CapSure VDD). One hundred seventy-eight patients have been implanted with the VDD system, representing 2100 device-months of experience. Mean follow-up (FU) time is 11.8 months. Mean age is 67.3 ± 14.8; range 10–87 years. Typical indications are 2° or 3° AV block with normal sinus node function.
Related to Primary Objective One, there were no unanticipated Adverse Device Events and the device related complication survival was not statistically different from that of the Control (P=
0.6035).
Related to Primary Objective Two, there were no postoperative CapSure VDD lead explants or cap/abandonments. Six leads were repositioned. 156 patients continue to be followed in the study (9 lost-to-follow-up and 13 patient deaths, none related to the pacing system or implantation). Mean minimum P-wave thresholds remained stable over time: 1.0 mV at 1 month, 1.1 mV at 3 months, 1.1 mV at 6 months and 1.1 mV at one year. Percent of synchronization was calculated from Event Counter data at the 6 month FU. More than half of the patients had 100% AS-VP (see Figure 1). Temporary P-wave undersensing was resolved for several patients at pre-discharge by adjusting sensitivity. Two patients were reprogrammed to VVIR due to loss of atrial sensing, but since the more sensitive 0.18 mV setting has become available, one has regained VDD mode. Twelve patients developed atrial arrhythmias (2 brady and 10 tachy) of which 3 were reprogrammed to VVI/R and the other 9 remained in VDD with or without Mode Switching ON.
Note: The ability to consistently sense P-waves with the VDD system is dependent upon proper lead positioning, with the atrial bipole in close proximity to the atrial wall. In the Thera VDD controlled clinical trial, the observed rate of programming out of VDD mode due to loss of atrial sensing was 1.1% higher in the VDD group compared to the standard two-lead DDD group.
3
This data was obtained with the Medtronic Model 5032 CapSure VDD lead in a clinical study which concluded in February 1995. Clinical studies were not performed on the Model 5038 CapSure VDD-2 lead due to its similarity in design and function to the Model 5032 lead. The clinical data collected for the Model 5032 lead therefore supports the safety and efficacy claims for the Model 5038 lead and is included here for reference purposes. The Model 5038 lead differs from the Model 5032 lead in the following minor ways: downsized bifurcation area, smaller introducer size (9 French without guidewire; with guidewire, 12 French), increased stylet socket depth.
6
Page 7

Figure 1. Percent of synchronization (AS-VP) at 6 months FU

0
25
50
75
100
Percent of AS-VP
Percent of patients
(cumulative)
n=126
100%
>98%
>95%
>90%
>80%
1

8 Directions for use

The implantation procedure generally includes using a stylet guide, stylets, and vein lifter; selecting an insertion site; positioning the lead; and connecting the lead to the pulse generator.
Some implantation techniques vary according to physician preference and the patient’s anatomy or physical condition. As described below, the techniques for selecting an insertion site, using a stylet guide or a stylet, positioning a lead, and anchoring a lead suggest one or two possible versions, while more may exist.

8.1 Using a stylet guide and stylets

The lead is packaged with the stylet guide attached to the connector pin and a stylet already inserted. If the stylet guide has been removed, replace it by gently pushing it as far as possible onto the connector pin of the ventricular connector leg (see Figure 2), which is indicated with an IS-1 BI VEN label.
Caution: A stylet can only be introduced via the ventricular connector pin.
Figure 2.
A stylet provides additional stiffness and controlled flexibility for maneuvering the lead into position. Stylets vary in diameter and length to accommodate a physician’s preference.
Insert a stylet wire through the stylet guide. If a slight curve is needed for the stylet, refer to the instructions in the section titled “Lead Positioning in the Right Ventricle.”
Caution: To avoid damage to the lead or body tissue, do not use excessive force or surgical instruments to insert a stylet into the lead. To avoid lead tip distortion, the stylet should always remain fully inserted into the lead during lead introduction and while advancing the lead, especially through tortuous veins that may cause the stylet to “back out” of the lead. When handling a stylet,
avoid overbending, kinking or blood contact. If blood is allowed to accumulate on a stylet, passage of the stylet into the lead may be difficult.

8.2 Selecting an insertion site

The lead may be inserted by venotomy through several different venous routes, including the right or left cephalic vein, other subclavian branches, or the external or internal jugular vein. The lead may also be inserted into a subclavian vein through a percutaneous lead introducer. Select the desired entry site (see Figure 3).
Caution: When using a subclavian approach, avoid placing the entry site in a location where the lead body can be clamped between the clavicle and the first rib. A more lateral approach is recommended to avoid first rib clavicular crush. Clamping of the lead may eventually cause the conductor to fracture and may cause other damage to the lead. Certain anatomical abnormalities, such as thoracic outlet syndrome, may also precipitate clamping of the lead.
Use fluoroscopy to facilitate accurate lead placement.

