Medtronic 2020C2 Reference Guide

CareLink Express
2
3
4
5
6
7
8
1
2
1
TM
Monitor
2020B/2020C
Warning: For use with Medtronic implantable cardiac devices only.
1
Caution: Federal Law (USA) restricts this device to sale by or on
the order of a physician.

Intended use

The Medtronic CareLink Express monitor Models 2020B and 2020C are portable interrogation and data transfer tools used with Medtronic implantable cardiac devices.

Monitor features

1. Power button
2. Power Status light (green)
3. Reading Status lights (green)
4. Sending Status lights (green)
5. Sending Complete Status light (green)
6. Battery Status light (orange)
7. Antenna Position Status light (orange)
8. Telephone Line Status light (orange
www.medtronic.com
Manufacturer
Medtronic, Inc. 710 Medtronic Parkway Minneapolis, MN 55432-5604 USA Tel. +1-763-514-4000 Fax +1-763-514-4879
Medtronic E.C. Authorized Representative
Medtronic B.V. Earl Bakkenstraat 10 6422 PJ Heerlen The Netherlands Tel. +31-45-566-8000 Fax +31-45-566-8668
The following list includes trademarks or registered trademarks of Medtronic in the United States and possibly in other countries. All other trademarks are the property of their respective owners.
CareLink, CareLink Express, Medtronic, Medtronic CareLink
© Medtronic, Inc. 2011 M947433A001D 2011-10-06
Europe/Africa/Middle East Headquarters
Medtronic International Trading Sàrl Route du Molliau 31 Case Postale 84 CH-1131 Tolochenaz Switzerland Tel. +41-21-802-7000 Fax +41-21-802-7900
*M947433A001*

Using the monitor

• Refer to the supported device card to determine the
2
appropriate monitor to use when interrogating the patient’s implanted cardiac device.
• Interrogate the patient’s device with the Model 2020B first. If the Model 2020B does not identify the patient’s device, then interrogate it with the Model 2020C.
• The monitor must be connected to an analog telephone line wall socket within 15 minutes of the interrogation to complete the transmission to the Medtronic CareLink network. If the monitor is not connected to an analog telephone line socket within 15 minutes, the monitor powers down, the patient’s data is lost, and the interrogation must be performed again.
• If a sterile barrier is required between the monitor antenna and the patient, cover the antenna and antenna cable with Medtronic Sterile Sleeve Model 6177, and follow standard sterile procedures.
1. Press the power button.
2. Hold the antenna directly over the patient’s implanted
cardiac device to begin interrogation. — When the monitor identifies the implanted cardiac
device, the first green Reading Status light turns solid and the monitor beeps twice.
— The patient’s implanted device may emit a tone when
the antenna is placed over it.
— The monitor completes the device interrogation when
Reading Status lights turn solid and the monitor beeps twice.
— Reposition the antenna if the orange Antenna Position
Status light illuminates.
— Remove the antenna when the interrogation is
complete.
3
3. Connect the monitor to an analog telephone line wall socket within 15 minutes of completing the device interrogation. Approximately every 25 seconds the monitor beeps 5 times, until the monitor is connected to an analog telephone line wall socket. The monitor stops beeping when data transmission begins. The monitor automatically begins to transmit the data from the implantable cardiac device to the Medtronic CareLink network.

Monitor setup

1. Install the four AA batteries, with proper polarity, into the battery compartment on the bottom of the monitor.
2. Select the appropriate telephone line settings.
4
• The dialing type selector switch is marked T (Tone) and P (Pulse). Position the switch to select the appropriate type of telephone dialing system for your facility.
• The dialing prefix selector configures the monitor to obtain a telephone line for placing a call. Select N (None) if no prefix is required to reach an external telephone line. If a number must be dialed to obtain an external line, position the switch to the position that corresponds to the required dialing prefix number.
• Put the monitor in a place where the patient can sit comfortably and the front of the monitor is visible.
• The monitor must be connected to an analog telephone line wall socket when updating the monitor software or sending a transmission to the Medtronic CareLink network.
• Before using the monitor, turn off any television sets or computer monitors that are within 6 feet of the monitor or its antenna.
• Do not place the monitor anywhere it could get wet.
Caution: Do not get the monitor wet. Water can damage the monitor. If the monitor is exposed to water, contact your Medtronic representative.

Connecting the monitor to a telephone line wall socket

The monitor does not need to be connected to an analog telephone line wall socket during the interrogation of a patient’s implanted cardiac device. However, if the monitor is not connected to an analog telephone line wall socket during interrogation, connect it within 15 minutes of the interrogation to complete the transmission to the Medtronic CareLink network.
1. Locate the monitor and the telephone cord.
2. Locate the sockets on the left side of the monitor.
3. Plug one end of the telephone cord into the telephone sockets on the left side of the monitor.

