(With optional cleaning cycle that eliminates manual cleaning)
1
CLEVO® is a registered trademark of Clevo Corporation
EPSON
®
is a registered trademark of Seiko Epson Kabushiki Kaisha
FUJIFILM™ is a trademark of Fujilm Corporation
HEPA
JUN-AIR
MEDIVATORS
ADVANTAGE PLUS
INTERCEPT
OLYMPUS
PENTAX
WINDOWS
XION
®
is a registered trademark of HEPA Corporation
®
is a registered trademark of Gast Manufacturing, Inc.
®
is a registered trademark of Medivators Inc.
®
®
is a registered trademark of Medivators Inc.
®
is a registered trademark of Olympus Corporation.
®
is a registered trademark of Hoya Corporation.
®
®
and WINDOWS EXPLORER® are registered trademarks of MICROSOFT Corporation.
is a registered trademark of XION Medical GmbH.
and RAPICIDE® are registered trademarks of Medivators Inc.
All rights reserved. This publication is protected by copyright. Copying, disclosure to others, or the use of
this publication is prohibited without the express written consent of Medivators Inc.
Medivators reserves the right to make changes in the specications shown herein without notice or
obligation. Contact your Medivators representative or Medivators customer service for more information.
2
TABLE OF CONTENTS
CHAPTER 1
Using this manual ...................................................................................................................5
APPENDIX A .......................................................................................................................171
Registration Forms
APPENDIX B .......................................................................................................................179
Replacement Part Numbers
APPENDIX C .......................................................................................................................181
Warranty
4
CHAPTER 1
INTRODUCTION
This chapter describes the MEDIVATORS® ADVANTAGE PLUS® Automated Endoscope Reprocessor.
Using This Manual
This manual describes the features of the MEDIVATORS ADVANTAGE PLUS endoscope reprocessor,
including the hardware, software, setup, operation, safety, maintenance, and troubleshooting procedures. It
is important to follow the instructions provided to keep the MEDIVATORS ADVANTAGE PLUS System
in proper operating order and to ensure that endoscopes receive adequate disinfection.
This is not a service manual and does not provide detailed instructions for service beyond general
maintenance. See Service Manual for servicing instructions. Contact your service representative for
additional service information.
Medivators ADVANTAGE PLUS Features
The MEDIVATORS ADVANTAGE PLUS System was designed for ease of use and maintenance in the
disinfection of endoscopes. Features include:
• Individual channel connectivity and channel blockage monitoring.
• Single use germicide: fresh disinfectant solution for each endoscope.
• Asynchronous operation of dual reprocessing basins with transparent glass lids.
• A dedicated personal computer (PC) for system operation of reprocessing cycle, cycle recording,
powerful quality assurance reporting, easy backups, networking availability, and remote diagnostics
• Easy-to-ll detergent and alcohol reservoirs
• Hookup blocks dedicated to a family of endoscopes ensure correct connectivity and ow rates to meet
manufacturers’ specications
• Diagnostics for operator assistance and troubleshooting
• Cleaning phase validated to eliminate manual cleaning of endoscopes prior to reprocessing (endoscopes
must be pre-cleaned immediately after procedures).
5
Safety
Intended Use
The MEDIVATORS
cleaning and high level disinfection of endoscopes and endoscope accessories.
®
ADVANTAGE PLUS® Automated Endoscope Reprocessor is intended for the
The MEDIVATORS ADVANTAGE PLUS System has been validated by MEDIVATORS for use only
with RAPICIDE
®
PA High-Level Disinfectant or INTERCEPT® Detergent.
Indications for Use of the ADVANTAGE PLUS System:
MEDIVATORS ADVANTAGE PLUS Endoscope Reprocessing System tests, cleans, disinfects and rinses
endoscopes, such as beroptic and video endoscopes between patient uses. The ADVANTAGE PLUS
system is indicated to provide high level disinfection of heat sensitive semi-critical endoscopes and related
accessories. Manual cleaning of endoscopes is not required prior to placement in the ADVANTAGE PLUS
system. The endoscopes must be pre-cleaned immediately after use.
The ADVANTAGE PLUS Endoscope Reprocessing System uses RAPICIDE PA High Level Disinfectant
to provide high level disinfection of endoscopes when used according to the directions for use. The system
uses INTERCEPT Detergent in its cleaning cycle at a concentration of 0.5%.
RAPICIDE PA Test Strips are used after the disinfection cycle to ensure that the used disinfectant is at or
above the minimum recommended concentration (MRC) of 850 ppm peracetic acid; this ensures that the
disinfectant was above MRC during the entire disinfection cycle. RAPICIDE PA should be used in the
ADVANTAGE PLUS System under the following contact conditions*:
Contact Time: 5 minutes
Temperature: 30°C
MRC: 850 ppm
Use-conditions may vary by country. Please contact your MEDIVATORS representative for information.
Caution: When using the cleaning cycle, endoscopes must be precleaned according to SGNA and facility
guidelines. Manual cleaning of the endoscopes is not required. It is recommended that reprocessing
using the cleaning cycle be performed within one hour after the procedure. If the cleaning cycle is not
chosen, endoscopes must be manually cleaned before placing in the Reprocessor for disinfection.
6
Regulatory Compliance
The MEDIVATORS
®
ADVANTAGE PLUS® Automated Endoscope Reprocessor complies with EN
The reprocessor is ETL and cETL certied and listed for safety per IEC Standard IEC 61010-1, UL
Standard UL 61010-1 and Canadian Standard CAN/CSA-C22.2 No. 61010-1 for laboratory equipment.
Notes, Cautions, and Warnings
Throughout the manual are notes, service notes, cautions, and warnings. These provide important
information and you should pay special attention to them. Examples are shown below.
Note: A note refers to important information to which you should pay special attention.
Service: A service note refers to operations or repairs that only a trained service technician may
perform.
Caution A caution describes actions or conditions that may cause damage to or destruction of the
equipment.
WARNING A warning describes actions and conditions that may cause severe personal injury
or death to the operator or patient.
Operator Safety
• Equipment must be operated solely by qualied personnel, and used for its intended use only. Never
use the reprocessor for any purpose other than the manufacturer’s specic intended purpose.
WARNING: To avoid biological contamination and chemical burns, always wear appropriate
personal protective equipment (PPE) when handling endoscopes or disinfectant solutions.
• Cleaning and disinfectant solutions must be used according to the regulations governing their use,
safety, and shelf life.
• Do not attempt to open the reprocessor lid during operation.
• Do not place heavy objects on the glass covers.
7
• For peracetic acid (PAA) disinfectant handling guidelines, refer to the American National Standard
recommended practice titled, Chemical Sterilants and High Level Disinfectants: A Guide to Selection
and Use (AAMI TIR7:1999) and/or Safe Handling and Biological Decontamination of Reusable
Medical Devices in Health Care Facilities and in Nonclinical Settings (AAMI/ANSI ST35:2003). The
documents are available from the Association for the Advancement of Medical Instrumentation.
• In the event of a disinfectant or other chemical leak, use personal protective equipment (clothing, gloves
& safety glasses) & follow the disinfectant manufacturer’s guidelines for spill clean-up procedures.
• Place guidelines for disinfectant clean-up in an easily visible area near the reprocessor to ensure users
will have access to the information when needed.
• It is recommended that the facility conduct regular training of all personnel concerned with the
operation and maintenance of this equipment, including emergency procedures for toxic, ammable
or pathogenic material released into the environment. Records of attendance at training should be
maintained and evidence of understanding demonstrated.
Handling and storage of 70% Isopropyl Alcohol.
Handling
Avoid prolonged or repeated contact with skin. Extinguish any naked ames. Remove ignition sources.
Avoid sparks. Do not smoke. Take precautionary measures against static discharges. Earth all equipment.
Do not empty into drains.
Handling Temperatures
Ambient.
Storage
Keep away from direct sunlight and other sources of heat or ignition. Do not smoke in storage areas. Keep
container tightly closed and in a well-ventilated place.
Storage Temperatures
Ambient.
Product Transfer
Take precautionary measures against static discharges. Earth all equipment. Avoid splash lling.
Respiratory Protection
No specic measures.
Eye Protection
Monogoggles.
Body Protection
Standard issue work clothes. Safety shoes or boots – chemical resistant.
8
Installation and Maintenance
Proper maintenance ensures effective disinfection and prolongs the life of the reprocessor.
• The MEDIVATORS
in accordance with applicable regulations.
• The system default is factory-set for the specic disinfectant immersion time for RAPICIDE
• All pressure regulators are factory-preset. Do not adjust the settings. Contact your Technical Support
representative for assistance.
• Do not allow the washing solution to contact metal components.
• Do not use alcohol or alcohol-based products to clean the reprocessor cabinet.
• MEDIVATOR endoscope hookups are not autoclavable. The hookups are automatically disinfected
with each use due to their exposure to the entire disinfection cycle; no further disinfection or rinsing
is required prior to their next use.
• Do not remove covers, doors, or panels fastened with screws while the reprocessor is in use.
• Replacement parts must be ordered from the manufacturer to maintain the warranty.
• Do not block ventilation openings.
• Regularly check all cables, tubing, and drains for damage or leaks.
• Do not carry out any maintenance that is not described in this manual.
