Weekly Maintenance
2. Inspect CER reprocessor and hook-up components for wear and tear
3. Lubricate O-rings with silicone oil
Monthly (as specied)
1. 1.0 micron water pre-lter (replace every 3 months or if below 40 PSI)
2. 0.2 micron absolute lter (replace every 6 months or if below 40 PSI) and perform waterline disinfection
3. Active Vapor Management System (if applicable): Charcoal lter (replace every 6 months)
MAINTENANCE
TROUBLESHOOTING
Troubleshooting
1. Review troubleshooting section in CER User Manual
2. Service Contract Coverage and Loaner Repair Program
Run the Reprocessing Cycle
1. Close CER lid and secure using lid locks
2. Cycle Selection: Automatic FULL or LCG/RINSE
3. Conrm Time Settings: Wash and LCG
4. Select START to begin cycle
5. Basin thermometer HLD temperature verication
Complete the Reprocessing Cycle
1. Verify STOP LED now illuminated
2. Alcohol Flushing Port w/air on (Manual, air/10, Start)
3. Record Endoscope Cycle completion in Log
4. Disconnect hook up and remove and store reprocessed endoscope(s) and hook up(s)
End of Day Shutdown
1. Water Supply Valve: End of Day “OFF”
2. Clean basin, lid and gaskets (Use lint-free cloth)
SYSTEM OPERATION
Medivators Website “Resource Center”
Go to: www.medivators.com, Select “Resource Center” and “User Library” for detailed user guides and hook up matrices, report forms and logs, and product bulletins
Medivators Customer Service Support Toll Free: Phone: 1-800-444-4729 · FAX: 1-866-421-7696
FOR FACILITY USE ONLY
Staff Name:
Training Date:
Trainer:
Endoscope Reprocessing System
IN-SERVICE GUIDE
Page 2
Intended Use
1. CER Endoscope Reprocessor - High-level disinfection of exible immersible endoscopes
2. Operator Safety Overview - appropriate PPE: mask, gown, gloves, eye protection
Guidelines for Reliable Disinfection
1. Detergent solution - used in wash cycle, must be low-foaming, free-rinsing, neutral in pH and diluted per manufacturer’s recommendation
2. Disinfectant solution - required for high-level disinfection cycle.
3. OLYMPUS
®
, PENTAX®, FUJIFILM®, STORZ® - Endoscope cleaning and disinfection guidelines
4. SGNA, ASGE, ASTM, APIC and AORN - Endoscope cleaning and disinfection guidelines
5. SGNA Approved Reprocessing Steps - Endoscope bedside pre-cleaning, leak testing, manual cleaning, high-level disinfection, nal drying and storage standards and guidelines
INTENDED USE AND GUIDELINES
Endoscope Reprocessor Components and their Functions
1. CER Front Panel: Control panel (keypads and LED indicators), cycle counter and alcohol ushing port
2. CER Basin: Lid and lid locks, channel connectors / endoscope hookup, uid level sensor, drain screen and cover, and basin thermometer (if applicable)
3. CER Side Panel: AC electrical connection, On/Off power switch
4. CER Rear Panel: Water Inlet, Drain, Solution In and Solution Out connections
5. Disinfectant Reservoir: Tubing connections, AC power, temperature control and display
6. Water Filtration System: 1.0 micron pre-lter and 0.2 micron absolute lter
7. CER Cart (if applicable)
8. Transfer Pump (optional)
9. Active Vapor Management System (if applicable)
Fluids
1. High-Level Disinfectant (HLD) - RAPICIDE® HLD, OPA or Glutaraldehyde
2. Detergent (optional) - recommended detergent is MEDIVATORS INTERCEPT
®
Detergent
3. Alcohol (70% isopropyl) - used to dry inside of endoscope channels
4. Test Strips - used to test the HLD prior to each cycle
PRODUCT FAMILIARIZATION
High-Level Disinfectant (HLD)
1. Amount of HLD disinfectant required: 4 gallons for CER-1, 5 gallons for CER-2
2. Loading disinfectant reservoir using Transfer Pump
3. Unloading disinfectant reservoir using Transfer Pump
4. Loading disinfectant reservoir using manual method
5. Unloading disinfectant reservoir using manual method
6. Setting/Adjusting disinfectant reservoir temperature
7. Disinfectant neutralization and disposal (optional)
8. Cleaning the disinfectant reservoir
FLUID LOADING AND UNLOADING
OPERATOR CONTROLS
1. Automatic Operaton: FULL Cycle
2. Automatic Operation: LCG/Rinse Cycle
3. LCG Time
4. Air: 10 minutes
5. Manual Operation: LCG In
6. Manual Operation: LCG Out
7. Manual Operation: Drain
8. LED’s Blinking
9. Start
10. Stop
11. Alarm Reset: LCG and Water Level Verication
12. Alcohol Flush
13. Cycle Counter
Start-up
1. Verify power is on
Daily Service
1. Perform Daily Quality Assurance Procedure
2. Water Supply Valve: Start of day “ON”
3. Water Flow Verication: 100cc’s (4 oz) in less than 15 seconds through each hook up port
4. Air Flow Verication: Conrm vigorous bubbling from each hook up port
5. Verify disinfectant level in reservoir
6. Check disinfectant reservoir temperature (2°C higher in reservoir for correct basin temp)
7. Check HLD expiration date, temperature and potency level with test strips
8. Check drain screen and cover for debris
9. Inspect hook ups for wear
Prepare Endoscope for Reprocessing
1. Place endoscope control section in lower right with light guide in the back left rear
2. Place accessory bag into basin away from drain
3. Locate proper hook up and connect hook up to endoscope ports and basin connection
Endoscope Reprocessing Cycle Summary
1. Wash Phase (optional)
2. Rinse Phase
3. Disinfection Phase
4. Rinse Phase (number of rinses depends on HLD used. Refer to your HLDs Instructions-For-Use)
5. Air Purge Phase
6. Manual Alcohol Flush Phase
SYSTEM OPERATION