Medtronic 27664 Instructions for Use

CSF-Ventricular Reservoirs

Skull
Burr hole
To ventricles
push into
t
ab groove
Bend catheter
Twist Drill or
Burr Hole
Ventricular
Catheter
Ventricular
Reservoir
Reservoir Dome
Button Valve
Prefill
Adapter
30 cc Syringe
16-gauge Blunt
Needle Adapter
Cut Here
Preformed Right
Angle Catheter
12 mm Convertible Reservoir
Radiopaque Tantalum Impregnated Silicone Plug
Flow Hole
Catheter
VentricleSkull
Burr Hole
Ventricular Catheter
Snap Assembly Tool
Incision
Integral Ventriculostomy Reservoir Dome
Snap Connector
Snap Shunt Assembly
(underside view)
Dura
Skull Surface
Radiopaque Silicone Dome
PlasticTubing Plug
Plastic Straight Connector
Plastic Base
Radiopaque Silicone Dome
PlasticTubing Plug
Plastic Straight Connector
Plastic Base
25-Gauge or Smaller Noncoring Needle
Straight Connector
Tubing
Description
CSF-ventricular reservoirs are designed to provide percutaneous access to the ventricular cerebrospinal uid (CSF) through a silicone elastomer reservoir connected to a ventricular catheter.
CSF-ventricular reservoirs are available in several models, characterized by inlet connector orientation, dome diameter, and reservoir conguration (Fig 1).
All reser voirs have transl ucent silicone elas tomer domes and are f abricated with a p recision-mo lded, rm polypropylene base invested in a smooth, exible silicone elastomer housing. The rm polypr opylene base contr ibutes to struc tural integr ity, and reduces the p otential for penet ration of the reservoir base by the needle during percutaneous puncture.
All reser voirs have an integ ral catheter conn ector with the e xception of the rese rvoir with inte gral right angl e catheter. The integ ral connector is d esigned to facili tate implantati on and decrease the possibility of catheter disconnection. Only one encircling ligature is required to secure the catheter to the rese rvoir. A radiopaqu e marker is located at t he base of the connec tor, which allows the physician t o visualize by X-ray the re lative positions o f the catheter and r eservoir in vivo.
Convert ible reservo irs may be integr ated with a Medt ronicNeurosur gery CSF shunt s ystem by cut ting o the outle t tube plug and conn ecting the ou tlet tube to a shun t valve or a valved cat heter.

Ventricular Catheters

All Medtr onic Neurosurger y CSF-ventric ular reservoi rs are indicated f or use with Medtr onic Neurosurgery ventricular catheters. See the Ventricular Catheter section o f this insert.
1
Fluid volume capacity of reservoirs (nominal):
30 mm, Side Inle t ........................................... 2.4 m L
28 mm, Side Inle t ............................................1.1 m L
28 mm, Bott om Inlet ......................................1.1 mL
28 mm, Bott om Inlet Convertib le ...................1.1 m L
18 mm, Side Inlet ........................................... 0.3 mL
12 mm, Burr Hole ........................................... 0.6 mL
12 mm, Burr Hole Con vertible ....................... 0.6 mL
12 mm, Convertible ........................................0.1 mL
Indications
All MedtronicNeurosurgery CSF-ventricular reservoirs provide access to the lateral cerebral ventricles, to cyst ic tumors, and to de bulked tumor cavi ties via hypode rmic punctur e for the inject ion of chemotherapeutic agents and/or radioisotopes. Reservoirs are useful in obtaining CSF samples for cytological and chemical studie s, for monitoring ventricular uid pressure, and for ventricular drainage. All conver tible reser voirs may be used as co mponents of Me dtronicNeurosu rgery CSF-sh unt systems.
Instructions for Use
CSF-Ventricular Reservoirs
The placem ent of the reservo ir may be accomplishe d through a variet y of surgical tec hniques. The surgica l method and site of pla cement are at the disc retion of the surg eon. The locatio n selected should pro vide adequate acces s for hypodermic i njection thro ugh the dome. The dome i s designed to allow inje ction via a 25-g auge or smaller nonco ring needle.
Complete p enetration of th e reservoir by the n eedle will be preve nted in models that con tain a rm polypropylene base.
The reser voir may be suture d to adjacent tissue by p assing a suture thr ough the fabric-reinf orced ange or thr ough the suture ho les on the ange of the 12 mm Bur r Hole models.
To adapt the Conver tible models to C SF-shunt compone nts, cut o the outl et tube plug and inse rt a value or ca theter connect or for assembly wi th value and/or dista l catheter (Fig 2). Refer to t he shunt system’s Instructions for Use for further information.
2
Ventricular Catheters–Surgical Technique
For ventricular catheter surgical technique see the Ventricular Catheter sec tion of this inser t.
Converti ble 12 mm Reservoir–Pr e-attached Ve ntricular Cathe ter Placement U sing Stylet
A stainle ss steel styl et and polypropy lene plug are suppli ed with the Conver tible 12 mm Reservoir with inte gral ventricular c atheter model. Th e stylet may be us ed as an aid to place the ca theter in the ventri cle as follows: cut t he radiopaque, ta ntalum-impre gnated silicone out let plug as shown in Figure2. Ins ert the styl et through the out let tube into the res ervoir, and caref ully straighten o ut the angled catheter, pushing the stylet to the proximal end of the catheter. Position the catheter in the ventric le and withdraw the s tylet, allowin g the angled cathe ter section to re turn to its pref ormed shape and th e reservoir to lie  at against the skull.
At this point t he outlet tube of th e reservoir may be p lugged with the po lypropylene tu bing plug, or a valved shun t may be attached . In either case, a sing le ligature is requi red to attach the o utlet tube to a ttin g (plug or straigh t connector).
If the product is packaged with a disposable quick release ventricular catheter introducer, the instru ctions for use ar e as follows:
Insert t he tip of the introd ucer stylet int o the proximal end of th e ventricular cat heter via the drain hole closest to the reservoir (Fig 3).
3
Once the de sired catheter lo cation is achieve d, the shaft grip i s held rmly as the st ylet grip is pulle d back appr oximately 1.5 centime ters (Fig 4). The catheter w ill be released f rom the introduce r by this action . The introducer may n ow be withdrawn.
4
Injec tion into the Rese rvoir
CSF-Ventric ular Reservoi rs are designed to al low injection th rough the dome by use o f a 25-gauge or smaller no ncoring or bevele d needle (Fig5). Reser voirs are designe d to allow multiple pu nctures, and it is rec ommended that the ne edle be insert ed at various loca tions to avoid repeate d punctures at a single point.
CSF sampl ing may be done by percu taneous access th rough the silicone r eservoir dome u sing a 25 gauge, or sm aller, non-coring ne edle. Puncture s can occur as oft en as desired up to 25 tot al punctures.
CAUTION: LOW TEAR S TRENGTH IS A CHARACTERIST IC OF MOST UNREINFORCED SILICONE ELASTOME R MATERIALS. CARE MUST BE TAKEN ON INSERTIO N AND REMOVAL OF THE NEEDLE.
CAUTION: TAKE CARE TO MAINTAIN STERILITY AND TO AVOID PARTICULATE CONTAMINATION.
5

