Appendix A System Components ............................................................................................ 173
Appendix B Product Specications ........................................................................................ 174
Appendix C EMC Manual Requirements Guidance Document ......................................... 176
Appendix D Symbol Denitions ............................................................................................... 180
HVAD™ Instructions for Use
1.0 Introduction
1
2
1.1 Introduction ..................2
1.2 Indications for Use .......2
1.3 Contraindications .........2
1.4 Warnings ......................3
1.5 Precautions ..................7
1.6 Potential
Complications............... 10
1.7 Pivotal US Clinical Study:
Bridge-to-Transplant ..11
1.8 Post Approval Study
Follow-up of the Pivotal
US Clinical Study:
Bridge-to-Transplant .. 20
Foreword
The HeartWare™ HVAD™ System is indicated for use
under the direct supervision of a licensed healthcare
practitioner or by personnel trained in its proper use.
Clinical users include physicians, registered nurses,
perfusionists and biomedical engineers. Implant of
the device must be performed by a qualied cardiac
surgeon trained by HeartWare-authorized personnel.
Clinical users of the HVAD System should attend
HeartWare training, should have a working knowledge
of the principles of ventricular assist devices (VADs),
and should be aware of the physical and psychological
needs of patients undergoing VAD support. Patients and
caregivers should complete a user training program
and demonstrate their ability to use the system prior to
independence.
Clinicians should read the entire Instructions for Use
before system operation. This manual may serve
as a reference for detailed information including
specic information on device function, system setup,
implant and maintenance. This manual is not intended
to replace comprehensive educational programs
or to supersede acquired knowledge or proper
medical judgment.
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
1.9 US Clinical Study:
Destination Therapy ....30
1.10 Destination Therapy
Supplemental Study 47
1.11 North American Clinical
Study: LATERAL ........64
Introduction
9
Reference
Guides
10
Appendix
1
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Accessories
1.0 Introduction
In this manual, there will be the following symbols:
Indicates there is more
information available in the
manual and the location in
the manual
WARNING! Carefully read this entire manual prior to implanting or operating the device. Improper
operation of the system and potential harm to the patient and to the user could result.
Identies information
as a Warning
A Warning is a statement
about the possibility of
injury, death or other serious
adverse reaction that is
associated with the use or
misuse of the device
Identies information
as a Caution or Precaution
A Caution is a statement
that not following the
instruction may lead to
device misuse, malfunction
or damage
Indicates there is
a quick reference
guide available
1.1 Introduction
Monitor
6
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
9
The HVAD System is designed to assist a weakened, poorly
functioning left ventricle. The HVAD System is designed
for in-hospital and out-of-hospital settings, including transportation via
xed wing aircraft or helicopter. The HVAD System utilizes
a centrifugal blood pump, the HVAD Pump (the “pump”), which is
implanted in the pericardial space with left ventricular apex to
ascending aortic cannulation for left ventricular support. The inow
conduit, which is partially sintered, is integrated with the pump and
Explant
a 10 mm gel impregnated outow graft with a strain relief is
attached to the pump. A percutaneous driveline connects the
pump to an external controller. The controller, powered by two
batteries or by one battery and electricity from a wall or car outlet,
regulates pump function and monitors the system. The monitor is used to
display system performance and to change controller operating
parameters. A battery charger is also included.
All components of the HVAD System are designed to be
used only in conjunction with each other. They are neither compatible
nor intended to be used with other manufacturer’s devices.
Figure 1:
HVAD System
1.2 Indications for Use
The HVAD System is indicated for hemodynamic support in patients with advanced, refractory
left ventricular heart failure; either as a Bridge to Cardiac Transplantation (BTT), myocardial
recovery, or as Destination Therapy (DT) in patients for whom subsequent transplantation is not
planned.
Reference
10
Guides
Appendix
1.3 Contraindications
The HVAD System is contraindicated in patients who cannot tolerate anticoagulation therapy.
2HVAD™ Instructions for Use
1.4 Warnings
1
WARNINGS
1. WARNING! Serious and life threatening adverse events, including stroke, have been
associated with use of this device. The risk of death as a result of stroke has been
observed in randomized clinical trials to be higher with the HVAD than with alternative
treatment options. The HVAD has been associated with a rate of stroke of 22% at one
year and 29.7% at two years. A blood pressure management protocol may reduce
the overall incidence of stroke to 16.9% at one year and may reduce the incidence
of disabling strokes at one year from 8.1% to 6.5%. A user must fully consider the risks of
this device with that of other treatment modalities before deciding to proceed with
device implantation. Please refer to section 1.8 (“Safety and Effectiveness Results”)
for a summary of the stroke data. To mitigate the risk of stroke, please adhere to the
following patient management guidelines:
• Monitor and treat mean arterial pressure. Maintain MAP less than 85 mmHg as
tolerated.
• Speed on the HVAD Pump should be set to maintain adequate pump ow index, this
generally does not need exceed 2.6 L/min/m2.
• Maintain anticoagulation within the recommended INR range of 2.0-3.0.
• Daily aspirin dose should be > 81 mg and platelet inhibition should be evaluated and
adjust ASA mono-therapy accordingly or consider combination therapy such as ASA
81 mg plus Aggrenox® (ASA plus extended –release dipyridamole) or daily ASA 81 mg
plus Plavix 75 mg.
2. WARNING! Carefully read this entire manual prior to implanting or operating the
device. Improper operation of the system and potential harm to the patient and to
the user could result.
3. WARNING! DO NOT use the HVAD System in pregnant women. Any woman receiving
an HVAD System who is of childbearing age and sexually active should use a
reliable method of birth control. Use of anticoagulants during pregnancy has been
associated with birth defects and bleeding.
4. WARNING! The Instructions for Use (IFU) is intended to be used by physicians, nurses,
and other clinical professionals. Setup and operation of this device should only be
undertaken by personnel who have completed an HVAD System product training
program. A thorough understanding of technical principles, clinical applications and
risks associated with the HVAD System is required before using this product. Failure to
understand these principles, applications and risks may result in improper operation of
the system and potential harm to the patient or to the user.
5. WARNING! ALWAYS connect an AC Adapter to the controller before relaxing or
sleeping. Power from an electrical outlet (AC Adapter) provides power for an
unlimited period of time.
6. WARNING! NEVER disconnect both power sources (batteries and AC or DC adapter)
at the same time since this will stop the pump. At least one power source must be
connected at all times.
7. WARNING! DO NOT allow patients to shower until they have received permission from
their clinician to do so. Patients who shower must use the HeartWare™ Shower Bag.
8. WARNING! DO NOT allow hearing impaired patients to shower unless their caregiver is
close by to hear alarms.
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
3Introduction
1
Introduction
1.4 Warnings (continued)
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Monitor
6
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
Accessories
Explant
WARNINGS
9. WARNING! DO NOT plug the controller into an AC wall outlet during showers; to eliminate
the possibility of a severe electrical shock, it should be connected to two batteries.
10. WARNING! DO NOT allow patients to take a bath or swim, as this may damage
HVAD System components and/or result in driveline exit site infection.
11. WARNING! DO NOT submerge HVAD System components in water or other uid as this
may damage them. If this happens, contact HeartWare.
12. WARNING! DO NOT allow water or other uids to enter the controller, power adapters,
batteries, battery charger or connectors, as this may damage HVAD System
components. If this happens, contact HeartWare.
13. WARNING! DO NOT use any components other than those supplied by HeartWare with
the HVAD System, as this may affect HVAD System operation.
14. WARNING! Damaged equipment should be reported to HeartWare and replaced.
15. WARNING! DO NOT rely only on ow estimation to assess cardiac output. An average
estimated ow on the monitor or Controller Display of less than 2.0 L/min, or greater than
10.0 L/min may indicate an electrical fault, incorrect hematocrit entry or an occlusion
and/or thrombus or other materials (e.g. tissue fragments) in the device. Inaccurate
assessment of HVAD Pump ow may lead to less than optimal treatment.
16. WARNING! DO NOT grasp the driveline cable as this may damage the driveline. To
remove the driveline from the controller, rst pull back the driveline cover then grasp and
pull the driveline connector.
17. WARNING! DO NOT disconnect the driveline from the controller or the pump will stop. If
this happens, reconnect the driveline to the controller as soon as possible to restart the
pump.
18. WARNING! DO NOT plug the HeartWare™ Battery Charger or monitor AC adapter into an
electrical outlet which is not properly grounded or you may receive a serious electrical
shock.
19. WARNING! DO NOT operate the controller in temperatures less than -20°C (-4°F) or
greater than +50°C (+122°F) or the controller may fail.
20. WARNING! AVOID devices and conditions that may induce strong static discharges
(e.g., television or computer monitor screens) as electrostatic discharges can damage
the electrical parts of the system and cause the VAD to perform improperly or stop.
21. WARNING! The HVAD System components should not be used adjacent to or stacked with
equipment other than specied in the IFU. If adjacent to or stacked use is necessary, the
HVAD System and other equipment should be observed to verify normal operation.
22. WARNING! ALWAYS have a back-up controller handy and, whenever possible, a
caregiver nearby when changing power sources or controllers. Be watchful for unusual
changes in power or ow alarms for a period of time following equipment changes.
23. WARNING! DO NOT drop the controller or other equipment. Dropping the controller
could cause sudden stoppage of the pump. Dropped equipment should be reported
to HeartWare and inspected.
24. WARNING! DO NOT disconnect the driveline or power sources from the controller while
cleaning it or the pump will stop. If this happens, reconnect the driveline to the controller
as soon as possible to restart the pump.
10
Appendix
4HVAD™ Instructions for Use
1.4 Warnings (continued)
1
WARNINGS
25. WARNING! NEVER clean the battery charger with the power on, as this may lead to an
electrical shock.
26. WARNING! AVOID areas with high magnetic forces such as theft detection devices or
airport security systems, as this may affect HVAD System operation.
27. WARNING! Keep mobile phones at least 20 inches (50 centimeters) away from the
controller, as mobile phones may interfere with controller operation.
28. WARNING! DO NOT let the patient have a magnetic resonance imaging (MRI)
procedure while implanted with the HVAD Pump. Doing so could cause harm to the
patient or could cause the pump to stop.
29. WARNING! DO NOT apply high power electrical treatment (e.g., deep tissue heating
which can be used for treatment of arthritis and/or some injuries) directly to the patient,
as this may affect HVAD System operation.
30. WARNING! AVOID therapeutic levels of ultrasound energy, as the device may
inadvertently concentrate the ultrasound eld and cause harm.
31. WARNING! AVOID therapeutic ionizing radiation since it may damage the device.
This damage may not be immediately detectable.
32. WARNING! ALWAYS investigate, and if possible, correct the cause of any alarm. Silencing
an alarm does not resolve the alarm condition.
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
33. WARNING! NEVER clean the monitor with the power on, as this may lead to an electrical
shock. DO NOT use alcohol or detergent on the monitor display. Gently wipe the display
with a soft, lint free cloth.
34. WARNING! The HVAD Pump may cause interference with AICDs. If electromagnetic
interference occurs, it may lead to inappropriate shocks, arrhythmia and possibly death.
The occurrence of electromagnetic interference with AICD sensing may require
adjustment of lead sensitivity, proximal placement of new leads or replacement of
an existing sensing lead.
35. WARNING! Keep power connected to the controller after setting up the primary controller
to minimize the risk of air embolus during implant. Disconnecting and then reconnecting
power will result in the controller starting the pump as soon as the driveline is connected.
36. WARNING! DO NOT use if package is damaged or opened. Sterile components are
intended for single use only. DO NOT re-sterilize or re-use as this will increase the risk of
infection.
37. WARNING! ALWAYS check for an audible click when connecting the driveline to the
controller or driveline extension cable. Failure to ensure a secure connection may cause
an electrical fault.
38. WARNING! NEVER turn on the HVAD Pump in air as this may damage the pump. DO
NOT use an HVAD Pump that was turned on without total submersion in uid during
the pre-implant test and prior to implantation: The HVAD Pump must be completely
submerged in uid before being turned on.
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
5Introduction
1
Introduction
1.4 Warnings (continued)
2
3
™
4
Peripherals
5
6
Surgical
7
Patient
8
Alarms and
9
Reference
10
PumpOverview
HVAD
and
Monitor
Implant and
Accessories
Explant
Management
Emergencies
Guides
WARNINGS
39. WARNING! DO NOT implant gel impregnated vascular prostheses in patients who exhibit
sensitivity to polyester or materials of bovine origin, as severe reactions may occur.
40. WARNING! The manufacturing process for gelatin sealed vascular grafts uses the cross-
linking agent formaldehyde to achieve the graft performance. All gelatin sealed grafts
are thoroughly rinsed with reverse osmosis water to reduce residual formaldehyde,
however residual amounts may be present in the nished graft. Formaldehyde is
also found at low levels naturally in the body, some of which is derived from food.
Formaldehyde is known to be mutagenic and carcinogenic. The risks of these potential
harms from the product have not been established clinically.
41. WARNING! DO NOT allow the Gelweave™ prostheses non-sterile foil pouch or outer tray
to be introduced to the sterile eld or the sterile eld will be contaminated. Only the
innermost tray is sterile.
42. WARNING! DO NOT preclot the outow graft. Preclotting may disrupt the gel matrix,
resulting in bleeding. Gelweave™ prostheses are sealed grafts and must not be
preclotted.
43. WARNING! DO NOT implant the Gelweave™ prostheses more than one month after
removal from the foil pouch. This may disrupt the gel matrix, resulting in bleeding.
44. WARNING! DO NOT allow anyone but a surgeon, physician’s assistant or surgical assistant
trained in the procedure to attach the outow graft to the pump, as a loose graft
connection may lead to bleeding and/or an air embolus.
45. WARNING! ALWAYS position the clamp screw so that it is located on the inner side of the
outow conduit to avoid tissue irritation or damage.
46. WARNING! DO NOT over-loosen the sewing ring’s screw or it may fall off the sewing ring
and be lost in the sterile eld.
47. WARNING! DO NOT cut the outow graft too short or too long, or it may kink. Prior to
chest closure, ensure that the graft is not kinked or compressed. A kinked or compressed
outow graft may lead to reduced ow and/or thrombus formation.
48. WARNING! DO NOT immerse the Gelweave™ grafts in saline for longer than 5 minutes.
Longer periods of soaking in saline may disrupt the gel matrix, resulting in bleeding.
49. WARNING! ALWAYS position the driveline exit site so that the tunneler does not contact
any vital organs or structures.
50. WARNING! DO NOT grasp the driveline and pull as this may damage the driveline. To
remove the driveline cap from the driveline, unscrew the outer sleeve, then pull back
on the grooved part of the connector.
51. WARNING! ALWAYS remove all air from the HVAD Pump and its conduits to reduce risk of
air embolus.
52. WARNING! DO NOT de-air the HVAD Pump when there is inadequate blood volume in
the HVAD Pump or leaks in the inow/outow connections, as air may enter the HVAD
Pump and outow graft resulting in a delay in de-airing and possible air embolism.
53. WARNING! At HVAD Pump explant the percutaneous driveline is not sterile; therefore
ensure that the driveline does not contaminate the sterile eld.
54. WARNING! ALWAYS check the Controller Display for any information regarding an
alarm when using loud machinery or in the vicinity of loud noises as the alarms may
not be audible.
55. WARNING! ALWAYS replace a controller with a blank display or no audible alarms. This
condition is predictive of a controller failure.
56. WARNING! ALWAYS switch to the back-up controller if there is a [Controller Failed] alarm
since the HVAD Pump may not be running.
57. WARNING! ALWAYS respond to low battery alarms. Silencing an alarm does not resolve
the alarm condition and will eventually deplete the batteries.
Appendix
6HVAD™ Instructions for Use
1.4 Warnings (continued)
WARNINGS
58. WARNING! DO NOT attach the alarm adapter to a controller that is connected to a
running pump. The alarm adapter silences the [No Power] alarm and should only be
attached to a controller that has failed or malfunctioned and is no longer connected to
a running pump.
59. WARNING! ALWAYS keep a spare controller and fully-charged spare batteries at a
temperature between 0°C and 50°C (+32°F to 122°F) available at all times in case of an
emergency.
60. WARNING! DO NOT remove the driveline cover from the driveline. Maintaining proper
driveline cover attachment prevents accidental disconnection which will lead to a
pump stop.
1.5 Precautions
1
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
PRECAUTIONS
1. CAUTION: Safety and effectiveness in persons less than 18 years of age and in
persons with a BSA of less than 1.5 m2 have not been established.
2. CAUTION: The HVAD System has had limited use in patients with articial mitral or
aortic valves and therefore the risks are currently unknown. Caution should be used
in selecting patients with articial mitral or aortic valves for HVAD System therapy.
3. CAUTION: ONLY use HVAD™ Controllers on one patient to avoid risks associated with
an inadvertent mismatch of controller pump speed settings.
4. CAUTION: The HeartWare™ Waist Pack and the HeartWare™ Shoulder Pack contain
magnetic closures. Patients with an internal cardiac debrillator (ICD) or pacemaker
should keep the pack away from their chest, including when sleeping. Per
pacemaker and ICD manufacturer guidelines, magnets should be kept at least 6
inches (15 centimeters) away from the pacemaker or ICD (please refer to manufacturer
guidelines for additional information).
5. CAUTION: DO NOT pull, kink or twist the driveline or the power cables, as these may
damage the driveline. Special care should be taken not to twist the driveline while
sitting, getting out of bed, adjusting controller or power sources, or when using the
shower bag.
6. CAUTION: DO NOT attempt to repair or service any components of the HVAD
System. If HVAD System equipment malfunctions, contact HeartWare.
7. CAUTION: Manual changes to the speed will immediately disable the [Ventricular
Suction Detection] alarm. An “Sx Off” will be displayed on the monitor screen below the
“Fixed” mode display. The [Ventricular Suction Detection] alarm will have to be
re-activated.
8. CAUTION: DO NOT enable the [Ventricular Suction Detection] alarm while the patient is
in a suction condition. To optimize operation of the suction detection the patient should
be hemodynamically stable prior to enabling the [Ventricular Suction Detection] alarm.
9. CAUTION: ALWAYS keep all connectors free of liquid, dust and dirt, or the HVAD
System may not function as intended.
5
Monitor
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
7Introduction
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Monitor
6
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
Accessories
Explant
1.5 Precautions (continued)
PRECAUTIONS
10. CAUTION: DO NOT force connectors together without proper alignment. Forcing
together misaligned connectors may damage the connectors.
11. CAUTION: ALWAYS conrm that the power cables are properly locked on the controller
by gently pulling the cable near the controller power connector or the power cables
may come loose and result in an alarm or the pump stopping.
12. CAUTION: ALWAYS recharge fully depleted batteries within 24 hours to avoid
permanent battery damage.
13. CAUTION: DO NOT expose batteries to temperatures outside the storage and
operational ranges or they may provide less runtime or may be unable to start a pump
in an emergency. To preserve battery life, batteries should be stored at room
temperature.
Battery operating and storage temperatures:
a. Operating: discharge (normal use with the HVAD System): 0°C to +50°C (+32°F to
+122°F). Operation at temperatures below 0°C will temporarily reduce battery
capacity but the battery will operate.
b. Storage: -20°C to +25°C (-4°F to +77°F). Long term storage outside of this range may
permanently reduce the battery capacity. Best condition for storage is at room
temperature.
14. CAUTION: ALWAYS keep batteries away from children. Children may be harmed by
damaged batteries or components.
15. CAUTION: DO NOT disassemble, crush, or puncture a battery.
16. CAUTION: DO NOT use a damaged battery. Battery function is unknown if the battery
is damaged.
17. CAUTION: DO NOT short circuit the external contacts on a battery since this may result
in battery damage.
18. CAUTION: DO NOT touch the uid if a battery pack is leaking uid. Dispose of a leaking
battery pack. In case of eye contact with uid, DO NOT rub eyes. Immediately ush
eyes thoroughly with water for at least 15 minutes, lifting upper and lower lids, until no
evidence of the uid remains. Seek medical attention.
19. CAUTION: DO NOT expose batteries to excessive shock or vibration since this may affect
battery operation.
20. CAUTION: DO NOT dispose of a battery in re or water. Dispose of batteries according
to federal, state, and local regulations.
21. CAUTION: DO NOT place batteries in water or liquid.
22. CAUTION: ONLY use the HeartWare™ Battery Charger to charge HeartWare™ Batteries.
Other battery chargers will not charge the batteries and may damage them.
23. CAUTION: ALWAYS wait until the “Ready” light turns on to disconnect the battery from
the battery charger. If this is not followed over consecutive charging cycles, the Battery
Capacity Display will not function properly and may convey misleading battery
capacity.
24. CAUTION: ALWAYS fully charge the monitor’s internal battery prior to patient use.
25. CAUTION: DO NOT allow patients to touch the monitor, as this may lead to the entering
of unwanted HVAD System parameters.
