Software Errata Sheet for the Thera/Thera i/Prodigy/ Legacy II/Visa, and InSync 8040 software for the Medtronic 2090 CareLink Programmer:
The following table lists certain conditions that may occur during use of the following software: Model 9891, Model 9849, and Model 9980. The table also provides the resolution for these conditions.
Condition Observed Behavior Suggested Corrective Action
Emergency Message When you select [Emergency] at the same
time that the device is gathering data for a Freeze or Measure ECG operation, the Emergency programmed message is not displayed. The emergency values are successfully programmed.
This occurs in the Thera/Thera i/Prodigy, Legacy II/Visa, and InSync 8040 applications.
The following are trademarks of Medtronic: Medtronic, Medtronic CareLink, Thera, Thera i Series, Legacy, InSync, Prodigy, Vitatron
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
Navigate to the Parameter screen to confirm emergency values.
© Medtronic, Inc. 2008 M931855A001B 2008-10-02