CSF-Ventriculostomy Reservoirs

Description
CSF-Ventriculostomy Reservoirs are designed to provide percutaneous access to the ventricular cerebrospinal uid (CSF). Any MedtronicNeurosurgery CSF-Ventriculostomy Reservoir can be used indepen dently as a reser voir or as a compone nt of a MedtronicNe urosurgery C SF-Flow Control Sh unt.
CSF-Ventric ulostomy Reser voirs have a 6 mm-d iameter burr hole, p olypropylene ba se, and a radiopaque, barium impregnated, silicone elastomer dome.
The rm po lypropylene ba se contributes t o structur al integrity, and re duces the potent ial for penetration of the reservoir base by the needle during percutaneous puncture.
The base is m ade with an integr al catheter conne ctor designe d to facilitate im plantation and de crease the possibility of catheter disconnection. One encircling ligature is required to secure the catheter to the inlet conn ector. The radiop aque reservo ir dome allows the phy sician to visualize b y X-ray the relative positions of the reservoir dome and the radiopaque catheter if disconnection is suspected. All reservoirs are speci ally designed to t i nto a 6 mm (1/4”) diam eter burr hole or t wist drill hole. A ny Medtronic Neurosur gery CSF-Flo w Control valve may be d irectly conne cted to the res ervoirs.
Pull-Over Ventriculostomy Reservoir
The Pull- Over Ventriculos tomy Reservoir (F ig 6) is available in two mo dels: 6 mm Burr Hole Ba se, Shallow, and 6 mm Bur r Hole Base, Regula r Depth.
The uid volume capacity of the reservoirs is:
Burr Hole, 6 m m Regular Depth: .15 m L
Burr Hole, 6 m m Shallow Depth: .10 mL
6
Unitized Ventriculostomy Reservoir
The Unitiz ed Ventriculosto my Reservoir is on e integral assemb ly. It has a 6 mm base, integra l to the reser voir dome. The uid volu me capacity of t he reservoir is 0.15 mL (Fig 7).
Snap Assembly Ventriculo stomy Reservoir
The Snap Assembly Ventriculostomy Reservoir facilitates connection of the proximal and distal componen ts of a CSF-Shunt and sho uld be used with a s nap assembly reservoir base. The uid volume cap acity of the res ervoir is 0.15 mL (Fig 8).
7
Indications
CSF-Ventriculostomy Reservoirs, when connected to ventricular catheters, provide access to the lateral cerebral ventricles. They may be used independently as reservoirs, or as components of CSF­Flow Control Shunts.
Instructions for Use
Ventricular Catheter Placement
A variet y of surgical tech niques may be used in p lacing the ventric ular catheter. Site of p lacement is at the disc retion of the surg eon. The ventricu lar catheter may be pl aced into the lateral ve ntricle using a sta inless steel st ylet as a guide. Af ter placement and tr imming (if necess ary) the catheter is attac hed to the assembl ed reservoir at t he base inlet conne ctor by an encircl ing ligature. The assembl ed CSF-Ventriculos tomy Reservoi r is placed in the 6 mm burr h ole or twist dri ll hole.
Alternately, the catheter length may be gauged before insertion and trimmed. After trimming, the catheter may be xed to the reservoir base by an encircling ligature (non-unitized versions only). The cath eter and base may then b e placed as an assemb ly using a stainles s steel style t. After catheter placement, the silicone elastomer cap is tted over the reservoir base, and the assembled CSF-Ventric ulostomy Reser voir is placed in the b urr hole or twis t drill hole.
CAUTION: DO NOT P UNCTURE THE SILICONE CAP WIT H THE STYLET DURING C ATHETER PLACEMENT. THIS MAY RESULT IN LEAK AGE THROUGH THE CAP AFTER RE SERVOIR IMPLANTATION.
Reservoir Placement
The placem ent of the reser voir can be accompli shed through a var iety of surgic al techniques. P lacement method is a t the discretio n of the surgeon. Th e location sele cted for the res ervoir should p rovide adequate access for i njection thr ough the reser voir dome, and if des ired, allow for the p ossible additi on of a CSF-Flow Control Sh unt. The reser voirs are design ed to t snugly into a 6 mm b urr hole or twis t drill hole.
Use of Reser voir Only
To use the reser voir without con nection to a CSF- shunt, push the plast ic tubing plug into th e reservoir dome outlet and secure with one encircling ligature (Fig 9).
8
Use of Reser voir in a CSF-Shunt
To connect the reservoir for use as a shunt component, push the straight connector into the reservoir dome outlet ar m and secure wit h one encircling li gature. The reser voir is now ready to b e connected to a C SF-Shunt by suturi ng the connecto r to the inlet tubing o f the shunt. A CSF-F low Control Valve or shun t with an integral inlet connector may be connected directly to the reservoir, without using the separate connector (Fig 10).
10
Injec tion into the Rese rvoir
The CSF-Ventr iculostomy Res ervoir is design ed to allow injecti on through the dome b y use of a 25-gauge or smaller noncoring needle.
CAUTION: LOW TEAR S TRENGTH IS A CHARACTERIST IC OF MOST UNREINFORCED SILICONE ELASTOME R MATERIALS. CARE MUST BE TAKEN ON INSERTIO N AND REMOVAL OF THE NEEDLE.
The reservoir is designed to allow multiple punctures. Inserting the needle at various locations to avoid repea ted punctures a t a single point on the c ap is recommended .
Produc t performanc e tests indicate t hat the CSF-Ventricu lostomy Reser voir will tolerate up t o 100 punct ures with a 25-ga uge or smaller noncor ing needle. Actua l product per formance may var y depending upon injection technique, needle type and size.
Snap Assembly Ventriculo stomy Reservoir
NOTE: The Snap Assembly Ventriculostomy Reservoir Dome must be used with a Snap Assembly Base.
NOTE: The Ventricular Catheter with Snap Assembly Base is available separately or as an integral unit with predetermined catheter lengths.
After th e ventricular cat heter has been place d, and attached to t he reservoir b ase, the reservo ir base may be snappe d together with t he reservoir do me. A Snap Assembly Tool is av ailable separate ly to simplif y and expedite t he snap procedure. Sl ip the “prongs” of the to ol under the reser voir base; then, snap th e dome onto the base (Fi g 11). After the dome a nd base are snappe d together, the reser voir is placed in the 6 mm b urr hole or twis t drill hole.
WARNING: VERIFY COMPL ETE SNAP ENGAGEMENT AND DO NOT DISA SSEMBLE THE RESERVOIR AND BASE ONCE THE Y ARE SNAPPED TOGETHER . THE RESERVO IR AND BASE ARE DESIGNED FOR “ONE-TIME-ONLY” SNAP A SSEMBLY. METICULOUS C ARE MUST BE TAKEN DURING ASSEMBLY TO PREVENT DAM AGE TO PLASTIC COMPONENT S. DISASSEMBLY AND REASSEMBLY AND/OR DAMAGE TO THE PL ASTIC COMPONENTS DURING A SSEMBLY CAN RESULT IN UNWANTED DISCONNEC TION OR IN LEAKAGE OF THE ASS EMBLY.
11

