Medtronic 439878 Technical Manual

ATTAIN PERFORMA™ STRAIGHT MRI SURESCAN™ 4398
Steroid-eluting, quadripolar electrode, transvenous, over-the-wire, cardiac vein pacing lead
Technical Manual
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
The following list includes trademarks or registered trademarks of Medtronic in the United States and possibly in other countries. All other trademarks are the property of their respective owners.
AccuRead, Attain, Attain Hybrid, Medtronic, Performa, SureScan

Contents

9 Implant procedure 8 10 Specifications 15 11 Medtronic warranty 16 12 Service 16

1 Description

The Medtronic Attain Performa Straight MRI SureScan 4398 steroid-eluting, quadripolar electrode, transvenous, over-the-wire LV lead is designed for pacing via a cardiac vein. The lead has been tested for use in the Magnetic Resonance Imaging (MRI) environment. All lead lengths for this lead model are MR Conditional. This lead contains 4 electrodes designed to function as cathodes or anodes, depending on how the device LV pacing polarity is programmed:
electrode LV1, the distal electrode, positioned near the distal tip of the lead
electrode LV2, positioned 21 mm proximal to electrode LV1
electrode LV3, positioned 1.3 mm proximal to electrode LV2
electrode LV4, the proximal electrode, positioned 21 mm proximal to electrode LV3
See Section 9.10, “Taking electrical measurements”, page 11 for information about LV pacing polarity selections.
The Medtronic IS4-LLLL1 four-pole inline connector on the lead facilitates device connection during implant. The tip of the connector pin has a white band indicator that can be used for visual confirmation of proper lead connection to the device.
The connector contacts align with the lead electrodes LV1 to LV1, LV2 to LV2, and so on:
connector contact LV1, the connector pin, positioned at the proximal tip of the lead (aligns to electrode LV1)
connector contact LV2, positioned distal to connector pin LV1 (aligns to electrode LV2)
connector contact LV3, positioned distal to connector contact LV2 (aligns to electrode LV3)
connector contact LV4, positioned distal to connector contact LV3 (aligns to electrode LV4)
See Section 10.2, “Specifications drawing (nominal)”, page 16 for an illustration of the lead electrodes and connector contacts.
The distal tip of the lead allows a guide wire to pass through to aid in cardiac vein selection. The tip contains a silicone rubber membrane, which seals the lead inner lumen to reduce blood ingress.
Each electrode contains a Monolithic controlled release device (MCRD) for elution of steroid to reduce inflammatory response within the cardiac vein. The MCRDs contain a combined-total target dosage of 288 µg of dexamethasone acetate steroid. The target dose of the steroid is 72 µg at each MCRD. Upon exposure to body fluids, the steroid elutes from the MCRDs. The steroid suppresses the inflammatory response that is believed to cause threshold rise typically associated with implanted pacing electrodes.
The lead features 2 polyurethane tines located between electrode LV1 and electrode LV2 to facilitate contact with the cardiac veins.
The outer insulation of the lead is polyurethane and the inner insulation is SI-polyimide (SI-PI)2. The SI-PI is applied as a coating to the conductor wire before coiling.
The Attain Performa Straight MRI SureScan 4398 LV lead can be positioned with the aid of a guide wire, a stylet, an inner catheter, or an inner catheter plus a guide wire or a stylet.
To implant the lead in a selected cardiac vein, a compatible delivery system is required. A compatible delivery system includes a guide catheter and either a hemostasis valve or an introducer valve that can be removed or that allows passage over the lead connector. Contact a Medtronic representative for further information regarding compatible delivery systems.

1.1 Medtronic SureScan system

The Model 4398 lead is part of the Medtronic SureScan system. The SureScan system includes a Medtronic SureScan device connected to Medtronic SureScan leads. Labeling for SureScan system components displays the SureScan logo and the MR Conditional symbol. To verify that components are part of a SureScan system, visit http://www.mrisurescan.com.
SureScan logo
MR Conditional symbol. The Medtronic SureScan sys­tem is MR Conditional and is designed to allow implan­ted patients to undergo an MRI scan under the speci­fied MRI conditions for use.
1
IS4-LLLL refers to an International Connector Standard (ISO 27186:2010) whereby pulse generators and leads so designated are assured of a basic mechanical fit. LLLL defines the lead connector contacts as low voltage (L).
2
Technology developed by NASA.
3
The MRI SureScan feature permits a mode of operation that allows a patient with a SureScan device to be safely scanned by an MRI machine while the device continues to provide appropriate pacing. When programmed to On, MRI SureScan operation disables arrhythmia detection, magnet mode, and all user-defined diagnostics. Before performing an MRI scan,
refer to the SureScan MRI technical manual for important information about procedures and MRI-specific warnings and precautions.

1.2 Contents of package

Leads and accessories are supplied sterile. Each package contains the following items:
1 lead with anchoring sleeve
1 guide wire insertion tool
1 guide wire clip
1 guide wire steering handle
4 stylets
2 AccuRead 2.0 analyzer cable interface tools
product documentation

1.3 Accessory descriptions

Dispose of all single-use accessories according to local environmental requirements.
AccuRead 2.0 tool – The AccuRead 2.0 analyzer cable interface tool guides the connection of analyzer cable terminals to the lead connector contacts. This connection facilitates accurate electrical measurements during implant and prevents possible connector damage.
Anchoring sleeve – An anchoring sleeve secures the lead to prevent it from moving and protects the lead insulation and conductors from damage caused by tight sutures.
Guide wire clip – A guide wire clip secures the excess guide wire and helps to protect and maintain the sterility of the guide wire.
Guide wire insertion tool – A guide wire insertion tool provides additional control when inserting a guide wire into the lead connector pin or the lead tip.
Guide wire steering handle – A guide wire steering handle is used only with guide wires 0.46 mm (0.018 in) or less in diameter. The steering handle provides additional guide wire steering and rotation control.
Stylet – A stylet provides additional stiffness and controlled flexibility for maneuvering the lead into position. Each stylet knob is labeled with the stylet diameter and corresponding lead length.