Figure 3. Suggested point of insertion

1 Suggested entry site

8.3 Using the vein lifter

A vein lifter facilitates lead introduction. Insert its tapered end into the incised vein and gently push the lead tip underneath and into the vein (see Figure 4).
Caution: Avoid placing the lead under extreme tension or angualtion to prevent lead dislodgement and possible lead fracture. Avoid gripping the lead with surgical instruments.

Figure 4. Using the vein lifter

7
Page 8

8.4 Lead positioning in the right ventricle

Advance the lead into the right atrium. For added control in maneuvering the lead tip through the tricuspid valve, the distal end of the lead may be slightly curved by inserting a gently curved stylet (see Figure 5). The lead tip may then be advanced directly through the valve, or it can be projected against the lateral atrial wall and the curved portion of the lead backed across the tricuspid valve.

Figure 5. Imparting a curve to the stylet with a smooth object

Caution: To avoid damange to the stylet, do not use a sharp
object to impart a curve to the distal end.
Due in part to the tines and the flexible nature of the lead, difficulty may be encountered when passing the lead through the tricuspid valve or its chordae tendineae. One way to minimize difficulty is to rotate the stylet or lead about on its axis as it is advanced or withdrawn through the valve.
After the lead tip is passed into the right ventricle, the curved stylet may be replaced with a straight stylet. To avoid using excessive tip force while achieving final electrode position, withdraw the stylet slightly or back the distal lead tip out of the pulmonary outflow tract. To minimize the occurrence of perforation, avoid known infarcted or thin ventricular wall areas.
Accurate positioning and wedging of the electrode tip are essential for stable endocaridal pacing of the right ventricle. A satisfacotry position is achieved when the lead tip points straight toward the apex or when the dispal end dips or bends slightly (see Figure 6).

Figure 6. Typical placement of the tined lead near the ventricular apex (A-P view).

After positioning and wedging the ventricular electrode tip, place the atrial electrode ring pair in close proximity to the myocardium. Base the optimum position on the P-wave amplitude measurements where either the most consistent P-wave amplitudes or the largest minimum P-wave are attained during deep breathing. This position may be verified by fluoroscopy.
Care should be taken to avoid placement in the proximity of the A-V valve or of the inflow tract of the vena cava. The system is completed with connection of this lead to a specific type of pulse generator which provides the necessary high atrial sensitivity (up to 0.25 mV level).
Use fluoroscopy (lateral position) to ensure that the tip is not in a retrograde position or is not lodged in the coronary sinus.

8.5 Taking electrical measurements

A notch in the stylet guide allows connection of a surgical cable on the ventricular connector pin for electrical measurements (see Figure 7).

Figure 7. Surgical cable connection

Low ventricular stimulation thresholds and adequate sensing of intracardiac signal amplitudes indicate satisfactory lead placement. Medtronic recommends using a voltage source such as a pacing system analyzer for obtaining electrical measurements.
Small P-wave amplitudes may be confused with other electrical signals, like far-field QRS complexes. This may result in false readings. Therefore, Medtronic advises the recording of the intra-atrial EGM to verify the correctness of the pacing system analyzer readings.
A low ventricular stimulation threshold provides for a desirable safety margin, allowing for a possible rise in ventricular thresholds that may occur within two months following implantation.
Adequate sensing amplitudes ensure that the lead is properly sensing intrinsic cardiac signals. Minimum signal requirements depend on the pulse generator’s sensitivity capabilities. Acceptable acute signal amplitudes for the lead must be greater than the minimum pulse generator sensing capabilities including an adequate safety margin to account for lead maturity. Recommended values are listed in the next table.
8
Page 9
Before taking final measurements, carefully remove the stylet and
1
stylet guide completely. When removing the stylet guide, grip the lead firmly just below the connector pin; this will help prevent possible lead tip dislodgement.

Table 1. Recommended electrical measurements at implant when using a pacing system analyzer

Measurement required
Maximum acute stimulation thresholds
Minimum acute sensing ampli­tudes
a
At pulse duration setting of 0.5 ms.
b
Take the minimum P-Wave during deep inspiration.
a
Ventri­cle Atrium
1.0 V
3.0 mA——
5.0 mV 1.0 mV
b
Initial ventricular electrical measurements may deviate from the recommendations because of acute cellular trauma. If this occurs, wait five to fifteen minutes and repeat the testing procedure. Values may vary depending upon pulse generator settings, cardiac tissue condition, and drug interactions. If electrical measurements do not stabilize to acceptable levels, it may be necessary to reposition the lead and to repeat the testing procedure.

