Medtronic 49341 Instructions for Use

Cardioblate™ LP 49341
Surgical Ablation Device
Instructions for Use
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
The following list includes trademarks or registered trademarks of Medtronic in the United States and possibly in other countries. All other trademarks are the property of their respective owners.
Cardioblate, Covidien, LigaSure, Medtronic, TissueFect, Valleylab
Explanation of symbols on package labeling
Nonpyrogenic
Sterilized using ethylene oxide
Do not reuse
Do not resterilize
Do not use if package is damaged
Use-by date
Lot number
Quantity
Open here
Consult instructions for use at this website
For US audiences only
Upper limit of temperature
Humidity limitation
Date of manufacture
Catalog number
Manufacturer
Type CF applied part
Contains di(2-ethylhexyl)phthalate (DEHP)
Authorized representative in the European Community
3

1 Model

49341

2 Product description

The Cardioblate LP surgical ablation device is a hand-held, single-use, bipolar, radiofrequency (RF) ablation device for use in cardiac surgery. It has a saline irrigation system to deliver fluid at the contact point between the tissue and electrode to cool the tissue during RF energy delivery. The device is intended for intermittent operation.
The Cardioblate LP surgical ablation device (Model 49341) is intended for use with the Covidien™ Valleylab™ FT10 FT series energy platform with TissueFect™ technology (Model VLFT10GEN) with software version 3.0.0 or higher.
Sterile, nonpyrogenic, disposable, single use only. Sterilized using ethylene oxide.
Rated accessory voltage: 382 V
Figure 1.
p
1 Rigid, slim line jaws, 7 cm (2.76 in) in length 2 7 cm (2.76 in) electrodes embedded in a porous polymer 3 300° rotation of jaw assembly 4 Malleable neck 5 Trigger 6 Trigger lock with raised thumb slide 7 Molded handle 8 Saline fluid delivery tubing 304.8 cm (120 in) in length with a female luer for connection to standard IV tubing systems 9 Electrical cable 304.8 cm (120 in) in length that terminates into a multi-pin connector

3 Indications for use

The Cardioblate LP surgical ablation device is intended to ablate cardiac tissue during cardiac surgery using radiofrequency energy.