2 Drug component description

The active ingredient in the Attain Performa Straight MRI SureScan 4398 LV lead is dexamethasone acetate [21-(acetyloxy)-9-fluoro-11β, 17-trihydroxy-16α-methylpregna-1,4-diene-3,20-dione]. The structural formula for this steroid is as follows:
Dexamethasone acetate (DXAC) - C24H31FO
6
Figure 1.
The target dosage of dexamethasone acetate is 288 µg per lead.
Cautions:
Drug interactions of dexamethasone acetate with this lead have not been studied.
Before implanting this lead, consider total patient exposure to dexamethasone acetate.

3 Indications

The Attain Performa Straight MRI SureScan 4398 steroid-eluting, quadripolar electrode, IS4 transvenous lead is indicated for chronic pacing in the left ventricle via the cardiac vein, when used with a compatible Medtronic Cardiac Resynchronization Therapy (CRT) system. Extended bipolar pacing is available using this lead in combination with a compatible CRT-D system and RV defibrillation lead.

4 Contraindications

Before performing an MRI scan, refer to the SureScan MRI technical manual for MRI-specific contraindications.
The Model 4398 lead is contraindicated as follows:
Coronary vasculature – This lead is contraindicated for patients with coronary venous vasculature that is inadequate for lead placement, as indicated by venogram.
Steroid use – The lead is contraindicated in patients for whom a single dose of 288 µg of dexamethasone acetate may be contraindicated.