• Ensure that the power supply connection is accessible in the event that emergency disconnection is
necessary.
®
ADVANTAGE PLUS® Automated Endoscope Reprocessor must be grounded
®
PA.
• Prior to undertaking any service or maintenance operations or when resetting the GFI, ensure that the
MEDIVATORS ADVANTAGE PLUS System is disconnected from the main power supply. If service
or maintenance operations are to be conducted on the water system, ensure that the MEDIVATORS
ADVANTAGE PLUS System is isolated from the main water supply.
For service or service-related questions, contact MEDIVATORS Technical Support at 1-800-444-4729 in
the United States or your local MEDIVATORS Distributor.
9
Guidelines for Reliable Disinfection
Water Quality
Potable water is the minimum standard, Medivators recommends the use of a high quality water pre
ltration system that lters incoming water to a minimum of 0.1 micron. A typical water pre-ltration
system, such as the one offered by Medivators, lters incoming water through a 1 micron, 0.4 micron, and
0.1 micron series of lters.
• The high performance 0.1 micron (absolute rated) water lter included with the reprocessor is highly
effective at removing microorganisms and particles larger than 0.1 microns. Appropriate pre-ltration
and regular disinfection are required to maintain the performance of this lter.
• The routine maintenance schedule recommends replacing the 0.1 micron water lter every six months
or sooner, depending on the pre-ltration system and the quality of the incoming water. If the lter
clogs to the point of non effectiveness, the reprocessor will alarm and will not continue until the lter
is replaced.
Detergent Solution
INTERCEPT
Detergent is low-foaming and formulated to use with medical instruments. INTERCEPT Detergent has
a neutral pH and also bacteriostatic properties to inhibit bacterial growth in the detergent reservoir and
detergent lines.
Monitoring Disinfectant Potency
The efcacy of a disinfection procedure is directly related to the disinfectant solution used, and the amount
of time the endoscope is exposed to that solution.
• RAPICIDE
in this manual.
®
Detergent must be used in the cleaning phase of the reprocessing cycle. INTERCEPT
®
PA must be monitored for potency every reprocessing cycle according to the instructions
• Use the RAPICIDE PA test strips to test the potency of the solution. If the potency of the solution is
below MRC minimum recommended concentration (MRC), discard and replace it with fresh solution.
• Never use disinfectant with unacceptable potency levels.
Disinfectant Solution
Use only with RAPICIDE PA High Level Disinfectant (Part A & Part B).
10
Endoscope Cleaning and Testing
The MEDIVATORS
®
ADVANTAGE PLUS® Automated Endoscope Reprocessor may be used with all
makes and sizes of exible, immersible endoscopes that can be internally and externally disinfected.
Proper endoscope disinfection is required because endoscopes have multiple internal valves and channels
that have the potential to harbor microorganisms that could cross-contaminate patients. Reprocessing
quality depends largely on a thorough ushing of all endoscope channels, valves and ttings where
contamination can occur.
This reprocessor includes a cleaning cycle that may be used to eliminate manual cleaning of the endoscopes.
Used endoscopes must be pre-cleaned according to manufacturers, guidelines and your internal facility
procedures before reprocessing. Follow the endoscope manufacturer’s, instructions and established
professional guidelines while pre-cleaning and disinfecting endoscopes. Leak test all endoscopes prior to
disinfection.
If the cleaning cycle is not chosen, endoscopes must be manually cleaned before placing in the reprocessor
for disinfection.
Only INTERCEPT
®
detergent has been validated for use in the cleaning cycle. It is recommended that
cleaning and reprocessing take place within one hour after the procedure is completed.
11
Professional Guidelines
The following organizations have published recommended guidelines for cleaning and disinfecting
endoscopes.
Society of GastroenterologyAssoc. for Professionals in Infection
Nurses and Associates, Inc.Control and Epidemiology, Inc.
401 North Michigan Ave.1275 K Street NW, Suite 1000
Chicago, IL 60611-4267Washington, DC 20005-4006
USAUSA
Tel: 800-245-7462Tel: 202-789-1890
Fax: 312-527-6658Fax: 202-789-1899
www.sgna.orgwww.apic.org
American Society forAmerican Society for
Gastrointestinal EndoscopyTesting and Materials
1520 Kensington Rd., Suite 202100 Barr Harbor Dr.
Oak Brook, IL 60523-2141West Conshohocken, PA 19428-2959
USAUSA
Tel: 630-573-0600Tel: 610-832-9585
Fax: 630-573-0691Fax: 610-832-9555
www.asge.orgwww.astm.org
Association of periOperativeCanadian Society of Gastroenterology
Registered Nurses, Inc.Nurses & Associates
2170 S. Parker Rd., Suite 30027 Nicholson Dr.
Denver, CO 80231-5711Lakeside, NS B3T 1B3
USACANADA
Tel: 800-755-2676Fax: 902-473-4406
Fax: 303-750-3462www.csgna.com
www.aorn.org
British Society ofSociété Française d’Hygiène Hospitalière S.F.H.H.
GastroenterologyHôpital de la Croix-Rousse
3 St. Andrews PlaceUnité d’Hygiène Hospitalière-Épidémiologie
Regents Park93 Grande Rue de la Croix-Rousse
London NW1 4LB69317 Lyon Cdx04
ENGLANDFRANCE
Tel: (+44) (0)20 7387 3534Tél: 04 72 07 19 83
www.bsg.org.ukFax: 04 72 07 19 85
www.sfhh.net
12
Reprocessor Specications
General Specications
The basic reprocessor specications required for installation and operation of the MEDIVATORS
ADVANTAGE PLUS® Automated Endoscope Reprocessor are summarized in Table 1 .
Table 1 Reprocessor Specications
Equipment TypeLaboratory Equipment
Electrical rating100-240 VAC, 50/60 Hz
Power rating1000 Watts
Air pressure range87 – 145 psi (6 - 10 bar)
Water pressure range 35 - 87 psi (2 - 6 bar)
Water ow range 2 – 8 Liters/minute
Water temperature operating range30ºC – 40ºC, set mixing valve to 35ºC
Humidity range for operation20–80%, non-condensing
Temperature range for operation50°F – 80.6°F (10°C - 27°C)
EnvironmentIndoor use only
Installation/Over voltage categoryII
Pollution degree2
®
Connection to power supplyCord connection
45”W X 31” D X 62” H (inches)
Dimensions
114 W X 79 D X 157 H (cm)
Weight400 lbs. (182 kg)
13
Dimensions and Clearances
Figure 1 Front Dimensions
Figure 2 Side Dimensions & Clearance
14
Installation Specications
Electrical
The MEDIVATORS
single electrical outlet capable of supplying 1000 watts at 100-240 volts AC, 50/60 Hz. The outlet must
be properly grounded.
Water
Potable water is the minimum standard. For optimum performance, water must be delivered at a minimum
ow rate of 5 L/min (1.32 gpm) at a dynamic pressure of 30 psi (2 bar) during the basin ll. For optimum
performance water must be delivered at 35ºC by use of a mixing valve (required). The mixing valve
must be purchased separately and installed by the facility. A back siphonage prevention device must be
purchased separately and installed between the water line & the reprocessor to prevent contamination of
the water supply in the event of a sudden drop in water pressure.
Drain
For optimum performance the MEDIVATORS ADVANTAGE PLUS System should be connected to a
vented sanitary drain system capable of draining at a minimum rate of 12 L/min (3.17 gpm).
Room Ventilation
The MEDIVATORS ADVANTAGE PLUS System must be installed in a room with a ventilation system
capable of delivering a minimum of 10 air exchanges per hour.
An optional vapor management system is available for ductless fume removal.
®
ADVANTAGE PLUS® Automated Endoscope Reprocessor must be connected to a
Air
The MEDIVATORS ADVANTAGE PLUS System must be connected to an air supply for proper
operation. The ADVANTAGE PLUS system includes an air compressor that requires a dedicated 120
VAC, 20 Ampere service or 230 VAC, 10 Ampere service. The air compressor delivers proper air ows
and pressure for correct system operation.
If the ADVANTAGE PLUS model purchased did not include the air compressor, the system may utilize
house compressed air (dry, oil free) which must meet these required air pressure specications : 58 to 125
PSI with a minimum ow volume of 1.5 CFM, maximum particle size is 5 microns, maximum dew point
is 7°C (45°F), maximum oil concentration is 5 mg/m
3
.
15
Storage
If the MEDIVATORS® ADVANTAGE PLUS® Automated Endoscope Reprocessor will be out of service
for an extended period of time, the following steps must be followed:
• Remove the side covers, purge all liquids from the system, and then replace the covers.
• Place a covering over the unit to protect it from dust and moisture.
• The storage temperature must not go below 0°C (32°F), or exceed 60°C (140°F).
• The maximum relative humidity must not exceed 80% during storage.
Transport
The MEDIVATORS ADVANTAGE PLUS System can be safely moved on a smooth, level oor. However,
if you encounter any thresholds or oor level changes, the base cover must be removed to avoid damage.
For all other moves, place the MEDIVATORS ADVANTAGE PLUS System on a pallet with a kraft liner.
Before moving MEDIVATORS ADVANTAGE PLUS System, ensure the electrical cord, drain line and
water supply are either disconnected or are appropriate lengths to accommodate the relocation of the
machine.
