Meatest M133C, M133Ci Operation Manual

M133C/133Ci Electric Power/Energy Calibrator
Operation manual
M133C Electric Power/Energy Calibrator MEATEST
Content
1 Basic information ....................................................................................................................... 5
2 Preparation for operation .......................................................................................................... 6
2.1 Inspecting package contents, selecting the installation location .................................... 6
2.2 Power-on .............................................................................................................................. 7
2.3 Warm-up time ..................................................................................................................... 8
2.4 Replacement of fuse ............................................................................................................ 8
2.5 Safety precautions .............................................................................................................. 9
3 Description of controls ............................................................................................................. 10
3.1 Front panel ........................................................................................................................ 10
3.2 Rear panel ......................................................................................................................... 15
4 Control of the calibrator .......................................................................................................... 16
4.1 Selection of function ......................................................................................................... 16
4.2 Setting the value of output signal .................................................................................... 17
4.3 Connection / disconnection of output terminals ............................................................ 18
4.4 Generation of electric power ........................................................................................... 20
Setting the power in modes Pdc Basic and Pac Basic ............................................................... 23
Setting the power in modes Pdc High I and Pac High I (Three phase version only) ................ 25
Setting the power in modes Pdc Extended and Pac Extended .................................................. 25
Setting the power in P Harmonic mode (model M133C only) ................................................. 27
Setting the power in P Interharmonic mode (model M133C only) ........................................... 29
Setting the power in P Dip/Swell mode (model M133C only) ................................................. 31
4.5 Generation of electric energy .......................................................................................... 33
Setting the energy in modes Edc Basic and Eac Basic .............................................................. 37
Setting the power in modes Edc High I and Eac High I (Three phase version only) ................ 38
4.6 Generation of calibrated voltage ..................................................................................... 38
Setting the voltage in modes Udc Basic and Uac Basic ............................................................ 39
4.7 Generation of calibrated current .................................................................................... 39
Setting the current in modes Idc Basic and Iac Basic ............................................................... 41
Setting the current in modes Idc High I and Iac High I (Three phase version only) ................. 41
5 Multimeter ................................................................................................................................ 42
5.1 Function selection ............................................................................................................. 42
6 Calibrator setup menu (Main menu) ...................................................................................... 43
6.1 General Menu ................................................................................................................... 43
6.2 Interface Menu ................................................................................................................. 44
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 4
6.3 Calibrator Menu ............................................................................................................... 45
6.4 Meter Menu ....................................................................................................................... 48
6.5 Calibration Menu.............................................................................................................. 48
7 Calibration mode ...................................................................................................................... 49
7.1 Calibration principles ....................................................................................................... 49
7.2 Access to the calibration procedure ................................................................................ 50
7.3 Selection of calibration type ............................................................................................. 51
7.4 Full calibration procedure ............................................................................................... 56
8 Error messages ......................................................................................................................... 60
9 Maintenance ............................................................................................................................. 62
10 Verification test ..................................................................................................................... 64
11 System control ....................................................................................................................... 70
11.1 IEEE488 bus properties ................................................................................................ 70
11.2 RS232 serial line setting ................................................................................................ 70
11.3 Command syntax ........................................................................................................... 71
11.4 Standard Status Data Structures ............................................................................... 138
12 Specification ........................................................................................................................ 142
13 Versions & Options ............................................................................................................. 148
14 Accessories ........................................................................................................................... 148
M133C Electric Power/Energy Calibrator MEATEST
1 Basic information
M133C Three (single) phase power calibrator is a multifunction calibrator, to be used primarily as a standard for calibration laboratories. It can be used for calibration of any measuring instrument which measures voltage, current, power, energy, phase and frequency. The calibrator is equipped with a built-in multimeter, which can be used simultaneously with output functions. Transducers of various types, regulators and sensing units can be therefore checked without the need for additional measuring instruments. M133C model is equipped with special functions for power line voltage analyzers testing. It can generate calibrated harmonic and interharmonic distortion, fluctuation harmonics, flickers, ramp signals and others. M133Ci model is delivered without the option of harmonic/interharmonic functions.