Shunt Patency Test, Snap Assembly Ventriculostomy Reservoir

A prell ada pter and a blunt needl e are provided wit h each Snap Shunt Asse mbly Ventriculos tomy Reservoir Dome to facilitate preimplantation processing (Fig 12).
Shunt Pate ncy Check
1. Attach t he translucent pre ll adapter to the inle t of the reservoi r dome.
2. Using the 16-gauge blunt needle adapter, ll a syringe with sterile, ltered, isotonic saline.
3. Insert t he blunt needle ada pter into the port o f the prell adapter, and us ing gentle syrin ge pressure, ush shunt assembly with the saline.
4. If the sali ne ows out of the dist al catheter end, t he shunt assembly is p atent.
Note: Pressure/ow test instruction are available upon request from the Medtronic Neurosurgery Customer Service department.
12
depth of penetration of the catheter into the lateral ventricle.
Ventricula r catheters are p ackaged with a ri ght angle clip (except fo r catheters desi gned for use with bur r hole reservoi rs and valves). The right an gle clip package d with the cathete r is designed to angulate the catheter to lie within the mid-portion of the lateral ventricle.
Each of the c atheters is kin k and compression r esistant, is impr inted with lengt h markers, and is package d with a stainles s steel style t inserted in the l umen. The proximal en d of the catheter ha s a single line of inlet holes spaced at 90º intervals around the circumference of the catheter tubing.
Barium Impregnated Ventricular Catheter
Barium Im pregnated Ventric ular Catheters ar e fabricated f rom relatively r m, barium impregna ted silicone elastomer tubing for radiopacity.
Standard Ventricular Catheters
The Standar d Ventricular Cathe ter has an outer diame ter of 2.5 mm and an inner di ameter of 1.3 mm (dimensions nominal).
Translucent Ventricular Catheter
The Translucent Ventricular Catheter is fabricated from translucent, non-radiopaque silicone elastomer tubing.
Translucen t Ventricular Cath eter with Radio paque Stripe
The Translucent Ventricular Catheter with Radiopaque Stripe is fabricated from translucent silicone elastomer tubing with a radiopaque stripe encased within the tubing wall.
Flanged Ventricular Catheter
The Flanged Ventricular Catheter, available in the small tubing size only, has eight exible silicone elastom er anges (Fig 13) and has 20 inlet ho les positioned a t 90º intervals ar ound the cathete r between the anges (Fig 14).
The anges are designed to bend and cover the inlet holes in the catheter tubing during insertion protec ting the inlet hole s from obstru ction by brain fr agments or bloo d clots. The ange s resume an extended conguration upon entering the ventricle and help prevent direct contact of the inlet holes with the c horoid plexus and v entricle wall.
Snap Assembly Ventricular Catheter
The Snap Assembly Ventricular Catheter is fabricated of silicone elastomer and polypropylene, and is desig ned to be used only wi th a snap assembly re servoir dome. Th e kink-and compress ion­resistant ventricular catheter is fabricated with smooth, radiopaque silicone elastomer. The rm polypr opylene base contr ibutes to struc tural integr ity, and reduces the p otential for penet ration of the reservoir base by the needle during percutaneous puncture.
13 14
Pliant Ventricular Catheter
The Pliant Vent ricular Cathete r is fabricated f rom a softer (lowe r durometer) silicon e elastomer tubing.
Ventricul ar Catheter with O pus
®
Stylet
The Opus Vent ricular Cathete rs are packaged w ith an Opus Style t. The Opus Stylet i s designed to facilit ate introduct ion of the cathete r into the lateral ventr icle. The Opus Sty let has a hollow shaf t, an entry p ort and an exit po rt, permitt ing visualizati on of CSF during cat heter placement (Fig 15). The Opus Sty let has a molded luer lock outlet gri p around the exit po rt. Flow of CSF thro ugh the exit por t may be stopp ed by covering the ex it port with a glo ved nger (Fig 16).
The luerlo ck outlet grip als o allows attachm ent of a syringe fo r CSF sampling.
CAUTION: DUE TO THE LE NGTH OF THE STYLET SHAFT, CSF FLOW THRO UGH THE STYLE SHAFT MAY NOT OCCUR IN PATIENTS WITH LOW INTR ACRANIAL PRESSURE.
CAUTION: CAR E SHOULD BE TAKEN TO MINIMIZE THE AMOUNT OF C SF ALLOWED TO ESCAPE THROUGH THE OPUS ST YLET.
15 16
Indications
Ventricula r catheters are d esigned for use a s proximal compon ents of reser voirs for acces s to cerebros pinal uid. They als o are used as a compo nent of CSF-Flow Con trol Shunts for u se in shunting cerebros pinal uid from t he ventricles of t he brain into the rig ht atrium of the hea rt or to the perit oneal cavit y. The small catheter is r ecommended for u se in patients w here a smaller diame ter is indicated.
Instructions for Use
A variet y of surgical tech niques may be used in p lacing the cathet ers into the lateral v entricle. Site of placement i s at the discretio n of the surgeon.
A stainle ss steel styl et, packaged wi th each catheter, is des igned to facilit ate introducti on of the cathete r into the lateral vent ricle. The cathete r is packaged wit h the stylet ins erted in the lumen . Once the ca theter is placed into t he ventricle, the s tylet is withdr awn from the cath eter.
The right a ngle clip may be used to b end the ventricula r catheter to an appr oximate 90º angle w here it exit s the twist drill o r burr hole (Fig 17). (The cli p must be removed whe n the catheter is use d with the Burr Ho le Reservoir.) The clip m ay be used as a marker fo r planned depth of ca theter insert ion by sliding it t he appropriate dis tance from the pr oximal tip of the cat heter prior to ins ertion; this can b e done with the stylet in the catheter (Fig 18). After the catheter is positioned properly in the ventricle, the ext racranial por tion of the cathe ter is pressed into t he split tubular se gment of the clip to fo rm the right a ngle bend (Figures 19 and 20). Avoid s tretching the ca theter when it is pre ssed into the clip. If the c atheter is to be place d in the ventricle th rough a tubular intr oducer, the clip must be removed prior to insertion of the catheter through the introducer.
17
18
19
20
CAUTION: LOW TEAR S TRENGTH IS A CHARACTERIST IC OF MOST UNREINFORCED SILICONE ELASTOME R MATERIALS. CARE MUST BE TAKEN WITH THE HANDLING A ND PLACEMENT OF THE SILICONE ELAS TOMER CATHETER TUBING TO AVOID CUTS, NICK S, OR TEARS.
Snap Assembly Ventriculostomy Catheter
NOTE: The Ventricular Catheter with Snap Assembly base is available separately or as an integral unit with predetermined catheter lengths.
After th e ventricular cat heter has been place d, and attached to t he reservoir b ase, the reservo ir base may be snappe d together with t he reservoir do me. A Snap Assembly Tool is av ailable separate ly to simplif y and expedite t he snap procedure. Sl ip the “prongs” of the to ol under the reser voir base; then, snap th e dome onto the base (Se e Fig 13, Snap Shunt Assembly). After the do me and base are snappe d together, the reser voir is placed in the 6 m m burr hole or twi st drill hole.
WARNING: VERIFY COMPL ETE SNAP ENGAGEMENT AND DO NOT DISA SSEMBLE THE RESERVOIR AND BASE ONCE THE Y ARE SNAPPED TOGETHER. THE RESER VOIR AND BASE ARE DESIGNED FOR “ONETIME ONLY” SN AP ASSEMBLY. METICULOUS C ARE MUST BE TAKEN DURING ASSEMBLY TO PREVENT DAM AGE TO PLASTIC COMPONENT S. DISASSEMBLY AND REASSEMBLY AND/OR DAMAGE TO THE PL ASTIC COMPONENTS DURING A SSEMBLY CAN RESULT IN UNWANTED DISCONNEC TION OR IN LEAKAGE OF THE ASS EMBLY.