16
Symbols and Icons
SymbolSymbol MeaningSymbolSymbol Meaning
LIO software shortcut
Management software
shortcut
Server software shortcut
Expert Operation
Cancel
Menu
Open/Close
OK/Enter
Cancel
Next
Release All
Start
Stop
Active report, lter applied.
Edit
Expand
Scroll Down
Scroll Up
Start
Stop
Import data
The icons in the grayed area are only used by Service Personnel.
Export data
Filter data
Generate report
Minimize report details
New
17
Warning Summary
WARNING Portable and mobile communication devices can affect electrically-operated medical
equipment
WARNING Connecting a chemical container with the incorrect uid may cause the endoscope to
be improperly disinfected. Do not use the endoscope on a patient if this occurs.
WARNING All endoscope hookups must be periodically inspected for wear. Discard and replace
damaged hookups otherwise it could result in an endoscope that is not disinfected and therefore
should not be used on a patient.
WARNING Verify all connections to the endoscope are secure and tubing is not linked. Failure
to do so could result in an endoscope which is not disinfected and therefore should not be used
on a patient.
WARNING Verify the connection tubing does not interfere with the spray head and conrm that
the spray function is working during the cycle. Failure to do this could result in an endoscope
that is not disinfected and therefore should not be used on a patient.
WARNING Before removing the endoscope, verify all connections to the endoscope are secure.
If an adapter is loose or disconnected, the disinfection was not complete and must be repeated.
Failure to do so could result in an endoscope which is not disinfected and therefore should not
be used on a patient.
WARNING Report all system errors to your supervisor.
WARNING Wear clean gloves to prevent soiling of the disinfected endoscope.
If a run error was generated, a channel is blocked, or the endoscope is leaking, the endoscope
must not be released for patient use. If there is any doubt as to whether an endoscope is
correctly disinfected, it must be disinfected again before use. Contact your supervisor for further
instructions.
Verify all endoscope connections are secure. Failure to do so could result in an endoscope which
is not disinfected and therefore should not be used on a patient.
WARNING Avoid possible chemical burns. Wear personal protective equipment (clothing, mask,
gloves & eye wear).
18
WARNING Use only uids approved by Medivators Inc.
Check the expiration date of the disinfectant in the reprocessor each day.
Never use disinfectant beyond the manufacturer’s expiration date, even if the MRC level is
acceptable.
Never use disinfectant that tests below the acceptable MRC level, even if it has not yet reached
the expiration date.
When replacing disinfectant, wear personal protective equipment (clothing, mask, gloves & eye
wear). The disinfectant and detergent manufacturers may recommend additional protection.
If leaking liquid is present and you suspect it is disinfectant, read and follow the disinfectant
instructions before attempting a cleanup.
WARNING Ensure that the reprocessor is installed and maintained in a position that allows
access to the power cord for emergency disconnection if necessary.
Never use the equipment in a manner inconsistent with the manufacturer’s guidelines. Using
the equipment in a manner inconsistent with the manufacturer’s guidelines may impair the
protections that are designed into the device and may cause injury to the user.
®
RAPICIDE
PA containers must be triple rinsed before disposal.
19
NOTE
20
CHAPTER 2
OPERATOR CONTROLS
This chapter describes the MEDIVATORS® ADVANTAGE PLUS® Automated Endoscope Reprocessor
operator controls, and how to set up and program the reprocessor.
21
Main Components
Exterior Components
The MEDIVATORS
of two independent reprocessing modules. Each has a basin, automatic cover, endoscope hookups, and
the necessary pumps and valves clean to disinfect, rinse, and dry most models of exible, immersible
Figure 1 MEDIVATORS ADVANTAGE PLUS System Front View
1. PC: Runs the programs on the MEDIVATORS ADVANTAGE PLUS System that disinfect the
machine and endoscopes.
2. PC Monitor: Displays messages, menu selections, and the operational status.
3. Basin Lid: Cover for the left or right basin of the MEDIVATORS ADVANTAGE PLUS System.
Open the lid to add or remove endoscopes.
4. Control Panel: Used to select settings, operate the reprocessor, and view system messages, errors
and warnings.
5. Access Door: Opens to allow access to the reservoirs, lters, transfer valves, and pressure gauges.
6. Foot Switch: Pressing the right foot switch opens the right basin lid, and pressing the left foot switch
opens the left basin lid.
7. Spray Arm Feeds: Provides the uid to the spray arms.
22
®
Figure 2 MEDIVATORS
ADVANTAGE PLUS® AER Back View
1. PC: Runs the programs on the MEDIVATORS ADVANTAGE PLUS System that disinfect the
reprocessor and endoscopes.
2. PC Monitor: Displays messages, menu selections, and the operation status.
3. Base Shroud: Covers and protects the bottom of the MEDIVATORS ADVANTAGE PLUS System.
4. Castors: Wheels which make the MEDIVATORS ADVANTAGE PLUS System easier to move.
Caster wheels can be locked.
5. Water Supply Inputs: Inlet for ltered water, one for each module (3/8” barb tting with hose
clamp, provided).
6. System Power: Electrical power supply connection.
7. Discharge: Fluids output to the drain (1” barb tting).
8. On-Board Air Compressor (Optional accessory not shown): provides pressurized air to the
system in the event that a pressurized air line from the facility is not available.
9. Compressed Air Hookup: Connects with facility pressured air line, if on-board air compressor
accessory is not purchased (1/4” NPT).
10. Circuit Breaker/Power Switch: Toggle switch turns MEDIVATORS ADVANTAGE PLUS on and
off.
11. Vent: Exhausts internal chemical vapor build-up.
23
12. Self-Disinfection Port: Connects to water line ltration system for water line disinfection.
Interior Components
Figure 3 MEDIVATORS® ADVANTAGE PLUS® System Top View
®
1. Basin Lid: Cover for the left or right basin of the MEDIVATORS
ADVANTAGE PLUS® Automated
Endoscope Reprocessor. Open the cover to load or remove endoscopes.
2. Basin: Containers where endoscopes are reprocessed.
3. Fluid Inlet: Port where liquids are dispensed into the basin.
4. Spray Head: Disperses a ne jet of uid into the basin.
5. Connection Manifold: Attaches endoscope hookup block to the system.
6. Drain: Outlet port through which waste uids are discarded.
7. Overow: Outlet used to drain off excess uids.
8. Control Panel: Used to select settings, operate the reprocessor, and view system messages, errors
and warnings.
9. Spray Arm: Provides uid spray for disinfection and rinsing of exterior surfaces of endoscopes.
24
Figure 4 Chemical Reservoirs
®
1. Main (High) Pressure Gauge: Measures air pressure in the MEDIVATORS
®
PLUS
Automated Endoscope Reprocessor. Must be set to 3 bar (45psi). Only to be adjusted by
ADVANTAGE
trained MEDIVATORS technicians.
2. Disinfectant Container: Holds disinfectant Part A used in the disinfection process.
3. Disinfectant Container: Holds disinfectant Part B used in the disinfection process.
4. Detergent Reservoir: Holds detergent used in the wash process.
5. Alcohol Reservoir: Holds alcohol that may be used in the alcohol/air purge process.
6. Leak Test (Low) Pressure Gauge: Measures air pressure in the endoscope sheath and MDS
system. Must be set to 0.25 bar (3.75 psi). Only to be adjusted by trained MEDIVATORS technicians.
®
7. Sample Ports: Dispenses RAPICIDE
PA High Level Disinfectant for MRC testing at the end of
each cycle.
25
PC Monitor and Connections
Figure 5 PC Monitor and Connections
1. On/Off Switch : Turns the computer on or off.
2. USB Ports : Connector between the PC and external peripherals, (i.e. mouse and keyboard)
3. Power Connection : Bottom of monitor in back (not shown)
4. Audio Out : Audio output connector for connection to monitor
5. Video Out DVI : DVI-I connector fro connection to monitor
6. USB : Extra USB connection for accessories (Printer or Barcode scanner)
7. Ethernet : Ethernet connection for Network or internet.
8. Main Power Switch : 10 Amp breaker with integrated main power switch.
9. Water Filter/OX Left : Low voltage output for connection to Water Filter System or left side of
Sterilox Control Box.
10. OX Right : Low voltage output for connection to right side of Sterilox Control Box.
11. AC Input : IEC 60320 C20 Inlet connector for Main input power connection.
12. AC Out : IEC 60320 C13 outlet connector for accessory power.
26
Control Panel
The control panel allows the operator to select settings, operate the reprocessor, and view system messages,
errors and warnings, and cycle information. It consists of ve main groups:
1. Left and right basin control buttons
2. Function buttons
3. Touchpad and mouse buttons
4. Selection buttons
5. Left and right basin LED lights
Figure 6 Control Panel
27
Basin Control Buttons
Each module is controlled by its own keypad grouping. The group shown in Figure 7 is for the right
module.
Figure 7 Basin Control Buttons
• Press the Open/Close button to open or close the lid on the basin. Note that during disinfection,
the lids will not open.
• Press the Start button to start a disinfection program. The button is also used to restart a program
after an error.
• Press the Cancel button whenever you have a reason to stop a disinfection program. If disinfectant
is in contact with the endoscope when the button is pressed, the program will run until conditions are
safe.
• Press the Menu
button to log in, select programs, and enter data.
