Medela Supplemental Nursing System, SNS, 8000794, 00901S, SNS20901 Instruction Manual

Supplemental Nursing System
(SNS™)
EN Instructions for use | ES Instrucciones de uso | FR Mode d’emploi
2
Important Safeguards
READ ALL INSTRUCTIONS BEFORE USING THIS PRODUCT.
The warning symbol identifi es all instructions that are important to safety. Failure to observe these instructions can lead to injury or damage to the breastpump. When used in conjunction with the following words, the warning symbols stand for:
WARNING Can lead to serious injury or death. CAUTION Can lead to minor injury. NOTE Can lead to material damage.
WARNING: To avoid health risk and reduce the risk of injury:
• The packaging should be inspected for holes, tears, broken seals, etc.
• If the sterile packaging is damaged, the contents must be cleaned prior to use (refer to cleaning instructions).
• This product is intended for use by a single user only and should not be shared between users.
• Sanitize all parts that come in contact with breastmilk or supplement once daily.
• Wash all parts that come in contact with breastmilk or supplement after every use.
• Inspect all components before each use.
• Do not use the components if they are damaged.
• Do not use the components if dirt, mold or other contamination is visible.
• Do not thaw frozen breastmilk in a microwave or in a pan of boiling water.
• Do not use the breastmilk or supplement if bottles or components become damaged.
• Always check the temperature of breastmilk or supplement before feeding to your baby.
SAVE THESE INSTRUCTIONS
CAUTION: Can lead to minor injury:
• Do not store wet or damp parts as mold may develop
• If tubing becomes moldy, discontinue use and replace.
Questions? Visit www.medela.com or call us at 1-800-435-8316.
3
1. Cleaning .................................................................... 6-11
1.1 Disassemble ..............................................................6
1.2 After each use ....................................................... 6-9
1.3 Sanitize once daily ............................................. 10-11
2. Assembly & Use ..................................................... 12-15
2.1 Assembly .......................................................... 12-14
2.2 Use ................................................................... 14-15
3. Tubing Guidelines ................................................... 16-17
3.1 Tube position ........................................................... 16
3.2 Tubing size guidelines ........................................ 16-17
4. Supplemental Information ...........................................18
Table of Contents
The intended use of the Supplemental Nursing System:
The Supplemental Nursing System is designed to administer supplemental nutrition to babies while they are being fed at the breast. Its intended use is to assist mothers in increasing and maintaining their milk supply.
4
Product Description
1x Feeding bottle
8110007
1x Attachment ring
8100500
1x Valve holder
2006620
Complete SNS
00901S(US)
Complete SNS
SNS20901(Canada)
SNS Container with lid
8000794
Questions? Visit www.medela.com or call us at 1-800-435-8316.
5
1x Neck Cord
6000155
1x Valve (translucent) with large tubing
2006620
1x Valve (red) with small tubing
2006620
1x Valve (white) with medium tubing
2006620
1x hypoallergenic
papertape roll
with holder
67138
1x hypoallergenic
papertape roll
8990007
1x Cover
8100502
6
1. Cleaning
Supplies needed:
• Mild Dish Soap
• Clean Dish Cloth
• Clean Sink or Bowl
• Drinking Quality Water
• Clean Towel
Parts to wash:
• Valve Holder
• Attachment Ring
• Feeding Bottle
• Valve with Tubing
• Cover
• Neck cord
1.1 Disassemble:
Before cleaning, dissassemble the parts below. Follow the cleaning and sanitizing instructions in Sections
1.2 and 1.3 that follow.
• Only use drinking-quality tap or bottled water for cleaning.
• Separate and wash all parts that come in contact with the breast, breastmilk or supplement immediately after use to avoid dry up of milk residues and to prevent growth of bacteria.
• Follow these cleaning instructions unless advised otherwise by your Healthcare Professional.
• Contents are sterile unless the package has been damaged or opened.
Caution
Warning
1
1.2 After each use
Clean all parts that come in contact with your breast, breastmilk or supplement. Refer to Section1.1.
Rinse all separated parts that came in contact with breast, breastmilk or supplement in cool water in order to remove breastmilk residue.
4
Replace attachment ring and valve with tubing and force water through tubing.
Questions? Visit www.medela.com or call us at 1-800-435-8316.
7
32
Soak all separated parts in warm soapy water for 5 minutes.
5
Fill SNS feeding bottle full of warm soapy water.
Refi ll feeding bottle with clean water.
6
Force water through tubing.
You may also wash SNS parts by the following the instructions on the Quick Clean
Breastmilk Removal Soap
bottle. (sold separately)
8
10
• Allow parts to air dry in a clean area.
• Store dry parts when not in use.
• Do not store wet or damp parts as mold may develop.
