Medtronic 419478 Technical Manual

ATTAIN® BIPOLAR OTW 4194
Steroid eluting, transvenous, bipolar, left ventricular, over the wire, cardiac vein pacing lead
Technical Manual
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
The following list includes trademarks or registered trademarks of Medtronic in the United States and possibly in other countries. All other trademarks are the property of their respective owners.

Contents

1 Required physician training 3 2 Device description 3 3 Indications for use 4 4 Contraindications 4 5 Warnings and precautions 4 6 Potential adverse events 5 7 Clinical trials 6 8 Directions for use 9
9 Detailed device description 15 10 Medtronic warranty 17 11 Service 17

1 Required physician training

In order to implant a Medtronic Attain Bipolar OTW 4194 lead, physicians are required to:
Thoroughly read this manual, and all associated device technical manuals.
Provide a copy of the patient manual to the patient and discuss it with him or her and any other interested parties.
Be trained on the following topics – Indications for use – Device operation to ensure therapy delivery – Measuring and managing biventricular thresholds – Assembly and use of LV lead implant tools – Placement of the LV lead – Patient management and system follow-up
Prior to implanting the system, Medtronic will certify that physicians received training.

2 Device description

The Medtronic Attain Bipolar OTW 4194 steroid eluting, transvenous, bipolar, left ventricular, over the wire, cardiac vein pacing lead is designed for pacing and sensing via a cardiac vein, as part of a Medtronic biventricular pacing system. The tip to anode spacing is 11 mm.
The lead features a tapered annular platinum alloy electrode tip and a platinum/iridium anode coil. The electrode tip contains a molded silicone rubber seal. This molded tip provides a fluid seal which allows guide wire passage and reduces blood ingress. The lead also features nickel alloy conductors, polyurethane outer insulation, silicone inner insulation, and an IS-11 Bipolar (BI) lead connector.
The lead’s distal tip contains a maximum of 1.0 mg dexamethasone sodium phosphate. Upon exposure to body fluids, the steroid elutes from the lead tip. The steroid is known to suppress the inflammatory response that is believed to cause threshold rises typically associated with implanted pacing electrodes.
The Model 4194 lead can be positioned with the aid of a guide wire or with a stylet. If a stylet is used, use only the stylets packaged with the lead or in a stylet kit (downsized knob).
Note: To implant the Model 4194 lead in a cardiac vein, a compatible delivery system is required, such as a Medtronic delivery system. A compatible delivery system includes a guide catheter and a hemostasis/introducer valve which allows passage through or removal from an IS-1 connector. Contact your Medtronic representative for further information regarding compatible delivery systems.

2.1 Contents of package

The lead and accessories are supplied sterile. Each package contains the following items:
1 lead with stylet and anchoring sleeve
2 guide wire insertion tools
1 guide wire clip
1 guide wire steering handle
Extra stylets
Product documentation

2.2 Accessory descriptions

Dispose of all single-use accessories according to local environmental requirements.
Anchoring sleeve – An anchoring sleeve secures the lead to prevent it from moving and protects the lead insulation and conductors from damage caused by tight sutures.
Guide wire clip – A guide wire clip secures the excess guide wire and helps to protect and maintain the sterility of the guide wire.
Guide wire insertion tool – A guide wire insertion tool provides additional control when inserting a guide wire into the lead connector pin or the lead tip.
Guide wire steering handle – A guide wire steering handle is used only with guide wires 0.46 mm (0.018 in) or less in diameter. The steering handle provides additional control and steerability of the guide wire.
Stylet – A stylet provides additional stiffness and controlled flexibility for maneuvering the lead into position. Each stylet knob is labeled with the stylet diameter and corresponding lead length.
1
IS-1 BI refers to an International Connector Standard (ISO 5841–3) whereby pulse generators and leads so designated are assured of a basic mechanical fit.
3

3 Indications for use

The lead has application as part of a Medtronic biventricular pacing system.

