Medtronic 2090 Reference Guide

MEDTRONIC CARELINK® 2090
Programmer
Reference Manual
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
MEDTRONIC CARELINK® 2090
Reference Manual
A guide for setting up and using the Medtronic CareLink 2090 Programmer.
CareLink, Jewel, Marker Channel, Medtronic, Medtronic CareLink, Paceart, RemoteView, Reveal, SessionSync, Vitatron
Medtronic MEDTRONIC CARELINK® 2090

Contents

1 Introduction to the programmer ....................................... 8
1.1 Explanation of packaging and product symbols ........................... 8
1.2 About this guide .................................................... 11
1.3 Description and intended use ......................................... 11
1.4 Contraindications ................................................... 12
1.5 Warnings .......................................................... 12
1.6 Precautions ........................................................ 13
1.7 Declaration of Conformity ............................................ 15
1.8 Regulatory compliance .............................................. 15
1.9 Programmer functions ............................................... 16
1.10 Security features of the programmer ................................... 18
1.11 Software requirements ............................................... 19
1.12 Obtain technical manuals ............................................ 19
2 Set up the programmer .............................................. 21
2.1 System components ................................................. 21
2.2 Basic setup ........................................................ 26
2.3 Connect peripheral devices .......................................... 32
2.4 Use external printers ................................................ 33
2.5 Install printer paper .................................................. 36
2.6 Printer buttons ...................................................... 38
2.7 Tear off a printout ................................................... 38
2.8 Low paper supply ................................................... 39
3 Configure the programmer ........................................... 40
3.1 Display screen features .............................................. 40
3.2 About the Between Patient Sessions tool palette ........................ 43
3.3 View and update programmer location and hardware information .......... 44
3.4 Adjust programmer time and date ..................................... 45
3.5 Select audible tones ................................................. 46
3.6 Check the software version ........................................... 47
3.7 Select other software ................................................ 48
Reference Manual 5
Medtronic MEDTRONIC CARELINK® 2090
3.8 Remove other software applications ................................... 48
3.9 Improve the detection of pacing artifacts ................................ 49
3.10 Start the Demonstrations option ....................................... 49
4 Update programmer software using the Software Distribution
Network ........................................................... 51
4.1 The Software Distribution Network ..................................... 51
4.2 Connect to the SDN using a network connection ........................ 51
4.3 Connect to the SDN using a dial-up connection ......................... 55
5 Conduct a patient session ........................................... 63
5.1 Prepare for a patient session ......................................... 63
5.2 Initiate a patient session ............................................. 71
5.3 Emergency VVI button ............................................... 75
5.4 End a patient session ................................................ 77
5.5 Store components .................................................. 77
6 Manage session data and reports ..................................... 78
6.1 Session data ....................................................... 78
6.2 Reports ............................................................ 78
6.3 Save to a PDF file ................................................... 78
6.4 Save to diskette ..................................................... 79
6.5 Save to USB ....................................................... 79
6.6 View reports that are saved to media ................................... 80
6.7 Set the interval for report deletion ..................................... 81
6.8 Manage patient data privacy .......................................... 83
6.9 Vitatron Manual-Guided Reset ........................................ 86
7 RemoteView™ Programmer Consultation ............................. 87
7.1 About RemoteView .................................................. 87
7.2 RemoteView status icon ............................................. 87
7.3 Use RemoteView ................................................... 88
7.4 Data privacy ........................................................ 90
8 SessionSync (Optional) .............................................. 91
8.1 About SessionSync ................................................. 91
8.2 Configure SessionSync .............................................. 91
6 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090
8.3 Enable and disable SessionSync ...................................... 92
8.4 SessionSync Status icon ............................................. 93
8.5 Use Automatic SessionSync .......................................... 94
8.6 Use Manual SessionSync for supported devices ........................ 95
8.7 SessionSync error message descriptions ............................... 96
8.8 View SessionSync Status screen ...................................... 97
8.9 Update SessionSync status .......................................... 97
8.10 Test the SessionSync network connection .............................. 97
9 Service the programmer ............................................ 100
9.1 Clean the system components ....................................... 100
9.2 Sterilize the programming head, ECG cable, and lead wires .............. 100
9.3 Replace a PC card ................................................. 101
9.4 Programmer specifications .......................................... 102
9.5 Special notice ..................................................... 105
9.6 Medtronic limited warranty .......................................... 105
Index ................................................................... 106
Reference Manual 7
Medtronic MEDTRONIC CARELINK® 2090

1 Introduction to the programmer

1.1 Explanation of packaging and product symbols

Refer to the package label and product to see which symbols apply to this product.
Conformité Européenne (European Conformity). This symbol means that the device fully complies with European Union Acts.
The use of this device might be subject to individual country licensing regimes in Europe.
Caution
Consult instructions for use
System meets the applicable Canadian and U.S. IEC safety standards.
Type BF applied part
Type CF applied part
Serial number
Temperature limitation
For U.S. audiences only.
Off
8 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090
On
Wireless communication enabled
Do not dispose of this product in the unsorted municipal waste stream. Dispose of this product according to local regulations. See http://recycling.Medtronic.com for instructions on proper disposal of this product.
RF transmitter
Caution: Strong magnet
The product fully complies with the Australian Communications and Media Authority (ACMA) and the New Zealand Ministry of Economic Develop­ment Radio Spectrum Management standards for radio communications products.
VGA monitor
Battery
Diskette
Network connection port
USB port
PCMCIA card slot
Microphone port
Reference Manual 9
Medtronic MEDTRONIC CARELINK® 2090
Headphones port
Authorized representative in the European community
Alternating current
Date of manufacture
Manufacturer
Reorder number
Lot number
Humidity limitation
Package contents
Programmer, software installed
Product documentation
Accessories
Magnetic Resonance (MR) Unsafe
Software series number
10 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090
UL recognized component
China RoHS
Caution: Federal Law (USA) restricts this device to sale by or on the order of a physician
The product complies with both Canadian and U.S. requirements for meet­ing UL safety standards

1.2 About this guide

This guide describes the features and functions of the Medtronic CareLink 2090 Programmer (referred to as the “programmer”).
Note: Screen images in this guide are for reference only. The content and presentation may vary depending on user selections, desktop, and device being interrogated.

1.3 Description and intended use

The Medtronic CareLink 2090 Programmer is a portable, line-powered (AC) microprocessor-based system with software to interrogate and program Medtronic and Vitatron implantable devices. Other features include:
Automated software updates using a dial-up or local area network (LAN) connection, depending on the hardware configuration. This connection allows the programmer to program new devices and to provide new features as they become available.
A large, bright screen that is adjustable for viewing when sitting or standing.
Keyboard to make entering information easier.
Fast printing speed of 50 mm/s on recorder paper.
ECG recording and diagnostic data reporting. Refer to the technical manuals supplied with the software and hardware accessories for specific details.
The Medtronic CareLink 2090 Programmer should be used by healthcare personnel only in a clinical or hospital environment.
Reference Manual 11
Medtronic MEDTRONIC CARELINK® 2090

1.4 Contraindications

There are no known contraindications to the use of the 2090 Programmer when not in a device application. For contraindications specifically related to the implantable devices that can be interrogated or programmed by the programmer, see the instructions for use for the implanted device.

1.5 Warnings

These warnings apply in general to using the programmer for programming implantable device parameter settings. For more information related to specific implantable device models, see the reference guides for the implantable device and the programmer software.
Damage due to impact – Do not use the programmer if it has sustained impact damage. Internal components may be damaged or exposed. Use of damaged equipment may impact user or patient safety.
Defective equipment – If technical and safety inspection reveals a defect that could harm the patient, clinicians, or third parties, the programmer should not be used until it has been properly repaired. The operator must immediately notify Medtronic or Vitatron of these defects.
Diagnostic ECG – Do not use the programmer ECG display for recording or diagnosis. Use a separate ECG device if recording or diagnostic ECG capabilities are required.
Equipment compatibility – The programmer must be used only for interrogating and programming compatible Medtronic or Vitatron implantable devices. If the programmer is used on other implanted devices, direct stimulation through energy coupling may occur. The programmer is not compatible with programmable devices of other manufacturers.
Flammable anesthetic mixture – The programmer is not suited for use in the presence of a flammable anesthetic mixture.
Importance of reference documentation – Implantable device programming should be done only after careful study of the reference guide for the implantable device and after careful determination of appropriate parameter values based on the patient’s condition and pacing system used. The implantable device reference guide contains a complete description of implantable device operation and important information, such as indications for use, contraindications, warnings, and precautions. The instructions contained in this reference guide and the reference guide supplied with the programmer software are limited to the mechanics of setting up the programmer and selecting the correct options for the desired programming function. Improper use of the programmer could result in erroneous or inadvertent programming and improper operation of telemetry and measurement functions.
12 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090
Internal electrodes – Do not connect the programmer to wires or electrodes internal to the body. The programmer is designed to be medically safe only when attached to surface electrodes.
Magnetic Resonance (MR) Unsafe – The programmer is MR Unsafe. Do not bring the programmer into Zone 4 (magnet room), as defined by the American College of Radiology.
Measurement function – The programmer is also designed to detect and measure pulse rate, AV interval and pulse width, and implantable device artifacts. The device takes these digital measurements with the assistance of optional skin electrodes. Medtronic and Vitatron make no claims or warranties as to the effectiveness of the programmer as a diagnostic tool to the physician.
Modification of equipment – Do not modify this equipment. Modifications may reduce system effectiveness and impact user or patient safety.
Telecom voltage limitation – When using a modem or combo card, make sure that the telecom voltage does not exceed 125 V. Excessive voltage may damage the programmer.

1.6 Precautions

VGA monitor use – To protect against interference or surge/leak currents, the use of a
secondary VGA monitor must meet an applicable safety standard such as UL 60950-1 or IEC 60950-1. The user is responsible for the safety of the resulting medical electrical system.
ECG cable integrity – Upon opening the package, if the ECG cable appears damaged, do not use it. Contact your local Medtronic or Vitatron representative.
Care in handling ECG cable wire – Do not pull on the insulated cable wire to disconnect the cable. Tension on the insulated cable wire may result in damage to the cable.
Electrocautery/external defibrillation – Do not position the programming head over an implanted device during electrocautery or external defibrillation procedures.
Do not immerse – Take care to prevent liquid from entering the programmer and programming head. Do not immerse the programmer or any accessories in any liquid or clean them with aromatic or chlorinated hydrocarbons.
Autoclaving – Do not autoclave the programming head or ECG cable and lead wires.
Electromagnetic interference (EMI) – The programming head has been tested for
compliance with industrial and medical EMI regulations. Any use outside the patient environment may result in the programming head malfunctioning.
Radio-frequency (RF) interference – Portable and mobile RF communications equipment can interfere with the operation of the programmer. Although this system has been approved, there is no guarantee that it will not receive interference or that any particular transmission from this system will be free from interference.
Reference Manual 13
Medtronic MEDTRONIC CARELINK® 2090
Damaged equipment – If the case of the programmer is cracked or if any of the connectors are damaged, contact your Medtronic or Vitatron representative. If there is insulation damage to the power cord or accessory cables or if any of the wall or equipment plugs are damaged, replace the part and dispose of it according to local regulations or return the part to Medtronic.
Electrode quality – Use of high-quality silver/silver chloride (Ag/AgCl) electrodes can minimize the occurrence of small DC voltages that can block the ECG signal. Use electrodes that are fresh and from the same box. Prepare the patient’s skin according to the directions provided with the electrodes.
Avoid damage from programming head – Keep the programming head away from any device or material that will be damaged by the magnetic field, including magnetic media, watches, and other electronic devices.
Product and packaging labels and information – If labels or information appear to be missing from the product or packaging, contact your local Medtronic representative at the address and telephone number located on the back cover of this document.

