Mecasys optizen 2120 uv User Manual

User's Guide
July 2012
Mecasys Co., Ltd.
Notice!
Spectrophotometer Optizen® and Software OptizenView™ are
No part of the Optizen Series, including this publication
may be reproduced, or modied in any form or by any means
without the prior written permission of Mecasys Co., Ltd.
For more detail information on Optizen® and OptizenView™,
you may refer to “Technical Support”
in the last Chapter of this guide or visit our websites below
.
http://english.mecasys.co.kr
2) Mode Setup
3
)
Measure (Step by Step
)
4) Flow Chart of Survey Scan Test
5
)
Flow Chart of Survey Scan Mode setup
3. Standard Curve
1
)
Initial Screen Test Mode
2
)
Mode Setup
3
)
Measure with Multi Cell
4
)
Flow Chart of Standard Curve Test
5
)
Flow Chart of Standard Curve Mode setup
4. ABS Ratio
1
)
Initial Screen Test Mode
2
)
Mode Setup
3
)
Measure
4
)
Flow Chart of Simple Kinetic Test/Mode Setup
5. Simple Kinetics
1
)
Initial Screen Test Mode
2
)
Mode Setup
3
)
Measure
4
)
Flow Chart of Simple Kinetics Test
5
)
Flow Chart of Simple Kinetics Test Mode Setup
6. Communication
1. Technical Supports
Part IV. Technical Supports
25
25
26
27
28
28
29
30
33
34
35
35
36
36
37
38
38
38
39
41
42
43
44
Part III. Measuring in Each Test Modes
1. ABS/%T/CONC
1
)
Initial Screen of Test Mode
2
)
Mode Setup
3
)
Measure (Step by Step
)
4) Flow Chart of ABS/%T/CONC Test
5
)
Flow Chart of ABS/%T/CONC Mode setup
2. Survey Scan
1
)
Initial Screen Test Mode
18
18
19
19
21
22
23
23
Part I. Introduction & General Information
1. Introduction
2. Specifications
3. Constructions
1
)
External Construction
2
)
Cell Holder
4. Accessories
1
)
Micro Volume Cell Holder
2
)
Long Path Cell Holder
Contents
6
7
8
8
9
10
10
11
1. Turning on Switch
2. Key Operation
3. Main Menu
4. Cell Moving Method
5. Utilites
Part II. Functions and Application
12
13
14
14
15
6
7
Part I.
Introduction & General Information
Part I.
Introduction & General Information
Part I. Introduction & General Information
1.
Introduction
Thank you for purch asing UV/VIS S pectrophotometer, Optizen Series.
Optizen Serie s can be applied var ious elds from a simple examinatio n like a water quality
analysis to a comp lex examination like a bio -chemic al analysis. User c an measure easier,
faster, and more accurate by c onvenient interf ace and auto- function.
In addition, user c an check and c ontrol experimental results on th e real time by OptizenView,
PC-Inter face sof tware of Optizen S eries.
Using Manual
This manual includes the ways of system installation, Optizen operating.
The manual is com posed of four par ts, which are c ategorized upo n the subjects; the refore,
user can searc h quickly the nec essary par ts.
Mecasys Co., Ltd. cont inuously suppor ts update via mai l, internet and e- mail.
1. Introduction
2. Specifications
2. Specications
Wavelength Ran ge 190nm to 1100nm
Spectr al Band Pass Width < 1nm
Wavelength Accuracy < ±0.5nm
Wavelength Reproducibility < ± 0.1n m
Monochromator
Modied Czerny- turner type with 1200 lines /mm Blazed grating
Stray Light 0.05 %T at 220 &340 nm
Photometric Range -0.1 to 3.0A, 0% to 125%
Photometric Accuracy < 1% of at 1 A
Photometric Reproducibility ± 0.001 A
Noise level < 0.001 A
Drift 0.002A /hr at 340nm
Stability ±0.002 A/ hr
Light Source Tungsten-Haloge n lamp and Deuterium lamp
Display Mode Backlight. Grap hical, liquid crys tal display with contrast cont rol
Sample Holder Automatic Rotar y Type 8-position multi- cell holder
Inter face Ports RS-232 and Centronic s ports
Wavelength Slew Rate 50 00nm/min
Wavelength Sc an Rate 1000nm/min
Standar d Capabilit y Software
Absorbance/%Transmittance, Concentration and File manager
Local Software
Real-t ime PC S/W
Automatic multi- cell & multi-wavele ngth measure
Advanced multi-survey scan
High order standard cur ves
Time based scan (simple kinetics)
Absorbance ratio
User dened functio n
Power Requirement 220V, 50/60Hz
Dimensions 514(W)mm X 430(D)mm X 243(H)mm
Weight 16kg (Only Main System)
8
9
Part I.
