Medtronic 304-01POR Instructions for Use

Heparin Assay Cartridge Heparin-Testkartusche Cartucho para análisis de heparina Cartouche test de concentration d'héparine Cartuccia per l'analisi dell'eparina
Instructions for Use • Gebrauchsanweisung • Instrucciones de uso • Mode d’emploi • Istruzioni per l’uso
Caution: Federal law (USA) restricts
Trademarks may be registered and are the property of their respective owners. Marken sind eventuell eingetragen und sind das Eigentum ihrer jeweiligen Inhaber. Las marcas comerciales pueden estar registradas y pertenecen a sus respectivos propietarios. Les marques commerciales mentionnées peuvent être déposées et appartiennent à leurs propriétaires respectifs. I marchi di fabbrica possono essere registrati e sono di proprietà dei rispettivi titolari.
Explanation of symbols on package labeling / Erläuterung der Symbole auf dem Verpackungsetikett / Explicación de los símbolos que aparecen en la documentación del envase / Explication des symboles des étiquettes sur l'emballage / Spiegazione dei simboli presenti sulle etichette della confezione
Conformité Européenne (European Conformity). This symbol means that the device fully complies with the European Council Directive IVDD: 98/79/EC. / Conformité Européenne (Europäische Konformität). Dieses Symbol besagt, dass das Produkt allen Vorschriften der Direktive 98/79/EG des Europäischen Parlaments und des Rates entspricht. / Conformité Européene (Conformidad europea). Este símbolo indica que el dispositivo cumple totalmente la Directiva del Consejo Europeo sobre dispositivos de diagnóstico in vitro (DDIV): 98/79/CE. / Conformité Européenne. Ce symbole signifie que le dispositif est entièrement conforme à la Directive européenne IVDD relative aux diagnostics in vitro : 98/79/CE. / Conformité Européenne (Conformità europea). Questo simbolo significa che il dispositivo è pienamente conforme alla Direttiva IVDD 98/79/CE del Consiglio europeo.
Sterilized Using Irradiation / Mittels Strahlung sterilisiert / Esterilizado mediante irradiación / Stérilisation par irradiation / Sterilizzato con radiazioni ionizzanti
Nonsterile / Unsteril / No estéril / Non stérile / Non sterile
Do Not Reuse / Nicht wiederverwenden / No reutilizar / Ne pas réutiliser / Non riutilizzare
Use-By Date / Zu verwenden bis einschließlich / No utilizar después de / À utiliser jusqu'au / Utilizzare entro
Manufactured In / Hergestellt in / Fabricado en / Lieu de fabrication / Fabbricato in
Lot Number / Chargennummer / Número de lote / Numéro de lot / Numero di lotto
Quantity / Menge / Cantidad / Quantité / Quantità
Consult Instructions for Use / Gebrauchsanweisung beachten / Consultar las instrucciones de uso / Consulter le mode d'emploi / Consultare le istruzioni per l'uso
1
For US Audiences Only / Gilt nur für Leser in den USA / Solo aplicable en EE. UU. / Uniquement pour les États-Unis / Esclusivamente per il mercato statunitense
Temperature Limit / Temperaturbereich / Límite de temperatura / Limite de température / Limiti di temperatura
Fragile, Handle with Care / Vorsicht, zerbrechlich / Frágil, manipular con cuidado / Fragile, manipuler avec précaution / Fragile, maneggiare con cura
Catalog Number / Katalognummer / Número de catálogo / Numéro de référence / Numero di catalogo
Manufacturer / Hersteller / Fabricante / Fabricant / Fabbricante
Authorized Representative in the European Community / Autorisierter Repräsentant in der Europäischen Gemeinschaft / Representante autorizado en la Comunidad Europea / Représentant agréé dans la Communauté européenne / Rappresentante autorizzato nella Comunità Europea
This Way Up / Diese Seite oben / Este lado hacia arriba / Placer ce côté vers le haut / Questa parte rivolta verso l'alto
In Vitro Diagnostic Medical Device / Medizinisches Gerät zur In-vitro-Diagnostik / Dispositivo médico para diagnóstico in vitro / Dispositif médical de diagnostic in vitro / Dispositivo medico diagnostico in vitro
Biological Risks / Biogefährdung / Riesgos biológicos / Risques biologiques / Rischi biologici
2

Heparin Assay Cartridge

1. Intended Use

For use in measuring the heparin concentration of a fresh whole blood sample by performing protamine titration.
For in vitro diagnostic use.