5 Warnings and precautions

A complete SureScan system is required for use in the MRI environment. Before performing an MRI scan, refer to the SureScan MRI technical manual for MRI-specific warnings and precautions.
Note: Medical procedure warnings and precautions that pertain
to the Medtronic implanted system are provided in the manual that is packaged with the device or on the Medtronic Manual Library website (www.medtronic.com/manuals).
4
Inspecting the sterile package – Inspect the sterile package with care before opening it.
If the seal or package is damaged, contact a Medtronic representative.
Store at 25 °C (77 °F). Excursions from this storage temperature are permitted in the range of 15 to 30 °C (59 to 86 °F). (See USP Controlled Room Temperature.) According to USP excursion conditions, transient spikes up to 40 °C (104 °F) are permitted as long as they do not exceed 24 hours.
Do not use the product after its expiration date.
Sterilization – Medtronic has sterilized the package contents with ethylene oxide before shipment. This lead is for single use only and is not intended to be resterilized.
Single use – The lead and accessories are for single use only.
Necessary hospital equipment – Keep external defibrillation
equipment nearby for immediate use during acute lead system testing, the implant procedure, or whenever arrhythmias are possible or intentionally induced during post-implant testing.
External defibrillation and cardioversion – External defibrillation and cardioversion are therapies that deliver an electrical shock to the heart to convert an abnormal heart rhythm to a normal rhythm.
Medtronic cardiac devices are designed to withstand exposure to external defibrillation and cardioversion. While damage to an implanted system from an external shock is rare, the probability increases with increased energy levels. These procedures may also temporarily or permanently elevate pacing thresholds or temporarily or permanently damage the myocardium. If external defibrillation or cardioversion are required, consider the following precautions:
Use the lowest clinically appropriate energy.
Position the patches or paddles a minimum of 15 cm (6 in) away from the device.
Position the patches or paddles perpendicular to the device and lead system.
If an external defibrillation or cardioversion is delivered within 15 cm (6 in) of the device, use a Medtronic programmer to evaluate the device and lead system.
Line-powered and battery-powered equipment – An implanted lead forms a direct current path to the myocardium. During lead implant and testing, use only battery-powered equipment or line-powered equipment specifically designed for this purpose to protect against fibrillation that may be caused by alternating currents. Line-powered equipment used in the vicinity of the patient must be properly grounded. Lead connector pins must be insulated from any leakage currents that may arise from line-powered equipment.
Concurrent devices – Output pulses, especially from unipolar devices, may adversely affect device sensing capabilities. If a patient requires a separate stimulation device, either permanent or temporary, allow enough space between the leads of the separate systems to avoid interference in the sensing capabilities of the devices. Previously implanted pulse generators and
implantable cardioverter defibrillators should generally be explanted.
Steroid use – It has not been determined whether the warnings, precautions, or complications usually associated with injectable dexamethasone acetate apply to the use of this highly localized, controlled-release device. For a list of potential adverse effects, refer to the Physicians’ Desk Reference.
Handling steroid monolithic controlled release devices (MCRDs) – Avoid reducing the amount of steroid available before
implanting the lead. Reducing the available amount of steroid may adversely affect low-threshold performance.
Do not allow the electrode surfaces to come in contact with surface contaminants.
Do not wipe or immerse the electrodes in fluid, except blood, at the time of implant.
Bipolar pacing – If the surface area of the selected anode electrode is equal to or less than the surface area of the selected cathode, higher pacing thresholds or anodal stimulation may result. This lead uses 4 equal-sized electrodes; therefore, bipolar pacing configurations may result in higher pacing thresholds or anodal stimulation.
If therapy cannot be delivered via the LV1 electrode, the LV2, LV3, and LV4 electrodes are available for use with specific Medtronic devices. Refer to the appropriate Medtronic Cardiac Resynchronization Therapy (CRT-D) system manual for use of available LV lead pacing polarity options.
Handling the stylet – Handle the stylet with care at all times.
To minimize the likelihood of trauma to the vein and to maintain lead flexibility while advancing the lead through the vein, keep the stylet withdrawn 1 to 2 cm or select a more flexible stylet.
Do not use excessive force or surgical instruments when inserting a stylet.
Avoid overbending, kinking, or blood contact on stylets.
Use a new stylet when blood or other fluids accumulate on the stylet. Accumulated fluids may cause damage to the lead or difficulty in passing the stylet through the lead.
Curving the distal end of the stylet prior to insertion into the lead will achieve a curvature at the distal end of the lead. Do not use a sharp object to impart a curve to the distal end of the stylet.
Handling the guide wire – Handle the guide wire with care at all times.
Do not insert the proximal end of the guide wire through the lead tip seal without using the guide wire insertion tool. Inserting the guide wire without the guide wire insertion tool may damage the lead.
Damage to a guide wire may prevent the guide wire from performing with accurate torque response and control and may cause vessel damage. For additional information about vessel damage and other potential adverse events, refer to the technical manual packaged with the appropriate guide wire.
5
If the distal end of the guide wire becomes severely kinked or twisted, it may be difficult to withdraw it back through the lead. Therefore, if there is an indication that the distal end of the guide wire has become damaged, or if there is significant resistance in guide wire passage, remove the lead and guide wire together as a unit. Remove the guide wire from the lead and insert a new guide wire into the lead. Do not use excessive force to retract the guide wire from the lead. Refer to the product documentation packaged with the guide wire for additional information.
Handling the lead – Handle the lead with care at all times.
Rust stylets are not recommended with this lead due to the risk of conductor coil or insulation perforation.
If a stylet is used for lead positioning, use only the stylets packaged with the lead or in a stylet kit (downsized knob). Other stylets may extend beyond the lead tip causing patient injury.
If the lead is damaged, do not implant it. Return the lead to a Medtronic representative.
Protect the lead from materials that shed particles such as lint and dust. Lead insulators attract these particles.
Handle the lead with sterile surgical gloves that have been rinsed in sterile water or a comparable substance.
Do not severely bend, kink, or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pin.
Do not immerse leads in mineral oil, silicone oil, or any other liquid, except blood, at the time of implant.
Use an anchoring sleeve with all leads. Ensure that the anchoring sleeve is positioned close to the lead connector pin, to prevent inadvertent passage of the sleeve into the vein. If it is necessary to wipe the lead before insertion, ensure that the anchoring sleeve remains in position.
Do not force the guide catheter or leads if significant resistance is encountered. Use of guide catheters or leads may cause trauma to the heart.
Chronic repositioning or removal – Chronic repositioning or removal of leads may be difficult because of fibrotic tissue development. The clinical study was not designed to evaluate the removal of left ventricular leads from the coronary venous vasculature. If a lead must be removed or repositioned, proceed with extreme caution. Return all removed leads to Medtronic.
Rust stylets are not recommended with this lead due to the risk of conductor coil/insulation perforation.
Verify lead length on the lead label on the connector to choose an appropriate stylet kit (downsized knob) length when repositioning. Always choose a stylet kit (downsized knob) 3 cm (1.2 in) shorter than the lead length. For example, choose a stylet kit (downsized knob) with stylets 75 cm long for a lead 78 cm long.
Lead removal may result in avulsion of the endocardium, valve, or vein.
Lead junctions may separate, leaving the lead tip and bare wire in the heart or vein.
Chronic repositioning may adversely affect the low-threshold performance of a steroid-eluting lead.
Cap abandoned leads to avoid transmitting electrical signals.
For leads that have been severed, seal the remaining lead end and suture the lead to adjacent tissue.
If a lead is removed and repositioned, inspect it carefully for insulator or conductor coil damage before repositioning.
Magnetic resonance imaging (MRI) – An MRI is a type of medical imaging that uses magnetic fields to create an internal view of the body. If certain criteria are met and the warnings and precautions provided by Medtronic are followed, patients with an MR Conditional device and lead system are able to undergo an MRI scan; for details, refer to the SureScan MRI technical manual that Medtronic provides for an MR Conditional device.
Diathermy treatment (including therapeutic ultrasound) –
Diathermy is a treatment that involves the therapeutic heating of body tissues. Diathermy treatments include high frequency, short wave, microwave, and therapeutic ultrasound. Except for therapeutic ultrasound, do not use diathermy treatments on cardiac device patients. Diathermy treatments may result in serious injury or damage to an implanted device and lead system. Therapeutic ultrasound (including physiotherapy, high intensity therapeutic ultrasound, and high intensity focused ultrasound), is the use of ultrasound at higher energies than diagnostic ultrasound to bring heat or agitation into the body. Therapeutic ultrasound is acceptable if treatment is performed with a minimum separation distance of 15 cm (6 in) between the applicator and the implanted device and lead system, as long as the ultrasonic beam is pointing away from the device and lead system.

