Medtronic 457430 Technical Manual

CAPSURE SENSE® 4574
Steroid-eluting, bipolar, implantable, tined, atrial, transvenous lead
Technical Manual
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
The following list includes trademarks or registered trademarks of Medtronic in the United States and possibly in other countries. All other trademarks are the property of their respective owners.

Contents

1 Description 3 2 Drug component description 3 3 Indications 3 4 Contraindications 3 5 Warnings and precautions 4 6 Drug information 5 7 Potential adverse events 6 8 Clinical data 6
9 Directions for use 9 10 Specifications (nominal) 12 11 Medtronic warranty 14 12 Service 14

1 Description

The Medtronic CapSure Sense Model 4574 steroid-eluting, bipolar, implantable, tined, atrial, transvenous lead is designed for atrial pacing and sensing. The platinum alloy tip and ring electrodes feature a high-active surface area of titanium nitride microstructure. This electrode configuration contributes to low polarization.
The tip electrode of the lead incorporates a steroid-eluting plug containing dexamethasone acetate. The tip electrode contains a target nominal dosage of 272 µg of dexamethasone. Upon exposure to body fluids, the steroid elutes from the electrode.
The lead is designed to provide low chronic pacing thresholds via steroid treatment of cardiac tissue near the lead tip. Steroid suppresses the inflammatory response that is believed to cause threshold rises typically associated with implanted pacing electrodes.
The distal portion of the lead is J-shaped. This facilitates placement of the electrode in or near the apex of the right atrial appendage.
The lead features four polyurethane tines near the electrode tip, MP35N nickel alloy conductors, polyurethane outer insulation, silicone inner insulation, and an IS-1 Bipolar (BI)1 lead connector.

1.1 Package contents

Leads and accessories are supplied sterile. Each package contains the following items:
1 lead with anchoring sleeve, stylet, and stylet guide
1 vein lifter
extra stylets
product documentation

1.2 Accessory descriptions

Dispose of all single-use accessories according to local environmental requirements.
Anchoring sleeve – An anchoring sleeve secures the lead to prevent it from moving and protects the lead insulation and conductors from damage caused by tight sutures.
Stylet – A stylet provides additional stiffness and controlled flexibility for maneuvering the lead into position. Each stylet knob is labeled with the stylet diameter and length.
Stylet guide – A stylet guide facilitates stylet insertion into the lead.
Vein lifter – A vein lifter facilitates lead insertion into a vein.

2 Drug component description

The active ingredient in the Model 4574 lead is dexamethasone acetate. Dexamethasone acetate is 9-Fluoro-11β, 17,21-trihydroxy-16α-methylpregna-1,4-diene-3,20-dione 21-acetate. Dexamethasone acetate has a molecular formula of C24H31FO6 and a molecular weight of 434.50. The MCRD excipient is silicone. See Figure 1 for the structural formula.
Figure 1.
The target dosage of dexamethasone acetate is 272 µg per lead.

3 Indications

The Model 4574 implantable, atrial, transvenous lead has application where implantable atrial single-chamber or dual-chamber pacing systems are indicated. The lead is intended for pacing and sensing in the atrium.

4 Contraindications

Use of atrial tined transvenous leads may be contraindicated in the absence of a right atrial appendage.
Use of steroid-eluting transvenous leads is contraindicated in patients for whom a single dose of 272 µg dexamethasone acetate may be contraindicated.
1
IS-1 BI refers to an International Connector Standard (ISO 5841-3) whereby pulse generators and leads so designated are assured of a basic mechanical fit.
3