4 Contraindications

Coronary vasculature – This lead is contraindicated for patients
with coronary venous vasculature that is inadequate for lead placement, as indicated by venogram.
Steroid use – The lead is contraindicated in patients for whom a single dose of 1.0 mg of dexamethasone sodium phosphate may be contraindicated.

5 Warnings and precautions

For single use only – Do not resterilize and reimplant an
explanted lead.
Necessary hospital equipment – Keep external defibrillation equipment nearby for immediate use during acute lead system testing, the implant procedure, or whenever arrhythmias are possible or intentionally induced during post-implant testing. Backup pacing should be readily available during implant. Use of the delivery system or leads may cause heart block.
Line-powered and battery-powered equipment – An implanted lead forms a direct current path to the myocardium. During lead implant and testing, use only battery-powered equipment or line-powered equipment specifically designed for this purpose to protect against fibrillation that may be caused by alternating currents. Line-powered equipment used in the vicinity of the patient must be properly grounded. Lead connector pins must be insulated from any leakage currents that may arise from line-powered equipment.
Diathermy – People with metal implants such as pacemakers, implantable cardioverter defibrillators (ICDs), and accompanying leads should not receive diathermy treatment. The interaction between the implant and diathermy can cause tissue damage, fibrillation, or damage to the device components, which could result in serious injury, loss of therapy, and/or the need to reprogram or replace the device.
Inspecting the sterile package – Carefully inspect the package prior to opening.
If the seal or package is damaged, contact your local Medtronic representative.
Do not use the product after its expiration date.
The lead has been sterilized with ethylene oxide prior to shipment. If the integrity of the sterile package has been compromised prior to the expiration date, resterilize the lead using ethylene oxide.
Ethylene oxide resterilization – If the sterile package seal is broken, resterilize the device using a validated ethylene oxide process. Avoid resterilization techniques that could damage the lead.
Refer to sterilizer instructions for operating instructions.
Use an acceptable method to determine sterilizer effectiveness, such as biological indicators.
Before resterilization, place the device in an ethylene oxide permeable package.
Do not exceed temperatures of 55 °C (131 °F).
Do not resterilize more than one time.
Allow proper aeration of ethylene oxide residues prior to implant.
Steroid use – It has not been determined whether the warnings, precautions, or complications usually associated with injectable dexamethasone sodium phosphate apply to the use of this highly localized, controlled-release device. For a list of potential adverse effects, refer to the Physicians’ Desk Reference.
Handling the steroid tip – Reducing the available amount of steroid may adversely affect low-threshold performance. Avoid reducing the amount of steroid available prior to lead implant.
Do not allow the electrode surface to come in contact with surface contaminants.
Do not wipe or immerse the electrode in fluid, except blood, at the time of implant.
Handling the lead – Leads should be handled with care at all times:
Do not withdraw the lead through a non-adjustable hemostasis/introducer valve. This action could cause distortion of the coil electrode and result in potential injury to the patient.
If a stylet is used for lead positioning, use only the stylets packaged with the lead or in a stylet kit (downsized knob). Other stylets may extend beyond the lead tip causing lead tip seal damage or patient injury.
If the lead is damaged, do not implant it. Return the lead to a Medtronic representative.
Protect the lead from materials that shed particles such as lint and dust. Lead insulators attract these particles.