6 Potential adverse events

The potential adverse events (listed in alphabetical order) related to the use of transvenous leads include, but are not limited to, the following conditions:
Air embolism
Avulsion or other damage to the endocardium, valve, or vein (particularly in fragile hearts)
Cardiac dissection or perforation
Cardiac tamponade
Coronary sinus dissection
Death
Endocarditis or pericarditis
Erosion through the skin
Extracardiac muscle or nerve stimulation
Fibrillation or other arrhythmias
Heart block
Heart wall or vein wall rupture
Hematoma/seroma
Infection
Lead conductor fracture or insulation failure
Lead dislodgement
Myocardial irritability
Myopotential sensing
Pericardial effusion or rub
Pneumothorax
Rejection phenomena (local tissue reaction, fibrotic tissue formation)
6
Threshold elevation or exit block
Thrombosis
Thrombotic embolism
Additional potential adverse events related to the lead and the programmed parameters include, but are not limited to, the following:
Potential adverse event
Lead dislodgementaIntermittent or contin-
Lead dislodgementaIntermittent or contin-
Lead conductor frac­ture
Lead conductor insu­lation failure
Threshold elevation or exit block
a
Transient loss of capture or LV EGM signal integrity (including sensing) may occur following surgery until lead stabilization takes place. If stabilization does not occur, lead dislodgement may be suspected.
Indicator of poten­tial adverse event
uous loss of capture or LV EGM signal integrity (including
a
sensing)
uous oversensing
Intermittent or contin­uous loss of capture or LV EGM signal integrity (including
a
sensing)
Intermittent or contin­uous loss of capture or LV EGM signal integrity (including
a
sensing)
Loss of capture
a
Corrective actions to consider
Reprogram the LV pacing polarity. Reposition the lead.
Reprogram the LV pacing polarity. Reposition the lead.
Replace the lead. Reprogram the LV pacing polarity.
Replace the lead. Reprogram the LV pacing polarity.
Adjust the implanta­ble device output. Reprogram the LV pacing polarity. Replace or reposition the lead.
Implant techniques that may damage the lead include, but are not limited to, the following techniques:
Implant techniques that may damage the lead
Forcing the lead through the intro­ducer/delivery system
Use of too medial of an approach with venous introducer resulting in clavicle and first rib binding
Using too stiff a stylet Conductor coil/insula-
Puncturing the perios­teum or tendon when using subclavian introducer approach resulting in binding
Possible effects on the lead
Electrode, conductor coil, or insulation damage
Conductor coil frac­ture, insulation dam­age
tion perforation
Conductor coil frac­ture, insulation dam­age
Corrective action to consider
Replace the lead.
Replace the lead.
Replace the lead.
Replace the lead.
Implant techniques that may damage the lead
Advancing the lead through the non-coro­nary central access veins without the stylet or guide wire fully inserted
Inserting the proximal end of the guide wire through the lead tip seal without using the guide wire insertion tool
Possible effects on the lead
Tip distortion or insu­lation perforation
Lead tip seal damage or conductor coil/insulation dam­age
Corrective action to consider
Replace the lead.
Replace the lead.

7 Drug information

7.1 Steroid mechanism of action

Steroid suppresses the inflammatory response that is believed to cause threshold rises typically associated with implanted pacing electrodes. Glucocorticoids decrease inflammation by stabilizing leukocyte lysosomal membranes. The membrane stabilization prevents the release of destructive acid hydrolases for the leukocytes and this inhibits the accumulation of macrophages in the inflamed area. The mechanism involves the activation of glucocorticoid receptors that increase or decrease the transcription of a number of genes involved in the inflammatory process. One of the key actions is the repression of cytokine gene transcription and other transcription factors activated in chronic inflammation.

7.2 Pharmacokinetics of leads using dexamethasone acetate steroid

Pharmacokinetics – The pharmacokinetics (local drug levels
and systemic levels) of dexamethasone acetate and its metabolites following implant were not evaluated in human clinical trials. When delivered intra-muscularly, the lipid-soluble dexamethasone acetate is slowly absorbed throughout the tissue.
Metabolism – The conversion of dexamethasone acetate to dexamethasone occurs within hours. The dexamethasone alcohol (dexamethasone) is the active glucocorticoid used in this Medtronic lead. Steroid is applied via monolithic controlled release device (MCRD) and eluted to the tissue interface where it will be used. The form of the steroid, whether it is a prodrug or the pharmacologically active dexamethasone, is irrelevant, as the steroid is directly present at the injury site to treat the inflammation. Dexamethasone acetate is hydrolyzed into dexamethasone, which is readily absorbed by the surrounding tissue and body fluids. Glucocorticoids, when given systemically, are eliminated primarily by renal excretion of inactive metabolites.
7

7.3 Mutagenesis, carcinogenicity, and reproductive toxicity

The mutagenesis, carcinogenicity, and reproductive toxicity of the Attain Performa Straight MRI SureScan 4398 LV lead have not been evaluated. However, the mutagenesis, carcinogenicity, and reproductive toxicity of dexamethasone acetate have previously been evaluated.
Mutagenesis – Genotoxicity evaluation of dexamethasone was undertaken using in vitro and in vivo assays. Analyses of chromosomal aberrations, sister-chromatid exchanges in human lymphocytes, and micronuclei and sister-chromatid exchanges in mouse bone marrow showed dexamethasone to be capable of attacking the genetic material. However, the Ames/Salmonella assay, both with and without S9 mix, did not show any increase His+ revertants.
Carcinogenicity – Although adequate and well-controlled animal studies have not been performed on Dexamethasone acetate, use in humans has not shown an increase in malignant disease.
Reproductive Toxicity – Adrenocorticoids have been reported to increase or decrease the number and motility of spermatozoa. However, it is not known whether reproductive capacity in humans is adversely affected.
Pregnancy – Adrenocorticoids cross the placenta. Although adequate studies have not been performed in humans, there is some evidence that pharmacologic doses of adrenocorticoids may increase the risk of placental insufficiency, decreased birth weights or stillbirth. However, tetrogenic effects in humans have not been confirmed.
Infants born to mothers who have received substantial doses of adrenocorticoids during pregnancy should be carefully observed for signs of hypoadrenalism and replacement therapy administered as required.
Prenatal administration of dexamethasone to the mother to prevent respiratory distress syndrome in the premature neonate has not been shown to affect the child’s growth or development adversely. Physiologic replacement doses of adrenocorticoids administered for treatment of adrenal insufficiency are also unlikely to adversely affect the fetus or neonate. Animal studies have shown that adrenocorticoids increase the instance of cleft palate, placental insufficiency, spontaneous abortions, and intrauterine growth retardation.
Lactation – Problems in humans have not been documented. Adrenocorticoids are excreted in breast milk and may cause unwanted defects such as growth suspension and inhibition of endogenous steroid production in the infant.