5 Warnings and precautions

Note: Medical procedure warnings and precautions that pertain
to the Medtronic implanted system are provided in the manual that is packaged with the device or on the Medtronic Manual Library website (www.Medtronic.com/manuals).
Line-powered and battery-powered equipment – An implanted lead forms a direct current path to the myocardium. During lead implant and testing, use only battery-powered equipment or line-powered equipment specifically designed for this purpose to protect against fibrillation that may be caused by alternating currents. Line-powered equipment used in the vicinity of the patient must be properly grounded. Lead connector pins must be insulated from any leakage currents that may arise from line-powered equipment.
Diathermy treatment (including therapeutic ultrasound) –
Diathermy is a treatment that involves the therapeutic heating of body tissues. Diathermy treatments include high frequency, short wave, microwave, and therapeutic ultrasound. Except for therapeutic ultrasound, do not use diathermy treatments on cardiac device patients. Diathermy treatments may result in serious injury or damage to an implanted device and leads. Therapeutic ultrasound is the use of ultrasound at higher energies than diagnostic ultrasound to bring heat or agitation into the body. Therapeutic ultrasound is acceptable if treatment is performed with a minimum separation distance of 15 cm (6 in) between the applicator and the implanted device and leads.
Vessel and tissue damage – Use care when positioning the lead. Avoid known infarcted or thin ventricular wall areas to minimize the occurrence of perforation and dissection.
Single use – The lead and accessories are for single use only.
Inspecting the sterile package – Inspect the sterile package
with care before opening it.
If the seal or package is damaged, contact a Medtronic representative.
Store at 25 °C (77 °F). Excursions from this storage temperature are permitted in the range of 15 to 30 °C (59 to 86 °F). (See USP Controlled Room Temperature.) According to USP excursion conditions, transient spikes up to 40 °C (104 °F) are permitted as long as they do not exceed 24 hours.
Do not use the product after its expiration date.
Sterilization – Medtronic has sterilized the package contents with ethylene oxide before shipment. This lead is for single use only and is not intended to be resterilized.
Steroid use – It has not been determined whether the warnings, precautions, or complications usually associated with injectable dexamethasone acetate apply to the use of this highly localized, controlled-release lead. For a list of potential adverse effects, refer to the Physicians’ Desk Reference.
Handling the steroid tip – Avoid reducing the amount of steroid available before implanting the lead. Reducing the available
amount of steroid may adversely affect low-threshold performance.
Do not allow the electrode surface to come in contact with surface contaminants.
Do not wipe or immerse the electrode in fluid, except blood, at the time of implant.
Handling a tined lead – Handle the lead with care at all times.
Do not implant the lead if it is damaged. Return the lead to a Medtronic representative.
Protect the lead from materials that shed small particles such as lint and dust. Lead insulators attract these particles.
Handle the lead with sterile surgical gloves that have been rinsed in sterile water or a comparable substance.
Do not severely bend, kink, or stretch the lead.
Do not immerse the lead in mineral oil, silicone oil, or any other liquid, except blood, at the time of implant.
Do not use surgical instruments to grasp the lead.
Do not force the lead if resistance is encountered during lead passage.
Handling the stylet – Handle the stylet with care at all times.
Curve the stylet before inserting it into the lead to achieve a curvature at the lead’s distal end. Do not use a sharp object to impart a curve to the distal end of the stylet.
Do not use excessive force or surgical instruments when inserting the stylet into the lead.
Avoid overbending or kinking the stylet.
Use a new stylet when blood or other fluids accumulate on the stylet. Accumulated blood or other fluids may damage the lead or cause difficulty in passing the stylet into the lead.
Necessary hospital equipment – Keep external defibrillation equipment nearby for immediate use during acute lead system testing, the implant procedure, or whenever arrhythmias are possible or intentionally induced during post-implant testing.
Magnetic resonance imaging (MRI) – An MRI is a type of medical imaging that uses magnetic fields to create an internal view of the body. Do not conduct MRI scans on patients who have this device or lead implanted. MRI scans may result in serious injury, induction of tachyarrhythmias, or implanted system malfunction or damage.
Concurrent devices – Output pulses, especially from unipolar devices, may adversely affect device sensing capabilities. If a patient requires a separate stimulation device, either permanent or temporary, allow enough space between the leads of the separate systems to avoid interference in the sensing capabilities of the devices. Previously implanted pulse generators and implantable cardioverter defibrillators should generally be explanted.
Chronic repositioning or removal of a tined lead – Proceed with extreme caution if a lead must be removed or repositioned. Chronic repositioning or removal of tined transvenous leads may be difficult because of fibrotic tissue development on the lead. In most clinical situations, it is preferable to abandon unused leads
4
in place. Return all removed leads, unused leads, or lead sections to Medtronic for analysis.
Lead removal may result in avulsion of the endocardium, valve, or vein.
Lead junctions may separate, leaving the lead tip and bare wire in the heart or vein.
Chronic repositioning of a lead may adversely affect a steroid lead’s low-threshold performance.
An abandoned lead should be capped so that the lead does not transmit electrical signals.
Severed leads should have the remaining lead end sealed and the lead body sutured to adjacent tissue.
Connector compatibility – Although the lead conforms to the IS-1 International Connector Standard, do not attempt to use the lead with any device other than a commercially available implantable pacing system with which it has been tested and demonstrated to be safe and effective. The potential adverse consequences of using such a combination may include, but are not limited to, undersensing cardiac activity and failure to deliver necessary therapy.

