Medtronic 227F5 Instructions for Use

FREEZOR® XTRA 227F1, 227F3, 227F5
Cardiac CryoAblation Catheter
Technical Manual
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
The following list includes trademarks or registered trademarks of Medtronic in the United States and possibly in other countries. All other trademarks are the property of their respective owners. Freezor, Medtronic

Explanation of symbols

Lot number
Reorder number
Use by
Sterilized using ethylene oxide
Do not reuse
Do not resterilize
t
Do not use if package is damaged
Do
Package contents
Cardiac cryoablation catheter
Consult instructions for use
Fragile: handle with care
Keep dry
Product documentation
Humidity limitation
Storage temperature
Transit temperature
Open here
Manufacturer

1 Description

The Freezor Xtra Cardiac CryoAblation Catheter is a flexible, steerable catheter used to ablate cardiac tissue. It is used together with the CryoConsole and related components. The tip of the Freezor Xtra catheter reaches cryoablation temperatures when refrigerant is injected from the CryoConsole to the tip of the catheter. The catheter tip has an integrated type T thermocouple for temperature reading capability. The Freezor Xtra catheter is available in 3 models, as described in the following table:
Model Curves available
227F1 49 mm
227F3 55 mm
227F5 60 mm
For details about the CryoConsole and how to use it with the catheter to perform cryoablation procedures, see the CryoConsole Operator’s Manual.

1.1 Contents of package

The Freezor Xtra Cardiac CryoAblation Catheter is supplied sterile. The package contains the following items:
1 Freezor Xtra Cardiac CryoAblation CatheterProduct documentation

2 Indications for use

The Freezor Xtra Cardiac CryoAblation Catheter, CryoConsole system, and related accessories are indicated for the cryoablation of the conducting tissues of the heart in the treatment of patients with atrioventricular nodal reentrant tachycardia (AVNRT). The Freezor Xtra catheter is also intended for minimally invasive cardiac surgery procedures, including surgical treatment of cardiac arrhythmias. The Freezor Xtra catheter freezes the target tissue and blocks the electrical conduction by creating an inflammatory response or cryonecrosis.

3 Contraindications

The Freezor Xtra Cardiac CryoAblation Catheter is contraindicated in patients with the following conditions:
Active systemic infectionsCryoglobulinemiaOther conditions where the manipulation of the catheter would be unsafe (for example,
intracardiac mural thrombus)

