Medtronic 4076110 Technical Manual

Page 1
CAPSUREFIX™ NOVUS 4076
Steroid eluting, bipolar, implantable, screw-in, ventricular/atrial, transvenous lead
Technical Manual
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
Page 2
The following list includes trademarks or registered trademarks of Medtronic in the United States and possibly in other countries. All other trademarks are the property of their respective owners.
CapSureFix, Medtronic
Page 3

Contents

9 Directions for use 8 10 Specifications 14 11 Medtronic warranty 14 12 Service 14
extra stylets
product documentation

1.2 Accessory descriptions

Dispose of all single-use accessories according to local environmental requirements.
Anchoring sleeve – An anchoring sleeve secures the lead to prevent it from moving and protects the lead insulation and conductors from damage caused by tight sutures.
Fixation tool – A fixation tool facilitates connector pin rotation.
Stylet – A stylet provides additional stiffness and controlled
flexibility for maneuvering the lead into position. Each stylet knob is labeled with the stylet diameter and corresponding lead length.
Stylet guide – A stylet guide facilitates stylet insertion into the lead.
Vein lifter – A vein lifter facilitates lead insertion into a vein.

1 Description

The Medtronic CapSureFix Novus 4076 steroid eluting, bipolar, implantable, screw-in, ventricular/atrial, transvenous lead is designed for pacing and sensing applications in either the atrium or ventricle. The platinum alloy tip and ring electrodes feature a high-active surface area of titanium nitride microstructure. This electrode configuration contributes to low polarization.
The helical tip electrode on the lead can be actively fixed in the endocardium. The helix electrode can be extended or retracted by rotating the lead connector pin with a fixation tool. An active fixation lead is particularly beneficial for patients who have smooth or hypertrophic hearts where lead dislodgment may be a potential problem. The lead also has a second, larger electrode proximal to the helical tip electrode and an IS-11 Bipolar (BI) connector with 1 terminal pin. The lead features MP35N nickel alloy conductors, an outer insulation of 55D polyurethane, and an inner insulation of silicone rubber.
The distal tip of the lead contains a target nominal dosage of 680 µg of dexamethasone acetate. Upon exposure to body fluids, the steroid elutes from the lead tip. The steroid is known to suppress the inflammatory response that is believed to cause threshold rises typically associated with implanted pacing electrodes.

1.1 Contents of package

The lead and accessories are supplied sterile. Each package contains:
1 lead with 1 radiopaque anchoring sleeve, stylet, and stylet guide
1 vein lifter
2 fixation tools

2 Drug component description

The active ingredient in the Model 4076 lead is dexamethasone acetate. Dexamethasone acetate is 9-Fluoro-11β, 17,21-trihydroxy-16α-methylpregna-1,4-diene-3,20-dione 21-acetate. Dexamethasone acetate has a molecular formula of C24H31FO6 and a molecular weight of 434.50. The MCRD (Monolithic controlled release device) excipient is silicone. See Figure 1 for the structural formula.
Figure 1.
The target dosage of dexamethasone acetate is 680 µg per lead.

3 Indications

The Model 4076 lead is designed to be used with a compatible implantable pulse generator as part of a chronic cardiac pacing system. The lead has application where implantable atrial or ventricular, single chamber or dual chamber pacing systems are indicated.
1
IS-1 refers to an International Connector Standard (ISO 5841-3) whereby pulse generators and leads so designated are assured of a basic mechanical fit.
3
Page 4

4 Contraindications

Use of a ventricular transvenous lead is contraindicated in patients with tricuspid valvular disease.
Use of a ventricular transvenous lead is contraindicated in patients with mechanical tricuspid heart valves.
Use of a steroid eluting lead is contraindicated in patients for whom a single dose of 1.0 mg of dexamethasone acetate may be contraindicated.