6 Drug information

6.1 Mechanism of action

Steroid suppresses the inflammatory response that is believed to cause threshold rises typically associated with implanted pacing electrodes. Dexamethasone acetate is a synthetic steroid of the glucocorticoid family. Glucocorticoids have potent anti-inflammatory actions via direct and indirect effects on major inflammatory cells. Glucocorticosteroids bind to a cytoplasmic glucocorticoid receptor as well as a membrane-bound receptor. Binding to the cytoplasmic receptor leads to receptor activation and translocation to the nucleus. The receptor interacts with specific DNA sequences within the regulatory regions of affected genes. Thus, glucocorticoids inhibit the production of multiple cell factors that are critical in generating the inflammatory response.

6.2 Pharmacokinetics and metabolism

Pharmacokinetics – The pharmacokinetics (local drug levels
and systemic levels) of dexamethasone acetate and its metabolites following implant were not evaluated in human clinical trials. When delivered intra-muscularly, the lipid-soluble dexamethasone acetate is slowly absorbed throughout the tissue.
Metabolism – The conversion of dexamethasone acetate to dexamethasone occurs within hours. The dexamethasone alcohol (dexamethasone) is the active glucocorticoid used in this Medtronic lead. Steroid is applied via MCRD (Monolithic controlled release device) and eluted to the tissue interface where it will be used. The form of the steroid, whether it is a prodrug or the pharmacologically active dexamethasone, is irrelevant, as the steroid is directly present at the injury site to treat the inflammation. Dexamethasone acetate is hydrolyzed into dexamethasone, which is readily absorbed by the surrounding
tissue and body fluids. Glucocorticoids, when given systemically, are eliminated primarily by renal excretion of inactive metabolites.

6.3 Mutagenesis, carcinogenicity, and reproductive toxicity

The mutagenesis, carcinogenicity, and reproductive toxicity of the Model 4574 lead have not been evaluated. However, the mutagenesis, carcinogenicity, and reproductive toxicity of dexamethasone acetate have previously been evaluated.
Mutagenesis – Genotoxicity evaluation of dexamethasone was undertaken using in vitro and in vivo assays. Analyses of chromosomal aberrations, sister-chromatid exchanges in human lymphocytes, and micronuclei and sister-chromatid exchanges in mouse bone marrow showed dexamethasone to be capable of attacking the genetic material. However, the Ames/Salmonella assay, both with and without S9 mix, did not show any increase His+ revertants.
Carcinogenicity – Although adequate and well-controlled animal studies have not been performed on Dexamethasone acetate, use in humans has not shown an increase in malignant disease.
Reproductive Toxicity – Adrenocorticoids have been reported to increase or decrease the number and motility of spermatozoa. However, it is not known whether reproductive capacity in humans is adversely affected.
Pregnancy – Adrenocorticoids cross the placenta. Although adequate studies have not been performed in humans, there is some evidence that pharmacologic doses of adrenocorticoids may increase the risk of placental insufficiency, decreased birth weights or stillbirth. However, tetrogenic effects in humans have not been confirmed.
Infants born to mothers who have received substantial doses of adrenocorticoids during pregnancy should be carefully observed for signs of hypoadrenalism and replacement therapy administered as required.
Prenatal administration of dexamethasone to the mother to prevent respiratory distress syndrome in the premature neonate has not been shown to affect the child’s growth or development adversely. Physiologic replacement doses of adrenocorticoids administered for treatment of adrenal insufficiency are also unlikely to adversely affect the fetus or neonate. Animal studies have shown that adrenocorticoids increase the instance of cleft palate, placental insufficiency, spontaneous abortions, and intrauterine growth retardation.
Lactation – Problems in humans have not been documented. Adrenocorticoids are excreted in breast milk and may cause unwanted defects such as growth suspension and inhibition of endogenous steroid production in the infant.
5
Loading...
+ 11 hidden pages