Medtronic 49260 Instructions for Use

Page 1
CARDIOBLATE
®
GEMINI
49260, 49261
®
Instructions for Use
Caution: Federal law (USA) restricts
this device to sale by or on the order of a physician.
Page 2
Medtronic®, Cardioblate®, Gemini®, and Navigator® are registered trademarks of Medtronic, Inc.
Page 3
Explanation of symbols on package labeling
Refer to the device labeling to see which symbols apply to this product
Nonpyrogenic
Sterilized Using Ethylene Oxide
Do Not Reuse
Do Not Resterilize
Do not use if package is damaged
Use By
Lot Number
Quantity
Consult Instructions for Use
For US Audiences Only
Upper Limit of Temperature
Humidity Limitation
Date of Manufacture
Catalog Number
Manufacturer
Type CF Applied Part
Contains di(2-ethylhexyl)phthalate (DEHP)
Authorized Representative in the European Community
1
Page 4
Gemini®
5
6
3
7
2
1
4
8
9
4926149260
Surgical Ablation Device, Models 49260 and 49261
Figure 1
Rigid, slim-line jaws with guide attachments
1.
2. 6.3 cm (2.49 in) electrodes with embedded weeping polymer
3. RF Trigger
4. Flexible neck
5. Handle
6. Handle lock release
7. Molded hand piece
8. Saline fluid delivery tubing 304.8 cm (10 ft) in length with a female luer for connection to standard IV tubing systems
9. Electrical cable 304.8 cm (10 ft) in length that terminates into a multi-pin connector
2
Page 5
CARDIOBLATE® GEMINI
®
Surgical Ablation Device
Models
49260, 49261

Description

The Medtronic Cardioblate® Gemini® Surgical Ablation Device is a hand-held, single-use, bipolar, radiofrequency ablation device intended to ablate cardiac tissue during cardiac surgery. It has a saline irrigation system to deliver fluid at the contact point between the tissue and electrode to cool the tissue during radiofrequency energy delivery. Two
jaw curvatures are provided: a standard curve (Model 49260) and
unique extra curve (Model 49261). The device is intended for intermittent operation.
Sterile, nonpyrogenic, disposable, single use only.

Indications for Use

The Cardioblate® Gemini® Surgical Ablation Device is intended to ablate cardiac tissue during cardiac surgery using radiofrequency energy. The system is indicated for use, under direct or endoscopic visualization, in surgical procedures, including minimally invasive surgical procedures.

Contraindications

The Cardioblate® Gemini® Surgical Ablation Device should not be used for:
patients that have active endocarditis at time of surgery
ablation in a pool of blood (eg, through a purse string suture on a beating heart) (Effects of this type of ablation are unknown.)

Adverse Effects

Possible adverse effects related to the ablation of cardiac tissue in combination with open heart surgery are:
tissue perforation
extension of extracorporeal bypass
perioperative heart rhythm disturbances (atrial and/or ventricular)
postoperative embolic complications
pericardial effusion or tamponade
injury to the great vessels
valve leaflet damage
conduction disturbances (SA/AV node)
acute ischemic myocardial event
thrombus formation
nerve damage
unintentional burns
pericarditis
pleural effusion
esophageal perforation
death
coronary sinus perforation
coronary artery spasm
atrial lead dislodgement
hypotension
pulmonary vein stenosis
cerebrovascular accident
transient ischemic attack
blood loss
Instructions for Use English 3
Page 6

Warnings

Do not activate device if saline is not flowing freely. Cooling of the outer tissue prevents the charring and tissue desiccation that may occur when applying radiofrequency energy without irrigation.
device may be damaged before or during the procedure by improper
The handling or other intervening acts. If damaged, the device may fail to function properly, and may result in the following medical complications including, but not limited to:
damage to the physiological conduction system
peripheral tissue ablation due to breach of insulation
covering Device shall be removed from the patient before defibrillation. No representation or warranty is made that failure or cessation of function
of the device will not result in an adverse event, or that medical complications (including perforation of cardiac tissue) will not follow the procedure, or that the use of the device will in all cases restore adequate cardiac function.