1.6.1 Environmental precautions

To ensure safe and effective operation, use the device with care to avoid damage to the programmer from environmental factors that may impair its function. Care is exercised in design and manufacturing to minimize damage to devices under normal use. However, electronic devices are susceptible to many environmental stresses including, but not limited to, the following examples.
The unit is designed to be used indoors in a clinic or hospital.
The unit should not be dropped or mishandled in such a manner as to cause physical damage to the unit. This may impair device function. Even if the unit works immediately after being dropped, operational damage may have occurred that may not be observed until some future time.
Fluid should not be spilled on the unit. Even though care is exercised in design and manufacture of the unit to minimize leakage, fluid incursion may occur, which could impair functioning of the unit.
The programmer may be affected by electrostatic discharge (ESD). In an environment likely to cause ESD, such as a carpeted floor, discharge any charge collected on your body before touching the device.
Printers and other connected office equipment should be placed at least 1.5 m (5 ft) from the patient environment.
14 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090
Electrically-operated medical devices, such as the programmer require special care (in terms of electromagnetic compatibility) when being installed. Refer to the accompanying insert: Electromagnetic Compatibility Declaration.
Do not open the device. The programmer is constructed to minimize risk from environmental factors. Opening the unit may make the unit susceptible to environmental factors and may expose the patient or user to hazardous voltage or current.
Rapid temperature changes may affect proper operation. Always allow the temperature to stabilize in the environment in which the device is used before using the device.
Prolonged storage or operation of the device in high humidity may affect proper operation.
If there is any concern that damage has occurred, the unit should be returned to Medtronic or Vitatron for inspection and any needed repair.
Besides these listed examples, various other environmental factors may impair proper performance of the unit in the hospital setting. Always use good health management practices to prevent environmental damage to the unit.

1.7 Declaration of Conformity

Medtronic declares that this product is in conformity with the essential requirements of Directive 1999/5/EC on Radio and Telecommunications Terminal Equipment and Directive 90/385/EEC on Active Implantable Medical Devices (AIMD).
For additional information, contact Medtronic or Vitatron at the telephone numbers and addresses provided on the back cover.

1.8 Regulatory compliance

1.8.1 Industry Canada

ID:3408D-MICS
Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device.
This device may not interfere with stations operating in the 400.150-406.000 MHz band in the meteorological aids, meteorological-satellite, and earth exploration-satellite services and must accept any interference received, including interference that may cause undesired operation.
Reference Manual 15
Medtronic MEDTRONIC CARELINK® 2090

1.8.2 Australia/New Zealand

The device fully complies with Australian Communications and Media Authority (ACMA) and the New Zealand Ministry of Economic Development Radio Spectrum Management standards. The C-tick symbol indicates that the product complies with the applicable EMC/Radio standard for radio communications products.

1.8.3 US Federal Communications Commission (FCC)

FCC ID:LF5MICS (for programmer)
FCC ID:LF59767 (for programming head)

1.8.4 The following provision applies to the low frequency communications system in the device:

This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. The user is cautioned that changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.

1.8.5 The following provision applies to the UHF communications system in the device:

This transmitter is authorized by rule under the Medical Device Radio Communications Service (47 C.F.R. Part 95) and must not cause harmful interference to stations operating in the 400.150 - 406.000 MHz band in the Meteorological Aids (i.e., transmitters and receivers used to communicate weather data), the Meteorological Satellite, or the Earth Exploration Satellite Services and must accept interference that may be caused by such aids, including interference that may cause undesired operation. This transmitter shall be used only in accordance with the FCC Rules governing the Medical Device Radio Communications Service. Analog and digital voice communications are prohibited. Although this transmitter has been approved by the Federal Communications Commission, there is no guarantee that it will not receive interference or that any particular transmission from this transmitter will be free from interference.

1.9 Programmer functions

The following list summarizes some of the programmer functions. Specific functions depend on the implantable device model being programmed or monitored and the software installed.
16 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090

1.9.1 Programming functions:

Permanent and temporary adjustment of parameter values.
Selection of nominal parameter values established by Medtronic, Vitatron or by the user.
Emergency buttons for VVI pacing.

1.9.2 Telemetry functions:

Automatic detection of the device model, and automatic application start-up, if the programming head is in proper position when the programmer is turned on.
Automatic confirmation of a programmed change.
Reporting of currently programmed parameter values in effect and battery status of the implanted device.
Reporting of real-time measurements of implantable device operating parameters such as battery voltage, output energy, etc.
Display and printout of Marker Channel telemetry to simplify EGM analysis.
Display and printout of an atrial and/or ventricular intracardiac electrogram (EGM) taken from the electrodes of the implantable device lead system.

1.9.3 ECG and other diagnostic functions:

ECG window on programming and telemetry data screens provides a continuous view of the patient’s ECG.
Full-window ECG display including a freeze option and an amplitude adjustment feature; ECG display includes Marker Channel telemetry, EGM waveforms, or both when available.
Continuous multi-channel recording (ECG and Marker Channel telemetry or ECG and EGM, for example).
Stimulation threshold test functions.
Direct measurement of pulse rate, AV interval, and pulse width.
Temporary implantable device inhibition.
Printout of programmed and measured information for permanent record.
Reference Manual 17
Medtronic MEDTRONIC CARELINK® 2090

1.9.4 Software update function:

Automated software updates using a network connection, depending on the hardware configuration. This connection allows the programmer to program new devices and to provide new features as they become available.
Updates available from Medtronic personnel.
Clinical software applications that have Uninstall Software capability may be removed using the programmer desktop.

1.10 Security features of the programmer

Good security practices are needed to protect patient data and the integrity of any network-connected product. The programmer incorporates features that facilitate management of security. These features work in conjunction with the security practices of hospitals and clinics to provide safe and secure operation of the programmer and protect the attached network.

1.10.1 How the programmer promotes security

All installed software has been approved by Medtronic. It is not possible to install general purpose software on the programmer. Controlling installed software minimizes the potential for vulnerabilities. Internal software that runs the programmer is locked from change. Every time the programmer is started, a clean version of the installed software is used.
The length of time that patient data can be stored on the programmer is limited. When patient data is removed from the programmer, it is completely erased so that it is no longer recoverable.
The programmer limits how it communicates on a network. When communicating on a network, the programmer uses industry-accepted protocols for authenticating servers and encrypting transmitted data. Only required network connections are open and they are only open when being used. Network communications are originated by the programmer. Unauthorized software is not permitted to originate communications with the programmer.
Unsupported hardware, including unsupported USB devices, is ignored by the programmer and is not accessed.
Medtronic continues to work with its partners to analyze emerging threats and evaluate potential impact on the programmer.
18 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090

1.10.2 What hospitals and clinics can do to promote the security of programmers

Maintain good physical controls over the programmer. Having a secure physical environment prevents access to the internals of the programmer.
Only connect the programmer to managed, secure networks.
Update the software on the programmer when Medtronic updates are available.

1.10.3 What to do if you suspect the programmer has been compromised

If you believe that the programmer has been compromised by a security threat, turn off the programmer, disconnect it from the network, then restart the system. Discontinue use of the programmer if it does not behave as expected. Contact your Medtronic or Vitatron representative for further assistance.

1.11 Software requirements

The programmer requires software from Medtronic and Vitatron to operate. Once installed, the software remains on the programmer hard drive.
Medtronic and Vitatron periodically update the software to add functions to the programmer.
The programmer will not operate properly without the appropriate software installed. If the programmer does not operate properly, check the version of software that is loaded on the programmer, and update it if necessary.

1.12 Obtain technical manuals

Medtronic technical manuals, including the manual you are reading, are available from the Medtronic Manual Library website listed on the back cover of this manual. The Medtronic Manual Library website offers real-time access to the latest versions of Medtronic technical manuals. Technical manuals can be viewed online, downloaded for viewing or printing, or ordered from the website.
New manuals are added to this site regularly. If you do not find the technical manual you want, contact your Medtronic or Vitatron representative.
Printed copies of technical manuals can be ordered. Refer the Medtronic Manual Library website for more information.
Reference Manual 19
Medtronic MEDTRONIC CARELINK® 2090

1.12.1 Access the Medtronic Manual Library website

To access the Medtronic Manual Library website and locate technical manuals, do the following:
1. Point your browser to the address listed on the back cover of this manual.
2. Select the United States.
3. Follow the instructions on the Medtronic Manual Library website to locate, view, print, or order technical manuals.
20 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090

2 Set up the programmer

2.1 System components

Figure 1. Programmer components - front view

1 Telephone cord (not supplied) 2 Emergency VVI button 3 Microphone jack 4 Display screen 5 Headphone jack 6 Power cord 7 Printer paper 8 Printer controls 9 Keyboard cover
10 Keyboard 11 Disk drive, PC card cover 12 Reference guide 13 Programming head (ordered separately) 14 Touch pen 15 Electrode leads 16 ECG cable with plug 17 Ethernet cable (not supplied)
Note: Only accessories approved by the manufacturer should be used.
Display screen – Display can be positioned smoothly from closed to nearly horizontal.
Programming options are selected on the screen with the touch pen.
Reference Manual 21
Medtronic MEDTRONIC CARELINK® 2090
Emergency VVI button – Used to deliver bradycardia VVI operation.
Microphone jack – Intended for future use.
Headphone jack – Intended for future use.
Keyboard cover – Slides forward to protect the keyboard.
Keyboard – Used to enter information.
Printer controls – Select paper speeds of 12.5, 25, or 50 mm/s. Push a button once to
select printing speed. Push it again to stop printing. The Paper Advance button allows the user to align the paper properly.
Telephone cord – Connects the programmer modem to a telephone jack. The telephone cord must be 26 wire gauge minimum. (Not supplied by Medtronic.)
Ethernet cable – Used to connect the programmer to the clinic’s network. The Ethernet cable must be Category 5 or better. (Not supplied by Medtronic.)
Printer paper – Paper for the internal printer.
Programming head – Provides the communication link between the programmer and the
patient’s implantable device. The programming head contains a strong permanent magnet, radio-frequency (RF) transmitter and receiver, and light array. It must be held over the implantable device during a program or interrogate operation. (Ordered separately; not supplied with the programmer.)
Touch pen – Used to select options on the display screen. Predetermined options are selected by applying the pen to the screen.
Electrode leads/ECG cable – Connects the programmer to skin electrodes on the patient for ECG and measurement functions requiring surface detection of cardiac and implantable device signals. Five color-coded lead wires connect the cable to standard, disposable skin electrodes applied to the patient. (Not supplied by Medtronic.)
Note: If you are using a five-lead cable with a plug, the plug can be removed for five-lead ECG applications.
Reference guide – Programmer Reference Guide, provides information about setting up the programmer and between session features.
Disk drive, PC card cover – Provides access to the disk drive and the PC card. If applicable to your programmer hardware, also provides access to USB port(s) and either Integrated Ethernet or parallel connector.
Power cord – Connects the programmer to an AC power outlet.
22 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090

Figure 2. Front connectors (keyboard turned up)

1 Programming head (yellow marker) 2 Analog Input/Output (green marker)
3 ECG cable (black or blue marker)
Programming head – The programming head connector has a yellow marker.
Analog Input/Output – Allows an external monitor or recorder to be connected to the
programmer. This connector has a green marker.
ECG cable – Connects the ECG cable to the programmer. This connector has a black or blue marker.