5 Warnings and precautions

Note: Medical procedure warnings and precautions that pertain to
the Medtronic implanted system are provided in the manual that is packaged with the device or on the Medtronic Manual Library website (www.medtronic.com/manuals).
Magnetic resonance imaging (MRI) – An MRI is a type of medical imaging that uses magnetic fields to create an internal view of the body. Do not conduct MRI scans on patients who have this device or lead implanted. MRI scans may result in serious injury, induction of tachyarrhythmias, or implanted system malfunction or damage.
Line-powered and battery-powered equipment – An implanted lead forms a direct current path to the myocardium. During lead implant and testing, use only battery-powered equipment or line-powered equipment specifically designed for this purpose to protect against fibrillation that may be caused by alternating currents. Line-powered equipment used in the vicinity of the patient must be properly grounded. Lead connector pins must be insulated from any leakage currents that may arise from line-powered equipment.
Diathermy treatment (including therapeutic ultrasound) –
Diathermy is a treatment that involves the therapeutic heating of body tissues. Diathermy treatments include high frequency, short wave, microwave, and therapeutic ultrasound. Except for therapeutic ultrasound, do not use diathermy treatments on cardiac device patients. Diathermy treatments may result in serious injury or damage to an implanted device and lead system. Therapeutic ultrasound (including physiotherapy, high intensity therapeutic ultrasound, and high intensity focused ultrasound), is the use of ultrasound at higher energies than diagnostic ultrasound to bring heat or agitation into the body. Therapeutic ultrasound is acceptable if treatment is performed with a minimum separation distance of 15 cm (6 in) between the applicator and the implanted device and lead system, as long as the ultrasonic beam is pointing away from the device and lead system.
Single use – The lead and accessories are for single use only.
Sterilization – Medtronic has sterilized the package contents
with ethylene oxide before shipment. This lead is for single use only and is not intended to be resterilized.
Necessary hospital equipment – Keep external defibrillation equipment nearby for immediate use during acute lead system testing, the implant procedure, or whenever arrhythmias are possible or intentionally induced during post-implant testing.
Inspecting the sterile package – Inspect the sterile package with care before opening it.
If the seal or package is damaged, contact a Medtronic representative.
Store at 25 °C (77 °F). Excursions from this storage temperature are permitted in the range of 15 to 30 °C (59 to 86 °F). (See USP Controlled Room Temperature.) According to USP excursion conditions, transient spikes up to 40 °C (104 °F) are permitted as long as they do not exceed 24 hours.
Do not use the product after its expiration date.
Concurrent devices – Output pulses, especially from unipolar devices, may adversely affect device sensing capabilities. If a patient requires a separate stimulation device, either permanent or temporary, allow enough space between the leads of the separate systems to avoid interference in the sensing capabilities of the devices. Previously implanted pulse generators and implantable cardioverter defibrillators should generally be explanted.
Steroid use – It has not been determined whether the warnings, precautions, or complications usually associated with injectable dexamethasone acetate apply to the use of this highly localized, controlled-release lead. For a listing of potential adverse effects, refer to the dexamethasone acetate manufacturer prescribing information or the Physicians’ Desk Reference.
Steroid elution and exit block – Although the addition of steroid to passive fixation leads has been shown to reduce pacing thresholds in patients with a history of exit block, the frequency of redevelopment of exit block was not statistically different between the steroid eluting and the non-steroid eluting, active fixation, screw-in leads in the Medtronic Model 4068 clinical trial.
Handling the steroid tip – Avoid reducing the amount of steroid available before implanting the lead. Reducing the available amount of steroid may adversely affect low-threshold performance.
Do not allow the electrode surface to come in contact with surface contaminants.
Do not wipe or immerse the electrode in fluid, except blood, at the time of implant.
Handling the stylet – Handle the stylet with care at all times.
Curve the stylet before inserting it into the lead to achieve a curvature at the lead’s distal end. Do not use a sharp object to impart a curve to the distal end of the stylet.
Do not use excessive force or surgical instruments when inserting the stylet into the lead.
Avoid overbending or kinking the stylet.
Use a new stylet when blood or other fluids accumulate on the stylet. Accumulated blood or other fluids may damage the lead or cause difficulty in passing the stylet into the lead.
Before inserting the lead – Use an anchoring sleeve with all leads. Ensure that the anchoring sleeve is positioned close to the lead connector pin. This will prevent inadvertent passage of the sleeve into the vein. If wiping the lead is necessary prior to insertion, ensure that the anchoring sleeve remains in position.
4
Page 5
Vessel and tissue damage – Use care when positioning the lead. Avoid known infarcted or thin ventricular wall areas to minimize the occurrence of perforation and dissection.
Handling a lead – Handle the lead with care at all times.
Do not implant the lead if it is damaged. Return the lead to a Medtronic representative.
Protect the lead from materials that shed small particles such as lint and dust. Lead insulators attract these particles.
Handle the lead with sterile surgical gloves that have been rinsed in sterile water or a comparable substance.
Do not severely bend, kink, or stretch the lead.
Do not immerse the lead in mineral oil, silicone oil, or any other liquid, except blood, at the time of implant.
Do not use surgical instruments to grasp the lead.
Do not force the lead if resistance is encountered during lead passage.
Chronic repositioning or removal of a screw-in lead –
Proceed with extreme caution if a lead must be removed or repositioned. Chronic repositioning or removal of screw-in transvenous leads may not be possible because of blood or fibrotic tissue development into the helix mechanism on the lead. In most clinical situations, it is preferable to abandon unused leads in place. Return all removed leads, unused leads, or lead sections to Medtronic for analysis.
Note: If a helix electrode does not disengage from the endocardium by rotating the connector pin, rotating the lead body counterclockwise may withdraw the helix electrode and decrease the possibility of damage to cardiovascular structures during removal.
Lead removal may result in avulsion of the endocardium, valve, or vein.
Lead junctions may separate, leaving the lead tip and bare wire in the heart or vein.
Chronic repositioning of a lead may adversely affect the low-threshold performance of a steroid lead.
An abandoned lead should be capped so that the lead does not transmit electrical signals.
Severed leads should have the remaining lead end sealed and the lead body sutured to adjacent tissue.
Connector compatibility – Although the lead conforms to the IS-1 International Connector Standard, do not attempt to use the lead with any device other than a commercially available implantable pacing system with which it has been tested and demonstrated to be safe and effective. The potential adverse consequences of using such a combination may include, but are not limited to, undersensing cardiac activity and failure to deliver necessary therapy.