8 Clinical study

Information regarding the Model 4398 lead clinical study is available on the Medtronic Manual Library website:
1. Point your browser to http://www.medtronic.com/manuals.
2. Follow the instructions on the website to locate, view, print, or order the Clinical Study Summary.
If you do not have web access, you can order a printed copy of the Model 4398 Clinical Study Summary from your Medtronic representative or by calling the toll-free number located on the back cover.

9 Implant procedure

Warning: Before implanting a SureScan system, consider the
risks associated with removing previously implanted leads. Abandoned leads or previously implanted leads not tested for MRI compatibility compromise the ability to safely scan the SureScan system during MRI scans.
Warning: Do not force the guide catheter or lead if significant resistance is encountered. Use of guide catheters or leads may cause trauma to the heart.
To implant the Attain Performa Straight MRI SureScan 4398 LV lead in a selected cardiac vein, a compatible delivery system is required, such as a Medtronic delivery system. A compatible delivery system includes a guide catheter and either a hemostasis valve or an introducer valve that can be removed or that allows passage over the lead connector. Contact a Medtronic representative for further information regarding compatible delivery systems.
Proper surgical procedures and sterile techniques are the responsibility of the medical professional. The implant procedures described in this manual are provided for information only. Each physician must apply the information in these instructions according to professional medical training and experience.

9.1 Placing the right ventricular lead

When deciding which ventricular lead to place first, consider the ease of coronary sinus cannulation and the need for backup pacing.

9.2 Preparing the delivery system

Prepare the delivery system for lead implantation according to the instructions in the product documentation packaged with the delivery system.

9.3 Venous access

8
Warning: Backup pacing should be readily available during implant. Use of the delivery system or leads may cause heart block.
1. To access the subclavian vein, use a preferred method
1
1
2
1
2
based on professional experience. Caution: Certain anatomical abnormalities, such as thoracic outlet syndrome, may precipitate pinching and subsequent fracture of the lead. Caution: Insertion should be done as far lateral as possible to avoid clamping the lead body between the clavicle and the first rib (Figure 2).
Figure 2.
1 Suggested entry site
2. Introduce a J-shaped introducer guide wire and percutaneous introducer sheath.
3. Introduce the guide catheter assembly to access the coronary sinus.
See the delivery system product documentation for additional information.

9.4 Obtaining venograms

Before placing a lead in the coronary sinus, obtain venograms. Venograms are recommended to assess a route for passage and a site for final placement based on the size, shape, location, and tortuosity of the veins. Also, venograms may be useful in identifying suspected coronary sinus trauma. For information on obtaining a venogram by using a venogram balloon catheter, see the product documentation packaged with an appropriate venogram balloon catheter.

9.5 Inserting the lead into the delivery system

1. Insert a stylet or a guide wire into the lead to vary the shape of the distal end of the lead. Note: When a stylet is fully inserted, the distal tip of the stylet does not reach the distal tip of the lead.
2. Insert the lead into the delivery system. See the delivery system product documentation for additional information.

9.6 Lead placement

The Attain Performa Straight MRI SureScan 4398 LV lead can be placed with the aid of a guide wire, a stylet, an inner catheter, or an inner catheter plus a guide wire or a stylet.
Stylet delivery – If the patient’s anatomy features gentle vein angulation off the coronary sinus and the cardiac vein branch is not tortuous (Figure 3), a stylet may be used for lead delivery.
Figure 3.
1 Coronary sinus 2 Cardiac vein
Guide wire delivery – If the patient’s anatomy features acute vein angulation off the coronary sinus and the cardiac vein branch is tortuous (Figure 4), a guide wire may be used for lead delivery.
Figure 4.
Warning: If a stylet is used for lead positioning, use only the
stylets packaged with the lead or in a stylet kit (downsized knob). Always use a stylet that is 3 cm shorter than the lead length listed on the IS4 connector label. Other stylets may extend beyond the lead tip causing injury or perforation of the cardiac vein or heart.
Warning: Do not force the lead if significant resistance is encountered during lead passage. The use of guide catheters or leads may cause trauma to the heart.
Caution: Use care when handling the lead during insertion.
Do not severely bend, kink, or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pin.
The following steps may be used to insert the lead:
1 Tortuous cardiac vein branch with gentle angulation from the
coronary sinus
2 Tortuous cardiac vein branch with acute angulation from the coronary
sinus

9.7 Placing the lead using a stylet

Warning: Do not force the lead if significant resistance is
encountered during lead passage.
9
Warning: To minimize the likelihood of trauma to the vein and to maintain lead flexibility while advancing the lead through the vein, keep the stylet withdrawn 1 to 2 cm or select a more flexible stylet.
Caution: Do not use a sharp object to impart a curve to the distal end of the stylet. Imparting a curve to the stylet can be accomplished with a smooth-surface, sterile instrument (Figure 5).
Figure 5.
Note: If it is difficult to advance the stylet around a bend, consider
using a different stylet. More flexible stylets are recommended for tortuous anatomies. Firmer stylets are recommended where additional support is needed.
There are many techniques that may be used to advance the lead into a cardiac vein using a stylet. The choice of technique is left to the discretion of the physician.

