MediSense Precision PCx User Manual

Page 1
Point-of-Care System
MediSense, Precision PCx, Precision and QC Manager are trademarks of Abbott Laboratories. © 2002 Abbott Laboratories.
Warning: This manual is protected by copyright law. Unauthorized reproduction or distribution of this manual or any portion of it may result in severe civil and criminal penalties, and will be prosecuted to the maximum extent under the law.
Healthcare Professional Operator’s Manual for Blood Glucose Monitoring
Page 2
Questions? Refer to Page 10-13 For Customer Care Information.
I
Introduction
1. Manual Organization................................................................... 1-1
Precision PCx Menu Tree.......................................................... 1-2
2. Overview: Understanding the Precision PCx
Point-of-Care System .................................................................. 2-1
Test Mode
3. Patient Test.................................................................................... 3-1
Operating Guidelines for all Samples....................................... 3-1
Collecting Blood Samples......................................................... 3-1
Obtaining Capillary Blood Samples ...................................... 3-1
Obtaining Venous Blood Samples.......................................... 3-2
Obtaining Arterial Blood Samples......................................... 3-2
Disposing of Waste .................................................................... 3-3
Test Procedure............................................................................ 3-3
4. Control Test .................................................................................. 4-1
Operating Guidelines for Control Solutions............................. 4-1
Test Procedure............................................................................ 4-2
Menu Mode
5. Data Review................................................................................... 5-1
Patient History ........................................................................ 5-1
Control History..................................................................... 5-15
Proficiency Data ................................................................... 5-25
Linearity Data....................................................................... 5-26
6. Proficiency Test ............................................................................. 6-1
7. Linearity Test ................................................................................ 7-1
8. Review/Change Setup................................................................... 8-1
Test Setup ................................................................................ 8-7
System Setup ........................................................................ 8-34
System Status........................................................................ 8-40
Date and Time....................................................................... 8-43
Edit Lists............................................................................... 8-46
Table of Contents
Page 3
Questions? Refer to Page 10-13 For Customer Care Information.
II
9. Data Upload................................................................................... 9-1
Docking Station (optional)........................................................ 9-2
10. Troubleshooting.......................................................................... 10-1
Patient Test Results.................................................................. 10-1
Out-of-Range Control Test Results......................................... 10-2
Error Messages ........................................................................ 10-2
Contacting MediSense for Service........................................ 10-13
Returning an Instrument........................................................ 10-16
Limited Warranty................................................................... 10-16
11. Maintenance................................................................................ 11-1
Cleaning the Monitor’s Exterior Surface................................ 11-1
Replacing the Batteries ............................................................ 11-2
Six-Month Maintenance Log .................................................. 11-3
12. Specifications............................................................................... 12-1
Other Information.................................................................... 12-2
Page 4
Questions? Refer to Page 10-13 For Customer Care Information.
III
Introduction
Welcome to the MediSense Precision PCx™Point-of-Care System, the portable hospital whole blood testing system. The Precision PCx System simplifies patient monitoring for point-of-care whole blood glucose testing by nurses, physicians, diabetes educators, laboratory staff and other healthcare professionals who need quick, accurate results. The Precision PCx System includes the following:
Features and Benefits
Hand-Held Monitor The hand-held design of the monitor allows
it to be easily brought to the patient for testing.
Simplified Data Entry The monitor contains a built-in barcode scanner
which allows the operator to enter patient identification, operator identification, test strip lot number and other pertinent data by simply scanning a barcode.
Easy Operation To perform a test, a test strip is inserted into the
monitor’s test strip port, and a small drop of blood is applied to the target area on the test strip.
Rapid Test Results The monitor automatically begins testing as soon as
the sample is detected on the test strip. Results are displayed in 20 seconds.
Easy Home Training The monitor’s design and operation are similar to
other MediSense monitors, making it easy to educate patients on home self-testing procedures.
Infection Control The risk of contamination to operator and patient
is minimized since the blood does not enter the monitor.
Improved Test Strip The Precision PCx System uses unique biosensor Technology technology to obtain accurate results over a broad
range of patient hematocrit levels and sample types. Precision PCx Test Strips and Precision PCx Plus Test Strips with TrueMeasure
Technology are approved for use with arterial, venous, capillary and neonatal whole blood samples.
Minimal Maintenance The monitor’s exterior surface needs only to be
wiped down once a day, or as needed, and the batteries need to be replaced or recharged when necessary.
MediSense Precision PCx Point-of-Care System
Page 5
IV
Accurate Calibration The barcode on each test strip packet label ensures
accurate calibration before each test begins.
Built-In Security Options Options in the Precision PCx Monitor enable the
requirement of Patient and Operator IDs, to include minimum and maximum lengths. Other security options include:
QC Interval – requires quality control tests at regular intervals. Operator Security – limits access to the list of certified operators. Strip Lot Security – test only with approved test strip lots. Test Security – reports control test as Pass/Fail only. This is to prevent patient testing from being performed while the monitor is in control test mode.
Patient and Quality The Precision PCx Monitor can store up to 4,000 Control Data Management patient tests and 1,000 control tests. All stored data
can be automatically uploaded to the MediSense
®
QC Manager™Data Management System using the docking station (optional) or serial cable.
Data Networking The QC Manager Data Management System provides
a simple and automated way to collect, report and send data to the Laboratory Information System (LIS) or Hospital Information System (HIS).
Battery Power The monitor uses 2 standard alkaline AA batteries,
or the rechargeable MediSense battery pack.
Page 6
Questions? Refer to Page 10-13 For Customer Care Information.
1-1
The Precision PCx Point-of-Care System Operator's Manual provides basic information about the Precision PCx Monitor and contains easy­to-follow, step-by-step procedures for using it.
The Precision PCx System operates using menu-driven software. The software consists of two main menus, Test Mode and Menu Mode. Test Mode is used to perform Patient Tests and Control Tests. Menu Mode is used to set up the monitor's options. (See page 1-2).
The menu portion of the software is set up like a tree, with various levels of sub-menus branching out from the main trunk. The menu diagram on the next page, called the Precision PCx Menu Tree, illustrates the multiple sub-menu layers. For ease of understanding in using the monitor, this manual is organized into sections which follow the menu.
It is recommended that the Introduction and Overview sections of this manual be read thoroughly prior to using the monitor for testing. These sections will acquaint users with the Precision PCx Point-of-Care System’s features and operating controls so it can be used more effectiv ely.
Manual Organization
1
Page 7
1-2
Test Mode
Patient
Test
Control
Test
Menu Mode
1. Data Review
2. Proficiency Test
3. Linearity Test
4. Review / 5.Change Setup
Upload
Inter
val
Operator ID Format Patient ID Format
Comment Code Memory Full
Results Format Comment Code Memory Full
QC Interval Operator ID Strip Lot
Glucose
Units Sample Type Action Range
1. Patient History sort by:
2. Control History
3. Proficiency Data
4. Linearity Data
1. Test Setup
2. System Setup
3. System Status
4. Date and Time
All T ests Patient ID Operator ID Action Range Comment Code Date
All T ests Operator ID Outside Range Comment Code Date
1. All Tests
2. Patient Test
3. Control Test
4. Security
5. Test Type
1. Operator List
5. Edit Lists
Add Delete Change
Add Delete Change
sort by:
2. Strip Lot List
1.
2.
Battery Power Temperature Memory Gauge Software Version
Language Beeper On/Off Date Format Time Format Password Factory Settings
Levels
Precision PCx Menu Tree
Page 8
Questions? Refer to Page 10-13 For Customer Care Information.
2-1
Understanding the Precision PCx Point-of-Care System
The Precision PCx Point-of-Care System is a portable whole blood testing system which performs tests with quick and accurate results. This section is intended to provide an overview of the Precision PCx System's components and basic operations. Use product and accessories only as directed. Failure to operate the product(s) in accordance with the manufacturer’s documentation may impair the product safety.
MediSense Precision PCx Point-of-Care System Components
The MediSense Precision PCx Point-of-Care System includes the following items:
• The Precision PCx Monitor
• Precision PCx and Precision PCx Plus Blood Glucose Test Strips
• Accessory / Carry Case (optional)
• Docking Station (optional)
The Precision PCx Monitor includes:
• Test Strip Port
• Large LCD Display
• Barcode Scanner
• Ke ypad
• Data Port
• Audible Indicator
Precision PCx Blood Glucose Test Strips
The Precision PCx and Precision PCx Plus Blood Glucose Test Strips offer the latest advances in biosensor technology. Each test strip is wrapped in a foil packet with a barcode label. This label holds the calibration information about the test strip including the lot number, the expiration date and expected control solution ranges.
Overview: 2
Page 9
Precision PCx Test Strips
There are now two types of Precision PCx test strips that may be used for testing with this system. The target area of the original Precision PCx biosensor test strip is on the top of the strip. The target area of the new Precision PCx Plus Test Strip with TrueMeasure™ Technology is at the end of the strip. Only Precision PCx branded test strips should be used for testing with this system. Note: For data management, Precision PCx Plus Test Strips can be used only in conjunction with QC Manager 3.0.
Inside each carton of test strips are test strip instructions for use. The instructions contain important information about the test strip such as: Storage and Handling, Sample Collection and Preparation, Quality Control and Limitations of the Procedure. The section “Limitations of the Procedure” contains the operating ranges for the system and the test strips. Please read the instructions for use carefully.
2-2
Precision PCx Test Strip
Lot Number
Expiration Date
Control Test Ranges
Barcode number
Barcode
Precision PCx PlusTest Strip with T rueMeasure Technology
T arget area:A small green and
white oval where you place a drop of blood on top of the test strip.
T arget area:A white area where you place a drop of blood at the end of the strip. You can apply blood to the edge of the strip or on the top of the strip within this white area.
Contact bars:At the opposite end of both test strips are three black contact bars to be inserted into the test strip port of the monitor.
Precision PCx PlusT est Strip with TrueMeasure T echnology For Use with Precision PCx v2.2 Only
Precision PCx T est Strip For Use with Precision PCx v2.0,v2.1 and v2.2
Lot Number
Barcode number
Barcode
Expiration Date
Control Test Ranges
Target Area
Contact Bars
Contact Bars
Target Area
Page 10
Questions? Refer to Page 10-13 For Customer Care Information.
2-3
Accessory / Carry Case (optional)
The light yet rugged accessory/carry case holds the hand-held monitor, test strips, glucose control solutions and other supplies in a convenient manner.
A transparent cover allows the operator to check the supply status without opening the case. The overall form is compact, allowing the convenient storage of many cases in a small area.
The case holds the Precision PCx Monitor and the following items:
• Precision PCx or Precision PCx Plus Test Strips
• Single-use Lancing Devices
• Gauze Pads
• Glucose Control Solutions
• Quick Reference Guide
Docking Station (optional)
The optional docking station will automatically upload the data from the monitor into a central computer via the MediSense QC Manager Data Management System.
To start the upload of data, simply place the Precision PCx Monitor into the Precision PCx Docking Station.
Page 11
2-4
The monitor will automatically upload data to the MediSense QC Manager Program. Please refer to the QC Manager Operator’s Manual for more information.
The docking station has an additional role as a recharge station for the MediSense rechargeable battery pack (optional). The rechargeable battery pack must be in the monitor to take advantage of this option.
Identifying Monitor Functions
Through the monitor, the operator can view information, make selections and enter information. The location of all the monitor functions are shown on the following page.
