Software Errata Sheet for the Thera/Thera i/Prodigy/
Legacy II/Visa, and InSync 8040 software for the Medtronic
2090 CareLink Programmer:
The following table lists certain conditions that may occur during use of the following software:
Model 9891, Model 9849, and Model 9980. The table also provides the resolution for these
conditions.
Condition Observed Behavior Suggested Corrective Action
Emergency Message When you select [Emergency] at the same
time that the device is gathering data for a
Freeze or Measure ECG operation, the
Emergency programmed message is not
displayed. The emergency values are
successfully programmed.
This occurs in the Thera/Thera i/Prodigy,
Legacy II/Visa, and InSync 8040
applications.
The following are trademarks of Medtronic: Medtronic, Medtronic CareLink, Thera,
Thera i Series, Legacy, InSync, Prodigy, Vitatron
Caution: Federal law (USA) restricts this device to sale by or on the order of a physician.
Navigate to the Parameter screen
to confirm emergency values.