2. Summary

Medtronic heparin assay cartridges perform a semi-quantitative measurement of heparin concentration in a whole blood sample. Heparin is measured by titration against known quantities of protamine added to the cartridge channels. The channel whose quantity of protamine neutralizes the heparin in the test sample will have the shortest clotting time. The concentration of heparin is calculated based on the known amount of protamine. Dilute thromboplastin is added to heparin assay cartridges to accelerate the reaction time. Heparin assay cartridges measure heparin levels from 0 to 6.0 mg/kg body weight or 0 to 8.2 units/mL. Heparin assay cartridges are available as 4 or 6 channel cartridges.

3. Principles

Each heparin assay cartridge channel contains an increasing quantity of protamine sulfate and an equal amount of dilute thromboplastin. The protamine in each channel neutralizes a specific amount of heparin. A fresh whole blood sample is delivered to each cartridge channel, and the reagents are mixed with the test sample by the Medtronic HMS Plus instrument to initiate coagulation. The first channel to clot is the channel in which the amount of protamine most closely neutralizes all of the heparin in the test sample without an excess of either heparin or protamine.
The end point of the test is the detection of clot formation. The HMS Plus instrument detects clot formation by measuring the rate of fall of the plunger contained in each cartridge channel. The plunger assembly falls rapidly through an unclotted sample. A fibrin web formed during clotting impedes the rate of plunger assembly descent. Clot formation is detected by a photo optical sensor located in the actuator assembly of the instrument.

4. Reagents

Heparin assay cartridges are formulated using United States Pharmacopoeia (USP) grade protamine sulfate for injection. The protamine-to-heparin neutralization ratio is precisely calibrated for each lot of protamine. Thromboplastin diluted in a buffer is used as an activator. Sodium azide is added as a bacteriostatic agent. Table 1 lists the quantity of protamine in each cartridge channel for a 1:110 neutralization ratio.
Table 1. Cartridge Protamine Concentrations
Color Range CH1 CH2 CH3 CH4 CH5 CH6
Red 0-0.9 mg/kg
0-1.2 u/mL
Yellow 0-1.5 mg/kg
0-2.0 u/mL
Tan 1.5-3.0 mg/kg
2.0-4.1 u/mL
Silver 2.0-3.5 mg/kg
2.7-4.8 u/mL
Micrograms of Protamine
0 0.8 1.6 2.4 - -
0 1.4 2.7 4.1 - -
4.1 5.4 6.8 8.2 - -
5.4 6.8 8.2 9.5 - -
Instructions for Use English 3
Color Range CH1 CH2 CH3 CH4 CH5 CH6
Blue 2.5-4.0 mg/kg
3.4-5.4 u/mL
Green 3.5-5.0 mg/kg
4.8-6.8 u/mL
Orange 0-2.5 mg/kg
0-3.4 u/mL
Gold 1.5-4.0 mg/kg
2.0-5.4 u/mL
White 2.5-5.0 mg/kg
3.4-6.8 u/mL
Purple 4.5-6.0 mg/kg
6.1-8.2 u/mL
Black 3.5-6.0 mg/kg
4.8-8.2 u/mL
Micrograms of Protamine
6.8 8.2 9.5 10.9 - -
9.5 10.9 12.2 13.6 - -
0 1.4 2.7 4.1 5.4 6.8
4.1 5.4 6.8 8.2 9.5 10.9
6.8 8.2 9.5 10.9 12.2 13.6
12.2 13.6 15.0 16.3 - -
9.5 10.9 12.2 13.6 15.0 16.3

5. Storage

Refer to cartridge package labeling for specific storage conditions and expiration dates.
Note: Do not freeze.

5.1. Room Temperature

Heparin assay cartridges may be stored at room temperature (10°C to 25°C [50°F to 77°F]) until the Use-By Date on the cartridge box. Cartridges should be stored in the box so that the Use-By Date can be referenced. Cartridges should be discarded if the Use-By Date is unknown, or if the reagent has evaporated or appears contaminated.

5.2. Refrigerated

Heparin assay cartridges are stable until the Use-By Date indicated on the cartridge box when stored at 2°C to 10°C (35.6°F to 50°F). Once a cartridge is removed from the refrigerator and left at room temperature, it must be used within 1 week or by the Use-By Date, whichever occurs first. Once removed, cartridges cannot be returned to refrigeration.