4 Warnings and precautions

Cryoadhesion – Do not pull on the catheter, sheath, umbilical cables, or console while the
catheter is frozen to the tissue, as this may lead to tissue injury. Improper connection – Do not connect the cryoablation catheter to a radio frequency (RF) generator or use it to deliver RF energy. Doing this may cause catheter malfunction or patient harm. RF ablation – Before powering up an RF generator or applying RF energy, disconnect the cryoablation catheter from the CryoConsole to avoid an error message and unnecessary catheter replacement. Pressurized refrigerant – The catheter contains pressurized refrigerant during operation. Release of this gas into the body or circulatory system due to equipment failure or misuse could result in gas embolism, pericardial tamponade, tissue emphysema or other patient injury.
Induced arrhythmias – Catheter procedures may mechanically induce arrhythmias. Fluoroscopy during catheter placement – The use of fluoroscopy during ablation
procedures presents the potential for significant x-ray exposure to both patients and laboratory staff. Extensive exposure can result in acute radiation injury and increased risk for somatic and genetic effects. Only perform ablation after giving adequate attention to the potential radiation exposure associated with the procedure, and taking steps to minimize this exposure. Give careful consideration before performing ablation procedures on or near pregnant women. Endocardial catheter positioning around the chordae tendinae – Avoid positioning the catheter around the chordae tendineae, as this increases the likelihood of catheter entrapment within the heart, which may necessitate surgical intervention or repair of injured tissues. Endocardial cryoablation near prosthetic heart valves – Do not pass the catheter through a prosthetic heart valve (mechanical or tissue). The catheter may become trapped in the valve, damaging the valve and causing valvular insufficiency or premature failure of the prosthetic valve. Do not resterilize – Do not resterilize this catheter for purpose of reuse. Resterilization may compromise the structural integrity of the catheter or create a risk of contamination from the catheter that could result in patient injury, illness, or death. For single use only – This catheter is intended only to be used once for a single patient. Do not reuse, reprocess, or resterilize this catheter for purpose of reuse. Reuse, reprocessing, or resterilization may compromise the structural integrity of the catheter or create a risk of contamination of the catheter that could result in patient injury, illness, or death.
Catheter handling –
Do not use excessive force to advance or withdraw the catheter, especially if resistance
is encountered.
Use extreme care when manipulating the catheter. Lack of careful attention can result in
injury such as perforation or tamponade.
Do not use the catheter if it is kinked, damaged, or cannot be straightened.Straighten the cooling segment before inserting or withdrawing the catheter. Do not at any time preshape or bend the catheter shaft or cooling segment. Bending or
kinking the catheter shaft may damage internal structures and increase the risk of catheter failure. Prebending of the distal curve can damage the catheter.
Catheter advancement should be performed under fluoroscopic guidance. Biohazard disp osal – Discard all used catheters and sterile components in accordance with hospital procedures. Leakage current from connected catheters – Use only isolated equipment (IEC 60601-1 Type CF equipment, or equivalent) with the CryoConsole and catheters or patient injury or death may occur. System compatibility – Use only Medtronic cryoablation catheters, refrigerant tanks, and components with the CryoConsole. The safety and use of other catheters or components has not been tested. Sterile package inspection – Inspect the sterile packaging and catheter prior to use. If the sterile packaging or catheter is damaged, do not use the catheter. Contact your Medtronic representative. Catheter integrity – Do not use the catheter if it is kinked or damaged. If the catheter becomes kinked or damaged while in the patient, remove it and use a new catheter. Prior to injecting, the physician should ensure that there is no kink in the catheter. Required use environment – Cryoablation procedures should be performed only in a fully equipped facility. Qualified users – This equipment should be used only by or under the supervision of physicians trained in surgical or endocardial cryoablation procedures. Fluid incursion – Do not expose the catheter handle or coaxial and electrical connectors to fluids or solvents. If these components become wet, the cryoablation system may not function properly, and connector integrity may be compromised.
Freezor Xtra Technical Manual English 3
Cardioversion/defibrillation during ablation procedure – Disconnect the catheter's electrical connection prior to cardioversion/defibrillation. Failure to do so may trigger system messages indicating a need for catheter exchange. Coronary vessels – Cryoablation involving coronary vessels with liquid nitrous oxide systems has been associated with subsequent clinically significant arterial stenosis. Care should be taken to minimize unnecessary contact with coronary vessels during cryoablation. Anticoagulation therapy – Administer appropriate levels of peri-procedural anticoagulation therapy for patients undergoing endocardial right-sided procedures. Administer anticoagulation therapy before and after the procedure according to the hospital standards. Embolism risk – Introducing any catheter into the circulatory system entails the risk of air or gas embolism, which can occlude vessels and lead to tissue infarction with serious consequences. Always advance and withdraw components slowly to minimize the vacuum created and therefore minimize the risk of air embolism.

5 Adverse events

Potential adverse events associated with cardiac catheter cryoablation procedures include, but are not limited to, the following conditions:
Access site complications (e.g.,
hematoma, infection, thrombosis, ecchymosis, AV fistula, bleeding from puncture site, hemorrhage)
Arrhythmia (including new or worsening
existing arrhythmias)
Cardiac arrestCardiac tamponade/perforationCatheter entrapment in cardia c structures
requiring surgical intervention
Chest discomfort, pain or pressureCoronary artery spasm, dissection,
thrombosis
Damage to adjacent organs/structuresDeathEndocarditisHeart block, partial or complete,
potentially requiring permanent pacemaker
HematomaHemothoraxInfection/sepsisMyocardial infarctionPericardial effusionPericarditisPleural effusionPneumothoraxPseudoaneurysmPulmonary edemaPulmonary embolismStoke/transient ischemic attack/embolismThrombosisValvular damageVascular complication (e.g., stenosis)Vasovagal reaction