Introduction & General Information
Part I.
Introduction & General Information
2) Cell Holder
3. Constructions
1) External Construction
3. Constructions
3. Constructions
Main
Top Side
Sample Compartment
Main Case
KEY PAD with LCD
Power Switch : Power ON /OFF
Fuse
AC Power Cable
COM 1: Interfacin g port for PC
(9 PIN RS-232 CON NECTOR)
Print (25PIN RS-232 CONNECTOR)
FAN
Rear Side
1
4
1
6
3
2
3
5
2
10
11
Part I.
Introduction & General Information
Part I.
Introduction & General Information
2) Long Path Cell Holder (50mm, 100mm)
Cell Holder Case
Leaf Spring
Screws
Screw
Installation of the Long path cell holder
1. Align the long path c ell holder with c ell holder case. (Three screws)
2. Push the long pat h cell holder to the bottom of cell ho lder case.
A difference of 50mm Path Cell Holder and 100mm Path Cell Holder
1. 50mm Path Cell Hol der : 50mm Path Cel l Holder has the 50 mm Block.
2. 100mm Path Cell H older : If you measur e the 100mm cell, you m ust remove the 50mm
block rst.
4. Accessories
1) Micro volume cell holder
Cell Holder Case
Micro Volume Cell
Screw
Screw
Screw
Installation of the Micro volume cell holder
1. Align the micro volu me cell holder w ith cell holder c ase. (Three sc rews)
2. Push the micro vo lume cell holder to the bottom of ce ll holder case.
Rotate three adjust ing screw to adjust th e position Mic ro volume cell at the M icro volume
cell holder.
4. Accessories
4. Accessories
12
13
Part II.
Functions and Application
Part II.
Functions and Application
Part II. Functions and Application
1. Turning on Switch
It is self-tunin g about CPU & ROM, Wave Moter, Cell Moter, Filter Moter, W Lamp, D2 Lamp
and D2 Wave. Results of self-t uning indicate Ok o r Error. Incase Error o ccurs at any step,
self-tuning is s topped. If you want to che ck next step, push [ENT ER] button. Only ever ything
is OK, automatically move to Main Mo de.
To get more stable data, pleas e warm up the system abou t 30 minutes afte r turning on the
switch.
CPU & ROM : Check CPU & ROM
WAVE MOTER : Wavelength motor dri ving test
CELL MOTER : Multi- Cell Holder dr iving test
FILTER MOTER : Filter dr iving test
W LAMP : Check the statu s of Tungsten-halogen La mp
D2 LAMP : Check the st atus of Deuterium L amp
D2 WAVE : Check the status of D euterium Lamp Wave
2. Key Operation
Key Description
Different Func tions in each scr een
~
To Log in gures from 0 to 9
To Back Space in Typing
To Insert Dot
To Input Minus
To Move cell location
Set up of Utilities m ode
To p ri nt
To Exit from Mode or To Cancel Input
To Test or To Set up Make Mode
To Conrm the Input, or To Execute
2. Key Ope ration
1. Turning on Sw itch
14
15
Part II.
Functions and Application
Part II.
Functions and Application
5. Utilites
You can change setup by selecting number.
1) Cell Mode
You can setup about Cell Mo de. Default is Multi c ell Mode. If you want to use S ingle cell
holder, have to change cell m ode to Single cell holder.
Multi Cell Mode : Use 8 hole cell ho lder.
Single Cell Mode : Use single Cell H older.
2) Unit
To change unit, push number 2 th en show up 12 concentrat ion unit. After moving with Up/
Down button, c onrm with [Ente r] button then uni t is changed. Default is CONC.
3. MAIN MENU (Selecting Mode)
Real time cloc k function is supp orted at only 2120UV p lus
Move to measuring mo de by selecting number key.
To move another mode after measuring one mode, c omeback to this ‘ Main Menu’ then se lect
num ber.
