Medacure Lincoln Expandable Bariatric Bed User Manual & Installation Guide Installation Instructions

USER MANUAL
LINCOLN EXPANDABLE BARIATRIC
5 FUNCTION ELECTRIC BED
WIDTH EXPANDS FROM 36” – 42” – 48”
FOR
SERIES MODEL # LX-BARI
MADE IN CHINA
TABLE OF CONTENTS
GE 2 INRODUCTION & PRODUCT FEATURES
PA
PAGE 3 PARTS LIST AND DESCRIPTION
PAGE 4 INSTALLLING THE LIFTING FRAME
& HEIGHT ADJUSTMENT MOTOR
PAGE 5
INSTALLING THE LENGTH EXTENSION FRAME
PAGE 6-8 INSTALLING BED BOARDS
PAGE 8 WIRING THE NURSE CONTROL
PAGE 9 CASTERS
PAGE 9-10 BED WIDTH EXTENSION
PAGE 10 BED LENGTH EXTENSION
PAGE 11 CORNER MATTRESS RETAINER
PAGE 11-12 BUMPER GUARD INSTALLATION AND USAGE
PAGE 12-13 SIDE RAIL INSTALLATION
PAGE 13 PATIENT REMOTE CONTROL
PAGE 13-15 PATIENT & NURSE REMOTE CONTROL
OPERATING FEATURES & FUNCTIONS
PAGE 15 BATTERY BACKUP
TRENDELENBURG AND IV POLE
1 | Page
The Med
Hospital & Home Care settings. It expands in both the width (from 36” to 48”) and length (from 80” to 88”) which
makes it ideal for standard, bariatric and tall patients. It has a low position of 14.9” and 5 patient position functions to
enhance the quality of care and for ultimate comfort. The Lincoln uses 4 sophisticated motors to control all functions
and has a weight capacity of 700 lbs. The bed comes standard with a patient remote control and an optional nurse
control on the foot board. The bed is a Class II Medical Device. The plug and the motors are IP43 rated.
Durable and built to last with all steel construction and removable head & foot boards. The Lincoln is a sturdy bed
with multi-functional features and an attractive upscale design that will enhance the décor of any room.
aCure Lincoln Expandable Bed is a fully electric, versatile and multi-functional bed for Long Term Care,
5 Func
tion Electric Bed
Head and Foot Positions, Hi/Lo, Trendelenburg & Reverse Trendelenburg
Expandable width from 36” – 42” – 48”
Adjusts for both Standard and Bariatric Patients
ODUCT FEATURES
PR
Patient remote hand control for all 5 positions (standard)Nurse control mounted on footboard with override feature (optional)Backup battery for all 5 functions in case of emergency
sition height travels to 29 ½”
Hi poLo
w position goes to 14.9”
Adjustable/expandable mattress deck length from 80” to 88”
 
Mattress keepers – keeps mattress from sliding
asters: 4 pcs. swivel & 4 locking casters
8 C2 Adju   IV pole mount on 2 head corners and the middle of the bed frameTrapDurable rectangle tube bed deck   Durable 1 piece reinforced steel frameE
  Fo
stable side rails
Under bed clearance of 5”
eze bracket mount on 2 head corners
Interchangeable head/foot boards
ach bed has four mattress retainers, 1 on each corner
Ea
ch bed has two bumper guards on the head of the bed
ldable head board
Weight ca
2 | Page
pacity: 700 lbs.
Width Expands & Retracts
without any Tools
PARTS LIST & DESCRIPTION
IMPORT keep in a safe place for future reference.
system should only be installed by an authorized technician or dealer representative.
Bed foot board (with optional nurse control
ANT: Read this manual completely before setting up or operating this system and
Bed frame
To avoid injury and for maximum safety, this
Electric bed installation parts listed below as per Figure # 1
Extension frame (right)
Extension frame (left)
Bed head board
Figure # 1
3 | Page
Φ 8 Flat washer
Bracket for bed foot part
M8 ×18 Plate nut
M8 × 20 Cup head
bolt
1. Installing the lifting frame
Remove theφ13.5 bolt from the lifting frame (1). Place aφ14 plastic washer between the lifting frame (1) and lifting frame (2), matching the bolt holes on both lifting frames. Fasten with a φ
13.5 bolt with cotter pin as per Figure # 2. Use 4 bolts to connect the lifting frames.