Test Strip Port
After the operator scans the barcode or manually enters* information via the monitor keypad, the operator will insert a new test strip into the test strip port to perform a test.
Display
Here is where menu options, error codes, and test results are viewed. Press and hold on the keypad to turn on or off the backlight for the monitor display.
Keypad
The keypad contains the following controls:
On/Off
Press this button to turn the monitor on or off. The monitor will auto­matically turn off after 4 minutes if the monitor is left on but has not been used.
Scan
Press the Scan button to use the laser barcode scanner. The barcode scanner enables the operator to scan the information into the monitor instead of manually entering the data by using the keypad. The scanner can automatically enter the test strip information (calibration, control range, lot number and expiration date), as well as the operator ID, patient ID, and control lot number.
* Manual entry of strip lots with alpha-numeric barcodes is not supported.
Page 12
Questions? Refer to Page 10-13 For Customer Care Information.
2-5
Numbers
The 10-digit keypad can be used to enter identification numbers or to select an option that appears on the display.
Clear
The operator can press Clear to perform the following functions:
• To back up one space while entering numbers on the keypad.
• Clear a numerical code from the monitor after pressing Enter.
• Clear a barcode scanned by mistake.
• Return to a previous screen.
• When Clear is pressed two times, all entered numbers shown in the display will be cleared.
Menu
When the Menu key is pressed, the following functions can be performed:
• Return to the top of the menu.
• Switch between Test Mode and Menu Mode.
Enter
When the Enter key is pressed, it will confirm the information that was manually entered.
Barcode Scanner
The Barcode Scanner is used to scan barcode information into the monitor. The operator will hold the barcode scanner 3-12 inches from the barcode to be scanned. When first using the barcode scanner, some precautions should be taken while mastering the device.
Page 13
2-6
Place the object to be scanned on a flat surface, by itself. This will prevent other items from being accidentally scanned.
Press and hold down the Scan button to start the barcode scanner. A visible red beam will be emitted from the monitor. Slowly move this beam over the barcode on the item to be scanned. When the monitor accepts the barcode, it will beep in acknowledgement. Without releasing the Scan button, look down at the monitor. The operator can view the data that was scanned by the monitor. Release the Scan button to stop the scan and continue with the next step.
CAUTION: Never look into the barcode scanner beam or point it toward anyone’s eyes. The beam could cause permanent damage to the eye.
Class II and IIa Laser Product: Avoid Long Term Viewing of Direct Laser Light.
Data Port
The Data Port is located at the bottom of the monitor. This port enables the monitor to automatically transfer the data through a cable connector to a PC, laptop, Laboratory Information System (LIS), or Hospital Information System (HIS).
Battery Compartment
The monitor holds two AA alkaline batteries to power the monitor. A rechargeable battery pack is also available from MediSense. The rechargeable battery pack can recharge itself while the monitor is in the optional docking station. To prolong the battery life of the monitor, the monitor should be turned off when not in use. If the monitor is left on and has not been used for 4 minutes, it will automatically shut itself off.
Page 14
Questions? Refer to Page 10-13 For Customer Care Information.
2-7
Audible Indicator (not visible)
The monitor will beep to indicate one of the following:
• A successful barcode entry.
• Sample detection.
• At three seconds before the result is to appear.
• When an error occurs.
• At thirty seconds before automatic shutoff.
Data Port
Page 15
This section describes the steps to be taken to test patient blood samples using the Precision PCx Monitor.
Operating Guidelines for All Samples
Please follow the recommended guidelines so the most accurate results will be obtained.
• Always wear gloves and follow your facility’s biohazard safety policies and procedures when performing tests involving patient blood samples.
• Use only control solutions approved for use on the test strip instructions for use to verify the performance of the Precision PCx and Precision PCx Plus Test Strips and the Precision PCx Monitor.
• Use the test strips before their expiration date.
• Do not use test strips that are wet, bent, scratched or damaged. Use the test strip immediately after opening its foil packet.
• Do not scan a packet’s barcode and use a test strip from another packet. This may cause incorrect assay results to be generated.
• Cover the entire target area of the test strip with the blood sample. The test results will not be affected if the target area has been briefly touched with the patient’s finger, a capillary tube or pipette.
• If the test fails to start, apply a second drop of blood to the target area within 30 seconds. If the test fails to start after the second drop is applied, or if more than 30 seconds have passed, discard the used test strip and repeat the test.
• After the blood is applied to the test strip and the count down begins, do not remove or disturb the test strip.
• Use a new test strip when repeating a patient test.
• Refer to the instructions for use in the test strip box for specific directions on storage and use of the test strips.
Questions? Refer to Page 10-13 For Customer Care Information.
3-1
Patient Test 3
Page 16
3-2
Collecting Blood Samples
How to Obtain a Capillary Blood Sample:
• Use only fresh whole blood samples.
• Make sure no water remains on the puncture site.
• Collect the capillary blood using a lancing device and an appropriate technique.
• Avoid squeezing the puncture site excessively.
• Apply the drop of blood directly to the target area of the test strip, covering the entire area.
• If necessary, the blood can be collected in a capillary tube coated with sodium heparin, lithium heparin, or EDTA, and then applied to the test strip.
How to Obtain a Venous Blood Sample:
• Use only fresh whole blood samples.
• Collect the venous blood sample in a collection tube containing sodium heparin, lithium heparin or EDTA ensuring that the test tube is completely filled.
• Do not use collection tubes that contain fluoride or oxalate.
• If the blood is collected from an intravenous line, clear the line before drawing the sample into a syringe that contains sodium heparin, lithium heparin or EDTA.
• Use the sample within 30 minutes of collection.
• Invert the tube with the sample several times immediately before removing the sample.
• Use a disposable transfer pipette to obtain a sample from the center of the collection tube.
Precision PCx PlusT est Strip with TrueMeasure T echnology For Use with Precision PCx v2.2 Only
Precision PCx T est Strip For Use with Precision PCx v2.0,v2.1 and v2.2.
Precision PCx PlusT est Strip with TrueMeasure T echnology For Use with Precision PCx v2.2 Only
Precision PCx T est Strip For Use with Precision PCx v2.0,v2.1 and v2.2.
Page 17
• Apply a drop of blood directly to the target area on the test strip, covering the entire area.
Note: Use VENOUS MODE only for venous samples that have been in capped tubes and have not been mixed with air. Use CAPILLARY or ARTERIAL MODE for all other samples.
How to Obtain an Arterial Blood Sample:
• Use only fresh whole blood samples.
• Clear the arterial line before drawing a blood sample into a syringe that contains sodium heparin, or lithium heparin or EDTA.
• Use the sample within 30 minutes of collection.
• Mix the syringe several times immediately before applying the sample to the target area on the test strip.
• Allow a drop of blood to form at the tip of the syringe.
• Cover the entire target area on the test strip with the blood sample. The syringe can briefly touch the test strip without affecting the test result.
Disposing of Waste
Observe the following guidelines when disposing of biohazardous waste:
• Dispose of used lancets in an approved sharps container.
• Discard used capillary tubes, disposable transfer pipettes or tips and test strips in an approved biohazard container.
• Follow your facility’s biohazard disposal policy.
Patient Test Procedure
Use the following procedure to perform patient tests.
Note: The word “Strip” on the display always refers to the test strip. If the monitor displays a message not indicated in these procedures, please refer to section 10, Troubleshooting.
Questions? Refer to Page 10-13 For Customer Care Information.
3-3
Precision PCx PlusT est Strip with TrueMeasure T echnology For Use with Precision PCx v2.2 Only
Precision PCx T est Strip For Use with Precision PCx v2.0,v2.1 and v2.2.
Page 18
What You See on the Display
What You Do Comments
3-4
3. Press Scan to scan the
Operator ID barcode, or manually enter the Operator ID via the keypad, then press
Enter.
The display will prompt for a Patient ID to be entered, if this setting has been enabled.
Note: If a Patient ID prompt is not seen, skip down to step 5.
1. Press On/Off to turn
on monitor.
The Abbott logo screen will appear for a few seconds followed by the product name, software version screen and then the Test Mode menu.
2. Press 1 to select Patient
T est.
After the patient test is selected, the Operator ID screen will appear, if this setting has been enabled.
Note: If an Operator ID prompt is not seen, skip down to step 4.
Page 19
What You See on the Display
What You Do Comments
Questions? Refer to Page 10-13 For Customer Support Information.
3-5
4. Press Scan to scan the
Patient ID barcode, or manually enter the Patient ID via the keypad, then press Enter.
After entering the Patient ID, the test strip lot screen appears next.
5. Press Scan to scan the
test strip barcode, or manually enter* the test strip lot number via the keypad, then press
Enter.
After the test strip lot number has been entered, the Sample Type screen will appear if this setting has been enabled. If it does not appear, skip down to step 7.
6. The operator can select
one of the following:
Press 1 – Capillary or Arterial.
Press 2 – Venous.
Note: Use VENOUS MODE only for venous samples that have been in capped tubes and have not been mixed with air. Use CAPILLARY or ARTERIAL MODE for all other samples.
7. Open the foil test strip
packet at the notch and tear up or down to remove the test strip.
* Manual entry of strip lots with alpha-numeric barcodes is not supported.
Page 20
What You See on the Display
What You Do Comments
3-6
8. With the contact bars
facing up, insert the test strip into the test strip port until it stops.
The Apply Sample screen will appear.
9. Apply a drop of blood
directly from the patient’s finger or from a syringe to the target area on the test strip.
The monitor beeps when the sample is accepted. The Sample Accepted screen will appear. For detailed description of the test strip target area, refer to page 2-2.
The test starts automatically, as the sample is accepted.
10. Wait for the monitor
to analyze the sample and display the test result.
The monitor counts down 20 seconds, then displays the test result.
11. Note the test result
and whether it falls outside the action range.
If the test result is above or below the action range, an up or down triangle appears in front of the test result.
(5.8 mmol/L)
Page 21
What You See on the Display
What You Do Comments
Questions? Refer to Page 10-13 For Customer Support Information.
3-7
For Out-of-Range results, refer to your institution’s protocol.
NOTE: The Outside Action Range screen will appear if the Patient Test Out-of-Range Comment Code is enabled and the following two conditions are met:
A. The Action Range is
set up.
B. The test result is out of
the Action Range.
The monitor may be enabled to scan or enter a 1- to 2-digit comment code. If there is no prompt to enter a comment code, skip down to step 12.
NOTE: The In-Range Action Range screen will appear if the Patient Test In-Range Comment Code is enabled and the following conditions are met:
A. The action range
is defined.
B. The test result is within
the Action Range.
12. Remove the test strip
from the monitor when finished testing.
(21.9 mmol/L)
(5.8 mmol/L)
Page 22
What You See on the Display
What You Do Comments
3-8
13. The operator can
select one of the following options:
Press 1 – Next Patient. Press 2 – Repeat Test. Press 3 – Patient History.
If 1 is selected, the Scan Patient ID screen will appear. Return to step 4 to continue with testing on the next patient.
If 2 is selected, the Scan or Enter Strip Lot screen will appear. Return to step 5 to repeat the test.
If 3 is selected, the monitor will display the last test result for this patient as well as allow the operator to view previous results.
Refer to the section entitled Data Review for Patient History for further information.