10
Appendix
8HVAD™ Instructions for Use
1.5 Precautions (continued)
PRECAUTIONS
26. CAUTION: DO NOT use the “Set Defaults” button on monitor REF1510 when a
controller is connected to a patient. Pressing it will erase all patient VAD parameter
information from the controller.
27. CAUTION: DO NOT use HeartWare equipment in the presence of a ammable
anesthetic mixture with air or with oxygen or nitrous oxide. (NOTE: Flammable
anesthetics are typically ether based).
28. CAUTION: A back-up controller should always be available and programmed
identically to the primary controller.
29. CAUTION: DO NOT exert excessive tension or force on the Gelweave™ prostheses as it will damage the polyester bers and the gelatin impregnation, which may result
in bleeding.
30. CAUTION: ALWAYS ensure the inow cannula position is pointed toward the mitral valve
and parallel to the interventricular septum to optimize HVAD Pump operation.
31. CAUTION: ALWAYS position the sewing ring to permit access to its screw after
cannulation.
32. CAUTION: ALWAYS use round body taper point needles when implanting Gelweave™
prostheses to minimize ber damage. A kinked or compressed outow graft may lead
to reduced ow and/or thrombus formation.
33. CAUTION: The driveline connector is made of nickel-coated brass which may cause
a rash in patients with a nickel allergy.
34. CAUTION: ALWAYS be aware of the position of the driveline to avoid damage by
surgical instruments and needles during HVAD Pump implantation and/or re-operation.
35. CAUTION: ALWAYS use the smallest possible needle for de-airing; 19-gauge is normally
sufcient. Hypodermic needles have a cutting point which may result in blood leakage
and may require repair by suturing.
36. CAUTION: DO NOT rely on HVAD Pump ow estimation during the de-airing procedure.
Flow estimation may not be accurate.
37. CAUTION: ALWAYS examine the driveline for evidence of tears, punctures or
breakdown of any of the material during exit site dressing changes. Driveline damage
may affect HVAD System performance.
38. CAUTION: DO NOT expose the driveline to direct or indirect sunlight. ALWAYS keep the
driveline completely covered when in the sun. Instruct patients not to use tanning lights
or black lights. The light from these sources contains ultraviolet radiation which may
damage the outer sheath of the driveline.
39. CAUTION: DO NOT use prophylactic topical antibiotic ointments such as silver
sulfadiazine, povidone iodine (betadine), or polymyxin-neomycin-bacitracin ointment
on the exit site. These ointments can injure the tissue next to the driveline.
40. CAUTION: Chest compressions may pose a risk due to pump location and position of
the outow graft on the aorta - use clinical judgment. If chest compressions have been
administered, conrm function and positioning of HVAD Pump.
41. CAUTION: Speeds below 2400 RPM or above 3200 RPM should be used with caution.
42. CAUTION: The safety and effectiveness of the Lavare™ Cycle has not been evaluated
clinically.
1
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
9Introduction
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Monitor
6
Surgical
Implant and
7
Patient
Accessories
Explant
Management
1.6 Potential Complications
Implantation of a VAD is an invasive procedure requiring general anesthesia and entry
into the thoracic cavity. These surgical procedures are associated with numerous risks. Risks
associated with the implant procedure and use of the device may include, but are not
limited to, the following1:
Other than death, the adverse events are listed
in alphabetical order according to INTERMACS
categories.
1.7 Pivotal US Clinical Study: Bridge-to-Transplant
1
Pivotal Clinical Study Design
This was a multi-center, prospective, contemporaneous control trial. The trial was non-randomized
and open label. Enrollment in the study is complete, subjects have all reached the primary
endpoint as described and specied in the protocol, but follow-up of subjects is ongoing.
Subjects were consented for participation and then assessed against the inclusion and
exclusion criteria for participation in the study and implantation of the HVAD Pump. After the
surgical recovery period, patients were allowed to leave the hospital if they met additional
criteria for hospital discharge. Each patient was followed to 180 days, death, device explant
for recovery, or cardiac transplantation, whichever occurred rst.
Patient outcomes were compared to a contemporaneously treated cohort of patients as
recorded in the Interagency Registry for Mechanical Assisted Circulatory Support (INTERMACS®).
All patients enrolled in the INTERMACS® registry over the same enrollment period as the trial that
met the control group inclusion and exclusion criteria comprised the control group.
Study Objectives
Primary Objective
The purpose of the HVAD System study was to evaluate the safety and effectiveness of the
HVAD System in patients listed for cardiac transplantation with refractory, advanced heart
failure at risk of death. The primary endpoint is success at 180 days which was dened as alive
on the originally implanted device or transplanted or explanted for recovery. If explanted for
recovery patients must have survived 60 days post-explant to be considered successful.
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
Effectiveness was measured by the primary endpoint. The proportion of study patients alive,
transplanted, or explanted for recovery at 180 days was compared to the same proportion
obtained from the INTERMACS® registry cohort and tested for non-inferiority.
Secondary Objectives including Safety
Secondary endpoints included: overall survival; incidence of all serious adverse events, including
neurocognitive status and unanticipated adverse device effects; incidence of all device
failures and device malfunctions; Quality of Life improvement, as measured by the Kansas City
Cardiomyopathy Questionnaire (KCCQ) and European Quality of Life Assessment (EuroQol)
EQ-5D; and functional status improvement, as measured by New York Heart Association (NYHA)
classication and 6-minute walk.
Safety measures included the frequency and rates of adverse events, overall and for each
specic event, which were collected throughout HVAD System support.
Study Population Demographics and Baseline Parameters
There were three analysis populations dened for this trial. These are the intent-to-treat
population, (ITT), the Safety population (SAF) and the Per Protocol population (PP).
Intent to Treat Population
N=140
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
Safety Population
N=140
Per Protocol Population
N=137
Major Protocol Violations
N=3
10
Appendix
11Introduction
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Accessories
1.7 Pivotal US Clinical Study: Bridge-to-Transplant (continued)
Subjects were predominately male (72.1%) and 53.3 ± 10.3 years of age. BSA and BMI were
2.1 ± 0.3 m2 and 28.6 ± 6.1 kg/m2 respectively. The principal etiology of heart failure was
ischemic heart disease (41%) and the average LVEF was 17.8 ± 7.1 %. Pulmonary Capillary
Wedge Pressure (PCWP) was elevated at 23 ± 9 mm Hg and pulmonary artery pressures were
also high: (49 ± 15)/(25 ± 9) mmHg. The majority of patients were classied as NYHA IV (95%).
Laboratory values at baseline were, in general, unremarkable except for an elevated BUN
(26 ± 14 mg/dL) and a depressed hematocrit (34 ± 5.8 %).
Eighty percent of subjects in the HVAD System treatment group were on inotropic therapy at
baseline. Some (23%) were on more than one inotrope. IABP therapy at baseline was reported
for 25% of subjects and 85% presented with an AICD. Subjects received typical medications for
congestive heart failure with diuretics (82%) most common.
Comparison of Selected Baseline Characteristics between Treatment and Control Groups
The mean age of implant recipients in the HVAD System group was 53.3 (range 22-70) and for
the control, 52.2. Other parameters available to compare included gender, BSA, BUN, right atrial
pressure and creatinine. In all cases, the values for both the HVAD treatment and control groups
were not statistically signicantly different (Table 1).
Monitor
6
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
9
Table 1: Select Baseline Characteristics for HVAD and INTERMACS® Groups
Characteristics
Age (years)53.3 ± 10.352.2 ± 12.20.19
Female Gender, n (%)39 (28%)120 (24%)0.36
Explant
BSA (m2)2.06 ± 0.282.07 ± 0.300.59
BUN (mg/deciliter)25.3 ± 13.528.9 ± 20.90.94
Right atrial pressure
(mmHg)
Serum creatinine
(mg/dL)
HeartWare™
HVAD™ System
N=140
10.8 ± 3.311.5 ± 5.00.53
1.3 ± 0.41.4 ± 0.60.89
INTERMACS
N=499
p-value
Effectiveness Results
Primary Endpoint
The analysis of the primary endpoint demonstrated HVAD non-inferiority to the control group
(Table 2). The difference in success rates between the HVAD group and controls was less than
the 15% non-inferiority margin (p <0.0001). The 95% one-sided UCL on the difference in success
rates was 4.5% for the Safety (SAF) population analysis and 0.9% for the Per Protocol (PP)
population analysis. The pre-specied primary endpoint was achieved.
Reference
10
Guides
Appendix
12HVAD™ Instructions for Use
1.7 Pivotal US Clinical Study: Bridge-to-Transplant (continued)
1
Table 2: Success Rates and Inference on Non-Inferiority
Implanted
(N)
Safety Cohort
HVAD™ 140127 (90.7)
Controls497448 (90.1)
Per Protocol Cohort
HVAD™ 137126 (92.0)
Controls497448 (90.1)
P-value: From signicance test of non-inferiority
UCL: 95% one-sided upper condence limit on the difference in success rates
Note: The table accounts for 497 of the 499 INTERMACS® patients; the remaining 2 patients, who withdrew
consent before 180 days, have a missing success/failure outcome.
Competing Outcomes
A competing risks analysis was performed (Figure 2), estimating the time-related probability
of experiencing each of the component events. These data are calculated from all events
occurring during the study duration, including deaths, transplants and exchanges occurring
after 180 days but ending with last-patient, last-visit.
There were eight subject deaths during the 180-day study period. Six deaths occurred in subjects
with their originally implanted device and two deaths occurred after device exchange.
Safety Results
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
This study was not randomized and used a contemporaneous control for the sole purpose of
comparing a pre-dened success outcome. The adverse events reported here are unique to the
HVAD System and have no randomized comparator arm.
13Introduction
10
Appendix
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Accessories
1.7 Pivotal US Clinical Study: Bridge-to-Transplant (continued)
Exposure
The total support (exposure) on the original HVAD System was 20,698 days or 56.7 patient-years.
The mean duration on device for the 140 subjects was 147.8 days (standard deviation 52.8)
with a median 180 (range 6 – 180 days). The mean duration on study was 222.5 days (standard
deviation 119) with a median of 196 (range 11 – 588 days). Duration on study exceeds duration
on device, because the follow-up post-transplant is included.
Adverse Events
A total of 776 events (Table 3) were reported by investigators during the 180 day period on the
original device. Of these 437 (437/776, 56.3%) were INTERMACS® dened specic events, and
338/776 (43.6%) events were recorded under the INTERMACS® category of “Other.” One UADE
was reported during the 180-day primary endpoint period.
Table 3: Summary of All Investigator-Reported Adverse Events
EventTotal%
INTERMACS® dened Events43756.3%
INTERMACS® “Other” AE’s33843.6%
UADE10.1%
Total776100%
Monitor
6
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
9
INTERMACS® Events
The INTERMACS® dened adverse events for the 180-day primary endpoint on original device are
summarized below and are separated into the perioperative (0-30 days) and post-perioperative
(31-180 days) periods. Events meeting INTERMACS® criteria are shown in Table 4. Bleeding,
Explant
infections and arrhythmia were the most common. Most bleeding events qualied due to
transfusions (see denition below). On the other hand, all reoperations due to bleeding were
in the rst 30-days post-op (23 vs. 0 events post-30 days).
Reference
10
Guides
Appendix
14HVAD™ Instructions for Use
1.7 Pivotal US Clinical Study: Bridge-to-Transplant (continued)
1
Table 4: INTERMACS® Events by Type and Time of Onset (HVAD System N=140)
Day of Event Onset
INTERMACS® dened AEs
Bleeding
Re Op
Transfusion Criteria2
>4 Units within 7 Days
Any Units at > 7 Days3125 (17.9)4620 (14.3)
Infections
Local (Non-device)2020 (14.3)1717 (12.1)
Driveline Exit55 (3.6)1411 (7.9)
Sepsis33 (2.1)87 (5.0)
Neurological Events
Ischemic CVA77 (5.0)33 (2.1)
Hemorrhagic CVA22 (1.4)22 (1.4)
TIA22 (1.4)54 (2.9)
Respiratory Dysfunction2622 (15.7)85 (3.6)
Arrhythmia
Ventricular1514 (10.0)1411 (7.9)
Supraventricular2521 (15.0)76 (4.3)
Right Heart Failure
Inotropes 1717 (12.1)87 (5.0)
RVAD33 (2.1)11 (0.7)
Arterial Thromboembolism0022 (1.4)
Venous Thromboembolism44( 2.9)33 (2.1)
Renal Dysfunction88 (5.7)65(3.6)
Psychiatric Event55 (3.6)44 (2.9)
Myocardial Infarction Event0011 (0.7)
Hypertension11 (0.7)00
Hepatic Dysfunction33 (2.1)11 (0.7)
Hemolysis Event
1
3
0-30 Days31-180 Days
Events
N
2320 (14.3)00
1010 (7.1)00
11 (0.7)11 (0.7)
Subjects
N (%)
Events
N
Subjects
N (%)
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
1
4 procedures were not included: elective hysterectomy, elective repair of hemorrhoids, HVAD™ exchange
and RVAD placement.
2
Transfusion criteria include: ≥ 20cc/kg packed red blood cells (PRBC) within any 24 hour period during the
rst 7 day post implant and any transfusion of packed red blood cells (PRBC) after 7 days following implant
with the Investigator recording the number of units given.
3
Two cases were excluded: 1 case hemolysis < 72 hours post-implant; 1 case hemolysis occurring in the
presence of tPA/Integrillin for VAD thrombosis.
9
Reference
Guides
10
Appendix
15Introduction
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Monitor
6
Surgical
Implant and
Accessories
Explant
1.7 Pivotal US Clinical Study: Bridge-to-Transplant (continued)
The majority of infections did not involve the driveline or cause sepsis. The local, non-device
category encompasses a host of sites, including the urinary tract, lungs, sinuses, IV punctures,
colon and skin. Infections involving the driveline exit site were more common after hospital
discharge (> 30 days). Similarly, subjects were somewhat more likely to experience sepsis
from 31-180 days (5.0% of subjects) than perioperatively (2.1%). Nearly a third (11/32) of the
supraventricular arrhythmias were bouts of atrial brillation, requiring drug therapy. Nearly all the
ventricular arrhythmias were ventricular tachycardia. AICD shocks were recorded in 24/29
episodes of ventricular arrhythmia and 2/29 received external cardioversion. Nearly all patients
with a reported episode of ventricular tachycardia were subsequently placed on amiodarone.
Respiratory problems were more common in the perioperative period, declining from 26/34
events at 0-30 days to about one-third that number (8/34) from 31-180 days. Subjects were more
likely to experience right heart failure events in the perioperative period (20/29). The most
common treatment for right heart failure was the use of inotropic drugs and the pulmonary
vascular dilator, nitric oxide (25/29). Three subjects required an RVAD and a fourth was
exchanged for a pneumatic biVAD at 75 days post-implant. Ischemic strokes (ICVA) were more
common overall (10/14 events) and occurred with greater frequency in the perioperative period
(7/9 perioperative strokes). Four hemorrhagic strokes (HCVA) were recorded. Three of these
resulted in deaths. TIAs were more common in the 31-180 day period (5/7 TIA events). While
HCVAs were generally fatal (75%) they were most often associated with hypertension (MAP >
90 mm Hg). Three of the 4 HCVAs had a mean arterial pressure of ≥ 95 mm Hg at the time of the
stroke and the one normotensive patient was septic and had an INR of 2.7 (high normal range).
Overall 70% of the patients who experienced ICVAs were transplanted or remained eligible. It
is noteworthy that 6/10 ICVA events occurred within 48 hours of implant and may have been
related to surgical procedural factors, such as ragged coring of the myocardium for inow
insertion or incomplete device de-airing. These issues were addressed by improvements to the
coring tool and by site retraining. The overall stroke survival for the combined ICVAs and HCVAs
on the original device was 77% (10/13 patients).
7
Patient
8
Alarms and
9
Reference
10
Appendix
Management
Emergencies
Guides
Venous thrombosis occurred in 5% of subjects. Most of these were cases of DVT in the lower
extremities. In the arterial thromboembolism category, a case of VAD thrombosis was treated
with tPA and resolved and in another case a clot was removed from the left main coronary
artery following cardiac catheterization. A third case appeared to involve a shower of small
emboli to the periphery.
No subject required permanent dialysis. Psychiatric events were recorded for nine subjects
(6.4%). All recovered without sequelae. Two hemolysis events were detected by strict
INTERMACS® criteria in the absence of VAD thrombosis. These resolved spontaneously.
One subject experienced a myocardial infarction and one subject had a hypertensive event
during the perioperative period. Hepatic dysfunction was noted in four subjects.
Adverse events were generally more common in the perioperative period.
Serious Adverse Events
A total of 452 serious adverse events on the original device occurred in 118 (84.3%) subjects
(Table 5). A total of 287 INTERMACS® dened events met the denition of an SAE, and 164
INTERMACS® “other” events met the denition of an SAE.
16HVAD™ Instructions for Use
1.7 Pivotal US Clinical Study: Bridge-to-Transplant (continued)
1
Table 5: Summary of Serious Adverse Events (HVAD System N=140)
Serious Adverse Events (SAEs)Number of SAEs
Total Serious Adverse Events452118 (84.3)
INTERMACS28798 (70.0)
“Other”16475 (53.6)
UADE11 (0.7)
Device Exchange
Device exchange occurred in 7 patients (7/140, 5.0%) in the SAF population during the period
180 days post-implant. Of these 7 exchanges, 3 were resultant from retained tissue being pulled
into the pump from the ventricle in the very early post-operative period and were deemed to be
procedure related, 2 were exchanged due to thrombus inside the pump, one was exchanged
for a high power event of unknown cause and one due to latent right heart failure which
caused the patient to require a biventricular support system.
Device Malfunctions
A device malfunction is dened as a failure of one or more of the components of the HVAD
System, which either directly causes or could potentially, cause or induce a state of inadequate
circulatory support (low cardiac output state) or death. There was information on 26 malfunctions
from 20 subjects entered into the clinical database during the study period (Table 6).
Subjects
N (%)
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
Table 6: Malfunctions by Suspected Component
HVAD System N=140
Device Component ID
Pump7 (5.0) *
Controller7 (5.0)
Battery1 (0.7)
Battery Charger0
Monitor0
Driveline2 (1.4)
Controller AC Adapter6 (4.3)
Other Component3 (2.1)
*Described in Pump Exchange section
Events
N (%)
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
17Introduction
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Accessories
1.7 Pivotal US Clinical Study: Bridge-to-Transplant (continued)
Quality of Life: KCCQ And EuroQol
Kansas City Cardiomyopathy Questionnaire (KCCQ): At baseline, 128/140 (91.4%) patients were
able to complete the KCCQ and at month 6 there were 88 patients available to complete
the test (39 had received a transplant, six had died, seven had met an endpoint receiving a
device exchange) (Table 7), Of the 88 patients available for assessment, 74 patients had data at
month 6, Reasons for missing the month 6 data included: 9 of 14 with poor compliance/missed
visit (8 of 9 of these from a single site and 1 of 9 had a prior ICVA with mRS score of 2), 2 were too
sick, 1 had no form available, 1 had been transplanted within the 14 day visit window, and 1 had
refused. Seventy patients (70) had both baseline and month 6 data. For these 70 patients who
were on HVAD™ therapy continuously for 180 days had a 31 point improvement in KCCQ Overall
Summary Score, over the 180 day period.
Table 7: KCCQ - Overall Summary Score
KCCQBaselineMonth 6Change from Baseline
N1287470
Mean (SD)34.9 (18.9)67.5 (20.4)30.9 (26.5)
Median31.571.434.5
Min, Max0.0, 84.119.3, 100.0-49.4, 80.5
95% CI31.6, 38.262.8, 72.224.6, 37.3
6
Surgical
7
Patient
8
Alarms and
9
Reference
10
Monitor
Implant and
Explant
Management
Emergencies
Guides
European Quality of Life (EuroQol): At baseline, 130/140 (92.9%) of patients were able to
complete the test, and at month 6 there were 88 patients available to complete the test,
(39 had received a transplant, six had died, seven had met an endpoint receiving a device
exchange) (Table 8). Of the 88 patients available 75 had data at month 6. Reasons for
missing the month 6 data included: 9 of 13 with poor compliance/missed visit (8 of 9 of these
from a single site and 1 of 9 had a prior ICVA with mRS score of 2), 2 were too sick, 1 had been
transplanted within the 14 day visit window, and 1 had refused. Seventy-two patients (72) had
both baseline and month 6 data showing an improvement of 30 points over the 180 day period.