Even though mode is ch anging, measured data remain without saving until turn of f Optizen.
On the upper lef t side is a title an d on the upper right s ide shows data infor mation, print
status, and the c ell position st atus. The central s creen shows the gr aphs or detail infor mation
on the measured dat a. The bottom si de of screen indic ates explanation, c ommand, and the
choice box for f unction.
4. Cell Moving Method
Cell Holder is d esigned to hold up to 8 C ells. Each Cell h as number “B, 1, 2, 3, 4, 5, 6, 7 ” and
it is possible to m ove a cell upward and downwa rd respectively by [ UP], [DOWN].
Please keep in mind that t he rst cell, “ B” is assigned to p ut Reference Cell.
3. MAIN M ENU (Selecti ng Mode)
5. Utili tes
16
17
Part II.
Functions and Application
Part II.
Functions and Application
9) Calibration Mode (Secret Key:’9’)
Optizen has self- correc tion mode for wavele ngth and cell hol der position. Wavelen gth
calibration mode checks the 0 th order posit ion of grating and th e peak wavelength (656 .1nm)
of Deuterium lamp.
- wp X : 0
th
order position
- lp X : # of step per 1nm
- cp : Cell Positio n
Wavelength Tuning : Self-correction f or wavelength.
Cell Tuning : Self-correct ion for cell pos ition.
Manual Input : Input modied va lues of calibration parameter.
Parameter Save : Save modied value.
Calibratin : Select Calibr ation Method. (One Po int or Two Point)
Warning : It is very strongly re commended th at ONLY expert or fully trai ned staff
engineers ope rate these advanced applicatio ns. Fatal errors c an be caused by
some incorrect operations.
3) D2 Lamp Save Mode
It automatical ly turns off the p ower of the D2 lamp that is n ot used by selecting number 3.
Incase this mo de ‘ON’, D2 Lamp is off at our o f wavelength range.
‘Wait a moment ’ message is showe d up at that time. A few seconds is need.
The wavelength rang e of Optizen 2120UV spect rophotometer is f rom ultraviolet ra diation
(190nm) to near infrared radiati on (1100nm). For this, deuteriu m lamp and tungsten -
halogen lamp is us ed as a light source. Two lamps are ch anged each other au tomatically
in 350~ 470nm range.
4) LCD Contrast Control
It controls br ightness level of LCD bac k light.
5) Lamp Change WL.
You can setup the wavelength for c hanging point of deuterium lamp and tungsten-halog en
lamp.
The range of the valu e that you can input is f rom 350nm to 470nm.
The default value is 370nm.
6) Initial Wavelength
You can set initial wavelength of Optizen. Incase set mainly used wavelength, you don’t need
to set wavelength every measuring time.
7) Clear Lamp Position
Clear Lamp posi tion of W Lamp and D2 Lamp.
8) Set Date & Time (Only 2120UV Plus)
Real time cloc k function is supp orted at only 2120UV p lus.
5. Utili tes
5. Utili tes
18
19
Part III.
Measuring in Each Test Modes
Part III.
Measuring in Each Test Modes
2) Mode Setup
Initialize Factor : Initialize all fac tor.
Modify Factor : Can modify S lope and Intercept .
Open Factor : Can open saved Factor by sel ecting number.
Open factor fu nction is gener ated only saved factor existing.
3) Measure (Step by Step)
With Multi Cell
1. Incase Push NEW W.L. [F1], you can see th e top left side g ure then input wavelength.
You can enter by 8 wavelengths.
2. Incase push CE LL [F3], show up the top right side gure. You can selec t the cell to
measure with numb er key.
3. MEASUR E [F2].
Part III. Measuring in Each Test Modes
1. ABS/%T/CONC
1) Initial Screen of Test Mode
The mode assign s the size of Wavelength and th e position of a cel l, then measures and
marks Absorbance, %Transmittance and Concentration.
<Upper Item>
NO. : TEST No.
C# : Cell number of each T EST
W.L. : Wavelength of each TEST
ABS : Absorbance of each TEST
%T : %Transmittance of each TEST
CONC : Concentratio n of each TEST
<Bottom Item>
NEW W.L [F1] : Input wavelength by maximum 8.
MEASURE [F2] : Start meas uring.
CELL [F3] : Select the cell by O N or OFF cell number.