Figure 3. Left view

1 ON/OFF switch 2 Cooling fan
3 Expansion slot (shown with Analyzer
installed)
4 Printer
ON/OFF switch – Controls power (AC) to the programmer. Once the programmer is turned off, wait 2 seconds before turning it on again.
Cooling fan – Internal fan provides continuous airflow to prevent the internal circuitry from overheating.
Reference Manual 23
Medtronic MEDTRONIC CARELINK® 2090
Expansion slot – Allows for additional features to be added, such as the Analyzer that is available as an option.
Printer – Integral thermal printer with text and graphic output capabilities. According to the selected function, the printer provides data reports or it can print out a continuous ECG with accompanying Marker Channel telemetry, EGM, or both when available.

Figure 4. Right view

1 Disk drive, PC card cover 2 Infrared window
Disk drive, PC card cover – Provides access to the disk drive and the PC card. If applicable to your programmer hardware, also provides access to USB port(s) and either Integrated Ethernet or parallel connector.
Infrared window – Intended for future use.

Figure 5. Disk drive, with parallel connector, PC card cover open

1 PC card slot 2 Parallel connector
3 Disk drive
24 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090

Figure 6. Disk drive, with USB port, PC card cover open

1 PC card slot 2 Disk drive
3 Integrated Ethernet 4 USB port
PC card slot – There are different types of PC cards that can be plugged into the PC card slot:
The Modem card.
The Combination Ethernet/Modem Network card, referred to as the combo card.
These PC cards can be used to connect to the Software Distribution Network or to connect the programmer to a consultant’s computer.
Warning: When using a modem or combo card, make sure that the telecom voltage does not exceed 125 V. Excessive voltage may damage the programmer.
Parallel connector – Allows a printer to be connected to the programmer.
Disk drive – Accommodates a 90 mm formatted diskette that is IBM-compatible.
Integrated Ethernet – Allows the programmer to connect to the Software Distribution
Network and the Paceart data management system using an Ethernet connection.
USB port(s) – Allows installation of software, software updates, and future device application installations. The USB port can also be used to connect to a USB printer, a USB flash drive, or a USB network adapter (contact a Medtronic service representative).
Reference Manual 25
Medtronic MEDTRONIC CARELINK® 2090

Figure 7. Back view (power cord door open)

1 USB port(s)
3 Power cord
2 VGA output port
VGA output port – Allows porting the screen image of the programmer to an external VGA monitor or for conversion of the output signal to NTSC/PAL format for presentation on a television monitor.
Power cord – Connects the programmer to an AC power outlet.

2.2 Basic setup

Before setting up the programmer, select a sturdy location for it without blocking the air vents on the right and left sides. The programmer uses a power cord, so the location must be near an AC outlet.
This section describes how to:
Position the display
Connect the ECG cable
Connect the programming head
Connect the power cord
Connect the telephone line
Connect the Ethernet cable
Turn on the programmer
Troubleshoot potential interference
26 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090

2.2.1 Position the display

1. Press inward on the two buttons on each end of the handle.
2. Lift the display
3. Place it at a comfortable viewing angle.
Reference Manual 27
Medtronic MEDTRONIC CARELINK® 2090

2.2.2 Connect the ECG cable

1. Slide the keyboard cover all the way back. Press the latch and flip up the keyboard.
2. Line up the arrow on the ECG cable with the red dot on the ECG connector.
3. Plug the cable into the connector with the black or blue marker on the right.
4. Flip down the keyboard, making sure that the cable passes through the notch on the right or left side.

2.2.3 About the ECG cable

The Medtronic Model 2090 EC/ECL ECG cable and lead wires connect the programmer to skin electrodes for ECG and measurement functions requiring surface detection of cardiac and implantable device signals.
The ECG cable is designed for use with five lead wires. However, some physicians prefer to use only four lead wires. If four lead wires are used, insert the chest ECG plug into the middle cable port of the ECG cable.
Note: Improper insertion of the cable plug may damage the connector pins.
28 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090
Cautions:
Upon opening the package, if the ECG cable appears damaged, do not use it. Contact your local Medtronic or Vitatron representative.
Do not pull on the insulated cable wire to disconnect the cable. Tension on the insulated cable wire may result in damage to the cable.

2.2.4 Connect the programming head

1. Slide the keyboard cover all the way back. Press the latch and flip up the keyboard.
2. Line up the red dots on the programming head cable and the programming head connector.
3. Plug the cable into the programming head connector with the yellow marker on the left.
4. Flip down the keyboard, making sure that the cable passes through the notch on the right or left side.
Reference Manual 29
Medtronic MEDTRONIC CARELINK® 2090

2.2.5 Connect the power cord

1. Open the back cover by pressing both latches.
2. Plug the power cord into the programmer.
3. Plug the power cord into AC power outlet. The programmer automatically adjusts to the available line power.
4. Make sure the power cord passes through the notch on the left side and close the cover.

2.2.6 Connect the telephone line

1. Locate the disk drive/PC card cover on the right side of the programmer and flip down the cover.
2. Connect the telephone line to the RJ-11 connector on the modem card or combo card. (The combo card is shown.)
30 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090
3. Connect the opposite end of the telephone line to an analog telephone jack.
4. If you replaced the card, reboot the programmer before using it.

2.2.7 Connect the Ethernet cable (as applicable to your hardware)

1. Locate the disk drive/PC card cover on the right side of the programmer and flip down the cover.
2. Connect the Ethernet cable to the Integrated Ethernet connection.
3. Alternatively, the Ethernet cable can be connected to the RJ-45 connector on the combo card instead of using the integrated Ethernet connection.
4. Connect the opposite end of the Ethernet cable to a network jack.

2.2.8 Turn on the programmer

Reference Manual 31
Medtronic MEDTRONIC CARELINK® 2090
1. Locate the ON/OFF switch on the left side of the programmer.
2. Press inward on the top of the ON/OFF switch.

2.2.9 About turning on the programmer

The first time the programmer is turned on following one of these changes, the start-up takes 2 minutes:
When a new keyboard language is installed
When the keyboard is removed (the programmer has been previously turned ON with a keyboard)
When a keyboard has been added (the programmer has been previously turned ON without a keyboard)
When the PC card has been removed (the programmer has been previously turned ON with a PC card)
When a PC card has been added (the programmer has been previously turned ON without a PC card)

2.2.10 Troubleshoot potential interference

To address possible harmful interference between the programmer and other devices, you are encouraged to take one or more of the following measures to address the situation:
Reorient or relocate the devices.
Increase the separation between the devices.
Connect the equipment to an outlet on a different circuit.
Consult Medtronic or Vitatron for help.

2.3 Connect peripheral devices

An analog input/output connector under the keyboard allows the use of a peripheral isolated medical grade recorder or monitor. A special adapter accessory (not included with the programmer) is needed to use the input/output connector. Contact your Medtronic or Vitatron representative for more information. The signals present at the output depend on the software application, but may include the following:
ECG
Marker Channel telemetry
32 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090
EGM
Software-specific waveforms
All electronic devices that are connected to the programmer must meet the electrical safety requirements of IEC-60950-1.

2.3.1 Locate the peripheral device connector

Figure 8. Locating the peripheral device connector

1 Analog Input/Output connector with green marker (under the keyboard)

2.4 Use external printers

Connecting a compatible printer to the programmer allows you to print full, page-size reports of session data when available. For more information, see the reference guide for the implanted device. This section describes how to connect a printer to your programmer.
All printers listed by this software are certified to IEC 60950, UL 60950 or equivalent. Only printers listed by this software may be connected to the programmer.

2.4.1 Printer compatibility

The programmer is compatible with many printers. A list of compatible printers can be accessed from the Print Queue screen.
Note: When programming a Vitatron device, refer to the applicable Vitatron reference guide for information about printing.
Reference Manual 33
Medtronic MEDTRONIC CARELINK® 2090

2.4.2 View a list of supported printers

Some devices only support the internal programmer printer. In these cases, the Print Queue window is not displayed during a session; it is displayed from the desktop when not in a session. Refer to the applicable device reference guide.
For those devices that support an external printer, refer to the following procedure.
1. If you are conducting a patient session, press the Reports icon, and then select Print Queue.
If you are not conducting a patient session, press the Print Queue icon.
2. On the Print Queue screen, select the Printer field to open the list of supported printers.

2.4.3 Materials you need

To connect a printer to the programmer, you need a USB printer cable or a Parallel Interface printer cable. For a USB printer cable, one end must be a USB Type A connector. The other end of the cable must fit the USB port on your printer. For a Parallel Interface printer, one end of the cable must fit the parallel interface port on your printer. The other end of the cable must be a standard 25-pin male D connector.

2.4.4 Connect the printer

Before connecting a printer to your programmer, exit the patient session and turn off the programmer.
Note: The connection method you use depends on your programmer hardware.
34 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090
2.4.4.1 Connect to a programmer that supports both parallel and USB printing
The following steps apply to programmers with hardware that support both parallel and USB printing.
1. To connect a printer, locate the correct port.
For USB cable printers, use a USB port located in the power cord compartment of the programmer.
For Parallel Interface printers, the Parallel Interface port is located under the disk drive/PC card cover on the right side of the programmer. Open the cover by pushing down on the small latch at the top center of the panel.
2. Connect the printer cable to the USB port or parallel connector on the programmer.
3. Connect the other end of the cable to the printer. Connect the printer power cord to an outlet and turn on the printer. Make sure that the printer has paper.
Note: Refer to the technical information provided with your printer for information about connecting and operating the printer.
4. Turn on the programmer and select the Print Queue icon.
Note: Be sure to select the correct printer driver from the options listed when you select the Printer field on the Print Queue window. You are now ready to use your programmer with the connected printer.
Reference Manual 35
Medtronic MEDTRONIC CARELINK® 2090
2.4.4.2 Connect to a programmer that supports only USB printing
The following steps apply to programmers with hardware that supports only USB printing.
1. To connect a printer, locate the correct port. There is one USB port located under the disk drive/PC card cover and two ports located under the power cord door on the back of the programmer.
2. Connect the printer cable to a USB port on the side or back of the programmer.
3. Connect the other end of the cable to the printer. Connect the printer power cord to an outlet and turn on the printer. Make sure that the printer has paper.
4. Turn on the programmer and select the Print Queue icon.
Note: Be sure to select the correct printer driver from the options listed when you select the Printer field on the Print Queue window. You are now ready to use your programmer with the connected printer.

2.5 Install printer paper

The internal printer requires a special thermal paper supplied in flat packs. You may obtain this paper directly from Medtronic or Vitatron. One package of printer paper contains six individual packs of paper. Each pack contains approximately 200 perforated sheets.
36 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090

2.5.1 How to load printer paper

1. Press the latch on the top of the tray on the left side of the programmer. Slide the tray out. Flip the paper holder over the end of the tray and remove any remaining paper.
Note: The tray may be taken out of the programmer to make removing any remaining paper easier. Press down on the lever on the side of the tray toward the front of the programmer.
2. Remove the wrapping from a new flat pack of printer paper, dispose of the cardboard, and unfold the top sheet.
Note: The shiny side of each sheet of paper with two black squares should be facing up as the paper is loaded into the tray. The black squares must be positioned toward the front of the programmer.
3. Slide the paper pack toward the back of the tray. Flip the paper holder over the paper pack.
4. Fold the top sheet of paper over the paper holder. Slide the tray back into position.
Reference Manual 37
Medtronic MEDTRONIC CARELINK® 2090

2.6 Printer buttons

The following four buttons control certain printer functions.