φ13.5 Bolt
φ14 Plastic
washer
Figure # 2
stalling height adjustment motor
2. In
Match the installation holes on height adjustment motor and motor base and fasten byφ10
bolt with latch as per Figure # 3. Before attaching the bolt, first attach a 10 plastic washer to the
bolt. Install the two height adjustment motors one by one.
Figure # 3
4 | Page
3. In
stalling the length extension frame
3.1. Insert extension frame (left) & extension frame (right) into the 60 x 30 mm rectangular bed
frame tube. See Figure # 4
The extension frame tubes
should face upward.
Insert the extension frame, tighten the
bolt to lock the extension frame, and
then fix the plastic head bolt.
Figure # 4
3.2. Install bracket for bed foot part as per Figure # 5.
Figure # 5
`
5 | Page
4. Installing the bed boards
4.1 Press the M8 × 18 plate nut into the bed board holes, as per Figure # 6.
Figure # 6
4.2.1 Mat
head bolt with φ 8 flat washer, as per Figure # 7.
ch the bed head board with the bed frame installation holes. Fasten by M8 X 20 cup
Figure # 7
4.2.2 Foldable head board installation as noted in Figure # 7.1
6 | Page
Unfold
edPosition
FoldedPosition
Figure # 7.1
4.3 Match the bed foot board with the extension frame installation holes.
Fasten by M8 X 20 cup head bolt with φ 8 flat washer.
Bed foot board can be installed in both the high and low level position. See Figure # 8 & Figure # 9.
Figure # 8 Bed foot board in low-level installation position
7 | Page
Figure # 9
Bed foot board in high-level installation position
4.4 Install the bed head board to the bed head frame, and bed foot board to the bed foot frame,
as per Figure # 10.
Figure # 10
4.5 Wire the nurse control on bed foot board as per Figure # 11
8 | Page
Connector for the nurse control
& remote control
Figure # 11
5. Casters
p
There are 4 pcs. locking casters and 4 pcs. swivel casters per set.
Step on the locking pedal (lower) to activate the locking position and step on the unlocking pedal
(upper) to deactivate the locking system. See Figure # 12 & 13.
Step on the pedal to lock
the caster in
lace
Step on the pedal to unlock the caster
Figure # 12 Locked position Figure # 13 Unlocked position
6. Operation for bed surface extension frame
6.1 First, loosen the “Locking Bolt”. Then, pull the “Pulling Button” with one hand and pull out
the extension frame until it clicks into your desired location hole (see 6.2 below). Tighten the
“Locking Bolt” to lock the extension frame in place. See figure # 14.
Pulling Button
Locking Bolt
Figure # 14
9 | Page
6.2 IMPORTANT NOTE: There are 3 location holes on each extension frame for 3 width
adjustment settings - 36”, 42” and 48” widths. There are 4 separate extension frames on each
side of the main frame. Repeat the above method in 6.1 to achieve the desired width and then
tighten the “Locking Bolt” to lock the extension frame in place. See Figure # 15.
Figure # 15
7. Bed foot length extension frame
Loosen the screw on the bed frame, and pull out the extension frame.
Tighten the screw until the ideal length is selected, as per Figure # 16.
Screw
Figure # 16
10 | Page
8. Each corner of the bed frame has two mattress retainer installation holes.
Insert the mattress retainer as noted in Figure # 16.1.
Mattress Retainer Installation for Standard Bed Deck

Figure # 16.1
Mattress Retainer Installation for Expandable Bed Deck
9. Bumper Guard Installation and Usage
There are two retractable rubber Bumper Guards on the head of the bed. When extended
they will protect the walls when raising or lowering the bed. Retract the bumper guards to
discreetly store under the frame.
9.1 Insert the bumper guard stem to the bracket. Align the mounting holes and fasten the
screws as noted in Figure # 16.2.
11 | Page
Figure # 16.2
9.2 Extended Position: Push the bumper guard stem upward and turn the stem outward until
the screw clicks with slot as noted in Figure # 16.3.
