Transvenous, bipolar, coronary sinus/cardiac vein pacing
lead
OTW 4193
Technical manual
Caution: Federal law (USA) restricts this device
to sale by or on the order of a physician.
The following list includes trademarks or registered trademarks of
Medtronic in the United States and possibly in other countries. All
other trademarks are the property of their respective owners.
Attain, InSync, Medtronic
Contents
Required physician training 5
Device description 5
Contents of package 6
Accessory descriptions 6
Indications for use 6
Contraindications 6
Warnings and precautions 7
Potential adverse events 10
Clinical trials 11
Attain OTW Model 4193 left-heart lead study 11
Objectives 13
InSync Registry study 18
Directions for use 18
Placing the right ventricular lead 19
Preparing the delivery system 19
Accessing the subclavian vein 19
Inserting the guide catheter assembly 20
Obtaining venograms 20
Inserting the lead into the delivery system 20
Selecting the lead placement method 21
Placing the lead using a stylet 22
Preparing the guide wire 24
Placing the lead using a guide wire 26
Fixating the lead 28
Taking electrical measurements 28
Removing the guide catheter from the lead 29
Anchoring the lead 29
Connecting the lead 30
Placing the device and leads into the pocket 30
Post-implant evaluation 31
In order to implant a Medtronic Attain OTW Model 4193 lead,
physicians are required to:
■
Thoroughly read this manual, and all associated device
technical manuals.
■
Provide a copy of the patient manual to the patient and discuss
it with him or her and any other interested parties.
■
Be trained on the following topics:
– Indications for use
– Device operation to ensure therapy delivery
– Measuring and managing bi-ventricular thresholds
– Assembly and use of LV lead implant tools
– Placement of the LV lead
– Patient management and system follow-up
Prior to implanting the system, Medtronic will certify that physicians
received training.
Device description
The Medtronic Attain OTW Model 4193 steroid eluting, transvenous,
unipolar, left ventricular, over the wire, cardiac vein pacing lead is
designed for pacing and sensing via a cardiac vein.
The lead features a tapered annular platinum alloy electrode with a
molded silicone rubber tip. This molded tip provides a fluid seal which
allows guide wire passage and reduces blood ingress. The lead also
features nickel alloy conductors, polyurethane insulation, and an
1
Unipolar (UNI) lead connector.
IS-1
The distal tip of the lead contains a maximum of 1.0 mg
dexamethasone sodium phosphate. Upon exposure to body fluids,
the steroid elutes from the lead tip. The steroid suppresses the
inflammatory response that is believed to cause threshold rises
typically associated with implanted pacing electrodes.
The Model 4193 lead can be positioned with the aid of a guide wire
or with a stylet. If a stylet is used, use only the stylets packaged with
the lead or in a stylet kit (downsized knob).
Note: To implant the Model 4193 in a cardiac vein, a compatible
delivery system is required. A compatible delivery system includes a
guide catheter and a hemostasis/introducer valve which allows
passage through or removal from an IS-1 connector. Contact your
Medtronic representative for further information regarding compatible
delivery systems.
1
IS-1 UNI refers to an International Connector Standard (ISO 5841-3) whereby
pulse generators and leads so designated are assured of a basic
mechanical fit.
4193 Technical ManualEnglish 5
Contents of package
The lead and accessories are supplied sterile. Each package
contains the following items:
■
One lead with stylet and anchoring sleeve
■
Two guide wire insertion tools
■
One guide wire clip
■
One guide wire steering handle
■
Extra stylets
■
Product literature
Accessory descriptions
Stylet – A stylet provides additional stiffness and controlled flexibility
for maneuvering the lead into position. Each stylet knob is labeled
with the stylet diameter and length.
Guide wire insertion tool – A guide wire insertion tool provides
additional control when inserting a guide wire into the lead connector
pin or the lead tip.
Anchoring sleeve – An anchoring sleeve secures the lead from
moving and protects the lead insulation and conductors from damage
caused by tight ligatures.
Guide wire steering handle – A guide wire steering handle is used
only with guide wires 0.018” in diameter or less. The steering handle
provides additional control and steerability of the guide wire.
Guide wire clip – A guide wire clip secures the excess guide wire
and helps to protect and maintain the sterility of the guide wire.
Indications for use
The lead has application as part of a Medtronic biventricular
pacing system.
Contraindications
Coronary vasculature – This lead is contraindicated for patients
with coronary venous vasculature that is inadequate for lead
placement, as indicated by venogram.
Steroid use – Do not use steroid eluting leads in patients for whom
a single dose of 1.0 mg of dexamethasone sodium phosphate may
be contraindicated.