6. Instrument

Medtronic heparin assay cartridges are designed specifically for use in the Medtronic HMS Plus instrument. For HMS Plus instrument instructions, consult the HMS Plus instrument Operator's Manual.

7. Specimen Collection and Preparation

Blood samples that are improperly drawn and handled can yield erroneous test results; samples must be collected carefully, and handling techniques must be consistent.
If blood is obtained by venipuncture, the venipuncture should be clean and the first several milliliters of blood discarded to avoid contamination of the sample with tissue thromboplastin. Blood should not be taken from heparinized indwelling catheters or other anticoagulated lines without thoroughly flushing the line.
Baseline samples, which contain no anticoagulant, should be tested within 60 seconds of collection. Heparinized samples should be tested within 2 minutes of collection.

8. Instructions for Use

This product is intended for use only by trained individuals.
4 Instructions for Use English
Materials supplied:
heparin assay cartridge
3-mL Monoject™ sterile, single-use syringe
19-ga, 36-mm (1 7/16-in) sterile, single-use, blunt needle
Materials required, not supplied:
test sample, fresh whole blood
Medtronic HMS Plus instrument
Steps:
1. Gently tap the heparin assay cartridge and place in the HMS Plus actuator heat block. Heparin assay cartridges do not require prewarming.
Note: If an HR-ACT (high-range activated clotting time) is tested with the heparin assay cartridge, the HR-ACT cartridge must be prewarmed at least 3 minutes.
2. Fill the supplied syringe with test sample in the following quantities:
4 channel heparin assay cartridge: 1.5 mL
4 channel heparin assay cartridge with HR-ACT: 2.5 mL
6 channel heparin assay cartridge: 2 mL
3. Attach and prime a clean blunt tip needle.
4. Insert the syringe into the dispenser of the HMS Plus instrument. Lock the syringe in place. The dispenser will automatically deliver the appropriate amount of test sample to each cartridge channel.
5. Press “Start/Stop” to initiate the test.
6. If the dispenser is not functioning, the test sample may be delivered manually into each cartridge channel (consult the HMS Plus instrument Operator's Manual for complete instructions).
Note: It is important to accurately deliver the test sample into each cartridge channel.

9. Results

Upon test completion, the following information can be obtained:
the sample heparin concentration in mg/kg, u/mL, or u/kg; the detection channel; as well as the run time of the detection channel
the average HR-ACT and individual HR-ACT channel clotting times (if an HR-ACT test is performed)
additional heparin required for the patient, pump, or total volume to maintain the desired heparin level
the protamine sulfate dose required to neutralize the patient, pump, or total heparin concentration
For additional information regarding heparin assay results, consult the HMS Plus instrument Operator's Manual.

10. Limitations

Medtronic heparin assay cartridges are for in vitro diagnostic use only and are to be used only with the Medtronic HMS Plus instrument.
Cartridges that have been used for testing should be considered hazardous. It is the responsibility of the user to dispose of the cartridge in accordance with local regulations and institution procedures.
The accuracy of the heparin assay is dependent on the cartridge channel resolution. This resolution varies from 0.3 to 0.5 mg/kg (0.4 to 0.7 u/mL) depending on the cartridge range.
The accuracy and precision of heparin measurements are dependent on the USP tolerances for heparin and protamine sulfate.
Instructions for Use English 5
The heparin measurement obtained is considered valid as long as the sample heparin level is within the range of the selected cartridge. For example, when a clot is detected in channel 1 of a heparin assay cartridge, an alert message appears, indicating the heparin concentration may be less than the measured value. When a clot is detected in the last channel of the cartridge, an alert message appears, indicating that the heparin concentration may be higher than the measured value.
Measurement of heparin concentration by protamine titration is generally not affected by hypothermia, hemodilution, or platelet inhibitory drugs.
To obtain accurate and consistent heparin measurements, the following must be observed:
1. Blood samples must be representative of the patient’s circulating heparin.
2. To optimize precision, all technique variables should be held constant.
3. Patient diagnosis and medications should be noted. Tests should be repeated if unexplained values are obtained. If necessary, a sample can be citrated and sent to the laboratory for further analysis.
For HMS Plus instrument precautions and limitations, consult the HMS Plus Operator's Manual.