Figure 9. Printer buttons

1 50 mm/s 2 25 mm/s
50, 25, or 12.5 mm/s – Three buttons allow for selection of a desired paper speed for ECG, Marker Channel telemetry, and EGM recording. Pressing a button selects the printer speed and turns on the indicator light. Pressing a lit button stops the printout. The selected paper speed is printed once along the top edge of the chart recording. Paper speed for text printing is not affected by these buttons.
Paper Advance – Pressing Paper Advance advances the printer paper to its next perforation for tearing off.
3 12.5 mm/s 4 Paper Advance

2.7 Tear off a printout

To tear off a printout, separate the paper strip at a perforation. To tear off the printout at the printer, press Paper Advance. This action advances the paper to the next perforation. Grasp the printout near the printer, and pull.
38 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090

Figure 10. Tearing off printouts

2.8 Low paper supply

As the paper supply nears its end, a red stripe appears along the edge of the paper. A new paper pack should be installed as previously described in this chapter. When the paper supply runs out, the printer stops operating.
Note: The thermal paper used in this programmer is intended for immediate use. As the quality of thermal paper diminishes over time, the image fades. Medtronic and Vitatron recommend that you make photocopies for your files.
Reference Manual 39
Medtronic MEDTRONIC CARELINK® 2090

3 Configure the programmer

3.1 Display screen features

The programmer display screen is an interface that displays text and graphics. It is also a control panel that displays buttons and menu options that you can select using the touch pen.

3.1.1 Features and conventions of the display screen

This section provides an overview of the features of the display screen. For more information, see the reference guide for the implanted device. The main elements of a typical display screen before you select a model, when you turn the programmer on, and when you end a patient session, are shown in Figure 11.

Figure 11. Main elements of the display screen

1 Task bar 2 Status bar 3 Live Rhythm Monitor window 4 Waveform adjustment bars
40 Reference Manual
5 Task area 6 Command bar 7 Buttons 8 Tool palette
Medtronic MEDTRONIC CARELINK® 2090
Vitatron display screens may be different. For more information, see the reference guide for the implanted device. If you see a Medtronic/Vitatron switch button, press it to display the Vitatron Select Model screen.
3.1.1.1 Task bar
The task bar can contain these icons/indicators:

Table 1. Task bar icons/indicators

Icon Name Function
Position head light array
SessionSync icon Provides information about the connection and data
Diskette indicator Turns green to indicate diskette drive is available for sav-
USB indicator Turns green to indicate USB flash drive is available for
RemoteView icon Provides information about the connection between the
Analyzer indica­tor/selector
Device indica­tor/selector
Turns green to indicate successful communication between the programming head and the device. The greater the number of green bars on the array, the better the communication. A minimum of two green bars should be lit.
transfer status between the Programmer and the data management system. SessionSync is an optional fea­ture. If your device does not support SessionSync, a red symbol will be superimposed over the icon. For more information, see Section 8.1.
ing PDF reports and patient data. When diskette indica­tor is green, USB flash drive is not available.
saving PDF reports and patient data. When USB indica­tor is green, diskette is not available. When inserting a USB flash drive, you may experience a slight delay before device is available for use.
programmer and the remote system. Allows the user to initiate and terminate a RemoteView connection, and indicates if the RemoteView feature is available. The icon changes to indicate the connection status between the programmer and the remote system. For more informa­tion, see Section 7.2.
Used to start an analyzer session or, if your device sup­ports concurrency, to switch to an analyzer session from a patient session. During an analyzer session, the indi­cator box turns green. (For more information on the optional Analyzer, see the 2290 Analyzer Reference Guide.)
Used to go to the Select Model screen on the program­mer desktop or, if your device supports concurrency, to switch to a patient session from an analyzer session. During a patient session, the indicator box turns green.
Reference Manual 41
Medtronic MEDTRONIC CARELINK® 2090
3.1.1.2 Status bar
Before selecting a model, the status bar has no information. For specific information about the status bar, refer to the reference guide for the implanted device. After model selection, the status bar may include:
The present pacing mode.
Test condition status.
The device model.
3.1.1.3 Live Rhythm Monitor window
This window is a partial view of the full-screen display of the ECG, and contains a Status bar and a Waveform adjustment bar that allows you to change the waveform display. You can expand this window to full size by selecting the small square button in the upper-right corner of the window or by pressing [Adjust…].
After model selection, Marker Channel and telemetered EGM waveform traces may be available.
Refer to the appropriate reference guide for the implanted device for more information about the Live Rhythm Monitor. Refer to the 2290 Analyzer Reference Guide for information about the controls on the Waveform Adjustment bar.
3.1.1.4 Task area
The portion of the screen between the Live Rhythm Monitor window near the top of the screen and the command bar at the bottom of the screen changes according to the task or function you select.
3.1.1.5 Command bar
The bar at the bottom of the screen shows the command buttons for automatically launching the proper software application and displaying the Vitatron Select Model screen. For information on what command buttons are available after selecting a model, see the reference guide for the implanted device.
3.1.1.6 Buttons
Buttons allow you to operate the programmer using the touch pen. You can “press” a button by touching it with the tip of the touch pen.
42 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090
Buttons may directly execute a command, such as [Freeze], or they may open a window that prompts another action. Buttons that open a window usually have a label ending with an ellipsis, such as [Strips…] or [Adjust…].
A procedure may instruct you to “press and hold” a button. Press the tip of the touch pen to the button and maintain pressure until it is time to “release” the button.
When a button is inactive, it appears a lighter color and does not execute a command when you press it with the touch pen.
3.1.1.7 Tool palette
The collection of buttons and icons along the edge of the screen is referred to as the “tool palette”. These buttons and icons are the controls you use to choose the task or function screen you want to display. Each of the icons acts like a button. To select an icon, touch the icon with the touch pen. For more information, see Section 3.2. For information about the session tool palette, see the reference guide for the implanted device.

3.2 About the Between Patient Sessions tool palette

The Between Patient Sessions tool palette is located on the Select Model screen. The Select Model screen appears before you select a model, when you turn the programmer on, and when you end a patient session.
The tools that are available between patient sessions are described in Table 2.
Note: When programming a Vitatron device, refer to the applicable reference guide for information about the tool palette.

Table 2. Between Patient Sessions tool palette.

Tool Selecting the tool (button or icon)…
Freezes a segment of the live rhythm display. Note: A frozen strip can be viewed, printed, or saved to PDF (but not saved to diskette or USB flash drive) between patient sessions. Markers and EGM traces are not present between patient sessions.
The [Strips…] button is not available between patient sessions. Saved rhythm strips can only be accessed during a patient session.
Opens a window of options for adjusting the live rhythm display. Note: Additional adjustment options are present during a patient session.
Displays the screen for selecting a model and starting a patient session.
Reference Manual 43
Medtronic MEDTRONIC CARELINK® 2090
Table 2. Between Patient Sessions tool palette. (continued)
Tool Selecting the tool (button or icon)…
Displays a queue of print requests from previous sessions as well as fro­zen waveform reports requested between sessions. Refer to the reference guide for the implanted device to determine if these features are available.
Displays the programmer setup options. Preferences Time and Date Artifact Detection Software Demonstrations Programmer Profile SessionSync Status SessionSync Network Configuration… RemoteView Network Configuration… Network Configuration Other Software Tools Licensing
Selects the Analyzer for analyzing the electrical performance of a cardiac lead system (requires optional hardware - for more information, see the 2290 Analyzer Reference Guide.)
Note: When some functions are active on the display, pressing a tool button or icon has no effect. Closing the active window restores operation of the tool palette.

3.3 View and update programmer location and hardware information

Information about the location of the programmer and its hardware is on the Programmer Profile screens.
The Programmer Profile location screen has the following information:
Clinic’s name, address, telephone number, contact person, and customer account number
Service representative’s name, telephone number, fax number, and e-mail address
The Programmer Profile hardware screen has the model number for the Analyzer, and model and serial numbers for the programmer and the programming head.
Information on the screen may be updated by selecting the appropriate field and then using the keyboard.
44 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090

3.3.1 Verify Programmer Profile information

Each programmer has a profile screen that contains identifying information about the installed hardware, the programmer location, and contact information for the Medtronic service representative.
Typically, the profile is completed when the programmer is first installed, and then updated only when necessary.
1. Press the Programmer icon, and then select Programmer Profile. Location Information appears by default.
2. Complete the location information or verify that the information shown is correct.
3. To view hardware information, select Hardware Information.

Figure 12. Programmer Profile screen

3.4 Adjust programmer time and date

If the time or date displayed and printed by the programmer is incorrect, use the following procedure to enter the correct settings. For Vitatron devices, see the applicable reference guide.
Reference Manual 45
Medtronic MEDTRONIC CARELINK® 2090

3.4.1 Set the time and date

1. Press the Programmer icon, and then select Time and Date.
2. From the Programmer Time and Date screen, press the up or down button to increase or decrease the value for the unit of time that you want to change. Press and release the button for single unit changes, or press and hold the button to make larger changes.
3. When all fields show the correct time and date, press [Apply]. Select another tool palette icon to close the Programmer Time and Date screen.

Figure 13. Programmer Time and Date screen

Note: Time must be entered based on a 24-hour clock, with 00:00 being midnight, and
12:00 being noon.

3.5 Select audible tones

Certain events in the operation of the programmer result in an audible signal. The following tones alert you to the success or failure of an action.
A two-tone beep (low-to-high) indicates confirmation of a successful Interrogate or a Program command.
A double low-tone beep indicates that an Interrogate, Program, or Emergency command was not confirmed. It can also indicate that the selected command cannot be executed.
46 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090
Note: For some devices, the tones may not be turned off. For more information, see the reference guide for the implanted device. For Vitatron devices, see the applicable reference guide.

3.5.1 Turn tones on or off

1. Press the Programmer icon, and then select Preferences.
2. From the Preferences screen, select [Audio ON] or [Audio OFF] as desired.

Figure 14. Audio Preferences screen

3.6 Check the software version

This section describes how to determine the version of software that is loaded on the programmer.
If you need to know what version of software is currently loaded on the programmer for any of the device models, use the following procedure.
For Vitatron devices, see the applicable reference guide.

3.6.1 To check the software version number

1. Press the Programmer icon, and then select Software.
2. For each device model with software loaded on the programmer, the screen displays the software version number next to the model number.
Reference Manual 47
Medtronic MEDTRONIC CARELINK® 2090

Figure 15. Software on This Programmer screen

Note: If the model that you require is not displayed, the software to support that model is not
currently loaded on the programmer. Contact your Medtronic or Vitatron representative.

3.7 Select other software

In addition to the standard application software, there are some programmers that have other applications installed. These applications may include supplemental software or software used in clinical studies for research. If you have other software installed, you may access the software, using the following procedure.
1. Press the Programmer icon, and then select Other Software.
2. When the programmer displays the list of available software, select the application and press [Start].

3.8 Remove other software applications

Programmers with other software installed, such as supplemental software or those used in clinical studies for research, may allow the applications to be removed from the programmer desktop. If you have software installed that permits removal, you may remove it using the following procedure.
1. Press the Programmer icon, and then select Software.
2. Press [Uninstall Software…].
48 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090
3. When the programmer displays the list of removable software, select the application to be removed, and then press [Uninstall].
4. Select the check box next to the acknowledgment statement, and then press [Continue].
5. The software is removed, and the programmer reboots.
6. Verify that the software has been removed.