Table 8: EuroQol (EQ-5D) - Summary of Quality of Life
EuroQolBaselineMonth 6Change from Baseline
Overall Summary Score
N1307572
Mean (SD)39.7 (23.5)69.8 (19.8)29.5 (25.2)
Median40.075.030.0
Min, Max0.0, 92.04.0, 100.0-36.0, 80.0
95% CI35.6, 43.765.2, 74.423.6, 35.4
Functional Analyses: 6 Minute Walk
6 Minute Walk: Of the 132 patients assessed for the 6-minute walk test, the mean distance
walked was 89.4 meters. Seventy-Five (75) of the 88 patients on pump at month 6 completed
the test (Table 9 and Figure 3). Reasons for missing the 6 minute walk test at month 6 included: 9
of 14 with poor compliance/missed visit (8 of 9 of these from a single site and 1 of 9 had a prior
ICVA with mRS score of 2), 2 were too sick, 1 had no form available, 1 had been transplanted
within the 14 day visit window, and 1 had refused. These 75 patients showed a mean distance
walked of 246 meters, a mean change of 150 meters from baseline.
Appendix
18HVAD™ Instructions for Use
1.7 Pivotal US Clinical Study: Bridge-to-Transplant (continued)
1
Table 9: Functional Status – 6 Minute Walk
6 Minute WalkBaselineMonth 6Change from Baseline
Distance Walked in Meters
N1327574
Mean (SD)89.4 (141.3)246.0 (203.9)150.1 (214.1)
Median0.0274.0108.3
Min, Max0.0, 600.20.0, 991.8-273.1, 700.9
95% CI65.1, 113.7199.1, 292.9100.5, 199.8
Figure 3:
6 Minute Walk Test
Table 10 shows a breakdown of results of patients who walked at both baseline and at 6 months
as well as those patients that did not walk at baseline but did walk at 6 months.
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
Table 10: 6 Minute Walk – Breakdown of Patients Walking vs. Not Walking at Baseline
HVAD System PatientsBaseline (m)Month 6 (m)
Patients walking at baseline and at
6 months
Patients NOT walking at baseline (for
any reason) but walking at 6 months
Overall Conclusions from Clinical Data
The HVAD System bridge-to-transplant study (ADVANCE) was a multi-center, prospective,
contemporaneous control trial. The purpose of this study was to evaluate the safety and
effectiveness in patients listed for cardiac transplantation with refractory, advanced heart
failure at risk of death. The primary endpoint was success at 180 days which is dened as alive
on the originally implanted HVAD Pump or transplanted or explanted for recovery.
The analysis of the primary endpoint yielded non-inferiority of the HVAD System to the
INTERMACS® control. The 95% one-sided UCL on the difference in success rates was 4.5% for
the Safety Group and 0.9% for the Per Protocol Group. Each of these limits was less than the
15% non-inferiority margin (p-value <0.0001).
• The pre-specied primary endpoint was achieved.
• Both quality of life and functional capacity showed improvements following implant of the
HVAD Pump.
260 ± 140 (n=25)338 ± 202 (n=25)
N/A333 ± 125 (n=30)
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
• The HVAD System has an adverse event prole that supports its safe use for bridge to transplant
patients.
19Introduction
10
Appendix
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Monitor
6
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
Accessories
Explant
1.8 Post Approval Study Follow-up of the Pivotal US Clinical Study:
Bridge-to-Transplant
Summary of the Post-Approval Study Methods
Study Objective
The purpose of the HW-PAS-03 follow-up study was to continue the evaluation of the longer-term
safety and effectiveness of the HeartWare Ventricular Assist System through 5 years in patients
who were enrolled in the ADVANCE pivotal study presented above in Section 1.7.
Study Design
This was an observational, prospective study, conducted at multiple study sites with no new
enrollment. As, no new patients were screened or implanted with the HeartWare VAS for this
study; it is a continued follow-up study only. HW-PAS-03 included, patients who were enrolled
from the ADVANCE Trial, and a continuation of that study, known as the continuous access
protocol (CAP).
Study Population and Data Source
Patient population
The original ADVANCE cohort implanted 140 bridge-to-transplant (BTT) patients, and the
Continued Access from the ADVANCE trial (CAP) implanted 242 additional patients using the
same inclusion criteria.
Patients who participated in the prior (BTT and CAP) were approached for participation in this
continued follow-up PAS if eligible according to the HW-PAS-03 protocol version 3.0 04Sep2013.
BTT and CAP patients eligible for participation in HW-PAS-03 were:
• Patients who were alive at the start of enrollment for the PAS who either
• were on continued HeartWare VAS support (original or exchange device), or
• had been explanted for transplant or recovery and had not yet completed six months of
follow-up.
At the time of enrollment, all surviving patients from the BTT cohort had been followed for at least
37.7 months, and all surviving patients from the CAP cohort had been followed for at least 4.4
months.
A total of 152 subjects (39.8% of the original combined BTT and CAP cohorts) survived to the
Premarket approval of the HeartWare VAS, were still enrolled in the original BTT or CAP trials and
were therefore eligible for enrollment into HW-PAS-03.
The PAS results below include three main cohorts:
• On Device: Patients in this cohort (N=84) were still on the HeartWare HVAS device upon
enrollment into the HW-PAS-03 study.
• All Enrolled: This cohort includes all 101 patients who enrolled into the HW-PAS-03 study. It
includes On Device patients (N=84), as well as Off Device patients.
• Off Device patients (N=17) were enrolled into HW-PAS-03 less than six months posttransplant or explant for recovery (no device in the body upon enrollment). These
patients only participated in the study through completion of their 6 months’ follow-up.
This PAS does not include 51 patients (33.6% of eligible subjects) from the total BTT+CAP
population who were eligible but declined. See Table 11 below for specic reasons.
10
Appendix
20HVAD™ Instructions for Use
1.8 Post Approval Study Follow-up of the Pivotal US Clinical Study:
Bridge-to-Transplant (continued)
Table 11: Reason for Not Participating in HW-PAS-03
1
2
Reason
Patients whose eligibility expired: they completed the required 6-month
post-explant visit between approval and enrollment
Patient declined participation12
Patient died between approval and enrollment8
Site declined participation7
Patient is lost to follow-up3
Patient transferred to another non HW-PAS-03 site/moved to another city2
Patient’s condition did not allow enrollment per PI1
Enrollment visit could not be performed within the enrollment period1
Total51
Key Study Endpoints
Endpoints for this study were observational only. The endpoints assessed included:
• Overall survival on device
• Final patient status
• Re-hospitalizations
• INTERMACS® adverse events
• Quality of Life measures
• Functional Status
Safety measures included the frequency of adverse events, which were collected throughout
HeartWare Ventricular Assist System support.
Total Number of
Patients
17
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Explant
Total number of Enrolled Study Sites and Subjects, Follow-up Rate
Summary of Study Progress
HW-PAS-03 protocol approval was received on November 20, 2012.
Enrollment into the study started on January 23, 2013 and was completed on May 23, 2013. A
total of 101 BTT and CAP subjects from 25 sites were enrolled. The study collected its nal data
and was closed on December 20, 2017.
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
21Introduction
1
382 BTT+CAP
subjects implanted
27* subjects
explanted
95 subjects
sll on device
157 subjects
transplanted
103 subjects
died on device
152 BTT+CAP
subjects eligible
51 subjects
not included
101 subjects
enrolled
84 subjects
sll on device
17 subjects
< 6 months
post-transplant
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
1.8 Post Approval Study Follow-up of the Pivotal US Clinical Study:
Bridge-to-Transplant (continued)
Figure 4. Flowchart for Enrollment into HW-PAS-03 Trial
Accessories
*2 subjects had withdrawn consent.
Monitor
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
9
Number of Eligible Sites
All 30 sites that participated in BTT and CAP were eligible for participation in HW-PAS-03 . Twenty-
ve of those sites enrolled at least one eligible subject and participated in HW-PAS-03.
The remaining ve centers did not participate in HW-PAS-03. The most frequent reason for not
participating was lack of center resources.
Study visits and length of follow-up
Assessments were conducted at enrollment into HW-PAS-03 and at visits according to the
following schedule:
• Subjects enrolled on device (either original device or HeartWare device exchange):
every six months until outcome or ve years post-initial implant of the original device.
• Subjects who were enrolled after being explanted for transplant or recovery: until six
months post-explant to record subject status only, at which point participation in the
study was considered complete.
Summary of the Post Approval Study Results
A total of 152 subjects were eligible for participation in the HW-PAS-03 study. Of those, 101 (66%)
were enrolled; 84 subjects enrolled while still on the HeartWare device, and 17 enrolled posttransplant. Subjects in the All Enrolled cohort (N=101) had rates of complete study visit follow-up
between 90.9% - 100%.
Reference
10
Guides
Appendix
22HVAD™ Instructions for Use
1.8 Post Approval Study Follow-up of the Pivotal US Clinical Study:
Bridge-to-Transplant (continued)
Table 12: Subject Baseline Demographics
1
2
Demographics
Age at enrollment into PAS03 (years)
Gender
Male73.3% (74/101)69.0% (58/84)94.1% (16/17)
Female26.7% (27/101)31.0% (26/84)5.9% (1/17)
Ethnicity
Hispanic or Latino5.9% (6/101)6.0% (5/84)5.9% (1/17)
Non-Hispanic or Non-Latino94.1% (95/101)94.0% (79/84)94.1% (16/17)
American Indian/Native Alaskan1.0% (1/101)0.0% (0/84)5.9% (1/17)
Other3.0% (3/101)3.6% (3/84)0.0% (0/17)
Note: Data is from the original ADVANCE and CAP trials for gender, ethnicity and race. Age is as of consent
into the HW-PAS-03 study.
For the All Enrolled cohort, 67 subjects (66%) were alive at the time of completion/exit; 26 (26%)
were still implanted with an HVAD (21 on original and 5 post-exchange). 41 (41%) subjects were
alive after being explanted for transplant. 34 (34%) subjects had died. On Device subjects spent
42.7 months on average implanted with the device through completion of study follow up (Table
14).
All Enrolled
(N=101)
54.4 ± 12.62 (101)
(22.0, 57.0, 74.0)
On Device
(N=84)
54.8 ± 12.35 (84)
(24.0, 56.0, 74.0)
Off Device
(N=17)
52.3 ± 14.08 (17)
(22.0, 58.0, 67.0)
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Explant
Figure 5: Subject Status
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
23Introduction
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Monitor
6
Surgical
Implant and
7
Accessories
Explant
1.8 Post Approval Study Follow-up of the Pivotal US Clinical Study:
Bridge-to-Transplant (continued)
Table 13: Subject Disposition (all enrolled through 5 years)
Disposition
Alive
On Original Device21 (20.8%)
Post-Exchange5 (5.0%)
Post-Explant for Transplant
Post-Explant for Recovery
Dead
On Original Device
Post-Exchange
Post-Explant for Transplant
Post-Explant for Recovery
Table 14: Summary of Duration on Device and In Study (all enrolled)
Parameter
Duration on Original
Devicea (months)
Duration on Device
(months)
Duration in Study
(months)
Note: Numbers are mean ± SD (min, median, max).a Duration on Original Device (months) = date of rst explant/transplant/exchange or last follow up – date of original implant + 1b Duration on Device (months) = date of last explant/transplant or last follow up – date of original implant + 1c Duration on Study (months) = date the subject exited from the study
– date of original implant +1
b
c
a
HW-PAS-03
(N=101)
33.6 ± 19.0
(0.1, 31.6, 62.1)
38.1 ± 18.3
(2.5, 41.5, 62.4)
40.5 ± 17.1
(7.8, 44.4, 65.4)
HW-PAS-03
(N=101)
67 (66.3%)14 (63.6%)53 (67.1%)
41 (40.6%)6 (27.3%)35 (44.3%)
0 (0.0%)0 (0.0%)0 (0.0%)
34 (33.7%)8 (36.4%)26 (32.9%)
22 (21.8%)4 (18.2%)18 (22.8%)
6 (5.9%)3 (13.6%)3 (3.8%)
6 (5.9%)1 (4.5%)5 (6.3%)
0 (0.0%)0 (0.0%)0 (0.0%)
BTT Cohort
(N=22)
39.9 ± 20.0
(0.2, 45.2, 61.5)
53.5 ± 6.9
(41.7, 56.9, 62.4)
55.2 ± 6.6
(43.4, 57.1, 63.4)
BTT Cohort
(N=22)
5 (22.7%)16 (20.3%)
3 (13.6%)2 (2.5%)
CAP Cohort
(N=79)
CAP Cohort
(N=79)
31.8 ± 18.4
(0.1, 30.3, 62.1)
33.8 ± 18.2
(2.5, 31.6, 62.1)
36.4 ± 16.8
(7.8, 33.6, 62.1)
Patient
8
Alarms and
9
Reference
10
Appendix
The Kaplan-Meier survival estimates at 5 years for all implanted BTT and CAP subjects (N=382)
was 37.1%.
Management
Emergencies
Guides
24HVAD™ Instructions for Use
1.8 Post Approval Study Follow-up of the Pivotal US Clinical Study:
Bridge-to-Transplant (continued)
Figure 6: Kaplan-Meier Survival on device
Note: Of the 51 subjects who did not rollover into the HW-PAS-03 trial from Table 11, 26 subjects were still eligible to
participate in the post market study at the time exit from the original trials (did not die or complete 6 months posttransplant).Their follow up was censored at the time of last follow up from the pre-market trials.
There was no statistically signicant difference in survival between the BTT and CAP Cohorts,
between males and females, or between white and non-white patients when analyzing the All
Enrolled (N=101) cohort, which only includes subjects who enrolled into the HW-PAS-03 trial. The
two most common causes of death were device malfunction (seven subjects) and neurological
dysfunction (four subjects).
Survival on device from time of consent into HW-PAS-03 for the On Device subjects (N=84)
is presented for the subject who were from the BTT cohort (N=21) and CAP cohort (N=63)
separately, as all subjects were enrolled into HW-PAS-03 after implant and given the difference
between implant times prior to enrollment. Time 0 was the date of consent for the HW-PAS-03
trial and subjects were censored at the earlier time of their last follow up or the end of LVAD
support.
Figure 7: Kaplan-Meier Survival on device from enrollment into HW-PAS-03, BTT cohort
1
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
25Introduction
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
1.8 Post Approval Study Follow-up of the Pivotal US Clinical Study:
Bridge-to-Transplant (continued)
Figure 8: Kaplan-Meier Survival on device from enrollment into HW-PAS-03, CAP cohort
Accessories
Monitor
6
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
Quality of life and functional status assessments demonstrated sustained improvements over
time. The overall summary score for KCCQ had an average improvement of at least 20 points
from baseline at all follow up visits, and the average EQ-5D-5L Visual Analog Scale was greater
than 65 at all visits. The 6-minute walk test showed an average increase of at least 90 meters
from baseline at all follow-up timepoints. At most timepoints for NYHA, over 80% of the subjects
who completed the assessment were improved to a NYHA classes I or II.
For the On Device subjects in this PAS (n=84), the three most common adverse events were
infection, device malfunction/failure and bleeding.
Explant
9
Reference
10
Guides
Appendix
26HVAD™ Instructions for Use
1.8 Post Approval Study Follow-up of the Pivotal US Clinical Study:
Bridge-to-Transplant (continued)
Table 15: INTERMACS® adverse events while on a HeartWare device during PAS03* (On Device
Subjects)
Right Heart Failure8 (9.5%)90.075 (23.8%)60.283 (4.8%)30.03
Inotropic
Therapy
RVAD1 (1.2%)10.011 (4.8%)10.05000
Inhaled Nitric
Oxide
Arterial Non-CNS
Thromboembolism
Venous
Thromboembolism
Wound
Dehiscence
Other
Note: Percentages are based on the number of subjects in the group. Subjects are counted once within each
INTERMACS® dened adverse event term. Summarized AEs include emergent AEs, and AEs that occurred while on any
HeartWare device (including pre and post-exchange AEs).
*Adverse Events that occurred and were not ongoing in the Premarket duration of the BTT and CAP studies are not
included in this table.
Subjects
with
Event
(%)
14
(16.7%)
7 (8.3%)80.064 (19.0%)50.243 (4.8%)30.03
000000000
1 (1.2%)10.010001 (1.6%)10.01
3 (3.6%)30.020003 (4.8%)30.03
000000000
61
(72.6%)
No. of
Events
150.114 (19.0%)40.19
2241.63
Event
Rate
per PY
(137.22)
Subjects
Event (%)
(76.2%)
with
16
BTT Cohort
(N=21)
No. of
Events
452.12
Event
Rate
per PY
(21.22)
Subjects
with
Event
(%)
10
(15.9%)
45
(71.4%)
CAP Cohort
(N=63)
No. of
Events
110.09
1791.54
Event
Rate
per PY
(115.99)
7
Patient
8
Alarms and
9
Reference
10
Appendix
Management
Emergencies
Guides
Neurological dysfunction in subjects on a HeartWare device included 16 subjects (19.0%)
who had 23 events. Of those, 6 subjects (7.1%) had 7 CT-conrmed ischemic cerebrovascular
event’s, 9 subjects (10.7%) had 12 CT-conrmed hemorrhagic cerebrovascular event’s, and 4
subjects (4.8%) had 4 TIAs. The proportion of On Device subjects who experienced neurological
dysfunction adverse events was higher in the BTT cohort (23.84%) than the CAP cohort (17.5%).
A total of 89 infection events occurred in 43 subjects (51.2%) while on a HeartWare device. Of
those, driveline infections occurred in 27 subjects (32.1%) who had 35 events and 9 subjects
(10.7%) experienced 12 sepsis events.
A total of 49 device malfunctions/failure events occurred in 35 subjects (42.9%) while on a
HeartWare device during HW-PAS-03 follow up. The most frequent events were related to the
pump, including outow graft and inow cannula issues (18 events, 23.1%) and suspected/
conrmed pump thrombus (9 events, 11.5%). The second most common event type was
controller faults and damage (17 events, 21.8%). Less frequent events included power
disconnection, connector issues, electrical faults, and battery issues.
Eight subjects (9.5%) had nine exchanges, with one subject having two exchanges, during HWPAS-03 follow up.
28HVAD™ Instructions for Use
1.8 Post Approval Study Follow-up of the Pivotal US Clinical Study:
Bridge-to-Transplant (continued)
Most subjects had at least one re-hospitalization, with more than half having three or more
re-hospitalizations. The mean cumulative LOS was 45.8 days. The two most common primary
reasons for re-hospitalization were adverse event (23 subjects, 77.4%) and explant (8 subjects,
27.4%).
1
2
Final safety ndings (key endpoints)
This post approval study followed the long-term safety and effectiveness of IDE trial subjects up
to ve years post original implant. All subjects had at least one re-hospitalization during the HW-
PAS-03 study. The longer term follow up for these subjects (more than three years on average)
were associated with, with infections, device malfunctions/failures and bleeding events as the
most common type of adverse event.
Final effectiveness ndings (key endpoints)
Of the 101 enrolled subjects, about two-thirds (67 subjects, 66.3%), were still alive at the time
of their study exit or HW-PAS-03 study completion/exit. Of those, 41 subjects were alive posttransplant (40.6% of the enrolled subjects) and 26 were alive on support (25.7% of those
enrolled). Fewer than 10% of subjects had an exchange during the HW-PAS-03 trial (8 subjects,
9.5%), 27.7% died while on the device (28 subjects), and 5.9% (6 subjects) died less than 6 months
post-transplant. Quality of life and functional status measurements improved over time.
Study Strengths and Limitations
There were several strengths to this study. It provided continued follow-up for patients who had
received an HVAD System, allowing observation of long-term outcomes in an initial bridge to
transplant approach. Additionally, the nal data demonstrated consistent results regarding
adverse event rate outcomes. A limitation of the study was that only subjects who were still alive
on a HeartWare device or post-transplant for less than 6 months were eligible for the HW-PAS-03
trial. Subjects also had varying follow up times prior to enrollment into HW-PAS-03. Additionally,
of the 152 eligible subjects, 33% (51/152) of those who potentially could have enrolled into HWPAS-03 did not (e.g., site declined participation, subject declined participation, lost to follow up,
etc.). These factors limit the interpretability of longer-term survival and adverse event results, as
the potential for selection bias and the inuence of competing risks must be considered.
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
29Introduction
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Monitor
6
Accessories
1.9 US Clinical Study: Destination Therapy
A. Study Design
Patients in the ENDURANCE trial were enrolled between August 4, 2010 and May 8, 2012. The
database for this Panel Track Supplement reected data collected through June 06, 2016, as
well as some additional updated data from March 27, 2017, and included 451 subjects enrolled
at 48 investigational sites.
The trial was a prospective, randomized, controlled, multicenter clinical trial. Subjects were
randomly assigned using a permuted block, central randomization scheme, in a 2:1 ratio, to
receive either the study (HVAD) or control (HeartMate II) device.
The objective of the trial was to compare the safety and effectiveness of HVAD for destination
therapy to the HeartMate II, which is legally marketed in the U.S. for destination therapy, in
patients with end-stage heart failure who are ineligible for heart transplantation.
The sample size for formal hypothesis testing was to be determined adaptively. Subjects were to
be randomized until 450 subjects were randomized and implanted.
It was pre-specied that after the rst 300 randomized subjects reached the two-year primary
endpoint, the success rate from the control subjects would be assessed. If the observed control
success rate was at least 55%, then the data from the rst 300 subjects would be analyzed. If the
observed control success was less than 55%, then no interim analysis would be performed and
the full 450 subjects would be subsequently analyzed. This adaptive sample size for statistical
analysis provides at least 90% power to establish non-inferiority.