DELETE [F4] : After measur ing, this functi on shows up. You can delete all data.
1. ABS/%T/CONC
1. ABS/%T/CONC
20
21
Part III.
Measuring in Each Test Modes
Part III.
Measuring in Each Test Modes
4) Flow Chart of ABS/%T/CONC Test
With Single Cell
1. Input Wavelength NEW W.L. [F1], at ‘Set new waveleng th _ nm’
2. Insert Bl ank then push AUTOZERO [F3].
3. After measur e AUTOZERO [F3], remove Blank then inse rt sample ce ll and MEASUR E [F2].
You can setup wavelength again or m easure AUTOZERO [F3] again.
With Standard Curve
1. At Mode setup step, you can sel ect and open a saved St andard Curve.
Ex) Incase you want to use Standard Curve AAA A in ST0, input No. 0.
2. Enter the CELL [F3] th en select the ce ll to measure.
3. MEASUR E [F2], Measured re sult is like the top right side gure.
4. Can measure again w ith MEASUR E [F2] and delete all dat a with DELETE [F4].
1. ABS/%T/CONC
1. ABS/%T/CONC
Move to
Mode Setup
1
Move to
ATC Mode
Input wavelength, [EN TER].
F1
Multi Cell Measure Mode
Measure by Standard Curve
Select the
Open Factor
On/Off t he Cell with No. Key.
F3
Measure Absorbance.
F2
Input wavelength, [EN TER].
F1
Single Cell Measure Mode
Insert Blank, AUTOZERO.
F3
Insert Sample then Measure.
F2
22
23
Part III.
Measuring in Each Test Modes
Part III.
Measuring in Each Test Modes
2. Survey Scan
1) Initial Screen of Test Mode
At this mode, you ca n survey scan abou t a sample in selected wavelength range. You can
see Absorba nce or Transmittance d ata with Graph or Table.
At the Survey Sc an Mode screen, you w ill nd 'New bas eline is needed' c ommand.
BASELINE is like a Bla nk of ATC Mode. You must measure BASELINE [F1] rst.
<Main Screen>
Baseline [F1] : On the proce ss of Baseline Sca nning, its status is i ndicated by '%'. When
the Baseline Sc an is over, the 'READY' sign will come up.
New Scan [F2]
Select the ce ll you intend to measure an d
push NEW SCAN [ F2] button, then it w ill scan
the wavelength sele cted in each cel l.
While survey sc an, you can see Abs orbance
or %Transmittance as wavelength on t he
upper side of scre en. The middle par t of
screen will indi cate real time gra phs.
If the scanning i s over, the graphs, which are
automatically readjusted to the out put range
of Y, will be drawn promptly. Based upon th is
data, all the gur es for graphs and tex ts of
Survey Scan M ode will be determi ned.
5) Flow Chart of ABS/%T/CONC Mode Setup
1. ABS/%T/CONC
2. Sur vey Scan
Move to
Mode Setup.
Initialize Factor
Initializing all factor.
(Slope: 1.0, y-Intercept: 0.0)
1
Open saved Standard Curve.
3
Open Factor
Set slope : _
Set Y-intersection : _
Modify Factor
Modify slope and y-Interc ept of
Graph.
2
24
25
Part III.
Measuring in Each Test Modes
Part III.
Measuring in Each Test Modes
ABS / %T [F4]
It is a Toggle Switch which d ecides whethe r to view the data in Ab sorbance or in
%Transmittance. The range of wavelength is maintained during toggling.
2) Mode Setup
Start Wavelength : Input start waveleng th to survey scan. Default value is 400.0 n m.
Finish Wavelength : Input nish wavelength to sur vey scan. Default val ue is 700.0 nm.
Input wavelength at ‘S et new wavelength _ nm’ with num ber key.
3) Measure step by step
1. Setup wavelength to measure at M ode setup step.
2. Select the c ell to measure with U p/Down butto n.
3. Measure Base Lin e [F1].
4. Measure New Scan [F2].
5. Conrm dat a with Graph, Ab s/%T.
• GRAPH [F3]
X-ZOOM [F1]
In order to look into g raphs in details, you c an
select the ran ge of wavelength. The modied
graphs by this mode w ill be kept.
- MIN X [F2] :
Input t he minimum wavelength to
zoom.
- MAX X [F3] :
Input the maximum wavele ngth to
zoom.