(5.8 mmol/L)
Important: If the blood glucose result appears to be inconsistent (lower or higher than expected), there may be a problem with the test strip. Repeat the test using a new test strip. Results that are incorrect may have serious medical consequences. Consult your healthcare professional before making any changes to diabetes medication plans if:
• The blood glucose results are not consistent with the physical symptoms AND you have ruled out common errors in technique.
• The blood glucose result is less than 50 mg/dL (2.8 mmol/L) or greater than 300 mg/dL (16.7 mmol/L).
Page 23
Questions? Refer to Page 10-13 For Customer Care Information.
4-1
This section describes the steps to be taken to test Control Solutions in order to verify the performance of the Precision PCx Monitor and Precision PCx and Precision PCx Plus Test Strips. The quality control features of the monitor can be customized to fit the requirements of each facility. The monitor can be set to require:
• Tests of Low, Normal, and High Control Solutions or any variation of this depending on the facility’s policy
• Tests of Control Solutions at regular times or time intervals, or by number of patient tests
• Control test results will appear as Numeric or Pass/Fail.
These quality control features can be set through the Menu Mode menu (Change Setup/Test Setup Menu).
Operating Guidelines for Control Solution Testing
Observe the following guidelines to obtain optimal quality control results using the Precision PCx System:
• Use only control solutions approved for use on the test strip instructions for use to verify the performance of the Precision PCx and Precision PCx Plus Test Strips and the Precision PCx Monitor.
• When opening a new bottle, write the current date on the bottle label. Each bottle of control solution is stable for 90 days after opening (replace cap tightly after each use). Discard all unused solutions 90 days after opening.
• Invert the control solution bottle several times to ensure thorough mixing before use.
• Invert and tap the capped control solution bottle to remove air bubbles from the nozzle of the bottle.
• Do not scan one test strip packet’s barcode and use a test strip from another packet. This may cause incorrect assay results to be generated.
• Replace the correct cap on the bottle and tighten the cap immediately after each use.
• Store the control solutions at temperatures between 39º and 86ºF (4º and 30ºC).
• Do not use control solutions after the expiration date printed on the bottles and the box. The Precision PCx Monitor does not accept control solutions that have passed their expiration date.
When To Test Control Solutions
• As required by your facility’s quality policy or local regulatory requirements.
• When you question blood glucose results
• When you open a new lot of test strips
Control Test 4
Page 24
4-2
What You See on the Display
What You Do Comments
Control Test Procedure
Use the following procedure to perform control tests.
Note: The word “Strip” on the display always refers to the test strip. If the monitor displays a message not indicated in these procedures, please refer to section 10, Troubleshooting.
1. Press On/Off to turn on
the monitor.
The Abbott logo screen will appear for a few seconds followed by the product name, software version and then the Test Mode menu.
2. Press 2 – Control Test. After the control test is
selected, the Operator ID screen may appear, if the setting has been enabled.
3. Press Scan to scan the
Operator ID barcode, or manually enter the Operator ID via the keypad, then press
Enter.
If the Operator ID is not required, just press enter.
After the Operator ID is entered, the Low Level Solution Lot screen will appear.
4. Scan or manually enter
the low control solution lot number via the keypad, then press
Enter.
The Strip Lot screen will appear next.
If the Unexpected Level screen appears, the operator may either:
1. Enter 1 to ReEnter the expected level.
2. Enter 2 to continue to test the unexpected level.
Page 25
Questions? Refer to Page 10-13 For Customer Care Information.
4-3
7. With the contact bars
facing up, insert the test strip into the test strip port until it stops.
The monitor will prompt the operator to apply the desired control solution to the test strip.
What You See on the Display
What You Do Comments
5. Scan or manually
enter* the test strip lot number via the keypad, then press Enter.
After the test strip lot number is entered, the monitor will prompt for a test strip to be inserted into the test strip port.
6. Open the foil test strip
packet at the notch and tear up or down to remove the test strip.
* Manual entry of strip lots with alpha-numeric barcodes is not supported.
Precision PCx PlusT est Strip with TrueMeasure T echnology For Use with Precision PCx v2.2 Only
Precision PCx T est Strip For Use with Precision PCx v2.0,v2.1 and v2.2.
Page 26
4-4
What You See on the Display
What You Do Comments
9. Wait for the monitor to
analyze the control solution and display the test result.
The monitor counts down 20 seconds then displays the test result.
The test starts automatically as the sample is accepted.
(2.7 mmol/L)
10. Note the test result
and whether it falls within the acceptable range.
Control test results can appear as numeric or PASS/FAIL, depending on how the test is setup.
The display also shows the control level, the control range, date and time.
8. Gently invert the
required control solution bottle 3-4 times then apply a drop to the target area on the test strip.
The monitor beeps when the sample is accepted. The Sample Accepted screen will be displayed. Recap the control solution bottle tightly. For a detailed description of the test strip target area, refer to page 2-2.
Page 27
Questions? Refer to Page 10-13 For Customer Care Information.
4-5
14. The operator may
select one of the following:
Press 1 – Next Level. Press 2 – Repeat Test Press Menu to return to
the Menu Mode menu. Press On/Off to turn off
the monitor.
If 1 is selected, the next level test will appear. The operator may perform the next level control test by returning to Step 4.
If 2 is selected, the Scan or Enter Strip Lot screen will appear.
Return to step 5 to repeat the test.
What You See on the Display
What You Do Comments
13. Remove the test strip
from the monitor when finished testing.
(2.7 mmol/L)
11. Note the test result
and whether it falls within the acceptable range.
The display shows the control level, PASS/FAIL, date and time.
12. If required, scan or
manually enter the comment code, and press Enter .
The monitor may be enabled to scan or enter a 1- to 2-digit comment code. If there is no prompt to enter a comment code, skip down to step 13.
An “X” will appear in front of a test result that is Out-of-Range.
(x3.7 mmol/L)
Page 28
Patient History
The Precision PCx Monitor stores up to 4,000 patient tests that can be retrieved by the following categories:
All Tests: This option displays all patient test results, starting with
the most recent test results (reverse chronological order).
Patient ID: This will display the test results for a particular patient,
starting with the most recent test.
Operator ID: This will show test results performed by a particular
operator, starting with the most recent test.
Action Range: This will display the test results that fall outside of
the action range.
Comment Code: This shows the test results for all tests tagged with
a specific comment code.
Date: This will display all the test results for a specific date, starting
with the most recent test done on that date.
5-1
Questions? Refer to Page 10-13 For Customer Care Information.
Data Review 5
Page 29
5-2
What You See on the Display
What You Do Comments
Data Review:
Follow these steps to access the Data Review menu: If the monitor displays a message not indicated in these procedures, please refer to section 10, Troubleshooting.
1. Press the On/Off
button to turn the monitor on.
The Abbott logo screen will appear for a few seconds, followed by the product name, software version and then the Test Mode menu.
2. Press the Menu button. The Menu Mode screen
will appear next.
3. Press 1 – Data Review. The Data Review menu
appears.
At this point, Patient History, Control History, Proficiency Data as well as Linearity Data can be viewed.
Page 30
Questions? Refer to Page 10-13 For Customer Care Information.
5-3
Data Review for Patient History by All Tests:
Please refer to page 5-2 of the Data Review section for the step-by-step instructions to access the Data Review menu.
Note: If no information is stored for the desired category that was selected, the following message will be displayed on the monitor: “No Stored Tests”.
Note: If no tests match the criteria specified, this message will be displayed on the monitor: “No Tests Match Request”. If the monitor displays a message not indicated in these procedures, please refer to section 10, Troubleshooting.
What You See on the Display
What You Do Comments
1. Press 1 – Patient
History.
The Patient History menu now appears.
2. Press 1 – All Tests. The monitor shows the
most recent patient test result.
Page 31
5-4
3. The operator can select
one of the following options:
Press 1 – Previous, view the previous test result.
Press 2 – Next, view the next test result.
The number that appears to the right of 1-Previous and 2-Next on the monitor indicates the number of tests available for review.
The display also shows: p: Patient ID o: Operator ID 12 : Comment Code ven: Venous Sample
4. When finished
reviewing the data, the operator can:
Press Clear to return to the Patient History menu.
Press Menu to return to the Menu Mode menu.
Press On/Off to turn off the monitor.
What You See on the Display
What You Do Comments
(5.8 mmol/L)
(21.9 mmol/L)
Page 32
Questions? Refer to Page 10-13 For Customer Care Information.
5-5
What You See on the Display
What You Do Comments
Data Review for Patient History by Patient ID:
Please refer to page 5-2 of the Data Review section for the step-by-step instructions to access the Data Review menu.
Note: If no information is stored for the desired category that was selected, the following message will be displayed on the monitor: “No Stored Tests”.
Note: If no tests match the criteria specified, this message will be displayed on the monitor: “No Tests Match Request”.
If the monitor displays a message not indicated in these procedures, please refer to section 10, Troubleshooting.
1. Press 1 – Patient
History.
The Patient History menu now appears.
2. Press 2 – Patient ID. The monitor will prompt
for the Patient ID to be scanned or entered.
3. Scan or manually enter
the Patient ID via the keypad, then press
Enter.
Page 33
5-6
What You See on the Display
What You Do Comments
5. When finished
reviewing the data, the operator can:
Press Clear to return to the Patient ID screen.
Press Menu to return to the Menu Mode menu.
Press On/Off to turn off the monitor.
4. The operator can select
one of the following options:
Press 1 – Previous, view the previous test result.
Press 2 – Next, view the next test result.
The number that appears to the right of 1-Previous and 2-Next on the monitor indicates the number of tests available for review.
The display also shows: p: Patient ID o: Operator ID 12 : Comment Code ven: Venous Sample
(5.8 mmol/L)
Page 34
Questions? Refer to Page 10-13 For Customer Care Information.
5-7
What You See on the Display
What You Do Comments
Data Review for Patient History by Operator ID:
Please refer to page 5-2 of the Data Review section for the step-by-step instructions to access the Data Review menu.
Note: If no information is stored for the desired category that was selected, the following message will be displayed on the monitor: “No Stored Tests”.
Note: If no tests match the criteria specified, this message will be displayed on the monitor: “No Tests Match Request”.
If the monitor displays a message not indicated in these procedures, please refer to section 10, Troubleshooting.
1. Press 1 – Patient
History.
The Patient History menu now appears.
2. Press 3 – Operator ID. The monitor will prompt
for the Operator ID to be scanned or entered.
3. Scan or manually enter
the Operator ID via the keypad, then press
Enter.
The monitor shows the result of the operator's most recent patient test result.
Page 35
5-8
What You See on the Display
What You Do Comments
4. The operator can select
one of the following options:
Press 1 – Previous, view the previous test result.
Press 2 – Next, view the next test result.
The number that appears to the right of 1-Previous and 2-Next on the monitor indicates the number of tests available for review.
The display also shows: p: Patient ID o: Operator ID 12 : Comment Code ven: Venous Sample
5. When finished
reviewing the data, the operator can:
Press Clear to return to the Operator ID screen.
Press Menu to return to the Menu Mode menu.
Press On/Off to turn off the monitor.
(5.8 mmol/L)
Page 36
Questions? Refer to Page 10-13 For Customer Care Information.
5-9
What You See on the Display
What You Do Comments
Data Review by Patient History for Action Range:
Please refer to page 5-2 of the Data Review section for the step-by-step instructions to access the Data Review menu.
Note: If no information is stored for the desired category that was selected, the following message will be displayed on the monitor: “No Stored Tests”.