11. Heparin Assay Cartridge Ranges

The following table lists the available heparin assay cartridges, the measurement range for each cartridge, and the channel resolution.
Table 2. Cartridge Ranges
Color Units CH1 CH2 CH3 CH4 CH5 CH6 Reso-
Red mg/kg 0.0 0.3 0.6 0.9 - - 0.3
u/mL 0.0 0.4 0.8 1.2 - - 0.4
Yellow mg/kg 0.0 0.5 1.0 1.5 - - 0.5
u/mL 0.0 0.7 1.4 2.0 - - 0.7
Tan mg/kg 1.5 2.0 2.5 3.0 - - 0.5
u/mL 2.0 2.7 3.4 4.1 - - 0.7
Silver mg/kg 2.0 2.5 3.0 3.5 - - 0.5
u/mL 2.7 3.4 4.1 4.8 - - 0.7
Blue mg/kg 2.5 3.0 3.5 4.0 - - 0.5
u/mL 3.4 4.1 4.8 5.4 - - 0.7
Green mg/kg 3.5 4.0 4.5 5.0 - - 0.5
u/mL 4.8 5.4 6.1 6.8 - - 0.7
Purple mg/kg 4.5 5.0 5.5 6.0 - - 0.5
u/mL 6.1 6.8 7.5 8.2 - - 0.7
Orange mg/kg 0.0 0.5 1.0 1.5 2.0 2.5 0.5
u/mL 0.0 0.7 1.4 2.0 2.7 3.4 0.7
Gold mg/kg 1.5 2.0 2.5 3.0 3.5 4.0 0.5
u/mL 2.0 2.7 3.4 4.1 4.8 5.4 0.7
White mg/kg 2.5 3.0 3.5 4.0 4.5 5.0 0.5
u/mL 3.4 4.1 4.8 5.4 6.1 6.8 0.7
Black mg/kg 3.5 4.0 4.5 5.0 5.5 6.0 0.5
u/mL 4.8 5.4 6.1 6.8 7.5 8.2 0.7
lution

12. Specific Performance Characteristics

Heparin assay cartridges are designed to provide results within 30 to 249 seconds from the time of test initiation. Clotting in all channels within 30 seconds of test initiation may indicate an activated test sample.
6 Instructions for Use English
Heparin assay run times greater than 249 seconds may indicate that heparin levels exceed the cartridge range.
When 2 channels of the same cartridge clot simultaneously, the HMS Plus instrument will display the result at the lower heparin concentration.
If the heparin concentration in a test sample is between the measured value of 2 cartridge channels, the sample may clot in either of the 2 channels.
Table 3 shows accuracy and precision obtained with the heparin assay cartridges. Tests were run on heparinized, recalcified plasma. The data shows the results of 6 to 9 sequential heparin assay cartridges run on the HMS Plus instrument.
Table 3. Cartridge Performance Characteristics
Heparin Level of Plasma
Mean Measured Hepa­rin Level
n
SDaCV
b
0.7 u/mL 0.7 u/mL 6 0 0
2.0 u/mL 2.0 u/mL 7 0 0
3.4 u/mL 3.4 u/mL 8 0 0
4.1 u/mL 4.1 u/mL 9 0 0
4.8 u/mL 4.8 u/mL 6 0 0
5.4 u/mL 5.3 u/mL 6 0.2 4.6%
a
Standard deviation
b
Coefficient of variation
The precision and accuracy of the heparin assay cartridges are dependent on the USP and IU tolerances for heparin. The accuracy is within the resolution of the applicable cartridge. The test accuracy is within ±½ channel (±0.2, or 0.4 u/mL for red, and ±0.35, or 0.7 u/mL for the other cartridges). Precision for the heparin assay cartridges is illustrated in the resolution column of Table 4.
Table 4. Cartridge Test Precision Example
Color Units CH1 CH2 CH3 CH4 CH5 CH6 Resolu-
tion
Red u/mL 0.0 0.4 0.8 1.2 - - 0.4 Yellow u/mL 0.0 0.7 1.4 2.0 - - 0.7 White u/mL 3.4 4.1 4.8 5.4 6.1 6.8 0.7

13. Quality Control

Heparin assay controls are used to verify instrument and cartridge performance. Pass/fail criteria for the controls tests are programmed into the HMS Plus instrument.
Note: If using only liquid controls, control tests must be performed every 8 hours of patient testing.
The HEPtrac electronic quality control is used to supplement liquid controls and does not replace them. Refer to the current regulatory agency guidelines on the acceptability and use of electronic controls.