6 Clinical study

6.1 Study design

The Freezor Xtra Catheter, Cryoablation CryoConsole system, and related accessories were evaluated in a clinical study for the treatment of atrioventricular nodal reentrant tachycardia (AVNRT) using an endocardial approach. The study was designed as a prospective multicenter, nonrandomized, single arm, unblinded, investigational clinical study to assess the safety and effectiveness of the catheter used percutaneously in the treatment of AVNRT. There were 34 total centers (8 in Canada, 26 in the US) that enrolled subjects. Subjects with a study procedure attempt were followed for a minimum of 6 months and up to 3 years post study cryoablation procedure. Postprocedure follow-up visits included phone follow-up at 8 days, in-person visits at 1, 3, and 6 months, and annual phone follow-up visits at 1, 2, and 3 years. Unscheduled visits occurred, as needed, for the continued evaluation of the safety and effectiveness of the device in the treatment of AVNRT in enrolled subjects.

6.2 Study endpoints

The primary endpoints for the study were as follows: Chronic safety - defined as the absence of serious complications associated with the catheter or use of the catheter during the procedure or within the six months post-ablation Chronic efficacy - defined as the following:
At the conclusion of the ablation procedure assessment: inability to induce more than one
echo beat by the same pacing maneuvers and drug provocation that induced AVNRT prior to the procedure
During the 6 months post ablation: no documented recurrence of clinical AVNRT Chronic safety and effectiveness were analyzed by calculating at a one-sided 0.025 significance level to meet the following pre-specified performance goals through 6 months of follow-up:
7% of subjects experiencing at least one safety event
83% of subjects achieving chronic effectiveness success

6.3 Description of subjects

A total of 572 subjects were enrolled in the study. Of the 572 enrolled subjects, 397 met study criteria and underwent cryoablation, which comprised the cohort used for primary endpoint analyses. Of the 175 ineligible subjects, 22 subjects exited the study prior to the EP study, 150 had an EP study but did not receive cryoablation due to not meeting post-EP inclusion or meeting post-EP exclusion criteria, 2 subjects met all inclusion and no exclusion criteria but were not treated due to an inability to induce AVNRT, and 1 subject was treated despite being ineligible for the study.
Subject description N
Consented 572
Exited prior to EP study 22
EP study 550
Exited at conclusion of EP study (not eligible) 150
Not exited at conclusion of EP study (treated, but not eligible) 1
Intent-to-treat (ITT) could not reproduce AVNRT 1
ITT exited due to extremely long time to induce AVNRT 1
Analysis Populations N
ITT:
All subjects who have signed the ICY-AVNRT consent form and have met all pre-EP and post-EP study inclusion and no exclusion criteria.
Modified Intent-to-Treat (mITT):
All subjects within the ITT set who had cryoablation application(s) applied with a Freezor Xtra cardiac cryoablation catheter.

6.4 Baseline demographics

The majority of subjects enrolled in the study were female (279/397; 70.3%) with a median age of 54 years. A large majority of subjects were of White or Caucasian Race (353/397;
88.9%).
Subject Characteristics N (%)
Gender (N, %) 397 (100.0%) Male 118 (29.7%) Female 279 (70.3%) Age (years)
399
397
4 English Freezor Xtra Technical Manual
N 397 (100.0%) Mean +/- Standard Deviation 52.6 +/- 3.9 Median 54.0 25th Percentile – 75th Percentile 44.0 - 62.0 Minimum – Maximum 18.0 - 84.0 Race/ Ethnic Origin (N, %) Subject/physician chose not to provide information 10 (2.5%) Not reportable per local laws or regulations 0 (0.0%) American Indian or Alaska Native 1 (0.3%) Asian 10 (2.5%) Black or African American 16 (4.0%) Hispanic or Latino 4 (1.0%) Native Hawaiian or Pacific Islander 0 (0.0%) White or Ca ucasian 353 (88.9% ) Two or more races 1 (0.3%) Other race 2 (0.5%)

6.5 Medical history AVN RT:

Subjects must have had an SVT compatible with AVNRT documented by ECG, TTM, Holter, or event monitor for inclusion in the study. The average SVT cycle length was 345.2 ± 65.2 ms with QRS duration of 86.7 ± 24.0 ms. No subject had exclusionary AV block (Atrioventricular block [1° (PR interval 220 ms), 2°, 3°] or left bundle branch block) present at baseline. Some form of bundle branch block presence was reported in 16 subjects (16/397; 4.0%), with the majority of those (10/397;
2.5%) reporting right bundle branch block. Four (4) subjects (1.0%) had electrical alternans present.