Note: If no tests match the criteria specified, this message will be displayed on the monitor: “No Tests Match Request”.
If the monitor displays a message not indicated in these procedures, please refer to section 10, Troubleshooting.
1. Press 1 – Patient
History.
The Patient History menu now appears.
2. Press 4 – Action Range. The monitor displays the
most recent test result that is outside the action range.
Page 37
5-10
What You See on the Display
What You Do Comments
3. The operator can select
one of the following options:
Press 1 – Previous, view the previous test result.
Press 2 – Next, view the next test result.
The number that appears to the right of 1-Previous and 2-Next on the monitor indicates the number of tests available for review.
The display also shows: p: Patient ID o: Operator ID 12 : Comment Code ven: Venous Sample
Indicates above action range
Indicates below action range
4. When finished
reviewing the data, the operator can:
Press Clear to return to the Patient History menu.
Press Menu to return to the Menu Mode menu.
Press On/Off to turn off the monitor.
(20.1 mmol/L)
Page 38
Questions? Refer to Page 10-13 For Customer Care Information.
5-11
What You See on the Display
What You Do Comments
Data Review for Patient History by Comment Code:
Please refer to page 5-2 of the Data Review section for the step-by-step instructions to access the Data Review menu.
Note: If no information is stored for the desired category that was selected, the following message will be displayed on the monitor: “No Stored Tests”.
Note: If no tests match the criteria specified, this message will be displayed on the monitor: “No Tests Match Request”.
If the monitor displays a message not indicated in these procedures, please refer to the section 10, Troubleshooting.
1. Press 1 – Patient
History.
The Patient History menu now appears.
2. Press 5 – Comment
Code.
The monitor will prompt for the comment code to be scanned or entered.
3. Scan or manually enter
the desired comment code via the keypad, then press Enter.
The monitor will show the most recent test result for the selected comment code.
Page 39
5-12
What You See on the Display
What You Do Comments
4. The operator can select
one of the following options:
Press 1 – Previous, view the previous test result.
Press 2 – Next, view the next test result.
The number that appears to the right of 1-Previous and 2-Next on the monitor indicates the number of tests available for review. The display also shows:
p: Patient ID o: Operator ID 12 : Comment Code ven: Venous Sample
5. When finished
reviewing the data, the operator can:
Press Clear to return to the Comment Code screen.
Press Menu to return to the Menu Mode menu.
Press On/Off to turn off the monitor.
(5.8 mmol/L)
Page 40
Questions? Refer to Page 10-13 For Customer Care Information.
5-13
What You See on the Display
What You Do Comments
Data Review for Patient History by Selected Date:
Please refer to page 5-2 of the Data Review section for the step-by-step instructions to access the Data Review menu.
Note: If no information is stored for the desired category that was selected, the following message will be displayed on the monitor: “No Stored Tests”.
Note: If no tests match the criteria specified, this message will be displayed on the monitor: “No Tests Match Request”.
If the monitor displays a message not indicated in these procedures, please refer to the section 10, Troubleshooting.
1. Press 1 – Patient
History.
The Patient History menu now appears.
2. Press 6 – Date. The monitor will now
prompt for the desired date to be entered.
3. Enter the desired date
for review then press
Enter.
The operator will enter in the date using two digits for month/day/year. For example, June 8, 2001 becomes 06/08/01.
Page 41
5-14
What You See on the Display
What You Do Comments
4. The operator can select
one of the following options:
Press 1 – Previous, view the previous test result.
Press 2 – Next, view the next test result.
The number that appears to the right of 1-Previous and 2-Next on the monitor indicates the number of tests available for review.
The monitor also shows: p: Patient ID o: Operator ID 12 : Comment Code ven: Venous Sample
5. When finished
reviewing the data, the operator can:
Press Clear to return to the Enter Date screen.
Press Menu to return to the Menu Mode menu.
Press On/Off to turn off the monitor.
(5.8 mmol/L)
Page 42
Control History
The Precision PCx monitor can store up to 1,000 control tests that can be retrieved by the following categories:
All Tests: This shows all control test results in memory starting with
the most recent test result in memory (reverse chronological order).
Operator ID: This shows the control test results that were performed
by a particular operator, starting with the most recent test.
Outside Range: This shows all the control test results that fell
outside of the acceptable ranges.
Comment Code: This shows all the control test results by comment
code.
Date: This shows the control test results performed on a specific
date, starting with the most recent test of the day.
Data Review for Control History for All Tests:
Please refer to page 5-2 of the Data Review section for the step-by-step instructions to access the Data Review menu.
Note: If no information is stored for the desired category that was selected, the following message will be displayed on the monitor: “No Stored Tests”.
Note: If no tests match the criteria specified, this message will be displayed on the monitor: “No Tests Match Request”.
If the monitor displays a message not indicated in these procedures, please refer to section 10, Troubleshooting.
Questions? Refer to Page 10-13 For Customer Care Information.
5-15
What You See on the Display
What You Do Comments
1. Press 2 – Control
History.
The Control History menu now appears.
2. Press 1 – All Tests. The monitor shows the
result of the most recent control test.
Page 43
5-16
What You See on the Display
What You Do Comments
3. The operator can select
one of the following options:
Press 1 – Previous, view the previous test result.
Press 2 – Next, view the next test result.
The number that appears to the right of 1-Previous and 2-Next on the monitor indicates the number of tests available for review.
The display also shows the acceptable range for this test and:
o: Operator ID 0 : Comment Code
4. When finished
reviewing the data, the operator can:
Press Clear to return to the Control History menu.
Press Menu to return to the Menu Mode menu.
Press On/Off to turn off the monitor.
(2.3 mmol/L)
Page 44
Data Review for Control History by Operator ID:
Please refer to page 5-2 of the Data Review section for the step-by-step instructions to access the Data Review menu.
Note: If no tests are stored for the desired category that was selected, the following message will be displayed on the monitor: “No Stored Tests”.
Note: If no tests match the criteria specified, this message will be displayed on the monitor: “No Tests Match Request”.
If the monitor displays a message not indicated in these procedures, please refer to section 10, Troubleshooting.
Questions? Refer to Page 10-13 For Customer Care Information.
5-17
What You See on the Display
What You Do Comments
1. Press 2 – Control
History.
The Control Test History menu now appears.
2. Press 2 – Operator ID. The monitor will prompt
for the Operator ID to be scanned or entered.
3. Scan or manually enter
the Operator ID via the keypad, then press
Enter.
The monitor will display the Operator's most recent control test result.
Page 45
5-18
What You See on the Display
What You Do Comments
4. The operator can select
one of the following options:
Press 1 – Previous, view the previous test result.
Press 2 – Next, view the next test result.
The number that appears to the right of 1-Previous and 2-Next on the monitor indicates the number of tests available for review.
The display also shows the acceptable range for this test and:
o: Operator ID 0 : Comment Code
5. When finished
reviewing the data, the operator can:
Press Clear to return to the Operator ID screen.
Press Menu to return to the Menu Mode menu.
Press On/Off to turn off the monitor.
(2.3 mmol/L)
Page 46
Questions? Refer to Page 10-13 For Customer Care Information.
5-19
Data Review for Control History by Outside Range Results:
Please refer to page 5-2 of the Data Review section for the step-by-step instructions to access the Data Review menu.
Note: If no tests are stored for the desired category that was selected, the following message will be displayed on the monitor: “No Stored Tests”.
Note: If no tests match the criteria specified, this message will be displayed on the monitor: “No Tests Match Request”.
If the monitor displays a message not indicated in these procedures, please refer to section 10, Troubleshooting.
What You See on the Display
What You Do Comments
1. Press 2 – Control
History.
The Control Test History menu now appears.
2. Press 3 – Outside
Range.
The monitor shows the most recent control test result that is outside the acceptable range.
Page 47
5-20
What You See on the Display
What You Do Comments
4. When finished
reviewing the data, the operator can:
Press Clear to return to the Control History menu.
Press Menu to return to the Menu Mode menu.
Press On/Off to turn off the monitor.
3. The operator can select
one of the following options:
Press 1 – Previous, view the previous test result.
Press 2 – Next, view the next test result.
The number that appears to the right of 1-Previous and 2-Next on the monitor indicates the number of tests available for review.
The display also shows the acceptable range for this test as well as:
o: Operator ID 2 : Comment Code X: Outside Range
(X3.5 mmol/L)
Page 48
Questions? Refer to Page 10-13 For Customer Care Information.
5-21
Data Review for Control History by Comment Code:
Please refer to page 5-2 of the Data Review section for the step-by-step instructions to access the Data Review menu.
Note: If no tests are stored for the desired category that was selected, the following message will be displayed on the monitor: “No Stored Tests.”
Note: If no tests match the criteria specified, this message will be displayed on the monitor: “No Tests Match Request”.
If the monitor displays a message not indicated in these procedures, please refer to section 10, Troubleshooting.
What You See on the Display
What You Do Comments
1. Press 2 – Control
History.
The Control Test History menu now appears.
2. Press 4 – Comment
Code.
The monitor will prompt for the comment code to be scanned or entered.
3. Scan or manually enter
the comment code via the keypad, then press
Enter.
The monitor will display the result of the most recent control test for the specified comment code.
Page 49
5-22
What You See on the Display
What You Do Comments
4. The operator can select
one of the following options:
Press 1 – Previous, view the previous test result.
Press 2 – Next, view the next test result.
The number that appears to the right of 1-Previous and 2-Next on the monitor indicates the number of tests available for review.
The display also shows the acceptable range for this test and:
o : Operator ID 2 : Comment Code X: Outside Range
5. When finished
reviewing the data, the operator can:
Press Clear to return to the Comment Code screen.
Press Menu to return to the Menu Mode menu.
Press On/Off to turn off the monitor.
(X3.5 mmol/L)
Page 50
Questions? Refer to Page 10-13 For Customer Care Information.
5-23
Data Review for Control History for Selected Date:
Please refer to page 5-2 of the Data Review section for the step-by-step instructions to access the Data Review menu.
Note: If no tests are stored for the desired category that was selected, the following message will be displayed on the monitor: “No Stored Tests”.
Note: If no tests match the criteria specified, this message will be displayed on the monitor: “No Tests Match Request”.
If the monitor displays a message not indicated in these procedures, please refer to section 10, Troubleshooting.
What You See on the Display
What You Do Comments
1. Press 2 – Control
History.
The Control Test History menu now appears.
2. Press 5 – Date. The monitor will prompt
for a specified date to be entered.
3. Enter a date then press Enter.
The monitor will display the most recent control test result for the specified date.
Page 51
5-24
What You See on the Display
What You Do Comments
4. The operator can select
one of the following options:
Press 1 – Previous, view the previous test result.
Press 2 – Next, view the next test result.
The number that appears to the right of 1-Previous and 2-Next on the monitor indicates the number of tests available for review.
The display also shows the acceptable range for this test and:
o : Operator ID 0 : Comment Code
5. When finished
reviewing the data, the operator can:
Press Clear to return to the Enter Date screen.
Press Menu to return to the Menu Mode menu.
Press On/Off to turn off the monitor.
(2.3 mmol/L)
Page 52
Questions? Refer to Page 10-13 For Customer Care Information.
5-25
What You See on the Display
What You Do Comments
Data Review for Proficiency Data
The monitor stores the last 60 proficiency test results. Please refer to the beginning of the Data Review for the step-by-step
instructions to access the Data Review menu.
