Medtronic 126 Errata

Errata Sheet for the Longevity Projection Update SW, version 1.1, for the Medtronic and the Medtronic
The following table lists certain conditions that may occur during use of the following software applications: Model 9879 v.36, Model 9849 v.36, Model 9847 v.36, and Model 9980 v.36 and the resolution for these conditions.
Condition Observed Behavior Suggested Corrective Action
Emergency message
System Error When a device is at ERI or POR condition, if the
When you select [Emergency] at the same time that the device is gathering data for a Freeze or Measure ECG operation, the Emergency programmed message is not displayed. The emergency values are sucessfully programmed.
This occurs in the Thera/Thera i/Prodigy, Legacy II/Visa, and InSync 8040 applications.
user attempts to navigate to a new screen, while the ERI/POR pop-up window is appearing, the selected page menu button at the top of the screen may be disabled.
This occurs in the Thera/Thera i/Prodigy, Legacy II/Visa, and InSync 8040 applications on the 9790 programmer only.
®
9790C Programmer.
®
2090 CareLink Programmer
Navigate to the Parameter screen to confirm emergency values.
Exit application and re-enter. Do not attempt navigation until interrogation is complete.
Caution: Federal Law (USA) restricts this device to sale by or on the order of a physician.
World Headquarters
Medtronic, Inc. 710 Medtronic Parkway Minneapolis, MN 55432-5604 USA www.medtronic.com/manuals Tel. +1-763-514-4000 Fax +1-763-514-4879
*M926738A001*
Medtronic USA, Inc.
Toll-free in the USA (24-hour technical consultation for physicians and medical professionals) Bradycardia: 1-800-505-4636 Tachycardia: 1-800-723-4636 Heart Failure: 1-800-723-4636
© Medtronic, Inc. 2006
M926738A001A 2006-08-10