9.8 Placing the lead using a guide wire

Warning: Do not insert the proximal end of the guide wire through
the lead tip seal without using the guide wire insertion tool. Inserting the guide wire without the guide wire insertion tool could damage the lead tip seal or the conductor coil or insulation.
Warning: Damage to a guide wire may prevent the guide wire from performing with accurate torque response and control and may cause vessel damage. For additional information about vessel damage and other potential adverse events, refer to the technical manual packaged with the appropriate guide wire.
Caution: Use care when positioning the guide wire. Refer to the product documentation packaged with the guide wire for additional information.
Notes:
Medtronic recommends using guide wires 0.36 mm to
0.46 mm (0.014 in to 0.018 in) in diameter. Contact a Medtronic representative for further information about recommended guide wires.
Consider soaking the guide wire in a heparin solution before insertion to minimize the risk of thrombus formation during use.
If you are using an Attain Hybrid guide wire, the procedure for preparing the guide wire for use differs from other guide wires because of the attached proximal knob.
Use the following steps to prepare the guide wire for use:
1. Select a guide wire. If the patient has tortuous anatomy, a more flexible guide wire is recommended. If additional support is needed, use a firmer guide wire. If you are using an Attain Hybrid guide wire, the steps involving the insertion tool (Step 2 through Step 4) do not apply.
2. Insert the guide wire into the lead by placing the distal (flexible) end of the guide wire into the lead connector pin using the guide wire insertion tool included in the package (Figure 6). To prevent the lead from dislodging from the guide wire insertion tool, grasp the lead and guide wire insertion tool firmly between thumb and forefinger. Caution: To minimize the risk of damaging the guide wire, be sure that the flexible section of the guide wire is fully inserted into the lead before removing the guide wire insertion tool from the lead. Note: Be sure to remove the guide wire insertion tool before lead implant.
Figure 6.
1 Guide wire 2 Lead connector pin
3. Disengage the guide wire insertion tool from the lead connector pin.
4. Remove the guide wire insertion tool by sliding the tool off the end of the guide wire.
5. Position the guide wire steering handle (Figure 7): For an Attain Hybrid guide wire, preload the guide wire steering handle onto the guide wire before loading the guide wire into the connector pin on the proximal end of the lead. For other guide wires:
a. Advance the guide wire steering handle over the proximal
(rigid) end of the guide wire.
b. Tighten the guide wire steering handle onto the guide
wire near the lead connector pin.
Figure 7.
1 Guide wire steering handle 2 Lead connector pin
6. Attach the guide wire clip to the guide wire and secure it within the sterile field. Medtronic recommends securing the guide wire clip to the patient’s sterile surgical drape.
10
As an alternate approach in situations where the guide wire is already in place, the lead can be loaded over the guide wire using the guide wire insertion tool.
Insert the guide wire into the lead by placing the proximal (rigid) end of the guide wire into the distal lead tip using the guide wire insertion tool included in the package (Figure 8).
Notes:
There may be slight resistance as the guide wire passes through the lead tip seal.
Grasp the lead and guide wire insertion tool between your thumb and forefinger to prevent the lead from dislodging from the guide wire insertion tool while you are inserting the guide wire.
Be sure to remove the guide wire insertion tool, through the slit in the guide wire insertion tool, before lead implant.
Figure 8.

9.9 Fixating the lead

Fixation is achieved the same way regardless of lead placement.
Using fluoroscopy for guidance, fixate the lead into the cardiac vein (Figure 9).
Figure 9.
1 Guide wire 2 Lead distal tip
Warning: Do not force the lead if significant resistance is encountered during lead passage.
Caution: If the distal end of the guide wire becomes severely kinked or twisted, it may be difficult to withdraw it back through the lead. Therefore, if there is an indication that the distal end of the guide wire has become damaged, or if there is significant resistance in guide wire passage, remove the lead and guide wire together as a unit. Remove the guide wire from the lead and insert a new guide wire into the lead. Do not use excessive force to retract the guide wire from the lead.
Notes:
If the lead is not advancing, or if the lead and guide wire seem to be sticking together, there may be thrombus on the guide wire at the lead tip. Remove and inspect the lead and guide wire. Consider using a new guide wire. Reinsert the lead and the guide wire as described in previous steps.
If it is difficult to advance the guide wire around a bend, consider using a different guide wire. More flexible guide wires are recommended for tortuous anatomies. Firmer guide wires are recommended where additional support is needed.
There are many techniques that may be used to advance the lead into a cardiac vein using a guide wire. The choice of technique is left to the discretion of the physician.
For optimal electrical performance and fixation, place the proximal electrode in the selected vein, not in the coronary sinus.
Note: If the selected vein is large, it may be necessary to position the lead in a smaller cardiac vein in order to achieve lead fixation. For optimal electrical performance, Medtronic recommends electrode-tissue contact.