• The green LED will display when system is operating normally, LED ashes green upon cycle
completion. If LED is red, system error has occurred. Verify the error condition. The endoscope may
not be disinfected. Do not use.
28
Touchpad and Mouse Buttons
Figure 8 Touchpad and Buttons
The touchpad works as a mouse to enter information. Moving your nger on the pad causes the cursor to
move on the monitor. Tapping the touchpad or pressing the left button has the same result as clicking the
left button on a mouse. The two buttons correspond to the left and right buttons on a mouse.
29
Function Buttons
Figure 9 Function Buttons
Table 1 Function Buttons
Function
Key
Press Button
(No. of times)
Result
F11I/O Inputs / Outputs,
left basin
F12I/O Inputs / Outputs,
right basin
F13I/O Inputs / Outputs,
both basins
F14Re-opens program window
F2——
F31Opens system parameters. Press again to close.
F41History report for the current day. Press again to close.
F5——
ABC1Get keyboard to appear on the monitor
ABC2Close keyboard screen
30
Help (?)——
Selection Buttons
• Pressing the OK
• The Cancel
button is the same as clicking on the button in the software.
button corresponds to the button in the software.
• Pressing the Scroll Up
• Pressing the Scroll Down
button allows you to scroll up through a list on the monitor.
button allows you to scroll down through a list on the monitor.
31
Figure 10 Selection Keys
Connector Blocks
The endoscope channels are connected to the MEDIVATORS
®
ADVANTAGE PLUS® Automated
Endoscope Reprocessor channel manifold via a connector block. Each endoscope model requires a unique
connector block to ensure that proper disinfection can be achieved. Refer to the hookup application guide
to identify the correct connector block to use for the endoscope to be reprocessed.
Note: Before a new type of endoscope is reprocessed or if you have any questions about a connection
block, contact your eld service representative.
Accessories
Endoscope accessories, buttons and valves may be disinfected in MEDIVATORS ADVANTAGE PLUS
system. Use the provided mesh bag.
Data Entry
Bar Code Reader
A bar code scanner can be used to automatically enter data rather than manually entering it with the control
panel. The system administrator may dene a list for any of the following items:
• Operator login name
• Endoscope code
• Doctor/specialist ID
• Assistant ID
• Patient number
32
Figure 11 Bar Code Reader for Operator Login and Endoscope ID
Software System
Available Programs
Three separate programs control the MEDIVATORS
Reprocessor:
1. LIO (Logic Input/Output) is the main software program that operators use to perform and run an
endoscope cycle. This part of the software controls the MEDIVATORS ADVANTAGE PLUS System
and is what runs and displays on the PC monitor. This program is used most often by daily operators
of the reprocessor.
2. Management is the software program for managing endoscopes, users and cycle information. It allows
an administrator access to input data, process reports, and can manage multiple MEDIVATORS
ADVANTAGE PLUS System reprocessors..
3. Print Service automatically prints a certicate to the cycle log printer which is dedicated to each
reprocessor. The certicate contains system information and reprocessing cycle details pertaining to a
specic reprocessing cycle.
®
ADVANTAGE PLUS® Automated Endoscope
33
Figure 12 MEDIVATORS ADVANTAGE PLUS System Programs
Re-Starting LIO
Before using the LIO software, you should have a working knowledge of PC computers and the Windows
operating system. Make sure you know how to use the control panel, including the touchpad and its
buttons. If you are uncertain how to proceed, contact your supervisor or the system administrator for
assistance. LIO and the server are initiated during normal PC power up.
1. Verify the PC is on, Windows is running.
2. Find the MEDIVATORS
®
ADVANTAGE PLUS® Automated Endoscope Reprocessor folder on the
Windows desktop and double click on it. It is titled “ADV”.
Figure 13 MEDIVATORS ADVANTAGE PLUS System Folder on the Desktop
34
®
3. Double click on the LIO icon to start the MEDIVATORS
ADVANTAGE PLUS® Automated
Endoscope Reprocessor operational program.
4. Double click the Print Server icon to start the cycle log print service.
35
Figure 14 LIO in the MEDIVATORS ADVANTAGE PLUS System Window
User Screen
The monitor displays all necessary information for operating and maintaining the MEDIVATORS
®
ADVANTAGE PLUS® Automated Endoscope Reprocessor. Use the buttons on the control panel, to access
different screens, programs and functions.
36
Figure 15 Typical Program Screen
Menu Bar
Each program has a separate release version shown in the upper left corner of the screen. If a system error
occurs, a red message bar appears here until the error is resolved.
The Expert Operation icon appears in the upper right of the menu bar. Double clicking the icon will
display a login dialog box. After logging in, a side bar is displayed on the left with the following icons.
Expert operation is only available to operators with the correct level of user rights.
Figure 16 Side Bar with icons
37
Module and Program Status
The status of each module is indicated in the main section of the screen. Program operations for the left
module are on the left side of the screen; operations for the right module are on the right side of the screen.
Running program operations for each module are indicated in the section. The information provided
includes:
• Operator name • Run time remaining
• Endoscope ID • Volume
• Program name • Temperature
• Parameter set • Program progress bar
• Number of cycles • Phase progress bar
• Program status • Step progress bar
• Expected run time • Cycle messages
Heartbeat
The red heartbeat indicator blinks on and off to show that the current program is active. Each module has
a heartbeat indicator.
Program Messages
Program messages and operator instructions are shown in the lower section of the screen. The messages
are presented with text and color. If an error occurs, an audible alarm also sounds.
The text shows the program time elapsed at which the message occurred, the message code, and a
description of the message or instructions for the operator to follow. For troubleshooting purposes, the
message codes and additional message details are listed in Chapter 5 Troubleshooting. The meaning of
each screen color is shown in Table 2:
Table 2 Screen Color and Program Status
Screen ColorProgram Status
1GrayProgram in standby
2OrangeStandby, with lid open or a warning is present
3BlueCycle in process
4RedError or Cycle Failed, endoscope not disinfected
The percentage of alcohol and detergent remaining in the reservoirs is provided at the bottom of the screen.
System Information
The bar at the bottom of the screen provides system communication information, reprocessor identication,
and the software version for your selected language.
MEDIVATORS
®
ADVANTAGE PLUS® Automated Endoscope Reprocessor software has two possible
system connections:
1. MEDIVATORS ADVANTAGE PLUS System PC to the MEDIVATORS ADVANTAGE PLUS
System (SCU).
2. MEDIVATORS ADVANTAGE PLUS System PC to a server application (may be networked)
The left eld on the bar indicates if the reprocessor is connected to a server. The next eld to the right
indicates if the PC is correctly connected to the reprocessor SCU. If the PC is disconnected the eld
appears red.
Login Name and Password
Many processes on the MEDIVATORS ADVANTAGE PLUS System require you to supply a login name
and password. This conrms that you have the proper authorization to run each program. Data gathered
from each process, including the operator’s name, is stored in the database and printed on reports.
To enter your login name, move the cursor to the window. Press the down arrow on the right side and scroll
down until your login name is highlighted. Click on the name to select it.
39
Figure 17 Login Name Selection
To enter your password, click on the appropriate numeric buttons in the window. If you enter an incorrect
password, click the CLR button and enter the password again. Press the OK button to continue.
Figure 18 Password Entry
40
Printing
Print Service Utility
1. Check for the printer icon in the system tray. The Print Server utility supports the printing of the Cycle
Logs on the certicate printer.
Figure 19 System Tray
2. If the icon is not present, display the Windows desktop by minimizing all open programs.
3. Open the MEDIVATORS
®
ADVANTAGE PLUS® Automated Endoscope Reprocessor folder by
5. The PrintService icon should now appear in the System Tray.
Set the Default Printer
1. Click on Start > Settings > Printers.
42
Figure 22 Printer Selection
2. Double-click the icon for your cycle log printer.
3. In the dialog box for the printer, double-click on Set As Default Printer. A checkmark appears
above the printer icon and in the submenu the next time it is accessed.
Figure 23 Set As Default Printer
Backups
The computer on the MEDIVATORS® ADVANTAGE PLUS® Automated Endoscope Reprocessor is
equipped with mirrored hard drives. Mirrored hard drives consist of a main hard drive and a second hard
drive which contains an exact copy of the rst hard drive. This conguration provides a backup of the
entire computer system.
The MEDIVATORS ADVANTAGE PLUS endoscope reprocessor uses MICROSOFT SQL Server 2008
Express to generate the database les.
The databases should be backed up on a regular basis. Although the mirrored hard drives do provide
protection against a single hard drive failure, regular backups of the database to a remote location (CD,
memory stick or remote server) will protect against database corruption or complete system failure. Follow
the procedures established by your facility for scheduling and storing backups.
MEDIVATORS provides a backup utility, SQLBackupAndFTP, preinstalled on the MEDIVATORS
ADVANTAGE PLUS system. This utility must be congured.
43
Fluid System
Fluid Reservoir Status
The system monitors the use of disinfectant, alcohol, and detergent. The percentage of uid remaining
appears in the footer of all screens.
Figure 24 Reservoir Levels
WARNING Relling the container with the incorrect uid may cause the endoscope to be
improperly disinfected. Do not use the endoscope on a patient if this occurs.