6 English4193 Technical Manual
Warnings and precautions
For single use only – Do not resterilize and reimplant an
explanted lead.
Inspecting the sterile package – Carefully inspect the package
prior to opening.
■
If the seal or package is damaged, contact your local
Medtronic representative.
■
Do not use the product after its expiration date.
■
The lead has been sterilized with ethylene oxide prior to
shipment. If the integrity of the sterile package has been
compromised prior to the expiration date, resterilize the lead
using ethylene oxide.
Ethylene oxide resterilization – If the sterile package seal is
broken, resterilize the device using a validated ethylene oxide
process. Avoid resterilization techniques that could damage the lead.
■
Refer to sterilizer instructions for operating instructions.
■
Use an acceptable method for determining sterilizer
effectiveness, such as biological indicators.
■
Before resterilization, place the device in an ethylene oxide
permeable package.
■
Do not exceed temperatures of 55 °C (131 °F).
■
Do not resterilize more than one time.
■
Allow proper aeration of ethylene oxide residues prior
to implant.
Steroid use – It has not been determined whether the warnings,
precautions, or complications usually associated with injectable
dexamethasone sodium phosphate apply to the use of this highly
localized, controlled-release device. For a list of potential adverse
effects, refer to the Physician’s Desk Reference.
Handling the steroid tip – Reducing the available amount of steroid
may adversely affect low-threshold performance. Avoid reducing the
amount of steroid available prior to lead implant.
■
Do not allow the electrode surface to come in contact with
surface contaminants.
■
Do not wipe or immerse the electrode in fluid, except blood at
the time of implant.
Handling the lead – Leads should be handled with care at all times.
■
If a stylet is used for lead positioning, use only the stylets
packaged with the lead or in a stylet kit (downsized knob).
Other stylets may extend beyond the lead tip causing lead tip
seal damage and/or patient injury.
■
If the lead is damaged, do not implant it. Return the lead to your
Medtronic representative.
■
Protect the lead from materials that shed particles such as lint
and dust. Lead insulators attract these particles.
4193 Technical ManualEnglish 7
■
Handle the lead with sterile surgical gloves that have been
rinsed in sterile water or a comparable substance.
■
Do not severely bend, kink, or stretch the lead.
■
Do not use surgical instruments to grasp the lead or
connector pin.
■
Do not immerse leads in mineral oil, silicone oil, or any other
liquid, except blood at the time of implant.
■
Use an anchoring sleeve with all leads. Ensure that the
anchoring sleeve is positioned close to the lead’s connector
pin, to prevent inadvertent passage of the sleeve into the vein.
If wiping the lead is necessary prior to insertion, ensure that the
anchoring sleeve remains in position.
■
Do not force the guide catheter and/or leads if significant
resistance is encountered. Use of guide catheters and/or leads
may cause trauma to the heart.
Handling the guide wire – Use care when handling guide wires.
■
Damage to a guide wire may prevent the guide wire from
performing with accurate torque response and control and may
cause vessel damage. For additional information about vessel
damage and other potential adverse events, refer to the
technical manual packaged with the appropriate product.
■
If the distal end of the guide wire becomes severely kinked or
twisted, it may be difficult to withdraw it back through the lead.
Therefore, if there is an indication that the distal end of the
guide wire has become damaged, or if there is a lot of
resistance in guide wire passage, remove the lead and guide
wire. Remove the guide wire from the lead and re-insert a new
guide wire into the lead. Do not use excessive force to retract
the guide wire from the lead.
Refer to the product literature packaged with the guide wire for
additional information.
Handling the stylets – Use care when handling stylets.
■
To minimize the likelihood of trauma to the vein and to maintain
lead flexibility while advancing the lead through the vein keep
the stylet withdrawn 1-2 cm or select a more flexible stylet.
■
Do not use excessive force or surgical instruments when
inserting a stylet.
■
Avoid overbending, kinking, or blood contact on stylets.
■
Use a new stylet when blood or other fluids accumulate on the
stylet. Accumulated fluids may cause damage to the lead or
difficulty in passing the stylet through the lead.
■
Curving the distal end of the stylet prior to insertion into the lead
will achieve a curvature at the distal end of the lead. Do not use
a sharp object to impart a curve to the distal end of the stylet.
8 English4193 Technical Manual
Necessary hospital equipment – Keep external defibrillation
equipment nearby for immediate use during the acute lead system
testing, implant procedure, or whenever arrhythmias are possible or
intentionally induced during post-implant testing. Back-up pacing
should be readily available during implant. Use of the delivery system
and/or leads may cause heart block.
