Software Errata Sheet for the Model 9986 Software Version 1.5.1 for the Medtronic 2090 CareLink Programmer:
The following table lists certain conditions that may occur during use of the Model 9986 Software Version 1.5.1. The table also provides the Resolution for these conditions.
When pacing artifacts are turned on, the artifacts will appear on the ECG traces in the various places where traces are displayed: Live Rhythm Monitor, Frozen Strip Window, Real-Time Chart Recorder. However, the artifacts will not appear on reports, such as the Frozen Strip report.
Programmer Overheat message fails to appear. If the programmer overheats, the expected behavior is that the programmer will shut down. If the user turns the programmer back on, the user will get no message indicating why the programmer shut down.
Pacing artifacts do not appear on printed reports.
The programmer shuts itself down, and the user reboots but receives no feedback regarding what caused the programmer to shut down.
Check the displayed waveforms for pacing artifact information.
Allow the programmer to cool down and then turn the programmer back on.