3.9 Improve the detection of pacing artifacts

The Artifact Detection function allows you to improve the detection of pacing artifacts when interference causes either false artifacts or no artifacts to appear on the patient’s ECG. Pacing artifacts are displayed on the patient’s ECG when the artifact detection option (Show Artifacts) has been enabled.
To determine if this feature is applicable, see the reference guide for the implanted device.

3.9.1 Enable artifact detection

1. Press the Programmer icon, and then select Artifact Detection.
2. Make sure the current settings include ARTIFACT DISPLAY IS ON.
3. Make sure the current settings include MV FILTER IS ON.

3.10 Start the Demonstrations option

The demonstrations option allows you to run a demonstration program on the programmer.
For Vitatron devices, see the applicable reference guide.
Note: Device applications and reference manuals may still refer to using the “demonstration disk” or “demonstration diskette” to run a demonstration program. The need for a demonstration diskette to access demonstration mode is no longer required. All references to a demonstration diskette can be ignored. If a demonstration diskette is inserted inadvertently, operation of the demonstration mode is not affected. All demonstration mode features are accessible with or without a demonstration diskette.
Reference Manual 49
Medtronic MEDTRONIC CARELINK® 2090

3.10.1 Access demonstrations

1. Press the Programmer icon, and then select Demonstrations.
2. From the Demonstration Model Selection screen, select the desired View option to list the available demonstration programs.
3. Select the desired demonstration program and press [Start].
50 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090

4 Update programmer software using the Software Distribution Network

4.1 The Software Distribution Network

Programmer software can be updated by Medtronic customers or Medtronic personnel by accessing the Medtronic Software Distribution Network (SDN) and downloading the software. The SDN uses a world-wide network to connect to servers in the United States. These servers are able to download software to many programmers simultaneously through secure connections.
The SDN is available 24 hours per day, 7 days per week and always contains the most current software. For this reason, it is recommended that you download the software from the SDN rather than from the flash drive.
You can connect to the SDN using a network connection (wired or wireless) or a dial-up connection. It is recommended that you use a network connection whenever possible, since software downloads are faster using a network connection than using a dial-up connection.
Notes:
It is recommended that the SDN be checked on a regular basis. Checking regularly reduces the size of the download and the time it takes to receive the software.
If the download was interrupted, the download will resume the next time the programmer attempts to access the SDN.
Normal programmer functions are unavailable during software installation.

4.2 Connect to the SDN using a network connection

You can connect to the SDN using your clinic’s network. By connecting through your network, software download time can be reduced.
Before you begin, make sure your network connection is configured properly.
Reference Manual 51
Medtronic MEDTRONIC CARELINK® 2090

4.2.1 How to connect to the SDN using a network connection

1. Press the Programmer icon, and then select Software.
The programmer displays the Software on This Programmer screen and lists the software already installed on the programmer. For each model, the screen displays the software version.
Note: The SDN cannot be accessed from Vitatron screens. Change to the Medtronic Select Model screen.
2. Press [Install from Medtronic…].
3. Press [Accept].
or
Press [Cancel]. The download process is cancelled and the programmer redisplays the Software on This Programmer screen.
52 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090
4. The programmer displays the Scheduled Software Update screen.
Either choose to start the download at a particular time by selecting a time from the Scheduled Update Time pull-down menu, or begin the download as soon as possible by pressing [Start].
5. The Scheduled Software Update window displays a countdown window showing how much time remains until the download begins. Press [Start Now] to override the countdown or press [Cancel] to interrupt the countdown and the download request and return to the Software on This Programmer screen.
Reference Manual 53
Medtronic MEDTRONIC CARELINK® 2090
6. The programmer displays a list of software that will download and install.
Note: Individual software cannot be selected or rejected.
You may press [Stop] during the software download and resume the download at a future time.
7. When the download is complete and the progress bar is approximately 50% filled, software installation begins.
8. The [Stop] button disappears.
Note: Do not power-off the programmer while software is being installed.
9. When software installation is complete, the programmer disconnects from the SDN, automatically reboots, and displays a screen listing the software that was downloaded and installed.
10. To obtain technical manuals for the new software, see Section 1.12.1, “Access the Medtronic Manual Library website”, page 20.
11. Press the Select Model icon. The programmer is then available for patient use.
54 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090
Note: The first time the newly downloaded software is accessed, some additional installation steps may be completed but these steps are automatic and no user intervention is required

4.3 Connect to the SDN using a dial-up connection

You can connect to the SDN using the programmer combo card or modem card, and a connection to an analog telephone line. In most cases, the modem connection to the SDN can be made using a local telephone number. There is a toll-free number provided, however, it should only be used if a local number is not available.
Before you begin, make sure that the telephone line is correctly connected to the combo or modem card. For more information, see Section 2.2.6.

4.3.1 How to connect to the SDN using dial-up

1. Press the Programmer icon, and then select Software.
Note: The SDN cannot be accessed from Vitatron screens. Change to the Medtronic
Select Model screen.
2. For each device model with software loaded on the programmer, the screen displays the software version number next to the model number.
3. Press [Install from Medtronic…].
Reference Manual 55
Medtronic MEDTRONIC CARELINK® 2090
4. Press [Accept].
or
Press [Cancel]. The download process is cancelled and the programmer redisplays the Software on This Programmer screen.
5. If the programmer has the combo card installed, the Scheduled Software Update window appears showing the LAN connection by default.
Press [Configure] to choose Dial-up Connection.
Then, select the Dial-up Connection radio button and press [OK].
The Scheduled Software Update window redisplays, now showing Dial-up Connection. Press [Start] to continue.
6. If the programmer has the modem card installed, the programmer automatically displays the Update Software screen.
56 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090
7. Review the Update Software screen.
To select a different clinic location, press the down arrow.
If your clinic location is not listed, refer to Section 4.3.2 for instructions on how to edit, create, or remove a location.
8. Enter a dialing prefix, if needed.
9. Press [Start] to begin the software download.
10. While the programmer is connecting to Medtronic, the screen temporarily goes blank.
Reference Manual 57
Medtronic MEDTRONIC CARELINK® 2090
11. The programmer reboots, connects to the SDN and then displays a list of software that will download and install.
Note: Individual software cannot be selected or rejected.
You may press [Stop] during the software download and resume the download at a future time.
12. When software downlad is complete and the progress bar is approximately 50% filled, software installation begins.
13. The [Stop] button disappears.
Note: Do not power-off the programmer while software is being installed.
14. When software installation is complete, the programmer disconnects from the SDN, automatically reboots, and displays a screen listing the software that was downloaded and installed.
15. To obtain technical manuals for the new software, see Section 1.12.1.
58 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090
16. Press the Select Model icon.
The programmer is then available for patient use.
Note: The first time the newly downloaded software is accessed, some additional installation steps may be completed but these steps are automatic and no user intervention is required.

4.3.2 Edit locations

Part of the dial-up software download process involves confirming the programmer location on the Update Software screen. If the clinic location does not appear in the clinic pull-down menu or if information about an existing location needs to be changed, select Edit Locations from the pull-down menu.
The following procedures describe how to create a new location, change an existing location, or remove a location from the programmer.
To create, edit or remove a location, you must be on the Update Software screen. The Update Software screen appears as one of a series of screens that are accessible only during the software update process.
Reference Manual 59
Medtronic MEDTRONIC CARELINK® 2090

4.3.3 How to create a new location

1. Using the clinic pull-down menu, select Edit Locations. When the Edit Locations screen appears, press [New].
2. When the new dialog window opens enter the name of the location.
3. Using the pull-down menus, identify the location by selecting the Country, Region, and Nearest City.
4. Using the pull-down menu for Number to dial, select the number to dial to access the SDN.
5. In the Dialing Prefix field, enter the digits of a dialing prefix, if a prefix is required to access an outside line.
6. To set this location as the default location on the Update Software screen, select Set as default location.
7. Press [OK] to save this location information.
8. When the Update Software screen is redisplayed, select the clinic and press [Start] to begin the download process or press [Cancel] to exit from the download operation.
60 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090

4.3.4 How to change location information

1. On the Edit Location screen use the pull-down menu to select an existing location. Then, beginning with the Nearest City, use the pull-down menu to select a different city if it needs to be changed.
2. To select a different SDN access number, use the pull-down menu opposite the Number to dial field. The list of SDN access numbers shown is based on the Nearest City selected.
3. In the Dialing Prefix field, enter the digits of a dialing prefix, if a prefix is required to access an outside line.
4. To set the currently displayed location as the default location on the Update Software screen, select Set as default location.
5. Press [OK] to save this location information.
6. When the Update Software screen is redisplayed, select the clinic and press [Start] to begin the download process or press [Cancel] to exit from the download operation.
Reference Manual 61
Medtronic MEDTRONIC CARELINK® 2090

4.3.5 Remove a location

1. On the Edit Location screen use the pull-down menu to select the name of the clinic you want to remove from the programmer.
2. Press [Remove].
3. Press [Yes] to confirm the deletion.
4. On the Edit Location screen, select the clinic to be used from the pull-down menu, then press [OK].
5. When the Update Software screen is redisplayed, press [Start] to begin the download process or press [Cancel] to exit from the download operation.
62 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090

5 Conduct a patient session

5.1 Prepare for a patient session

Familiarize yourself with the information in this section before beginning a patient session.

5.1.1 Connect the programmer to skin electrodes

At the start of each patient session, ECG cable leads must be connected to the patient to detect cardiac and pulse artifact signals.
Note: The quality of disposable skin electrodes used with the programmer is important to the performance of the programmer signal sensing functions. Chemical reactions occur at the electrode/paste interface and produce small DC voltages that can block the ECG signal. Using high quality silver/silver chloride (Ag/AgCl) electrodes can minimize this problem. Electrodes should be fresh and from the same box. The patient’s skin should be prepared according to the directions provided with the electrodes.
Protocols covering attachment of leads to disposable skin electrodes may vary. Leads may be attached to the electrodes either before or after the electrodes are applied to the patient. The order of the following procedure is arbitrary.
Warning: Do not connect the programmer to wires or electrodes internal to the body. The programmer is designed to be medically safe only when attached to surface electrodes.
Reference Manual 63
Medtronic MEDTRONIC CARELINK® 2090

5.1.2 Attach electrodes

Attach five standard, disposable electrodes to the patient in the positions shown.

5.1.3 Connect the ECG cable

64 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090
1. As shown, attach a color-coded lead wire to each of the five electrodes. Match a color to each electrode as in Table 3.
Note: Connecting the chest lead is optional. If the chest lead is not used, insert the chest ECG plug into the middle cable port of the ECG cable.
2. Connect each lead wire to the ECG cable as in Table 4. Match each lead connector to the proper cable port.

Table 3. Electrode lead wire color coding

AHA Coding
a
IEC Coding
Black Yellow to left arm
Red Green to left leg
Brown White to left chest area
Green Black to right leg
White Red to right arm
a
American Hospital Association
b
International Electrotechnical Commission
c
Connecting the chest lead is optional. If the chest lead is not used, insert the chest ECG plug into the middle cable port of the ECG cable.
b
Body Area
c

Table 4. ECG cable color coding

AHA Coding IEC Coding
Black to LA Yellow to L
Red to LL Green to F
Brown to V
Green to RL Black to N
White to RA Red to R
a
Labeled C on some cables
a
White to C
Note: Occasionally, mutual interference occurs between the programmer skin electrode signals and signals from an external ECG recorder or monitor attached directly to the patient. This interference may cause erratic operation of the programmer functions that depend on surface signal detection. If interference occurs, temporarily disconnect the leads from the attached ECG recorder or monitor. This interference does not affect the programming functions of the programmer.
Reference Manual 65
Medtronic MEDTRONIC CARELINK® 2090

5.1.4 Use the touch pen

The touch pen is used to select programming functions provided by the software. Proper use of the pen is described in Figure 16 and in Section 5.1.5.