- RECALL [F4] : Init ia liz e sh ape of g rap h.
Y-ZOOM [F2]
You can see y-axis of the g raph in detail.
CURSOR [F3]
At this mode, you ca n search Absor bance
and %Transmittance of each waveleng th step
by step.
Wavelength, Absor bance and %Trans-
mittance w ill be indicated like the left side
gure.
<<-- [F1], -->>[F4] is a key to search Peaks
and Valleys values only. With <- [F2],
-
>[F3] key can move by 1 nm and show the
respective dat a.
TABLE [F4]
You can see data by table that indicates
Peaks, Valleys, wavelength, AB S, %T.
2. Sur vey Scan
2. Sur vey Scan
26
27
Part III.
Measuring in Each Test Modes
Part III.
Measuring in Each Test Modes
4) Flow Chart of Survey Scan Test
2. Sur vey Scan
2. Sur vey Scan
5) Flow Chart of Survey Scan Mode Setup
2
Move to Survey
Scan Test Mode
Scan Baseline.
F1
Measurement with Multi cell
Start Measuring.
F2
Conrm data with X-ZOOM, Y-ZOOM, CURSOR, and TABLE.
F3
Insert Blank, Scan Baseline.
F1
Measurement with Single cell
Conrm measur ed data with ABS or %T
F4
ABS/%T Toggle Mode
Select the Ce ll with
Up/Down.
UP DOWN
Remove Blank, Inser t sample then st art
New Scan.
F2
Move to
Mode Setup.
Start Wavelength
Set new wavelength _ nm Input wavelength, [Ente r].
1
Finish Wavelength
Set new wavelength _ nm Input wavelength, [Ente r].
2
28
29
Part III.
Measuring in Each Test Modes
Part III.
Measuring in Each Test Modes
GRAPH [F3]
After measur ing, GRA PH [F3] button
shows up. You can see standard c urve
graph.
FACTOR [F4] : You can open, modif y or delete saved fact or.
At ABS/%T/CONC mode, you c an get concentr ation of samples us ing a saved standard
curve. Open st andard cur ve name is indicated at u pper side of scree n.
2) Mode Setup
Wavelength : Setup wavelength to measure. In case select num ber 1, the message of ‘Set
new wavelength _ nm’ shows up. H ere input wavelength wi th number key
then [ENTER].
Select Type : Select the type of Sta ndard Curve you pla n to make. There are three k inds of
Curve t ype.
- Linear(0) Standard C urve : Linear ty pe that passes zero point
- Linear(N) Standard Curve : Gener al linear type
- Segmented Stan dard Curve : A Type of graphs that l inks certai n points.
3. Stand ard Curve
3. Stand ard Curve
3. Standard Curve
1) Initial Screen of Test Mode
At initial scree n, you can see only AD D [F1], FACTOR [F4]. To make a standard curve, select
ADD [F1] and input concent ration. Then ME ASURE [F2] wi ll appear. With FACTOR [F4] you
can add data at saved st andard curve. You can also FACTOR OPEN or DELETE.
At this mode you ca n make a standard cur ve with standar d solution. You can save standar d
curve by ten.
<Upper side Item>
NUM. : TEST No.
CELL : CELL No. of each TEST
CONC : Concentration of standard solution
ABS : Absorbance of eac h standard solution
<Bottom side Item>
ADD [F1] : Select ADD [F1] and input Conc entration value.
You can input up to 7 concentr ations conse cutively.
To make standard curve, you have to measure minimum 2~3 standa rd solutions.
MEASURE [F2] : Measure conc entration of sta ndard solution. A fter measuri ng, you can
add another con centration by AD D [F1] and measure again by MEASU RE
[F2].
30
31
Part III.
Measuring in Each Test Modes
Part III.
Measuring in Each Test Modes
3. Measure con centration of eac h cell with ME ASURE [F2].
Results show up like the b ottom gure.
Single cell Mo de : After measuring Reference t hen insert st andard solution a nd measure
in order.
After measur ing, can add concentration wi th ADD [F1].
4. You can conrm measured data with GR APH [F3].
Graph type i s selected at Mode setup.
Manual Input
- NORMAL :
Incase make a stand ard curve by measur ing concent ration of standar d solutions.