• If parts become moldy discontinue use and replace parts. Contact Medela Customer Service at customer.service@ medela.com or 800-435-8316.
• When you are cleaning the SNS, check feeding bottle and components for damage. Throw away and replace at the fi rst sign of damage or weakness.
Caution
1. Cleaning (cont.)
7
Squeeze the empty feeding bottle a few times to remove water droplets from the tubing.
Questions? Visit www.medela.com or call us at 1-800-435-8316.
9
8
Clean remaining parts with a clean dish cloth in warm soapy water.
9
Rinse all separated parts with clear water.
Reassemble dry parts and store when not in use.
11
10
1. Cleaning (cont.)
4
Allow water to cool in open pot and gently remove parts from water with tongs.
• Take care not to damage SNS parts during cleaning.
• Store dry SNS parts in a clean bag or container until next use.
• Wash SNS parts after each use.
• Sanitize SNS parts once daily.
Note
1
1.3 Sanitize once daily
Wash hands thoroughly.
Questions? Visit www.medela.com or call us at 1-800-435-8316.
11
• Place parts on a clean surface and/or towel.
• Allow all parts to air dry.
• Store dry parts when not in use (in a clean bag/container).
• Do NOT store wet or
damp parts.
5
3
• Fill a pot with enough water to cover all parts.
• Bring water to a boil.
• Place parts in boiling water for 5 minutes.
2
Separate all parts that come in contact with your breast, breastmilk or supplement. Be sure to separate valve and valve holder.
Refer to Section 1.1
12
2. Assembly & Use
1
4
Wash hands thoroughly.
• Slide tubing and valve holder through attachment ring and push into the ring until it snaps.
• Do not pull on tubing.
• Inspect parts for cracks, chips, holes, tears or warping. If you notice any damage to these parts, throw away and replace them immediately.
Note
• Do not pull on tubing.
Caution
2.1 Assembly
Questions? Visit www.medela.com or call us at 1-800-435-8316.
13
2 23
Choose proper size tubing. Refer to section 3.2 and consult with your Healthcare Professional.
Push valve with attached tubing over valve holder.
5
Fill feeding bottle with breastmilk or supplement recommended by your Healthcare Professional.
6
Screw attachment ring onto feeding bottle.
10
Position baby at breast and release just the tubing on the breast you are about to nurse. Place all your fi ngers underneath the breast and off the areola. Place your thumb on top of the breast and gently on the tube (don’t press hard with your thumb or the tube will point upward). Tickle baby’s lower lip with your nipple and wait till the baby opens wide. Position the tube under the baby’s upper lip as you pull baby close to the breast so that baby accepts both the breast and the tube. The tip of the baby’s nose should just touch the breast. This should help baby accept both your breast and the tube at the same time. If you have diffi culty, don’t worry as this may take a little practice. Remove baby and start again.
2. Assembly & Use (cont.)
14
87
2.1 Assembly (cont.) 2.2 Use
Hang the assembled SNS feeding bottle around your neck.
Gently clamp both tubings into attachment ring to avoid milk fl ow as you place the system around your neck.
Do not stretch or pull on tubing.
Questions? Visit www.medela.com or call us at 1-800-435-8316.
15
Questions?Questions?
11
9
12
Place tubing on the top of the nipple making sure it extends about 1/4” inch beyond nipple and secure vertically with a piece of tape down to the base of the areola. Another piece of tape may be placed horizontally at base of areola. Put tape on breast and areola but not on the nipple itself.
If the fl ow is not started by infant’s suck, it may be necessary to prime the system by squeezing gently on the feeding bottle. This will cause the tubing to be fi lled with breastmilk or supplement and prevent the infant from swallowing air.
Follow cleaning instructions in Section 1.
16
3.1 Tube position
Position the tube such that the baby keeps the tube in the center of its upper lip or in the corner of its mouth.
3.2 Tubing size guidelines
Consult with your Healthcare Professional to choose the proper tubing size.
Small Size Suggestions (red valve):
1. Baby gaining well but sucking needs to be improved
2. Thinner fl uids
Medium Size Suggestions (white valve):
1. Beginning Use
2. Adopted Babies
3. Slow Weight Gaining Babies
4. Most Sucking Problems
5. Premature Babies
Large Size Suggestions (translucent valve):
1. Sucking Problems characterized by a tight mouth
(Faster fl ow prevents tightening of mouth)
2. Poor action by the mouth
3. Weak, Lazy Sucking Babies
3. Tubing Guidelines
Questions? Visit www.medela.com or call us at 1-800-435-8316.
17
The versatility of the Medela SNS allows you to vary the rate of fl ow:
To summarize
Increase Flow Decrease fl ow
1. Use larger tubing 1. Smaller size tubing
2. Raise device 2. Lower device
3. Squeeze feeding bottle
4. Warm supplement
Loading...