Medtronic 459888 Technical Manual

ATTAIN PERFORMA™ S MRI SURESCAN™ 4598
Steroid-eluting, quadripolar electrode, transvenous, over-the-wire, cardiac vein pacing lead
Technical Manual
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
The following list includes trademarks or registered trademarks of Medtronic in the United States and possibly in other countries. All other trademarks are the property of their respective owners.
AccuRead, Attain, Attain Hybrid, Medtronic, Performa, SureScan

Contents

9 Implant procedure 8 10 Specifications 15 11 Medtronic warranty 16 12 Service 16

1 Description

The Medtronic Attain Performa S MRI SureScan 4598 steroid-eluting, quadripolar electrode, transvenous, over-the-wire LV lead is designed for pacing via a cardiac vein. The lead has been tested for use in the Magnetic Resonance Imaging (MRI) environment. All lead lengths for this lead model are MR Conditional. This lead contains 4 electrodes designed to function as cathodes or anodes, depending on how the device LV pacing polarity is programmed:
electrode LV1, the distal electrode, positioned near the distal tip of the lead
electrode LV2, positioned 21 mm proximal to electrode LV1
electrode LV3, positioned 1.3 mm proximal to electrode LV2
electrode LV4, the proximal electrode, positioned 21 mm proximal to electrode LV3
See Section 9.10, “Taking electrical measurements”, page 11 for information about LV pacing polarity selections.
The Medtronic IS4-LLLL1 four-pole inline connector on the lead facilitates device connection during implant. The tip of the connector pin has a white band indicator that can be used for visual confirmation of proper lead connection to the device.
The connector contacts align with the lead electrodes LV1 to LV1, LV2 to LV2, and so on:
connector contact LV1, the connector pin, positioned at the proximal tip of the lead (aligns to electrode LV1)
connector contact LV2, positioned distal to connector pin LV1 (aligns to electrode LV2)
connector contact LV3, positioned distal to connector contact LV2 (aligns to electrode LV3)
connector contact LV4, positioned distal to connector contact LV3 (aligns to electrode LV4)
See Section 10.2, “Specifications drawing (nominal)”, page 16 for an illustration of the lead electrodes and connector contacts.
The distal tip of the lead allows a guide wire to pass through to aid in cardiac vein selection. The tip contains a silicone rubber membrane, which seals the lead inner lumen to reduce blood ingress.
Each electrode contains a Monolithic controlled release device (MCRD) for elution of steroid to reduce inflammatory response within the cardiac vein. The MCRDs contain a combined-total target dosage of 288 µg of dexamethasone acetate steroid. The target dose of the steroid is 72 µg at each MCRD. Upon exposure to body fluids, the steroid elutes from the MCRDs. The steroid suppresses the inflammatory response that is believed to cause threshold rise typically associated with implanted pacing electrodes.
The electrodes are positioned relative to the 3 distal curves of the lead to facilitate contact with the cardiac veins.
The outer insulation of the lead is polyurethane and the inner insulation is SI-polyimide (SI-PI)2. The SI-PI is applied as a coating to the conductor wire before coiling.
The Attain Performa S MRI SureScan 4598 LV lead can be positioned with the aid of a guide wire, a stylet, an inner catheter, or an inner catheter plus a guide wire or a stylet.
To implant the lead in a selected cardiac vein, a compatible delivery system is required. A compatible delivery system includes a guide catheter and either a hemostasis valve or an introducer valve that can be removed or that allows passage over the lead connector. Contact a Medtronic representative for further information regarding compatible delivery systems.

1.1 Medtronic SureScan system

The Model 4598 lead is part of the Medtronic SureScan system. The SureScan system includes a Medtronic SureScan device connected to Medtronic SureScan leads. Labeling for SureScan system components displays the SureScan logo and the MR Conditional symbol. To verify that components are part of a SureScan system, visit http://www.mrisurescan.com.
SureScan logo
MR Conditional symbol. The Medtronic SureScan sys­tem is MR Conditional and is designed to allow implan­ted patients to undergo an MRI scan under the speci­fied MRI conditions for use.
1
IS4-LLLL refers to an International Connector Standard (ISO 27186:2010) whereby pulse generators and leads so designated are assured of a basic mechanical fit. LLLL defines the lead connector contacts as low voltage (L).
2
Technology developed by NASA.
3
The MRI SureScan feature permits a mode of operation that allows a patient with a SureScan device to be safely scanned by an MRI machine while the device continues to provide appropriate pacing. When programmed to On, MRI SureScan operation disables arrhythmia detection, magnet mode, and all user-defined diagnostics. Before performing an MRI scan,
refer to the SureScan MRI technical manual for important information about procedures and MRI-specific warnings and precautions.