Cardiac/Peritoneal Catheter with Closing Pressure

Description
Cardiac /Peritoneal Cat heter, Closing Pressu re, Low, Medium, or High, is avai lable in a variet y of congurations. They are fabricated from radiopaque silicone elastomer tubing. Relatively rm catheter tubing is incorporated into the catheter design to provide resistance to catheter kinking and occlusio n. Black length mar kers are made of graph ite-impregna ted silicone elast omer material. The shape of the t ip is designed for c ardiac or perito neal placement. Cere brospinal uid ex its the cathete r through t he distal slit open ings. The slit openi ngs are coated with g raphite during ma nufacture to minimize the possibility of slitted surface adhesion during storage. The slit openings are tested to conform t o a specic closin g pressure range. Th ey are not designed as pressure/ow control valves.
They are designed to provide a dened closing pressure range only.
In produc tion, the group ave rage closing pres sure is measured and c ompared to speci cations as an average cl osing pressure f or the lot of cathete rs. The closing pr essure range is ide ntied by a dot code at the di stal end of the cat heter as shown below :
Dot Code Closing Pressure Closing Pressure Range (cmH2O)
Low 1.5 – 5.4
•• Medium 5.5 – 9.4
••• High 9.5 – 14.4
Indications
The Cardi ac/Peritoneal C atheter, Closing Pres sure Low, Medium or High, is d esigned as a dist al componen t of a CSF-Flow Contro l Shunt for use in shun ting cerebrospin al uid (CSF) from the ventric les of the brain into th e right atrium of th e heart or to the per itoneal cavity.
Instructions for Use
A variet y of surgical tech niques may be used to p lace the catheter s into the right atri um, peritoneal cavit y, or other C SF diversion site. Si te of placement is at the d iscretion of th e surgeon. The surge on may trim the p roximal end for the s elected place ment site at the time of su rgery. The cathet er should be c hecked for patenc y at the time of surge ry including vi sual determinatio n that all slit openings will allow ow of cerebral spinal uid.
If the surge on desires to veri fy that the closi ng pressure of the c atheter is within t he range of perfo rmance as labeled, a v ariety of techn iques may be used. Th e method is at the dis cretion of the surgeon.
Closing Pressure Test
CAUTION: TAKE CARE TO MAINTAIN STERILITY AND TO AVOID PARTICULATE CONTAMINATION.
Test Method
I. Equipmen t required for thi s test:
1. Suppor t for catheter
2. Sterile 30 cm w ater manometer, gradua ted in millimeters , with 3-way stopcoc k at base
3. Sterile sy ringe, 30 cc to 50 cc
4. Sterile 5 µ syringe lter
5. Sterile tubing adapter
6. Sterile silicone tubing
7. Catheter Connector
II. Equipme nt setup
1. Set up mano meter so that the zero l evel on the manometer a nd the support f or the catheter are at the same h eight (Fig 21a).
2. Fill syringe with sterile water using 5 µ syringe lter. (When relling syringe, always use 5 µ syringe lter.) After lling syringe, detach syringe lter.
3. Connect syringe, manometer, and silicone tubing as shown in Figure 21a using straight connector as needed.
4. To purge all air f rom the assembled s terile test app aratus, turn sto pcock as shown in Fig ure 22.
5. Flush with s terile water fro m syringe.
X369-1E
2018 - 10
© 2018 Medtronic, Inc.
9
Ventricular Catheters
Description
Ventricula r Catheters are ava ilable in a variet y of conguration s: small and standar d sizes; dierent lengths; w ith or without ang es; barium impreg nated; translucent o r translucent wit h stripe; pliant; and snap ass embly. See package la bels for congura tion details.
The catheter has a bullet-shaped tip lled with radiopaque silicone elastomer. Black length markers m ade of a graphite-i mpregnated silicon e elastomer are posi tioned on the cath eter at points 5, 10, and 15 cm (+/- 0.25 cm) from t he proximal tip to enab le the surgeon to gauge t he
See Snap Assembly Ventriculostomy Reservoir section for snap shunt assembly.
It is recomm ended that the rig ht angle clip be secu red to adjacent tissu e by passing suture s through t he two suture hole s on the sides of the cli p (Fig 17). If the right angle cli p is not used, it is recommen ded that the surge on trim the rim of the t wist drill or bu rr hole to provide a be veled notch where the catheter emerges and curves to lie adjacent to the skull.
Medtronic, Inc.
710 Medtronic Parkway NE Minneapolis, MN 55432 USA +1 800 468-9710 USA/Canada +1 901 344-0645 International +1 800 468-9713 FAX USA/Canada +1 901 396-2698 FAX International medtronic.com manuals.medtronic.com
Polypropylene Handle
Stainless Steel Shaft
Catheter Connection
Polypropylene Obturator Tip
Off
From Syringe
To Catheter
0
Off
From Syringe
To Catheter
0
0 cm
30 cm
Manometer
Stopcock
Syringe Filter
Silicone Tube
Silicone Tubing and Catheter Connector
Off
From Syringe
To Catheter
0
0 cm
30 cm
Manometer
Stopcock
Syringe Filter
Silicone Tube
Silicone Tubing and Catheter Connector
21a
21b
22
23
24
III. Equipment calibration
1.
Turn stopcoc k as shown in Figure 2 3 and ll manometer to a t least 5 cm H2O.
2. With cat heter held at zero level, t urn stopcock to i solate syringe f rom the manometer (F ig 24).
3. Allow water co lumn in the manomete r to fall.
4. Water column sh ould stop at zero leve l of the manometer (Fig 21a).
5. Manomete r is now calibrated to z ero level. Fix or mount t he manometer to maint ain this reference position.
IV. Test procedure
NOTE: For cor rect result s, the zero level of t he manometer mu st be properl y aligned with the distal slit openings of the catheter.
1. Connec t the sterile cath eter to be tested to th e assembled ster ile test apparat us (Fig 21b). The catheter tip should not be touching any substance.
2. Turn stopcoc k as shown in Figure 23 a nd ll manometer to at le ast 30 cm H2O.
3. Turn stopcoc k to isolate manomet er from ow path as sh own in Figure 22.
4. Purge all air f rom the cathete r and the assemble d test apparatus by g ently ushing wit h sterile water fro m syringe.
5. Gently es tablish ow throu gh the catheter wi th sterile water f rom the syringe.
6. While ge ntly maintaining  ow through the cat heter, turn stopco ck to isolate syr inge from ow path as show n in Figure 24. After st opcock is placed in p roper positio n, the water column in t he manomete r should start t o fall. The syrin ge is now isolated f rom the catheter a nd continuing ow with the s yringe is no longe r necessary. If wa ter column does not f all, repeat steps 2 t hrough 6.
7. Allow water column in manometer to stabilize. The column of uid in centimeters from the slit openings (zero level) to t he top of the uid colum n in the manometer is re corded as the clos ing pressure.

