MEDTOX SURE-SCREEN Quick Reference Manual

Page 1
SURE-SCREEN
®
Quick Reference Guide
The device detects one or more of the following drug classes at the following concentrations:
Amphetamine (AMP)
300 ng/ml
Barbiturates (BAR)
200 ng/ml
Buprenorphine (BUP)
10 ng/ml
Benzodiazepines (BZO)
200 ng/ml
Cocaine (COC)
100 ng/ml
Methadone (MTD)
200 ng/ml
Methamphetamine (mAMP)
1000 ng/ml
Opiates (OPI)
100 ng/ml
Oxycodone (OXY)
100 ng/ml
Phencyclidine (PCP)
25 ng/ml
Propoxyphene (PPX)
300 ng/ml
Cannabinoids (THC)
40 ng/ml
Note: If LFAS adulterant strip is present (to the right of drug test strips), refer to the Color Comparator Chart to interpret results.
Obtain urine. Urine level must be between the Min / Max lines on cup. If urine was refrigerated, bring to room temp and mix well before testing.
Screw lid onto cup --
TIGHTEN FIRMLY.
Tear open pouch and remove device. Align device and cup as shown.
Place the device securely on to the lid---the parts should interlock easily and move freely clockwise.
Rotate the device clockwise ¼ of a turn (i.e. from 9 o’clock to 12 o’clock) until it snaps into place.
Place cup on its side as shown above to start urine flow.
For examples at right:
C = Control T1 Tests = AMP, THC, BZO T2 Tests = COC, OPI
VALID:
A reddish-purple line appears in the “C” area for each test.
Example Sufficient URINE Present
--about ¼ inch evenly distributed-- is all that is needed for the device to migrate properly. More urine is NOT necessary! Allow test device to rest undisturbed for 5 minutes.
INVALID:
No control line “C” present. Retest urine with a new device.
NEGATIVE:
Any reddish-purple line visible at 5 min. in the test area is a negative result (even a faint or broken line).
Example above: all 5 test results are negative.
PRELIMINARY POSITIVE:
The absence of a line in the test area is a preliminary positive result.
Example above: the AMP (Amphetamine) test is a preliminary positive.
Seal Properly if Transporting to Laboratory for confirmation:
Tighten the lid  Use dual seal provided
with cup
o Seal sample port o Seal vent hole
For complete product information refer to the PACKAGE
INSERT available at http://www.medtox.com/ProductTraining.aspx
or contact Technical Support at 1-877-643-5703.
SURE-SCREEN
®
12-panel Forensic Use Only Package
INSERT Rev. 6/10, P/N 102109.
MEDTOX Diagnostics, Inc. • 1238 Anthony Road
Burlington, NC 27215
Phone: 1-877-457-0613 • Fax: 651-286-6222
E-mail: sales@medtox.com
READING THE TEST RESULTS Read Time: 5 minutes
C = Control Line T1 = Test Line for drug
name on top portion
T2 = Test Line for drug
name below (if no print on lower color, then no T2 drug tested)
Bring pouched device to room temp (18-25°C / 64­77°F) before opening.
For Forensic Use Only
Instructions
Strips with two tests will be labeled with two colors. The top color will indicate the T1 test with drug test name=T1. The bottom color will indicate the T2 test with test name=T2. Strips with only one color will have test results appear at the T1 position only.
© 2010 MEDTOX Diagnostics, Inc. All rights reserved. P/N: 102113 Rev. 8/10
Loading...