6 Drug information

6.1 Mechanism of action

Steroid suppresses the inflammatory response that is believed to cause threshold rises typically associated with implanted pacing electrodes. Dexamethasone acetate is a synthetic steroid of the
glucocorticoid family. Glucocorticoids have potent anti-inflammatory actions via direct and indirect effects on major inflammatory cells. Glucocorticosteroids bind to a cytoplasmic glucocorticoid receptor as well as a membrane-bound receptor. Binding to the cytoplasmic receptor leads to receptor activation and translocation to the nucleus. The receptor interacts with specific DNA sequences within the regulatory regions of affected genes. Thus, glucocorticoids inhibit the production of multiple cell factors that are critical in generating the inflammatory response.

6.2 Pharmacokinetics and metabolism

Pharmacokinetics – The pharmacokinetics (local drug levels
and systemic levels) of dexamethasone acetate and its metabolites following implant were not evaluated in human clinical trials. When delivered intra-muscularly, the lipid-soluble dexamethasone acetate is slowly absorbed throughout the tissue.
Metabolism – The conversion of dexamethasone acetate to dexamethasone occurs within hours. The dexamethasone alcohol (dexamethasone) is the active glucocorticoid used in this Medtronic lead. Steroid is applied via MCRD (Monolithic controlled release device) and eluted to the tissue interface where it will be used. The form of the steroid, whether it is a prodrug or the pharmacologically active dexamethasone, is irrelevant, as the steroid is directly present at the injury site to treat the inflammation. Dexamethasone acetate is hydrolyzed into dexamethasone, which is readily absorbed by the surrounding tissue and body fluids. Glucocorticoids, when given systemically, are eliminated primarily by renal excretion of inactive metabolites.

6.3 Mutagenesis, carcinogenicity, and reproductive toxicity

Mutagenesis, carcinogenicity, and reproductive toxicity –
The mutagenesis, carcinogenicity, and reproductive toxicity of the CapSureFix Novus 4076 lead have not been evaluated. However, the mutagenesis, carcinogenicity, and reproductive toxicity of dexamethasone acetate have been evaluated previously.
Carcinogenesis, mutagenesis, and impairment of fertility –
No adequate studies have been conducted in animals to determine whether corticosteroids have a potential for carcinogenesis (tumor initiation or promotion). Dexamethasone was genotoxic in assays for clastogenicity (including sister chromatid exchange in human lymphocytes) but not in an assay for mutagenicity in salmonella (Ames test). Adrenocorticoids have been reported to increase or decrease the number and mobility of spermatozoa in some patients.

6.4 Pregnancy

Pregnancy – Dexamethasone acetate has been shown to be
teratogenic in many species when given in doses equivalent to the human dose. There are no adequate and well-controlled studies in pregnant women. Dexamethasone acetate should be used during pregnancy only if the potential benefit justifies the potential risk to the fetus. Studies in mice, rats, and rabbits have shown that adrenocorticoids increase the incidence of cleft palate, placental insufficiency, and spontaneous abortions, and can decrease the intrauterine growth rate.
5
Page 6
Nursing mothers – Systemically administered corticosteroids appear in human milk and could suppress growth, interfere with endogenous corticosteroid production, or cause other untoward effects in nursing infants. Because of the potential for serious adverse reactions in nursing infants from corticosteroids, a decision should be made whether to discontinue nursing or to use a non-steroidal lead, taking into account the importance of the lead and the drug to the mother.

6.5 Lactation

Corticosteroids are secreted into human milk, and there is a potential for serious adverse reactions. A decision should be made whether to nurse or to discontinue the drug, taking into account the importance of the drug to the mother. These potential risks of corticosteroids should also be considered along with any other steroidal therapy being received by the patient.

7 Adverse events

The potential adverse events related to the use of transvenous leads include, but are not limited to, the following patient related conditions that can occur when the lead is being inserted and/or repositioned: valve damage (particularly in fragile hearts), fibrillation and other arrhythmias, thrombolytic and air embolism, cardiac perforation, heart wall rupture, cardiac tamponade, muscle or nerve stimulation, pericarditis, pericardial rub, infection, myocardial irritability, thrombosis, and pneumothorax.
Potential complications related to the lead and the programmed parameters include, but are not limited to, the following:
Potential complica­tion Symptom
Lead dislodgement Intermittent or contin-
Lead conductor or helix electrode frac­ture or insulation fail­ure
Threshold elevation or exit block
a
Evidence indicates that there is a higher frequency of exit block in the ventricle when using a screw-in lead. This should be considered when selecting a screw-in lead for use in the ventricle.
a
uous loss of capture or sensing
Intermittent or contin­uous loss of capture or sensing
Loss of capture Adjust the pulse gen-
Potential acute/chronic complications associated with lead placement include, but are not limited to, the following:
Corrective action to be considered
Reposition the lead
Replace the lead. In some cases with a bipolar lead, the pulse generator may be pro­grammed to a unipo­lar configuration or the lead may be unipolar­ized
erator output. Replace or reposition the lead
Implant technique
Forcing the lead through the introducer
Use of too medial an approach with venous introducer resulting in clavicle and first rib binding
Puncturing the perios­teum and/or tendon when using subcla­vian introducer approach
Advancing the lead into the venous inser­tion site and/or through the veins without the stylet fully inserted
Potential complica­tion Corrective action
Screw electrode dam­age and/or insulation damage
Conductor coil frac­ture, insulation dam­age
Conductor coil frac­ture, insulation dam­age
Tip distortion and/or insulation perforation
Replace the lead
Replace the lead
Replace the lead
Replace the lead
In addition, prolonged implant procedures or multiple repositions can allow blood or body fluids to build up on the helix electrode mechanism. This may result in an increased number of rotations required to extend or retract the helix electrode, which may damage the lead.