Handle the lead with sterile surgical gloves that have been rinsed in sterile water or a comparable substance.
Do not severely bend, kink, or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pin.
Do not immerse leads in mineral oil, silicone oil, or any other liquid, except blood, at the time of implant.
Use an anchoring sleeve with all leads. Ensure that the anchoring sleeve is positioned close to the lead connector pin, to prevent inadvertent passage of the sleeve into the vein. If wiping the lead is necessary prior to insertion, ensure that the anchoring sleeve remains in position.
Do not force the guide catheter or leads if significant resistance is encountered. Use of guide catheters or leads may cause trauma to the heart.
4
Handling the guide wire – Handle the guide wire with care at all times.
Damage to a guide wire may prevent the guide wire from performing with accurate torque response and control and may cause vessel damage. For additional information about vessel damage and other potential adverse events, refer to the technical manual packaged with the appropriate guide wire.
If the distal end of the guide wire becomes severely kinked or twisted, it may be difficult to withdraw it back through the lead. Therefore, if there is an indication that the distal end of the guide wire has become damaged, or if there is significant resistance in guide wire passage, remove the lead and guide wire together as a unit. Remove the guide wire from the lead and insert a new guide wire into the lead. Do not use excessive force to retract the guide wire from the lead. Refer to the product documentation packaged with the guide wire for additional information.
Handling the stylet – Handle the stylet with care at all times.
To minimize the likelihood of trauma to the vein and to maintain lead flexibility while advancing the lead through the vein, keep the stylet withdrawn 1 to 2 cm or select a more flexible stylet.
Do not use excessive force or surgical instruments when inserting a stylet.
Avoid overbending, kinking, or blood contact on stylets.
Use a new stylet when blood or other fluids accumulate on the stylet. Accumulated fluids may cause damage to the lead or difficulty in passing the stylet through the lead.
Curving the distal end of the stylet prior to insertion into the lead will achieve a curvature at the distal end of the lead. Do not use a sharp object to impart a curve to the distal end of the stylet.
Concurrent devices – Output pulses, especially from unipolar devices, may adversely affect device sensing capabilities. If a patient requires a separate stimulation device, either permanent or temporary, allow enough space between the leads of the separate systems to avoid interference in the sensing capabilities of the devices. Previously implanted pulse generators and implantable cardioverter defibrillators should generally be explanted.
Chronic repositioning or removal – Chronic repositioning or removal of leads may be difficult because of fibrotic tissue development. Studies have not specifically evaluated the removal of left ventricular leads from the coronary venous vasculature. If a lead must be removed or repositioned, proceed with extreme caution. Return all removed leads to Medtronic.
Verify lead length on the IS-1 label on the connector to choose an appropriate stylet kit (downsized knob) length when repositioning. Always choose a stylet kit (downsized knob) 3 cm shorter than the lead length listed on the IS-1 connector label. For example, choose a stylet kit (downsized knob) with stylets 75 cm long for a lead 78 cm long.
Lead removal may result in avulsion of the endocardium, valve, or vein.
Lead junctions may separate, leaving the lead tip and bare wire in the heart or vein.
Chronic repositioning of a lead may adversely affect a steroid eluting lead’s low-threshold performance.
Cap abandoned leads to avoid transmitting electrical signals.
For leads that have been severed, seal the remaining lead end and suture the lead to adjacent tissue.
If a lead is removed and repositioned, inspect it carefully for insulator or conductor coil damage prior to repositioning.