Basic features of the calibrator include generation of calibrated AC voltage in the range of 1 V to 600 V and DC voltage in the range of 1 V to 280 V, DC and AC current in the range of 5 mA to 30 A. Phase shift for all outputs (3x voltage and 3x current) can be set independent for each output. The best accuracy of the calibrator is 0.025 % for voltage and 0.035 % for current. Internal ranges (voltage 10, 30, 70, 140, 280 V and 600 Vac, current 0.3, 1, 2, 5, 10 and 30 A) are suitable for accurate testing of power transducers. Maximum frequency range is 15 Hz to 1000 Hz.
Internal multimeter with 20 mA, 10 V and 10 kHz (0.02 % accuracy) can be used to measure normalized signals coming from transducers.
The calibrator includes many other features which facilitate easy use. For example the possibility to save the actual setting into the internal memory, currently displayed uncertainty of the output signal, calibration and testing procedures etc. The concept of calibrator control and indication of its status is based on flat luminiscent display, which provides all necessary information. The calibrator is controlled by opening menus on the display and selection from menus. Frequently used functions are assigned direct-control keys. The calibrator comes with standard IEEE488 bus, RS-232 serial line and Ethernet connector, which allow the calibrator to be controlled from a PC.
The calibrator can easily fit within calibration systems featuring Power software support.
Calibrator is available in Three or Single phase version. Both versions look the same. The only difference is that the Single phase version doesn’t contain channels 2 and 3. Single phase version can be upgraded to Three phase version by the manufacturer.
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 6
2 Preparation for operation
2.1 Inspecting package contents, selecting the installation location
Basic package includes the following items:
M133C
Power/Energy calibrator Power cord Spare fuse Operational manual Test report Test cable 1000V/32A – 12 pcs (Three phase version), 4 pcs (Single phase version) RS232 cable M133C-01 High Current Adapter (Three phase version only)
The calibrator should be powered by 230/115 V – 50/60 Hz mains. It is a laboratory instrument whose parameters are guaranteed at 232 oC. Before powering on the instruments, place it on a level surface. Do not cover the vents at the bottom side and the fan opening at the rear panel.
M133C-01 High Current Adapter (Three phase version only)
Standard part of delivery the three­phase power calibrator is the option M133C-01. Using this adapter all currrent outpus are connected in parallel and the configuration can deliver up to 90 A in signle phase mode.
M133C Electric Power/Energy Calibrator MEATEST
2.2 Power-on
Before connecting the calibrator to the mains, check the position of the mains voltage selector
located at the rear panel.
Plug one end of the power cord into the connector located at the rear panel and connect the
other end of the power cord into a wall outlet.
Switch on the mains switch located at the rear panel. Flat display is lit.
The calibrator performs internal hardware checks for 5 seconds. Configuration of the
instrument is displayed in the end of this test (connected phase channels).
- phase unit is connected and ready
N.A. - phase unit is not connected
- phase is connected, but failed
After the tests conclude, the calibrator resets to its reference state, i.e. the following parameters
are set: Function AC power Pac Basic Voltage range 10 V Set value 10 V Current range 1 A Set value 1 A Phase Frequency 50 Hz Output terminals OFF
GPIB address of the calibrator is factory-preset to 2. This value is valid until the user changes it.
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 8
2.3 Warm-up time
The calibrator works after it is switched on and the initial checks complete. Specified parameters are only guaranteed after the instrument warms up for 60 minutes. During this period, the
instrument cannot be calibrated. The display shows “cannot access the calibration” message if
calibration is attempted during this period.
2.4 Replacement of fuse
The calibrator includes a fuse located in the mains connector at the rear panel. Replace the fuse as follows:
Switch off the calibrator Remove the end of power cord from the mains connector at the rear panel. Insert the blade of a flat screwdriver into the opening cut in the mains voltage selector and pull
out the fuse holder.