The ENDURANCE trial was conducted under the oversight of an independent Clinical Events
Committee, which adjudicated all the adverse events according to the Interagency Registry
of Mechanically Assisted Circulatory Support (INTERMACS®) denitions; and an independent
Data Safety Monitoring Board reviewed study compliance and monitored adverse events and
outcomes.
Surgical
7
Patient
8
Alarms and
9
Reference
10
Implant and
Explant
Management
Emergencies
Guides
1. Clinical Inclusion and Exclusion Criteria
Enrollment in the ENDURANCE trial was limited to patients who met the following inclusion
criteria:
- Patients ≥18 years old with chronic, advanced left ventricular failure with New York Heart
Association (NYHA) functional class IIIB or IV limitations despite optimal medical therapy
and were transplant ineligible at the time of enrollment in whom informed consent was
obtained.
Patients were not permitted to enroll in the ENDURANCE trial if they met any of the following
exclusion criteria:
- Patients eligible for cardiac transplant or with prior cardiac transplant.
- Patients with recent (within 14 days) acute myocardial infarction or stroke within 180
days.
- Patients with a mechanical heart valve.
- Patients with severe right heart failure in whom right ventricular support is anticipated.
- Patients who might be unwilling or unable to comply with the study criteria.
- Additional exclusion criteria available in the Clinical Study Report.
Appendix
30HVAD™ Instructions for Use
1.9 US Clinical Study: Destination Therapy (continued)
1
2. Follow-up Schedule
All patients were scheduled to return for follow-up examinations at 3, 6, 12, 18, and 24
months with a window of ± 7days, and at 30, 36, 42, 48, 54, and 60 months with a window of
± 14 days postoperatively.
Preoperative baseline assessments included demographics, medical history, physical
examination, concurrent medications, laboratory tests, electrocardiogram (ECG), New
York Heart Association (NYHA) classication, The National Institutes of Health (NIH) stroke
scale, neurocognitive exam, quality of life, and functional status. Postoperative assessments
included LVAD parameters, hemodynamics, concurrent medications, laboratory tests,
neurocognitive exam, six-minute walk test, NYHA status, and health status.
3. Clinical Endpoints
The primary endpoint was a composite of two-year survival free of disabling stroke (i.e.,
modied Rankin score ≥ 4 assessed 24 weeks post-event), while alive on the originally
implanted device, electively transplanted or explanted due to left ventricular recovery.
Success in meeting the primary endpoint was tested for non-inferiority of the experimental
group against the control device. The non-inferiority margin of 15% was based on the
observed success rate of the control device at >55%. Estimates of stroke-free survival were
performed for each treatment using Kaplan-Meier non-inferiority log-rank methodology,
comparing study device to control using a one-sided alpha of 0.05; that is, non-inferiority will
be established if the one-sided upper condence limit on the difference in proportions is less
than the non-inferiority margin. Analysis of the primary endpoint was conducted on the Per
Protocol (PP) population.
Patients were considered a success if at 730 days post implantation, the subject was alive,
did not have a stroke of mRS ≥ 4 assessed 24 weeks post-stroke, and remained on the
originally implanted device, unless the device was removed due to heart recovery, or the
subject was electively transplanted. Patients were considered a failure if at 730 days post
implantation, they expired, had a stroke with a modied Rankin score ≥ 4 assessed 24 weeks
post-stroke, or were urgently transplanted or had surgery for LVAD removal or replacement
due to failure of the original device.
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Explant
7
Patient
Management
There were seven (7) secondary endpoints, of which the following three (3) were to be
assessed inferentially to test for superiority in a xed-sequence procedure if non-inferiority
was established for the primary endpoint: incidence of bleeding (per INTERMACS®
denition), incidence of major infections (per INTERMACS® denition), and overall survival
(time to death). In addition, a number of subgroup analyses were pre-specied, including
gender and BSA (<1.5 m2 vs. ≥1.5 m2).
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
31Introduction
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Accessories
1.9 US Clinical Study: Destination Therapy (continued)
B. Accountability of PMA Cohort
Pre-specified Interim Analysis
Per the pre-specied analysis plan, the interim analysis cohort (N=300) was to serve as the
principal analysis cohort if the Control group success rate for the primary endpoint was at least
55%; as shown below, the observed success rate for the Control group was 59%. A total of 451
patients (inclusive of the initial 300 patients) were enrolled, of which 445 were implanted with a
device. This summary presents the ENDURANCE trial results using both the pre-specied interim
analysis and full enrollment cohorts. FDA considered the interim analysis to be the principal
analysis of the ENDURANCE trial, but considered all analyses when evaluating the safety and
effectiveness of the HVAD. The analyses from the full enrollment cohort are included in the Other
Results section.
At the time of database lock for the interim analysis 100% of the pre-specied interim analysis
cohort (300 patients) had been followed through the 2-year primary endpoint time point. The
disposition of the patients is shown in Figure 9.
The Randomized population (HVAD N=200 and Control N=100) included all subjects who were
consented (Intent-to-Treat (ITT)) and then enrolled in the study.
The Anesthetized Population (AP) included all randomized subjects who receive induction of
anesthesia for implantation.
Monitor
6
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
9
The Anesthetized and Implanted Population (AIP) population, equivalent to an As Treated
population, consisted of all randomized subjects who received induction of anesthesia for
implantation and received an implant of an LVAD. In the full cohort (N=445), four (4) patients
crossed over from HVAD to Control and three (3) patients crossed over from Control to HVAD
after randomization but before receiving a device, and one (1) patient in the Control arm
did not receive any device. As such, the AT population for the interim analysis consists of 300
patients, 197 in the HVAD arm and 103 in the Control arm.
Explant
The Per Protocol (PP) population included all subjects in the AIP population analyzed according
to the LVAD to which they were randomized. This denition is more consistent with the ICH
denition of what a modied ITT population would be.
The Inclusion Compliant (IC) population included all randomized subjects who received the
LVAD to which they were randomized and who did not violate certain inclusion and exclusion
criteria that would likely have an effect on outcome.
The primary analysis was performed on the Per Protocol (PP) population. All safety analyses were
performed on the AIP population.
Reference
10
Guides
Appendix
32HVAD™ Instructions for Use
1.9 US Clinical Study: Destination Therapy (continued)
1
Figure 9: Disposition of First 300 Subjects in the ENDURANCE Trial
Randomized Subjects
N = 300
Received HeartWare
Device
N = 197
Did not complete
first 2 Years
N = 85
Death
N = 83
HeartWare N = 200
(96 Non-Sintered,
104 Sintered)
Voluntary
Withdrawal
N = 2
Completed first
2 Years
N = 115
Did not complete
first 2 Years
N = 32
Death
N = 32
Control
N = 100
Voluntary
Withdrawal
N = 0
Received Control
Device
N = 103
Completed first
2 Years
N = 68
C. Study Population Demographics and Baseline Parameters
The demographics and baseline characteristics, as summarized in Table 16, are typical for an
LVAD study performed in the U.S. The HVAD and Control groups did not differ signicantly.
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Table 16: Patient Demographics and Baseline Characteristics in the rst 300 Subjects in the
Explant
ENDURANCE Trial
Demographics and Baseline
Characteristics
Age (years)64.4 ± 12.066.1 ± 10.40.25
Male gender (%)77.5%80.8%0.66
Race (%)
White
Black or African American
Other
Height (cm)173.5 ± 9.8175.2 ± 9.30.15
Body Surface Area (m2)2.0 ± 0.32.0 ± 0.30.98
INTERMACS Prole (%)
1
2
3
4
5-7
Ischemic Etiology of Heart Failure 59.5%59.0%> 0.99
HVAD
(N=200)
79.5%
19.5%
1.0%
3.5%
27.5%
39.5%
21.5%
8.0%
Control
(N=100)
75.0%
25.0%
0.0%
1.0%
38.0%
41.0%
13.0%
7.0%
P-value
0.37
0.17
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
33Introduction
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
1.9 US Clinical Study: Destination Therapy (continued)
D. Safety and Effectiveness Results
1. Primary Endpoint
The pre-specied interim analysis was conducted on the rst 300 patients to reach two (2)
years post implantation. The Kaplan-Meier estimate for stroke-free success at 2 years for the
Control arm was 59.0%; as such, the interim analysis represented the primary analysis for the
ENDURANCE trial. The Kaplan-Meier estimate for stroke-free success at 2 years for the HVAD
arm was 51.1%. The results of the interim analysis are shown in Figure 10. The upper bound
of the condence interval around the difference exceeded the 15% non-inferiority margin
(17.9%), resulting in a p-value of 0.1219. The interim analysis showed that the trial failed to
demonstrate non-inferiority of the HVAD to the Control.
Figure 10: ENDURANCE Trial Primary Endpoint. Survival at 2 years free from disabling stroke (mRS ≥ 4)
and alive on the originally implanted device, or transplanted or explanted for recovery.
Accessories
Monitor
6
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
9
Explant
A binary analysis from the pre-specied interim analysis is presented in Table 17.
Reference
10
Guides
Appendix
34HVAD™ Instructions for Use
1.9 US Clinical Study: Destination Therapy (continued)
1
Table 17: Binary Analysis of the Primary ENDURANCE Endpoint and its Components for Subjects
Receiving Study or Control Device
Event Free Survival at 2 years
Success51.5% (103)59 (59.0%)
Failure48.5% (97)41.0% (41)
If Failure, reason:
Patient dies35.5% (71)25.0% (25)
Device malfunction or failure
requiring exchange, explant or
urgent transplant
Exchange
Explant
Urgent Transplant
Disabling stroke
(mRS ≥ 4 at 24 weeks)
Imputed failure*0.5% (1)0.0% (0)
A subject may have multiple reasons for not completing the rst two (2) years, only the rst failure type for
each subject is specied.
*Patient experienced a stroke prior to their 2-year endpoint, and died beyond the 2 year endpoint, but
before the 24 week mRS assessment.
HVAD
(N=200)
11.0% (22)
9.5% (19)
0.0% (0)
1.5% (3)
1.5% (3)0.0% (0)
Control
(N=100)
16.0% (16)
14.0% (14)
0.0% (0)
2.0% (2)
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
2. Secondary Endpoints
Because the primary endpoint was not met per the pre-specied interim analysis, the
hypotheses associated with the secondary endpoints of incidence of bleeding (per
INTERMACS® denition), incidence of major infections (per INTERMACS® denition), and
overall survival (time to death) could not be tested. As such, the secondary endpoints were
not reported.
3. Other Results - Adjunctive analysis: Primary Endpoint Using Expanded Dataset
Following the interim analysis at 300 patients, the trial was expanded to enroll additional
patients to further investigate various device, procedural, and clinical changes introduced
during the trial. A total of 451 patients (inclusive of the initial 300 patients) were enrolled,
of which 445 were implanted with a device. The patient disposition is summarized in Figure
11.The results of the expanded dataset are summarized below.
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
35Introduction
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
1.9 US Clinical Study: Destination Therapy (continued)
Figure 11: Disposition of Subjects in the ENDURANCE Expanded Dataset
Accessories
Monitor
6
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
9
The demographics and baseline characteristics of the ENDURANCE expanded dataset is
summarized in Table 18. The demographics and baseline characteristics are typical for an LVAD
study performed in the U.S. The HVAD and Control groups did not differ signicantly with respect
to severity of illness, baseline hemodynamic characteristics, or treatment with evidence-based
Explant
therapy for heart failure at the time of enrollment. However, subjects in the control group were
slightly older (66.2 versus 63.9, control versus HVAD, P=0.04).
Reference
10
Guides
Appendix
36HVAD™ Instructions for Use
1.9 US Clinical Study: Destination Therapy (continued)
1
Table 18: Patient Demographics and Baseline Characteristics of the ENDURANCE Expanded
Dataset
Demographics and Baseline
Characteristics
Age (years)63.9 ± 11.666.2 ± 10.20.044
Male gender (%)76.4%82.4%0.178
Race (%)
White
Black or African American
Other
Height (cm)173.8 ± 9.4175.5 ± 9.10.068
Body Surface Area (m2)2.0 ± 0.32.0 ± 0.30.615
INTERMACS Prole (%)
1
2
3
4
5-7
Ischemic Etiology of Heart Failure 57.9%60.1%0.684
Smoker68.0%62.2%0.243
Diabetic44.4%43.9%> 0.999
Previous Stroke/TIA19.2%16.2%0.515
Hypertension requiring medication65.3%70.9%0.241
Serum creatinine (mg/dL)1.5 ± 0.51.4 ± 0.50.760
Severe tricuspid valve insufciency
Left ventricular ejection fraction (LVEF, %)17.1 ± 4.616.2 ± 4.80.055
HVAD
(N=297)
76.8%
22.2%
1.0%
3.4%
29.0%
40.4%
19.9%
7.4%
12.0%
(N=292)
Control
(N=148)
77.7%
21.6%
0.7%
3.4%
31.1%
40.5%
18.2%
6.8%
5.5%
(N=146)
P-value
0.962
0.989
0.040
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Explant
7
Patient
Management
Survival free from disabling stroke (mRS ≥4) and alive on the originally implanted device, or
transplanted or explanted for recovery for the complete ENDURANCE population is shown
below in Figure 12. The expanded dataset includes a higher proportion of HVAD devices
having titanium-sintered inow cannulae, a device modication that was introduced during
ENDURANCE and designed to decrease thromboembolic adverse event rates. Post hoc oneyear comparisons of all sintered HVADs (pooled from both ENDURANCE and ENDURANCESupplemental) to pooled Control subjects were also performed, as shown in Figure 12.
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
37Introduction
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
1.9 US Clinical Study: Destination Therapy (continued)
Figure 12: ENDURANCE Trial Expanded Dataset: Survival free from disabling stroke (mRS ≥4) and alive on
the originally implanted device, or transplanted or explanted for recovery in the overall study
dataset.
Accessories
Monitor
6
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
9
The post hoc comparison of sintered and non-sintered HVAD™ Pumps in the interim analysis
cohort did not demonstrate markedly different results (See Figure 13A, 13B).
Figure 13: Comparison of Outcomes from the Interim analysis of Subjects with Sintered Pumps Compared to
Control: Survival free from disabling stroke (mRS ≥4) and alive on the originally implanted device,
or transplanted or explanted for recovery in A) the subset of subjects receiving a sintered
HVAD ™ Pump, compared to Control, and in B) the subset of subjects receiving the non-sintered
Explant
HVAD™ Pump. This analysis is based on the as-treated population.
A) Sintered:
Reference
10
Guides
Appendix
38HVAD™ Instructions for Use
1.9 US Clinical Study: Destination Therapy (continued)
1
B) Non-Sintered:
Table 19: Binary Analysis of ENDURANCE Expanded Dataset: Survival at 2 years free from
disabling stroke (mRS ≥4) and alive on the originally implanted device, or transplanted
or explanted for recovery.
Event Free Survival at 2 years
Success55.2% (164)57.4% (85)
Failure 44.8% (133)42.6% (63)
If Failure, reason:
Patient dies34.7% (103)26.4% (39)
Device malfunction or failure
requiring exchange, explant or
urgent transplant
Exchange
Explant
Urgent Transplant
Disabling stroke
(mRS ≥ 4 at 24 weeks)
Imputed failure*0.3% (1)0
HVAD
(N=297)
8.8% (26)
7.7% (23)
0.0% (0)
1.0% (3)
1.0% (3)0
Control
(N=148)
16.2% (24)
13.5% (20)
0.7% (1)
2.0% (3)
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
A subject may have multiple reasons for not completing the rst two (2) years, only the rst failure type for
each subject is specied.
*Patient experienced a stroke prior to their 2-year endpoint, and died beyond the 2 year endpoint, but
before the 24 week mRS assessment.
9
Reference
Guides
10
Appendix
39Introduction
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Monitor
6
Accessories
1.9 US Clinical Study: Destination Therapy (continued)
In the analyses presented on the entire ENDURANCE trial cohort, the secondary endpoints were
analyzed and descriptive data include the following:
• The incidence of bleeding was 60.1% for the HVAD compared to 60.4% for the Control.
• The incidence of major infections was 69.3% for the HVAD and 62.4% for the Control.
• Overall survival was 60.2% for the HVAD and 67.6% for the Control.
The CEC adjudicated causes of death for the entire ENDURANCE trial cohort are shown in Table 20.
Table 20: ENDURANCE Expanded Dataset Cause of CEC Adjudicated on Device Death within 730
days (AIP as Received)
Cause of Death
Total38.5% (114)30.9% (46)
Bleeding 0.3% (1)0.7% (1)
Cardiovascular procedure1.4% (4)1.3% (2)
Heart failure16.2% (48)14.8% (22)
Infection3.0% (9)2.7% (4)
Malignancy1.4% (4)0.7% (1)
Multisystem organ failure0.0% (0)0.7% (1)
Respiratory failure0.0% (0)0.7% (1)
Stroke8.4% (25)6.0% (9)
Sudden death3.7% (11) 2.0% (3)
Trauma0.7% (2)0.0% (0)
Other cardiovascular2.7% (8)1.3% (2)
Other non-cardiovascular0.7% (2)0.0% (0)
HVAD
(N=296)
Control
(N=149)
Surgical
7
Patient
8
Alarms and
9
Reference
10
Implant and
Explant
Management
Emergencies
Guides
Appendix
40HVAD™ Instructions for Use
1.9 US Clinical Study: Destination Therapy (continued)
1
Overall survival for the ENDURANCE trial expanded dataset beyond the two (2) year timepoint is
included below in Figure 14. Aggregate 5-year mortality results for all ENDURANCE subjects were
similar.
Figure 14: Kaplan Meier Survival (Time to Death) in ENDURANCE through 5 years (PP, Per Protocol).
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
Adverse events
The key safety/adverse event outcomes for the ENDURANCE trial expanded dataset are
presented in Table 21 below. Patients in the HVAD arm had a higher rate of ischemic and
hemorrhagic stroke, sepsis, and right heart failure compared to control. An analysis of the
patient-level data indicated that elevated blood pressure appeared to be a risk factor for
stroke, particularly hemorrhagic stroke.
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
41Introduction
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Monitor
Accessories
1.9 US Clinical Study: Destination Therapy (continued)
Table 21: Summary of INTERMACS® Adverse Events Occurring Through 2 Years in Subjects in the
ENDURANCE Trial Expanded Dataset
Adverse Event
Overall Bleeding events
GI Bleed
Cardiac Arrhythmia37.8% (112)40.9% (61)
Hepatic Dysfunction4.7% (14)8.1% (12)
Hypertension15.9% (47)16.8% (25)
Sepsis23.6% (70)15.4% (23)
Driveline Exit Site Infection19.6% (58)15.4% (23)
Stroke
Ischemic Cerebrovascular Event
Hemorrhagic Cerebrovascular Event
TIA
Renal Dysfunction14.9% (44)12.1% (18)
Respiratory Dysfunction29.1% (86)25.5% (38)
Right Heart Failure
Need for RVAD*
Pump Replacement
Exchange for Pump Thrombosis
Device Malfunction or Failure31.4% (93)25.5% (38)
HVAD
(N=296)
60.1% (178)
35.1% (104)
29.7% (88)
17.6% (52)
14.9% (44)
8.4% (25)
38.5% (114)
2.7% (8)
7.8% (23)
6.4% (19)
Control
(N=149)
60.4% (90)
34.2% (51)
12.1% (18)
8.1% (12)
4.0% (6)
4.7% (7)
26.8% (40)
3.4% (5)
13.4% (20)
10.7% (16)
6
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
*Site-reported event.
Abbreviations: GI - gastrointestinal; RVAD=right ventricular assist device; TIA= transient ischemic attack (<24 hours).
Note: The event of device thrombosis reported is not an INTERMACS®-dened event.
Explant
Stroke-related Deaths
Per CEC adjudication, among the full AIP population 12.5% (37/296) of HVAD patients and
6.7% (10/149) of Control patients had stroke-related deaths (data lock date of May 30, 2017, all
patients with follow-up > 4 years or censored). HVAD subjects in the ENDURANCE trial had a risk
of death from stroke that was 87% greater than the risk of Control patients. The rate of strokerelated death within 2 years of implantation was 8.4% (25/296) for HVAD patients and 6.0%
(9/149) for Control patients. The rate of later-onset stroke-related death (i.e., stroke occurring
after 2 years of LVAD support) was 3.7% (11/296) for HVAD patients and 0.7% (1/149) for Control
patients. The majority of HVADs which were involved with stroke-related deaths had sintered
inlet cannulae.
10
Appendix
42HVAD™ Instructions for Use
1.9 US Clinical Study: Destination Therapy (continued)
1
Device Failures and Malfunctions
The incidence of device failures and device malfunctions within 730 days was 31.4% in the
HVAD arm vs. 25.5% in the Control arm. The rates of pump thrombosis were similar in both arms,
though sintering of the HVAD did appear to decrease this event. Device malfunctions related to
controller faults were substantially more frequent in the HVAD arm.