Note: If no tests are stored for the desired category that was selected, the following message will be displayed on the monitor: “No Stored Tests”.
If the monitor displays a message not indicated in these procedures, please refer to section 10, Troubleshooting.
1. Press 3 – Proficiency
Data.
The monitor now shows the most recent proficiency test result.
2. The operator can select
one of the following options:
Press 1 – Previous, view the previous test result.
Press 2 – Next, view the next test result.
3. When finished
reviewing the data, the operator can:
Press Clear to return to the Data Review screen.
Press Menu to return to the Menu Mode menu.
Press On/Off to turn off the monitor.
The number that appears to the right of 1-Previous and 2-Next on the monitor indicates the number of tests available for review.
The display shows the following:
s : Sample Type or ID o : Operator ID 0 : Comment Code
(2.5 mmol/L)
Page 53
5-26
What You See on the Display
What You Do Comments
Data Review for Linearity Data
The monitor stores the last linearity panel that includes a maximum of 4 replicates for 5 levels (20 tests).*
Please refer to the beginning of the Data Review section for the step­by-step instructions to access the Data Review menu.
Note: If no tests are stored for the desired category that was selected, the following message will be displayed on the monitor: “No Stored Tests”.
If the monitor displays a message not indicated in these procedures, please refer to section 10, Troubleshooting.
1. Press 4 – Linearity
Data.
The monitor now shows the most recent Linearity Test.
At the top of the display the test level is indicated. The Level indicates Level 1-5 while Test indicates 1-4 replicates at that level.
2. The operator can select
one of the following options:
Press 1 – Previous, view the previous test result.
Press 2 – Next, view the next test result.
3. When finished
reviewing the data, the operator can:
Press Clear to return to the Data Review screen.
Press Menu to return to the Menu Mode menu.
Press On/Off to turn off the monitor.
The number that appears to the right of 1-Previous and 2-Next on the monitor indicates the number of tests available for review.
The display also shows the following:
o: Operator ID
(15.3 mmol/L)
*It is outside the intended use of the PCx/QCM system to perform a test in linearity mode on a PCx
instrument and change the test result record to a patient result in QCM Data Management System. Linearity test results outside the test strip assay range should only be used to set linearity performance on the PCx instrument. The performance of the test strips has not been verified outside the assay range listed on the test strip instructions for use.
Page 54
The Proficiency Test menu enables the operator to perform unknown sample test challenges if required by facility policy.
Use the following procedures to perform a proficiency test. If the monitor displays a message not indicated in these procedures,
please refer to section 10, T roubleshooting.
6-1
Questions? Refer to Page 10-13 For Customer Care Information.
Proficiency Test 6
What You See on the Display
What You Do Comments
1. Press the On/Off
button to turn on the monitor.
The Abbott logo screen appears for a few seconds, followed by the product name, the software version screen, then the Test Mode menu.
2. Press Menu. The Menu Mode screen
now appears.
3. Press 2 – Proficiency
T est.
The monitor will prompt for the Operator ID to be scanned or entered.
4. Scan or manually enter
the Operator ID via the keypad, then press
Enter.
The monitor will prompt for the Sample ID to be scanned or entered.
5. Scan or manually enter
the Sample ID via the keypad, then press
Enter.
The monitor will prompt for the Strip Lot number to be scanned or entered.
Page 55
What You See on the Display
What You Do Comments
6-2
6. Scan or manually
enter* the test strip lot number via the keypad, then press Enter.
The monitor will prompt the operator to insert the test strip.
7. Open the foil test strip
packet at the notch and tear up or down to remove the test strip.
8. With the contact bars
facing up, insert the test strip into the test strip port until it stops.
The monitor will prompt the operator to apply the sample to the test strip.
* Manual entry of strip lots with alpha-numeric barcodes is not supported.
Precision PCx PlusT est Strip with TrueMeasure T echnology For Use with Precision PCx v2.2 Only
Precision PCx T est Strip For Use with Precision PCx v2.0,v2.1 and v2.2.
Page 56
Questions? Refer to Page 10-13 For Customer Care Information.
What You See on the Display
What You Do Comments
The test starts automatically, as the sample is accepted.
10. Wait for the monitor
to analyze the sample and display the test result.
The monitor counts down 20 seconds then displays the test result.
11. If the operator is
prompted to enter a comment code, continue to step 12; otherwise skip forward to step 13.
12. Scan or manually
enter the comment code via the keypad, then press Enter.
The monitor returns to the Results menu. The comment code number that was entered will appear in a box to the left of the test result.
13. The operator will
select one of the following options:
Press 1 – Next Test. Press Menu to return to
the Menu Mode. Press On/Off to turn off
the monitor.
The operator can refer back to step 5 for the steps to perform any additional proficiency testing at this time.
(14.6 mmol/L)
(14.6 mmol/L)
(14.6 mmol/L)
9. Gently invert the
sample bottle 3-4 times then apply a drop of the sample to the target area on the test strip.
The monitor beeps when the sample is accepted. The Sample Accepted Screen will be displayed.
6-3
Page 57
The Linearity Test menu enables the operator to verify method linearity using a Linearity Assessment Kit.*
Please use the following procedures to perform a Linearity Test. If the monitor displays a message not indicated in these procedures,
please refer to section 10, Troubleshooting.
Questions? Refer to Page 10-13 For Customer Care Information.
1. Press On/Off to turn on monitor.
The Abbott logo screen will appear for a few seconds, followed by the product name, the software version, then the Test Mode menu.
2. Press Menu. The Menu Mode menu
will appear.
3. Press 3 – Linearity T est.
The monitor will prompt for the Operator ID to be scanned or entered.
4. Press Scan or manually enter the Operator ID via the keypad, then press Enter .
The monitor will next prompt for the Linearity Kit lot number to be scanned or entered.
What You See on the Display
What You Do Comments
*It is outside the intended use of the PCx/QCM system to perform a test in linearity mode on a PCx
instrument and change the test result record to a patient result in QCM Data Management System. Linearity test results outside the test strip assay range should only be used to set linearity performance on the PCx instrument. The performance of the test strips has not been verified outside the assay range listed on the test strip instructions for use.
7-1
Linearity Test 7
Page 58
What You See on the Display
What You Do Comments
7-2
5. Scan or manually enter
the linearity kit lot number via the keypad, then press Enter. If the monitor prompts for a level to be entered, skip to step 7.
If the linearity kit lot number that was entered is different from the previous one, the monitor will prompt the operator to do one of two actions: either re-enter the existing linearity kit lot number or replace the existing linearity kit lot number with the new linearity kit lot number.
6. The operator can
perform one of the following actions:
Press 1 – ReEnter Kit Lot.
Press 2 – Replace Panel.
If 1 is pressed, the monitor will return to the previous screen so the operator can scan or enter the linearity kit lot number.
The monitor will prompt the operator to select a level.
7. Enter the number of
the level of the next test to be run.
There can be up to 4 replicates per level. The number in the parenthesis indicates the number of replicates already run for that level. When a level is full, the menu item (1­Level 1) will no longer appear.
8. If the operator presses
6 for a new panel, the monitor will again prompt the operator to select a level.
The monitor will prompt the operator to scan or manually enter the test strip lot number.
Page 59
Questions? Refer to Page 10-13 For Customer Care Information.
What You See on the Display
What You Do Comments
9. Scan or manually enter* the test strip lot number via the keypad, then press Enter.
The monitor will prompt for a new test strip to be inserted into the test strip port.
10. Open the foil test strip
packet at the notch and tear up or down to open the packet, then remove the test strip.
11. With the contact bars
facing up, insert the test strip into the test strip port until it stops.
The monitor will prompt the operator to apply the sample to the test strip.
* Manual entry of strip lots with alpha-numeric barcodes is not supported.
Precision PCx PlusT est Strip with TrueMeasure T echnology For Use with Precision PCx v2.2 Only
Precision PCx T est Strip For Use with Precision PCx v2.0,v2.1 and v2.2.
7-3
Page 60
What You See on the Display
What You Do Comments
7-4
15. Press Scan or
manually enter the comment code via the keypad then press Enter .
The monitor will return to the results screen. The comment code number that was entered will appear in a box to the left of where the result is displayed.
16. The operator can do
one of the following options:
Press 1 – Next Level. Press 2 – Same Level. Press the Menu button to
return to the Menu Mode. Press On/Off to turn off
monitor.
The operator can press 2 to perform more tests for the same level (return to step 9) or press 1 to perform tests for another level (return to step 8).
14. Note the result. If the
operator is prompted to enter a comment code, continue to step 15, otherwise skip to step 16.
The display includes the level and test numbers, date and time. Level indicates Level 1-5. Test indicates replicates 1-4 for that level.
The test starts automatically, as the sample is accepted.
(14.6 mmol/L)
(14.6 mmol/L)
(14.6 mmol/L)
13. Wait for the monitor
to analyze the sample and display the test result.
The monitor counts down 20 seconds then displays the test result.
12. Gently invert the
sample 3-4 times then apply a drop of the sample to the target area on the test strip.
The monitor beeps when the sample is accepted. The Sample Accepted Screen will appear.
Page 61
The Precision PCx Monitor provides setup options to customize the testing procedures and requirements. These standards promote good laboratory practices for operators.
Since the steps to review or change setup options are similar, the two sub-menus are combined in one section. Any operator can review the setup information; only an administrator (manager) with the password can change the setup information.
The Review Setup feature enables an operator to review only the settings for Test Setup, System Setup, System Status and Date and Time.
The Change Setup feature enables an administrator to review or change the settings for Test Setup, System Setup, System Status, Date and Time, and Edit Lists.
The Test Setup menu includes: All Tests, Patient Test, Control Test, Security and Test Type.
The System Setup menu defines the look or appearance of the data as it is displayed on the monitor, such as language and formats. The menu includes: language, beeper on/off, date format, time format, set password and restore defaults.
The System Status menu allows a review of the current status of the monitor's operating system: battery power, temperature, memory gauge and software version.
The Date and Time menu displays the current date and time. The Edit Lists menu changes the approved Operator ID or Test Strip
Lot lists. The Precision PCx Menu Tree on the following page will outline the
above menus and the sub-menus.
Questions? Refer to Page 10-13 For Customer Care Information.
8-1
Review / Change Setup
8
Page 62
8-2
Precision PCx Menu Tree
Test Mode
Patient
Test
Control
Test
Menu Mode
1. Data Review
2. Proficiency Test
3. Linearity Test
4. Review / 5.Change Setup
Upload
Inter
val
Operator ID Format Patient ID Format
Comment Code Memory Full
Results Format Comment Code Memory Full
QC Interval Operator ID Strip Lot
Glucose
Units Sample Type Action Range
1. Patient History sort by:
2. Control History
3. Proficiency Data
4. Linearity Data
1. Test Setup
2. System Setup
3. System Status
4. Date and Time
All T ests Patient ID Operator ID Action Range Comment Code Date
All T ests Operator ID Outside Range Comment Code Date
1. All Tests
2. Patient Test
3. Control Test
4. Security
5. Test Type
1. Operator List
5. Edit Lists
Add Delete Change
Add Delete Change
sort by:
2. Strip Lot List
1.
2.
Battery Power Temperature Memory Gauge Software Version
Language Beeper On/Off Date Format Time Format Password Factory Settings
Levels
Page 63
Moving Around the Review/Change Setup Menu
The Review and Change Setup menu options offer the same additional sub-menus (as shown in the Precision PCx Menu Tree). The operators may have to select additional options to access the desired change or review option.