Figure 16. Using the touch pen

66 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090

5.1.5 Select an option on the screen

1. Move the tip of the pen to a position directly over the desired option. If the desired option is a displayed key or button, position the pen tip within the rectangular outline. If the desired option is a name or number, such as a parameter or parameter value, position the pen directly over the letters or numbers forming the option.
2. Touch the pen to the screen to select an option.

5.1.6 Use the keyboard

Certain fields on the screen allow you to enter data, such as the patient’s name or chart number. The programmer has a compact computer keyboard for entering data.
Reference Manual 67
Medtronic MEDTRONIC CARELINK® 2090

Figure 17. Programmer keyboard

5.1.7 Position the programming head

During most applications of the programmer, the programming head must be positioned over the implantable device. Positioning the programming head is required for any interaction between the programmer and the implantable device.

5.1.8 When to position the programming head

Caution: Do not position the programming head over an implanted device during
electrocautery or external defibrillation procedures.
During a patient session, properly position the programming head over the implanted device before any of the following actions:
Selection of any command that initiates a programming transmission. The programming head must be held in position until completion of the transmission, which usually is indicated by a confirmation message.
Selection of any command that initiates data transmission from the implantable device. The programming head should be held steady until data reception is complete, which usually is indicated by a confirmation message.
Selection of a measurement function that requires the implantable device to be operating asynchronously as a result of the programming head magnet.
68 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090
For any temporarily programmed state or function or for reception of continuous data such as Marker Channel telemetry or EGM waveforms, the programming head must be held in place over the implantable device for the duration of the function or until termination is desired. Lifting the programming head cancels a temporary program and terminates continuous telemetry. The implantable device reverts to permanently programmed values.

5.1.9 Determine the correct position

For an implantable device, the programming head should be held directly against the patient’s skin. The face of the programming head must be parallel to and typically within 5 cm of the implantable device. Optimum position of the programming head may not be directly centered over the implantable device.

Figure 18. Position the programming head

1 Light array 2 Green
3 Amber/green
Correct placement of the programming head is indicated in two places: the position head array in the top left corner of the screen and the array of seven lights on the programming head (see Figure 18).
Programming and Interrogation are not recommended when fewer than two green lights are lit.
Reference Manual 69
Medtronic MEDTRONIC CARELINK® 2090

5.1.10 Program and interrogate the implanted device

1. Select the appropriate software parameters according to the reference guide for the implanted device.
2. Position the programming head near the implanted device.
Programming and Interrogation can begin when the LED lights on the position head array indicate satisfactory positioning and telemetry strength.
Notes:
For Medtronic devices, the programming head array shows the signal strength of the communication link. Medtronic recommends moving the programming head to maximize the number of green lights. All lights may not illuminate for all models. For more information, see the reference guide for the implanted device.
Misalignment of the programming head could result in failure of a programming transmission and/or failure to receive data from the implantable device. Medtronic recommends that you interrogate the device after programming to confirm that any setting changes were successful.
Reveal Plus and some older Vitatron devices respond with only one light in the position head array. In programming mode, the light turns green, to indicate that the programming head is in the correct position and that it is communicating with the device; the light turns solid amber if the programming head is misaligned.

5.1.11 Use the P and I buttons

The P button and I button on the programming head (Figure 19) correspond to commands that appear on the display screen of the programmer. Refer to the appropriate reference manual to determine which commands the buttons correspond to. For convenience, you may press the P button or the I button on the programming head in place of selecting the corresponding option from the display.
70 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090

Figure 19. Programming head buttons

1 P button 2 I button

5.1.12 The programming head magnet

The programming head contains a strong magnet. For more information about the effects of a magnet, see the reference guide for the implanted device.
The programming head may attract metal instruments or may be attracted to metal surfaces. The magnet is susceptible to partial demagnetization when it is subjected to opposing magnetic fields, such as those present when forcing the programming head against another magnet. The programming head should be stored as shown in Figure 21 when not in use.
Caution: Keep the programming head away from any device or material that will be damaged by the magnetic field, including magnetic media, watches, and other electronic devices.

5.2 Initiate a patient session

A patient session involves the application of the various programmer functions to such procedures as programming implantable device parameters, analyzing or assessing implantable device operation, troubleshooting, and routine follow-up. The instructions for using each programmer function are covered in the reference guide for the implanted device.
Note: Before proceeding, ensure that all preparations covered in Section 2.2 and Section 5.1 have been completed.
Reference Manual 71
Medtronic MEDTRONIC CARELINK® 2090

5.2.1 Programmer checklist

1. Is the programmer set up according to the procedures in Section 2.2?
2. Are the ECG cable, touch pen, and programming head connected to the programmer?
3. Does the power cord connect the programmer to an AC power outlet?
4. Has the appropriate software been installed? Refer to Section 3.6 for a description of how to verify the software version.
5. Are the programmer ECG cable leads connected to electrodes on the patient as described in Section 5.1.1?
Specific information related to each implantable device model or family of models is included in the reference guide for the device.
Refer to the reference guide for the implanted device before beginning a patient session.

5.2.2 Model identification

Because the programmer collects and stores data on a session-by-session basis, it is important to start and end each session correctly.
The programmer supports both a Medtronic and Vitatron desktop. Whichever desktop is in use when the programmer is powered down, that same desktop appears when the programmer is powered on. To switch from the Vitatron desktop to the Medtronic desktop and vice versa, press the Vitatron/Medtronic switch button that appears on the bottom of the screen.
There are two ways to begin a patient session:
Before turning on the programmer, position the programming head over the patient’s device. When you turn on the programmer, the programmer attempts to interrogate the device. Depending on the device, either the software application is launched automatically or a message appears with further instructions.
After turning on the programmer, position the programming head over the patient’s device. During the first 5 minutes, the Medtronic desktop displays the Find Patient screen. Afterward, it displays the Select Model screen. The Vitatron desktop displays the Select Model screen immediately. A patient session can begin at either the Find Patient screen or the Select Model screen. Follow the instructions on the screen that is displayed.
72 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090
5.2.2.1 Find Patient screen
When the programmer is first turned on, the Medtronic desktop displays the Find Patient screen. If it does not detect a device within about 5 minutes, the programmer removes the Find Patient screen to reveal the Select Model screen.
When the Find Patient screen is displayed, you may begin a patient session.
Place the programming head over the patient’s device and hold it steady. For most devices, the programmer identifies the device model and automatically starts up the proper software application. If a device cannot be automatically identified, the programmer displays a message at the top of the Find Patient screen. Perform one of the following steps, depending on the message instructions:
Press [Cancel] and manually select the software application from the Select Model screen.
Press [Cancel] and then press the Vitatron/Medtronic switch button to go to the Vitatron desktop.
If the message indicates that the needed software application has not been installed, contact your Medtronic or Vitatron representative.
Reference Manual 73
Medtronic MEDTRONIC CARELINK® 2090
5.2.2.2 Select Model screen
A patient session may also begin from the Select Model screen. The Select Model screen appears after one of the following actions:
Shortly after the programmer has been turned on
After you end a patient session
If the Select Model screen is not displayed, use the touch pen to press the Select Model icon. If the Select Model icon is not displayed, a patient session is in progress. You must end that session before starting a new session.
If you are between patient sessions, you can access other screens by using the icons and buttons described in Section 3.2.
If the device is a Vitatron device and it is not listed on the Select Model screen, refer to the Vitatron Software Programming Guide.
If the Select Model screen does not look like this example and you see a different button in the command bar, press the Vitatron/Medtronic switch button to display this screen.
Position the programming head over the patient’s device and hold it steady. Press [Find Patient] shown on the Medtronic desktop or manually select the device from the displayed list of devices and press [Start].
74 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090
When a device is manually selected from the list of devices, the programmer starts up the application that corresponds to your selection, not the device that is under the programming head. The Starting Application screen is displayed briefly as the programmer starts up the proper software application. If the software application has not been installed, the programmer displays a message indicating that the software must be installed before proceeding.
The programmer may automatically interrogate the patient’s implanted device to retrieve most of the data that might be needed during the session. To take advantage of this automatic interrogation, position the programming head over the implanted device and continue to hold it in place until the interrogation is complete.
For more information about determining the model, see the reference guide for the implanted device.
Note: If you are using the optional Analyzer and your device does not support concurrency with the Analyzer, you must exit the Analyzer before you can start a patient session. If you use the task bar to switch from the Analyzer to the Select Model screen and then press [Find Patient], the programmer displays the message: “This application is not able to run concurrently with the Analyzer. Please exit the Analyzer and try again.”

5.3 Emergency VVI button

The red emergency VVI button on the display panel provides immediate access for emergency VVI pacing during a patient session (Figure 20). (Specific parameter values are determined by each application.)
For all Medtronic pacemaker applications, the red emergency VVI button is active when the [Emergency] button is displayed on screen.
Note: The [Emergency] button is also implemented in the software and appears on the display. It functions similarly to the VVI button on the panel. For more information on the [Emergency] button, see the reference guide for the implanted device.
For some ICD applications, the red emergency VVI button is active at all times. When pressed, this button delivers VVI pacing and displays the emergency screen options.
Reference Manual 75
Medtronic MEDTRONIC CARELINK® 2090

Figure 20. VVI button on the display panel

1 Red emergency VVI button
Warning: The red emergency VVI button is disabled for the following devices: PCD, Jewel PCD, Micro Jewel II, MicroJewel, Jewel CD, Jewel Plus, Model 7202 Jewel CD, Models 7218 and 7211, Model 7201 CD, and Jewel AF. Use the touch pen to press the on-screen [Emergency] button.

5.3.1 Deliver emergency bradycardia pacing

To initiate emergency pacing, correctly position the programming head over the implanted device and press the red emergency VVI button. A message confirms programming, and emergency VVI operation begins.

5.3.2 Deliver emergency tachyarrhythmia therapy

To deliver therapy, press the red emergency VVI button to display the emergency screen on the programmer and press the on-screen [Deliver] button with the touch pen.
For specific applications, refer to the appropriate device Reference Guide, System Reference Guide, or Clinician Manual for complete instructions regarding the use of the [Deliver] button.
76 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090

5.4 End a patient session

When you want to end a patient session, you may save data to a supported storage device or end the session without saving.
Refer to the reference guide for the implanted device for specific information on saving device data.

5.5 Store components

The following diagrams show the proper way to store components.

Figure 21. Storing components

Reference Manual 77
Medtronic MEDTRONIC CARELINK® 2090

6 Manage session data and reports

6.1 Session data

Patient session data may be saved to a diskette or a USB flash drive.

6.2 Reports

Depending on the implanted device model, various types of reports can be created. Refer to the reference guide for the implanted device for specific information on report types and contents. During an active session, reports may be printed, or saved as PDF files on a diskette or USB flash drive. Reports held for later printing may be printed while at the desktop or when returning to a session. Reports might not be available for later printing from the desktop, depending on the device application and on the current print queue deletion schedule. For more information, see Section 6.7.