Line-powered equipment – An implanted lead forms a direct
current path to the myocardium. During lead implant and testing, use
only battery-powered equipment or line-powered equipment
specifically designed for this purpose, to protect against fibrillation
that may be caused by alternating currents. Line-powered equipment
used in the vicinity of the patient must be properly grounded. Lead
connector pins must be insulated from any leakage currents that may
arise from line-powered equipment.
Concurrent devices – Output pulses, especially from unipolar
devices, may adversely affect device sensing capabilities. If a patient
requires a separate stimulation device, either permanent or
temporary, allow enough space between the leads of the separate
systems to avoid interference in the sensing capabilities of the
devices. Previously implanted pulse generators and implantable
cardioverter defibrillators should generally be explanted.
Diathermy – People with metal implants such as pacemakers,
implantable cardioverter defibrillators (ICDs), and accompanying
leads should not receive diathermy treatment. The interaction
between the implant and diathermy can cause tissue damage,
fibrillation, or damage to the device components, which could result
in serious injury, loss of therapy, and/or the need to reprogram or
replace the device.
Chronic repositioning or removal – Chronic repositioning or
removal of leads may be difficult because of fibrotic tissue
development. Studies have not specifically evaluated the removal of
left ventricular leads from the coronary venous vasculature. If a lead
must be removed or repositioned, proceed with extreme caution.
Return all removed leads to Medtronic.
■
Verify lead length on the IS-1 label on the connector to choose
an appropriate stylet kit (downsized knob) length when
repositioning. Always choose a stylet kit (downsized knob)
3 cm shorter than the lead length. For example, choose a stylet
kit (downsized knob) with stylets 75 cm long for a lead 78 cm
long.
■
Lead removal may result in avulsion of the endocardium, valve,
or vein.
■
Lead junctions may separate, leaving the lead tip and bare wire
in the heart or vein.
■
Chronic repositioning may adversely affect the low-threshold
performance of a steroid eluting lead.
■
Cap abandoned leads to avoid transmitting electrical signals.
■
For leads that have been severed, seal the remaining lead end
and suture the lead to adjacent tissue.
4193 Technical ManualEnglish 9
■
If a lead is removed and repositioned, inspect it carefully for
insulator or conductor coil damage prior to repositioning.
Potential adverse events
The potential adverse events (listed in alphabetical order) related to
the use of transvenous leads include, but are not limited to, the
following patient-related conditions:
Additional potential adverse events related to the lead and the
programmed parameters include, but are not limited to, the following:
Potential adverse
event
Lead dislodgment
Lead dislodgment
Lead conductor
fracture or insulation
failure
Threshold elevation or
exit block
a
Transient loss of capture or sensing may occur following surgery until lead
stabilization takes place. If stabilization does not occur, lead dislodgment may
be suspected.
Indicator of potential
adverse event
a
Intermittent or continuous
loss of capture or sensing
a
Intermittent or continuous
oversensing
Intermittent or continuous
loss of capture or sensing
Loss of capture
Corrective action
to be considered
Reposition the lead.
a
Reposition the lead.
Replace the lead.
a
a
Adjust the implantable
device output. Replace
or reposition the lead.
10 English4193 Technical Manual
Implant techniques that may damage the lead include, but are not
limited to, the following:
Implant techniques that may
damage the lead
Forcing the lead through
the introducer/delivery system
Use of too medial of an
approach with venous
introducer resulting in clavicle &
first rib binding
Puncturing the periosteum and/
or tendon when using
subclavian introducer approach
Advancing the lead through the
veins without the stylet or guide
wire fully inserted
Inserting the proximal end of the
guide wire through the lead tip
seal without using the guide
wire insertion tool
Advancing a stylet tip beyond
the distal end of the lead tip seal
A prospective, multi-center trial conducted at 28 investigative sites in
the United States evaluated the safety and effectiveness of the
Medtronic InSync III cardiac resynchronization device and the
Attain OTW Model 4193 left-heart lead in heart failure patients.
An investigational protocol pre-specified the performance criteria for
both safety and effectiveness based on previous studies. For this
study, patients who satisfied inclusion and exclusion criteria
underwent a baseline evaluation to determine study eligibility and
then underwent an implant attempt of the InSync III cardiac
resynchronization system including an Attain OTW Model 4193 lead.
Key study inclusion criteria included patients with the following:
■
Symptomatic congestive heart failure (NYHA classification III
or IV).
■
QRS duration of ≥ 130 ms.
■
Left ventricular end diastolic diameter of ≥ 55 mm.
■
Left ventricular ejection fraction ≤ 35%.
■
A stable pharmacological medical regimen prior to implant of
the cardiac resynchronization system. This included ACE-I or
substitute for at least one month and a beta blocker for at least
three months if tolerated.
4193 Technical ManualEnglish 11
Loading...
+ 25 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.