6 Potential adverse events

The potential adverse events (listed in alphabetical order) related to the use of transvenous leads include, but are not limited to, the following conditions:
cardiac dissection
cardiac perforation
cardiac tamponade
coronary sinus dissection
death
endocarditis
erosion through the skin
extracardiac muscle or nerve stimulation
fibrillation or other arrhythmias
heart block
heart wall or vein wall rupture
hematoma/seroma
infection
myocardial irritability
myopotential sensing
pericardial effusion
pericardial rub
pneumothorax
rejection phenomena (local tissue reaction, fibrotic tissue formation, pulse generator migration)
threshold elevation
thrombosis
thrombotic or air embolism
valve damage (particularly in fragile hearts)
Additional potential adverse events related to the lead and the programmed parameters include, but are not limited to, the following:
Potential adverse event
Lead dislodgementaIntermittent or contin-
Lead dislodgementaIntermittent or contin-
Indicator of poten­tial adverse event
uous loss of capture or sensing
a
uous oversensing
Corrective action to be considered
Reposition the lead.
Reposition the lead.
5
Potential adverse event
Lead conductor frac­ture or insulation fail­ure
Threshold elevation or exit block
a
Transient loss of capture or sensing may occur following surgery until lead stabilization takes place. If stabilization does not occur, lead dislodgement may be suspected.
Indicator of poten­tial adverse event
Intermittent or contin­uous loss of capture or sensing
Loss of capture
a
a
Corrective action to be considered
Replace the lead.
Adjust the implanta­ble device output. Replace or reposition the lead.
Implant techniques that may damage the lead include, but are not limited to, the following:
Implant techniques that may damage the lead
Forcing the lead through the intro­ducer/delivery sys­tem
Use of too medial of an approach with venous introducer resulting in clavicle & first rib binding
Puncturing the perios­teum and/or tendon when using subcla­vian introducer approach
Advancing the lead through the veins without the stylet or guide wire fully inser­ted
Inserting the proximal end of the guide wire through the lead tip seal without using the guide wire insertion tool
Advancing a stylet tip beyond the distal end of the lead tip seal
Possible effects on the lead
Electrode, conductor coil, and/or insulation damage
Conductor coil frac­ture, insulation dam­age
Conductor coil frac­ture, insulation dam­age
Tip distortion and/or insulation perforation
Lead tip seal damage Replace the lead.
Lead tip seal damage Replace the lead.
Corrective action to be considered
Replace the lead.
Replace the lead.
Replace the lead.
Replace the lead.

7 Clinical trials

7.1 Attain Bipolar OTW Model 4194 left-heart lead study

A prospective, multi-center trial conducted at 22 investigative sites in the United States and Canada evaluated the safety and effectiveness of the Medtronic Attain Bipolar OTW Model 4194 left-heart lead in heart failure patients.
An investigational protocol pre-specified the performance criteria for both safety and effectiveness based on previous studies. For this study, patients who satisfied inclusion and exclusion criteria underwent a baseline evaluation to determine study eligibility and then underwent a cardiac resynchronization system implant attempt, which could include the Model 4194 left heart lead.
Key study inclusion criteria included patients with the following:
Patient provides written informed consent
Patients 18 years of age or older
Patient is diagnosed with New York Heart Association functional classification three (NYHA III) or New York Heart Association functional classification four (NYHA IV) despite optimal medical therapy which is defined as: – ACE inhibitor or Angiotensin Receptor Blocker (ARB), if
tolerated, for at least 1 month prior to implant
– Beta-blockers for at least 3 months preceding implant, if
tolerated, and stable for 1 month. Stable is defined as no upward titration of beta-blockers.
Demonstrated intrinsic QRS duration ≥ 130 ms (test documented within 6 months of baseline)
Left ventricular ejection fraction (EF) ≤ 35% (test documented within 12 months of baseline)
Patient is expected to remain available for follow-up visits
Patient is willing and able to comply with the protocol
Patient is indicated for ICD implantation for the treatment of life threatening ventricular arrhythmias (required only if a patient will receive an ICD)
Key study exclusion criteria included patients with the following:
Patients with a previous complete atrial based biventricular CRT system
Patients with a previous LV lead implanted or previous implant attempt within 30 days of implant or ongoing adverse events from previous unsuccessful attempt
Patients contraindicated for < 1.0 mg dexamethasone sodium phosphate
Patients with unstable angina or who have experienced an acute myocardial infarction (MI) within the past month
Patients who have had a coronary artery revascularization (CABG) or coronary angioplasty (PTCA) within the past 3 months
Patients on more than 2 inotropic infusions per week within 1 month of implant
Patients with contraindications for standard transvenous cardiac pacing (for example, mechanical right heart valves)
Patients with chronic (permanent) atrial arrhythmias
Post heart transplant patients (patients waiting for heart transplant are allowed in the study)
Patients enrolled in any concurrent drug and/or device study that may confound the results of this study
Patients with a terminal illness who are not expected to survive more than 6 months
6
Loading...
+ 14 hidden pages