Medtronic 29150 Instructions for Use

AccuMist
Blower/Mister
29150
®
Instructions for Use
Caution: Federal law (USA) restricts this
device to sale by or on the order of a physician.
Explanation of symbols on package labeling
Refer to the outer package label to see which symbols apply to this product
Conformité Européenne (European Conformity). This symbol means that the device fully complies with European Council Directive 93/42/EEC.
Nonpyrogenic
Sterilized Using Ethylene Oxide
Do Not Reuse
Use By
Lot Number
Quantity
Open Here
Attention, See Instructions for Use
For US Audiences Only
Manufacturer
12
10
11
9
8
8
7
7
6
5
4
3
2
1
1. Tip
2. Malleable Shaft
3. Handpiece
4. CO2/Medical Air/Fluid “On” Switch
5. CO2/Medical Air/Fluid “Off” Switch
6. Roller Clamp
7. Fluid Line
8. CO2 /Medical Air Line
9. In-line Filter
10. CO2/Medical Air Connector
11. Line Pressure Vent
12. Spike
Figure 1
1
3
7
6
5
2
4
Figure 2
1. Sterile Fluid (not included)
2. Pressure Cuff (mm Hg) (not included)
3. Fluid Line
4. CO2/Medical Air Line
5. Flow Meter (L/min) (not included)
6. CO2/Medical Air Source (not included)
7. Pressure Regulator

Instructions for Use

AccuMist® Blower/Mister
Model
29150

Description

The AccuMist® Blower/Mister consists of a handpiece, a malleable shaft, and a tubing set (CO2/medical air line tubing co-joined to fluid line tubing). The proximal end of the tubing has a spike for fluid attachment and a vented connection for CO2/medical air source attachment. The roller clamp provides a means to adjust fluid flow. The tubing set is approximately 304.8 cm (10 ft) in length. The handpiece has controls for on/off of CO2/medical air/fluid. Sterile, nonpyrogenic, disposable.

Indications

This device is intended for use during procedures when a wound or surgical site must be cleared by non-contact means for improved visibility.

Contraindications

This device is not intended for use except as indicated above. Do not use where the effects of a CO2/medical air stream or irrigation mist are contraindicated.

Warnings

DO NOT USE OXYGEN WITH THIS DEVICE. Use caution when moving the tip of the device near the wound or surgical site. DO NOT ALLOW TIP TO CONTACT TISSUE.

Precautions

To release compressed CO2/medical air in a controlled fashion, point tip of device away from the wound or surgical site before turning device on. This device is designed and intended for single-use only. DO NOT RESTERILIZE OR REUSE.

Adverse Effects

None known.
Instructions for Use
1. Inspect the package and product for damage and expiration date. Open the package and transfer product onto the sterile field utilizing aseptic technique.
2. Push the CO2/medical air connector onto the CO2/medical air source regulator's tapered adapter.
Note: When connected to the CO2/medical air source, an audible sound may be heard at the line pressure vent.
3. Use the following recommended settings:
If Using:
Pressure Regulator Only 83 – 172 kPa (12 – 25 psi) Flow Meter Only 4 – 6 L/min Both Pressure Regulator
and Flow Meter
4. Use the Off switch on the handpiece to turn off the CO2/medical air/fluid.
5. Aseptically connect the spike to a new bag of sterile fluid (not included) and place in a pressure cuff (not included). Inflate the pressure cuff to approximately 150 mm Hg.
Set At:
276 – 345 kPa (40 – 50 psi) and 4 – 6 L/min
4 Instructions for Use English
6. To release compressed CO2/medical air in a controlled fashion, point tip of device away from the wound or surgical site before turning device on. Compressed CO2/medical air results whenever the device is cycled off, then on. This compressed CO2/medical air is released by turning the handle control to the “ON” position.
7. Use roller clamp to adjust fluid flow.
8. Bend the malleable shaft to optimize wound or surgical site access and visualization.
9. Use caution when moving the tip of the device near the wound or surgical site. DO NOT ALLOW TIP TO CONTACT TISSUE.
Instructions for Use English 5
The following disclaimer of warranty applies to United States customers only:

