Medtronic 301-000-000 Instructions for Use

301-000-000
340-000-000
McGRATHTM MAC
Video Laryngoscope
Addendum
This addendum provides the appropriate specication tables for the next-generation McGRATH™ MAC video laryngoscope per IEC 60601-1-2: 2014 Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – Requirements and tests.
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
Table 1 – Electromagnetic Emissions
The device is intended for use in the electromagnetic environment specied below. The operator of the device should assure that it is used in such an environment.
Emissions Test Compliance Level
Radiated RF emissions CISPR 11:2009 +A1:2010
Radiated RF emissions ISO 7137:1995
Table 2 – Electromagnetic Immunity
The device is intended for use in the electromagnetic environment specied below. The operator of the device should assure that it is used in such an environment.
Immunity Test Compliance Level
Electrostatic Discharges IEC 61000-4-2:2008
Radiated RF EM Fields IEC 61000-4-3:2006 +A1:2007 +A2:2010
Radiated RF EM Fields and Proximity Wireless Fields IEC 61000-4-3:2006 +A1:2007 +A2:2010
Rated Power-frequency Magnetic Field IEC 61000-4-8:2009
Group 1 Class B
100MHz – 6GHz
±8kV contact ±2kV, ±4kV, ±8kV, ±15kV air
10V/m Modulation of 80% AM at 1kHz. Frequency 80MHz – 2700MHz
Refer to Table 3
30A/m. Frequency 50Hz and 60Hz
Table 3 – Proximity elds from RF wireless communications equipment
Test Frequency (MHz)
385 18 Hz 27
450 18 Hz 28
710 745 780
810 870 930
1720 1845 1970
2450 217 Hz 28
5240 5500 5785
The following tests are not applicable as the device is battery powered:
• Terminal disturbance voltages CISPR 11:2009+A1:2010
• AC-Mains Harmonics IEC 61000-3-2:2005 +A1:2008+A2:2009
• AC-Mains uctuations and icker IEC 61000-3-3:2013
• Electrical fast transient / bursts IEC 61000-4-4:2012
• Surges IEC 61000-4-5:2005
• Conducted disturbances induced by RF elds IEC 61000-4­6:2013
• Voltage dips IEC 61000-4-11:2004
• Voltage interruptions IEC 61000-4-11:2004
Modulation Complies with
IMMUNITY Level (V/m)
217 Hz 9
18 Hz 28
217 Hz 28
217 Hz 9
Consult
instructions
for use
Part No. PT00147312 Rev A 2021-06
COVIDIEN, COVIDIEN with logo, and Covidien logo and Positive Results for Life are U.S. and internationally registered trademarks of Covidien AG. ™* brands are trademarks of their respective owners. Other brands are trademarks of a Covidien company.
© 2021 Covidien. www.medtronic.com/patents
Covidien llc, 15 Hampshire Street, Manseld, MA 02048 USA.
Medtronic B.V., Earl Bakkenstraat 10, 6422 PJ Heerlen, The Netherlands.
www.covidien.com