9.10 Taking electrical measurements

The Attain Performa Straight MRI SureScan 4398 LV lead was designed to provide pacing via selectable electrodes. The 16 available pacing polarities are shown in Figure 10.
11
Figure 10.
RV coil
RV lead
LV lead
LV lead
RV coil
RV lead
LV3
LV3
LV2
LV2
LV1
LV1
LV4
LV4
1 Extended bipolar pacing polarities 2 Bipolar (reversible) pacing polarities
Extended bipolar pacing is available using this lead in combination with a compatible CRT-D system and RV defibrillation lead. The 4 available pacing polarities include:
LV1 to RVcoil
LV2 to RVcoil
LV3 to RVcoil
LV4 to RVcoil
Bipolar pacing is available using this lead in combination with a compatible CRT-D system. The polarity of each of the 6 electrode pairs can be reversed to yield a total of 12 bipolar pacing polarities:
LV1 to LV2, LV2 to LV1
LV1 to LV3, LV3 to LV1
LV1 to LV4, LV4 to LV1
LV2 to LV3, LV3 to LV2
LV2 to LV4, LV4 to LV2
LV3 to LV4, LV4 to LV3 Note: If you are monitoring the LV EGM signal when the LV2 to LV3 or LV3 to LV2 pacing polarity is selected, the signal amplitude may be attenuated compared to other pacing configurations such as LV1 to RVcoil. This signal attenuation is an expected characteristic of the Attain Performa Straight MRI SureScan 4398 LV lead with the short spacing between electrodes LV2 and LV3.
Caution: Before taking electrical or defibrillation efficacy measurements, move objects made of conductive materials, such as guide wires or stylets, away from all electrodes.
Note: Before taking electrical measurements, it is recommended that you retract the stylet or guide wire, inside the lead lumen, to a point proximal of all electrodes. This action allows the lead tip to
12
resume its normal shape, enabling appropriate electrode-tissue contact to occur.
The AccuRead 2.0 analyzer cable interface tool is used to facilitate accurate electrical measurements during implant.
Caution: Use the AccuRead 2.0 cable interface tool when connecting analyzer cable terminals to the lead connector. This tool enables you to take accurate electrical measurements at implant while reducing the risk of connector damage, electrical bridging, or electrical shorting. The potential for connector damage, bridging, or shorting is due to variations in analyzer cable terminals and due to the width and proximity of the contacts (rings and pin) on the IS4 connector.
Use the following steps to take electrical measurements:
1. Ensure that the lead connector is inserted into the AccuRead 2.0 tool completely. If the AccuRead 2.0 tool is properly attached, the connector pin is accessible (see Figure 11).
Figure 11.
2. Attach a surgical cable to the lead connector. Use the AccuRead 2.0 cable interface tool to guide the alignment of the cable clips with the contacts on the lead connector. The tool helps to ensure that accurate readings are obtained. See the lead drawing in Chapter 10 for information about the alignment of the lead connector contacts with the lead electrodes.
3. Use an implant support instrument, such as a pacing system analyzer, for obtaining electrical measurements. For information on the use of the implant support instrument, consult the product documentation for that device. Satisfactory lead placement is indicated by low stimulation thresholds and adequate sensing of intracardiac signal amplitudes.
Notes:
A low stimulation threshold provides a desirable safety margin, allowing for a possible rise in thresholds that may occur within 2 months following implant.
Adequate acute LV EGM signal amplitudes ensure that the lead is properly sensing intrinsic cardiac signals. Minimum signal requirements depend on the sensitivity capabilities of the device. Acceptable acute signal amplitudes for the lead must be greater than the minimum device sensing capabilities. Be sure to include an adequate safety margin to account for lead maturity.

Table 1. Recommended measurements at implant

Measurement recommended Left ventricle
Maximum acute stimulation thresholdsb3.0 V
Minimum acute LV EGM signal amplitude 4.0 mV
a
Assuming 500 Ω resistance.
b
At a pulse width setting of 0.5 ms.
c
The LV EGM signal amplitude for the LV2 to LV3 and LV3 to LV2 pacing polarities may be attenuated compared to other pacing polarities such as LV1 to RVcoil.
a
c
4. If electrical measurements do not stabilize to acceptable levels, it may be necessary to reposition the lead and repeat the testing procedure.
5. Check for phrenic nerve stimulation by pacing at 10 V and a pulse width setting greater than 0.5 ms. Then observe for diaphragmatic contracting either by fluoroscopy or direct abdominal palpitation. Further testing may include patient positional changes to simulate upright chronic conditions. If phrenic nerve stimulation occurs, reduce the voltage until a phrenic nerve stimulation threshold is determined. Phrenic nerve stimulation may necessitate changing the pacing polarity or repositioning of the lead.
6. The AccuRead 2.0 tool may be removed from a guide wire or stylet using the slit on the side of the tool (see Figure 12).
Figure 12.
3
1 Removing the AccuRead 2.0 tool from the connector pin 2 Removing the AccuRead 2.0 tool from the stylet using the slit on the
side of the tool

9.11 Removing the guide catheter from the lead

Note: If an Attain Hybrid guide wire was used for placement of the
lead, go to Step 2. It is not necessary to replace the Attain Hybrid guide wire with a stylet for catheter removal.
Once the lead is in the final position, remove the guide catheter from the lead:
1. If used, remove the guide wire and guide wire insertion tool. Replace the guide wire with a straight stylet (downsized knob). Insert the straight stylet into the lead to the mid-coronary sinus.
2. Remove the guide catheter from the lead. See the delivery system product documentation for details. Note: For Medtronic slittable delivery systems, use a slitter compatible with a 1.75 mm (5.3 French) lead body.
3
The LV EGM signal amplitude for the LV2 to LV3 and LV3 to LV2 pacing polarities may be attenuated compared to other pacing polarities such as LV1 to RVcoil.
13
3. Carefully and completely remove the stylet or Attain Hybrid guide wire. When removing the stylet, grip the lead firmly close to the distal end of the connector pin; gripping the lead at this location helps prevent possible lead tip dislodgement.
4. Ensure that the lead connector is inserted into the AccuRead 2.0 analyzer cable interface tool and then repeat the electrical measurements. See Section 9.10, “Taking electrical measurements”, page 11.

9.12 Anchoring the lead

Caution: Use care when anchoring the lead.
Use only nonabsorbable sutures to anchor the lead.
Do not attempt to remove or cut the anchoring sleeve.
Do not use the anchoring sleeve tabs for suturing.
During lead anchoring, take care to avoid dislodging the lead tip.
Do not secure sutures so tightly that they damage the vein, lead, or anchoring sleeve (Figure 13).
Do not tie a suture directly to the lead body (Figure 13).
Figure 13.
Anchor the lead using all 3 grooves:
1. Position the anchoring sleeve against or near the vein.
2. Secure the anchoring sleeve to the lead body by tying a suture firmly in each of the 3 grooves (Figure 14).
Figure 14.
3. Use at least 1 additional suture in 1 of the grooves to secure the anchoring sleeve and lead body to the fascia.