14. References

1. Anido G, Freeman DJ. Heparin assay and protamine titration. Am J Clin Pathol. 1981;76:410.
2. Bowie JE, Kemna GD. Automated management of heparin anticoagulation in cardiovascular surgery. Proc Amer Acad Perfus. 1985;6:85.
3. Despotis GJ, et al. Comparison of activated coagulation time and whole blood heparin measurements with laboratory plasma anti-xa heparin concentration in patients having cardiac operations. J Thorac
Cardiovasc Surg. 1994;108:1076-82.
Instructions for Use English 7
4. Donahoo KM, et al. A promising new multi-functional instrument for monitoring heparin therapy. Adv Therapy. 1985;2:150.
5. Fox DJ, et al. Vehicles of heparin management: A comparison. J Extracorp Tech. 1979;11:137.
6. Hill AG, Lefrak EA. Monitoring heparin and protamine therapy during cardiopulmonary bypass procedures. AMSECT Proceedings. 1978;6:8.
7. Kemna GD, Bowie JE. Heparin protamine management for cardiopulmonary bypass patients with abnormal heparin dose response. Proc Amer Acad Perfus. 1985;6:89.
8. NCCLS Document H21-A. 1986;6(20):596-597.
9. Pifarré R, et al. Management of postoperative heparin rebound following cardiopulmonary bypass. J Thorac Cardiovasc Surg. 1981;81:378.
10. Triplett DA, Smith C. Sensitivity of the activated partial thromboplastin time. Results of the CAP survey and a series of mild and moderate factor deficiencies. In: Triplett DA, ed. Standardization of Coagulation Assays: An Overview. Coll Am Pathol; 1981:142.

15. Disclaimer of Warranty

THE FOLLOWING DISCLAIMER OF WARRANTY APPLIES TO UNITED STATES CUSTOMERS ONLY:
ALTHOUGH THE HEPARIN ASSAY CARTRIDGES, HEREAFTER REFERRED TO AS “PRODUCT,” HAVE BEEN MANUFACTURED UNDER CAREFULLY CONTROLLED CONDITIONS, MEDTRONIC HAS NO CONTROL OVER THE CONDITIONS UNDER WHICH THIS PRODUCT IS USED. MEDTRONIC, THEREFORE, DISCLAIMS ALL WARRANTIES, BOTH EXPRESS AND IMPLIED, WITH RESPECT TO THE PRODUCT, INCLUDING, BUT NOT LIMITED TO, ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. MEDTRONIC SHALL NOT BE LIABLE TO ANY PERSON OR ENTITY FOR ANY MEDICAL EXPENSES OR ANY DIRECT, INCIDENTAL, OR CONSEQUENTIAL DAMAGES CAUSED BY ANY USE, DEFECT, FAILURE, OR MALFUNCTION OF THE PRODUCT, WHETHER A CLAIM FOR SUCH DAMAGES IS BASED UPON WARRANTY, CONTRACT, TORT, OR OTHERWISE. NO PERSON HAS ANY AUTHORITY TO BIND MEDTRONIC TO ANY REPRESENTATION OR WARRANTY WITH RESPECT TO THE PRODUCT.
The exclusions and limitations set out above are not intended to, and should not be construed so as to, contravene mandatory provisions of applicable law. If any part or term of this Disclaimer of Warranty is held to be illegal, unenforceable, or in conflict with applicable law by a court of competent jurisdiction, the validity of the remaining portions of this Disclaimer of Warranty shall not be affected, and all rights and obligations shall be construed and enforced as if this Disclaimer of Warranty did not contain the particular part or term held to be invalid.

16. Disclaimer of Warranty

THE FOLLOWING DISCLAIMER OF WARRANTY APPLIES TO CUSTOMERS OUTSIDE THE UNITED STATES:
ALTHOUGH THE HEPARIN ASSAY CARTRIDGES, HEREAFTER REFERRED TO AS “PRODUCT,” HAVE BEEN CAREFULLY DESIGNED, MANUFACTURED, AND TESTED PRIOR TO SALE, THE PRODUCT MAY FAIL TO PERFORM ITS INTENDED FUNCTION
8 Instructions for Use English
SATISFACTORILY FOR A VARIETY OF REASONS. THE WARNINGS CONTAINED IN THE PRODUCT LABELING PROVIDE MORE DETAILED INFORMATION AND ARE CONSIDERED AN INTEGRAL PART OF THIS DISCLAIMER OF WARRANTY. MEDTRONIC, THEREFORE, DISCLAIMS ALL WARRANTIES, BOTH EXPRESS AND IMPLIED, WITH RESPECT TO THE PRODUCT. MEDTRONIC SHALL NOT BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES CAUSED BY ANY USE, DEFECT, OR FAILURE OF THE PRODUCT, WHETHER THE CLAIM IS BASED ON WARRANTY, CONTRACT, TORT, OR OTHERWISE.
The exclusions and limitations set out above are not intended to, and should not be construed so as to, contravene mandatory provisions of applicable law. If any part or term of this Disclaimer of Warranty is held by any court of competent jurisdiction to be illegal, unenforceable, or in conflict with applicable law, the validity of the remaining portion of the Disclaimer of Warranty shall not be affected, and all rights and obligations shall be construed and enforced as if this Disclaimer of Warranty did not contain the particular part or term held to be invalid.
Instructions for Use English 9