6.3 Save to a PDF file

Printable reports, frozen strips, and other data may be saved to a PDF file. A PDF file is an electronic version of a printed document; therefore, the feature is accessible under the printing commands.
Note: The Save to PDF File feature works for all applications that can print to a full-size printer.
To save to a PDF file, perform the following steps:
1. Open or create the report or file.
2. Press [Print…] or [Print Options…] to display the Print – Options dialog box.
Note: If the Print – Options dialog box does not display, open Preferences, and then select the Printing: Pop up these options when any print button is selected check box.
3. From the Print – Options dialog box, select the Printer: Full Size radio button. A list of supported printers is displayed.
4. From the list of printers, select the Save to PDF File option. The report is saved to an attached USB flash drive, or to a diskette, if a diskette is present in the diskette drive and a USB flash drive is not connected. For more information on saving to USB, see Section 6.5.
78 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090

6.4 Save to diskette

For specific information on saving and reading device data using a diskette, see the reference guide for the implanted device . The diskette drive is functional only when a USB flash drive is not in use. Before using the diskette drive, ensure that a USB flash drive is not connected to the programmer.
When no USB flash drive is connected, the diskette indicator is green to indicate that the diskette drive is available for use. Inserting a diskette allows all session data and PDF file saving operations to take place on the diskette.

6.5 Save to USB

Many device applications support the use of diskettes for saving and loading session data. The programmer is also able to perform these operations using a USB flash drive. When a USB flash drive is connected to the programmer, the diskette drive becomes unavailable.
Note: Device applications and reference manuals may still use the terms “disk” or “diskette” in the context of saving and retrieving device data. When a USB flash drive is connected to the programmer and available for use as described below, the terms disk or diskette should be interpreted as applying to the USB flash drive, rather than the diskette.

6.5.1 Supported USB storage devices

In order to ensure the integrity and security of patient health information, it is recommended that you use USB flash drives that are dedicated to storing programmer data only.

6.5.2 Operation

A USB flash drive should be connected or disconnected while at the desktop or in a session. Connect a writable USB flash drive to the programmer using any available USB port. A slight delay may occur while the USB flash drive is authorized. The USB indicator on the task bar turns green to indicate that the USB flash drive is available for use. The diskette icon turns gray to indicate that the diskette drive is not available.
Reference Manual 79
Medtronic MEDTRONIC CARELINK® 2090
USB flash drives should not be connected or disconnected while the following actions are in progress:
Programming a device
Performing a Save to Disk or Media
Performing a Read from Disk or Media
Saving a report as a PDF file
Notes:
While a Save to Disk or Media action is in progress, the progress indicator and the message “Saving…” display. The progress indicator displays the completion percentage. Before removing the USB flash drive, wait a few seconds after the progress indicator shows 100%.
After a report is saved as a PDF file, the message “PDF report(s) saved to media” is displayed for about 5 seconds. Wait a few seconds after the message goes away before removing the USB flash drive.
If an active session is ended while reports are currently printing or pending, the reports are cancelled and may not be available from the desktop print queue.
Any operation that normally uses a diskette to read or write data (such as Save to Disk or Media, Read from Disk or Media, save reports to a PDF file) will use the USB flash drive after it is connected. Refer to the reference guide for the implanted device for specific information on saving device data. For more information on saving reports to PDF file, see Section 6.3.
Connect only one writable USB flash drive at a time. Connecting two or more USB flash drives results in an error during data-saving operations. This condition is indicated by the USB disabled icon.

6.6 View reports that are saved to media

Reports that have been saved to media may only be viewed on a computer; they cannot be viewed on the programmer itself. After saving, remove the storage media (diskette or USB flash drive) containing the reports and insert it into a computer equipped to display files that are in PDF format.
80 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090
All reports from one patient’s session are contained in one PDF file. File names are automatically assigned according to a naming convention that ensures uniqueness on the storage media:
Patient’s name (if previously provided in Patient Information)
Device serial number
“Session Report”
Clinic visit date in MM_DD_YY format
Version number (the first PDF saved to this storage media gets “1”)
For example: John Q Patient_aaannnnnna_Session Report_06_25_10_1.PDF

6.6.1 Recommendations for viewing and printing PDF files

Due to computer and software variations, some PDF files may not be displayed properly when viewed on a computer monitor.
The use of Adobe Reader 9 or later is recommended. Adjusting the following settings may reduce or eliminate display imperfections:
Replace document colors with white page background and black text (in Adobe Reader 9, select: Edit > Preferences > Accessibility > Replace Document Colors > Custom Color)
Deselect the option to enhance thin lines (in Adobe Reader 9, select: Edit > Preferences > Page Display)
Imperfections that may be seen on screen:
On graphs that contain rectangles drawn with thin lines, e.g., bar graphs, the thin lines may not be displayed at various zoom levels.
On Pacing and Tachy Trigger Episode reports, unfilled circles may be displayed as filled circles.
PDF reports print properly at a resolution of 300 dpi or greater.

6.7 Set the interval for report deletion

For patient data security, the programmer permanently deletes reports automatically from the Medtronic desktop print queue at the time when the programmer is powered up. You can control how long reports are retained in the print queue before automatic deletion.
Reference Manual 81
Medtronic MEDTRONIC CARELINK® 2090

6.7.1 Select a report deletion interval

1. Press the Programmer icon, and then select Preferences.
2. From the Preferences screen, select Delete Reports. The programmer displays the Delete Reports screen as shown in Figure 22.

Figure 22. Delete Reports screen

3. Select a radio button to specify which reports the programmer deletes:
All Reports
Reports older than 1 Day
Reports older than 2 Days
Reports older than 7 Days (Default)
Reports older than 14 Days
The age of a report is determined by the date and time it was created. When the programmer is powered on, reports that meet the deletion criteria are permanently deleted.
82 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090

6.7.2 Delete a report immediately

To delete a report immediately, directly access it from the print queue and press [Delete].

6.8 Manage patient data privacy

You can immediately remove from the programmer all Protected Health Information (PHI). This feature deletes all of the following files:
Contents of the Print Queue (unless files are currently being printed or copied)
Temporary files residing on the programmer
Memory dumps (applies to Vitatron devices only)
You cannot delete any PHI data if a session is in progress or while files are being printed or copied to media. If deletion is interrupted manually, some PHI remains on the programmer.
Note: The user of the programmer is responsible for the use of this feature, as well as for management of patient data that has already been removed from the programmer (for example on paper, diskette, or a USB flash drive).

6.8.1 Delete Protected Health Information

1. Press the Programmer icon, and then select Tools. The programmer displays the Tools screen as shown in Figure 23.

Figure 23. Tools screen with Patient Data Privacy index item selected

Reference Manual 83
Medtronic MEDTRONIC CARELINK® 2090
2. Press [Delete Protected Health Information]. The programmer displays the dialog box as shown in Figure 24.

Figure 24. Delete Protected Health Information confirmation dialog box

3. Press [Delete] to continue.
One of the following events may occur:
The programmer displays an “In progress…” dialog box as shown in Figure 25. Deletion may last several minutes, depending on the amount of data to delete.

Figure 25. In progress dialog box

Note: On the In progress dialog box, press [Stop] if you want to halt further deletion. For
example, you may want to end a deletion process that is taking too long.
If you press [Stop], some protected health information remains on the programmer.
Or,
If there are reports currently printing, the programmer displays a message directing you to wait until printing is complete, as shown in Figure 26. Press [Close].
84 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090

Figure 26. Printing in progress dialog box

4. If pressing [Delete] in Step 3 resulted in protected health information deletion, the programmer displays a message stating that deletion was successful, as shown in Figure 27. Press [Close].

Figure 27. Deletion successful dialog box

Or,
If the programmer is unable to complete deletion of files, it displays a message stating that there was an error and that some data may remain on the programmer, as shown in Figure 28. Press [Close]. Contact Medtronic Technical Support if the message recurs.

Figure 28. Error deleting files dialog box

Reference Manual 85
Medtronic MEDTRONIC CARELINK® 2090

6.9 Vitatron Manual-Guided Reset

Manual-guided reset is a diskette-based tool that allows recovery of Vitatron DDD3, DIVA, and (Co)DEMA devices that have lost data such as losing trim values on a device reset. Before attempting to use the MGR diskette, ensure that no USB flash drives are connected.
86 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090

7 RemoteView™ Programmer Consultation

7.1 About RemoteView

Medtronic RemoteView provides the ability for a programmer user to share the programmer screen with a remote viewer, located in a different physical location anywhere in the world via the Internet. RemoteView uses software on both the programmer and the remote viewer’s system. The programmer user and remote viewer each establish a network connection to a Medtronic RemoteView server. After the server provides authentication, screen sharing occurs.
Notes:
RemoteView does not allow remote programming or remote control of the programmer; it is only for viewing.
Also, patient data may be viewed on the remote system during a RemoteView connection. For a review of potential data privacy issues, see Section 7.4.

7.2 RemoteView status icon

The RemoteView status icon on the programmer provides information about the connection between the programmer and the remote system. The icon changes to indicate the connection status between the programmer and the remote system.

Table 5. RemoteView status icon states

RemoteView status icon state Color What the icon indicates
Grayed out (not selecta­ble)
Gray RemoteView is available but not connec-
RemoteView is disabled. Clicking the icon will cause the programmer to beep. RemoteView is not available because the device application is not supported or you have a programmer that does not meet the hardware system require­ments. If your programmer does not meet the hardware system require­ments, contact Medtronic Technical Support.
ted. There is no connection between the programmer and the remote system.
Reference Manual 87
Medtronic MEDTRONIC CARELINK® 2090
Table 5. RemoteView status icon states (continued)
RemoteView status icon state Color What the icon indicates
Gray with flashing green
Green A successful connection has been
Gray with redXThe connection attempt has failed. You
The connection is being established between the programmer and the remote system, but it is not yet connected.
established between the programmer and the remote system.
can attempt to reconnect by clicking the icon again. Any of the following condi­tions may cause a connection attempt to fail: Network card error Network cable connection error. Check to see if the network cable is connected. No internet access. Check with clinic net­work support staff to ensure HTTPS on port 443 is allowed to connect to the Internet. Incorrect server configuration selected. The server on the programmer and the remote viewer’s system need to match. See Section 7.3.1 for more specific infor­mation related to the Medtronic servers. Invalid session key entered. Try to re-enter the session key.

7.3 Use RemoteView

To use RemoteView, the programmer needs to be connected to your clinic’s network. Before you begin, make sure your network is configured correctly.
Note: The remote viewer must have the appropriate remote viewing software installed to use RemoteView.

7.3.1 Configure RemoteView network

The programmer user needs to configure the RemoteView Network the first time the RemoteView Network is used by selecting the correct geographic designation as follows:
United States and World–choose this designation if the programmer user is physically located in the United States or anywhere else in the world except for Europe, as defined below.
88 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090
Europe–choose this designation if the programmer user is physically located in Europe.
“Europe” means: Austria, Belgium, Bulgaria, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Ireland, Italy, Latvia, Liechtenstein, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, or the United Kingdom.
1. Press the Programmer icon, and then select RemoteView Network Configuration….
2. From the RemoteView Network Configuration screen, select the available field to display the geographic designation options.
3. Select either “United States and World” or “Europe”, depending on your location.
4. Press [OK].
7.3.1.1 Server locations
These geographical designations correspond to the location of the Medtronic server, through which the connection between the programmer user and the remote viewer is established. The server corresponding to the “United States and World” designation is located in the United States. The server corresponding to the “Europe” designation is located in a Western European country.
7.3.1.2 Match programmer to the remote viewer’s system
When installing the remote viewing software, the remote viewer must select the geographical location that matches the designation chosen by the programmer user. For example, if the programmer user has chosen “Europe,” the remote viewer must also have chosen “Europe” when installing the remote viewing software, regardless of the remote viewer’s physical location. In this example, if the remote viewer has initially chosen “United States and World,” the remote viewer will need to download a second copy of the remote viewing software and choose “Europe” during this subsequent installation.