Table 22: Device Failure or Malfunctions in the ENDURANCE Trial Expanded Dataset
The average number of re-hospitalizations within 730 days after the initial hospitalization
was similar between the HVAD arm and the Control arm, as shown in Figure 15. For the AIP
population, the HVAD subjects were re-hospitalized on average, 4.1 times, compared to 3.6
times in the Control group.
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
43Introduction
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Monitor
6
1.9 US Clinical Study: Destination Therapy (continued)
Figure 15: Average Number of Rehospitalizations over Two Years in the ENDURANCE Trial Expanded Dataset
Accessories
Functional Status
Functional status was assessed by the NYHA class and the 6-minute walk test (6MWT), as
shown in Figures 16 and 17. Following LVAD implant, approximately 70-80% of subjects in both
arms improved to NYHA class I or II by Month 12. The median baseline 6-minute walk distance
(6MWD) was 0 meters for both study and control subjects. At 3 months following LVAD implant,
6MWD increased to a median of 210 meters and 201 meters for study and control subjects,
respectively. These improvements were sustained through two (2) years.
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
Figure 16: ENDURANCE Trial Expanded Dataset Six-Minute Walk Test
Explant
10
Appendix
44HVAD™ Instructions for Use
1.9 US Clinical Study: Destination Therapy (continued)
The quality of life was assessed by the EQ-5D-5L and the KCCQ questionnaires, as summarized
in Figure 18. At baseline, subjects in both cohorts had poor quality of life and health status
assessed by KCCQ and EuroQOL EQ-5D. At 3 months, median KCCQ score had improved by
27.3 points and 24.2 points for study and control subjects, respectively. EuroQOL EQ-5D VAS
improved an average of 1.6 points at 3 months for subjects in the study arm and 1.7 points at
3 months for subjects in the control arm. Improvements in KCCQ and EuroQOL EQ-5D were
sustained during the follow-up period.
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
Figure 18: Improvements in Quality of Life and Functional Capacity in the ENDURANCE Trial Expanded
Dataset. A) Change over time of the KCCQ Overall Summary Score. B) Change over baseline in the EQ-5D
Visual Analog Scale.
A. KCCQ
100.0
80.0
60.0
40.0
20.0
Overall Summary Score
0.0
HVAD
Control
3 monthBaseline
6 month 12 month 24 month
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
45Introduction
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Monitor
Accessories
1.9 US Clinical Study: Destination Therapy (continued)
B. EQ-5D
3.0
2.5
2.0
1.5
1.0
0.5
Change from Baseline
0.0
Subgroup Analyses
The following preoperative characteristics were evaluated for potential association with
outcomes: gender, and BSA (< 1.5 m2, ≥ 1.5 m2). The pre-specied sub-group analyses showed
no major clinical differences in outcomes based on gender or BSA.
HVAD
Control
3 months
6 months12 months 24 months
6
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
9
Conclusions from the ENDURANCE Destination Therapy Trial
The ENDURANCE trial did not meet its pre-specied primary endpoint, a demonstration of non-
inferiority of the HVAD to the control device for patients alive on original device at two (2) years
free from disabling stroke (mRS >4). However, an adjunctive analysis using the full-enrollment
Explant
dataset demonstrated similar endpoint results, with 57.4% success for control and 55% success
for HVAD. Following LVAD implant, approximately 80% of subjects in both arms improved to
NYHA class I or II symptomatology. At 3 months following LVAD implant, median 6 minute walk
distance increased in both arms (210 meters and 201 meters for study and control subjects,
respectively). Patients in both arms also showed comparable improvement in quality of life
from baseline to 3 months as measured by EQ-5D-5L and KCCQ, and the results were sustained
through 2 years.
Reference
10
Guides
Appendix
46HVAD™ Instructions for Use
1.10 Destination Therapy Supplemental Study
1
A. Study Design
The objective of the ENDURANCE Supplemental trial was to evaluate the safety and
effectiveness of a prospective blood pressure management strategy in HVAD DT patients.
The purpose of implementing the prospective blood pressure management strategy was
to investigate its effect on the stroke rates in HVAD subjects. The trial was a prospective,
randomized, controlled, un-blinded, multicenter clinical study. Subjects were randomly assigned
using a permuted block, central randomization scheme, in a 2:1 ratio, to receive either the study
(HVAD) or control (HeartMate II) device. All HVAD subjects, in addition to receiving standard
of care management, were required to adhere to a blood pressure management protocol
that aimed to maintain mean arterial pressure (MAP) ≤ 85 mmHg (automated pneumatic cuff
method) or < 90 mmHg (Doppler cuff method). Control patients were not managed with a
blood pressure management protocol.
Patients in the ENDURANCE Supplemental trial were enrolled between October 25, 2013 and
August 7, 2015. 475 subjects were randomized, with 465 patients implanted at 47 investigational
sites.
Similar to the ENDURANCE trial, the ENDURANCE Supplemental trial was conducted under the
oversight of an independent Clinical Events Committee and monitored by an independent
Data Safety Monitoring Board.
1. Clinical Inclusion and Exclusion Criteria
Enrollment in the ENDURANCE Supplemental trial was limited to patients who met the same
inclusion and exclusion criteria as in the ENDURANCE trial.
2. Follow-up Schedule
All patients were scheduled to return for follow-up examinations at 3 and 6 months with a
window of ± 7 days, at 12 months with a window of + 7 days, and at 18, 24, 30, 36, 42, 48, 54,
and 60 months with a window of ± 14 days postoperatively.
The pre- and post-operative assessments were the same the in the ENDURANCE trial.
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Explant
3. Clinical Endpoints
The primary endpoint was the incidence of neurologic injury at 12 months. Neurologic injury
was dened as an ICVA or HCVA with mRS > 0 at 24 weeks post-stroke, or a TIA, or as a
spinal cord infarct (SCI).
The HVAD was to be considered non-inferior to the HeartMate II if the upper bound of the
two-sided 90% exact binomial condence interval of the difference in the primary endpoint
between the HVAD arm and the control arm was less than 6%.
There were two secondary endpoints. The rst secondary endpoint was incidence of HVAD
stroke and TIA by 12 months on the originally implanted HVAD. Unlike the primary endpoint,
this secondary endpoint included those strokes that were classied as mRS=0 at 24 weeks
post-stroke. This endpoint was to be tested by comparison to a performance goal of 17.7%;
the performance goal was equivalent to the lower 95% condence interval of the one-year
stroke/TIA rate among sintered HVAD patients in the ENDURANCE trial.
The second secondary effectiveness endpoint was analogous to the ENDURANCE trial’s
primary endpoint, in that it compared the composite of stroke-free (mRS < 4 at 24 weeks
post-stroke) survival while on the original device between HVAD and Control arms;
however, the time point for this endpoint was one year, unlike the ENDURANCE trial’s 2-year
endpoint. This endpoint was to test for non-inferiority of the HVAD to the control device, with
a non-inferiority margin of 15%.
Additional endpoints included adverse events, device malfunctions and failures, as well as
health status and functional improvements.
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
47Introduction
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
1.10 Destination Therapy Supplemental Study (continued)
B. Accountability of PMA Cohort
At the time of database lock, of the 494 patients enrolled in the ENDURANCE Supplemental trial,
93.7% (463) patients were available for analysis of the primary objective at the completion of the
study, the 12-month post-operative visit. The disposition of the patients is shown in Figure 19.
Figure 19: Disposition of Subjects in the ENDURANCE Supplemental Trial
Accessories
Monitor
6
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
9
Explant
The Modied Intent-to-Treat Population (mITT; Total N=465; HVAD, N=308 and Control, N=157)
included all subjects who received a device. It was analyzed according to the device to which
the subjects were randomized.
All safety analyses were performed on the safety population (SAF), which assigned subjects to
the device they actually received. The SAF was equivalent to the mITT population.
The Complete Case Population includes all subjects in the mITT population except those who
withdraw, are lost to follow-up, or have missing outcomes (any subject with missing post-event
mRS) on original device. It differs for each objective. For the primary endpoint, the Complete
Case Population was dened as the mITT population excluding any subjects who withdrew or
were lost to follow-up, and any subjects who were missing CEC adjudicated mRS scores (both
day of event and 24 weeks post-event) for the latest stroke event on original device. For the
secondary endpoint of stroke/TIA incidence at 12 months on the originally implanted HVAD,
the Complete Case Population was dened as the mITT population excluding the subjects who
withdrew or were lost to follow-up. For the secondary endpoint of stroke-free success (mRS < 4
at 24 weeks post-stroke) at 12 months, the Complete Case Population was dened as the mITT
population excluding subjects who withdrew or were lost to follow-up, and those subjects who
were missing a 24 week mRS score for their last stroke on original device (within 1 year post
original implant).
Reference
10
Guides
Appendix
48HVAD™ Instructions for Use
1.10 Destination Therapy Supplemental Study (continued)
1
C. Study Population Demographics and Baseline Parameters
The demographics and baseline characteristics of the study population, as summarized in Table 23,
are typical for an LVAD study performed in the U.S. The baseline characteristics of the two
(2) arms were similar; there was no clinically signicant difference in the severity of illness or
treatments at the time of enrollment.
Table 23: Patient Demographics and Baseline Characteristics in the ENDURANCE Supplemental Trial.
Demographics and Baseline
Characteristics
Age (years)63.3 ± 11.464.2 ± 11.10.39
Female gender (%)18.2%20.4%0.62
Race (% White)71.8%75.2%0.51
Height (cm)175.0 ± 9.4175.1 ± 9.80.91
Body Mass Index (kg/m2)28.2 ± 5.527.4 ± 5.20.13
INTERMACS Prole (%)
1
2
3
4-7
Ischemic Etiology of Heart Failure 55.2%58.0%0.62
History of smoking68.2%65.6%0.60
Diabetic49.4%48.4%0.92
Previous Stroke10.4%8.3%0.51
Hypertension requiring medication75.0%72.0%0.50
Atrial Fibrillation50.6%51.0%> 0.99
Mean arterial blood pressure (mmHg)
Tricuspid regurgitation (≥ moderate)
Left ventricular ejection fraction (LVEF, %)17.3 ± 5.118.2 ± 4.50.07
Note: P-values are post-hoc and are included for information purposes only.
P-values comparing categorical values are from the Fisher’s Exact Test. P-values comparing continuous
values are from a two-sample t-test.
HVAD
(N=308)
3.9%
32.8%
43.3%
20.0%
78.9 ± 11.5
(N=296)
40.4%
(N=302)
80.8%
28.9%
19.2%
Control
(N=157)
2.5%
32.5%
43.3%
21.7%
77.6 ± 11.1
(N=153)
44.2%
(N=154)
82.2%
28.7%
15.9%
P-value
0.90
0.23
0.48
0.80
> 0.99
0.45
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
49Introduction
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Monitor
6
Surgical
Implant and
Accessories
Explant
1.10 Destination Therapy Supplemental Study (continued)
D. Safety and Effectiveness Results
1. Primary Endpoint
The outcome and analysis of the primary endpoint are shown in Table 24 and Figure 20.
The results show that 14.7% of the HVAD subjects experienced endpoint-dened neurologic
injury as compared to 12.1% of the control subjects, with a difference of 2.6% between the
two arms. The upper bound of the two-sided 90% exact binomial condence interval of the
difference in the neurologic injury incidence was 10.7%, which was above the pre-specied
non-inferiority margin of 6%. Thus, the primary endpoint of the ENDURANCE Supplemental
trial was not met.
Table 24: Analyses of the Primary Endpoint
HVAD
(N=306)
Number of subjects who had a stroke/TIA at 12 months5824
Number of subjects who had a stroke at 12 months5123
Number of subjects who had a TIA at 12 months
Number of subjects who had mRS > 0 at 24 weeks
post-stroke
Number of subjects who had spinal cord infarction
at 12 months
1.10 Destination Therapy Supplemental Study (continued)
1
2. Secondary Endpoints
Because the primary endpoint was not met, the hypotheses associated with the secondary
endpoints of stroke/TIA incidence and stroke-free success rate could not be tested. Thus,
only descriptive data are presented for the two secondary endpoints.
The incidence of stroke/TIA (inclusive of strokes with mRS = 0 at the 24 week time point) in
HVAD patients was 19.2% at 12 months. The Time to event curve is shown in Figure 21.
Figure 21: ENDURANCE Supplemental Trial Survival Free from Stroke or TIA
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
The proportion of subjects who survived to one year on the original device in the absence of
“disabling” stroke (mRS ≥4), death, device exchange or urgent transplantation was 75.3% in the
HVAD arm and 66.7% in the Control arm. A freedom from event analysis is shown in Figure 21, using
data from March 27, 2017. The magnitude of the rate differential for this composite decreased
with follow-up more analogous to the ENDURANCE trial’s 2-year endpoint time frame.
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
51Introduction
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
1.10 Destination Therapy Supplemental Study (continued)
Figure 22: ENDURANCE Supplemental Trial Survival Free from Death, Disabling Stroke or Device Malfunction/
Failure Requiring Exchange
Accessories
Monitor
6
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
9
In the ENDURANCE Supplemental trial, freedom from ischemic stroke was numerically greater
in the Control arm, as shown in Figure 22; freedom from hemorrhagic stroke was similar in HVAD
and Control, as shown in Figure 23.
Figure 23: ENDURANCE Supplemental Trial Survival Free from Ischemic Stroke
Explant
Reference
10
Guides
Appendix
52HVAD™ Instructions for Use
1.10 Destination Therapy Supplemental Study (continued)
1
Figure 24: ENDURANCE Supplemental Trial Survival Free from Hemorrhagic Stroke
3. Adverse Events
Table 25 lists all the adverse events that occurred in the safety cohort.
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
Table 25: Summary of Adverse Events at 1 Year in the ENDURANCE Supplemental Trial.
Adverse Event
Major Bleeding51.6% (159)56.7% (89)
Cardiac Arrhythmia34.1% (105)31.2% (49)
Hepatic Dysfunction3.9% (12)3.8% (6)
Hypertension13.0% (40)12.7% (20)
Major Infection53.9% (166)59.2% (93)
Driveline Exit Site Infection16.2% (50)12.1% (19)
Device Malfunction/Failure24.0% (74)24.2% (38)
Hemolysis1.3% (4)5.7% (9)
Stroke
Ischemic Cerebrovascular Event
Hemorrhagic Cerebrovascular Event
TIA
Renal Dysfunction10.4% (32)14.6% (23)
Respiratory Failure19.8% (61)19.7% (31)
Right Heart Failure35.4% (109)38.2% (60)
Pump Replacement5.2% (16)11.5% (18)
Exchange for Pump Thrombosis4.5% (14)10.2% (16)
HVAD
(N=308)
16.9% (52)
13.0% (40)
5.2% (16)
4.2% (13)
Control
(N=157)
14.6% (23)
7.6% (12)
7.0% (11)
0.6% (1)
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
53Introduction
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Accessories
1.10 Destination Therapy Supplemental Study (continued)
Stroke-related Deaths
Within the mITT population, the CEC-adjudicated rate of stroke-related death within 1 year of
implantation was 3.2% (10/308) for HVAD patients and 2.5% (4/157) for Control patients.
Comparing the results of ENDURANCE and ENDURANCE Supplemental, the rates of strokerelated death decreased by the same proportions (approximately 58%) for both HVAD and
Control arms; only the HVAD arm was exposed to the trial’s investigational intervention of a
blood pressure management protocol. The stroke-related deaths are compared in Table 26. The
MAP over time from the ENDURANCE Supplemental trial for the HVAD compared to the Control is
shown in Figure 25.
Table 26: Stroke-related Deaths in ENDURANCE and ENDURANCE Supplemental Trials
HVAD
HMII (control)
ENDURANCE
Within 2 years of implant (AIP)
25/296 (8.4%)10/308 (3.2%)
9/149 (6.0%)4/157 (2.5%)
ENDURANCE Supplemental
Within 1 year of implant (mITT)
Monitor
6
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
9
Figure 25:ENDURANCE Supplemental Trial MAP over Time
Explant
Health Status and Functional Improvements
The improvements in quality of life, as measured by the KCCQ and EQ-5D-5L, and functional
capacity, as measured by the 6 minute walk test and NYHA Class improvement, in the
ENDURANCE Supplement trial patients are presented in Figure 26.
Reference
10
Guides
Appendix
54HVAD™ Instructions for Use
1.10 Destination Therapy Supplemental Study (continued)
1
Figure 26: Improvements in Quality of Life and Functional Capacity in ENDURANCE Supplemental Subjects.
A) Change over time of the KCCQ Overall Summary Score. B) Change over time in the EQ-5D
Visual Analog Scale. C) Change over time of total distance walked in the Six Minute Walk Test. D)
Percent of patients with 2 or more class increase in NYHA Classication at 12 months compared
to baseline.
A. KCCQ
B. EQ-5D
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
55Introduction
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Accessories
1.10 Destination Therapy Supplemental Study (continued)
C. Six-Minute Walk
D. NYHA Classication Improvement
Monitor
6
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
9
Explant
Figure 27: Average Number of Rehospitalizations in the First-Year Post Implant in the ENDURANCE
Supplemental Trial
Reference
10
Guides
Appendix
56HVAD™ Instructions for Use
1.10 Destination Therapy Supplemental Study (continued)
1
Subgroup Analyses
The following preoperative characteristics were evaluated for potential association with
outcomes: gender, BSA (<1.5 m2, ≥1.5 m2). No associations to outcomes of the primary and
secondary endpoints were found for these two preoperative characteristics.
Conclusions from the ENDURANCE Destination Therapy Supplemental Trial
The ENDURANCE Supplemental trial did not meet its pre-specied primary endpoint, a
demonstration of non-inferiority of the HVAD to the control device for freedom from neurologic
injury (stroke with mRS >0 at 24 weeks post stroke or a transient ischemic attack) at 12 months
(HVAD: 14.7% vs control: 12.1%). The combined rate of stroke and TIA in HVAD patients at one
year did not meet a performance goal derived from the rate observed in ENDURANCE. Survival
at 1 year free from the composite of disabling stroke or device exchange favored the HVAD
System (HVAD: 75.3% vs control: 66.7%), though the trend diminished in magnitude over time
(at 2 years, HVAD: 59.2% vs Control: 55.2%). The HVAD System and Control both demonstrated
sustained improvements in quality of life, functional capacity, and NYHA classication. Finally,
although the HVAD failed to demonstrate non-inferiority compared to Control for incidence of
neurological injury at one year, the implementation of a blood pressure management strategy
for HVAD recipients did demonstrate a reduction in the overall stroke rates in patients receiving
an HVAD System in the ENDURANCE Supplemental trial as compared to the rst ENDURANCE
trial.
Additional long-term follow-up from the ENDURANCE Destination Therapy Supplemental Trial
As of July 30, 2018 additional long-term data from the ENDURANCE Supplemental trial showed
the comparative incidences and event-rates for adverse events between the HVAD and Control
groups remained similar to those observed in the initial data analysis. The incidence of TIA in the
HVAD cohort (9.1%) is signicantly higher than the Control cohort (3.8%). However, the overall
myocardial infarction event rate was statistically higher in the Control group, while the overall
ICVA event rate was statistically higher in the HVAD group. So, while it appears that the number
of subjects having ICVA events is similar, HVAD subjects are more frequently having multiple
occurrences of these events.
There was no statistically signicant difference between treatments for freedom from thrombus
on original device, freedom from exchange, freedom from stroke on original device, freedom
from ICVA on original device, or freedom from HCVA on original device.
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
57Introduction
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
1.10 Destination Therapy Supplemental Study (continued)
The Kaplan-Meier estimates for freedom from CEC adjudicated stroke, ICVA and HCVA on
original device are shown in Figure 28, Figure 29, and Figure 30, below. The log-rank p-values
show no signicant differences between HVAD and Control for stroke, ICVA, or HCVA.
Figure 28: ENDURANCE Supplemental Trial Freedom from Stroke on Original Device
Accessories
Monitor
6
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
9
Figure 29: ENDURANCE Supplemental Trial Freedom from ICVA on Original Device
Explant
Reference
10
Guides
Appendix
58HVAD™ Instructions for Use
1.10 Destination Therapy Supplemental Study (continued)
1
Figure 30: ENDURANCE Supplemental Trial Freedom from HCVA on Original Device
Figure 31: ENDURANCE Supplemental Trial Time to Device Malfunction
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
While the incidence of device malfunction/failure was numerically higher in the HVAD group
(39.9% vs. 36.9%), the incidence of pump thrombosis and pump replacement was numerically
higher in the Control group (21.0% vs. 20.1%, and 17.2% vs. 13.3%, respectively). Kaplan-Meier
estimates for freedom from thrombus on original device are presented in Figure 32 below. There
is no signicant difference between the two groups.