Remove the fuse and replace it with new fuse of the same rating.
M133C Electric Power/Energy Calibrator MEATEST
2.5 Safety precautions
The instrument has been designed in Safety Class I according to EN 61010-1. The design reflects the requirements of A2 amendment of the standard.
Safety is ensured by the design and by the use of specific component types. The manufacturer is not liable for the damage caused by modification of the construction or
replacement of parts with non-original ones.
Safety symbols used on the equipment
Warning, reference to the documentation
Warning - risk of electric shock
Danger - high voltage
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 10
3 Description of controls
3.1 Front panel
The front panel of the calibrator includes a TFT color display, control buttons and output terminals. The following picture shows the control part of the front panel.
1 Display buttons
There are five buttons below the display, whose meaning changes depending on the contents of the display. These buttons usually call-up the “local Menu” (units, modes, etc.).
2 Cursor buttons
Using these buttons, the cursor can be controlled within allowed limits on the display. The keyboard includes two buttons (<, >) which allow the cursor to be set to the required position at the display. The cursor can be moved to the left or right. These buttons are usually used to step through the options and to move from one option to another or between the menu levels. Numeric values can be set in some control modes as well. In these cases, the buttons marked (, ) allow the user to increase or decrease the number at the cursor button.
3
2 1 4 6 5 8 9
10
11
12
7
M133C Electric Power/Energy Calibrator MEATEST
The central button is used to select value you want to change (TAB).
3 Rotary knob
The rotary knob integrates several functions. By turning the knob to the left or right, the user can:
step through the options enter numeric values
The function of the rotary knob can usually be performed by the cursor buttons. The central button is used to confirm the selection (Select) or change the function of the rotary knob (value change or position change).
4 Numeric keyboard
The keyboard allows the entry of numeric values on the display. ENTER button is used to confirm the selection. CANCEL button can be used to cancel the entry.
5 Direct keyboard
MENU button is used to open the Main menu of the instrument (parameters petting). SAVE (RECALL) buttons can be used to save (recall) the actual setting of the instrument (all values including parameters). Capacity of the memory i sup to 100 different settings. Each setting contains simple text description.
6 Function buttons
Function buttons can be used to call-up the functions of the calibrator directly. The following buttons are provided:
function
key
Power
P
Energy
E
Voltage
U
Current
I
AC
AC
DC
DC
Output ON (Operate)
ON
Output OFF (Standby)
OFF
After the function mode is changed, the parameters of the respective function are restored. If the respective function was never used, the calibrator resets to its reference values.
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 12
7 Display
The display is divided to three horizontal sections:
A. Information line
Selected function (Pac Basic – Power AC basic mode) Date and time Remote control state (Local – keyboard, Remote – computer)
B. Main area
This section displays the set-up values of generated signals and the data related to the calibrator status. The section includes the following types of data:
1. Main value
There is displayed main output value with the unit and indication of AC or DC (picture). There is displayed also achal position of cursor ▼▲ if the parameter is in edit mode. Position of cursor can be changed using keys <, > and parameter can be changed using keys , , or using the rotary knob. If output voltage exceeds 50V sign of danger voltage is displayed before main value.
2. Auxiliary parameters
This section displays auxiliary parameters of actually selected function:
Voltage (functions power and energy) Current (functions power and energy) Phase / Power factor (functions power and energy) Frequency (AC functions) Power (energy) Dose of energy (energy)
3. Measured value
Value measured using internal multimeter with unit. Value „over“ is displayed if measured
signal is out of range.
A B1B2
B2
B2
B2
B3
C
B4
B5
B6
B7
M133C Electric Power/Energy Calibrator MEATEST
4. Outputs state
Window displays which phase units are connected (one phase to three phase), which terminals are used for signal generation and also informs about output voltage higher than 50V. Symbols displayed in window:
- warning, displayed if output voltage is higher or equal 50V.