- INPUT : By inputting C oncentration a nd Absorbanc e by maximum 30, you c an make
Standard Cur ve. After input ting, you can make sta ndard curve auto matically by
clicking ME ASURE [F2] in Test Mode.
Data Clear : Incase selec t the [Data Clear], you can see the mes sage ‘Warning! ! All data
will be lost. Pres s [ENTER] to Conrm.’ then all data is deleted.
3) Measure with Multi cell
1. Setup wavelength and graph t ype at Mode Setu p step.
2. Select AD D [F1] at initial screen of standar d mode then input c oncentration o f standard
solution in order.
You can add by 7 concentrati ons. After addi ng over 2, then shows up ME ASURE [F2] key.
Incase select Input mode
Select
ADD [F1] then ‘Set concentration: _’ shows up. Here input concentrat ion then [ENTER].
After that, t he message of ‘Set a bsorbance: _’ shows up t hen input absorb ance. Conrm
with [ENTER]. Repeat as number of sample s.
3. Stand ard Curve
3. Stand ard Curve
32
33
Part III.
Measuring in Each Test Modes
Part III.
Measuring in Each Test Modes
4) Flow Chart of Standard Curve Test5. After measu ring, select FACTOR [F4] then SAVE [F3] shows up like the bottom g ure.
You can save standard cur ve by 10. Also you can open saved st andard cur ve with [OPEN]
button. By selec ting the number you c an delete save standa rd curve.
6. If you want to delete measured data, select data clear at Mode setup step.
3. Stand ard Curve
3. Stand ard Curve
3
Move to
STC Mode
Measure with multi cell
Standard solutions are measured as input or der.
F2
You can conrm measured data with
graph.
F3
You can save or conrm made st andard
curve.
F4
1. Insert Referenc e cell then measur e.
2. Insert sample cell then m easure in
or der.
F2
Measure with single cell
Input concent ration of standa rd solutions in o rder.
At input mode, input concentration
and absorbance.
F1
34
35
Part III.
Measuring in Each Test Modes
Part III.
Measuring in Each Test Modes
5) Flow Chart of Standard Curve Mode Setup
4. ABS Ratio
User input Factor and Wavelength at mode setup step t hen ABS Ratio is me asured and
calculated auto matically. The ratio means A
260nm/A280nm
.
1) Test Mode initial screen
It automatical ly calculates the ratio per 2 or 3 Wavelength s.
Start the m easurement by click ing MEASUR E [F2] in the beginning. Prior to measurement,
please input the Wavelength, Cell Number a nd Factor at the Mode Set up.
You can measure maximum 7 sa mples.
More time is nee ded compare wit h other mode, bec ause measure at 2~3 wavelengt h
about one sample.
<Upper side Item>
TN-C : TN means total measuri ng times, C means Ce ll number.
ABS1 : Measured absorbanc e at Wavelength 1 (260nm).
ABS2 : Measured absorbanc e at Wavelength 2 (280nm).
RATIO : The ratio ABS 1and A BS 2
RESULT : Ratio result considere d factor Ratio. Inc ase no factor; it ’s a same value with ratio
column.
3. Stand ard Curve
4. ABS R atio
Move to
Mode Setup.
Wavelength
Input wavelength to measure, [Enter]
1
Select Type
Select the cur ve type wit h number 2 key then conrm wi th [Enter]
2
Manual Input
Select whethe r measuring or inputting
absorbance w ith number 3.
3
Data Clear
Delete all measure d data.
4
36
37
Part III.
Measuring in Each Test Modes
Part III.
Measuring in Each Test Modes
4) Flow Chart of ABS Ratio Test/ Mode Setup<Bottom Item>
MEASURE [F2]
Measure using inp ut the Wavelength, Cell
Number and Factor at the Mode Setup.
The result is indic ated like the left side gure.
2) Mode Setup
Reference Wavelength : Wavelength that will measure Reference, and default i s 320nm.
Wavelength 1 (numerator) : Wavelength that will measure t he numerator value of r atio.
Default is 260 nm.
Wavelength 2 (denominator) : Wavelength that will measure t he denominator valu e of
ratio. Default is 28 0nm.
Wavelength Referencing : Select whether measure or not Reference.
Factor : Factor that will be multiplied by Ratio.
3) Measure
You can measure with Measur e [F2] button af ter mode setup.