8 Clinical trials

Note: Clinical studies were not performed on the 4076 lead due to
its similarity in design and function to the Model 5076 lead. The clinical data collected for the Model 5076 lead therefore supports the safety and efficacy claims for the 4076 lead and is included here for reference purposes. The lead used for the Model 5076 clinical study did not include titanium nitride coated platinum alloy tip and ring electrodes or an outer insulation of 55D polyurethane.

8.1 Summary

A multi-center, prospective, randomized control clinical study conducted at 31 investigational sites (in the United States, Europe, and Australia) compared the Model 5076 steroid eluting lead to the Model 5068 steroid eluting lead. Both leads are silicone extendible/retractable active fixation leads. The study design included an initial phase where each implanting physician implanted 2 patients with 5076 leads. This initial phase was then followed by the randomized clinical study. During the study, 119 randomized dual chamber patients received Model 5076 leads, 119 randomized dual chamber patients received Model 5068 leads, and 99 dual chamber patients were included in the initial phase.
Primary objectives: Compare the Model 5076 lead to the Model 5068 lead control lead for:
Survival from lead related complications and events at three months
Pacing threshold performance at three months
6
Page 7
Compare the Model 5076 lead to the Model 5072 lead (historical
0.40
0.50
0.30
0.20
0.10
0.00
Note: Mean Plus/Minus 2 Standard Errors
Follow-Up Time in Weeks
Thresholds in ms
5068
Impedance in Ohms
Note: Mean Plus/Minus 2 Standard Errors
Follow-Up Time in Weeks
Follow-Up Time in Weeks
Sensing in mV
Note: Mean Plus/Minus 2 Standard Errors
Thresholds in ms
Note: Mean Plus/Minus 2 Standard Errors
Follow-Up Time in Weeks
control) for:
Sensing performance at three months
Results:The lead Model 5076 was found to be clinically equivalent to the lead Model 5068 with respect to freedom from lead related complications and total lead related events at three months. The freedom from lead-related complications for the Model 5076 is 97.6% (5 complications) and the Model 5068 is
94.7% (11 complications). The freedom from atrial lead related events for the Model 5076 is 97.4% (3 events) and the Model 5068 is 94.7% (6 events). The freedom from ventricular lead related events for the Model 5076 is 91.0% (10 events) and the Model 5068 is 86.7% (12 events). The 16 lead-related Model 5076 events (one was past three months, therefore it is not included in the survival analysis) observed in the study are similar in kind and rate to those commonly experienced with endocardial, active fixation pacing leads.
There were five perforations reported with 5076 leads during the initial phase of the study (2.5%), and one perforation reported in the randomized portion of the study with 5076 leads (0.42%). No perforations were reported with the 5068 control lead (0%). Please refer to Chapter 7 of the technical manual for information regarding risks associated with endocardial leads and the Section 9.5 for recommended techniques to minimize tip pressure during lead implant.
The electrical analysis included comparison of pulse width thresholds and sensing performance. The mean Model 5076 pulse width thresholds measured at 2.5 V was found to be clinically equivalent to the Model 5068 from implant to three months post implant. The observed difference between the ventricular leads (5076-5068) is -0.007 ms and the observed difference between the atrial leads (5076-5068) is -0.009 ms. The mean Model 5076 R/P-wave amplitudes were found to be clinically equivalent to the mean Model 5072 (from previous clinical study) R/P-wave amplitudes from implant to three months post implant. The observed difference between R-wave amplitudes (5072-5076) is 1.00 mV and -0.03 mV between P-wave (5072-5076) amplitudes.

Figure 2. Atrial Pulse Width Thresholds at 2.5 V

Figure 3. Atrial bipolar impedance

Figure 4. Atrial P-wave sensing

Figure 5. Ventricular pulse width threshold at 2.5 V

7
Page 8

Figure 6. Ventricular bipolar impedance

Impedance in Ohms
Note: Mean Plus/Minus 2 Standard Errors
Follow-Up Time in Weeks
Note: Mean Plus/Minus 2 Standard Errors
Follow-Up Time in Weeks
Sensing in mV
1

Figure 7. Ventricular R-wave sensing

2. Leave the stylet inserted in the lead. Pull the stylet guide off the connector pin and slide the stylet guide towards the stylet knob.
3. Press both legs of the fixation tool together and place the most distal hole of the fixation tool on the connector pin (Figure 8).

Figure 8. Attachment of the fixation tool to the connector pin

1 Most distal hole of the fixation tool
4. While holding the lead as shown in Figure 9, and ensuring that the stylet is fully inserted, rotate the tool clockwise until the helix electrode is completely exposed (Figure 9). Maximum helix electrode exposure reveals approximately 1 1/2 to 2 coils.