Choose a place to set up the monitor

• Consider the following recommendations for choosing a place to set up the monitor:
• Put the monitor in a place where the patient can sit comfortably and the front of the monitor is visible.
• The monitor must be connected to an analog telephone line wall socket when updating the monitor software or sending a transmission to the Medtronic CareLink network.
• Before using the monitor, turn off any television sets or computer monitors that are within 6 feet of the monitor or its antenna.
• Do not place the monitor anywhere it could get wet. Caution: Do not get the monitor wet. Water can damage the monitor. If the monitor is exposed to water, contact your Medtronic representative.
5

Connecting the monitor to a telephone line wall socket

6
The monitor does not need to be connected to an analog telephone line wall socket during the interrogation of a patient’s implanted cardiac device. However, if the monitor is not connected to an analog telephone line wall socket during interrogation, connect it within 15 minutes of the interrogation to complete the transmission to the Medtronic CareLink network.
1. Locate the monitor and the telephone cord.
2. Locate the sockets on the left side of the monitor.
3. Plug one end of the telephone cord into the telephone sockets on the left side of the monitor.
4. Connect the monitor to an analog telephone line wall socket within 15 minutes of completing the device interrogation.
7
• If the monitor is turned on and plugged into an analog telephone line wall socket, and if you have successfully interrogated a device within the last 15 minutes, the monitor will beep twice and automatically begin dialing the programmed telephone number. The transmission completes when all of the Sending Status lights turn solid, the Sending Complete Status light turns solid, and the monitor beeps twice.
• The monitor turns off automatically after two minutes, once the transmission completes.

Update the monitor software to support newer devices

1. Verify that the telephone cord is connected to one of the monitor telephone sockets and an analog telephone line wall socket.
• Press and hold the power button for 10 seconds.
• The Telephone Line Status indicator light illuminates.
• The first Sending Status light turns solid, and then the
remaining two lights turn solid, as the monitor connects to the Medtronic CareLink network and updates the monitor software.
• The software download begins.
• The Sending Complete light turns solid and the monitor
8
beeps twice when the software update completes.
• The download completes in approximately four minutes.

Troubleshooting

The monitor status lights indicate system status and errors. Orange lights indicate battery, antenna, or telephone line errors. Green lights indicate normal monitor status, but can also indicate errors or unsuccessful operation.

Orange Battery Status light blinks

The orange Battery Status light should not be lit when you are using the monitor. If the monitor beeps and the orange Battery Status light blinks:
• The batteries are low and should be replaced.
• Verify that the batteries are installed correctly, if no status lights come on after pressing the Power button.

Orange Antenna Position Status light blinks

The orange Antenna Position Status light should not be lit when you are using the monitor. If the orange Antenna Position Status light begins blinking and you hear short beeps:
• Verify that the hand strap on the top of the antenna is facing out and away from the patient.
• Verify that the antenna is placed over the patient’s implanted cardiac device. If necessary, reposition the antenna over the patient’s implanted cardiac device.

Notes:

• If the monitor does not identify the patient’s device, the orange Antenna Position Status light blinks and you hear short beeps.
• When the antenna is repositioned correctly, the orange antenna Position Status light turns off, the green status lights blinks, and the monitor resumes data collection.

Orange Telephone Line Status light blinks

• The orange Telephone Line Status light should not be lit when you are using the monitor. If the monitor beeps and the orange Telephone Line Status light blinks:
• Check that the telephone cord is fully inserted into the telephone line socket on the left side of the monitor and the analog telephone line wall socket.
• Check that the telephone line settings are correctly positioned. If you need help determining the type of telephone line, contact your local telephone carrier.
• Verify that the correct prefix number is selected if necessary for your facility.
• Verify that the monitor is connected to an analog telephone line and not a digital VoIP or fiber optic telephone line. If necessary, contact your telephone company to verify the type of telephone line at the facility’s location.

Green Power Status light does not illuminate

The green Power Status light illuminates when the monitor is powered on. If the green Power Status light does not illuminate:
• Verify that the batteries are installed correctly. If necessary, replace the batteries.

Green Reading Status lights do not illuminate during device interrogation

The green Reading Status lights verify that the monitor has successfully interrogated the patient’s implanted cardiac device. The first green light illuminates and the monitor beeps twice when the monitor identifies the patient’s implanted cardiac device. The remaining 4 lights first blink, and then turn solid as the monitor interrogates the patient’s device. The interrogation completes when all of the lights turn solid and the monitor beeps twice.
9
• If the Reading Status lights stop blinking and remain lit for
10
more than 3 minutes, or if the monitor turns off before the data is read, repeat steps 1 through 4 in “Using the monitor”.
• If the monitor shuts off after the third Reading Status light illuminates, the monitor has not successfully interrogated the device. Add additional fabric layers, such as a towel folded over multiple times, between the antenna and the patient’s implanted device, and repeat steps 1 through 4 in “Using the monitor”.

Green Sending Status or Sending Complete Status lights do not illuminate

The green Sending Status and Sending Complete Status lights verify that the monitor has successfully transmitted the data from the interrogation of the patient’s implanted cardiac device to the Medtronic CareLink network. If the green Sending Status lights do not illuminate, if the Sending Status lights remain lit for more than 1 hour, if the Sending Complete Status light does not illuminate, or if the monitor turns off before the patient’s data is transmitted:
• Verify all telephone settings.
• Repeat steps 1 through 4 in “Using the monitor”.
11
Loading...