- phase unit (channel) is connected, output is off
- phase unit (channel) is connected, output is on
- phase unit (channel) is not connected
5. Specification
In the field specification of set-up output signal is shown. Uncertainty is related to the main parameter. Label BUSY instead of accuracy is displayed, if calibrator is not within specified parameters or if internal reconnection is in process. This can occur during changing the functions, changing set parameters, switching output terminals ON and OFF. Specification is not available in power functions „P Harmonic“ and „P Interharmonic“.
6. Information section
The information section located in the right part of the display and displays additional information related to the selected function:
information about the grounding method of output terminals: GndU, GndI as set up
using the menu „Calibrator“.
information about voltage terminal configuration (2-wire / 4-wire) as set up using the
menu „Calibrator“.
information about the use of 25-turn coil or 50-turn coil (COIL x50) as set up using the
menu „Calibrator“.
information about number of controlled output channels (only functions „Power Pac“
and „Energy Eac“).
7. Meter’s information section
C. Display softkeys
The functions of these keys change during operation (depends on actual display mode).
8 Power switch
Turns instrument AC power ON and OFF.
9 Auxiliary input
Auxiliary input terminals for energy counting and for AC output signal synchronization.
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 14
10 Current output
Current outputs HI and LO terminals (three phases L1, L2 and L3).
11 Voltage output
Voltage outputs HI and LO terminals (three phases L1, L2 and L3).
12 Meter input
Input terminals for internal multimeter. The COM terminal is common for all measuring. The V terminal is for DC voltage (10V) and frequency (10kHz) measuring. The mA terminal is for DC current (20mA) measuring.
Colors on display
Common rules are used for applied color of labels and values.
1. Red color is applied, when displayed value is measured by the calibrator.
2. Blue color is applied for parameters or values, which can be set-up or modified directly
from front panel keyboard or via Main menu.
3. Black color is used for fix values, labels, notes, parameters which cannot be modified and
for other fix text with general information purpose.
4. Meaning of display buttons is always shown the lowest line. If there is no description
above display button, the button is not active in selected function.
M133C Electric Power/Energy Calibrator MEATEST
3.2 Rear panel
Rear panel includes ventilation holes, power cord socket, power line fuse, power line voltage selector 115/230V, fuse for internal multimeter, interface connectors GPIB, RS232 and RJ45, BNC outputs connectors and metal central ground terminal.
1 forced ventilation holes 2 energy pulse output - reference output (see Energy pulse output in Chapter 4.5) 3 sample reference output - manufacturing and service only 4 phase reference output - internal reference signal for 0 ° phase (all output channels are
synchronized with this signal) 5 GPIB, RS-232 and RJ45 connectors 6 central ground terminal 7 power line switch 8 fuse for internal multimeter (T100mA / 250V) 9 power line fuse 10 power line entry
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 16
4 Control of the calibrator
4.1 Selection of function
After the power is switched on and the initial checks complete, the calibrator resets to its reference state:
Function: Pac Basic Voltage: 10V Current: 1A Phase: Frequency: 50Hz
The state of the calibrator can be changed using the buttons located at the front panel in one of the following ways:
1. Change of function by pressing one of direct function buttons
After pressing one of the buttons P, E, U, I, DC, AC calibrator switches to the selected function and set the last set-up parameters. Whenever function is changed, calibrator always goes over to STANBY position. Output terminals are disconnected. All functions with direct access has submenus, where various basic modes of the selected function can be chosen. Access to the submenu is always vie display button Mode.
2. Connection /disconnection of output terminals
After pressing the ON button, the output terminals of the calibrator are connected. Only those terminals are activated which are aimed for DUT connection. To disconnect active terminals push the OFF button.
3. Access to the Main menu
After pressing the MENU button the calibrator displays the basic level of the setup menu. In this menu you can change instruments setting and calibration data. Previous function is recalled after pressing Exit softkey.