Catheter Connectors

Description
Catheter Connector, Straight
The Cathe ter Connector, Strai ght (Fig 25), is a nylon connec tor designed f or use with ventric ular and distal catheters.
Catheter Co nnector, Right An gle
The Cathe ter Connector, Right A ngle (Fig 26), is a nylon connec tor designed fo r use with ventric ular catheters.
Catheter Connector, 3-Way
The Cathe ter Connector, 3-Way (Fig 27), is a pol ypropylene and si licone elastomer co nnector designe d to connect two ve ntricular cathe ters to one CSF-ow c ontrol valve inlet c onnector.
Indications
Catheter Connector, Straight
The Cathe ter Connector, Strai ght, is indicated f or use when distal o r ventricular cat heter revision is require d, or to connect a vent ricular cathet er to the inlet por t of a valve.
Catheter Co nnector, Right An gle
The Cathe ter Connector, Right A ngle, is indicated f or use to angulate the ve ntricular cathe ter to a 90º angle (nomina l) where it exits t he twist drill or b urr hole.
25
26
27
Catheter Connector, 3-Way
The Cathe ter Connector, 3-Way, is indic ated for use when t wo ventricular cat heters are conne cted to one CSF- ow control valve.
Instructions for Use
The surgic al technique emp loyed in the use of cath eter connecto rs is at the discret ion of the surgeon.