Wear protective clothing, gloves and eye wear. The disinfectant and detergent manufacturers
may recommend additional protection.
If the uid reservoir status indicates that the detergent or alcohol volume is low, then you must follow the
alcohol or detergent loading process to ll the appropriate uid reservoir.
Note: A program does not stop if a 0% volume level is reached during a cycle. An error message appears
and you must replace the uids.
The uid levels in the detergent and alcohol reservoirs are monitored with oat switches. A cycle cannot
be initiated if sufcient uid is not available. The replace uid utilities must be run once out of uid is
detected even though the reservoirs have been relled.
44
Detergent and Alcohol (Optional)
The detergent and alcohol reservoirs serve both the left and right modules. Each reservoir is threaded
into a captive header and is removed by unscrewing the body. The alcohol must be 70% isopropyl. The
detergent must be INTERCEPT
®
Detergent manufactured by Medivators Inc.
Figure 25 Alcohol and Detergent Reservoirs
NOTE: Upon relling of the detergent or alcohol reservoirs, the Replace Fluids Utility must be run. If a
container empty condition has been detected, the Replace Fluids utility must be run to correct the error
condition.
Because the two reservoirs are identical, you must pay attention to the labeling to ensure that the correct
uid is in each reservoir. To prevent errors, complete relling one reservoir before removing the other.
WARNING To prevent the accidental switching of the Alcohol and Detergent reservoirs always
remove, rell and replace each reservoir bottle separately. Return the reservoir (bottle) to its
original position by matching up the reservoir label to the label above the reservoir header.
WARNING Relling the container with the incorrect uid may cause the endoscope to be
improperly disinfected. Do not use the endoscope on a patient if this occurs.
Wear personal protective equipment (clothing, mask, gloves & eye wear). The disinfectant and
detergent manufacturers may recommend additional protection.
45
Disinfectant
Rapicide PA Part A and Part B
The two disinfectant containers serve both the left and right modules. One container holds Part A while
the other container holds Part B of the disinfectant solutions. The MEDIVATORS
®
ADVANTAGE PLUS®
Automated Endoscope Reprocessor automatically mixes the two parts with water at the appropriate time
during the disinfection cycle.
Note: Upon replacing RAPICIDE
®
PA Part A and B containers, the Replace Fluids utility must be run.
A reprocessing cycle cannot be initiated if an out of disinfectant has been detected. The Replace Fluids
utility must be run to clear the error condition.
Part A uses a blue cap, and is positioned to the left as you look into the front of the MEDIVATORS
ADVANTAGE PLUS System. Part B uses a white cap, and is positioned to the right as you look into the
front of the MEDIVATORS ADVANTAGE PLUS System. Part A and B must always be positioned in the
correct location in the machine, and must always be attached to the appropriately colored cap. To facilitate
the correct connection, the blue cap is larger than the white cap, and only ts the Part A container. The
smaller white cap only ts the Part B container.
Contact Time
Prior to replacing Part A or B the Replace Fluids utility must be run. Information from the containers is
required. The disinfectant contact time and disinfectant temperature is pre- set during manufacturing. The
contact time and temperature eld is not congurable.
46
Figure 26 Parts A and B
WARNING Use only cleaning and disinfectant uids approved by Medivators Inc.
Water Line and Water Filter Disinfection
The ISO 15883-4:2008 (Washers-disinfectors Part 4) water quality standard, requires that chemical
disinfection of the water treatment (ltration) and delivery system must be performed on a weekly basis
and whenever the 0.1 micron bacterial retentive lter is replaced. This standard requires that all rinse
water delivered to the washer/disinfector contain no more than 10 cfu (colony forming units) per 100 ml.
Microbiological quality of even potable water varies considerably. If you are unsure of the microbiological
quality of your washer/disinfector supply water, you should have it tested according to either local or
internal guidelines.
The Water Line Disinfection program or utility is used to disinfect both the water line and the water lter.
Activating this utility lls the water line and water lter with disinfectant, and circulates it through the
channels and the basin.
1. Ensure that the disinfectant reservoirs are lled and the water supply is turned on.
2. Open the cover by pressing the Open/Close
button.
3. Remove any endoscope(s).
4. Install a Disinfection Block on each basin connection manifold.
Figure 27 Disinfection Block
5. Press the Menu
ADVANTAGE PLUS
button for the left module on the control panel. The left side of the MEDIVATORS®
®
System controls this utility.
6. Log in to the system.
7. When the Program Selection screen opens, click on Program Selection and click the OK
button.
47
Figure 28 Program Selection
8. Select Water Line Disinfect from the Select Program screen and click the OK
button.
Figure 29 Water Line Disinfect
9. When the Program Set Selection screen opens, choose the self-disinfection cycle and then click on
the OK button:
S Disinfect 4 hours
48
Figure 30 Program Set Selection
10. Select either Normal Start or Delayed Start and click on the OK
Figure 31 Normal Start
button.
49
Figure 32 Delayed Start
11. If a delayed start is selected, a screen opens called Select the Time. It lists start times in 15 minute
increments for the next 7 days. Note that the dates are listed in MM/DD/YYYY, where M = month, D
= day, and Y = year. Scroll down to select the desired start time and click the OK
button.
Figure 33 Select Delayed Start Time
12. An informational screen is displayed to verify that the disinfection blocks are installed. Select the OK
button if the blocks are in place or select the Cancel button to install the water line disinfect
blocks.
50
Figure 34 Summary/Information Screen
13. Press the Start
button to initiate the water line disinfection. Disinfectant is ushed through the
water line and water lled. The disinfectant remains in the water lter and water lines while the system
is disinfected. The water lines and lter are ushed, the basin is rinsed and the utility terminates.
51
NOTES
52
CHAPTER 3
OPERATION
This chapter explains how to prepare and disinfect an endoscope in the MEDIVATORS® ADVANTAGE
®
PLUS
Automated Endoscope Reprocessor.
53
System Startup
1. Verify that the MEDIVATORS® ADVANTAGE PLUS® Automated Endoscope Reprocessor PC and
main power source are on.
2. Open the shutoff valve to the incoming water line. For optimum performance the water supply must
be capable of delivering a minimum of 1.32 gpm (5 L/min) at 35 psi (2 bar).
3. Turn the external air source on (if applicable). External air supply must be capable of delivering house
compressed air (dry / oil free) at a minimum rate of 10.58 G/min (40 L/min) at 87 psi (6 bar).
Daily Service
System Air Pressures
The main air pressure and leak test air pressure are maintained by regulators. The air pressure adjustment
controls and corresponding gauges are located behind the front access door, just above the disinfectant
reservoirs.
Proper operation of the MEDIVATORS ADVANTAGE PLUS System reprocessor requires that the air
pressure be set and maintained at the values shown in the table below.
54
Figure 1 Air Pressure Controls
Table 1 Acceptable Pressure Ranges
Pressure ControlAcceptable Pressure Range
Main Air Pressure42 – 45 psi (2.8–3.0 bar)
Leak Test Air Pressure3.6 – 3.75 psi (0.24–0.25 bar)
Channel Connectivity25.5 – 30 psi (1.7 – 2.0 bar)
If the pressure control readings do not fall within the acceptable ranges, contact technical service. Do
®
not attempt to adjust the settings yourself as damage may occur to the MEDIVATORS
®
PLUS
Automated Endoscope Reprocessor.
ADVANTAGE
O-Rings
Check the connection manifolds to verify that all the O-rings are present and undamaged. The manifolds
are located in each basin and are the studs onto which the connector block slides down and seals.
Drain Filters
Replace the lters in the rinse basin if they are excessively soiled (See Appendix B for replacement
part numbers).
Figure 2 Connection Manifold
55
Figure 3 Drain Filters
Fluids
WARNING: Relling a reservoir with the incorrect uid may cause the endoscope to be improperly
disinfected. Do not use the endoscope on a patient if this occurs.
Always wear appropriate personal protective equipment (PPE) when handling endoscopes or
disinfectant solutions. The detergent and disinfectant manufacturers may recommend additional
protection.
Loading Alcohol (Optional)
Alcohol is an optional feature and may not be congured in the system. If congured, the alcohol must be
70% isopropyl.
1. Press the Menu
button on the control panel.
2. Log in to the system.
3. When the Program Selection screen opens, select Replace Fluids and click the OK
button.
56
Figure 4 Program Selection Screen
4. When the Select Fluid Type screen opens, select Alcohol and click the OK
Figure 5 Select Fluid Type Screen
5. Enter the code found on the manufacturer’s container and press OK
button.
button.
Endoscope repr
57
Figure 6 Container Code
6. Remove, rell and replace the reservoir.
Loading Detergent (Optional)
Detergent is an optional feature and may not be congured in the system.
Note: The MEDIVATORS® ADVANTAGE PLUS® Automated Endoscope Reprocessor cleaning phase
®
requires the use of INTERCEPT
detergent.
58
Remove, rell and replace the detergent reservoir.
1. Press the Menu
button on the control panel. Log in to the system.
2. When the Program Selection screen opens, select Replace Fluids and click the OK
Figure 7 Program Selection Screen
button.
3. When the Select Fluid Type screen opens, select Detergent and click the OK
Figure 8 Select Fluid Type Screen
4. Enter the code found on the manufacturer’s container and press the OK
button.
button.