Figure 9. Rotating the fixation tool

9 Directions for use

Proper surgical procedures and sterile techniques are the responsibility of the medical professional. The following procedures are provided for information only. Some implant techniques vary according to physician preference and the patient’s anatomy or physical condition. Each physician must apply the information in these instructions according to professional medical training and experience.

9.1 Opening the package

Within the sterile field, open the sterile package and remove the lead and accessories.

9.2 Verifying the mechanical functioning of the helix electrode

Before implant, verify the mechanical functioning of the helix electrode, as described by the following:
1. Within the sterile field, remove the lead and the
8
accompanying stylets from the sterile packaging. The lead is packaged with a stylet already inserted.
The recommended maximum number of rotations of the fixation tool needed to extend or retract the helix electrode for initial placement is stated in Section 10.1. The maximum number of rotations depends on the particular lead model but will increase or decrease proportionately for longer or shorter leads. Any additional curvatures introduced to the stylet may increase the number of rotations needed to extend or retract the helix electrode. Caution: Exceeding the recommended maximum number of rotations required to extend or retract the helix electrode may damage the lead.
5. Disconnect the fixation tool from the connector pin and release the proximal end of the lead body. Allow several seconds for the residual torque in the lead to be relieved.
6. After allowing for relief of the residual torque, reattach the fixation tool and rotate it counterclockwise until the helix electrode tip is retracted into the sheath.
Page 9

9.3 Using the stylet guide and stylets

1
The lead is packaged with the stylet guide attached to the connector pin and a stylet already inserted.
A stylet provides additional stiffness and controlled flexibility for maneuvering the lead into position. Stylets vary in stiffness to accommodate a physician’s preference for lead and stylet flexibility.
If the stylet guide has been removed, replace it by gently pushing it as far as possible onto the connector pin (Figure 10). Insert a stylet through the stylet guide and into the lead body.2 If a slight curve is needed for the stylet, refer to the procedure in Section 9.5, “Positioning the lead in the ventricle”, page 9.

Figure 10. Stylet guide attachment.

Caution: To avoid damage to the lead or body tissue, do not use
excessive force or surgical instruments to insert a stylet into the lead. To avoid lead tip distortion, the stylet should always remain fully inserted into the lead during lead introduction and while advancing the lead, especially through tortuous veins, that may cause the stylet to “back out” of the lead. When handling a stylet, avoid overbending, kinking, or blood contact. If blood is allowed to accumulate on a stylet, passage of the stylet into the lead may be difficult.

9.4 Selecting an insertion site

Figure 11. Suggested insertion site

1 Suggested insertion site

Caution: When using a subclavian vein approach, avoid placing the entry site in a location where the lead body can be clamped between the clavicle and the first rib. A more lateral approach is recommended to minimize the risk of first rib clavicular crush.
Clamping the lead may eventually cause the conductor to fracture, may cause damage to the insulation, or may cause other damage to the lead. Certain anatomical abnormalities, such as thoracic outlet syndrome, may also precipitate clamping of the lead.
Use fluoroscopy to facilitate accurate lead placement.

9.5 Positioning the lead in the ventricle

Caution: Use care when handling the lead during positioning.
Do not severely bend, kink, or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pin.
1. Insert the tapered end of a vein lifter into the incised vein and gently push the lead tip underneath and into the vein (Figure 12). A vein lifter facilitates lead insertion.

Figure 12. Using the vein lifter

The lead may be inserted by venotomy through several different venous routes, including the right or left cephalic vein, other subclavian branches, or the external or internal jugular vein. The lead may also be inserted into a subclavian vein through a percutaneous lead introducer. Select the desired entry site (Figure 11).
2. Advance the lead into the right atrium using a straight stylet to facilitate movement through the veins.
2
If additional stylets are needed, Medtronic recommends using the same type of Medtronic stylet that is packaged with the lead.
9
Page 10
3. Advance the lead through the tricuspid valve. Replacing the straight stylet with a gently curved stylet may add control in maneuvering the lead through the tricuspid valve. Advance the lead directly through the tricuspid valve, or project the lead tip against the lateral atrial wall and back the curved portion of the lead body through the tricuspid valve until the lead tip enters the ventricle.
4. Position the lead in the ventricle using the following techniques. Accurate positioning of the electrode is essential for stable pacing. Caution: If there is reason to believe the patient has an unusually thin wall at the apex of the right ventricle, the implanter may wish to consider another site for placement of the lead. Caution: If placing the lead in or near the right ventricular apex, use caution if passing the distal end of the lead directly from the valve to the apex. This technique may result in excessive tip pressure. Caution: If an awake patient feels a twinge of pain, this may be an early sign of perforation. Using one of the following techniques may help minimize transmission of pressure directly toward the tip of the lead:
Partially withdraw the stylet so that the stylet tip is proximal to the electrode ring while positioning the lead, to minimize tip stiffness. The stylet can then be gently advanced to the tip of the lead before securing the electrode in the endocardium.
A curved stylet may be used during positioning to minimize direct pressure on the apex.
Using a curved stylet, or partially withdrawing the stylet to allow the lead to be carried by blood flow, the lead may be curved up toward the outflow tract and then allowed to fall gently into position near the apex by pulling back on
the lead body. Use fluoroscopy (lateral position) to ensure that the tip is not in a retrograde position or is not lodged in the coronary sinus.
5. After placing the lead in a satisfactory position, extend the helix electrode by following the procedure in Section 9.7.