Disclaimer of Warranty

ALTHOUGH THE ACCUMIST® BLOWER/MISTER, HEREAFTER REFERRED TO AS “PRODUCT” HAS BEEN MANUFACTURED UNDER CAREFULLY CONTROLLED CONDITIONS, MEDTRONIC HAS NO CONTROL OVER THE CONDITIONS UNDER WHICH THIS PRODUCT IS USED. MEDTRONIC, THEREFORE DISCLAIMS ALL WARRANTIES, BOTH EXPRESS AND IMPLIED, WITH RESPECT TO THE PRODUCT, INCLUDING, BUT NOT LIMITED TO, ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. MEDTRONIC SHALL NOT BE LIABLE TO ANY PERSON OR ENTITY FOR ANY MEDICAL EXPENSES OR ANY DIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES CAUSED BY ANY USE, DEFECT, FAILURE OR MALFUNCTION OF THE PRODUCT, WHETHER A CLAIM FOR SUCH DAMAGES IS BASED UPON WARRANTY, CONTRACT, TORT OR OTHERWISE. NO PERSON HAS ANY AUTHORITY TO BIND MEDTRONIC TO ANY REPRESENTATION OR WARRANTY WITH RESPECT TO THE PRODUCT.
The exclusions and limitations set out above are not intended to, and should not be construed so as to, contravene mandatory provisions of applicable law. If any part or term of this Disclaimer of Warranty is held to be illegal, unenforceable or in conflict with applicable law, by a court of competent jurisdiction, the validity of the remaining portions of this Disclaimer of Warranty shall not be affected, and all rights and obligations shall be construed and enforced as if this Disclaimer of Warranty did not contain the particular part or term held to be invalid.
The following disclaimer of warranty applies to customers outside the United States:
Disclaimer of Warranty
ALTHOUGH THE ACCUMIST® BLOWER/MISTER, HEREAFTER REFERRED TO AS “PRODUCT” HAS BEEN CAREFULLY DESIGNED, MANUFACTURED AND TESTED PRIOR TO SALE, THE PRODUCT MAY FAIL TO PERFORM ITS INTENDED FUNCTION SATISFACTORILY FOR A VARIETY OF REASONS. THE WARNINGS CONTAINED IN THE PRODUCT LABELING PROVIDE MORE DETAILED INFORMATION AND ARE CONSIDERED AN INTEGRAL PART OF THIS DISCLAIMER OF WARRANTY. MEDTRONIC, THEREFORE, DISCLAIMS ALL WARRANTIES, BOTH EXPRESS AND IMPLIED, WITH RESPECT TO THE PRODUCT. MEDTRONIC SHALL NOT BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES CAUSED BY ANY USE, DEFECT, OR FAILURE OF THE PRODUCT, WHETHER THE CLAIM IS BASED ON WARRANTY, CONTRACT, TORT OR OTHERWISE.
The exclusions and limitations set out above are not intended to, and should not be construed so as to, contravene mandatory provisions of applicable law. If any part or term of this Disclaimer of Warranty is held by any court of competent jurisdiction to be illegal, unenforceable or in conflict with applicable law, the validity of the remaining portion of the Disclaimer of Warranty shall not be affected, and all rights and obligations shall be construed and enforced as if this Disclaimer of Warranty did not contain the particular part or term held to be invalid.
6 Instructions for Use English
Europe
*M931110A001*
Europe/Africa/Middle East Headquarters
Medtronic International Trading Sàrl Route du Molliau 31 Case Postale 84 CH - 1131 Tolochenaz Switzerland Internet: www.medtronic.co.uk Tel. 41-21-802-7000 Fax 41-21-802-7900
Medtronic E.C. Authorized Representative/Distributed by
Medtronic B.V. Earl Bakkenstraat 10 6422 PJ Heerlen The Netherlands Tel. 31-45-566-8000 Fax 31-45-566-8668
Asia-Pacific
Japan
Medtronic Japan Comodio Shiodome 5F 2-14-1 Higashi-Shimbashi, Minato-ku Tokyo 105-0021 Japan Tel. 81-3-6430-2011 Fax 81-3-6430-7140
Australia
Medtronic Australasia Pty. Ltd. Business Address: 97 Waterloo Road North Ryde NSW 2113 Mailing Address: PO Box 945 North Ryde NSW 1670 Australia Tel. 61-2-9857-9000 Fax 61-2-9878-5100
Asia
Medtronic International Ltd. Suite 1602 16/F, Manulife Plaza The Lee Gardens, 33 Hysan Avenue Causeway Bay Hong Kong Tel. 852-2891-4068 Fax 852-2591-0313
Americas
Latin America
Medtronic Latin America 3750 NW 87th Avenue Suite 700 Miami, FL 33178 USA Tel. 305-500-9328 Fax 786-709-4244
Canada
Medtronic of Canada Ltd. 6733 Kitimat Road Mississauga, Ontario L5N 1W3 Canada Tel. 905-826-6020 Fax 905-826-6620 Toll-free in Canada: 1-800-268-5346
United States
Manufacturer: Medtronic, Inc. 710 Medtronic Parkway Minneapolis, MN 55432-5604 USA Internet: www.medtronic.com Tel. 763-514-4000 Fax 763-391-9100 Toll-free in the USA: 1-800-328-2518 (24-hour consultation service)
© Medtronic, Inc. 2004, 2007, 2008 All Rights Reserved M931110A001 Rev. A
Loading...