7.3.2 Enter RemoteView session key

A RemoteView session key is used by the remote viewer’s server to authenticate the connection with the programmer user. At the start of each session, the remote viewer will generate the session key on the remote user’s system and then share it with the programmer user by telephone. Programmer users use the following steps:
Reference Manual 89
Medtronic MEDTRONIC CARELINK® 2090
1. Call the remote user to start the consultation conversation and obtain the session key.
2. Press the RemoteView icon on the programmer task bar.
3. Enter the 7-digit session key generated by the remote system using the number keys and press [Enter].

7.3.3 End RemoteView connection

A RemoteView connection can be ended by either the programmer user or the remote viewer using one of the following methods:
The programmer user ends a device application session and presses [End Session] on the programmer screen. An active RemoteView connection will automatically terminate when the application software is exited.
The programmer user presses the RemoteView icon when RemoteView is connected to terminate a RemoteView connection.
The remote viewer clicks the “X” in the upper right corner of the connection window and clicks End Session.
When the connection has ended, the RemoteView icon on the programmer task bar will change from the green, connected state back to the gray, not connected state. The programmer screen image will be blank for the remote viewer.

7.4 Data privacy

Patient data may be viewed on the remote system during a RemoteView session. Medtronic provides the software on the programmer as well as the connectivity via a Medtronic server, which enables this to occur. Medtronic does not, however, collect, use, or retain any patient’s personal or health data via RemoteView.
The programmer user is responsible for complying with all applicable data privacy and data protection laws and regulations when choosing to allow a remote viewer to view the patient data via RemoteView, including obtaining consents and other agreements that may be required. Allowing the viewing of patient data by a remote viewer in another country could constitute a data transfer under the laws where the programmer user is located. For this purpose, please see Section 7.3.1 for the RemoteView server locations.
90 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090

8 SessionSync (Optional)

8.1 About SessionSync

SessionSync is an optionally installed feature that provides network connectivity between the programmer and the Medtronic Paceart data management system. Using your clinic’s network, the programmer can send downloaded device data and reports through SessionSync to the data management system. Reports are only sent to Paceart systems that accept them.
The SessionSync status icon and the SessionSync status screen provide information on the connection status of the programmer to the data management system.
You must configure the programmer network settings to allow for this data transfer.

8.2 Configure SessionSync

8.2.1 Configure the SessionSync network connection

Before you begin, make sure your network is configured correctly.
1. Press the Programmer icon, and then select SessionSync Network Configuration….
2. Enter the Clinic Name.
3. Enter the SessionSync network configuration information.
Note: If you do not have your SessionSync network configuration information, contact your clinic’s technical support or Medtronic Technical Support.
4. Press [OK].
Reference Manual 91
Medtronic MEDTRONIC CARELINK® 2090

Figure 29. SessionSync Network Configuration

8.3 Enable and disable SessionSync

1. Select Programmer > Preferences.
2. Select SessionSync from the Index menu.
3. Select [Enabled] to enable SessionSync or Select [Disabled] to disable SessionSync.
Note: The SessionSync icon in the task bar will be grayed out when the feature is disabled. SessionSync functions are not available within a patient session unless you have enabled this feature prior to starting a patient session.
92 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090

Figure 30. Preferences screen with Session Sync selected

8.4 SessionSync Status icon

The SessionSync status icon provides information on the network connection between the programmer and the data management system.

Figure 31. The task bar with the SessionSync Status icon

1 The SessionSync Status icon
If SessionSync is not installed on the programmer, the icon is not visible in the task bar.

Figure 32. Parts of the SessionSync Status icon

1 The data management system status 2 The connection status between the programmer and the data management system 3 The programmer status
Reference Manual 93
Medtronic MEDTRONIC CARELINK® 2090
The sections of the SessionSync Status icon change colors to indicate data ready for transfer, a valid connection between the programmer and the data management system, and successful data transfer to the data management system.
Note: When the whole icon is grayed out, SessionSync has been disabled under the programmer preferences.

Table 6. SessionSync Status icon states

Part of SessionSync Status Icon Color What the color indicates
Programmer Gray No session data files are in the Transfer
Queue. Note: The transfer queue is the list of session data files that have been saved to the programmer hard disk but are wait­ing for transfer.
Blue Session data files exist in the Transfer
Queue
Connection Not visible No valid connection exists between the
programmer and the data management system
Green Valid connection exists between the pro-
Red circle with a line through it
Data Management System Gray No session data has been transferred to
Blue All session data has been successfully
grammer and the data management sys­tem
Device application in use does not sup­port SessionSync
the data management system
transferred to the data management sys­tem

8.5 Use Automatic SessionSync

Automatic SessionSync allows you to perform a SessionSync automatically at the end of a patient session. This feature is available for all SessionSync enabled devices.

8.5.1 Create SessionSync Report(s) to send to Paceart

1. Press the Reports icon, and then select Print Queue.
2. Select SessionSync PDF printer from full-size printer menu.
94 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090
Notes:
SessionSync PDF is not available in the Print Queue window on the desktop.
SessionSync PDF is available only if SessionSync is installed and enabled, and is only availabe to SessionSync enabled device applications.

8.5.2 Save the patient session with Manual SessionSync

1. Press the Session icon.
2. Select SessionSync….
3. The SessionSync - Saving Session Data On Programmer window is opened and for some devices an interrogation is automatically started. The SessionSync - Saving Session Data On Programmer window shows the progress of the save. The Programmer side of the SessionSync Status icon turns blue after the data has been saved on the Programmer hard disk. If the subsequent transfer is successful, the data management system side of the SessionSync Status icon turns blue.
4. Any reports that were printed using SessionSync PDF printer will be transmitted to Paceart systems that accept them.

8.5.3 End a patient session with Automatic SessionSync enabled

1. Press [End Session…]. If an interrogation is required before the SessionSync data transfer, the Interrogation Required window is displayed.
2. Verify that the Automatic SessionSync box is checked, and then press [End Now].
3. The SessionSync - Saving Session Data On Programmer window is opened. For some devices, an interrogation is automatically started. The SessionSync - Saving Session Data On Programmer window shows the progress of the save. The programmer side of the SessionSync Status icon turns blue after the data has been saved on the programmer hard disk. If the subsequent transfer is successful, the data management system side of the SessionSync Status icon turns blue.
4. Any reports that were printed using SessionSync PDF printer will be transmitted to Paceart systems that accept them.

8.6 Use Manual SessionSync for supported devices

Manual SessionSync allows you to send interrogated data to the Paceart data management system without ending the patient session on the Programmer.
Reference Manual 95
Medtronic MEDTRONIC CARELINK® 2090
Manual SessionSync is not available for all SessionSync supported devices. If manual SessionSync is available for a device, the SessionSync… option appears in the Session menu.

8.7 SessionSync error message descriptions

You may receive error or information messages at different times in the SessionSync process. For a list of error messages, see Table 7. If you have any issues with the programmer contact Medtronic Technical Support at the telephone number on the back cover of this manual.

Table 7. SessionSync error messages

Error Message What it means
Data Transfer Failed A device communication error has occurred during the interrog-
ation and you have cancelled out of the interrogation window. The session data has not been saved on the programmer hard disk. Do one of the following: Press [Retry] to retry the operation. Press [Cancel] to close the window.
Ending a Session without Auto­matic SessionSync
Interrogation Required You must conduct an interrogation before starting a SessionSync
Interrogate - Unsuccessful The programmer cannot interrogate the device. You must reposi-
Unable to Save Session Data The session data cannot be saved on the programmer hard disk.
You have cleared the Automatic SessionSync check box on the End Session window before pressing the [End Now] button.
data transfer for this device. Press [OK] to close the window.
tion the programming head. Do one of the following: Press [Retry] or [Continue] after repositioning the programming head. Press [Cancel] to close the window.
Do one of the following: Press [Save to Media] or [Save Session] to save the session data to media. Press [End Now] to end the session without saving the device data. Press [Cancel] to close the window without saving the device data.
96 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090

8.8 View SessionSync Status screen

The SessionSync Status screen displays information on the data files being transferred to the data management system using SessionSync. Each message includes the date, time, and event information for the associated SessionSync event.

8.9 Update SessionSync status

1. Select Programmer > SessionSync Status.
2. Select [Update Status].
Note: SessionSync status does not dynamically update when the window is open. To update, press the [Update Status] button.

8.10 Test the SessionSync network connection

Use the Test SessionSync feature when instructed to do so by Medtronic Technical Support, or if your clinic’s network is having difficulty connecting to the SessionSync network. If the SessionSync is enabled but no connection is made, clinic network support staff may wish to use results of the network connection tests to troubleshoot the connection. Contact Medtronic Technical Support if you need assistance in interpreting results.
Reference Manual 97
Medtronic MEDTRONIC CARELINK® 2090

8.10.1 Run the tests

1. Select Programmer > SessionSync Status.
2. Press [Test SessionSync].
The programmer displays the Test SessionSync dialog.
98 Reference Manual
Medtronic MEDTRONIC CARELINK® 2090
3. Press [Test SessionSync].
Several diagnostic tests are run, and their results are listed on the dialog. The tests include ipconfig command, nslookup command, tracert command, ping command, and telnet command.
After tests are complete, the programmer screen displays “Finished executing SSL call”.

8.10.2 Save network connection test results to media

1. Insert a diskette or connect a USB flash drive to the programmer.
2. On the Test SessionSync dialog, press [Save To Media].
Note: This button is available only if the Test SessionSync tests have been run.
The programmer displays the Save to Media dialog.
3. Press [Save].
A text file (ntwk_debug.log) containing the test results is saved to the diskette or USB flash drive. Download the file to a compatible computer for viewing and/or transfer to Medtronic personnel.
Reference Manual 99
Medtronic MEDTRONIC CARELINK® 2090

9 Service the programmer

9.1 Clean the system components

The exterior surfaces of the programmer and its accessories may be cleaned with a sponge or soft cloth moistened with water, mild detergent, hydrogen peroxide, or alcohol.
Thoroughly clean the surfaces of the programmer wires; the bottom surface of the programming head; and the wire connecting the programming head to the programmer.
Caution: Take care to prevent liquid from entering the programmer and programming head. Do not immerse the programmer or any accessories in any liquid or clean them with aromatic or chlorinated hydrocarbons.
The lead wires may be cleaned by wiping each lead wire with a sponge or soft cloth moistened with the cleaning material, then wiped with a sponge or soft cloth moistened with clean water, and then wiped dry.
The exterior surfaces of the lead wires can be cleaned up to 15 times with each of the following materials without functional degradation:
Green soap, green soap tincture, or alcohol-free hand soap
2% gluteraldehyde solution (such as Cidex)
Sodium hypochlorite (bleach solution 10% in water)

9.2 Sterilize the programming head, ECG cable, and lead wires

Except for the programming head or ECG cable and lead wires, the programmer and its accessories cannot be sterilized.
Caution: Do not autoclave the programming head or ECG cable and lead wires.
Visually inspect the cable and connections of the programming head after sterilizing. Do not use the programming head if it appears damaged. Damage includes, but is not limited to, deterioration of the cable insulation (brittleness, cracking, thinning, or bare spots). Do not use the programming head if the conductive wires are exposed.
100 Reference Manual
Loading...