4. ABS R atio
4. ABS R atio
4
Move to ABS
ratio mode.
Start measuring after mo de setup.
F2
Mode Setup
Move to Mode Setup.
Input Reference wavelength then [EN TER].
1
Input factor that w ill be
multiplied by Rati o then [ENTER]
5
Select whethe r measure or
not Reference.
4
Input wavelength to measure the deno minator
ratio [ENTER].
3
Input wavelength to measure the numer ator of
ratio then [ENT ER].
2
Number of sample s to measure.
6
Measurement
38
39
Part III.
Measuring in Each Test Modes
Part III.
Measuring in Each Test Modes
5. Simple Kinetics
1) Test Mode Screen
At this mode, you ca n measure continuously absorbanc e variation at a wavele ngth during a
certain period of time.
2) Mode Setup
Wavelength : Input wavelength to measure. D efault Wavelength value is 6 00nm.
Total Scan Time : Input total scan time. Default dur ation
Interval Time : Input interval tim e to measure. But, Inter val time must lon ger than Total
Scan Time/241.
3) Measure
After Mode S etup, start measu ring with ME ASURE [F2].
You can see progressing r esult with absor bance as time variation.
After measur ing,MEAS URE [F2], GR APH [F3],
ABS/% [F4] butto n show up like the left s ide
gure.
MEASURE [F2]
You can measure again.
GRAPH [F3]
X-ZOOM [F1]
In order to look into g raphs in details, you c an
select the ran ge of wavelength. The modied
graphs by this mode w ill be kept.
- MIN X [F2] :
I n p u t t h e m i n i m u m v a l u e o f X - a x i s
to zoom.
- MAX X [F3] :
Input the maximum valu e of X-axis
to zoom.
- RECALL [F4] : I ni tia liz e gr ap h.
Y-ZOOM [F2]
Indicate the graph in detail.
5. Simple Kinetics
5. Simple Kinetics
40
41
Part III.
Measuring in Each Test Modes
Part III.
Measuring in Each Test Modes
4) Flow Chart of Simple Kinetics Test
CURSOR [F3]
You can see absorbance or transmittance as
time variation.
<<-- [F1], -->>[F4] key moves by 15s and <-
[F2], ->[F3] key moves by 1s.
TABLE [F4]
You can see absorbance or transmittance as
time by table.
5. Simple Kinetics
5. Simple Kinetics
5
Move to
Kinetics
Tes t M ode
Select a cell w ith
Up/Down but ton.
UP DOWN
X-ZOOM, Y-ZOOM, CURSOR, TABLE but ton shows up.
F3
Start measuring.
Incase using a Single cell, rst in sert
reference and me asure then inser t sample and measur e.
F2
Can conrm m easured data with
absorbance or transmittance
F4
You can setup the maximum an d minimum value of X-axis to zoom in. Also can recall gr aph with [RECALL] button.
F1
Indicate Y-axis in det ail.
F2
You can conrm data wi th
table.
F4
You can conrm data wi th cursor.
F3
F3F1 F2 F4
42
43
Part III.
Measuring in Each Test Modes
Part III.
Measuring in Each Test Modes
6. Communication
It is a mode for PC -Interfa ce Communications.
Press number key [6] in mai n menu before using O ptizenView(Optizen’s PC-Int erface
Software). You can exit this mode with [ESC] keys or [Disc onnect] butto n in OptizenView.
5) Flow Chart of Simple Kinetics Mode Setup
5. Simple Kinetics
6. Communication
Move to
Mode Setup
Wavelength
Input wavelength the n [Enter]
1
Input time to measur e then [Enter] (1~12 0M in)
2
Input interval t ime to measure, [Enter] (1~120Sec) : _ Sec
3
44
Part IV. Technical Supports
Part IV. Technical Supports
1. Technical Support
If you come acro ss any problems whi le installing or using Optizen, please refer to this manual
rst. If the pro blem still exists, p lease contact u s via telephone, fa x, e-mail an d our web sites.
We will solve it for you imm ediately.
1. Technical Su pport
Mecasys Co., Ltd.
Postal code: 30 5-150
10F, 640-3 Bansuk dong , Yuseong-gu, Daej eon, Rep. of KOREA.
Fax : +82-42- 485-7118
For technical se rvice E-mail : maintenance@ mecasys.co.kr
http://english.mecasys.co.kr/