Figure # 16.3
9.3 Retracted Position: when bumper guard is not in use, push the stem upward and turn it
inward until screw clicks with slot as noted in Figure # 16.4
Figure # 16.4
10. Side rail installation & use
10.1 Loosen the screw on the side rail mounting base. Clip the mounting base with the 60 x
30mm rectangular bed frame tube and tighten the screw when you finish adjusting the
position. See Figure # 17
Screw Pulling button
Figure # 17
12 | Page
10.2 To raise the side rails - Pull up until the rail clicks in place.
To lower the side rails - release the pulling button as per Figure # 18.
Figure # 18
11. Patient remote control only - on bed models without the nurse control
Raise bed head part
Lower bed head part
Raise bed foot part Lower bed foot part
Raise both bed head
and foot part
Raise the entire bed
Trendelenburg
Lower both bed head
and foot part
Lower the entire bed
Reverse Trendelenburg
12. Patient remote control and nurse control – on bed models with nurse control
First ensure the power cord is connected. Patient can easily control the bed positions by
using the patient remote control or nurse control, including adjusting the bed height, raising
and lowering the back and foot positions, trendelenburg and reverse trendelenburg. Press
the remote control button until the ideal position is selected. See Figure # 19.
13 | Page
re # 19 Nurse Control (optional) Patient Remote Control
Figu
When the power is on, the locking option is released automatically. The LED1
indicator light
turns on and the patient remote control and nurse control are ready for use.
To lock the position functions, press the lock button (S12). The indicator light will turn off and the
S1-S10 position buttons on the patient remote control and nurse control are locked and
inoperable.
To unlock the position functions, press the unlock button (S11). The indicator light will turn on
and the patient remote control and nurse control functions are unlocked and operable.
S1-S10: to control the raising/lowering of the head, raising/lowering of the foot, raising/lowering
of both the head and foot, the entire bed height adjustment, trendelenburg and reverse
trendelenburg, under control of the lock function.
REMOTE CONTROL OPE RATING FEATURES & FUNCTIONS – on bed models with nurse control
S1: controls the raising of the head position S2: controls the lowering of the head position S3: controls the raising of the foot position S4: controls the lowering of the foot position S5: controls the raising of both the head position and foot position S6: controls the lowering of both the head position and foot position S7: controls the raising of the entire bed S8: controls the lowering of the entire bed S9: controls the Trendelenburg S10: controls the reverse Trendelenburg
S11: to Unlock the position functions S12: to Lock the position functions
S13: controls the lowering of both the head and foot position to the bottom and simultaneously
controls the reverse trendelenburg to the highest position; out of control of the locking button.
14 | Page
S14 CPR button: controls the lowering of the head and foot position and lowers the entire bed to
the lowest position; out of control of the lock function.
S15 Night Light button (optional feature): turns the night light on and off; out of control of the lock
function.
When the plug is connected to an outlet using AC power, the indicator light LED2 for the
nurse control plug and patient remote control backlight comes on. Without AC power or when the
plug is not plugged in, the indicator light LED2 for the nurse control plug and patient remote control
backlight goes out.
13. BATTERY BACKUP
The MedaCure Lincoln Bed comes with an emergency battery backup which enables the caregiver
to control all functions by the nurse control or patient remote control in case of a power outage or
during patient transport. Battery type: lead-acid cell
If the backup battery is not connected, the green/orange indicator light f or the nurse control
panel will be off. If the battery power is sufficient to maintain normal operation, the green
indicator light LED3 will turn on and the orange indicator light LED4 will turn off. If the battery
voltage is unable to maintain normal operation, the green indicator light LED3 will turn off and the
orange indicator light LED4 will turn on.
When the battery power is low the alarm will sound to alert the user until the battery discharge
function is shut down. During long distance transportation or when not in use for a long period of
time it is recommended to shut down the storage battery function when AC is off by pressing the
red button at the right side for 5 seconds.
Figure # 20 Trendelenburg position Figure # 21 Reverse trendelenburg position
14. Lifting pole and IV pole are optional
There are two (2) lifting pole sockets and two (2) IV pole sockets on the head part of the bed
frame.
15 | Page
© 2020
MADE IN CHINA