1.2 Contents of package

Leads and accessories are supplied sterile. Each package contains the following items:
1 lead with anchoring sleeve
1 guide wire insertion tool
1 guide wire clip
1 guide wire steering handle
4 stylets
2 AccuRead 2.0 analyzer cable interface tools
product documentation

1.3 Accessory descriptions

Dispose of all single-use accessories according to local environmental requirements.
AccuRead 2.0 tool – The AccuRead 2.0 analyzer cable interface tool guides the connection of analyzer cable terminals to the lead connector contacts. This connection facilitates accurate electrical measurements during implant and prevents possible connector damage.
Anchoring sleeve – An anchoring sleeve secures the lead to prevent it from moving and protects the lead insulation and conductors from damage caused by tight sutures.
Guide wire clip – A guide wire clip secures the excess guide wire and helps to protect and maintain the sterility of the guide wire.
Guide wire insertion tool – A guide wire insertion tool provides additional control when inserting a guide wire into the lead connector pin or the lead tip.
Guide wire steering handle – A guide wire steering handle is used only with guide wires 0.46 mm (0.018 in) or less in diameter. The steering handle provides additional guide wire steering and rotation control.
Stylet – A stylet provides additional stiffness and controlled flexibility for maneuvering the lead into position. Each stylet knob is labeled with the stylet diameter and corresponding lead length.

2 Drug component description

The active ingredient in the Attain Performa S MRI SureScan 4598 LV lead is dexamethasone acetate [21-(acetyloxy)-9-fluoro-11β, 17-trihydroxy-16α-methylpregna-1,4-diene-3,20-dione]. The structural formula for this steroid is as follows:
Dexamethasone acetate (DXAC) - C24H31FO
6
Figure 1.
The target dosage of dexamethasone acetate is 288 µg per lead.
Cautions:
Drug interactions of dexamethasone acetate with this lead have not been studied.
Before implanting this lead, consider total patient exposure to dexamethasone acetate.

3 Indications

The Attain Performa S MRI SureScan 4598 steroid-eluting, quadripolar electrode, IS4 transvenous lead is indicated for chronic pacing in the left ventricle via the cardiac vein, when used with a compatible Medtronic Cardiac Resynchronization Therapy (CRT) system. Extended bipolar pacing is available using this lead in combination with a compatible CRT-D system and RV defibrillation lead.

4 Contraindications

Before performing an MRI scan, refer to the SureScan MRI technical manual for MRI-specific contraindications.
The Model 4598 lead is contraindicated as follows:
Coronary vasculature – This lead is contraindicated for patients with coronary venous vasculature that is inadequate for lead placement, as indicated by venogram.
Steroid use – The lead is contraindicated in patients for whom a single dose of 288 µg of dexamethasone acetate may be contraindicated.