To access these sub-menus or options in both the Review and Change Setup menus, the message 1-Next appears at the bottom of the display. The operator presses 1 to move to the next option. The monitor also allows the operator to go back to the previous option by pressing Clear. The current status or selection for the option appears in bold type on the display .
To change an option, the administrator must access the Change Setup menu (this is password protected). In addition to the 1-Next message that appears at the bottom of the option screen, there is the 2-Change message. The administrator presses 2 to change the option. The active selection for the option now appears in bold type.
When finished reviewing or changing the options, the operator or administrator can press Menu once to return to the beginning of Menu Mode or press Menu twice to go to Test Mode.
In this section the majority of the directions refer specifically to how to change various settings. However, these same directions can be followed as well for reviewing settings under its own menu option.
Differentiation is made between the password protected sections in the Change Setup Menu Option, where only administrators can access the screens, and the non-password protected sections in the Review Setup Menu Option, where operators and administrators can access the screens.
Questions? Refer to Page 10-13 For Customer Care Information.
8-3
Page 64
What You See on the Display
What You Do Comments
8-4
Review Setup:
If the monitor displays a message not indicated in these procedures, please refer to section 10, Troubleshooting.
Use the following steps to access the Review Setup menu:
1. Press On/Off to turn
on the monitor.
The Abbott logo screen will appear for a few seconds followed by the product name, the software version screen and then the Test Mode menu.
2. Press Menu. The Menu Mode menu
will appear.
3. Press 4 for Review
Setup.
The Review Setup Menu will appear.
The operator can review Test Setup, System Setup, System Status and Date and Time by following the appropriate steps for these procedures.
Page 65
Questions? Refer to Page 10-13 For Customer Care Information.
8-5
What You See on the Display
What You Do Comments
Change Setup:
If the monitor displays a message not indicated in these procedures, please refer to section 10, Troubleshooting.
Use these steps to access the Change Setup menu:
1. Press On/Off to turn on
the monitor.
The Abbott logo screen will appear for a few seconds followed by the product name, the software version screen and then the Test Mode menu.
3. Press 5 for Change
Setup.
The display prompts the operator to enter a password of up to 8 characters. The password is displayed as asterisks on the monitor.
4. Key in the password, then press Enter. If the password is entered incorrectly, the monitor displays error message: “Incorrect Password.” Press Clear twice, verify the password then re-enter the correct password and press
Enter.
The factory set password is 123. Once in the Change Setup Mode, it is suggested that the administrator change the password. This can be done in the System Setup/Set Password option.
2. Press Menu. The Menu Mode menu
will appear.
Page 66
What You See on the Display
What You Do Comments
8-6
The administrator can change Test Setup, System Setup, System Status, Date and Time and Edit Lists by following the appropriate steps for these procedures.
Page 67
Questions? Refer to Page 10-13 For Customer Care Information.
8-7
Review/Change Setup for Test Setup
The Review/Change Setup menus enable the operator to review/change the test settings.
Under the Test Setup Menu, the following options can be viewed or changed:
• All Tests
• Patient Test
• Control Test
• Security
• Test Type In the All Tests menu, the following options can be viewed or changed:
• Upload Interval for sending stored data to other systems.
• Establish Operator ID requirements and barcode format. In the Patient Test menu, the following options can be viewed or
changed:
• Establish Patient ID requirements and barcode format.
• Set the requirement for out-of-range and/or in-range comment codes.
• Management of memory full overwrite. In the Control Test menu, the following options can be viewed or
changed:
• Enable high, normal and/or low control testing.
• Control test results presented as numeric or Pass/Fail.
• Set the requirement for in-range and/or out-of-range comment codes.
• Management of memory full overwrite. In the Security menu, the following options can be viewed or changed:
• Set QC interval requirements.
• Set requirements to prevent testing by non-certified operators.
• Set requirements to prevent testing using test strip lots that have not been qualified for use.
In the Test Type menu, the following options can be viewed or changed:
• Set units of measurement.
• Set sample type prompts.
• Set action ranges. The action range can also be changed in QC Manager. Refer to the QC Manager Operator’s Manual for more information. Note: Default action range is 20-600.
Page 68
8-8
Review/Change Test Setup for All Tests:
The options that are associated with the Review/Change Test Setup for All Tests apply to all test types.
Use this menu to review or change these options for all tests.
Upload Interval
The monitor can be set up to warn or lockout a task if the monitor has not been uploaded at a regularly scheduled interval. The upload interval options are:
• Allow Test–allows the testing to continue. No message will appear on the monitor.
• Warn–allows the testing to continue but the operator will receive a message that the upload interval (as specified below) has expired.
• Lockout–does not allow any testing until the data has been uploaded. This option is set to designate a required maximum number of hours or days (from 1 to 99) between data uploads. When a task is selected and the upload interval has expired, the operator will receive a message that the Upload Interval has expired. If Upload Interval is expired, the monitor will prevent an operator from doing other tasks until the stored data is uploaded.
Operator ID
The Operator ID menu has the following options which can be viewed or changed:
• Disabled, Optional or Required entry of an Operator ID for all testing.
• Require a minimum and a maximum number of digits for the length of an Operator ID.
• Allow any combination of five common barcode formats used in the facility:
I2 of 5 (with or without check digit) Code 39 (with or without check digit, alphanumeric or
ASCII format) Code 128 Codabar Code 93
• Truncate (ignore) leading and/or trailing barcode digits.
Note: The display shows selected options in bold type.
Page 69
Questions? Refer to Page 10-13 For Customer Care Information.
8-9
What You See on the Display
What You Do Comments
1. Press 1 for Test Setup. The display shows the
Test Setup menu.
2. Press 1 for All Tests. The display shows the
Upload Interval menu.
3. The administrator will
select one of the following options:
Press 1 – Next; skip forward to step 8.
Press 2 – Change.
The display shows the current settings in bold. If 2 is pressed, the display shows the Upload Interval options.
4. The administrator will
select one of the following options:
Press 1 – Allow T est; skip forward to step 7.
Press 2 – Warn. Press 3 – Lockout.
If 2 or 3 is pressed, the display will show the upload interval hours and days option.
5. The administrator will
select one of the following options:
Press 1 – Hours. Press 2 – Days.
The display prompts for the number of days/hours to be entered.
Review/Change Setup for All Tests:
Refer to the beginning of the Review/Change Setup Section for step-by­step instruction to access the appropriate setup menu.
If the monitor displays a message not indicated in these procedures, please refer to section 10, Troubleshooting.
Page 70
What You See on the Display
What You Do Comments
8-10
6. Enter 1-99 hours or
1-99 days then press
Enter.
The All Tests Upload Interval screen will be displayed.
7. Press 1 to continue to
the All Tests Operator ID screen.
The display shows the All Tests Operator ID menu.
8. The administrator will
select one of the following options:
Press 1 – Next, return to the Test Setup menu.
Press 2 – Change.
The display shows the Operator ID options.
9. The administrator will
select one of the following options:
Press 1 – Disabled; skip forward to step 13.
Press 2 – Optional. Press 3 – Required.
If Disabled is selected, the display will not prompt for Operator ID. If Optional or Required is selected, the administrator can enter a minimum and maximum number of digits, between 1-15 digits.
10. The administrator will
enter minimum and maximum numbers between 1-15.
Key in the minimum number of digits for an operator ID then press
Enter.
Key in the maximum number of digits for an operator ID then press
Enter.
The display shows the Operator ID Barcode options.
Page 71
Questions? Refer to Page 10-13 For Customer Care Information.
8-11
What You See on the Display
What You Do Comments
11. The administrator may customize I2 of 5 by pressing 1 if I2 of 5 is bolded (enabled). Press 1 again to access the Operator ID Barcode screen.
Activated barcodes appear in bold type. More than one Barcode format can be active at one time. Any Barcode can be deactivated by selecting the corresponding number.
Note: Only Barcodes 1 and 2 can be customized.
12. The administrator will
select one of the following:
1. No Check Digit
2. USS Check Digit
3. OPCC Check Digit
The display returns to the Operator ID Barcode screen.
USS or OPCC will appear after the format, if selected.
13. The administrator may customize Code 39 by pressing 2 if Code 39 is bolded (enabled). Press 2 again to access the Operator ID Barcode screen.
14. The administrator will select one of the following:
1. No Check Digit
2. Use Check Digit
The display shows the Alphanumeric, Full ASCII screen.
15. The administrator will select one of the following:
1. Alphanumeric
2. Full ASCII
The display returns to the Operator ID Barcode screen.
C will appear after the format if Check Digit is selected.
F will appear if Full ASCII is selected.
Page 72
8-12
17. Press Enter twice
for no truncation of leading or trailing digits; otherwise enter the number of leading and trailing barcode digits to truncate (ignore).
The monitor can be set to ignore a specific number of leading and/or trailing barcode digits.
The display returns to the All Tests, Operator ID screen.
18. The administrator will
select one of the following options:
Press 1 – Next, return to the Test Setup menu.
Press Menu to return to the Menu Mode.
Press On/Off to turn off the monitor.
What You See on the Display
What You Do Comments
16. To access the Truncate Digits screen, press 6 and continue forward to step 17.
The display shows the Operator ID Barcode Truncate Digits screen.
Page 73
Questions? Refer to Page 10-13 For Customer Care Information.
8-13
Review/Change Test Setup for Patient Test:
The options set with this menu will apply to all Patient Tests. The Patient ID option uses a patient identification number to perform
tests. Patient ID uses the following options:
• Disable, Require or make Optional the entry of a Patient ID.
• Require a minimum and maximum number of digits for the length of the Patient ID.
• Allow any combination of five common barcode formats used in the facility.
I2 of 5 (with or without check digit) Code 39 (with or without check digit, alphanumeric or
ASCII format) Code 128 Codabar Code 93
• Truncate (ignore) leading and/or trailing barcode digits.
Comment Code option allows the following to be selected:
• Disable the prompt for a Comment Code.
• Prompt for a Comment Code with the option of not requiring it.
• Require a Comment Code for every test result out-of-range and/or in-range of the action range. (See Test Type setup options).
The Memory Full option will notify the operator that the patient test memory has reached its limit (4000 tests that have not been uploaded). Memory Full provides the following options:
• Lockout additional tests until the data is uploaded.
• Prompt the operator to upload data before overwriting old test results with new ones.
• Automatically overwrite oldest test results with new ones.
Page 74
8-14
Review/Change Test Setup for Patient Test:
Refer to the beginning of the Review or Change Setup section for step-by-step instructions to access the appropriate setup menu.
If the monitor displays a message not indicated in these procedures, please refer to section 10, Troubleshooting.
What You See on the Display
What You Do Comments
1. Press 1 – Test Setup. The display shows the
Test Setup menu.
2. Press 2 – Patient Test. The display shows the
Patient Test, Patient ID Barcode screen.
3. The administrator will
select one of the following options:
Press 1 –
Next; skip
forward to step 14. Press 2
Change; to
change Patient ID.
The display shows the Patient ID options.
4. The administrator will
select one of the following options:
Press 1 –
Disabled; (returns to the previous screen).
Press 2
Optional.
Press 3 – Required.
If Disabled is selected, the display will not prompt for Patient ID. If Optional or Required is selected, the operator can enter a minimum and maximum number of digits, between 1-15 digits.