9.6 Positioning the lead in the atrium

Caution: Use care when handling the lead during positioning:
Do not severely bend, kink, or stretch the lead.
Do not use surgical instruments to grasp the lead or connector pin.
The following procedure is suggested for atrial placement of the lead:
1. Insert the tapered end of a vein lifter into the incised vein and gently push the lead tip underneath and into the vein (Figure 13). A vein lifter facilitates lead insertion.

Figure 13. Using the vein lifter

2. Advance the lead into the right atrium or the inferior vena cava using a straight stylet to facilitate movement through the veins. After the lead tip is passed into the atrium or inferior vena cava, replace the straight stylet with a gently curved stylet or the J-shaped stylet supplied with the lead.
3. Direct the lead tip into an appropriate position. Accurate positioning of the helix electrode is essential for stable pacing and sensing. Generally, a satisfactory position has the lead tip situated against the atrial endocardium in or near the apex of the appendage. As viewed on the fluoroscope (A-P view), the lead tip points medially and forward toward the left atrium. A successful position is usually achieved with an anterior, medial, or lateral tip location. Caution: An awake patient who feels a twinge of pain may indicate an early sign of perforation.
4. After placing the lead tip in a satisfactory position, extend the helix electrode by following the procedure in Section 9.7.

9.7 Securing the helix electrode into the endocardium

The following procedure is recommended for helix electrode fixation:
1. Leave the stylet inserted in the lead. Pull the stylet guide off the connector pin and slide the stylet guide towards the stylet knob.
2. Press both legs of the fixation tool together and place the most distal hole of the fixation tool on the connector pin (Figure 8).
3. Press the lead tip against the endocardium using the appropriate technique:
Ventricular placement: Press the lead tip against the endocardium by gently pushing the stylet and lead at the vein entry site.
Atrial placement: With the lead tip advanced into the atrium and a J-shaped or gently curved stylet in the lead, press the lead tip against the endocardium by gently pulling the stylet and the lead at the vein entry site.
10
Page 11
4. Rotate the fixation tool clockwise until the helix electrode is
Retracted
Extended
completely exposed (Figure 9). Maximum electrode exposure reveals approximately 1½ to 2 coils of the helix. The maximum number of rotations of the fixation tool needed to extend or retract the helix electrode for initial placement is stated in the section Section 10.1, “Specifications (nominal)”, page 14.” The maximum number of rotations depends on the particular lead model but will increase or decrease proportionately for longer or shorter leads. Caution: Prolonged implant procedures or multiple repositionings can allow blood or body fluids to build up on the helix electrode mechanism. This may result in an increased number of rotations required to extend or retract the helix electrode. Caution: Exceeding the number of rotations required to extend or retract the helix electrode may damage the lead.
5. Use fluoroscopy to verify helix electrode extension. Extension of the space between the indicator ring (A) and the drive mechanism (B) implies complete exposure of the helix electrode (Figure 14).

Figure 14. Possible views of the electrode

6. Disconnect the fixation tool from the connector pin and release the proximal end of the lead body. Allow several seconds for the residual torque in the lead to be relieved.
7. Carefully and partially, withdraw the stylet.
8. Verify that the helix electrode is affixed.
a. For a lead placed in the ventricle: Gently pull back on
the lead and check for resistance to verify affixation. A properly affixed helix electrode will remain in position. If the helix electrode is not properly affixed, the lead tip may become loose in the right ventricle. If the helix electrode does not remain affixed, it may be possible to fixate it during a subsequent attempt by rotating the whole lead body clockwise approximately one rotation after allowing the residual torque to be relieved in step 4. Caution is recommended if turning of the whole lead body is employed during or after fixation of the helix electrode.
b. For a lead placed in the atrium: Use frontal
fluoroscopy to check for lateral movement of the atrial tip, which reflects atrial and ventricular contractions. Check for constancy of the movement by rotating the lead body (up to 180 degrees in either direction) while the patient breathes deeply. Poor fixation is suspected when tip movement seems random. After the lead tip is fixated, allow lead slack to build up in the atrium. Lead slack helps prevent tip dislodgement. Enough slack is present if, under fluoroscopy, the lead assumes an “L” shape during deep inspiration. Avoid excessive slack buildup that may cause the loop of the lead to drop near the tricuspid valve.
9. If repositioning is required, reattach and rotate the fixation tool counterclockwise until the helix electrode is withdrawn. Use fluoroscopy to verify withdrawal. Again, as previously stated for final positioning of a lead placed in the ventricle, avoid transmission of pressure directly toward the tip of the lead to prevent the lead from being pushed directly into the apex. Caution: Do not rotate the fixation tool more than the number of rotations required to fully retract the helix electrode.
10. Remove the stylet guide and stylet completely. When removing the stylet guide, grip the lead firmly just below the connector pin; this will help prevent possible lead dislodgement.
11. Obtain final electrical measurements.