59
Figure 9 Container Code
5. Remove, rell and replace the reservoir.
60
Loading Disinfectant
WARNING Use only RAPICIDE
Check the expiration date of the disinfectant on the machine each day.
Never use disinfectant beyond the manufacturer’s expiration date, even if the MRC level is
acceptable.
Never use disinfectant that tests below the acceptable MRC level, even if it has not yet reached
the expiration date.
When replacing disinfectant, always wear protective clothing, gloves and eye wear. The
disinfectant manufacturer may recommend additional protection.
If leaking liquid is present and you suspect it is disinfectant, read and follow the manufacturer’s
instructions before attempting a cleanup.
Service: If leaking uids are present, contact a qualied service technician.
®
PA High Level Disinfectant.
1. Press the Menu
button on the control panel.
2. Log in to the system.
3. When the Program Selection screen opens, select Replace Fluids and click the OK button.
Figure 10 Program Selection Screen
4. When the Select Fluid Type screen opens, select Disinfectant and click OK.
5. Enter the code found on the manufacturer’s container and press OK.
6. Remove, and discard the empty containers of Part A and Part B solutions.
7. Place the new containers of Part A and Part B solutions into the machine.
61
Prepare the Endoscope
Caution: When using the cleaning cycle, endoscopes must be precleaned according to SGNA and facility
guidelines. Manual cleaning of the endoscopes is not required. It is recommended that reprocessing
using the cleaning cycle be performed within one hour after the procedure. If the cleaning cycle is not
chosen, endoscopes must be manually cleaned before placing in the Reprocessor for disinfection.
Place the Endoscope in the Basin
1. Endoscopes t best in the machine if they are installed in a particular fashion. The nal position of the
endoscope in the rinse basin depends primarily on the type of endoscope. To a lesser extent, the brand
of endoscope may also affect the nal position in the basin. Do not forcibly bend the endoscope to get
it to t. When positioned properly, the endoscope will t appropriately in the basin.
2. Hold the control housing in your left hand and the insertion tube and light guide in your right hand.
3. Place the control housing in the basin with the angulation controls facing up.
Figure 11 Place Control Housing in Basin
4. Set the insertion tube in a clockwise circle in the basin.
62
Figure 12 Set the Insertion Tube in Basin
5. Fit the light guide tube in a counterclockwise circle in the basin.
6. Select the proper connection block and channel separator from the endoscope hookup guide.
Figure 13 Fit the Light Guide Tube in Basin
Install the Channel Separator
Note: Reference the endoscope hookup application guide for the correct channel separator.
Select the correct channel separator for the endoscope to be disinfected. Position the adapter so that the
stem with the O-ring ts properly in the water/air valve.
Because the valve design may vary depending on the brand and type of endoscope, attachment of the
adapters may differ.
63
Figure 14 Fitting the Separator
Install the Connector Block
1. Hold the connector block with the handle positioned to the right in Figure 15.
Figure 15 Fitting the Connector Block
2. Place the block lightly over the manifold as shown in Figure 16 making sure that none of the hoses
run or are pinched under the block.
64
Figure 16 Press Block Over the Manifold
3. Lock the connector block in place by moving the handle to the left as shown in Figure 17.
Figure 17 Close the Handle
WARNING Inspect the seating of the connector block to the manifold. Failure to achieve a proper
connection could result in an endoscope that is not disinfected and therefore should not be used
on a patient.
65
Connection of the Endoscope Channels
1. Connect the endoscope channels. The names of the connections of the different channels may be
found in the endoscope user manual. See Figure 18 for details on how to hookup the endoscope to the
connectors.
2. Connect each tube from the labeled port on the connector block to the appropriate endoscope channel.
WARNING All endoscope hookups must be periodically inspected for damage, and replaced if
damaged, otherwise it could result in an endoscope that is not disinfected and therefore should
not be used on a patient.
Figure 18 Connecting the Channels
3. Make sure the hoses do not have any sharp bends or kinks.
4. Position the connection hoses in such a way that they will not be caught by the spray tower.
Verify the Connections
WARNING Verify all connection to the endoscope are secure. Failure to do so could result in an
endoscope which is not disinfected and therefore should not be used on a patient.
• Before closing the basin lid, inspect the hookup to ensure that all connections are made properly and
the tubing does not interfere with the sprayer.
• All channels must be connected correctly.
• The channel separator must be installed.
• The leak test tubing must be secured under the endoscope so that it does not oat or interfere with
the sprayer.
• The connector hoses must not be kinked.
• Move any connector tubes away from the head of the sprayer so that it can turn freely.
66
WARNING Verify the connection tubing does not interfere with the spray head and conrm that
the spray function is working during the cycle. Failure to do this could result in an endoscope
that is not disinfected and therefore should not be used on a patient.
Figure 19 Sprayer Must Turn Freely
67
Disinfection Process Summary
The complete reprocessing cycle in the MEDIVATORS® ADVANTAGE PLUS® Automated Endoscope
Reprocessor is comprised of a number of specic steps designed to ensure proper endoscope disinfection.
These steps include: startup, washing, disinfectant, rinse, alcohol, and air phases. All steps must be properly
followed and should not be modied beyond manufacturer recommendations.
Cycle Startup Phase
During the start-up phase, the software reads and monitors all system sensors. If an error is detected, a
message is displayed and the process is halted. The system operator then needs to resolve the error before
the cycle can be continued.
Next the leak test is activated. The endoscope is pressurized and the pressure is monitored for a period of
time. If the pressure drops below a set reading, an error message is displayed and the process is halted.
WARNING: If the machine detects a leak in the endoscope, do NOT use the endoscope on patients.
Refer to Chapter 5: Troubleshooting for appropriate instructions to resolve errors.
Cleaning Phase (Optional)
During the cleaning phase, detergent and water enter the basin. This basin uid is then ushed through
the endoscope channels. At the completion of the phase, the basin is drained and the endoscope channels
are ushed with fresh water. A rinse soak period without detergent then follows. The manual cleaning step
may be eliminated if the cleaning phase using Intercept detergent is programmed.
Disinfectant Phase
During the disinfectant phase, the endoscope channels are ushed with disinfectant. The basin is lled
with disinfectant solutions and the endoscope is soaked for the disinfectant contact time as recommended
by the disinfectant manufacturer. After the soak, the disinfectant is drained from the basin.
Rinse Phase
During the rinse phase, the basin is partially lled with water and then drained to eliminate foam residue
from inside the endoscope and from the basin. The basin is then lled with fresh water while the endoscope
channels are being ushed. Once the basin lls up, water ushes through the endoscope channels. The
basin is then drained while the channels are purged with fresh water.
Alcohol Purge Phase (Optional)
During the alcohol phase, alcohol is injected through the endoscope channels. The alcohol is used to help
purge any residual uids from the endoscope.
Air Purge Phase
The air phase is simply a programmed time during which air is purged through the endoscope channels.
The air is used to aid in drying of the endoscope.
68
Run the Disinfection Program
Normal Start
Data entry must be carried out carefully. Each selection or input must be checked and conrmed by the
operator. Entering erroneous information may lead to machine faults or inadequate disinfection of an
endoscope channel.
1. Close the basin lid by pressing the Open/Close
2. Press the Menu
button.
button or the foot switch.
3. When the Operator Selection window opens, select your login name by using the Scroll Down
button, or use the touchpad, bar code reader. Then click the OK
button.
4. If the Authorization window opens. Enter your password with the touchpad, bar code reader and then
click the OK
button (option).
5. When the Program Selection window opens, highlight Endoscope Selection and click the OK
button.
Figure 20 Program Selection
69
6. From the Endoscope Selection window, choose the instrument code or endoscope that you wish to
disinfect by using the scroll buttons or, bar code reader. Click the OK
button.
Note: The endoscope code or ID is required for the disinfection program to run. If pre-programmed
endoscope codes have been entered by your system administrator, the specic disinfection program for
that endoscope is selected automatically.
Figure 21 Endoscope Selection
7. Select the correct parameter set from the Endoscope Type Parameter Set Selection window by
scrolling up or down with the Scroll Down
button or Scroll Up button. Then click the OK
button.
70
Figure 22 Parameter Set
WARNING Ensure the correct parameter set is selected for the type of endoscope to be processed.
An endoscope is not disinfected and should not be used on a patient if the incorrect parameter
set is used during a cycle.
8. The Patient Code window opens. Enter the patient ID with the touchpad or bar code reader and click
the OK
button.
Figure 23 Patient Code
9. When the Physician Selection window opens, select the desired name by scrolling with the Scroll
Down
button, or use the bar code reader. Then click the OK button.
10. When the Physician Assistant Selection window opens, select the desired name by scrolling with the
Scroll Down button, or use the bar code reader. Then click the OK button.
®
Note: The patient code, physician code, and assistant code on your MEDIVATORS
ADVANTAGE PLUS®
Automated Endoscope Reprocessor are congurable. If your system administrator has determined that
these items do not have to be tracked, the screens will not be shown.
11. If the system is congured to automatically start the operator needs to do nothing more and the cycle
will start. If not congured the Start Program window opens. Verify that the Start data on the screen
is correct, and then click the OK
button. The process LED indicator light should turn green.
12. A system check is carried out. If a red error report appears at the bottom of the screen, attempt to
correct the error by consulting Chapter 5: Troubleshooting.