Precautions

Caution: Federal law (USA) restricts this device to sale by or on the order
of a physician. Proper surgical procedures and techniques are necessarily the
responsibility of the medical professional. Each surgeon must evaluate the appropriateness of any procedure based on his or her own medical training and experience, and the type of surgical procedure.
Patient and procedure selection is solely a medical responsibility and the outcome is dependent on many variables including patient pathology, surgical and perfusion procedures.
Use caution to visualize the target area to avoid trauma to tissues not within the target area of ablation.
Purge air from system to ensure saline fluid flows from the electrodes during use.
Separation of electrodes from tissue contact during activation will result in a high impedance shutdown placing generator power condition to OFF. Placing electrodes in direct contact during radiofrequency energy activation results in a low impedance shutdown and places the generator power condition to OFF. In both situations, three short audible alarm beeps will be heard. Refer to the Cardioblate 68000 generator technical manual for reset instructions.
Maximum storage temperature: 40°C (104°F) Storage relative humidity: 5% to 85%, noncondensing This device was designed for single patient use only. Do not reuse,
reprocess, or resterilize this product. Reuse, reprocessing, or resterilization may compromise the structural integrity of the device and/or create a risk of contamination of the device, which could result in patient injury, illness, or death. Multiple uses can result in the occlusion of the irrigation openings and affect the device’s performance. To prevent this from occurring, the generator has the capability to detect previously used ablation devices and will not allow their reuse.
It is the responsibility of the user to dispose of the devices in accordance with local regulations and hospital procedures.
Due to the presence of phthalates in the product, the clinician must weigh the medical benefits of product use against the drawbacks of phthalate exposure for male children and pregnant or nursing women.

Connections to External Devices

1. Connection to the Cardioblate 68000 Generator is as follows:
a. Position the multi-pin connector of the bipolar device so the
alignment indicator matches the generator port indicator.
4 Instructions for Use English
Page 7
b. Insert
1
2
the multi-pin connector of the bipolar device into the device port on the generator until it locks in place. Do not force the connection.
Note: If additional devices are to be used, refer to their specific Instructions for Use.
2. Sterile normal saline, commercially available pressure cuff, and IV tubing set are required for use with the bipolar device. These are not included with the Cardioblate Gemini Surgical Ablation Device.