5 Warnings and precautions

A complete SureScan system is required for use in the MRI environment. Before performing an MRI scan, refer to the SureScan MRI technical manual for MRI-specific warnings and precautions.
Note: Medical procedure warnings and precautions that pertain
to the Medtronic implanted system are provided in the manual that is packaged with the device or on the Medtronic Manual Library website (www.medtronic.com/manuals).
4
Inspecting the sterile package – Inspect the sterile package with care before opening it.
If the seal or package is damaged, contact a Medtronic representative.
Store at 25 °C (77 °F). Excursions from this storage temperature are permitted in the range of 15 to 30 °C (59 to 86 °F). (See USP Controlled Room Temperature.) According to USP excursion conditions, transient spikes up to 40 °C (104 °F) are permitted as long as they do not exceed 24 hours.
Do not use the product after its expiration date.
Sterilization – Medtronic has sterilized the package contents with ethylene oxide before shipment. This lead is for single use only and is not intended to be resterilized.
Single use – The lead and accessories are for single use only.
Necessary hospital equipment – Keep external defibrillation
equipment nearby for immediate use during acute lead system testing, the implant procedure, or whenever arrhythmias are possible or intentionally induced during post-implant testing.
External defibrillation and cardioversion – External defibrillation and cardioversion are therapies that deliver an electrical shock to the heart to convert an abnormal heart rhythm to a normal rhythm.
Medtronic cardiac devices are designed to withstand exposure to external defibrillation and cardioversion. While damage to an implanted system from an external shock is rare, the probability increases with increased energy levels. These procedures may also temporarily or permanently elevate pacing thresholds or temporarily or permanently damage the myocardium. If external defibrillation or cardioversion are required, consider the following precautions:
Use the lowest clinically appropriate energy.
Position the patches or paddles a minimum of 15 cm (6 in) away from the device.
Position the patches or paddles perpendicular to the device and lead system.
If an external defibrillation or cardioversion is delivered within 15 cm (6 in) of the device, use a Medtronic programmer to evaluate the device and lead system.
Line-powered and battery-powered equipment – An implanted lead forms a direct current path to the myocardium. During lead implant and testing, use only battery-powered equipment or line-powered equipment specifically designed for this purpose to protect against fibrillation that may be caused by alternating currents. Line-powered equipment used in the vicinity of the patient must be properly grounded. Lead connector pins must be insulated from any leakage currents that may arise from line-powered equipment.
Concurrent devices – Output pulses, especially from unipolar devices, may adversely affect device sensing capabilities. If a patient requires a separate stimulation device, either permanent or temporary, allow enough space between the leads of the separate systems to avoid interference in the sensing capabilities of the devices. Previously implanted pulse generators and
implantable cardioverter defibrillators should generally be explanted.
Steroid use – It has not been determined whether the warnings, precautions, or complications usually associated with injectable dexamethasone acetate apply to the use of this highly localized, controlled-release device. For a list of potential adverse effects, refer to the Physicians’ Desk Reference.
Handling steroid monolithic controlled release devices (MCRDs) – Avoid reducing the amount of steroid available before
implanting the lead. Reducing the available amount of steroid may adversely affect low-threshold performance.
Do not allow the electrode surfaces to come in contact with surface contaminants.
Do not wipe or immerse the electrodes in fluid, except blood, at the time of implant.
Bipolar pacing – If the surface area of the selected anode electrode is equal to or less than the surface area of the selected cathode, higher pacing thresholds or anodal stimulation may result. This lead uses 4 equal-sized electrodes; therefore, bipolar pacing configurations may result in higher pacing thresholds or anodal stimulation.
If therapy cannot be delivered via the LV1 electrode, the LV2, LV3, and LV4 electrodes are available for use with specific Medtronic devices. Refer to the appropriate Medtronic Cardiac Resynchronization Therapy (CRT-D) system manual for use of available LV lead pacing polarity options.
Handling the stylet – Handle the stylet with care at all times.
To minimize the likelihood of trauma to the vein and to maintain lead flexibility while advancing the lead through the vein, keep the stylet withdrawn 1 to 2 cm or select a more flexible stylet.
Do not use excessive force or surgical instruments when inserting a stylet.
Avoid overbending, kinking, or blood contact on stylets.
Use a new stylet when blood or other fluids accumulate on the stylet. Accumulated fluids may cause damage to the lead or difficulty in passing the stylet through the lead.
Curving the distal end of the stylet prior to insertion into the lead will achieve a curvature at the distal end of the lead. Do not use a sharp object to impart a curve to the distal end of the stylet.
Handling the guide wire – Handle the guide wire with care at all times.
Do not insert the proximal end of the guide wire through the lead tip seal without using the guide wire insertion tool. Inserting the guide wire without the guide wire insertion tool may damage the lead.