13. The Program Screen opens and allows you to follow the progress of the disinfection process.
71
Figure 24 Process Window
Delayed Start
When selecting an delayed start, the endoscope is placed in the basin but the disinfection cycle will not
begin until a pre-determined time. This is an administrative function and is available only during the time
period as set up and allowed by the system administrator.
1. Prepare and connect the endoscope as described in “Run the Disinfection Program”, steps 1-12 for
the normal start above.
2. The Select the Time window opens. Select Delayed Start and press the OKSelect the Time window does not open, the delayed start features is not available.
72
button. If the
Figure 25 Delayed Start
3. When the Select the Time window opens, scroll down to select the desired start time and date. Click
the OK
button.
Figure 26 Select the Time
4. The Delay Start program window opens. Verify that the data on the screen is correct, and then click
the Start
button.
Figure 27 Delayed Start Progress
73
5. When the designated start time is reached, a system check is carried out. If a yellow error report appears
at the bottom of the screen, attempt to correct the error by consulting Chapter 5: Troubleshooting.
6. The reprocessing cycles is initiated at the selected start time.
Cycle Interruptions
Introduction
At startup and during the disinfection program cycle, the MEDIVATORS
Automated Endoscope Reprocessor monitors process parameters for the endoscope, machine, and software
to ensure accurate disinfection of the endoscope. Reports, program errors, and system errors appear on the
screen to communicate the cycle progress to you.
Operator messages and program errors appear at the bottom of the screen in the Reports eld. A code
number and short description of the report or error is shown. The elapsed time at which the message or
error is generated is also provided.
System errors appear at the top of the screen in white letters on a red background.
Detailed error reports and troubleshooting steps are provided in Chapter 5.
WARNING If system power is lost during a reprocessing cycle, the system will not complete the
reprocessing cycle once the power is restored. The endoscope is not disinfected and should
not be used on a patient. You must start the reprocessing cycle over so that the endoscope is
properly disinfected.
®
ADVANTAGE PLUS®
74
Program Errors
Restarting a Program
In a certain instances, a program may be restarted after an error is displayed to restart the program if the
following conditions are true:
To restart the program:
1. Press the Start
button.
2. If the program restarts and progresses without any further errors, the disinfection cycle is valid.
3. If the program does not restart, or one or more errors appear, press the Cancel button. Follow
the instructions below for the steps to take next.
If a program is stopped by an error condition and lasts longer than 60 minutes, the program can not be
restarted, the LED turns red, and the program status is ASC (Reject).
Canceling a Program
If the Cancel
button is pressed, the program runs until a safe state is reached.
The alarm sounds and cancel message appears in the Report section of the screen.
Every time the Cancel button is pressed, the current program step is stopped and the next step in
the cancel sequence is called. By pressing the Cancel button multiple times, the program can be
forced to the end of the cycle. However, caution must be taken because the endoscope channels remain
lled with uid.
One of two status messages appears on the monitor. Follow the instructions below for the status message
on your monitor.
Status: Failure Recovery
1. Press the Cancel button.
2. Press the Open/Close
3. Select your login name and press the OK
4. Enter your login name and password.
5. The basin lid opens.
6. Remove the endoscope or correct the cause of the error.
7. Re-start the disinfection program.
button.
button.
75
Status: Shutdown
The endoscope cannot be removed from the machine. Continue the disinfection program.
After the current program is completed, run the disinfection program again. Except for the operator login
name, all entered data are automatically copied from the previous run if these steps are followed:
1. Press the Open/Close
2. Select your login name and press the OK
button.
button.
3. Enter your login name and password.
4. The basin cover opens.
5. Within 30 seconds, press the Menu
6. Select your login name and press the OK
button.
button.
7. Enter your login name and password.
8. Select “Repeat last run” and press the OK button.
System Errors
If a system error occurs during standby mode, a warning message appears as the fault occurs but disappears
as the situation is corrected.
If the error occurs during a disinfection cycle, the program ends and uids are drained from the basin.
A recovery program runs until a safe state is reached and the endoscope can be removed. The recovery
program cannot be stopped by the operator.
WARNING Report all system errors to your supervisor.
76
Taking a Sample of Disinfectant – Check MRC
At the end of a cycle, the system will prompt to collect disinfectant.
1. Open the center door to access the sample port.
2. Wearing protective gloves, take a sample of the disinfectant from the sample port by pressing a sample
cup up against the metal prong.
Figure 28 Taking a Disinfectant Sample
3. Close the access door.
4. Test the sample using RAPICIDE
®
PA Test Strip to ensure that the disinfectant is above MRC.
77
5. Follow the prompt on the screen indicating whether the MRC passed or failed.
If Test Strip Fails
1. If the test strips indicates a fail, rerun the cycle and retest:
2. If the test strip indicates a fail again, open a new bottle of test strips and test the concentration again
with a new test strip.
3. If it continues to fail, contact Medivators Technical Support or your Medivators Distributor.
78
Complete the Disinfection Process
When the disinfection process is complete, the process LED indicator light blinks green, the message
“Completed” displays on the LCD screen, and the alarm sounds.
Table 2 Post-Disinfection Program Status
StatusDescription
ACTIVEThe program is progressing correctly.
CORRECTThe endoscope was disinfected without any faults or program interruptions.
CORRECT!!!The endoscope was disinfected with errors reported.
STOP (fault)The program is stopped due to an error.
STOP (ASC)The safety control program stopped the program.
STOP (OPERATOR)The operator stopped the program without a fault report
STOP (LIO)The program stopped automatically after an error report or the maximum
error time (50 minutes) was reached.
SHUTDOWN (fault)After an error, the endoscope is rinsed so that it can be safely removed.
SHUTDOWN (OPERATOR)The operator stopped the program. Run down program.
SHUTDOWN (ASC)The safety control program stopped the disinfection. Completion program.
ASC (REJECT)The safety control program rejected the disinfection. Open the lid to close
the program.
LIO (REJECT)Disinfection was rejected due to an automatic program stop.
79
Table 3 Post-Disinfection Instrument Status
StatusDescription
DISINFECTEDThe endoscope was run through a complete disinfection program.
NOT APPLICABLEThe program is not an endoscope disinfection program. Warning: The
endoscope can be used only if the connectors are correctly connected
after the program.
NOT SUITABLE FOR USEThe endoscope is insufciently disinfected.
WARNING Wear clean gloves to prevent soiling of the disinfected endoscope.
If a run error was generated, a channel is blocked, or the endoscope is leaking, the endoscope
must not be released for patient use. If there is any doubt as to whether an endoscope is
correctly disinfected, it must be disinfected again before use. Contact your supervisor for further
instructions.
Verify all connections to the endoscope are secure. Failure to do so could result in an endoscope
which is not disinfected and therefore should not be used on a patient.
1. Open the basin lid by pressing the Open/Close
button or by using the foot switch. The run status
report disappears and the Operator selection window opens.
2. Enter your login name and password.
3. Check that all adapters are still connected to the channel connections. If an adapter is loose or
disconnected, the disinfection was not complete and must be repeated.
WARNING Before removing the endoscope, verify all connections to the endoscope are secure.
If an adapter is loose or disconnected, the disinfection was not complete and must be repeated.
Failure to do so could result in an endoscope which is not disinfected and therefore should not
be used on a patient.
80
4. Disconnect the channel adapters
5. Remove the channel separator(s).
6. Take the connector block out of the basin and hang up the block to dry.
7. Take the endoscope out of the basin.
8. If appropriate, t the valves.
9. The endoscope is now ready for a new examination. If it will not be used right away, the endoscope
must be stored in an airing cupboard or on a suspension rack.
10. If automatic cycle log printing is enabled, the log is printed.
Reports
Run Data Recorded
For each disinfection program run, the following data is recorded and stored on the PC hard drive:
• Machine ID
• Run number
• Date and time
• Endoscope Code
• Operator ID
• Patient ID
• Process step times
• Errors
• Disinfectant MRC Pass/Fail
• Additional information as set up by the system administrator
81
Printing Reports
Current Report
A cycle log prints automatically whenever the MEDIVATORS® ADVANTAGE PLUS® Automated
Endoscope Reprocessor module cover opens at the end of a disinfection cycle. To redisplay the current
cycle log, double-click the printer icon in the system tray.
Earlier Reports
All completed cycle logs may be viewed and reprinted at any time from the Management program.
1. Display the Windows desktop by minimizing all running programs.
2. Double click on the MEDIVATORS ADVANTAGE PLUS System ADV icon to open the folder.
Figure 29 MEDIVATORS ADVANTAGE PLUS System Desktop Folder
3. Double-click on the Management icon.
82
Figure 30 Management Icon
4. Log in to the Management system.
5. On the Management screen sidebar, click on the General tab.
Figure 31 Management Screen
6. On the sidebar, click on the Cycles icon.
7. On the blank Cycle Reports screen, set the From Date and To Date parameters, and then click
on the Generate icon in the toolbar.
8. The Cycle Reports screen opens. Double-click on the desired cycle. Click on the
display the Cycle Log Screen. Click on to display the detailed Cycle Report.
icon to
83
Figure 32 Cycle Log Screen
Figure 33 Cycle Reports Screen
9. The Cycle Details screen appears. Click on the plus sign to the left of each line to expand the report.
Click on the printer icon to print a copy of the report.
