Medtronic 407658 Technical Manual

CapSureFix Novus MRI™ SureScan™ 4076
Steroid eluting, bipolar, implantable, screw-in, ventricular/atrial, transvenous lead
Technical Manual Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
The following list includes trademarks or registered trademarks of Medtronic in the United States and possibly in other countries. All other trademarks are the property of their respective owners.
CapSureFix, CapSureFix Novus MRI, Medtronic, SureScan

Contents

9 Directions for use 9 10 Specifications 15 11 Medtronic warranty 16 12 Service 16

1 Description

The Medtronic CapSureFix Novus MRI SureScan 4076 steroid eluting, bipolar, implantable, screw-in, ventricular/atrial, transvenous lead is designed for pacing and sensing applications in either the atrium or ventricle. The platinum alloy tip and ring electrodes feature a high-active surface area of titanium nitride microstructure. This electrode configuration contributes to low polarization. The lead has been tested for use in the Magnetic Resonance Imaging (MRI) environment. The following lead lengths are MR conditional: 35 cm to 85 cm. Other lead lengths are not MR conditional.
The helical tip electrode on the lead can be actively fixed in the endocardium. The helix electrode can be extended or retracted by rotating the lead connector pin with a fixation tool. An active fixation lead is particularly beneficial for patients who have smooth or hypertrophic hearts where lead dislodgment may be a potential problem. The lead also has a second, larger electrode proximal to the helical tip electrode and an IS-11 Bipolar (BI) connector with 1 terminal pin. The lead features MP35N nickel alloy conductors, an outer insulation of 55D polyurethane, and an inner insulation of silicone rubber.
The distal tip of the lead contains a target nominal dosage of 680 µg of dexamethasone acetate. Upon exposure to body fluids, the steroid elutes from the lead tip. The steroid is known to suppress the inflammatory response that is believed to cause threshold rises typically associated with implanted pacing electrodes.

1.1 Medtronic SureScan system

A complete SureScan pacing system is required for use in the MR environment. A complete SureScan system includes a Medtronic SureScan device with the appropriate number of Medtronic SureScan leads. Any other combination may
result in a hazard to the patient during an MRI scan.
The Model 4076 lead is part of the Medtronic SureScan system. The SureScan system includes a Medtronic SureScan device connected to Medtronic SureScan leads. Labeling for SureScan system components displays the SureScan logo and the MR Conditional symbol. To verify that components are part of a SureScan system, visit http://www.mrisurescan.com.
SureScan logo
MR Conditional symbol. The Medtronic SureScan sys­tem is MR Conditional and is designed to allow implan­ted patients to undergo an MRI scan under the speci­fied MRI conditions for use.
The MRI SureScan feature permits a mode of operation that allows a patient with a SureScan device to be safely scanned by an MRI machine while the device continues to provide appropriate pacing. When programmed to On, MRI SureScan operation disables arrhythmia detection, magnet mode, and all user-defined diagnostics. Before performing an MRI scan,
refer to the SureScan MRI technical manual for important information about procedures and MRI-specific warnings and precautions.

1.2 Contents of package

The lead and accessories are supplied sterile. Each package contains:
1 lead with 1 radiopaque anchoring sleeve, stylet, and stylet guide
1 vein lifter
2 fixation tools
extra stylets
product documentation

1.3 Accessory descriptions

Dispose of all single-use accessories according to local environmental requirements.
Anchoring sleeve – An anchoring sleeve secures the lead to prevent it from moving and protects the lead insulation and conductors from damage caused by tight sutures.
Fixation tool – A fixation tool facilitates connector pin rotation.
Stylet – A stylet provides additional stiffness and controlled
flexibility for maneuvering the lead into position. Each stylet knob is labeled with the stylet diameter and corresponding lead length.
1
IS-1 refers to an International Connector Standard (ISO 5841-3) whereby pulse generators and leads so designated are assured of a basic mechanical fit.
3
Stylet guide – A stylet guide facilitates stylet insertion into the lead.
Vein lifter – A vein lifter facilitates lead insertion into a vein.