6.6 Cryoablation procedure data Procedure time:

The average total procedure time (period of time between puncture of the skin to obtain venous access for the EP study to the final ECG performed at the end of the cryoablation procedure) was 149.1 ± 55.5 minutes (median 142.0 minutes). Fluoroscopy time for the total procedure was 11.51 ± 9.4 minutes (median 9.0 minutes). The ablation procedure time is the time from first ablation catheter insertion to last ablation catheter removal. The portion of the procedure in which ablations were performed averaged
89.48 ± 40.3 (median 80.0 minutes). These times include a 30-minute waiting period to evaluate slow pathway recovery. Fluoroscopy time from the first cryoapplication to the last cryoapplication averaged 4.8 ± 5.9 minutes (median 3.1 minutes) in the ablated subjects.
Procedure Characteristics Index Procedures
Total procedure time (Min) N 397 (100.0%) Mean +/- Standard Deviation 149.1 +/- 55.49 Median 142.0 25th Percentile – 75th Percentile 109.0 - 180.0 Minimum – Maximum 52.00 - 417.0 Total fluoroscopy time (Min) N 394 (99.2%) Mean +/- Standard Deviation 11.51 +/- 9.36 Median 9.00 25th Percentile – 75th Percentile 5.70 - 14.40 Minimum – Maximum 0.02 - 95.98 Total ablation procedure time (Min) N 397 (100.0%) Mean +/- Standard Deviation 89.48 +/- 40.32 Median 80.00 25th Percentile – 75th Percentile 62.00 - 110.0 Minimum – Maximum 24.00 - 390.0 Total ablation fluoroscopy time (Min) N 367 (92.4%) Mean +/- Standard Deviation 4.78 +/- 5.89 Median 3.10 25th Percentile – 75th Percentile 1.10 - 6.20 Minimum – Maximum 0.02 - 57.00
Energy applications:
Of 3628 energy applications, the majority reported sinus rhythm (3037, 83.7%) during the application. The average application duration was 146.5 ± 97.5 seconds and temperature was -76.2 ± 12.4 C. Ablation locations were defined as 6 numbered zones within Koch’s triangle (the area bounded by the coronary sinus orifice, tendon of Todaro, and atrioventricular node); locations and energy application percentages as follows:
1 – Superior (6.8%) 2 – Mid (45.0%) 3 – Inferior (34.9%) 4 – Below coronary sinus ostium (3.4%) 5 – Antrum of coronary sinus ostium (9.3%)
6 – Coronary sinus (0.5%) The majority of Freezor Xtra Catheter energy applications (1634/3628, 45.0%) were at Location 2 (Mid), followed by inferior (Location 3, 1267/3628, 34.9%).
AV block assessment:
No subjects received a permanent pacemaker due to AV block as a result of the study procedure or Freezor Xtra Catheter. There were 3 (3/397, 0.76%) adverse events of 1st degree AV block in 3 subjects, first observed at the pre-discharge visit. These events were adjudicated by the Adverse Events Adjudication Committee (AEAC) as not serious, and related to both the Freezor Xtra Catheter and the cryoablation procedure. Those 3 subjects had 1st degree AV block observed during at least one subsequent follow-up visit over the 6 month follow-up period.
Acute procedural success:
Following a postprocedure 30-minute waiting period, investigators assessed procedural success (the inability to induce more than one echo beat by the same pacing maneuvers that induced AVNRT before cryoablation). Acute procedural success was reported in 95.2% (378/397) of cases. Of the 378 acute procedural successes, 71.7% had no remaining AV node echo beats.
Freezor Xtra Catheter Acute Procedural Success Acute Procedural
Number of remaining AV node echo beats 0 271 (71.7%) 1 107 (28.3%) Freeze-thaw-freeze performed at the successful site? Yes 357 (94.4%)
(N = 397)
Successes (N=378)
Freezor Xtra Technical Manual English 5
Did the physician report that the catheter location changed between the 2 applications?
No 353 (98.9%) Yes 4 (1.1%)