84
Figure 34 Cycle Details Screen
Shutdown
Use the following process to shutdown the reprocessor at the end of the day.
1. Clean the system by wiping down all surfaces with a non-scouring cleansing agent. Use a lime
dissolving, non-scouring agent for the rinse basin and inside the lids. Use soap and water to clean the
rubber gaskets.
WARNING Avoid possible chemical burns. Always wear personal protective equipment (gloves,
goggles) when handling disinfectant and/or detergent.
2. Check the detergent and alcohol reservoirs for appropriate levels. Rell them if necessary.
3. Clean the detergent cap, bracket and reservoir. Flush the reservoir thoroughly with hot water. Rell the
detergent reservoir, if necessary.
4. Close the incoming water line shutoff valve (unless either a delayed start program, water-line disinfect
or endoscope reprocessing cycle is to be run).
85
Procedure Summary
1. Open the cover.
2. Connect the endoscope.
3. Close the cover.
4. Enter your login name and password, the endoscope ID, the patient number, the physician ID, and the
assistant ID.
5. Start the program.
6. After disinfection, open the basin lid.
7. Enter your login name and password.
8. Check the sprayer and the connections.
9. Disconnect the connector block.
10. Remove the endoscope from the basin.
11. Print a report.
Downtime
Brief Downtime
• If the MEDIVATORS
a brief period time—less than seven days—it should remain powered on.
®
ADVANTAGE PLUS® Automated Endoscope Reprocessor will be unused for
• To prevent moisture in the leak test system, t disinfection blocks in the MEDIVATORS ADVANTAGE
PLUS System when it is not is use.
Prolonged Downtime
• If the MEDIVATORS ADVANTAGE PLUS System will be unused for more than seven days, it should
be powered off.
• The tubing, blocks, and lters in the machine contain a lot of water where bacterial growth may occur.
After a long downtime, replace the bacterial lter and disinfect the MEDIVATORS ADVANTAGE
PLUS System before running any endoscope disinfection procedures.
• To shut down the MEDIVATORS ADVANTAGE System, close all running programs rst before
pressing the main PC power switch. Failure to do this may result in software errors.
• See storage section for further information.
86
CHAPTER 4
MAINTENANCE
This chapter contains basic service procedures required for maintaining the MEDIVATORS® ADVANTAGE
®
PLUS
continuous quality in endoscope reprocessing. Failure to perform required maintenance may result in
injury or death to the operator or patient.
Automated Endoscope Reprocessor. Strict adherence to these requirements guarantees optimum
87
Daily Maintenance
Details and instructions for the daily service requirements are included in Chapter 3: Operation.
The following daily maintenance procedures must be completed at the start of each day:
• Conrm that the main air pressure regulator is set between 2.8–3.0 bar (42.0 – 45.0 psi).
• Conrm that the leak test air pressure regulator is set at 0.24–0.25 bar (3.6 – 3.75 psi).
• Verify a sufcient ow of water into the basin or replace the system lters.
• Check that all O-rings on the connection columns are present and undamaged.
• Replace the screens in the rinse basin if they are soiled.
At the end of each day, clean the MEDIVATORS
®
ADVANTAGE PLUS® system:
• For the top and outside covers, use a non-scouring cleansing agent.
• On the endoscope basin and inside covers, use a non-scouring, lime-dissolving cleansing agent.
• Use soap and water to clean the cover gaskets and prevent them from sticking.
Weekly Maintenance
Inspection
Inspect the following on a weekly basis. Adjustments or changes should be made by a qualied service
technician.
• Hookup Connector Blocks: Check for cracks in the hookup hoses, especially at the connections
to the connector block. If cracks have formed, the hose should be replaced.
• Glass Covers: The basin lids should open and close easily.
• Door: The front door should open and close easily.
• MEDIVATORS ADVANTAGE PLUS System Tubing, Connections, Fittings, Hydraulics and Dosing System: Check for wear and leakage and replace as needed.
88
O-Ring Lubrication
1. Using a cloth, dry the O-rings on the connection columns. Check for wear and replace as needed.
2. Take a small amount of O-ring grease between your thumb and index nger and spread it diagonally
on the connector plate along the rst two columns.
3. Repeat these two steps for the other three pairs of columns.
4. Spread the grease lengthwise along the O-rings with your thumb and index nger.
Monthly Maintenance
Basin Drain Screen
Using a screen hook, remove and clean the screen. Check for wear and replace as needed.
Air Filter Replacement
The MEDIVATORS® ADVANTAGE PLUS® Automated Endoscope Reprocessor has four (4) air lters.
Each lter should be replaced every 6 months.
1. Acquire new inline lter assemblies (See Appendix B for replacement part numbers).
2. Open the right and left blue-doors and locate the lters.
89
Figure 1 Filter Location
1. Remove one lter at a time. Move the lter away from the machine to allow access to the ttings.
90
Figure 2 Quick Connect Fitting
4. Press on Quick Connect ttings to release the lter.
Figure 3. Pressing Quick Connect button.
5. Remove and discard the lter assembly.
6. Replace the lter by attaching the Quick Connect ttings. Ensure that the color coded ttings match.
7. Close the doors.
91
Water Filter Replacement
WARNING: Only replace one water lter at a time. When replacing the water lter, make sure that
the lter being removed is the same as the lter replacing it.
The water ltration system removes particulates and biological contaminants as small as 0.1 micron from
the supply water. During the water line disinfection procedure, the entire water ltration system and the
incoming water line for the MEDIVATORS
is disinfected.
Note: A water line and water ltration disinfection process must be performed whenever a water lter is
replaced. Refer to Chapter 2 Waterline and Water Filter Disinfection.
Water lters should be replaced at a minimum every three months for the 1.0 micron and 0.4 micron lters
and every six months for the 0.1 micron lter. If the water pressure drops below 40 PSI or the difference
between the gauges is greater than 5 PSI the corresponding lter needs to be replaced.
The following conditions must be met before changing a lter:
• No endoscope reprocessing cycles are running.
• No water line disinfection procedure is running.
• The inlet water supply is turned off.
You will need the following materials:
• A container to catch water.
• A lter bowl wrench.
• The necessary lter cartridges (see Table 1) (See Appendix B for replacement part numbers).
1. Relieve the pressure in the ltration system by opening the bleeder valve on the lter system (rotate
down).
2. Check the inlet pressure gauge to ensure that the pressure is at zero.
3. Place an empty container under the 1 micron lter housing and disconnect the Quick Connect tting
at the bottom of the housing. Allowing all of the water to empty out.
4. Using a lter bowl wrench, remove the 1 micron lter housing and discard the used lter cartridge.
94
Figure 7 Loosen Filter Housing
Figure 8 Remove Housing and Filter
5. Place a new lter in the housing, ensuring that the O-ring is in place.
Figure 9 Check O-Ring
6. Thread the housing onto the header and tighten by hand only.
7. Repeat the procedure to replace the 0.4 micron and 0.1 micron lter.
95
Figure 10 Open Bleed Valve
8. To remove the 0.1 micron lter rotate the used lter cartridge counterclockwise, pull downward,
and discard.
9. To install a new 0.1 micron lter insert a new lter cartridge into the housing header and rotate
clockwise until the tabs lock into place gure 11.
10. Return the bleed valve to the closed position (rotate up).
11. Slowly open the main water supply valve and check for any leaks.
96
Figure 11 Filter Mounting Tabs
12. Run the Water Sample Utility
• Push Menu button
• Log in
• Select: Program
• Select: Utility Program
• Choose: Take Water Sample
13. Place a container under the bleed valve outlet tube and slowly open the bleed valve (13 g. 4). Let the
water run until it is a steady stream.
14. Close the bleed valve.
15. Push cancel to stop water utility or allow to complete. No actual sample needs to be taken.
16. Perform a water line and water ltration disinfection procedure. Refer to Chapter 2.
97
NOTE
98
CHAPTER 5
TROUBLESHOOTING
This chapter lists error messages you may encounter while using the MEDIVATORS® ADVANTAGE
®
PLUS
error recurs, contact your Technical Service representative.
Automated Endoscope Reprocessor. If the suggested action does not solve a problem or if the
99
Introduction
Error Reports
The MEDIVATORS
display different types of errors which signal and identify specic areas requiring attention. Error types
include:
• System Errors: A communication problem identied by the prex C with errors ranging from C-001
through C-099. This is an error caused by the Windows based PC or MEDIVATORS ADVANTAGE
PLUS system.
• Bar Code Errors: A bar code problem identied by the prex M with errors ranging from M-910
through M-940.
• Endoscope Errors: An endoscope problem identied by the prex S with errors ranging from
S-102 through S-177.
• Reprocessing Cycle Errors: A problem identied during the endoscope disinfection cycle
identied by the prex S with errors ranging from S-201 through S-297, and S-511 through S-832
• Control Errors (SCU): Lists any deviations in the disinfection program as monitored by the SCU
controller, and identied by the pre-x S with errors ranging from S-300 through S-399.
• Warning Errors: An alert or warning identied so that the operator is aware of a machine condition
that requires attention. Warnings are identied by the prex W and range from W-701 through W-821.
®
ADVANTAGE PLUS® Automated Endoscope Reprocessor system can report and
100
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.