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
59Introduction
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
1.10 Destination Therapy Supplemental Study (continued)
Figure 32: ENDURANCE Supplemental Trial Freedom from Thrombus on Original Device
Accessories
Figure 33: ENDURANCE Supplemental Trial Freedom From Exchange
Monitor
6
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
9
Explant
Overall, the incidences of the adverse events still appear similar between the two groups.
Reference
10
Guides
Appendix
60HVAD™ Instructions for Use
1.10 Destination Therapy Supplemental Study (continued)
1
Table 27: Summary of Adverse Events on Original Device in the ENDURANCE Supplemental Trial
Adverse Event
Major Bleeding 63.6% (190) 66.2% (104)
Cardiac Arrhythmia 39.% (122) 36.3% (57)
Hepatic Dysfunction 5.8% (18) 5.1% (8)
Hypertension 14.3% (44) 15.9% (25)
Major Infection 73.7% (227) 73.2% (115)
Device Malfunction/Failure 39.9% (114) 36.9% (58)
Pump Thrombosis20.1% (62)21.0% (33)
Hemolysis 7.1% (22) 6.4% (10)
Stroke
Ischemic Cerebrovascular
Event
Hemorrhagic Cerebrovascular
Event
TIA
Renal Dysfunction 14.6% (45) 20.4% (32)
Respiratory Failure 24.0% (74) 28.7% (45)
Right Heart Failure 36.7% (113) 40.8% (64)
Pump Replacement 12.3% (38) 17.2% (27)
Additional post hoc one-year comparisons of all sintered HVADS (pooled from both ENDURANCE
and ENDURANCE-Supplemental) to pooled Control subjects were also performed, and analyzed
against the primary endpoint denition of the ENDURANCE Trial (at one year, Figure 34B) and
against the primary endpoint of the ENDURANCE Supplemental Trial (Figure 34A).
An Analysis of Patients Receiving Sintered HVAD™ Pumps (Pooled ENDURANCE and ENDURANCE
Supplemental) Compared to Control. A) the Primary Endpoint of the ENDURANCE Supplemental
Trial, and B) the Primary Endpoint of the ENDURANCE Trial at 1 year.
HVAD
(N=308)
25.3% (78)22.3% (35)
17.5% (54)12.7% (20)
10.4% (32)11.5% (18)
9.1% (28)3.8% (6)
Control
(N=157)
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
61Introduction
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
1.10 Destination Therapy Supplemental Study (continued)
Figure 34A: Survival on Original Device Free from Neurologic Events (Strokes with mRS>0, TIA or SCI)
Accessories
Figure 34B: Survival on Original Device Free from Disabling Stroke (mRS ≥4)
Monitor
6
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
9
Explant
Reference
10
Guides
Appendix
62HVAD™ Instructions for Use
1.10 Destination Therapy Supplemental Study (continued)
1
Overall Conclusions
The overall safety comparisons in both the ENDURANCE and ENDURANCE Supplemental trials
resulted in similar mortality rates and adverse event proles. Pump thrombosis rates for the
sintered HVAD and the Control LVAD were similar, but a higher proportion of Control pump
thrombosis events resulted in device exchange. The incidence of stroke was 2.5 times higher in
the patients receiving an HVAD compared to control in the ENDURANCE trial. The ENDURANCE
Supplemental trial, which included implementation of a blood pressure management strategy
for HVAD recipients, demonstrated a reduction in the overall stroke rates in patients receiving
an HVAD System, although a reduction in overall stroke rates was also demonstrated in Control
patients who were not subject to the blood pressure management strategy. In ENDURANCE
Supplemental, the HVAD failed to demonstrate non-inferiority compared to Control for
incidence of neurological injury at one year.
The data supports the reasonable assurance of safety and effectiveness of the HVAD System for
hemodynamic support in patients with advanced, refractory left ventricular heart failure; either
as a bridge to cardiac transplantation (BTT), myocardial recovery, or as destination therapy (DT)
in patients for whom subsequent transplantation is not planned.
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
63Introduction
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Accessories
1.11 North American Clinical Study: LATERAL
HeartWare™ HVAD™ System Clinical Trial Overview: LATERAL Trial
This was a multi-center, prospective, open-label non-randomized single arm trial conducted
in collaboration with the InterAgency Registry for Mechanically Assisted Circulatory Support
(INTERMACS®) to Evaluate the Thoracotomy Implant Technique of the HVAD System in Patients
with Advanced Heart Failure. Enrollment in the study is complete, subjects have all reached
the primary endpoint as described and specied in the protocol, but follow-up of subjects is
ongoing.
A. Study Design
Patients were enrolled between January 15, 2015 and April 26, 2016. The study data was
collected through June 15, 2017 and included 144 subjects treated per protocol who were
enrolled at 26 investigational sites.
The study was a prospective, single arm, multi-center clinical study in collaboration with the
InterAgency Registry for Mechanically Assisted Circulatory Support (INTERMACS®) to evaluate
the thoracotomy implant technique of the HVAD System in patients with advanced heart failure.
The study device was the HVAD (HeartWare Ventricular Assist Device) System. The treatment was
open-label.
Following implantation, device performance, follow-up visits and visit windows for the LATERAL
Study are dictated by the INTERMACS® protocol.
Monitor
6
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
Adverse events (AE) were reported through INTERMACS®, according to the INTERMACS® AE
denitions. An independent Data Safety Monitoring Board monitored and reviewed study
compliance, adverse events and outcomes.
An NHLBI-appointed (independent) Observational Study Monitoring Board (OSMB) was
established in 2006 and meets, at minimum, on an annual basis. The principal role of the OSMB
is to monitor data from the Registry, review and assess the performance of its operations, assure
patient safety, and make recommendations to the NHLBI and INTERMACS® co-investigators.
Explant
1. Clinical Inclusion and Exclusion Criteria
Enrollment in the LATERAL Study was limited to subjects who met the following inclusion
criteria:
-Subjects >19 years old with chronic, advanced left ventricular failure who were
transplant eligible at the time of enrollment in whom informed consent was obtained.
Subjects were not permitted to enroll in the LATERAL Study if they met any of the following
exclusion criteria:
-Subjects with a body surface area < 1.2 m
-Subjects with prior cardiac transplant
-Subjects with a mechanical heart valve
-Subjects with severe right heart failure or receiving biventricular or the device as a
right ventricular assist device
-Subjects with a planned concurrent procedure
-Subjects with known LV thrombus
-Additional exclusion criteria are available in the Clinical Study Report.
2
10
Appendix
64HVAD™ Instructions for Use
1.11 North American Clinical Study: LATERAL (continued)
1
2. Follow-up Schedule
Preoperative baseline assessments included demographics, medical history, concurrent
medications, laboratory tests, echocardiogram, NYHA, neurocognitive status, quality of
life, and functional status.
All patients were scheduled to return for follow-up examinations at Week 1 +/- 3 days,
Month 1 +/- 7 days, Month 3 +/- 30 days, Month 6 +/- 60 days (primary endpoint), and
every 6 months +/- 60 days through 5 years of follow-up.
Week 1 and Month 1 visits included clinical laboratory tests for hematology, chemistry
and INR, LVAD parameters, concurrent medications, echocardiogram, and assessments
of NYHA. Months 3, 6 and onward also included health status, six-minute walk and an
assessment of neurocognitive status.
Adverse events and complications were recorded at all visits.
The key timepoints are shown below in the tables summarizing safety and effectiveness.
3. Clinical Endpoints
The primary endpoint was a composite of six-month survival free of disabling stroke (i.e.,
modied Rankin score ≥ 4 assessed 12 weeks post-event), while alive on the originally
implanted device, transplanted or explanted due to left ventricular recovery.
Success in meeting the primary endpoint was tested comparing to a performance goal
(77.5%) using a one-sided exact binomial test. Success will be met if the lower bound of
the one-sided exact 95% condence limit is greater than 77.5%. Success at six months
is estimated to be 86% compared to a performance goal of 77.5%. The target success
estimate was based on the primary endpoint observed in the ADVANCE BTT+CAP Trial,
post-approval data on HVAD outcomes from the INTERMACS® Registry (through Q2 2014),
and the INTERMACS® report from Q1 2014 indicating an 85% survival estimate with LVAD
support. Using an exact binomial test, with a one-sided alpha of 0.05, and 80% Power, a
sample size of 145 implanted subjects was planned.
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Explant
The prespecied secondary endpoint is an improvement in the mean length of initial
hospital stay as compared to a performance goal of median sternotomy subjects. The
mean length of initial hospital stay is estimated to be 26.1 days with a standard deviation
of 22.8 days and a median of 20 days based on data from the Bridge to Transplantation
Continued Access Protocol (BTT CAP) population (N=242). Using a one sample t-test, with
a one-sided alpha of 0.05, a sample of 145 implanted subjects with an average value of
21.3 days or less will result in Power greater than 80%.
Other secondary endpoints included the incidence of major adverse events (classied
according to the INTERMACS® denitions), overall survival, changes in quality of life and
health status as assessed by the KCCQ and the EQ-5D VAS, and functional status, as
measured by NYHA functional class and 6-minute walk distance. The safety analysis
focused on adverse events. Survival analysis was performed using the Kaplan-Meier
method.
With regards to safety, predetermined secondary endpoints included the incidence
bleeding, major infections, and overall survival. Additionally, incidence of device
malfunctions/failures, quality of life and health status changes, as well as major adverse
events (classied according to the INTERMACS® denitions) were analyzed.
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
65Introduction
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Monitor
6
Accessories
1.11 North American Clinical Study: LATERAL (continued)
B. Accountability of Cohort
At the time of the data cut-off for this analysis (June 15, 2017), of the 178 patients enrolled
in the LATERAL study, with 158 subjects qualifying for the analysis population, 98.7% (156)
subjects were available for analysis of the primary objective at the completion of the
study, the 6-month post-operative visit. The disposition of patients is shown in Figure 35.
The primary analysis population is the Per Protocol (PP) population, including a total of 144
subjects implanted with an HVAD via a thoracotomy approach and meeting all inclusion
criteria without violating any exclusion criteria. The mean duration of subjects on original
device in the PP population is 12.1 months. Eleven (11) additional subjects were included
in the intent-to-treat population. Of the 23 screen failures, 3 subjects were implanted via a
thoracotomy and were thus included in additional secondary endpoint analyses.
Figure 35: Subject Enrollment and Study populations in the LATERAL study.
Signed Informed
Consent
(N=178)
Screen Failure (N=23)
12 = Excluded from
2 = Consent withdrawn for
1 = Transplant prior to implant
5 = Planned Sternotomy
3 = Other LATERAL
Analysis Poplulation N=158
All Enrolled (N=11)
Use of sternotomy and/or
descending aorta outow
graft placement
Intent to Treat
(N=155)
INTERMACS
LATERAL
exclusion criteria met,
but implanted via
thoracotomy
Surgical
7
Patient
8
Alarms and
9
Reference
10
Implant and
Explant
Management
Emergencies
Guides
Per Protocol (PP)
Study Population
(N=144)
The purpose of the LATERAL Study was to evaluate the safety and effectiveness of the implanting
the HVAD System by a thoracotomy approach, therefore patients in whom an implant via
sternotomy was planned were screen failed from the study. Additionally, subjects in whom
thoracotomy was planned, but then were implanted via a sternotomy, or those in whom
the outow was implanted in the descending aorta rather than the ascending aorta were
included in the intent to treat population. Finally, those patients who violated one or more of
the exclusion criteria, but in whom a thoracotomy implant was still completed, were included
in the analysis population. All thoracotomy implant procedures were also recommended to be
completed on cardiopulmonary bypass.
Study Population Demographics and Baseline Parameters
The demographics of the primary study population are typical for an advanced heart failure
with LVAD therapy study performed in the US. The subjects in the trial had advanced heart
failure associated with a substantial risk of death, as evidenced by over 80% of subjects classied
as INTERMACS® Prole 1-3, almost 20% with chronic renal disease, and more than 60% with
ejection fractions lower than 20%. See Table 28.
Appendix
66HVAD™ Instructions for Use
1.11 North American Clinical Study: LATERAL (continued)
1
Table 28: Baseline Demographics and Parameters in LATERAL
Demographics and Baseline CharacteristicsStudy Device N=144
Age (years)54.2 ± 11.5
Male gender (%)77.1%
Race (%)
White
Black or African American
Asian
American Indian or Alaskan Native
Other, none of the above
Unspecied, Undisclosed
Height (cm)175.2 ± 8.8
Body Surface Area (m2)2.0 ± 0.3
Intended use
Bridge to Transplant
Possible Bridge to Transplant
INTERMACS® Prole (%)
1
2
3
4
5-7
Ischemic Etiology of Heart Failure32.6%
Prior Cardiac Surgery22.9%
Previous Major Stroke4.9%
Chronic Renal Disease18.8%
History of Atrial Arrhythmias30.6%
Serum creatinine (mg/dL)1.3 ± 0.74
Mean Arterial Blood Pressure (mmHg)79.5 ± 10.5
Left ventricular ejection fraction <20% (LVEF, %)61.1%
Cardiac Index (L/min/m2)2.1 ± 0.54
62.5%
20.8%
4.9%
2.1%
4.9%
7.6%
73.6%
26.4%
3.5%
31.3%
47.2%
15.3%
2.8%
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Explant
7
Patient
Management
C. Safety and Effectiveness Results
1. Primary Endpoint
The outcome and analysis of the primary endpoint are shown in Table 29. The primary
endpoint success in the LATERAL Study was dened as alive on original device,
transplanted or explanted for recovery without a stroke with an mRS score of ≥ 4 (assessed
≥ three months post-stroke event). One subject had missing data for this endpoint, which
resulted in 143 evaluable subjects. The results show that 88.1% of the HVAD subjects
treated per protocol achieved primary endpoint success, which was signicantly greater
than the 77.5% performance goal (P=0.0012). The most common reason for primary
endpoint failure was death on original device, seen in 7.7% (11/143) of subjects (Table 29).
Protocol-mandated stroke assessments were not carried out in all subjects who
experienced a stroke. In addition, 6 patients were reported as having a stroke within the
rst 6 months of implant. 3 of these 6 patients do not have sufcient data on mRS scores
and the severity of their strokes could not be determined. The Primary Endpoint for these 3
patients are based upon post hoc imputation of mRS scores.
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
67Introduction
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Accessories
1.11 North American Clinical Study: LATERAL (continued)
Table 29: Primary Endpoint at Six Months*
Thoracotomy (N=143)N (%)95% CI *
Primary Endpoint Success126 (88.1%)
Alive on original device
Transplanted
Explanted for Recovery
Primary Endpoint Failure17 (11.9%)
Death on original device
Stroke with mRS>=4
(as assessed >=3 Months post
event)
Exchange
Explanted (not for recovery)
*Table is summarized by subjects’ rst failure event and includes all subjects with an endpoint event prior to 6
months or known to be alive on original device at 6 months.
**P-value of one-sided Binomial Exact test comparing to a performance goal of 77.5%
*** Subject considered a success if all strokes have day of and follow-up mRS scores <4.
97 (67.8%)
29 (20.3%)
0 (0.0%)
11 (7.7%)
4 (2.8%)
1 (0.7%)
1 (0.7%)
0.0012
(82.7%, NA)
Monitor
6
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
9
2. Secondary Endpoints
For the PP population, the mean length of initial hospital stay (initial recovery and step
down unit) after the implant procedure (date of implant to rst discharge) was 18 ± 12.36
days, which is signicantly less than the 26.1 day performance goal estimate for median
sternotomy subjects. (P<0.0001). Additionally, the mean length of stay in the intensive
care unit (ICU) was 8 ± 9.82 days. Results in the ITT population were similar.
Rehospitalization in the PP population was documented in 70.1% of subjects within 6
Explant
Figure 36: Time to Rehospitalization in LATERAL (PP Population, N=144)
months of initial hospitalization. The most common reasons for re-hospitalization were
transplant (27.7%), anticoagulation adjustment (19.8%), and major bleeding (17.8%). A
freedom from rehospitalization is presented in Figure 36.
Reference
10
Guides
Appendix
68HVAD™ Instructions for Use
1.11 North American Clinical Study: LATERAL (continued)
1
Overall survival on the HVAD System was also assessed. A Kaplan-Meier estimate for freedom
from death on original device for the PP population was 91.8% at 6 months, 88.8% at one year,
and 87.4% at two years (Figure 37). The Kaplan-Meier estimates in the ITT population were similar.
Figure 37: Kaplan Meier Survival Analysis in the LATERAL Study (N=144)
Six of 11 deaths occurring within the rst 6 months post-implant were related to circulatory
causes, with two due to right heart failure (1.4%) and two due to sudden unexplained death
(1.4%). The most common non-cardiovascular cause of death was neurological dysfunction,
which occurred in three subjects (2.1%). Additionally, one subject died of multi-system organ
failure and another after withdrawal of life support.
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
69Introduction
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
1.11 North American Clinical Study: LATERAL (continued)
Health Status and Functional Improvements
The improvements in quality of life, as measured by the KCCQ and EQ-5D-5L, and functional
capacity, as measured by the 6-Minute Walk Test and NYHA Class improvement, in the LATERAL
Study subjects are presented in Figure 38A-D.
Figure 38: Improvements in Quality of Life and Functional Capacity in ENDURANCE Supplemental
Subjects. A) Change over time of the KCCQ Overall Summary Score. B) Change over time in the
EQ-5D Visual Analog Scale. C) Change over time in subjects’ NYHA Classication.
D) Change over time of total distance walked in the Six Minute Walk Test.
A. KCCQB. EQ-5D
Accessories
Monitor
6
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
C. NYHA Classication
Improvement
Explant
3. Adverse Events
The analysis of safety was based on the per protocol cohort of 144 patients analyzed
through the primary endpoint. A Clinical Events Committee (CEC) was not utilized in this
study for adverse event adjudication. A total of 537 INTERMACS® adverse events were
reported within 6 months on original device. Adverse events were most often reported
within the rst 30 days post-implant, with 87.5% (126/144) subjects having at least one
INTERMACS®-dened adverse event.
The most common adverse events were cardiac arrhythmia, bleeding, and infections.
Adverse events are summarized in Table 30.
D. Six-Minute Walk
10
Appendix
70HVAD™ Instructions for Use
1.11 North American Clinical Study: LATERAL (continued)
1
Table 30: Summary of INTERMACS® Adverse Events Occurring Through 6 Months
<=30 Days
(N=143)
INTERMACS® Category
Adverse Events
Total Adverse Events126 (87.5)89 (63.6)
Bleeding
Re-Hospitalization
Re-Operation
Transfusion: >=4 within 7
days
GI
Cardiac Arrhythmia
Ventricular
Supraventricular
Device Malfunction/Failure9 (6.3)10 (7.1)
Hepatic Dysfunction1 (0.7)0 (0.0)
Infection
Line Sepsis
Driveline Exit Site
Myocardial Infarction0 (0.0)0 (0.0)
Neurological
Ischemic CVA
Hemorrhagic CVA
TIA
Psychiatric3 (2.1)2 (1.4)
Renal Dysfunction8 (5.6)6 (4.3)
Respiratory Dysfunction11 (7.6)2 (1.4)
Arterial non-CNS
Thromboembolism
Venous Thromboembolism4 (2.8)1 (0.7)
Wound Dehiscence1 (0.7)2 (1.4)
Other20 (13.9)22 (15.7)
abbreviations: GI - gastrointestinal; TIA= transient ischemic attack (<24 hours)
Patients with
Event (%)
15 (10.4)
5 (3.5)
5 (3.5)
13 (9.0)
6 (4.2)
32 (22.2)
20 (13.9)
13 (9.0)
20 (13.9)
0 (0.0)
2 (1.4)
12 (8.3)
3 (2.1)
3 (2.1)
1 (0.7)
0 (0.0)0 (0.0)
> 30 Days – 6 Months
(N=140)
Patients with
Event (%)
20 (14.3)
14 (10.0)
3 (2.1)
16 (11.4)
11 (7.9)
13 (9.3)
9 (6.4)
2 (1.4)
32 (22.9)
1 (0.7)
7 (5.0)
12 (8.6)
3 (2.1)
3 (2.1)
4 (2.9)
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
Nineteen of 143 subjects were reported to have experienced a stroke within six months post-
implant, of which two were severely disabling with a mRS ≥ 4 at three months post-stroke.
Functional assessment of stroke was specied in the protocol; degree of follow-ups is shown in
Table 31.
