Starlight p GB 11 / 16
MANUAL OF USE AND MAINTENANCE
05.2 - Sterilisation procedure
WARNING: Carry out sterilisation only in a steam autoclave
at a maximum temperature of 135° for 20 minutes.
Do not use any other sterilisation procedures (dry heat, radiation, ethylene oxide, gas, low-temperature plasma, etc.).
DANGER: Infection control - Sterilisable parts.
To avoid infection caused by bacteria or viruses, always clean the
following components after each treatment:
1 Optical fibre;
2 Optical protection.
NOTE: All the stages of sterilisation should be followed by the operator, in accordance with EN standard 554.
05.3 - Cleaning, disinfection e sterilisation of the optical
fibre
WARNING: Do not use sharp-edged objects to clean the opti-
cal fibre.
Carry out the following operations:
1 Eliminate any residues of polymerised composites from the sur-
face of the optical fibre with alcohol.
2 Disinfect the surface using a cloth moistened with a solution of
mild detergent/disinfectant having a neutral pH (pH=7).
3 Dry.
4 Seal the optical fibre in a disposable bag on its own.
5 The optical fibre should autoclave sterilised for 20 minutes at a
maximum temperature of 135°C.
05.4 - Cleaning, disinfection and sterilisation of the
optical protection
WARNING: Do not use sharp-edged objects to clean the opti-
cal protection.
Proceed as follows:
1 Clean and disinfect the surface using a cloth moistened with a
solution of mild detergent/disinfectant having a neutral pH (pH=7).
2 Dry.
3 Seal the optical protection in a disposable bag on its own.
4 The optical protection should be autoclave sterilised for 20 min-
utes at a maximum temperature of 135°C.