Heparin-Testkartusche

1. Vorgesehener Verwendungszweck

Für die Verwendung zur Bestimmung der Heparinkonzentration einer frischen Vollblutprobe nach der Protamintitrationsmethode.
Nur für die Verwendung in der In-vitro-Diagnostik.

2. Kurzbeschreibung

Medtronic Heparin-Testkartuschen führen eine semiquantitative Bestimmung der Heparinkonzentration in einer Vollblutprobe durch. Die Bestimmung der Heparinkonzentration erfolgt durch Titration gegen bekannte Mengen Protamin, das den Kartuschenkanälen zugegeben wird. Der Kanal, dessen Menge Protamin das Heparin in der Probe neutralisiert, hat die kürzeste Gerinnungszeit. Die Heparinkonzentration wird auf Grundlage der bekannten Protaminmenge berechnet. Zur Beschleunigung der Reaktionszeit wird den Heparin-Testkartuschen Thromboplastinverdünnung zugegeben. Heparin-Testkartuschen messen Heparinspiegel von 0 bis 6,0 mg/kg Körpergewicht bzw. von 0 bis 8,2 E/ml. Heparin-Testkartuschen sind als 4- oder 6-Kanal-Kartuschen erhältlich.

3. Anwendungsprinzip

Jeder Kanal der Heparin-Testkartusche enthält Protaminsulfat in zunehmender Menge und eine gleichbleibende Menge Thromboplastinverdünnung. Das Protamin in den einzelnen Kanälen neutralisiert eine bestimmte Menge Heparin. Jedem Kartuschenkanal wird eine frische Vollblutprobe zugegeben, und die Reagenzien werden durch das Medtronic HMS Plus-Gerät mit der Testprobe vermischt, um die Koagulation einzuleiten. Der erste Kanal, bei dem eine Gerinnung auftritt, ist der Kanal, dessen Protaminmenge am besten die in der Testprobe enthaltene Heparingesamtmenge neutralisiert, ohne dass ein Überschuss an Heparin oder Protamin vorliegt.
Der Test endet, sobald eine Gerinnungsbildung erkannt wird. Das HMS Plus-Gerät erkennt die Gerinnung an der Senkgeschwindigkeit des in jedem Kartuschenkanal enthaltenen Kolbens. Bei einer nicht geronnenen Probe senkt sich der Kolben schnell. Bei der Gerinnung bildet sich ein Fibrinnetz, wodurch der Kolben mit geringerer Geschwindigkeit sinkt. Die Gerinnung wird von einem fotooptischen Sensor im Betätigungselement des Geräts registriert.

4. Reagenzien

Heparin-Testkartuschen enthalten Protaminsulfat für Injektionszwecke in USP-Qualität (United States Pharmacopoeia). Das Protamin-zu-Heparin­Neutralisierungsverhältnis ist für jede Protamincharge exakt kalibriert. Als Aktivator wird Thromboplastinverdünnung in einem Puffer verwendet. Als Bakeriostatikum wird Natriumazid zugesetzt. Tabelle 1 führt die Protaminmenge in den einzelnen Kartuschenkanälen für ein Neutralisierungsverhältnis von 1:110 auf.
Tabelle 1. Protaminkonzentrationen der Kartusche
Farbe Bereich KA1 KA2 KA3 KA4 KA5 KA6
Rot 0–0,9 mg/kg
0–1,2 E/ml
Gelb 0–1,5 mg/kg
0–2,0 E/ml
Protamin [Mikrogramm]
0 0,8 1,6 2,4
0 1,4 2,7 4,1
10 Gebrauchsanweisung Deutsch
Loading...
+ 28 hidden pages