Medtronic 2020C2 Technical Manual

CARELINK EXPRESS™ MONITOR
Model 2020B/2020C
Electromagnetic Compatibility Declaration
Caution: Federal Law (USA) restricts this device to sale by or on the order of a physician.
CareLink, CareLink Express, Medtronic, Medtronic CareLink
1 Electromagnetic Compatibility Declaration
The CareLink Express monitor should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the CareLink Express monitor should be observed to verify normal operation in the configuration in which it will be used.
Portable and mobile RF communications equipment can affect the CareLink Express monitor.
The CareLink Express monitor contains RF transmission and receiving capabilities. Consequently, it is possible that other equipment may interfere with the CareLink Express monitor even if that other equipment complies with CISPR emission requirements. The following is a technical summary of the RF communication properties:
Transmitting and receiving at 175 kHz:
Receiving bandwidth of the CareLink Express monitor: NA - NOTE 1
Modulation characteristics (2020B): Transmit using BFSK, receive using BPSK.
Modulation characteristics (2020C): Transmit and receive using pulsed AM (OOK).
Field strength (2020B): -2,5 dBuA/m at 10 meters - NOTE 2
Field strength (2020C): 11,5 dBuA/m at 10 meters - NOTE 2
NOTE 1 Receiver detects switching magnetic fields and thus has the capability to respond to any frequency in the LF (Low Frequency) area as long as the incidence angle is correct.
NOTE 2 Effective radiated power has been expressed as field strength given that it has been the regulated parameter measured for this frequency.
Guidance and manufacturer’s declaration – electromagnetic emissions
The CareLink Express monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the CareLink Express monitor should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment – guidance
RF emissions Group 1 The CareLink Express monitor uses RF
energy only for its internal function. Therefore,
CISPR 11
its RF emissions are very low and are not likely to cause interference in nearby electronic equipment.
RF emissions Class A The CareLink Express monitor is suitable for
use in all establishments other than domestic
CISPR 11
Harmonic emissions
Not Applicable
and those directly connected to the public low­voltage power supply network that supplies buildings used for domestic purposes.
IEC 61000-3-2
Voltage fluctuations/flicker
Not Applicable
emissions
IEC 61000-3-3
3