DISASSEMBLY/ASSEMBLY
M708348B486E Rev. A
INSTRUCTIONS FOR DIVERGENCEL™ INSERTERS
2015-08-07
DISASSEMBLY INSTRUCTIONS:
Refer to the Medtronic Reusable Instruments Instructions for Use prior to reprocessing this instrument. After the instrument is
disassembled, refer to the Medtronic Reusable Instruments Instructions for Use.
DISASSEMBLY
1. Start pulling Inserter Shaft (Item 2) out of Handle Inserter (item 1) until it stops, then turn counter clockwise until it is loose.
Then, separate the Inserter Shaft from the Handle Inserter.
2. Hold the Guide (Item 3) with one hand and the Handle Inserter with the other hand, then unthread the Handle Inserter from
the Guide.
3. Separate the Guide from the Handle Insterter.
Item Part # Description
1 2151100 Handle Inserter
2 2151200 Inserter Shaft

Item Part # Description
3
OLIF
3
ALIF
2150210 O51 Screw Guide 10mm
2150212 O51 Screw Guide 12mm
2150214 O51 Screw Guide 14mm
2150216 O51 Screw Guide 16mm
2150218 O51 Screw Guide 18mm
2150220 O51 Screw Guide 20mm
2150200 O51 Simple IB Inserter
2150400 O51 Freehand inserter
2160210 ALIF Screw Guide 10mm
2160212 ALIF Screw Guide 12mm
2160214 ALIF Screw Guide 14mm
2160216 ALIF Screw Guide 16mm
2160218 ALIF Screw Guide 18mm
2160220 ALIF Screw Guide 20mm
2160200 ALIF Simple IB Inserter
2160400 ALIF Freehand inserter
ASSEMBLY
1. Insert Guide (Item 3) into Handle Inserter (Item 1), and then thread the Guide and Handle Inserter together.
2. Insert Inserter Shaft (Item 2) into Handle Inserter until it stops, then turn clockwise until it falls the rest of the way into the
handle.
3. Handle Inserter, Inserter Shaft, and Guide in the assembled condition.
©2015 Medtronic Sofamor Danek USA, Inc. All rights reserved.