2 Drug component description

The active ingredient in the Model 4076 lead is dexamethasone acetate. Dexamethasone acetate is 9-Fluoro-11β, 17,21-trihydroxy-16α-methylpregna-1,4-diene-3,20-dione 21-acetate. Dexamethasone acetate has a molecular formula of C24H31FO6 and a molecular weight of 434.50. The MCRD (Monolithic controlled release device) excipient is silicone. See Figure 1 for the structural formula.
Figure 1.
The target dosage of dexamethasone acetate is 680 µg per lead.

3 Indications

The Model 4076 lead is designed to be used with a compatible implantable pulse generator as part of a chronic cardiac pacing system. The lead has application where implantable atrial or ventricular, single chamber or dual chamber pacing systems are indicated.

4 Contraindications

Before performing an MRI scan, refer to the SureScan MRI technical manual for MRI-specific contraindications.
Use of a ventricular transvenous lead is contraindicated in patients with tricuspid valvular disease.
Use of a ventricular transvenous lead is contraindicated in patients with mechanical tricuspid heart valves.
Use of a steroid eluting lead is contraindicated in patients for whom a single dose of 1.0 mg of dexamethasone acetate may be contraindicated.

5 Warnings and precautions

A complete SureScan system is required for use in the MRI environment. Before performing an MRI scan, refer to the
SureScan technical manual for MRI-specific warnings and precautions.
Note: Medical procedure warnings and precautions that pertain
to the Medtronic implanted system are provided in the manual that is packaged with the device or on the Medtronic Manual Library website (www.medtronic.com/manuals).
Magnetic resonance imaging (MRI) – An MRI is a type of medical imaging that uses magnetic fields to create an internal view of the body. If certain criteria are met and the warnings and precautions provided by Medtronic are followed, patients with an MR Conditional device and lead system are able to undergo an MRI scan; for details, refer to the SureScan MRI technical manual that Medtronic provides for an MR Conditional device.
Line-powered and battery-powered equipment – An implanted lead forms a direct current path to the myocardium. During lead implant and testing, use only battery-powered equipment or line-powered equipment specifically designed for this purpose to protect against fibrillation that may be caused by alternating currents. Line-powered equipment used in the vicinity of the patient must be properly grounded. Lead connector pins must be insulated from any leakage currents that may arise from line-powered equipment.
Diathermy treatment (including therapeutic ultrasound) –
Diathermy is a treatment that involves the therapeutic heating of body tissues. Diathermy treatments include high frequency, short wave, microwave, and therapeutic ultrasound. Except for therapeutic ultrasound, do not use diathermy treatments on cardiac device patients. Diathermy treatments may result in serious injury or damage to an implanted device and lead system. Therapeutic ultrasound (including physiotherapy, high intensity therapeutic ultrasound, and high intensity focused ultrasound), is the use of ultrasound at higher energies than diagnostic ultrasound to bring heat or agitation into the body. Therapeutic ultrasound is acceptable if treatment is performed with a minimum separation distance of 15 cm (6 in) between the applicator and the implanted device and lead system, as long as the ultrasonic beam is pointing away from the device and lead system.
Single use – The lead and accessories are for single use only.
Sterilization – Medtronic has sterilized the package contents
with ethylene oxide before shipment. This lead is for single use only and is not intended to be resterilized.
Necessary hospital equipment – Keep external defibrillation equipment nearby for immediate use during acute lead system testing, the implant procedure, or whenever arrhythmias are possible or intentionally induced during post-implant testing.
Inspecting the sterile package – Inspect the sterile package with care before opening it.
If the seal or package is damaged, contact a Medtronic representative.
4
Store at 25 °C (77 °F). Excursions from this storage temperature are permitted in the range of 15 to 30 °C (59 to 86 °F). (See USP Controlled Room Temperature.) According to USP excursion conditions, transient spikes up to 40 °C (104 °F) are permitted as long as they do not exceed 24 hours.
Do not use the product after its expiration date.
Concurrent devices – Output pulses, especially from unipolar devices, may adversely affect device sensing capabilities. If a patient requires a separate stimulation device, either permanent or temporary, allow enough space between the leads of the separate systems to avoid interference in the sensing capabilities of the devices. Previously implanted pulse generators and implantable cardioverter defibrillators should generally be explanted.
Steroid use – It has not been determined whether the warnings, precautions, or complications usually associated with injectable dexamethasone acetate apply to the use of this highly localized, controlled-release lead. For a listing of potential adverse effects, refer to the dexamethasone acetate manufacturer prescribing information or the Physicians’ Desk Reference.