6.7 Safety Results

Cryoablation treatment was performed on 397 mITT subjects, of which 4 had a primary safety event through 6 months of follow-up. A primary safety event is defined as the occurrence of any adverse event that meets the following criteria:
Adjudicated by the AEAC as being serious and study ablation procedure-related and/or
Freezor Xtra Catheter-related that results in death, a life-threatening illness or injury, permanent impairment of a body function or permanent damage to a body structure
Necessitates significant intervention, such as major surgery or even intravenous medical
therapy (e.g. vasopressors), to prevent permanent impairment of a body function or permanent damage to a body structure
Requires in-patient hospitalization or a prolongation of an existing hospital stay All 4 primary safety events were classified as serious and procedure-related. None of the 4 events was considered related to the Freezor Xtra Catheter.
1. Cardiac tamponade: adjudicated by the AEAC to be related to the EP procedure and diagnostic catheter.
2. Incision site hemorrhage: adjudicated by the AEAC to be related to the procedure and the sheath.
3. Pulmonary embolism: adjudicated by the AEAC to be related to the procedure (occurred 25 days postprocedure).
4. Device difficult to use: adjudicated by the AEAC to be related to the EP procedure and the diagnostic catheter.
The rate of primary safety events through 6 months was 1.0%, which is below the predefined performance goal of 7.0%.
Table 1. Primary safety objective
Chronic safety Kaplan-Meier
Treated subjects with at least one chronic safety event through 6 months
1
Kaplan-Meier log-log confidence interval

6.8 Effectiveness

Subjects must have met both of the following acute and chronic conditions to be considered a chronic effectiveness (treatment) success:
Acute success: The inability to induce more than one echo beat by the same pacing
maneuvers that induced AVNRT before cryoablation (with drug provocation if required for induction before cryoablation) at the conclusion of the study cryoablation procedure assessment
Chronic success: Lack of documented recurrence of clinical AVNRT during the 6-month
follow-up period after the study cryoablation procedure Cryoablation treatment was performed on 397 mITT subjects, of which 29 had a chronic effectiveness failure through six months of follow-up. There were 19 failures due to acute failure (at the conclusion of the ablation procedure assessment, performed 30 minutes post­procedure), and 10 due to recurrence of AVNRT documented between procedure and 6 months post procedure. Freedom from AVNRT through 6 months was 92.6%, which is above the pre-defined performance goal of 83.0%.
Table 2. Primary effectiveness objective
Chronic effectiveness Kaplan-Meier
Treated subjects without acute procedural failure or AVNRT recurrence through 6 months
1
Kaplan-Meier log-log confidence interval
estimate at 6 months
(%)
1.0% (0.4%, 2.7%) <0.0001
estimate at 6 months (%)
92.6% (89.5%, 94.8%) <0.0001
95% Two-sided
confidence
interval
95% Two-sided confidence
1
interval
1
P-value for
Ha: PS < 0.07
P-value fo r Ha: PE > 0.83

7 Instructions for use

7.1 Connect the catheter

To connect the catheter, follow these steps. (For more detailed instructions, refer to the CryoConsole Operator’s Manual.)
1. Connect the auto connection box to the CryoConsole.
2. Connect the catheter to a sterile coaxial umbilical cable and a sterile electrical umbilical cable.
3. Connect the coaxial umbilical cable to the CryoConsole, and connect the electrical umbilical cable to the auto connection box. Note: Use adequate signal clipping on distal ablation channel to allow continuous monitoring of other signals on the EP recording system.