Disposable Subcutaneous Catheter Passer, 38 and 60 cm

Description
The Dispo sable Subcutan eous Catheter Pas sers are dispos able surgical ins truments (Fig 28). They a re available i n two lengths, 38 cm an d 60 cm, and consist of t wo components:
1. A tubular sh aft with polyp ropylene D-shap ed handle
2. An obtura tor with bullet-shap ed tip
The instruments are designed as malleable subcutaneous tunneling devices for passing cerebrospinal uid shunt catheters.
28
Indications
The Dispo sable Subcutan eous Catheter Pas sers are used for t he placement of the pe ritoneal cathe ter component of a cerebrospinal uid shunt or for the placement of the lumboperitoneal catheter.
Instructions for Use
The surgic al technique emp loyed in the use of the Di sposable Subcu taneous Cathet er Passers is at the discr etion of the surge on. The surgeon is ad vised to ensure that t he catheter will p ass through the inst rument prior to be ginning the proced ure. The tubular shaf t is malleable and m ay be hand­formed b y the surgeon as indi cated for the oper ative procedure s. When the instr ument is passed subcut aneously, the obtura tor must be inser ted in the tubular sha ft and secure d in the handle by snapping t he obturator’s lock ing tip into positi on. After removal o f the handle and obtu rator, the cathete r may be hand-fed thr ough the lumen of the tu bular shaft.
The obtur ator also includes a c atheter atta chment site for at tachment of an open e nded distal cathete r. Attac hment of the cathe ter to the obturator a llows catheter s of appropriate I.D. (in ner diameter) to b e pulled through th e passer shaft.
How Supplied
Each Med tronic Neurosur gery produc t is packaged ster ile and non-pyrogenic in a double-wr ap packagi ng system and is inte nded for single (one-time) use only. The double w rap system facilit ates the prefer red method of ster ile product tr ansfer from th e circulating area to t he sterile eld. Do not u se if package has b een opened previo usly or damaged. Do n ot resterilize. Re sterilizatio n can damage the product, potentially leading to patient injury. MedtronicNeurosurgery is not respons ible for the perf ormance of any prod uct that has been re sterilized.
Storage Conditions
Store devic es in dry, clean condit ions at normal indo or ambient room tem perature.
Contraindications
CSF-Ventricular Reservoirs should not be implanted if there is infection in the scalp or other areas through w hich the catheter w ill traverse, or if i nfection is pre sent in any area of the b ody. The exceptio n to the foregoing co ntraindicatio n is the use of the devic e to treat inammator y diseases involving the central nervous system, e.g., mycotic or other meningitides or brain abscess.
Shunting o f CSF into the right at rium, peritoneal c avity, or other area s of the body should no t be carrie d out if there is infe ction in any areas i n which the variou s components of the s hunt system will be impl anted. These incl ude infection s of the scalp and oth er sin area through wh ich the shunt system will traverse, the meninges and cerebral ventricles, peritoneum, and intraperitoneal and retrop eritoneal organs , pleura and blood s tream. CSF shuntin g is contraindica ted if there is infec tion present i n any area of the body. Addit ionally, shunting into t he atrium of patien ts with congenit al heart disease or other serious cardiopulmonary abnormalit ies is contraindicated.
Patient Education
It is the resp onsibility of t he physician to educ ate the patient and/or t heir representa tive(s) regarding the use of ventricular reservoirs and CSF shunting. This should include a description of the compli cations associ ated with implant able shunt system s, and an explanatio n of potential alternative products and treatments.
Warnings and Precautions
If the ventr icular catheter b ecomes detach ed from the reser voir, it may be withdra wn from, or lost in, the later al ventricle of the b rain.
Physicians who infuse chemotherapeutic drugs in this device must assure themselves of the compatib ility of the drug w ith silicone elas tomers and polyp ropylene material s. Some change in drug pote ncy may occur when i t is in contacted wi th reservoir an d catheter materi als.
Overdra inage of CSF may predi spose developmen t of a subdural hemato ma or hydroma or collaps e of the lateral ventricular walls leading to obstruction of the ventricular catheter.
The per formance charac teristics o f this device may be alter ed if component s or devices of other manufac turers are use d in conjunction w ith this device.
The appro priate produc t and size must be chos en for the speci c patient’s needs, b ased on diagnos tic tests and phys ician experien ce. Incorrect se lection may lead to c omplications. P roduct labeling species applicable product per formance levels or ranges.
The use of co mponents or dev ices of other manu facturer s in conjuncti on with Medtro nic valves has not been veri ed. The perf ormance chara cteristic s of this device may b e altered if comp onents or devic es of other manufacturers are used in conjunction with this device. The perfor mance chararcterstics are veried for all Med tronic shunt com ponents, whic h are designed to be i nterchangeab ly connected w ith each other.
Avoid contac ting implantab le products w ith lint, glove talc, o ily residue from s kin, oil based soa ps, synthetic detergents or other surface contaminants.
Improper use of instruments in the handling or implantation of reservoir products may result in the cutti ng, slitting or cr ushing of compone nts. Such damage may le ad to loss of reser voir integrit y, and neces sitate premature s urgical revisio n of the reservoi r system. Do not ben d or fold the device during ins ertion beca use this may cause de vice damage.
Care must be taken to ensure that particulate contaminants are not introduced into shunt components during preimplanation testing or handling. Introduction of contaminants could result in improper performance (overdrainage or underdrainage) of the shunt system. Particulate mater that enters the shunt system also may hold pressure/ow-controlling mechanisms open, resulting in overdrainage.