Page 75
Questions? Refer to Page 10-13 For Customer Care Information.
8-15
What You See on the Display
What You Do Comments
5. The administrator will
enter minimum and maximum numbers between 1-15.
Key in the minimum number of digits for a Patient ID number then press Enter .
Key in the maximum number of digits for a Patient ID number then press Enter .
The display returns to the Patient Test, Patient ID Barcode Menu.
6. The administrator may customize I2 of 5 by pressing 1 if I2 of 5 is
bolded (enabled). Press 1 again to access the
Patient ID Barcode screen.
Activated barcodes appear in bold type. More than one Barcode format can be active at one time. Any Barcode can be deactivated by selecting the corresponding number.
Note: Only Barcodes 1 and 2 can be customized.
7. The administrator will
select one of the following:
1. No Check Digit
2. USS Check Digit
3. OPCC Check Digit
The display returns to the Patient ID Barcode screen.
USS or OPCC will appear after the format, if selected.
8. The administrator may customize Code 39 by pressing 2 if Code 39 is bolded (enabled). Press 2 again to access the Patient ID Barcode screen.
9. The administrator will
select one of the following:
1. No Check Digit
2. Use Check Digit
The display shows the Alphanumeric, Full ASCII screen.
Page 76
8-16
What You See on the Display
What You Do Comments
12. Press Enter twice
for no truncation of leading or trailing digits; otherwise enter the number of leading and trailing barcode digits to truncate (ignore).
The monitor can be set to ignore a specific number of leading and/or trailing barcode digits.
The display returns to the Patient Test, Patient ID Barcode screen.
13. Press 1 Next;
continue to the next menu.
The display shows the In­Range Comment Code screen.
14. The administrator will
select one of the following options:
Press 1 – Next, and continue to step 17.
Press 2 – Change, to change the In-Range Comment Code.
The display shows the In-Range Comment Code options. The Comment Code options can be set to Disabled, Optional or Required for test results that fall in the action range.
10. Select one of the
following:
1. Alphanumeric
2. Full ASCII
The display returns to the Patient ID Barcode screen.
C will appear after the format if Check Digit is selected.
F will appear if Full ASCII is selected.
11. To access the Truncate
Digits screen, press 6 and continue forward to step 12.
The display shows the Patient ID Barcode Truncate Digits screen.
Page 77
Questions? Refer to Page 10-13 For Customer Care Information.
8-17
15. The administrator will
select one of the following options:
Press 1 – Disabled. Press 2 – Optional. Press 3 – Required.
The display is returned to the Patient Test Comment Code If In-Range screen.
16. Press 1 Next;
continue to the next menu.
The display shows the Out-of-Range Comment Code menu. The comment code options can be set to Disabled, Optional or Required for test results that fall outside the action range.
What You See on the Display
What You Do Comments
17. The administrator will
select one of the following options: Press 1 –
Next, and
continue to step 20. Press 2
Change, to
change the Out-of­Range Comment Code.
The display shows the Out-of-Range Comment Code options.
18. The administrator will
select one of the following options:
Press 1 – Disabled. Press 2 – Optional. Press 3 – Required.
The display shows the Patient Test Comment Code If Out-of-Range screen.
Page 78
8-18
What You See on the Display
What You Do Comments
19. Press 1 – Next;
continue to the next menu.
The display shows the Memory Full screen.
20. The administrator will
select one of the following:
Press 1 – Next, and continue to next menu. Press 2 – Change, to change Memory Full option.
The display shows the Memory Full options. Once the Patient Test storage bank is full, the following can be set by the monitor: lockout additional tests; prompt for overwrite option; or automatically overwrite the oldest stored tests.
21. The administrator will
select one of the following options:
Press 1 – Allow Overwrite.
Press 2 – Prompt for Overwrite.
Press 3 – Lockout.
The display is returned to the Patient Test Memory Full screen.
22. The administrator will
select one of the following options:
Press 1 to return to Test Setup menu.
Press Menu to return to the Menu Mode.
Press On/Off to turn off the monitor.
Page 79
Questions? Refer to Page 10-13 For Customer Care Information.
8-19
Review/Change Test Setup for Control Tests
The options associated with this menu apply to all Control Tests. Control levels can be any combination of low, normal, and high.
At least one level must be selected. The Results Format has the following options:
Show test results in a Numeric or Pass/Fail format. Comment Code prompts have the following options:
• Disable the prompt for a comment code
• Prompt for a comment code with the option of not requiring it
• Require a comment code for every in-range and/or out-of-range test result.
The Memory Full option can notify the operator that the control test memory has reached its limit (1000 tests that have not been uploaded). Memory Full has the following options:
• Lockout additional tests until data is uploaded
• Prompt to overwrite old test results with new ones
• Automatically overwrite oldest test results with new ones.
Page 80
8-20
Review/Change Test Setup for Control Tests:
Refer to the beginning of the Review or Change Setup section for step­by-step instructions to access the appropriate setup menu.
If the monitor displays a message not indicated in these procedures, please refer to section 10, Troubleshooting.
What You See on the Display
What You Do Comments
1. Press 1 – Test Setup. The display shows the
Test Setup menu.
2. Press 3 – Control Test. The display shows the
Control Test Lev e l s screen. The monitor can be set to perform any combination of low, normal and high control tests.
3. The administrator will
select one of the following options:
Press 1 – Next, to review Control Test Result; skip forward to step 6.
Press 2 – Change, to change Control Test Levels option.
The display shows the Control Test Levels options.
4. The administrator will
select one or a combination of the following options:
Press 1 – Low, to activate low control.
Press 2 – Normal, to activate normal control.
Press 3 – High, to activate high control.
Press 4 – to return to Control Test Levels screen.
The activated control level(s) are in bold type.
Page 81
Questions? Refer to Page 10-13 For Customer Care Information.
8-21
What You See on the Display
What You Do Comments
5. Press 1 – Next, to
continue to the Control Test Result menu.
The display shows the Control Test Result option. The control test results can be set to appear as numeric or Pass/Fail.
6. The administrator will
select one of the following options:
Press 1 – Next, to review In-Range Comment Code; skip forward to step 9.
Press 2 – Change, to change the result to Numeric or Pass/Fail.
The display shows the Control Test Result menu.
Pass/Fail indicates whether the control test result is in-range or out­of-range.
7. The administrator will
select one of the following options:
Press 1 – Numeric. Press 2 – Pass/Fail.
The display is returned to the Control Test Result menu.
8. Press 1 – Next. The display shows the
Control Test Comment Code If In-Range screen.
9. The administrator will
select one of the following options:
Press 1 – Next, to continue to Control Test Comment Code If Out-of­Range menu. Skip forward to step 12.
Press 2 – Change, to change the In-Range Comment Code option.
The display shows the Comment Code If In­Range options.
Page 82
8-22
What You See on the Display
What You Do Comments
10. The administrator will
select one of the following options:
Press 1 – Disabled. Press 2 – Optional. Press 3 – Required.
The display returns to the Control Test Comment Code If In­Range screen.
11. Press 1 to continue to
the next option.
The display shows the Control Test Comment Code If Out-of-Range screen.
12. The administrator will
select one of the following options:
Press 1 – Next, to continue to Memory Full menu. Skip forward to step 15.
Press 2 – Change, to change the Comment Code If Out-of-Range options.
The display shows the Comment Code If Out­of-Range options.
13. The administrator will
select one of the following options:
Press 1 – Disabled. Press 2 – Optional. Press 3 – Required.
The display returns to the Control Test Comment Code If Out-of-Range screen.
14. Press 1 – Next, to
continue to the Memory Full menu.
The display shows the Control Test Memory Full options.
Page 83
Questions? Refer to Page 10-13 For Customer Care Information.
8-23
What You See on the Display
What You Do Comments
15. The administrator will
select one of the following options:
Press 1 – Next, return to the Test Setup menu.
Press 2 – Change, change the Memory Full options.
Allow Overwrite will automatically overwrite the existing data. Prompt for Overwrite will notify the operator that the data will be overwritten, giving the operator the choice to overwrite the data or not. Lockout will lockout the operator from overwriting any data until the data has been uploaded.
16. The administrator will
select one of the following options:
Press 1 – Allow Overwrite.
Press 2 – Prompt for Overwrite.
Press 3 – Lockout.
The display is returned to the Control Test Memory Full screen.
17. The administrator will
select one of the following options:
Press 1 – Next, return to Test Setup menu.
Press Menu to return to the Menu Mode.
Press On/Off to turn off the monitor.
Page 84
8-24
Review/Change Test Setup for Security
The Test Setup for Security menu enables the monitor to establish that certain criteria be met before a test can be performed.
• In all situations, the monitor can be set to Allow Test (allow the operator to test), Warn (tell the operator about the situation and still test) or Lockout (no testing can be done).
• QC Interval can require control tests to be run at certain time intervals. For example, every 12 hours, at specified times (7 a.m., 3 p.m. and 11 p.m.), or after a specified number of patient tests (1-250).
• The Operator ID can require that the operator ID is on the list of qualified/certified operators to continue.
• The Strip Lot (Test Strip Lot) can require that the test strip lot is on the list of qualified test strip lots to continue.
Page 85
Questions? Refer to Page 10-13 For Customer Care Information.
8-25
Review/Change Test Setup for Security:
Refer to the beginning of the Review or Change Setup section for the step-by-step instructions to access the appropriate setup menu.
If the monitor displays a message not indicated in these procedures, please refer to section 10, Troubleshooting.
What You See on the Display
What You Do Comments
1. Press 1 – Test Setup. The display shows the
Test Setup menu.
2. Press 4 – Security. The display shows the
Security QC Interval menu.
3. The administrator will
select one of the following options:
Press 1 – Next, to continue to Security Operator ID Not Certified screen and skip forward to step 10.
Press 2 – Change, to change the QC Interval Controls Expired option.
The display shows the options for the QC Interval Controls Expired menu. The options are to Allow Test, even if the QC interval has expired; Warn the operator that even though the QC interval has expired, the test may continue; or Lockout the system until the required control tests are performed.
Page 86
8-26
What You See on the Display
What You Do Comments
4. The administrator will
select one of the following options:
Press 1 – Allow T est. Press 2 – Warn. Press 3 – Lockout.
The display shows the QC Interval Security menu. The operator can choose between QC intervals. The settings are:
• Time Interval – QC Interval based on a set number of hours between control tests.
• Time of Day – choose up to three times during the day that QC Interval will take place.
• Number of Patient Tests – QC Interval based on the number of patient tests specified.
5. The administrator will
select one of the following options:
Press 1 – Time Interval: continue to step 6.
Press 2 – Time of Day: skip to step 7.
Press 3 – Patient Tests: skip to step 8.
QC Time Interval option, the Time of Day option, or the Number of Patient Tests option.
6. Enter a Time Interval
number between 1-99, then press Enter. Skip forward to step 9.
The display returns to the Security QC Interval menu.
7. Enter up to 3 times
of the day in chronological order, starting at midnight. Press Enter to confirm each time. Skip forward to step 9.
The display returns to the Security QC Interval menu.
Page 87
Questions? Refer to Page 10-13 For Customer Care Information.
8-27
What You See on the Display
What You Do Comments
10. The administrator will
select one of the following options:
Press 1 – Next, to continue; skip forward to step 13.
Press 2 – Change, to change Operator ID Security.