7.2 Perform cryoablation

To perform cryoablation, follow these steps. (For CryoConsole operating instructions, refer to the CryoConsole Operator’s Manual.)
Notes:
– If an unanticipated event occurs, the injection of refrigerant can be stopped at any
time by pressing the Stop Current Action button on the console control panel, by pressing the Stop CryoAblation button on the screen, or by pushing down on the foot switch.
– Because of the complexity of arrhythmia electrophysiology, fluoroscopy and ECG
guidance are commonly used to assess catheter tip position in AVNRT ablations.
– The catheter tip can be deflected to facilitate positioning by using the thumb knob to
vary tip curvature. Pulling the thumb knob back causes the catheter tip to bend; pushing the knob forward causes the tip to straighten.
– Prior to introducing the catheter into the patient, test the deflection mechanism to
ensure that it is operational.
– This procedure applies to both surgical and endocardial AVNRT approaches.
1. Follow aseptic techniques to introduce and position the catheter at the intended cryoablation site, using a compatible introducer or sheath or using appropriate surgical methods. Comply with the following practices: – Use fluoroscopic and electrogram guidance for endocardial catheter placement. – Ensure good tip contact. Note: Cryoadhesion will ensure that the catheter tip does not move during energy delivery.
2. Set the treatment duration on the CryoConsole screen. Note: The default preset duration is 240 seconds. Physicians may modify the preset cryoablation duration based on clinical judgment.
3. Perform cryoablation. Note: Do not manipulate the catheter during therapy.
6 English Freezor Xtra Technical Manual
4. Wait for the cryoablation to finish; however, the injection may be terminated immediately, in cases such as the following: – if AV nodal dysfunction occurs – if AV block occurs – if the desired cryoablation effect is not achieved
Notes:
– To terminate the injection before the preset cryoablation duration, press the Stop
Current Action button on the console control panel, press the Stop CryoAblation button on the screen, or push down on the foot switch.
– Evaluate slow pathway effect early during the freeze.
5. For endocardial AVNRT ablation: If interruption of the slow pathway is observed, perform a second cryoablation at the successful site (confirmed by fluoroscopy, electrograms or 3D mapping), unless AV block or dysfunction occurred with the initial cryoablation. Note: After the catheter reaches body temperature, ensure that there is no shaft deflection before removing the catheter from the point of cryoablation.
6. Ensure that the catheter is not adhered to the tissue, and then remove the catheter from the point of cryoablation.
7. If necessary, perform additional consolidation lesions or cryotherapy treatments; otherwise, remove the catheter from the patient.

8 Specifications

Catheter shaft size 2.5 mm (7 Fr)
Recommended introducer size 2.7 mm (8 Fr) minimum
Tip length 6 mm
Shaft length 108 cm
Number of electrodes on tip 4
Spacing between electrodes 2 mm, 5 mm, 2 mm
Number of thermocouples 1
Curves available 227F1 - 49 mm
Environmental parameters
Recommended transit temperature
Recommended storage temperature
Operation 15 °C to 30 °C (59 °F to 86 °F) at altitudes less than
Refrigerant flow
Flow ± 200 sccm 2550 sccm

9 Medtronic limited warranty

For complete warranty information, see the accompanying limited warranty document.

10 Service

Medtronic employs highly trained representatives and engineers located throughout the world to serve you and, upon request, to provide training to qualified hospital personnel in the use of Medtronic products. Medtronic also maintains a professional staff to provide technical consultation to product users. For more information, contact your local Medtronic representative, or call or write Medtronic at the appropriate telephone number or address listed on the back cover.
227F3 - 55 mm 227F5 - 60 mm
-35 °C to 58 °C (-31 °F to 136 °F); up to 85% relative humidity (non-condensing)
15 °C to 30 °C (59 °F to 86 °F)
2400 meters (8000 feet) above sea level
Freezor Xtra Technical Manual English 7
8 English Freezor Xtra Technical Manual
Medtronic, Inc.
710 Medtronic Parkway Minneapolis, MN 55432 USA www.medtronic.com +1 763 514 4000
Manufacturer
Medtronic CryoCath LP 9000 Autoroute Transcanadienne Pointe-Claire, Quebec H9R 5Z8, Canada +1 514 694 1212 Technical Supp ort +1 877 464 2796
Europe/Middle East /Africa
Medtronic International Trading Sàrl Route du Molliau 31 Case Postale 84 CH-1131 Tolochenaz Switzerland +41 21 802 7000
Technical manuals
www.medtronic.com/manuals
© 2017 Medtronic M426506B001 1B 2017-02-01
*M426506B001*
Loading...