9.8 Taking electrical measurements

Attach a surgical cable to the connector pin for taking electrical measurements.
Note: A notch in the stylet guide allows connection of a surgical cable for electrical measurement.
Low stimulation thresholds and adequate sensing of intracardiac signal amplitudes indicate satisfactory lead placement. Medtronic recommends using a voltage source such as a pacing system analyzer for obtaining electrical measurements.
A low stimulation threshold provides for a desirable safety margin, allowing for a possible rise in thresholds that may occur within 2 months following implantation.
Adequate sensing amplitudes ensure that the lead is properly sensing intrinsic cardiac signals. Minimum signal requirements depend on the pulse generator’s sensitivity capabilities. Acceptable acute signal amplitudes for the lead must be greater than the minimum pulse generator sensing capabilities including an adequate safety margin to account for lead maturity.
11
Page 12

Table 1. Recommended electrical measurements at implant when using a pacing system analyzer

Ventricle Atrium
Maximum acute stimulation thresholds
Minimum acute sensing amplitudes 5.0 mV 2.0 mV
a
At pulse duration setting of 0.5 ms.
a
1.0 V
3.0 mA
1.5 V
4.5 mA
Initial electrical measurements may deviate from the recommendations because of acute cellular trauma. If this occurs, wait 5 to 15 minutes and repeat the testing procedure. Values may vary depending upon lead type, pulse generator settings, cardiac tissue condition, and drug interactions.
If electrical measurements do not stabilize to acceptable levels, it may be necessary to reposition the lead and to repeat the testing procedure.
Check for diaphragmatic stimulation by pacing at 10 V and observing on fluoroscopy whether the diaphragm contracts with each paced stimulus. If diaphragmatic pacing occurs, reduce the voltage until a diaphragmatic pacing threshold is determined. If the diaphragmatic threshold is less than the required programmed pacing output, the lead should be repositioned.
Pacing impedance (or resistance) is used to assess pulse generator function and lead integrity during routine pulse generator patient follow-up and to assist in troubleshooting suspected lead failures. (Additional troubleshooting procedures include ECG analysis, visual inspection, measurement of thresholds, and electrogram characteristics.)
Pacing impedance values are affected by many factors including lead position, electrode size, conductor design and integrity, insulation integrity, and the patient’s electrolyte balance. Apparent pacing impedance is significantly affected by the measurement technique; therefore, comparison of pacing impedance should be done using consistent methods of measurement and equipment.
An impedance higher or lower than the typical value is not necessarily a conclusive indication of lead failure. Other causes must be considered as well. Before reaching a conclusive diagnosis, the full clinical picture must be considered: pacing artifact size and morphology changes in 12-lead analog ECGs, muscle stimulation with bipolar leads, sensing and/or capture problems, patient symptoms, and pulse generator characteristics.
In addition to measuring impedance values, non-invasive blood pressure monitoring and the use of echocardiographic methods may be valuable during implantation.
Recommendations for clinically monitoring and evaluating leads in terms of impedance characteristics are given as follows.
For pulse generators with telemetry readout of impedance:
Routinely monitor and record impedance values, at implant and follow-ups, using consistent output settings. (Be aware that impedance values may be different at different
programmable output settings [for example, pulse width or pulse amplitude] of the pulse generator or pacing system analyzer.)
Establish a baseline chronic impedance value once the impedance has stabilized, generally within 6 to 12 months after implant.
Monitor for significant impedance changes and abnormal values.
Where impedance abnormalities occur, closely monitor the patient for indications of pacing and sensing problems. The output settings used for measuring impedance should be the same as that used for the original measurements.
For patients at high risk, such as pulse generator dependent patients, physicians may want to consider further action, such as: increased frequency of monitoring, provocative maneuvers, and ambulatory ECG monitoring.
For pulse generators without telemetry:
Record impedance value at implant. Also record the measurement device, its output settings, and the procedure used.
At the time of pulse generator replacement, if pacing analyzer system-measured impedance is abnormal, carefully evaluate lead integrity (including thresholds and physical appearance) and patient condition before electing to reuse the lead.
Bear in mind that impedances below 250 Ω may result in excessive battery current drain, which may seriously compromise pulse generator longevity, regardless of lead integrity.
For more information on obtaining electrical measurements, consult the technical manual supplied with the testing device.

9.9 Anchoring the lead

Note: The anchoring sleeve contains a radiopaque substance,
which allows visualization of the anchoring sleeve on a standard x-ray and may aid in follow-up exams.
Use the triple groove anchoring sleeve to secure the lead and to protect the lead insulation and conductor coil from damage caused by tight ligatures (Figure 15, Figure 16, and Figure 17).
Anchor the lead with nonabsorbable sutures.
Caution: Tabs on anchoring sleeves are provided to minimize the possibility of the sleeve entering the vein. Do not remove the tabs (Figure 15). If using a large diameter percutaneous lead introducer (PLI) sheath, extreme care should be taken to prevent passage of the anchoring sleeve into the PLI lumen and/or the venous system.
12
Page 13

Figure 15. Triple groove anchoring sleeve with tabs.