9
Reference
Guides
10
Appendix
71Introduction
1
2
3
™
4
Peripherals
5
6
Surgical
7
Patient
8
Alarms and
9
Reference
10
Introduction
PumpOverview
HVAD
and
Monitor
Implant and
Accessories
Explant
Management
Emergencies
Guides
1.11 North American Clinical Study: LATERAL (continued)
Table 31: Summary of Stroke Events
Patient IDPrimary
End-point
Population
(Y/N)
102
899
103
374
103
481
103
573
103
799
103
825
103
984
105
103
105
184
105
534
105
799
106
048
106
239
106
358
106
667
107
394
107
558
108
106
108
131
108
257
* Grey indicates prior to stroke event
Primary
Success
N
YN
YY
YNDeath5
YY
YY
YN
YY
YN
YY
YN
N
YY
YN
YY
YNDeath
YY
YY
YNDeath
YY
End-
point
(Y/N)
Endpoint
Type
Stroke
with
mRS
>= 4
Alive
Original
Device
Alive
Original
Device
Alive
Original
Device
Stroke
with
mRS
>= 4
Trans-
planted
Stroke
with
Missing
mRS
Alive
Original
Device
Stroke
with
Missing
mRS
Explanted
for
Recovery
Stroke
with
mRS
>= 4
Alive
Original
Device
Alive
Original
Device
Alive
Original
Device
Alive
Original
Device
mRS at
Stroke
Event
Time to
Stroke
Event
(Months)
2.726
955
505444444
21.74
993
3.876
876
0.032
4
855
6.340
993
0.032
5
855
0.032
1
855
3.646
0
892
13.99
1
618
4.928
233
4.041
151
16.72
2
314
2.529
826
0.427
2
113
0.919
937
15.11
325
10.97
353
2.891
7.688
043
23
1 Week
Implant
Follow
up mRS
1 Month
Post
Implant
Follow
up mRS
00
4000000
00
5
120
00
1
23
1111
Post
3 Months
Post
Implant
Follow
up mRS
6 Months
Post
Implant
Follow
up mRS
12
Months
Post
Implant
Follow
up mRS
18
Months
Post
Implant
Follow
up mRS
24
Months
Post
Implant
Follow
up mRS
Appendix
72HVAD™ Instructions for Use
1.11 North American Clinical Study: LATERAL (continued)
1
INTERMACS® data collection methods for hemolysis, hypertension and right heart failure
(RHF) adverse events differ from other adverse events. Specically, events were triggered
based on data entered at each patient visit rather than site-reporting of specic events as
they occur. This method results in a potential difference in reporting of event frequency.
These events are no longer site-reported events per se.
A major hemolysis event was triggered in 11.2% (16/143) of subjects at one-week postimplant based on INTERMACS®-dened criteria, and to date this value has not changed
substantially over the follow-up period. Severe hemolysis incidence appears greater
than previously reported in HVAD studies, however, these are not site-reported events so
comparisons cannot be made.
Moderate and severe right heart failure (RHF) events were triggered in 69.9% (100/143)
and 0.7% (1/143) of subjects, respectively, at one-week post-implant based on
INTERMACS®-dened criteria. Though the incidence of severe RHF was low, the incidence
of moderate or severe RHF appears to be greater than RHF adverse events previously
reported in HVAD studies. However, once again it should be noted that these are
triggered events and not site- reported, so comparisons to previous reports cannot be
appropriately made.
No unanticipated adverse device effects were reported in the LATERAL Study.
The prevalence of device failure/malfunction in the per protocol population was 12.5%.
In a majority of these cases, no cause was identied. There were 5 cases of suspected
or conrmed pump thrombosis identied by hemolysis (2/5) and/or abnormal pump
parameters (4/5). One case was conrmed as pump thrombosis.
Table 32: Device Malfunction/Failure or Pump Thrombosis within 6 months
Adverse EventN=144 % (N)
Device Malfunction/Failure and/or
Pump Thrombus
Outcome
Death0% (0)
Serious Injury0% (0)
Urgent Transplantation0% (0)
Explant without Replacement0% (0)
Exchange2.8% (4)
Breach of Integrity of Driveline that
Required Repair
None of the Above9.7% (14)
Causative or Contributing Factors
Patient Accident0.7% (1)
Patient Non-Compliance0% (0)
Sub Therapeutic Anticoagulation0.7% (1)
Prothrombotic States0% (0)
End of Component Expected Life0% (0)
Technical/Procedural Issues2.8% (4)
No Cause Identied9.0% (13)
Thrombus (Suspected or Conrmed)3.5% (5)
12.5% (18)
0% (0)
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
73Introduction
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Monitor
6
1.11 North American Clinical Study: LATERAL (continued)
4. Subgroup Analyses
Subgroup analyses were planned for analysis of the primary endpoint analysis using
stratication factors site, on- vs. off-cardiopulmonary bypass pump, and outow graft
location. An analysis of site homogeneity found that the primary endpoint results were
signicantly different by site (P = 0.035). Further analyses found no predictive factors.
Across all populations, only one ITT subject had alternative outow location. The subject
is currently alive on original device. Similarly, only one subject had no indication in his
records regarding use of cardiopulmonary bypass, and that subject was transplanted.
Due to the sparse nature of these data, no additional analyses were performed.
Overall Conclusions from the LATERAL Trial
The LATERAL Trial was a multi-center, prospective, contemporaneous controlled study. The
purpose of this study was to evaluate the safety and effectiveness of HVAD implantation via the
thoracotomy approach. The analysis of the primary endpoint suggested success of the HVAD
System implanted via thoracotomy compared to the performance goal. The most common
reason for primary endpoint failure was death on original device, seen in 7.7% (11/143) of
subjects.
Accessories
An improvement in mean length of initial hospital stay was also observed, with a mean length of
stay of 18 ± 12.36 days as compared to the 26.1 day performance goal.
HVAD implantation via thoracotomy appears to be safe, with adverse event rates that are
comparable to previous HVAD studies, with the exception of hemolysis and right heart failure
which may be partially related to data collection methods.
Overall, HVAD implantation via the thoracotomy approach appears to be effective, with
the achievement of statistical signicance in the LATERAL primary and secondary endpoints.
Additionally, overall quality of life and functional capacity were meaningfully improved in
LATERAL subjects.
Surgical
7
Patient
8
Alarms and
9
Reference
10
Implant and
Explant
Management
Emergencies
Guides
Appendix
74HVAD™ Instructions for Use
2.0 HeartWare™ HVAD™ System Overview
1
2
2.1 HVAD™ Pump and
Surgical Tools .............75
2.2 HVAD™
Controller ....................76
2.3 HeartWare™
Monitor .......................77
2.4 HVAD™ Controller
Power Sources ...........77
2.5 HeartWare™
Battery Charger .........78
2.6 Carrying Cases and
Shower Bag ................78
System Component Overview
See Appendix A for a complete list of system components.
The primary components of the HVAD System (excluding
the Monitor) are intended for single patient use.
STERILE: All HeartWare implantable components, surgical
tools and accessories used at implant are provided
sterile.
2.1 HVAD™ Pump and Surgical Tools
The HVAD System consists of a blood pump with an
integrated, partially sintered inow cannula; a 10 mm
diameter gel impregnated polyester outow graft, and
a percutaneous driveline. A strain relief is used on the
outow graft to prevent kinking and secures the outow
graft to the pump. The driveline cable is wrapped with
woven polyester fabric to encourage tissue in-growth
at the skin exit site. The small, wearless pump has a
displaced volume of 50cc and weighs 160 grams. The
pump has one moving part, an impeller, which spins
blood to generate up to 10.0 L/min of ow. There are
two motors in the pump housing with one motor providing
redundancy.
A short integrated inow cannula is inserted into the left
ventricle and the outow graft connects the HVAD Pump
to the aorta. A sewing ring attaches to the myocardium
and allows for pump orientation adjustments
intraoperatively. The device size and short inow cannula
allow for pericardial placement, which eliminates the
need for abdominal surgery and device pockets
(Figure 39).
Figure 39: HVAD Pump and Left Ventricular (LV) Cannulation
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
For additional information about the HVAD Pump,
see Section 3.0.
9
Reference
Guides
10
Appendix
75HeartWare™ HVAD™ System Overview
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Monitor
Accessories
2.1 HVAD™ Pump and Surgical Tools (continued)
Surgical Tools and Accessories are Provided Sterile for Surgical Implantation.
Figure 40: Surgical Tools
1. Tunneler – to tunnel the pump’s percutaneous driveline
through the skin to the exit site
2. Sewing ring wrench – to tighten the screw on the
sewing ring
3. Driveline cover – to cover the driveline connection to
the controller
4. Apical coring tool – to core the LV apex
5. Strain relief wrench – to secure the strain relief and
outow graft to the HVAD Pump
All tools and accessories used during implantation are for single-use only.
Figure 41: Components used at Implant
1. HVAD Pump
2. Outow graft
3. Sewing ring
4. Driveline cap
4
1
4
2
5
Figure 40
1
6
5
2
3
6
Surgical
7
Patient
8
Alarms and
9
Reference
10
Implant and
Explant
Management
Emergencies
Guides
5. Strain relief
6. Inow cap
7. Driveline extension cable
For additional information on implantation, see Section 6.0.
3
2.2 HVAD™ Controller
The controller (Figure 42) is a microprocessor unit that
controls and manages HVAD System operation. It sends
power and operating signals to the blood pump and
collects information from the pump. The percutaneous
driveline is connected to the controller, which must always
be connected to two power sources - an AC adapter or
DC adapter and/or rechargeable batteries. The controller’s
internal, non-replaceable, rechargeable battery is used
to power an audible [No Power] alarm when both power
sources are disconnected. The controller interfaces with
the monitor through a data port.
7
Figure 41
Figure 42
Appendix
76HVAD™ Instructions for Use
2.2 HVAD™ Controller (continued)
CAUTION: ONLY use HVAD™ Controllers on one patient to avoid risks associated with an
inadvertent mismatch of controller pump speed settings.
1
2
For additional information about the HVAD™ Controller, see Section 4.2.
2.3 HeartWare™ Monitor
The monitor (Figure 43) is a touch screen tablet that uses proprietary software to display system
performance and to permit adjustment of selected controller parameters. When connected
to a controller, the monitor receives continuous data from the controller and displays real-time
and historical pump information. The monitor also displays alarm conditions and can provide
notication of available controller software updates.
Figure 43
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Explant
For additional information about the HeartWare™ Monitor, see Section 5.0.
2.4 HVAD™ Controller Power Sources
The controller requires two power sources for safe operation: either two batteries, or one
battery (Figure 44) and an AC adapter (Figure 45) or DC adapter (Figure 46). While active,
patients will typically use two batteries. While relaxing or sleeping, patients should use power from
an electrical outlet (AC adapter) because it provides power for an unlimited period of time.
Figure 44Figure 45
WARNING! ALWAYS connect an AC Adapter to the controller before relaxing or sleeping. Power from an
electrical outlet (AC Adapter) provides power for an unlimited period of time.
Figure 46
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
77HeartWare™ HVAD™ System Overview
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
2.4 HVAD™ Controller Power Sources (continued)
WARNING! NEVER disconnect both power sources (batteries and AC or DC adapter) at the same time
since this will stop the pump. At least one power source must be connected at all times.
For additional information on HeartWare™ Batteries, see Section 4.3. For additional information on
the AC/DC adapter, see Section 4.5.
2.5 HeartWare™ Battery Charger
The battery charger (Figure 47) is used to simultaneously recharge up to four batteries. It takes
approximately 5 to 6 hours to fully charge a depleted battery. Each battery slides into a bay and
is connected to the battery charger. It is safe to leave the batteries in the charger.
Accessories
6
Surgical
7
Patient
8
Alarms and
9
Reference
10
Monitor
Implant and
Explant
Management
Emergencies
Guides
Figure 47
For additional information on the battery charger, see Section 4.4.
2.6 Carrying Cases and Shower Bag
The HeartWare™ Waist Pack, HeartWare™ Shoulder Pack, and HeartWare™ Convertible Patient
Pack are used to safely secure, store and carry the controller and batteries. They can be used in
or out of the hospital, when resting, sleeping or ambulating. One controller and two batteries t
into each of the carrying cases.
The expected Useful Life of the HeartWare™ Shoulder Pack, Waist Pack, and Convertible Patient
Pack is 12 months. Carry cases should always be inspected prior to use. DO NOT use a carry
case if it shows signs of damage. Contact HeartWare for a replacement.
CAUTION: The HeartWare™ Waist Pack and the HeartWare™ Shoulder Pack contain magnetic closures.
Patients with an internal cardiac debrillator (ICD) or pacemaker should keep the pack away from their
chest, including when sleeping. Per pacemaker and ICD manufacturer guidelines, magnets should
be kept at least 6 inches (15 centimeters) away from the pacemaker or ICD (please refer to
manufacturer guidelines for additional information).
Appendix
78HVAD™ Instructions for Use
2.6 Carrying Cases and Shower Bag (continued)
1
HeartWare™ Shower Bag
A shower bag is available for use in conjunction with the HVAD System. To ensure safe and
appropriate use of the shower bag, all patients and caregivers should be trained on shower bag
operation prior to use.
WARNING! DO NOT allow patients to shower until they have received permission from their clinician
to do so. Patients who shower must use the HeartWare™ Shower Bag.
WARNING! DO NOT allow hearing impaired patients to shower unless their caregiver is close by to
hear alarms.
WARNING! DO NOT plug the controller into an AC wall outlet during showers; to eliminate the possibility
of a severe electrical shock, it should be connected to two batteries.
WARNING! DO NOT allow patients to take a bath or swim, as this may damage HVAD System
components and/or result in driveline exit site infection.
WARNING! DO NOT submerge HVAD System components in water or other uid as this may damage
them. If this happens, contact HeartWare.
WARNING! DO NOT allow water or other uids to enter the controller, power adapters, batteries,
battery charger or connectors, as this may damage HVAD System components. If this happens, contact
HeartWare.
WARNING! DO NOT use any components other than those supplied by HeartWare with the HVAD System,
as this may affect HVAD System operation.
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Explant
WARNING! Damaged equipment should be reported to HeartWare and replaced.
CAUTION: DO NOT pull, kink or twist the driveline or the power cables, as these may damage the
driveline. Special care should be taken not to twist the driveline while sitting, getting out of bed,
adjusting controller or power sources, or when using the shower bag.
CAUTION: DO NOT attempt to repair or service any components of the HVAD System. If HVAD System
equipment malfunctions, contact HeartWare.
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
79HeartWare™ HVAD™ System Overview
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Notes
Accessories
Monitor
6
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
9
Explant
Reference
10
Guides
Appendix
80HVAD™ Instructions for Use
3.0 HeartWare™ HVAD™ System Pump
1
2
3.1 Principles of
Operation ....................81
3.2 Physiologic Control
Algorithms ..................83
3.2.1 Flow Estimation .......... 83
3.2.2 [Ventricular Suction
Detection] Alarm .............. 84
3.2.3 LavareTM Cycle ............ 86
3.3 HVAD™ Pump Operating
Guidelines ...................87
3.4 Expected Useful Life
of the HVAD™ Pump ....87
3.5 Device Tracking
and Reporting
Requirements .............87
3.1 Principles of Operation
Background
The HVAD Pump is a continuous ow pump. It contains
a rotating impeller that adds energy to the blood by
converting the rotational kinetic energy into mechanical
energy (Figure 48). Impeller blades push the uid through
the pump using hydrodynamic and centrifugal forces.
The net effect is to build up the uid pressure, sometimes
referred to as pump head (i.e., related to the differential
pressure across the device) or just head, such that the
uid is moved from the inlet to the outlet of the pump.
Pump head is the difference between the afterload and
the preload. Energy to rotate the impeller is provided
through electromagnetic coupling between permanent
magnets (rotor magnet) attached or enclosed within
the impeller and the motor stators. The motor stators
consist of coils of wire that are sequentially charged by
electrical current, turning the coils into electromagnets.
These electromagnets have the effect of spinning the
rotor magnets around an axis of rotation. The HVAD
Pump is efcient at pumping moderate quantities of
blood against moderate amounts of resistance.
1
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
3
2
Figure 48: Open View
of HVAD Pump
1. Inow Cannula
2. Impeller
3. Center Post
Emergencies
9
Reference
Guides
10
Appendix
81HeartWare™ HVAD™ System Pump
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Monitor
6
3.1 Principles of Operation (continued)
Blood Flow Characteristics
The amount of ow a rotary pump can generate is dependent upon the diameter of the
impeller, the geometry of the impeller blades, housing design, motor capacity, rotational
speed, and pressure differential that exists across the pump. This allows for in-vitro pump
characterization for a specic pump and is the basis for blood ow estimation.
The HVAD System estimates blood ow rate using HVAD Pump characteristics (electrical current,
impeller speed) and blood viscosity. Viscosity is calculated from the patient’s
hematocrit. To obtain the most accurate estimate of blood ow, the patient’s hematocrit
must be entered into the HeartWare™ Monitor. Flow estimation should be used as a trending tool
only, as it cannot adapt to changing uid conditions.
For additional information on Flow Estimation, see Section 3.2.1.
Accessories
The volume of ow generated by the HVAD Pump is determined by the rotational speed of
the impeller and by the pressure differential across the pump. The pressure that the HVAD Pump
must work against is similar to the mean arterial pressure. If the pump speed (RPM) is set too
low then the device may not generate enough forward pressure. This can lead to retrograde
ow (ow from the aorta back through the device and into the left ventricle). The maximum
rotational speed is determined by how much ow is available from the right heart. If the speed
is set too high and the pump attempts to pump more blood than is available, ventricular suction
may occur.
Surgical
7
Patient
8
Alarms and
9
Reference
10
Implant and
Explant
Management
Emergencies
Guides
The controller operates in “Fixed” mode, which maintains a constant motor speed. The motor
speed range is between 1800 and 4000 RPM. The appropriate speed should be determined
based on the patient condition.
NOTE: Recommended HVAD Pump speeds are between 2400 RPM and 3200 RPM. HVAD Pump speeds
outside this range may result in less than optimal HVAD Pump operation.
Appendix
82HVAD™ Instructions for Use
3.2 Physiologic Control Algorithms
1
The HVAD Pump control algorithms provide clinicians information about device performance
and HVAD Pump blood ow estimation.
3.2.1 Flow Estimation
Estimated HVAD Pump blood ow is calculated using VAD power, speed parameters, and
hematocrit, based on a blood sample from the patient. The default hematocrit setting is 30%,
but for accurate ow estimation, the patient’s hematocrit should be entered into the monitor.
Adjustments to the hematocrit setting on the monitor should be made for hematocrit changes
of ± 5% or greater.
NOTE: Update hematocrit settings on the monitor whenever hematocrit changes by plus or minus 5%
or more.
The valid range of estimated blood ow is -2.0 to 10.0 L/min. The table below shows monitor and
Controller Display messages and what they mean.
Table 33: Monitor and Controller Display Messages
Monitor and Controller DisplayEstimated Flow Range
“--“Invalid, not available
“< -2.0 L/min“less than -2.0 L/min
“-2.0 L/min” up to “10.0 L/min”-2.0 to 10.0 L/min
“> 10.0 L/min”greater than 10.0 L/min
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Explant
The error of the estimated ow is the maximum of either 1.0 L/min or 20%, whichever is greater.
Flow estimation accuracy can be maintained only if accurate hematocrit values are entered.
Out of range values on the low side (less than -2.0 L/min), are invalid in terms of estimated ow
but could indicate an incorrect hematocrit value used in the ow or occlusion of the inow or
outow conduits. Out of range values on the high side (greater than 10.0 L/min), may occur due
to thrombus or other materials (e.g. tissue fragments) in the device, due to an incorrect
hematocrit value used in the ow or during an electrical fault.
NOTE: Flow estimation should only be used as a trending tool. Actual ow may differ from readout due
to variability of patient’s hematocrit.
WARNING! DO NOT rely only on ow estimation to assess cardiac output. An average estimated ow
on the monitor or Controller Display of less than 2.0 L/min, or greater than 10.0 L/min may indicate an
electrical fault, incorrect hematocrit entry or an occlusion and/or thrombus or other materials (e.g. tissue
fragments) in the device. Inaccurate assessment of HVAD Pump ow may lead to less than optimal
treatment.
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
83HeartWare™ HVAD™ System Pump
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Accessories
3.2 Physiologic Control Algorithms (continued)
3.2.2 [Ventricular Suction Detection] Alarm
A suction condition may occur due to ventricular collapse or inow occlusion. Ventricular
collapse occurs when a continuous ow VAD attempts to pump more blood from the left
ventricle than is available, resulting in considerable reduction in ventricular volume. Left ventricular
collapse can be the result of clinical events affecting left ventricular preload, including hypovolemia
(bleeding), right heart failure, arrhythmia or pulmonary embolus. An inow occlusion occurs when
the inow cannula is obstructed, causing a suction condition. Temporary inow obstruction can
occur as a result of surgical positioning, patient position or during straining (valsalva).
The [Ventricular Suction Detection] alarm functions by monitoring the estimated ow for sudden
decreases in ow rate. A ow baseline is established by continuously tracking the minimum ow
values. A trigger value is established at 40% below the estimated ow baseline. An indication
of suction is obtained when the minimum ow falls below this trigger level. The alarm will be
triggered if this condition is maintained for 10 seconds.
The ow minimum that triggers the suction alarm is also used to dene the suction clear limit. The
estimated ow baseline is continuously compared to this limit. The suction alarm will be cleared if
the ow baseline is maintained above the trigger level for 20 seconds. (Figure 49) This is an
indication that the suction condition has cleared.
