M-series Quick Reference Guide, art nr 1504436
4
C. Manual Daily Performance Verification
1. Perform a background check
It is recommended to run a background check at the beginning
of each day on the different analysis modes that will be used.
From the main screen press [NEW SAMPLE].
Press [NEXT PROFILE] or [PREV PROFILE] to scroll to
BACKGROUND.
Press the Start plate. The aspiration time is approximately 10
seconds. After ~ 10 seconds the analyzer will time out due to no
detection of blood, and continue its cycle.
Check that the background is not higher than the figures shown
below. Rerun sample if values are not acceptable.
Parameters Values accepted
RBC ≤ 0.01 (1012/ L)
WBC ≤ 0.2 (109/ L)
HGB ≤ 0.2 (g/ dL)
PLT ≤ 10 (109/ L)
2. Perform a quality control check
Good laboratory practice indicates that the performance of the
Medonic M-Series system is checked daily with certified blood
controls authorized by Boule. Comparing the analyzer results to
the known values on the Boule control assay sheet is a good
assurance that the system is functioning properly.
M-series Quick Reference Guide, art nr 1504436
13
D. Clot Prevention (cont’d)
From Main Menu press [ADVANCED], then
[MAINTENANCE] and then press [CLOT PREVENTION].
o For Cap Piercer: Place filled cleaner tube into cap piercer, same
as a normal sample analysis, close the door and continue to the
next step.
o For Sampling Device: Place filled cleaner tube into Position 1 on
wheel, lock wheel into place, and continue to the next step.
Hold the container (with cleaner) under the OT needle,
submerged in cleaner, press [OK] to confirm. Do not remove
container (with cleaner) for at least 5 seconds after aspiration
has stopped. (This is important as Cap Piercer and Sampling
Devices will take a few extra seconds to perform aspiration
before the OT begins to aspirate.)
The system will then perform the cleaning process for all
analysis modes simultaneously, and upon completion
instrument is ready for next analysis.
Perform a background check to verify all values are within
range. See Section C for more details.
E. Six (6) Month Cleaning Procedure
To increase the life of the analyzer’s internal tubing, the following
cleaning procedure is strongly recommended.
Press [ADVANCED] from Main menu, then press
[MAINTENANCE], and then press [CLEANING MENU] to
enter the Cleaning Menu.
Follow the instruction for the Boule Cleaning kit to clean the
analyzer. (Instructions for use are supplied with the Boule
Cleaning kit solutions).
The Boule Cleaning kit contains: Hypochlorite (2%),
Enzymatic Cleaner, and Detergent Cleaner
The Annual Cleaning procedure takes approximately one hour
and 15 minutes to complete.
F. LCD Display
When necessary, gently clean the display with a soft cloth, slightly
moistened with water and a mild soap. Dry carefully.
G. Maintenance
The maintenance should be performed every year or 20,000
samples by local distributor or authorized service technician.