1
1 Anchoring sleeve tab
With a triple groove anchoring sleeve, generally 2 or 3 of the grooves may be used with the following procedure (Figure 16 or Figure 17).
The triple groove anchoring sleeve is situated at the connector end of the lead. Partially insert the anchoring sleeve into the vein.
Use the most distal suture groove to secure the anchoring sleeve to the vein.
Use the middle groove to secure the anchoring sleeve to the fascia and lead. First, create a base by looping a suture through the fascia underneath the middle groove and tying a knot. Continue by firmly wrapping the suture around the middle groove and tying a second knot.
Use the third and most proximal groove to secure the anchoring sleeve to the lead body.
Alternatively, only 2 of the 3 grooves may be used on the triple groove anchoring sleeve to tie down the lead. In that case, follow the anchoring procedure for the distal and middle groove (Figure 17).
Caution: Do not use the anchoring sleeve tabs for suturing.
Figure 16. Triple groove anchoring sleeve secured to the lead and
fascia using 3 grooves.
(Figure 18). During anchoring, take care to avoid dislodging the lead tip.

Figure 18. Do not secure the sutures too tightly and do not tie a suture to the lead body.

9.10 Connecting the lead to the pulse generator

Connect the lead to the pulse generator according to the instructions in the pulse generator manual.
Caution: Always remove the stylet before connecting the lead to the pulse generator. Failure to remove the stylet may result in lead failure.
Caution: To prevent undesirable twisting of the lead body, wrap the excess lead length loosely under the pulse generator and place both into the subcutaneous pocket (Figure 19).

Figure 19. While rotating the device, loosely wrap the excess lead length around the device.

Caution: When placing the pulse generator and lead into the subcutaneous pocket:
Do not coil the lead. Coiling the lead can twist the lead body and may result in lead dislodgment (Figure 20).
Do not grip the lead or pulse generator with surgical instruments.

Figure 20. Do not coil or twist the lead body.

Figure 17. Triple groove anchoring sleeve secured to the lead and
fascia using 2 grooves.
Tie the sutures securely but gently to prevent damage to the triple groove anchoring sleeve.
Caution: Do not secure the ligatures so tightly that they damage the vein or lead. Do not tie a ligature directly to the lead body

9.11 Post-implant evaluation

After implant, monitor the patient’s electrocardiogram until the patient is discharged. If a lead dislodges, it usually occurs during the immediate postoperative period.
Recommendations for verifying proper lead positioning include x-rays and pacing and sensing thresholds taken at pre-hospital discharge, 3 months after implant, and every 6 months thereafter.
13
Page 14
In the event of a patient death, explant all implanted leads and devices and return them to Medtronic with a completed Product Information Report form. Call the appropriate phone number on the back cover if there are any questions on product handling procedures.

10 Specifications

10.1 Specifications (nominal)

Parameter Model 4076
Type Bipolar
Chamber Ventricle/Atrium
Fixation Screw-in
Length 20–110 cm
Connector IS-1 BI
Materials Conductor: MP35N nickel alloy
Outer Insula-
Inner Insula-
Connector pin: Stainless steel
Connector
Electrode materi­als
Electrode surface area
Tip to ring spacing 10 mm
Diameter Lead body: 1.9 mm
Lead introducer (recommended size)
without guide
with guide
Helix length 1.8 mm (fully extended)
Lead conductor resistance
Steroid Dexamethasone acetate
Amount of steroid 680 µg (target dosage)
Unipolar: 33.5 Ω (58 cm)
Bipolar: 61.4 Ω (58 cm)
55D polyurethane
tion
Treated silicone rubber
tion:
Stainless steel
ring:
Helix: Titanium nitride coated platinum
alloy
Ring: Titanium nitride coated platinum
alloy
Helix: 4.2 mm
Ring: 22 mm
wire:
wire:
2
2
2.3 mm (7 French)
3.0 mm (9 French)

Table 2. Maximum number of rotations to extend or retract the helix electrode

Straight stylet Lead length J-shaped stylet
10 45 cm 15
11 52 cm 17
12 58 cm 18
14 65 cm
18 85 cm
23 110 cm

11 Medtronic warranty

For complete warranty information, see the accompanying warranty document.

12 Service

Medtronic employs highly trained representatives and engineers located throughout the world to serve you and, upon request, to provide training to qualified hospital personnel in the use of Medtronic products. Medtronic also maintains a professional staff to provide technical consultation to product users. For more information, contact your local Medtronic representative, or call or write Medtronic at the appropriate telephone number or address listed on the back cover.
The following table lists the recommended maximum rotations required to extend or retract the helix electrode for initial placement.
14
Page 15
Page 16
Medtronic, Inc.
*M955297A001*
710 Medtronic Parkway Minneapolis, MN 55432 USA www.medtronic.com +1 763 514 4000
Medtronic USA, Inc.
Toll-free in the USA (24-hour technical consultation for physicians and medical professionals) Bradycardia: +1 800 505 4636 Tachycardia: +1 800 723 4636
Europe/Middle East/Africa
Medtronic International Trading Sàrl Route du Molliau 31 Case Postale 84 CH-1131 Tolochenaz Switzerland +41 21 802 7000
Technical manuals
www.medtronic.com/manuals
© 2015 Medtronic, Inc. M955297A001 C 2015-09-23
Loading...