Damage to a guide wire may prevent the guide wire from performing with accurate torque response and control and may cause vessel damage. For additional information about vessel damage and other potential adverse events, refer to the technical manual packaged with the appropriate guide wire.
5
If the distal end of the guide wire becomes severely kinked or twisted, it may be difficult to withdraw it back through the lead. Therefore, if there is an indication that the distal end of the guide wire has become damaged, or if there is significant resistance in guide wire passage, remove the lead and guide wire together as a unit. Remove the guide wire from the lead and insert a new guide wire into the lead. Do not use excessive force to retract the guide wire from the lead. Refer to the product documentation packaged with the guide wire for additional information.
Handling the lead – Handle the lead with care at all times.
Rust stylets are not recommended with this lead due to the risk of conductor coil or insulation perforation.
If a stylet is used for lead positioning, use only the stylets packaged with the lead or in a stylet kit (downsized knob). Other stylets may extend beyond the lead tip causing patient injury.
If the lead is damaged, do not implant it. Return the lead to a Medtronic representative.
Protect the lead from materials that shed particles such as lint and dust. Lead insulators attract these particles.
Handle the lead with sterile surgical gloves that have been rinsed in sterile water or a comparable substance.
Do not severely bend, kink, or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pin.
Do not immerse leads in mineral oil, silicone oil, or any other liquid, except blood, at the time of implant.
Use an anchoring sleeve with all leads. Ensure that the anchoring sleeve is positioned close to the lead connector pin, to prevent inadvertent passage of the sleeve into the vein. If it is necessary to wipe the lead before insertion, ensure that the anchoring sleeve remains in position.
Do not force the guide catheter or leads if significant resistance is encountered. Use of guide catheters or leads may cause trauma to the heart.
Chronic repositioning or removal – Chronic repositioning or removal of leads may be difficult because of fibrotic tissue development. The clinical study was not designed to evaluate the removal of left ventricular leads from the coronary venous vasculature. If a lead must be removed or repositioned, proceed with extreme caution. Return all removed leads to Medtronic.
Rust stylets are not recommended with this lead due to the risk of conductor coil/insulation perforation.
Verify lead length on the lead label on the connector to choose an appropriate stylet kit (downsized knob) length when repositioning. Always choose a stylet kit (downsized knob) 3 cm (1.2 in) shorter than the lead length. For example, choose a stylet kit (downsized knob) with stylets 75 cm long for a lead 78 cm long.
Lead removal may result in avulsion of the endocardium, valve, or vein.
Lead junctions may separate, leaving the lead tip and bare wire in the heart or vein.
Chronic repositioning may adversely affect the low-threshold performance of a steroid-eluting lead.
Cap abandoned leads to avoid transmitting electrical signals.
For leads that have been severed, seal the remaining lead end and suture the lead to adjacent tissue.
If a lead is removed and repositioned, inspect it carefully for insulator or conductor coil damage before repositioning.
Magnetic resonance imaging (MRI) – An MRI is a type of medical imaging that uses magnetic fields to create an internal view of the body. If certain criteria are met and the warnings and precautions provided by Medtronic are followed, patients with an MR Conditional device and lead system are able to undergo an MRI scan; for details, refer to the SureScan MRI technical manual that Medtronic provides for an MR Conditional device.
Diathermy treatment (including therapeutic ultrasound) –
Diathermy is a treatment that involves the therapeutic heating of body tissues. Diathermy treatments include high frequency, short wave, microwave, and therapeutic ultrasound. Except for therapeutic ultrasound, do not use diathermy treatments on cardiac device patients. Diathermy treatments may result in serious injury or damage to an implanted device and lead system. Therapeutic ultrasound (including physiotherapy, high intensity therapeutic ultrasound, and high intensity focused ultrasound), is the use of ultrasound at higher energies than diagnostic ultrasound to bring heat or agitation into the body. Therapeutic ultrasound is acceptable if treatment is performed with a minimum separation distance of 15 cm (6 in) between the applicator and the implanted device and lead system, as long as the ultrasonic beam is pointing away from the device and lead system.

6 Potential adverse events

The potential adverse events (listed in alphabetical order) related to the use of transvenous leads include, but are not limited to, the following conditions:
Air embolism
Avulsion or other damage to the endocardium, valve, or vein (particularly in fragile hearts)
Cardiac dissection or perforation
Cardiac tamponade
Coronary sinus dissection
Death
Endocarditis or pericarditis
Erosion through the skin
Extracardiac muscle or nerve stimulation
Fibrillation or other arrhythmias
Heart block
Heart wall or vein wall rupture
Hematoma/seroma
Infection
Lead conductor fracture or insulation failure
Lead dislodgement
Myocardial irritability
Myopotential sensing
Pericardial effusion or rub
Pneumothorax
Rejection phenomena (local tissue reaction, fibrotic tissue formation)
6
Loading...
+ 12 hidden pages