In securing catheters to connectors, the encircling ligatures should be fastened securely but not too tightly, least they eventually cut through the silicone tubing.
Care must be taken in the routing of catheters to prevent kinking and needless abrasion along their course. T he rim of the twis t drill or burr hole may b e trimmed to provi de a beveled notch whe re the ventric ular catheter emer ges and is curved t o lie adjacent to the skul l.
Patients with hydrocephalus shunt systems or implantable ventricular reservoirs must be kept under close obs ervation in the p ostoperative p eriod for signs an d symptoms that sug gest malfunc tion. The clinical ndings may indicate infection or other complications.
Obstr uction may occur i n any of the component s of the ventricula r reservoir or shu nt system. The system m ay become internall y occluded due to tis sue fragments, b lood clots, tumo r cell aggregates, bacter ial colonizatio n, or other debris. C atheters that con tact internal bo dy structu res can become kinked, or b locked at their tips (e.g., in vestment of a ventr icular catheter t ip into the choroid p lexus). Finally, obst ruction may occ ur due to growth of an i nfant or child, or phy sical activi ties, which resul t
in disconn ection of the com ponents, or with drawal of a distal c atheter from it s intended access si te.
Ventricular reservoir systems may fail due to mechanical malfunction.
If the ventr icular catheter b ecomes bound to th e choroid plexus o r adjacent brain tis sue by brous tissue adh esions, it is sugge sted that it should n ot be forcibly remo ved. It is suggeste d that gentle rotatio n of the catheter may he lp to free it. It is adv ised that the cath eter be left in pl ace rather than risk intraventricular hemorrhage, which may be caused by forcible removal.
“Small” si ze catheters have th inner walls and lower ove rall strength as co mpared with “Sta ndard” size cathe ters. These cha racteristic s result in a compa ratively greater po tential failure (f racture) rate and, there fore, shorter li fe expectan cy for “Small” siz e catheters. Phys icians who implant “ Small” size cathe ters for cosmet ic reasons must ack nowledge the pot entially higher rate o f catheter revisi on and weigh thi s against the cosme tic benets.
Clotti ng around the atria l portion of a cat heter may lead to embol ization of the pulm onary arteri al tree with resulting cor pulmonale and pulmonary hypertension.
Disconnected shunt components may further migrate into the heart, or into the peritoneal cavity.
Shunt systems may fail due to mechanical malfunction, leading to underdrainage or overdrainage.
Malfunc tion or obstr uction of the shun t system may lead to sig ns and symptoms of in creased intracranial pressure if the hydrocephalus is not compensated. In the infant, common ndings are increas ed tension of the anter ior fontanelle, co ngestion of sc alp veins, listless ness, drowsiness a nd irritability, vomiting, and nuchal rigidity. Other children and adults will develop signs and symptoms commonly a ssociated with i ncreased intra cranial pressur e such as headaches, vo miting, blurri ng of vision, nuchal rigidity, deterioration of consciousness and variable abnormal neurological ndings.
Subcut aneous catheter p assers can brea k at welds or componen t assembly point s, or as a result of extr eme deformatio n of the malleable shaf t. Sudden break age can lead to traum a of tissues or organs, an d damage to the implant able system. Ins truments mus t be inspected p rior to use to ensure continued integrity and functionalit y. Disposable instruments must never be reused, or injury to the patie nt and physician is po ssible.
Care should be taken to avoid plugging the holes in the ventricular catheter with brain fragments during it s passage throu gh the cerebral cor tex into the ventr icle. The tip of the ca theter should not be advan ced anterior to the f oramen of Monro if it i s introduced into t he lateral ventric le via a parieto -occipital a pproach. This may c ause the catheter t o become embedde d in the frontal lob e if the ventri cle becomes redu ced in size.
It has been r eported that a ver y small populatio n of patients may show a n acute allergic-type reacti on to the ventricula r reservoir sy stem. This may be due to i ts materials of fab rication, and can lead to patient discomfort tissue erosion, or other complications.
Complications
Complica tions associate d with the use of the se products may b e similar to those ex perienced in any surgica l procedure carr ied out under loc al and/or general anes thesia. These in clude reactio ns to drugs and a nesthetic agen ts, electrol yte imbalance, car diac arrhythm ias, and excessive blo od loss, parti cularly in infant s. The patient may ra rely exhibit a reac tion due to sensi tivity to the imp lant.
Additional complications associated with neurosurgical procedures include seizure, hemorrhage, and neurologic decit/dysfunction.
Use of non- ltered or non-co ntrolled uids fo r testing may resul t in improper devic e performance a nd necessitate device revision. Devices must be handled in controlled hospital environments to prevent the intro duction of par ticles, bers o r other contaminan ts into the valve.
In CSF shunt ing procedures, t he most common comp lications are due t o obstructio n of the system as descr ibed under “Warni ngs and Precauti ons.” Obstruct ion may occur in any comp onent of the system d ue to plugging by brain f ragments, blo od clots, and/or tum or cell aggregates a t some point along its c ourse. Obstr uction also may occ ur because of sep aration of the sys tem components o r kinking and/or coiling of the catheter. This may predispose migration of the ventricular catheter into the late ral ventricle and th e distal cathet er into the heart an d pulmonary ar terial tree, the peritoneum, or other structure in which the catheter is implanted. Migration into the pleural cavity may lead to pleu ral eusion. Shun t disconnecti on may also occur due to g rowth of an infant o r child, or physic al activities w hich result in disc onnection of th e shunt component s or withdrawal of a di stal cathete r from its intend ed drainage site. Dis connection may c ause complicat ions. Disconnec ted shunt components may migrate.