Monitor
6
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
9
Explant
Figure 49
The [Ventricular Suction Detection] alarm can only be activated from the System Screen of the
monitor. Therefore, only the clinician has access to control the state of this alarm. The default
setting for Suction Response is “Off”. In this mode, there will be no alarm during a ventricular
suction condition. An “Sx Off” message will be displayed on the lower left-hand corner of the
monitor screen below the “Fixed” mode display. When Suction Response is enabled (via the
“Alarm” button), the “Sx On” message will be displayed on the lower left-hand corner of the
monitor screen below the “Fixed” mode display.
For additional information on the monitor, see Section 5.0.
The Suction Response “Alarm” mode must not be turned on if the patient is in a suction
condition. If the mode is turned on during a suction condition, the “Sx On” message will be
displayed on the monitor and the [Ventricular Suction Detection] alarm will be enabled but will
be inaccurate due to the fact that normal baseline parameters could not be established during
a suction condition. The algorithm attempts to establish a baseline detection level to distinguish
abnormal conditions. This is not possible if the patient is experiencing ventricular suction when
the algorithm is initiated. Once the suction condition clears, an accurate baseline will be
obtained automatically and the suction detection will proceed.
NOTE: Ventricular suction detection may be activated once the patient’s intravascular volume and
pump ow have been stabilized.
If a [Ventricular Suction Detection] alarm is triggered, the clinician should evaluate whether
the alarm was triggered by a transient, reversible condition which corrects itself, or whether the
alarm is more serious and requires intervention. Transient alarms often occur at certain times
during the day and/or under particular circumstances such as bending over or lying on one
side. They usually resolve quickly without problems. If the [Ventricular Suction Detection] alarm
is persistent and there are clinical symptoms of decreased blood ow, such as dizziness or
hypotension, or if a [Low Flow] alarm is active, then the patient should be evaluated. This can be
accomplished by checking the pump ow waveform on the monitor for evidence of suction, or
if necessary, by visualizing the left ventricle with echocardiography. The clinician should
attempt to identify and treat the underlying cause of the suction event. If the cause for the
suction event cannot be determined, or if the cause is refractory to treatment, then the
clinician should manually adjust the speed to resolve the suction condition. Manual changes
to the speed will immediately disable the [Ventricular Suction Detection] alarm. An “Sx Off” will
be displayed on the monitor screen below the “Fixed” Mode display. The clinician will have to
reactivate the alarm after adjusting the speed.
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Explant
7
Patient
Management
CAUTION: Manual changes to the speed will immediately disable the [Ventricular Suction Detection]
alarm. An “Sx Off” will be displayed on the monitor screen below the “Fixed” mode display. The
[Ventricular Suction Detection] alarm will have to be re-activated.
CAUTION: DO NOT enable the [Ventricular Suction Detection] alarm while the patient is in a suction
condition. To optimize operation of the suction detection the patient should be hemodynamically
stable prior to enabling the [Ventricular Suction Detection] alarm.
The ventricular suction detection function will temporarily deactivate if:
• The estimated ow value becomes invalid. Once the ow estimation is within valid range, the
ventricular suction detection will resume.
• The baseline ow value is less than 1.8 L/min – the algorithm loses sensitivity if the baseline and,
therefore, the suction detection level gets too low. Once the baseline value is above 1.8 L/min,
then the ventricular suction detection will resume.
• The clinician changes the hematocrit input – the algorithm recognizes that a change in the
uid viscosity will cause a change in the estimated ow. The ventricular suction detection
reactivates once a new baseline is established.
• Lavare™ Cycle is active – the Lavare™ Cycle has a direct impact on the Suction Alarm tracking
parameters, so the algorithm is temporarily disabled. The ventricular suction detection
re-activates with the previous baseline once the Lavare™ Cycle is completed.
3.2.3 Lavare™ Cycle
The Lavare™ Cycle is a speed modulation algorithm designed to reduce areas of potential blood
stasis within the left ventricle. As depicted in Figure 50, the Lavare™ Cycle decreases the pump
speed by 200 RPM below the set speed for 2 seconds then increases the pump speed to 200
RPM above the set speed for 1 second followed by the return to the original set speed for 60
seconds. Once activated the Lavare™ Cycle continues until deactivated.
6
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
Figure 50: Lavare™ Cycle Example: full cycle (left), close up of speed modulation (right)
Explant
Lavare™ Cycle is limited by pump speed range of 1800 – 4000 RPM during the low and high
speed portions of the cycles. Accounting for the ±200 RPM change with respect to the set speed
during the Lavare™ Cycle, set speeds below 2000 or above 3800 will not allow for the full ±200
RPM change. For example, if the set speed is 1900 RPM, the Lavare™ Cycle will operate between
1800 RPM and 2100 RPM, instead of 1700 RPM and 2100 RPM.
10
Appendix
86HVAD™ Instructions for Use
3.2 Physiologic Control Algorithms (continued)
1
3.2.3 Lavare™ Cycle (continued)
The Lavare™ Cycle has two settings, OFF and ON which can be set by the clinician via the
HeartWare™ Monitor. It is recommended that the Lavare™ Cycle be initiated once the patient
is hemodynamically stable and it is conrmed that the patient can tolerate the 2 seconds of
reduced support.
Hemodynamically Stable Suggests:
• HVAD Pump ow is maintained within the targeted range for each patient.
• The patient’s intravascular volume is stable requiring no serial blood product transfusions (no
active bleeding).
• Inotropic, vasoactive and anti-arrhythmic drugs are at constant dosages or being decreased.
It is recommended that the “Lavare Cycle” be deactivated if use of the Lavare™ Cycle has a
detrimental effect on the patient such as increased suction events or [Low Flow] alarms.
NOTE: If thrombus is suspected within the device, the Lavare™ Cycle should be turned “Off”
until the thrombus is resolved.
3.3 HVAD™ Pump Operating Guidelines
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
The HVAD Pump speed can be set between 1800 and 4000 RPM, however the recommended
clinical operating speed range is 2400 to 3200 RPM. Speeds below 2400 RPM should only be used
during the implant procedure when weaning from cardiopulmonary bypass. Speeds above 3200
RPM are seldom needed and may increase the risk of suction events. HVAD Pump power ranges
from 2.5 to 8.5 Watts when operating within the recommended speed range. Power values > 8.5
Watts suggests a problem which should be evaluated by log le analysis. The table below shows
the expected average values for speed, power and ow at 2400 and 3200 RPMs.
Table 34: Expected Average HVAD Pump Parameters
Speed (RPMs)Power (Watts)Flow (L/min)
24002.53.0
32008.58.0
3.4 Expected Useful Life of the HVAD™ Pump
The HVAD Pump was designed and tested to function for two years.
3.5 Device Tracking and Reporting Requirements
The HVAD Pump is considered a life-sustaining medical device and must be tracked per US Food
and Drug Administration (FDA) and other foreign regulatory agency regulations. Compliance
is mandatory. Accordingly, all device tracking paperwork must be completed and promptly
returned to HeartWare. In addition, any device malfunctions must be reported to HeartWare by
the implanting center.
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
87HeartWare™ HVAD™ System Pump
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Notes
Accessories
Monitor
6
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
9
Explant
Reference
10
Guides
Appendix
88HVAD™ Instructions for Use
4.0 HeartWare™ HVAD™ System Peripherals and
Accessories
4.1 HVAD™ Controller Connections
4.1 HVAD™ Controller
Connections ...............89
4.2 HVAD™
Controller ....................93
Controller Connector Layout
There are four ports on the controller (Figure 51)
Data Cable Connection
• Usually covered with a
dust cap
Driveline Connection
• Connects the pump
driveline to the controller
1
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
4.3 Using the HeartWare™
Batteries ....................97
4.4 Using the HeartWare™
Battery Charger .......103
4.5 Using the HVAD™ Controller
AC Adapter or
DC Adapter ................106
4.6 Carrying Cases ......... 107
4.7 Recommended Equipment
for Use at Home ........108
• Accepts the data cable
from the monitor or the
red alarm adapter with
the blue ring
Figure 51
Power Connection
• Connects the controller to the power source
• Accepts battery, AC, or DC adapter power
• Never disconnect both power sources at the
same time or the pump will stop
• Should never be removed
unless performing a
controller exchange
5
Monitor
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
CAUTION: ALWAYS keep all connectors free of liquid,
dust and dirt, or the HVAD System may not function as
intended.
9
Reference
Guides
10
Appendix
89HeartWare™ HVAD™ System Peripherals and Accessories
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Accessories
4.1 HVAD™ Controller Connections (continued)
Driveline Connection to the Controller:
To Disconnect the Driveline from the Controller:
1. Slide the driveline cover away from the controller so you can see
the whole silver connector.
2. Place your ngers on the silver connector, over the ringed area.
3. Pull back on the ringed area to release the locking mechanism.
DO NOT remove the driveline cover from the driveline.
NOTE: if you pull back on any other area of the driveline or
connector it will not release the driveline from the controller.
Figure 52
Figure 53
Figure 54
Monitor
6
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
9
WARNING! DO NOT remove the driveline cover from the driveline. Maintaining proper driveline cover
attachment prevents accidental disconnection which will lead to a pump stop.
To Connect the Driveline:
Explant
1. Line up the red dots on the driveline connector and on the
controller driveline port.
Figure 55
2. Push the driveline connector straight into the port.
NOTE: Verify the pump is running to ensure proper
connection.
Figure 56
3. Slide the driveline cover over the driveline connector.
Figure 57
Reference
10
Guides
Appendix
WARNING! DO NOT grasp the driveline cable as this may damage the driveline. To remove the driveline
from the controller, rst pull back the driveline cover then grasp and pull the driveline connector.
WARNING! DO NOT disconnect the driveline from the controller or the pump will stop. If this happens,
reconnect the driveline to the controller as soon as possible to restart the pump.
90HVAD™ Instructions for Use
4.1 HVAD™ Controller Connections (continued)
1
Connecting Power Sources and Monitor Cable to the Controller:
To Connect a Power Source:
1. To connect all power supplies (battery, AC adapter or DC
adapter) grasp the power cable near its connector. Leave the
connector free to rotate.
2. Line up the solid white arrow on the cable connector with the
dot on the controller (Figure 59).
3. Gently push the cable into the controller. DO NOT twist the
connector, but allow it to naturally lock in place. A good
connection will turn on the battery or AC/DC indicator on the
controller, as well as beep. If an alarm is active or muted, the beep
will not be heard.
Figure 58
Figure 59
Figure 60
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
NOTE: When pushing the connector into the controller the white
arrow will shift slightly into the correct locking position.
4. Conrm that the power cable is properly locked to the controller by gently pulling on the cable
near the connector. Repeat steps above for second power source.
CAUTION: DO NOT force connectors together without proper alignment. Forcing together misaligned
connectors may damage the connectors.
CAUTION: ALWAYS conrm that the power cables are properly locked on the controller by gently
pulling the cable near the controller power connector or the power cables may come loose and result
in an alarm or the pump stopping.
6
Surgical
Implant and
Explant
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
91HeartWare™ HVAD™ System Peripherals and Accessories
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Accessories
4.1 HVAD™ Controller Connections (continued)
Disconnecting Power Sources:
Disconnecting from the AC Adapter or DC Adapter
Before switching from AC or DC power to battery power, make sure that a fully charged
battery is available. Connect the fully charged battery after disconnecting the AC or DC
adapter. To disconnect power cables from the controller:
1. Turn the connector counterclockwise until it stops.
Figure 61
2. Pull the connector straight out from the controller.
NOTE: If another power source is not connected within 20 seconds,
the [Power Disconnect] message will be displayed on the
Controller Display and an alarm will sound.
Monitor
6
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
9
Figure 62
NOTE: The alarm will automatically clear when another power
source is connected to the controller.
3. Connect a fully charged battery to the controller power connector.
Explant
4. If a charged battery is not connected to the controller within 20 seconds, the [Power
Disconnect] message will be displayed on the Controller Display and an alarm will sound.
Reference
10
Guides
Appendix
92HVAD™ Instructions for Use
4.2 HVAD™ Controller
WARNING! DO NOT operate the controller in temperatures less than -20°C (-4°F) or greater than +50°C
(+122°F) or the controller may fail.
1
2
3
HVAD
Figure 63: The controller face incorporates a number of visual indicators and function buttons.
Alarm
Battery Indicator 1
Power Source 1
Alarm Mute Button
AC/DC Indicator
Controller Settings Menu
The Controller Settings Menu may be used to view controller information when a monitor is not
available.
• To access the menu press and hold for two (2) seconds and then release.
• Press again to navigate through the controller settings. After the last setting in the menu,
navigation will return to the top of the Settings Menu.
The controller will exit the settings menu automatically after 60 seconds.
Indicator
Controller Display
Battery Indicator 2
Power Source 2
Scroll Button
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Explant
7
Patient
Management
• To exit the settings menu manually, press and hold for two (2) seconds.
• Once a user reaches the end of the menu, the next press will return to the Home screen on
the Controller Display.
The Controller Settings Menu includes the following information:
1. Battery Cycles
2. [Low Flow] alarm (Setting)
3. [High Power] alarm (Setting)
4. Hematocrit (Setting)
5. RPM Setting
6. Suction Response (On/Off)
7. “Lavare Cycle”, if enabled
8. Peak Flow (L/min)
9. Trough Flow
10. Implant Date
11. Controller Date
12. VAD ID
13. Patient ID
14. Controller Software Version
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
93HeartWare™ HVAD™ System Peripherals and Accessories
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
Monitor
6
4.2 HVAD™ Controller (continued)
Table 35: Guide to Controller Display, Buttons, and Indicators
Accessories
The CONTROLLER DISPLAY gives pump information including impeller
speed (RPM), power (Watts), and blood ow (L/min). When an alarm
occurs, the pump information is replaced by two lines of text that tell
you what the alarm is and what to do. The Controller Display also provides
additional pump information in the controller settings menu.
For additional information on alarms, see Section 8.1.
The AC/DC INDICATOR will be green if you are using the AC adapter or
DC adapter to power the controller.
The two BATTERY INDICATORS located on the top of the controller are
labeled “1” and “2”. Either the “1” or “2” will be lit, depending upon which
port is providing primary power. If an AC or DC adapter is connected,
this will be the primary power source. The Battery Indicators tell you
approximately how much power remains in each battery.
Note: If the AC adapter or DC adapter is connected to the
controller, the corresponding Battery Indicator will not display
lights but the corresponding “1” or “2” will be lit.
The ALARM INDICATOR lights when one or more alarms occur. The Alarm
Indicator changes colors depending on the severity of the alarm and
always displays the most severe alarm in the case of multiple alarms.
Surgical
7
Patient
8
Alarms and
9
Reference
10
Implant and
Explant
Management
Emergencies
Guides
For additional information on alarms, see Section 8.1.
The “ALARM MUTE” button will silence (mute) a low or medium alarm for
5 minutes or until a new alarm occurs. A high alarm cannot be silenced.
Follow the instructions on the display screen, including to call your clinician
for all medium and high alarms.
The “SCROLL” button on the right side of the controller has 4 functions:
1. used to see all active alarms as well as pump information (RPM, L/min,
Watts) on the Controller Display.
2. will clear resolved medium alarms from the Controller Display.
3. will brighten the Controller Display.
4. will show additional information if pressed and held for 2 seconds.
Pressing and holding the “ALARM MUTE” button and the “SCROLL” button
for 5 seconds at the same time will prevent the [No Power] alarm from
sounding when power is removed during a controller exchange (use only
on a controller not connected to a pump).
For additional information on changing controllers, see Section 8.7.
Appendix
94HVAD™ Instructions for Use
4.2 HVAD™ Controller (continued)
1
Controller Power-up Sequence
When rst adding power to the controller the battery and alarm indicator lights will go on and then off. Both
the green and red lights will be turned on and then off. Although the red alarm indicator will turn on for 2.5
seconds, this is normal and does not mean there is a problem with the system. The power-up sequence is
complete when the controller screen shows the pump information.
Electrostatic Discharge (ESD)
Static electricity is widely present and more so in certain conditions such as in drier environments and in the
vicinity of certain materials and fabrics such as silk clothing and carpeting. Discharge of static electricity,
commonly referred to as electrostatic discharge (ESD), may interfere with electronic equipment. The
HVAD™ Controller, as a piece of electronic equipment, is
susceptible to ESD. Be aware of ESD and its potential to cause disruptive and possibly fatal faults in
susceptible patients.
The controller may alarm in certain situations as a result of ESD. These alarms include:
1. A [Controller Failed] alarm.
2. A high audible alarm without accompanying alarm text on the Controller Display.
• If either of those alarms occur: The controller should be switched to the back-up controller.
3. A [Controller Fault] alarm.
• If that alarm occurs: It should be treated as directed in Section 8.3 (“Medium Priority Alarms”), since
there are a number of potential causes for this alarm.
WARNING! AVOID devices and conditions that may induce strong static discharges (e.g., television or
computer monitor screens) as electrostatic discharges can damage the electrical parts of the system
and cause the LVAD to perform improperly or stop.
WARNING! The HVAD System components should not be used adjacent to or stacked with equipment
other than specied in the IFU. If adjacent to or stacked use is necessary, the HVAD System and other
equipment should be observed to verify normal operation.
2
3
HVAD
™
PumpOverviewIntroduction
4
Peripherals
and
Accessories
5
Monitor
6
Surgical
Implant and
Explant
WARNING! ALWAYS have a back-up controller handy and, whenever possible, a caregiver nearby when
changing power sources or controllers. Be watchful for unusual changes in power or ow alarms for a
period of time following equipment changes.
In order to avoid or minimize the potential for ESD occurrence, follow good power/battery
connection techniques as described in the IFU and patient manual.
To reduce the chance of ESD damage to the controller instruct patients to:
1. Make good connections when changing power sources
• Do not touch the controller connector ports, or let foreign objects or materials come near a
disconnected controller power port.
• Have new battery within reach before disconnecting power source and when possible, have a
caregiver nearby in case an alarm occurs.
• Use 2 power sources. Only leave power source ports on controller open for the time it takes to change
the power sources.
2. Be careful near materials and electronic devices prone to static electricity, such as: carpeted oors, silk
clothing, TV screens, microwaves when in operation, and laptop or computer screens.
• Avoid changing power sources in these areas.
• Avoid vacuuming and removing clothes from the dryer.
• Use anti-static dryer sheets and fabric softener.
• Consider humidier in your house.
7
Patient
Management
8
Alarms and
Emergencies
9
Reference
Guides
10
Appendix
95HeartWare™ HVAD™ System Peripherals and Accessories
1
Introduction
2
3
PumpOverview
™
HVAD
4
Peripherals
and
5
4.2 HVAD™ Controller (continued)
In patients who may be at risk of catastrophic cardiovascular collapse associated with a pump
shutdown (fused aortic valve, aortic valve that has been sewn shut due to aortic valve regurgitation,
or patients with very poor endogenous ventricular function) ESD education is extremely important
and controller exchanges should be performed in a controlled clinical setting whenever possible.
Controller Care
Once a week: Instruct the patient to inspect the controller power connections and connector
pins for dirt. This inspection can be done while the patient is changing batteries or when
changing from batteries to the AC adapter. Check the power connections on the controller
one at a time. DO NOT disconnect both power sources to examine the connections. DO NOT
disconnect the pump to examine the percutaneous lead/controller connection. This connector
should be inspected only during a controller exchange. The patient should not attempt to clean
the controller connectors, but should be instructed to contact their VAD coordinator if they
notice the connectors are dirty. Exterior surfaces of the controller should be cleaned using a
clean cloth. A damp cloth may be used but a wet cloth should not.
Accessories
Periodically or as needed:
The controller may be cleaned with the following agents:
Monitor
6
Surgical
Implant and
7
Patient
Management
8
Alarms and
Emergencies
• Alcohol (Isopropyl 90% or Ethyl 70%)
• Hydrogen peroxide solution (1.4%)
• n-Alkyl Dimethyl Dibenzyl Ammonium Chloride combined with n-Alkyl Dimethyl Ethybenzyl
Ammonium Chloride (active agent in some disinfecting wipes).
• Silver, ethanol, acetic acid, peroxyacetic acid, hydrogen peroxide solution (active ingredients
of some sporicidal disinfecting wipes)
• UV-C disinfecting wand, one that radiates “short wave” UV-C band rays (100 to 280
Explant
nanometers) from a 4 watt bulb (or stronger)
Expected Useful Life of the HVAD™ Controller
The HVAD™ Controller was designed and tested to function for two years.
WARNING! DO NOT drop the controller or other equipment. Dropping the controller could cause sudden
stoppage of the pump. Dropped equipment should be reported to HeartWare and inspected.
WARNING! DO NOT disconnect the driveline or power sources from the controller while cleaning it or the
pump will stop. If this happens, reconnect the driveline to the controller as soon as possible to restart the pump.
9
Reference
10
Guides
Appendix
96HVAD™ Instructions for Use
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.