Steroid elution and exit block – Although the addition of steroid to passive fixation leads has been shown to reduce pacing thresholds in patients with a history of exit block, the frequency of redevelopment of exit block was not statistically different between the steroid eluting and the non-steroid eluting, active fixation, screw-in leads in the Medtronic Model 4068 clinical trial.
Handling the steroid tip – Avoid reducing the amount of steroid available before implanting the lead. Reducing the available amount of steroid may adversely affect low-threshold performance.
Do not allow the electrode surface to come in contact with surface contaminants.
Do not wipe or immerse the electrode in fluid, except blood, at the time of implant.
Handling the stylet – Handle the stylet with care at all times.
Curve the stylet before inserting it into the lead to achieve a curvature at the lead’s distal end. Do not use a sharp object to impart a curve to the distal end of the stylet.
Do not use excessive force or surgical instruments when inserting the stylet into the lead.
Avoid overbending or kinking the stylet.
Use a new stylet when blood or other fluids accumulate on the stylet. Accumulated blood or other fluids may damage the lead or cause difficulty in passing the stylet into the lead.
Before inserting the lead – Use an anchoring sleeve with all leads. Ensure that the anchoring sleeve is positioned close to the lead connector pin. This will prevent inadvertent passage of the sleeve into the vein. If wiping the lead is necessary prior to insertion, ensure that the anchoring sleeve remains in position.
Vessel and tissue damage – Use care when positioning the lead. Avoid known infarcted or thin ventricular wall areas to minimize the occurrence of perforation and dissection.
Handling a lead – Handle the lead with care at all times.
Do not implant the lead if it is damaged. Return the lead to a Medtronic representative.
Protect the lead from materials that shed small particles such as lint and dust. Lead insulators attract these particles.
Handle the lead with sterile surgical gloves that have been rinsed in sterile water or a comparable substance.
Do not severely bend, kink, or stretch the lead.
Do not immerse the lead in mineral oil, silicone oil, or any other liquid, except blood, at the time of implant.
Do not use surgical instruments to grasp the lead.
Do not force the lead if resistance is encountered during lead passage.
Chronic lead removal and the SureScan pacing system –
When implanting a SureScan pacing system, consider the risks associated with removing previously implanted leads before doing so. Abandoned leads or previously implanted non-SureScan labeled leads compromise the ability to safely scan the SureScan pacing system during MRI scans.
Chronic repositioning or removal of a screw-in lead –
Proceed with extreme caution if a lead must be removed or repositioned. Chronic repositioning or removal of screw-in transvenous leads may not be possible because of blood or fibrotic tissue development into the helix mechanism on the lead. In most clinical situations, it is preferable to abandon unused leads in place. Return all removed leads, unused leads, or lead sections to Medtronic for analysis.
Note: If a helix electrode does not disengage from the endocardium by rotating the connector pin, rotating the lead body counterclockwise may withdraw the helix electrode and decrease the possibility of damage to cardiovascular structures during removal.
Lead removal may result in avulsion of the endocardium, valve, or vein.
Lead junctions may separate, leaving the lead tip and bare wire in the heart or vein.
Chronic repositioning of a lead may adversely affect the low-threshold performance of a steroid lead.
An abandoned lead should be capped so that the lead does not transmit electrical signals.
Severed leads should have the remaining lead end sealed and the lead body sutured to adjacent tissue.
Connector compatibility – Although the lead conforms to the IS-1 International Connector Standard, do not attempt to use the lead with any device other than a commercially available implantable pacing system with which it has been tested and demonstrated to be safe and effective. The potential adverse consequences of using such a combination may include, but are not limited to, undersensing cardiac activity and failure to deliver necessary therapy.
External defibrillation and cardioversion – External defibrillation and cardioversion are therapies that deliver an electrical shock to the heart to convert an abnormal heart rhythm to a normal rhythm.
5
Medtronic cardiac devices are designed to withstand exposure to external defibrillation and cardioversion. While damage to an implanted system from an external shock is rare, the probability increases with increased energy levels. These procedures may also temporarily or permanently elevate pacing thresholds or temporarily or permanently damage the myocardium. If external defibrillation or cardioversion are required, consider the following precautions:
Use the lowest clinically appropriate energy.
Position the patches or paddles no closer than 15 cm (6 in) to the device.
Position the patches or paddles perpendicular to the device and lead system.
If an external defibrillation or cardioversion is delivered within 15 cm (6 in) of the device, use a Medtronic programmer to evaluate the device and lead system.