8.6 Pacing impedance for ventricular electrode

Ventricular pacing impedance (or resistance) is used to assess pulse generator function and lead integrity during routine pulse generator patient follow-up and to assist in troubleshooting suspected lead failures (additional troubleshooting procedures include ECG analysis, visual inspection, measurement of thresholds, and electrogram characteristics).
Pacing impedance values are affected by many factors including lead position, electrode size, conductor design and integrity, insulation integrity, and the patient’s electrolyte balance. Apparent pacing impedance is significantly affected by measurement technique; therefore, comparison of pacing impedance should be done using consistent methods of measurements and equipment.
An impedance higher or lower than the typical values is not necessarily a conclusive indication of a lead failure. Other causes must be considered as well. Before reaching a conclusive diagnosis, the full clinical picture must be considered: pacing artifact size and morphology changes in 12-lead analog ECGs, muscle stimulation with bipolar leads, sensing and/or capture problems, patient symptoms, and pulse generator characteristics.
Medtronic has developed the following recommendations for clinically monitoring and evaluating leads in terms of impedance characteristics.
For pulse generators with telemetry readout of impedance:
Routinely monitor and record impedance values at implant and follow-ups. All current Medtronic VDD pulse generators provide impedance values using consistent output settings (note, impedance values may be different at different
programmable output settings [e.g. pulse width or pulse amplitude] of the pulse generator or pacing system analyzer).
Establish a baseline chronic impedance value once the impedance has stabilized, generally within 6-12 months after implant.
The Medtronic VDD pulse generator will automatically report lead impedance values which fall outside of the established range (see the technical manual delivered with the Medtronic VDD pulse generator for more information).
Where impedance abnormalities occur, closely monitor the patient for indications of pacing and sensing problems.
For patients at high risk, such as pulse generator-dependent patients, physicians may want to consider further action such as increased frequency of monitoring, provocative maneuvers, and ambulatory ECG monitoring.
For more information on obtaining electrical measurements, consult the technical manual supplied with the testing device.

8.7 Anchoring the lead

Use the slotted triple groove anchoring sleeve to secure the lead from moving and to protect the lead insulation and conductor coil from damage caused by tight ligatures.
Anchor the lead with nonabsorbable sutures.
Caution: Tabs on anchoring sleeves are provided to minimize the possibility of the sleeve entering the vein. Do not remove the tabs (see Figure 8 ). If using a large diameter percutaneous lead introducer (PLI) sheath, extreme care should be taken to prevent passage of the anchoring sleeve into the PLI lumen and/or venous system.

Figure 8. Slotted triple groove anchoring sleeve with tabs

1 Anchoring sleeve tab
With a slotted triple groove anchoring sleeve, generally two or three of the grooves may be used with the following procedure (see Figure 9 and Figure 10).
The slotted triple groove anchoring sleeve is situated at the connector end of the lead. Partially insert the anchoring sleeve into the vein.
Use the most distal suture groove to secure the anchoring sleeve to the vein.
Use the middle groove to secure the anchoring sleeve to the fascia and lead. First, create a base by looping a suture through the fascia underneath the middle groove and tying a knot. Continue by firmly wrapping the suture around the middle groove and tying a second knot.
9
Page 10

Figure 9. Slotted triple groove anchoring sleeve secured to the lead and fascia using three grooves.

Use the third and most proximal groove to secure the anchoring sleeve to the lead body.
Alternatively, the physician may decide to use only two of the three grooves on the anchoring sleeve to tie down the lead. In that case, follow the anchoring procedure for the distal and middle groove (see Figure 10).
Caution: Do not use the anchoring sleeve tabs for suturing.
Figure 10. Slotted triple groove anchoring sleeve secured to the
lead and fascia using two grooves.
Tie the sutures securely but gently to prevent damage to the slotted triple groove anchoring sleeve.
Caution: Do not secure the ligatures so tightly that they damage the vein or lead. Do not tie a ligature directly to the lead body (see Figure 11). During anchoring, take care to avoid dislodging the lead tip.

Figure 11. Do not secure the sutures too tightly and do not tie a suture to the lead body.

named IS-1 since the atrial electrode ring pair is only designed for sensing and not for pacing, as required by the IS-1 standard to allow IS-1 identification.
Caution: The Medtronic CapSure VDD-2 Model 5038 is intended for use with current Medtronic VDD pulse generators.
Caution: Always remove the stylet before connecting the lead to the pulse generator. Failure to remove the stylet may result in lead failure.
Caution: To prevent undesirable twisting of the lead body, wrap the excess lead length loosely under the pulse generator and place both into the subcutaneous pocket (see Figure 12).