Local and systemic infections due to organisms inhabiting the skin, particularly Staphylococcus epidermidis, may occur. Other p athogens circu lating in the bloods tream may colonize t he shunt and, in the maj ority of patie nts, require it s removal. Rigorous p erioperative co ntrol of the opera ting environm ent and the use of antib iotics at the phys ician’s discretion may av ert infecti on occurrence.
Infec tion rates may be dec reased with the us e of antibiotics , short duratio n of surgery (surgi cal experience) and control of the operating room environment (e.g., designated operating room, limited pe rsonnel and tra c, covered skin sur faces). Results als o can be obtained wi thout the use of antibiotics, but with rigorous perioperative control of the environment.
The use of pr ophylactic ant ibiotics is some what controversi al because their u se may predispose infec tion by resistan t organisms. There fore, the decisio n to use antibiotic s prophylactic ally rests with the at tending physici an and/or surgeon.
CSF overdr ainage may result in e xcessive reduct ion of intracran ial pressure and pr edispose the developm ent of a subdural hema toma or hygroma, and exce ssive reducti on of ventricular si ze leading to obstruction because of impingement of the ventricular walls on the inlet holes in the cathe ter. In rare cases, spi nal abnormalitie s have been associa ted with overdrain age. In infant s, excessive press ure reduction d ue to overdrainage wi ll cause a marked dep ression of the anterior fontanelle, overriding of cranial bones and may convert communicating into obstructive hydrocephalus.
Intra-abdominal complications associated with peritoneal shunting include perf oration of the small or large bowel with resultant peritonitis, perforation of other viscera, ureteral obstruction, adhesions, migration of the catheter into unintended anatomical locations, and development of ascites and pseudocysts.
CSF overdr ainage has in rare ca ses been associ ated with spinal abn ormalities such a s myelopathy.
Neurosur gical procedur es such as CSF shunti ng carry risk s associated wi th brain tissue mani pulation, includin g neurologic dec it/dysfunc tion, seizure, and s troke. Other surg ical risks incl ude bleeding complications, inadvertent introduction of air into the pleural space (pneumothorax), and wound healing problems such as wound dehiscence.
In rare cas es, seeding of tum or cells along cathe ter lengths has be en reported.
Returned Goods Policy
Produc ts must be retur ned in unopened p ackages, with man ufacturer ’s seals int act, to be accepte d for replac ement or credit, u nless returned d ue to a complaint of pro duct defec t or mislabeling
Determi nation of a produc t defect or misl abeling will be made by M edtronic Neuro surgery, which determination will be nal.
Produc ts will not be accept ed for replacement o r credit if they have b een in possessio n of the custom er for more than 90 days .
Warranty
A. Standa rd Limited Warr anty. Medtronic Ne urosurgery wa rrants to the ori ginal end user
purchaser (“Purchaser”) that the enclosed single use product (“Product”) purchased by Purchaser, at the time of d elivery to Purch aser, shall be substant ially free from d efects in mate rial and workmans hip. Medtronic Neu rosurgery make s no warranty (exp ress, implied or s tatutory) for Produc ts that are modi ed (except as express ly contemplated her ein) or subjected to un usual physical stress, misuse, improper operation, neglect, improper testing, use in combination with other pro ducts or compo nents other than th ose for which the Pr oducts were de signed, or use in any manner or medical procedure for which the Products are not indicated.
B. Remedy. Purc haser’s exclusi ve remedy and Med tronic Neurosu rgery’s sole lia bility for br each of the foregoi ng warranty s hall be, at Medtron ic Neurosurger y’s sole option a nd election, t o replace the Prod uct or credi t Purchaser for t he net amount ac tually paid for any s uch Product , provided that (i) Me dtronic Neurosur gery is noti ed in writing wi thin ninety (90) day s after Purcha ser’s receipt of th e Product that s uch Produc t failed to confo rm, including a d etailed expl anation in English o f any alleged nonco nformity; (i i) such Prod uct is return ed to Medtronic N eurosurger y within ninet y (90) days after Pur chaser’s receip t of the Produc t F.O.B., 125 Cremona D rive, Goleta, Ca lifornia 93117, USA., or as other wise design ated by Medtron ic Neurosurgery; and (iii) Medtronic Neurosurgery is reasonably satised that the claimed nonconformities actual ly exist. Exce pt as expressl y provided in this p aragraph, Purc haser shall not have t he right to retur n Products to Medtronic Neurosurgery without Medtronic Neurosurgery’s prior written consent.
C. Exclu sion of Othe r Warranties. E XCEPT FOR THE LIMITED WARRANT Y PROVIDED IN (A) ABOVE, MEDTRONI C NEUROSURGERY GRANTS NO O THER WARRANTIES OR CONDITIONS, E XPRESS OR IMPLIED AND MANUFACTURER SPECIFICALLY DISCLAIMS THE IMPLIED WARRAN TIES AND CONDITIONS OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. MEDTRONIC NEUROSURGERY NEITH ER ASSUMES NOR AUTHORIZES A NY OTHER PERSON TO ASSUME AN Y OTHER LIABILITIES ARISING OU T OF OR IN CONNECTION WITH THE SALE O R USE OF ANY PRODUCT.
CSF-Ventricular Reservoir
Products and Accessories
Instructions for use

Symbols

Caution
Use by
Sterilization: Ethylene-Oxide Gas
Do Not Reus e
Do Not Resterilize
Sterilize
Reference Number
Lot Number
Package Contents
Cautio n: U. S. federal law re strict s this device to s ale by or on the ord er of a physicia n.
Manufacturer
Consult Instructions for Use.
Do not use i f package is op en or damaged.
The foll owing is a trade mark or regis tered trade mark of Medtr onic, Inc. in th e United States a nd other count ries: Delta™. All ot her trademar ks, servi ce marks, reg istered tr ademarks or r egistered s ervice mar ks are the pro perty of their re spective ow ners in the Uni ted States and o ther countri es.
Loading...