6 Drug information

6.1 Mechanism of action

Steroid suppresses the inflammatory response that is believed to cause threshold rises typically associated with implanted pacing electrodes. Dexamethasone acetate is a synthetic steroid of the glucocorticoid family. Glucocorticoids have potent anti-inflammatory actions via direct and indirect effects on major inflammatory cells. Glucocorticosteroids bind to a cytoplasmic glucocorticoid receptor as well as a membrane-bound receptor. Binding to the cytoplasmic receptor leads to receptor activation and translocation to the nucleus. The receptor interacts with specific DNA sequences within the regulatory regions of affected genes. Thus, glucocorticoids inhibit the production of multiple cell factors that are critical in generating the inflammatory response.

6.2 Pharmacokinetics and metabolism

Pharmacokinetics – The pharmacokinetics (local drug levels
and systemic levels) of dexamethasone acetate and its metabolites following implant were not evaluated in human clinical trials. When delivered intra-muscularly, the lipid-soluble dexamethasone acetate is slowly absorbed throughout the tissue.
Metabolism – The conversion of dexamethasone acetate to dexamethasone occurs within hours. The dexamethasone alcohol (dexamethasone) is the active glucocorticoid used in this Medtronic lead. Steroid is applied via MCRD (Monolithic controlled release device) and eluted to the tissue interface where it will be used. The form of the steroid, whether it is a prodrug or the pharmacologically active dexamethasone, is irrelevant, as the steroid is directly present at the injury site to treat the inflammation. Dexamethasone acetate is hydrolyzed into dexamethasone, which is readily absorbed by the surrounding tissue and body fluids. Glucocorticoids, when given systemically, are eliminated primarily by renal excretion of inactive metabolites.

6.3 Mutagenesis, carcinogenicity, and reproductive toxicity

Mutagenesis, carcinogenicity, and reproductive toxicity –
The mutagenesis, carcinogenicity, and reproductive toxicity of the
CapSureFix Novus 4076 lead have not been evaluated. However, the mutagenesis, carcinogenicity, and reproductive toxicity of dexamethasone acetate have been evaluated previously.
Carcinogenesis, mutagenesis, and impairment of fertility –
No adequate studies have been conducted in animals to determine whether corticosteroids have a potential for carcinogenesis (tumor initiation or promotion). Dexamethasone was genotoxic in assays for clastogenicity (including sister chromatid exchange in human lymphocytes) but not in an assay for mutagenicity in salmonella (Ames test). Adrenocorticoids have been reported to increase or decrease the number and mobility of spermatozoa in some patients.

6.4 Pregnancy

Pregnancy – Dexamethasone acetate has been shown to be
teratogenic in many species when given in doses equivalent to the human dose. There are no adequate and well-controlled studies in pregnant women. Dexamethasone acetate should be used during pregnancy only if the potential benefit justifies the potential risk to the fetus. Studies in mice, rats, and rabbits have shown that adrenocorticoids increase the incidence of cleft palate, placental insufficiency, and spontaneous abortions, and can decrease the intrauterine growth rate.
Nursing mothers – Systemically administered corticosteroids appear in human milk and could suppress growth, interfere with endogenous corticosteroid production, or cause other untoward effects in nursing infants. Because of the potential for serious adverse reactions in nursing infants from corticosteroids, a decision should be made whether to discontinue nursing or to use a non-steroidal lead, taking into account the importance of the lead and the drug to the mother.

6.5 Lactation

Corticosteroids are secreted into human milk, and there is a potential for serious adverse reactions. A decision should be made whether to nurse or to discontinue the drug, taking into account the importance of the drug to the mother. These potential risks of corticosteroids should also be considered along with any other steroidal therapy being received by the patient.

7 Adverse events

The potential adverse events related to the use of transvenous leads include, but are not limited to, the following patient related conditions that can occur when the lead is being inserted and/or repositioned:
cardiac perforation
cardiac tamponade
fibrillation and other arrhythmias
heart wall rupture
infection
muscle or nerve stimulation
myocardial irritability
pericardial rub
pericarditis
6
Loading...
+ 12 hidden pages