The display shows the Operator ID Not Certified options.
11. The administrator will
select one of the following options:
Press 1 – Allow T est. Press 2 – Warn. Press 3 – Lockout.
The display is returned to the Security Operator ID Not Certified screen.
12. Press 1 to continue to
the Security Strip Lot Not on List screen.
The display shows the Security Strip Lot Not on List screen.
9. Press 1 to continue to
the Security Operator ID Not Certified screen.
The display shows the Security Operator ID Not Certified screen.
8. Enter number of patient
tests between 1-250, then press Enter.
The display returns to the Security QC Interval menu.
Page 88
8-28
What You See on the Display
What You Do Comments
15. The administrator will
select one of the following options:
Press 1 to return to the Test Setup menu.
Press Menu to return to the Menu Mode.
Press On/Off to turn off the monitor.
14. The administrator will
select one of the following options:
Press 1 – Allow T est. Press 2 – Warn. Press 3 – Lockout.
The display returns to the Security Strip Lot Not on List screen.
13. The administrator will
select one of the following options:
Press 1 – Next, to return to the Test Setup menu.
Press 2 – Change, to change the Strip Lot Not on List options.
The display shows the Strip Lot Not on List options.
1.Allow Test – a test strip
lot not on the test strip lot list.
2.Warn – the test strip lot
is not on the test strip lot list (operator may continue with test).
3.Lockout – a test
strip lot that is not on the test strip lot list (operator may not continue).
Page 89
Review/Change Test Setup for Test Type
The options set with the Review/Change Test Setup for Test Type apply to different tests that may be performed, such as glucose. Use this menu to review or change these setups for Test Type:
Units:
Display test results in certain measurement units. The options are:
• mg/dL
• mmol/L
Sample T ype:
Distinguish between different types of blood samples used for patient tests. The options are:
• Prompt to select the blood sample type for each patient test.
• Default to Capillary or Arterial for all patient tests.
• Default to Venous for all patient tests.
Action Range:
Set the range (low and high) when action should be taken based on the patient result obtained. The options are:
• Disable the action range.
• Enable and set the action range.
Questions? Refer to Page 10-13 For Customer Care Information.
8-29
Page 90
8-30
What You See on the Display
What You Do Comments
Review/Change Test Setup for Test Type:
Refer to the beginning of the Review or Change Setup section for the step-by-step instructions to access the appropriate setup menu.
If the monitor displays a message not indicated in these procedures, please refer to section 10, Troubleshooting.
1. Press 1 – Test Setup. The display shows
the Test Setup menu.
2. Press 5 – Test Type. The display shows the
Test Setup, Test Type menu.
3. Press 1 to continue to
Glucose Units option.
The display shows the Glucose Units screen.
4. The administrator will
select one of the following options:
Press 1 – Next, to continue to Glucose Sample Type menu; skip forward to step 7.
Press 2 – Change, to change the test measurement units.
The display shows the Units options.
Page 91
Questions? Refer to Page 10-13 For Customer Care Information.
8-31
What You See on the Display
What You Do Comments
5. Choose one of the
following options:
Press 1 – mg/dL. Press 2 – mmol/L.
The display is returned to the Glucose Units screen.
6. Press 1 to continue
to the next menu.
The display shows the Glucose Sample Type screen.
7. The administrator will
select one of the following options:
Press 1 – Next, to continue to Glucose Action Range menu; skip forward to step 10.
Press 2 – Change, to change Sample Type prompt.
The display shows the Sample Type options.
8. The administrator will
select one of the following options:
Press 1 – Prompt. Press 2 – Capillary or
Arterial. Press 3 – Venous.
The monitor can be set to always prompt for the sample type or to always default to either Capillary or Arterial, or Venous blood samples.
The display is returned to the Glucose Sample Type screen.
9. Press 1 to continue to
the Glucose Action Range menu.
The display shows the Glucose Action Range screen.
Page 92
(1.1 mmol/L,
-27.8 mmol/L)
Precision PCx Plus Test Strip
8-32
What You See on the Display
What You Do Comments
10. The administrator will
select one of the following options:
Press 1 – Next, to return to Test Setup Test Type.
Press 2 – Change, to change action range.
If the action range is disabled, the display is returned to the Test Setup Test Type screen. If the action range is enabled, the display shows the action range.
The display shows the Glucose Action Range options.
11. The administrator will
select one of the following options:
Press 1 – Disabled. Skip forward to step 13.
Press 2 – Enabled.
The display shows the Action Range screen.
When the action range is enabled, the display shows an Action Range Low and High values screen. Enter a low action range value and a high action range value.
13. Press 1 – Next to
return to the Test Setup Test Type Screen.
12. Enter the low value and press Enter. Enter the high value and press Enter.
The display is returned to the Glucose Action Range screen.
Refer to test strip instructions for use for Action Range.
(1.1 mmol/L,
-33.3 mmol/L)
Precision PCx Test Strip
Page 93
Questions? Refer to Page 10-13 For Customer Care Information.
8-33
What You See on the Display
What You Do Comments
14. The administrator will
select one of the following options:
Press Clear to return to Test Setup menu.
Press Menu to return to the Menu Mode menu.
Press On/Off to turn off the monitor.
Page 94
8-34
Review/Change Setup for System Setup
System Setup enables the look or appearance of the data shown on the display to be customized. The monitor can be formatted to present information in a way that is more comfortable or familiar to work with. Use this menu to review or change the setup for these features:
Language
This option presents the data in English or another language.
Beeper
This option turns the monitor beeper on or off. (The monitor beeps to alert the operator of certain functions).
Date Format
This option presents the date in mm/dd/yy, dd/mm/yy, or yy/mm/dd format. The user can select the punctuation that separates month, day and year.
Time Format
This option presents time in a 12 or 24-hour clock and separates numbers with different punctuation.
Set Password
This option allows the password to be changed.
Restore Defaults
This option restores the monitor to factory settings.
Page 95
Questions? Refer to Page 10-13 For Customer Care Information.
8-35
What You See on the Display
What You Do Comments
Review/Change Setup for System Setup:
Refer to the beginning of the Review or Change Setup section for the step-by-step instructions to access the appropriate setup menu.
If the monitor displays a message not indicated in these procedures, please refer to section 10, Troubleshooting.
1. Press 2 – System Setup. The display shows the
System Setup Language menu.
2. The administrator will
select one of the following:
Press 1 – Next, to continue to System Setup Beeper menu; skip forward to step 5.
Press 2 – Change, to change the language.
The display shows the Language setup options.
3. The administrator will
select the appropriate language.
The display returns to the System Setup Language menu.
4. Press 1 – Next, to go to
System Setup Beeper.
The display shows the System Setup Beeper menu.
5. The administrator will
select one of the following options:
Press 1 – Next, to go to System Setup Date Format menu; skip forward to step 8.
Press 2 – Change, to change the beeper setting.
The display shows the Beeper option.
Page 96
8-36
What You See on the Display
What You Do Comments
6. The administrator will
select one of the following options:
Press 1 – Enabled. Press 2 – Disabled.
The display returns to the System Setup Beeper menu.
7. Press 1 – Next. The display shows the
System Setup Date Format menu.
8. The administrator will
select one of the following options:
Press 1 – Next, to continue to System Setup Time Format menu. Skip forward to step 12.
Press 2 – Change, to change the Date Format option.
The display shows the Date Format options. The date is listed using 2 digits for the month, day and year. For example, June 27, 1999 becomes 06 27 99.
9. The administrator will
select one of the following options:
Press 1 – month/day/year. Press 2 – day/month/year. Press 3 – year/month/day.
The display shows the Date Format Separator options.
The numbers can be separated with a slash (/), dash (-), period (.) or no punctuation.
10. The administrator will
select one of the following options:
Press 1 – slash (/). Press 2 – dash (-). Press 3 – period (.). Press 4 – none.
The display is returned to the System Setup Date Format menu.
Page 97
Questions? Refer to Page 10-13 For Customer Care Information.
8-37
What You See on the Display
What You Do Comments
11. Press 1 to continue to
System Setup Time Format menu.
The display shows the System Setup Time Format menu.
12. The administrator will
select one of the following options:
Press 1 – Next, to continue to System Setup Password; skip forward to step 19.
Press 2 – Change, to change the Time Format.
The display shows the Time Format options. The four options to use are: 12 hour clock using capital AM/PM; 12 hour clock using lowercase am/pm; a 0-24 hour clock; and a 01-24 hour clock.
13. The administrator will
select one of the following options:
Press 1 – 1-12 AM/PM. Press 2 – 1-12 am/pm.
The display shows the Time Format Separator screen for hours. The 5 options are separators for hours.
14. The administrator will
select one of the following options:
Press 1 – colon (:). Press 2 – period (.). Press 3 – h. Press 4 – H. Press 5 – none.
The display returns to the System Setup Time Format screen.
15. If the administrator
selects one of the following options:
Press 3 – 0-23. Press 4 – 00-23.
The display shows the Time Format Separator screen for hours. The 5 options are separators for hours.
Page 98
8-38
What You See on the Display
What You Do Comments
18. Press 1 to continue to
System Setup Password.
The display shows the System Setup Password screen.
19. The administrator will
select one of the following options:
Press 1 – Next, to continue to the System Setup Change all settings to Factory Defaults menu. Skip forward to step 24.
Press 2 – Change, to change the Password.
The display will prompt the operator to enter the current Password.
20. Enter the current
Password, up to 8 digits, then press
Enter.
The display prompts the operator to enter a New Password.
16. The administrator will select one of the following options:
Press 1 – colon (:). Press 2 – period (.). Press 3 – h. Press 4 – H. Press 5 – none.
The display shows the Time Format Units screen for minutes. The 4 options are separators for minutes.
17. The administrator will select one of the following options:
Press 1 – m. Press 2 – apostrophe (). Press 3 – h. Press 4 – none.
The display returns to the System Setup Time Format screen.
Page 99
Questions? Refer to Page 10-13 For Customer Care Information.
8-39
21. Enter the New
Password, then press
Enter.
The display asks the operator to re-enter the password to confirm the New Password, then press
Enter.
22. Re-enter New
Password; press Enter to confirm.
The display returns to the System Setup Password menu.
23. Press 1 to continue to
System Setup Change all settings to Factory Defaults menu.
The display shows the System Setup Change all settings to Factory Defaults menu.
What You See on the Display
What You Do Comments
24. The administrator will
select one of the following options:
Press 1 – Next, to return to Change Setup menu.
Press 2 – Change, to change all settings to Factory Defaults.
The display shows the Confirm options.
25. The administrator will
select one of the following options:
Press 1 – Change to Factory Defaults.
Press 2 – Don't Change.
The display is returned to the System Setup Change all settings to Factory Defaults menu.
26. The administrator will
select one of the following options:
Press 1 to return to Change Setup menu.
Press Menu to return to the Menu Mode.
Press On/Off to turn off the monitor.
Page 100
8-40
Review/Change Setup for System Status
The Review/Change Setup menu allows the operator or administrator to review important information about the monitor operating status. The operator scrolls through the various screens, in the order they are listed below, by pressing 1
Next.
Battery Power
This option checks the amount of battery power available for a rechargeable or disposable battery.
Temperature
This option measures the current temperature of the area where the monitor is.
Memory
This option indicates the percentages of memory available for Patient Test results and Control Test results.
Software Version
This option displays the software version in the monitor.
Loading...