Instructions for Use

1. Inspect the package and device to ensure the expiration date has not passed and damage has not occurred to the device during shipping and handling. Do not use the device if the packaging or device is damaged or if the expiration date has elapsed. Open the package and transfer the contents onto the sterile field utilizing aseptic technique.
2. Uncoil both the saline delivery tubing and the electrical cable carefully to avoid kinking and tangling.
3. Attach the saline delivery tubing female luer connector to an IV tubing set and 1000 mL bag of sterile normal saline. (0.9% normal saline is recommended because it provides a low impedance pathway for the radiofrequency energy to the tissue.)
4. Insert a 1000 mL bag of 0.9% normal saline into commercially available pressure cuff and inflate to between 150 and 300 mm Hg.
5. Open the IV line and squeeze the handle to purge the system of air until fluid is seen coming from the electrodes on the Gemini device. Release the handle to stop IV flow.
1. 6.3 cm (2.49 in) Active Electrodes
Guide Attachments
2.
Figure 2. Jaw/Electrode
6. Attach the Gemini electrical cable to the Cardioblate 68000 Generator per instructions outlined in “Connections to External Devices.”
7. Optional: Navigator Guide Connection (Requires Cardioblate Gemini Model 49261 and Cardioblate Navigator Model 68015)
Insert the surgical instrument guide (supplied with the
a.
Cardioblate Navigator) into the flexible guide attachment so the tip is visible, then twist the guide lock clockwise to secure the surgical instrument guide into place (Figure 3).
Figure 3. Insert Surgical Instrument Guide into Flexible Guide
b.
To remove the flexible guide attachment from the guide, firmly grab the flexible portion of the guide attachment and pull away from the guide (Figure 4).
Instructions for Use English 5
Page 8
Figure 4.
8.
Connect the guides to the device (Figure 5) to assist in positioning the electrodes on the targeted tissue.
Figure 5. Connect Jaw/Guide
Prior to tissue ablation:
9. a. Squeeze the handle and make certain that saline is flowing freely
out of the electrodes. Continuous flow of saline is necessary during ablation to cool the outer tissue. Cooling of the outer tissue prevents charring and tissue desiccation and allows radiofrequency energy to be delivered deeper into the tissue to more effectively create transmural lesions.
Note: While irrigation increases the effectiveness of radiofrequency ablation, ablating in a pool of fluid can have a detrimental effect. Radiofrequency energy can shunt between the 2 electrodes and the transmurality algorithm may be less effective. Suctioning of excessive fluid at the ablation site is recommended to prevent this from occurring.
b. Confirm the Cardioblate 68000 Generator is set to the bipolar
mode.
c. The mode of operation will default to the TRANSMURALITY
control mode.
d. In the TRANSMURALITY mode, all generator parameters will be
automatically adjusted.
10. Tissue Ablation: a. Once the electrodes are correctly positioned, squeeze the handle
until the parallel jaws lock and the electrodes are in full contact with the tissue. The locking mechanism built into the handle must be fully engaged to ensure accurate transmurality readings.
b. The RF energy is activated in the device by either pressing the
RF energy button on the Cardioblate 68000 Generator display
6 Instructions for Use English
Page 9
panel, squeezing the RF Trigger on the device, or using the foot
0
2
0
3
04
0
5
06
07
08
0
9
0
01
0
2000
4000
6000
8000
10000
12000
14000
16000
18000
20000
0
1
5
1
02
5
2
03
53
04
54
05
1
3
2
3
1
switch provided with the generator. When using the RF Trigger, the current flows between the electrodes only when the handle is in the locked position.
Note: It is not necessary to disconnect the guides before tissue ablation.
Table 1. Anticipated Energy Delivery at Various Tissue Thicknesses
Tissue Thickness (6 cm length) Energy Delivered (W)
c. The lesion becomes visible by a white coloration on the tissue.
In TRANSMURALITY control mode, the Cardioblate 68000 Generator will emit a solid tone once a transmural lesion is created. This is the indication to the surgeon to stop ablating by releasing RF energy button on the generator.
Note: Ablation times will vary due to tissue thickness but will not exceed 45 seconds per ablation cycle.
Note: Ablating beyond recommended ablation times may result in tissue perforation.
Note: To prevent the porous jaws from drying out and clogging, maintain the jaws immersed in saline or covered with a wet sponge between ablations.
2 mm 833 ±32 6 mm 889 ±135
10 mm 1023 ±163
the foot switch, releasing the RF trigger, or pressing the
1. Average Impedance (Ohms)
Time (msec)
2.
3. Power (Watts)
Corresponding Power Levels.
11. After tissue ablation: a. Press
b. Disconnect the guides from the device (Figure 7).
12.
Upon completion of the surgical procedure, disconnect the device from the generator and discard.
Clinical Data on Conduction Block
In vitro testing was conducted creating lesion sets on porcine atria. In vivo testing was performed on porcine tissue using the Gemini
Figure 6. Typical Tissue
Impedance Values with
(Approximate total energy
delivered 570 W)
the handle lock release on the handle to unlock the parallel
jaws and allow the device to be retracted from the target tissue.
Figure 7. Disconnect Jaw/Guide
Instructions for Use English 7
Page 10
device. Transmural lesions were demonstrated with a 95/95 confidence level. The tables below provide an example of lesion depth, width, and length using the Gemini device:
In vitro lesion evaluation study Cardiac Tissue
Max Transmurality Depth (mm)
Lesion Length (cm) 4.1 ±0.6 6.0* Lesion Width (mm) 4.5 ±0.4 2.9 ±1.1
*In vitro transmurality and length variables were controlled by tissue cut to 3 mm thickness and full 6 cm bites, therefore there are no standard deviations on soft tissue lesions.
In vivo lesion observations in a single por-
Cardiac
Tissue
Max Transmural­ity Depth (mm)
Lesion Length (cm)
Lesion Width (mm)
**Data was obtained in a nonstatistical sample size, in vivo porcine study with 3 lesion burns in each tissue. The Gemini device can achieve transmurality of up to 10 mm in cardiac tissue as demonstrated in vitro.
These results are consistent with a small, retrospective study conducted at 2 investigational centers of 69 patients during 2006-2007. Thirty-five of those patients received cardiac tissue
using the Medtronic Cardioblate system. Of the 35 patients,
ablation 29 (82.9%) underwent conduction block assessment confirming transmurality of cardiac tissue lesions. The remaining 6 patients were not assessed.
(n=3)
3.0 ±2.5** 6.5 ±1.9 7.4 ±2.9 8.5 ±2.4
3.8 ±0.6 3.8 ±1.0 5.3 ±0.4 4.8 ±1.0
4.3 ±1.2 5.7 ±2.0 5.4 ±1.0 3.6 ±1.3
(n=42)
5.6 ±1.2 3.0*
cine animal study
Pulmonary
Tissue
(n=3)
Hepatic
Tissue
(n=3)
Soft Tissue (n=44)
Renal
Tissue
(n=3)
8 Instructions for Use English
Page 11
The following disclaimer of warranty applies to United States customers only:

Disclaimer of Warranty

ALTHOUGH THE CARDIOBLATE® GEMINI® SURGICAL ABLATION DEVICE, HEREAFTER REFERRED TO AS “PRODUCT” HAS BEEN MANUFACTURED UNDER CAREFULLY CONTROLLED CONDITIONS, MEDTRONIC HAS NO CONTROL OVER THE CONDITIONS UNDER WHICH THIS PRODUCT IS USED. MEDTRONIC, THEREFORE DISCLAIMS ALL WARRANTIES, BOTH EXPRESS AND IMPLIED, WITH RESPECT TO THE PRODUCT, INCLUDING, BUT NOT LIMITED TO, ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. MEDTRONIC FOR ANY MEDICAL EXPENSES OR ANY DIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES CAUSED BY ANY USE, DEFECT, FAILURE OR MALFUNCTION OF THE PRODUCT, WHETHER A CLAIM FOR SUCH DAMAGES IS BASED UPON WARRANTY, CONTRACT, TORT OR OTHERWISE. NO PERSON HAS ANY AUTHORITY TO BIND MEDTRONIC TO ANY REPRESENTATION OR WARRANTY WITH RESPECT TO THE PRODUCT.
The exclusions and limitations set out above are not intended to, and should not be construed so as to, contravene mandatory provisions of applicable law. If any part or term of this Disclaimer of Warranty is held to be illegal, unenforceable or in conflict with applicable law, by a court of competent jurisdiction, the validity of the remaining portions of this Disclaimer shall be construed and enforced as if this Disclaimer of Warranty did not contain the particular part or term held to be invalid.
SHALL NOT BE LIABLE TO ANY PERSON OR ENTITY
of Warranty shall not be affected, and all rights and obligations
Instructions for Use English 9
Page 12
Europe
*M940542A001*
Europe/Africa/Middle East Headquarters
Medtronic International Trading Sàrl Route du Molliau 31 Case Postale 84 CH - 1131 Tolochenaz Switzerland Internet: www.medtronic.co.uk Tel. 41-21-802-7000 Fax 41-21-802-7900
Authorized Representative in the European Community
Medtronic B.V. Earl Bakkenstraat 10 6422 PJ Heerlen The Netherlands Tel. 31-45-566-8000 Fax 31-45-566-8668
Asia-Pacific
Japan
Medtronic Japan Comodio Shiodome 5F 2-14-1 Higashi-Shimbashi, Minato-ku Tokyo 105-0021 Japan Tel. 81-3-6430-2011 Fax 81-3-6430-7140
Australia
Medtronic Australasia Pty. Ltd. 97 Waterloo Road North Ryde NSW 2113 Australia Tel. 61-2-9857-9000 Fax 61-2-9878-5100
Asia
Medtronic International Ltd.
1602 16/F, Manulife Plaza
Suite The Lee Gardens, 33 Hysan Avenue Causeway Bay Hong Kong Tel. 852-2891-4068 Fax 852-2591-0313
Americas
Latin America
Medtronic Latin America 3750 NW 87th Avenue Suite 700 Miami, FL 33178 USA Tel. 305-500-9328 Fax 786-709-4244
Canada
Medtronic of Canada Ltd. 99 Hereford Street Brampton, Ontario L6Y 0R3 Canada Tel. 905-460-3800 Fax 905-826-6620 Toll-free: 1-800-268-5346
United States
Manufacturer: Medtronic, Inc. 710 Medtronic Parkway Minneapolis, MN 55432 USA Internet: www.medtronic.com Tel. 763-514-4000 Fax 763-391-9100 Toll-free: 1-800-328-2518 (24-hour consultation service)
© 2007, 2011 Medtronic M940542A001 Rev. 1B
Loading...