Figure 12. While rotating the pulse generator, loosely wrap the excess lead length and place it under the pulse generator

Caution: When placing the pulse generator and leads into the subcutaneous pocket:
Do not coil the lead. Coiling the lead can twist the lead body and may result in lead dislodgement (see Figure 13).
Do not grip the lead or pulse generator with surgical instruments.
After implantation, monitor the patient’s electrocardiogram continuously. If a lead dislodges, it usually occurs during the immediate postoperative period.

Figure 13. Do not coil the lead body

8.8 Connecting the lead to the pulse generator

Connect the ventricular IS-1 Bipolar (BI) lead connector, identified by the IS-1 BI VEN label, and the atrial bipolar connector, identified by the ATRIAL SENSE label, to the pulse generator, according to the instructions in the pulse generator manual.
Note: The bipolar atrial connector meets the mechanical and functional requirements of the IS-1 standard, but cannot be
10

9 Detailed device description

Table 2. Specifications (nominal)

Parameter Model 5038
Type Quadripolar
Chamber Dual (atrial sense and ventricu-
lar pace and sense)
Page 11
Table 2. Specifications (nominal) (continued)
Parameter Model 5038
Fixation 4 tines, each 3.0 mm in length
Length 58 cm or 65 cm (5038)
58 cm or 52 cm (5038S)
65 cm (5038L)
Connectors Ventricular: IS-1 BI
Atrial: Bipolar
Material Conductor: MP35N
Insulator: Treated silicone rubber
Ring electrode: Platinum alloy
Tip electrode: Platinum alloy
Tines: Silicone rubber
Tip electrode configuration Hemispherical, platinized,
Diameters Lead body: 2.65 mm
Atrial ring section
body:
Ventricular tip elec-
trode:
Ventricular ring
electrode:
Atrial ring electro-
des:
Lead introducer (recommended size)
without guide wire: 3.0 mm (9 French)
with guide wire: 4.0 mm (12 French)
Electrode sur-
Ventricular tip: 5.8 mm
face area
Ventricular ring: 36.0 mm
Atrial rings: 12.5 mm2 (each)
Resistance Unipolar ventricu-
lar:
Bipolar ventricular: 98 Ω (58 cm)
Bipolar atrial : 743 Ω (58 cm)
Distance between elec-
Ventricular tip and
ventricular ring:
trodes
Ventricular tip and
center of atrial ring
pair:
Atrial rings: 8.6 mm
Steroid Dexamethasone sodium phos-
Amount of steroid <1.0 mg
Steroid binder Silicone rubber
a
The Atrial sense connector meets the mechanical and functional requirements of the IS-1 standard (ISO 5841-3), but is only designed for sensing functions.
b
See Chapter 1 for a listing of available lead models, including lead length and V-A spacing.
a
porous, steroid eluting
3.00 mm
1.70 mm
2.85 mm
3.00 mm
2
2
45 Ω (58 cm)
28 mm
b
135 mm
phate
Figure 14.
1 Tip electrode: 1.7 mm 2 Atrial ring pair to ventricular tip distance: 115 – 155 mm 3 Ventricular tip to ring spacing: 28 mm 4 Proximal ventricular ring electrode area: 36 mm 5 Distal ventricular tip electrode area: 5.8 mm 6 Atrial ring spacing: 8.6 mm 7 Proximal atrial ring electrode surface area: 12.5 mm 8 Distal atrial ring electrode surface area: 12.5 mm
2
2
2
2
11
Page 12
9 Lead body: 2.65 mm 10 Lead length: 52 – 65 cm 11 Slotted 12 6.1 mm

10 Service

Medtronic employs highly trained representatives and engineers located throughout the world to serve you and, upon request, to provide training to qualified hospital personnel in the use of Medtronic products. Medtronic also maintains a professional staff to provide technical consultation to product users. For more information, contact your local Medtronic representative, or call or write Medtronic at the appropriate telephone number or address listed on the back cover.

11 Medtronic warranty

For complete warranty information, see the accompanying warranty document.
12
Page 13
Page 14
Medtronic, Inc.
*M955236A001*
710 Medtronic Parkway Minneapolis, MN 55432 USA www.medtronic.com +1 763 514 4000
Medtronic USA, Inc.
Toll-free in the USA (24-hour technical consultation for physicians and medical professionals) Bradycardia: +1 800 505 4636 Tachycardia: +1 800 723 4636
Europe/Middle East/Africa
Medtronic International Trading Sàrl Route du Molliau 31 Case Postale 84 CH-1131 Tolochenaz Switzerland +41 21 802 7000
Technical manuals
www.medtronic.com/manuals
© 2014 Medtronic, Inc. M955236A001 A 2014-06-09
Loading...