9.13 Connecting the lead

Caution: Before connecting the lead to the device, always remove
implant tools such as a stylet, guide wire, or AccuRead 2.0 cable interface tool. Failure to remove implant tools may result in lead failure.
Connect the lead to an implantable device.
1. Insert the lead connector into the device connector block. Consult the product documentation packaged with the implantable device for instructions on proper lead connections.
2. Take electrical measurements through the device.

9.14 Placing the device and lead into the pocket

Caution: Use care when placing the device and leads into
the pocket.
Ensure that the leads do not leave the device at an acute angle.
Do not grip the lead or device with surgical instruments.
Do not coil the lead. Coiling the lead can twist the lead body and may result in lead dislodgement (Figure 15).
Figure 15.
Use the following steps to place the device and leads into the pocket:
1. To prevent undesirable twisting of the lead body or bending of the lead body at an acute angle, rotate the device to wrap the excess lead length loosely (Figure 16).
Figure 16.
2. Insert the device and leads into the pocket.
3. Before closing the pocket, verify sensing, pacing, cardioversion, and defibrillation efficacy.

9.15 Post-implant evaluation

After implant, monitor the patient’s electrocardiogram until the patient is discharged. If a lead dislodges, it usually occurs during the immediate postoperative period.
Recommendations for verifying that the lead is properly positioned include x-rays and pacing thresholds taken at pre-hospital discharge, 3 months after implant, and every 6 months thereafter.
In the event of a patient death, explant all implanted leads and devices and return them to Medtronic with a completed Product Information Report form.
14

10 Specifications

10.1 Specifications (nominal)

Parameter 4398
Serial number prefix QUB
Type Quadripolar elec-
Chamber paced Left ventricle
Length 78 and 88 cm (30.71
Connector IS4-LLLL
Material Conductor: 25% Ag-core-
Insulators: Polyurethane (outer)
Electrodes: Platinum iridium
Connector pin: MP35N
Connector rings:
Molded tip seal:
Electrode configuration Radiused, titanium-
Diameter Lead body: 1.75 mm
Electrodes: 1.70 mm
Spacer between elec­trodes LV1 (distal) and LV2:
Spacer between elec­trodes LV2 and LV3:
Spacer between elec­trodes LV3 and LV4 (proximal):
Medtronic delivery sys­tem (recommended inner diameter)
Diagnostic guide wire (recommended diame­ter)
Electrode surface area 5.8 mm
Distance between elec­trodes
Electrode LV1 (distal) to LV2:
Electrode LV2 to LV3:
trode
and 34.65 in)
MP35N
SI-polyimide (SI-PI)
a
(inner)
alloy with titanium nitride coating
MP35N
Silicone rubber
nitride-coated, ste­roid-eluting
(5.3 French)
(5.1 French)
1.30 mm (3.9 French)
1.57 mm (4.7 French)
1.30 mm (3.9 French)
1.90 mm (5.7 French)
0.36 mm to 0.46 mm (0.014 in to 0.018 in)
2
21 mm
1.3 mm
Parameter 4398
Electrode LV3 to LV4 (proxi­mal):
Conductor resistance LV1 22 ±5 Ω (78 cm)
LV2 19 ±4 Ω (78 cm)
LV3 18 ±4 Ω (78 cm)
LV4 17 ±4 Ω (78 cm)
Steroid Dexamethasone
Target dose of steroid 72 µg at each mono-
Steroid binder Silicone rubber
a
Technology developed by NASA.
21 mm
24 ±6 Ω (88 cm)
21 ±5 Ω (88 cm)
21 ±4 Ω (88 cm)
20 ±4 Ω (88 cm)
acetate
lithic controlled release device (MCRD) 288 µg target com­bined amount
15

10.2 Specifications drawing (nominal)

Figure 17.
6 IS4 connector 7 Contact LV4 8 Contact LV3 9 Contact LV2
10 Contact LV1

11 Medtronic warranty

For complete warranty information, see the accompanying warranty document.

12 Service

Medtronic employs highly trained representatives and engineers located throughout the world to serve you and, upon request, to provide training to qualified hospital personnel in the use of Medtronic products. Medtronic also maintains a professional staff to provide technical consultation to product users. For more information, contact your local Medtronic representative, or call or write Medtronic at the appropriate telephone number or address listed on the back cover.
1 Electrode LV1 (the distal electrode): nominally 5.8 mm2 geometric
pacing surface area 2 Electrode LV2: nominally 5.8 mm2 geometric pacing surface area 3 Electrode LV3: nominally 5.8 mm2 geometric pacing surface area 4 Electrode LV4 (the proximal electrode): nominally 5.8 mm2 geometric
pacing surface area 5 Anchoring sleeve
16
Medtronic, Inc.
*M959361A001*
710 Medtronic Parkway Minneapolis, MN 55432 USA www.medtronic.com +1 763 514 4000
Medtronic USA, Inc.
Toll-free in the USA (24-hour technical consultation for physicians and medical professionals) Bradycardia: +1 800 505 4636 Tachycardia: +1 800 723 4636
Europe/Middle East/Africa
Medtronic International Trading Sàrl Route du Molliau 31 Case Postale 84 CH-1131 Tolochenaz Switzerland +41 21 802 7000
Technical manuals
www.medtronic.com/manuals
© 2015 Medtronic, Inc. M959361A001 B 2015-06-01
Loading...