301-000-000
340-000-000
McGRATHTM MAC
Video Laryngoscope
Addendum
This addendum provides the appropriate specication tables for the next-generation McGRATH™ MAC video laryngoscope per
IEC 60601-1-2: 2014 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance –
Collateral Standard: Electromagnetic disturbances – Requirements and tests.
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
Table 1 – Electromagnetic Emissions
The device is intended for use in the electromagnetic
environment specied below. The operator of the device
should assure that it is used in such an environment.
Emissions Test Compliance Level
Radiated RF emissions CISPR
11:2009 +A1:2010
Radiated RF emissions
ISO 7137:1995
Table 2 – Electromagnetic Immunity
The device is intended for use in the electromagnetic
environment specied below. The operator of the device
should assure that it is used in such an environment.
Immunity Test Compliance Level
Electrostatic Discharges
IEC 61000-4-2:2008
Radiated RF EM Fields
IEC 61000-4-3:2006 +A1:2007
+A2:2010
Radiated RF EM Fields and
Proximity Wireless Fields
IEC 61000-4-3:2006 +A1:2007
+A2:2010
Rated Power-frequency
Magnetic Field
IEC 61000-4-8:2009
Group 1 Class B
100MHz – 6GHz
±8kV contact
±2kV, ±4kV, ±8kV, ±15kV air
10V/m Modulation of 80%
AM at 1kHz.
Frequency 80MHz –
2700MHz
Refer to Table 3
30A/m. Frequency 50Hz
and 60Hz
Table 3 – Proximity elds from RF wireless communications
equipment
Test Frequency
(MHz)
385 18 Hz 27
450 18 Hz 28
710
745
780
810
870
930
1720
1845
1970
2450 217 Hz 28
5240
5500
5785
The following tests are not applicable as the device is battery
powered:
• Terminal disturbance voltages CISPR 11:2009+A1:2010
• AC-Mains Harmonics IEC 61000-3-2:2005 +A1:2008+A2:2009
• AC-Mains uctuations and icker IEC 61000-3-3:2013
• Electrical fast transient / bursts IEC 61000-4-4:2012
• Surges IEC 61000-4-5:2005
• Conducted disturbances induced by RF elds IEC 61000-46:2013
• Voltage dips IEC 61000-4-11:2004
• Voltage interruptions IEC 61000-4-11:2004
Modulation Complies with
IMMUNITY Level (V/m)
217 Hz 9
18 Hz 28
217 Hz 28
217 Hz 9
Consult
instructions
for use
Part No. PT00147312 Rev A 2021-06
COVIDIEN, COVIDIEN with logo, and Covidien logo and Positive Results for
Life are U.S. and internationally registered trademarks of Covidien AG.
™* brands are trademarks of their respective owners. Other brands are
trademarks of a Covidien company.
© 2021 Covidien. www.medtronic.com/patents
Covidien llc, 15 Hampshire Street, Manseld, MA 02048 USA.
Medtronic B.V., Earl Bakkenstraat 10, 6422 PJ Heerlen, The Netherlands.
www.covidien.com