M133C Electric Power/Energy Calibrator MEATEST
4.2 Setting the value of output signal
Edit mode
Parameters of output signal can be changed in Edit mode. Only parameters displayed in blue color can be changed. Display can be switched to edit mode in different ways:
- Pressing numeric button
- Pressing „Sel“ button (in the middle of cursors buttons)
- Pressing cursor button
- Pressing the rotary knob
In edit mode is edited value highlighted by blue background. Pressing the „Sel“ button you can change among „blue“ parameters. Edit mode is finished by pressing „Exit“ softkey.
Entry of the value using numeric keyboard
Use the numeric keyboard to select the desired value. After the first digit is entered,
input box is displayed. In the upper row of the input box is the name of edited parameter. Using softkeys you can enter the new value in different units.
Enter desired value.
After the entry is complete press softkey with requested unit or press „Enter“ button.
„Enter“ button inputs the value in basic units (V, A, W, …).
Calibrator sets new value.
The value is copied to the appropriate field in the screen and the input box disappears.
Entry of the value using cursor buttons
Press , , or button. The display now includes cursor marks which points to the
active digit.
and buttons can be used to change the active digit. , buttons can be used to
change the position of the cursor marks
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 18
To get to the default screen, press „Exit“ button. All values can be set using the buttons
or the rotary knob.
Entry of the value using the rotary knob
Press the rotary knob. The display now includes cursor marks which points to the
active digit.
Turn the knob to change the active digit Press the rotary knob to change to the mode which allows to change the value of the
active digit. ← and → symbols are displayed above the active digit. Active digit can be changed by turning the knob.
Turn the knob to change the position of the active digit. To get to the default screen press „Exit“ button. All values can be set using the buttons
or the rotary knob.
Note:
All parameters have limits (high and low). If the entered value is outside these limits
error message is displayed („Value too large (small)“) and new value is not accepted.
4.3 Connection / disconnection of output terminals
After switching the AC power supply on the output terminals are disconnected in all modes.
Press the “On” button to connect the output signal to the terminals. Green LED above the “On”
button is lit. Press the “Off” button to disconnect the output terminals. Green LED goes off. Output terminals are disconected automaticaly under following conditions:
Function mode change Output voltage exceeds 100V. It occurs if you change the output value that is below
100V to the new value above 100V.
Calibrator is overloaded. Changing some parameters of output signal (P Harmonic, etc.) Changing frequency for voltage above 280V Energy counting is finished. Output voltage exceeds 280V while current outputs are floating. It occurs if you change
the output value that is below 280V to the new value above 280V and current outputs
M133C Electric Power/Energy Calibrator MEATEST
are not grounded (GndI Off). Calibrator disconnects output terminals, grounds the current terminals and displays following message:
Grounding state of current outputs will be recovered according to the Setup Menu automatically as soon as the voltage will drop below 280V.
Control sequence when output voltage over 100 V is selected When output voltage over 100 V is selected, the information section of the display shows the
symbol which informs that a life-threatening voltage will be present at the output terminals. If the output terminals are currently connected, they will be disconnected when output voltage over 100 V is selected. ON button must be pressed to reconnect the output signal to the output terminals. After the ON button is pressed, ON LED is lit and the information section of the display shows the symbol notifying the user about the connection of the dangerous output signal to the output terminals.
Voltage, and frequency can be set without the outputs being disconnected. The output terminals are automatically disconnected when changing between AC and DC ranges or when changing the function mode.
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 20
4.4 Generation of electric power
The calibrator can generate exact value of electric power and energy. Power function modes provides output voltage at VOLTAGE OUTPUT HI – LO terminals and output current at CURRENT OUTPUT HI – LO terminals. Depending on the setting of the calibrator, voltage up to 600 Vef can be present at the terminals.
Power setting range: 0.005 VA to 18 kVA (each channel) Voltage setting range: 1 to 600 Vac, 1 to 280 Vdc Current setting range: 5 mA to 30 A Power factor setting range: -1 to +1 (phase 0 to +359.99 °) Frequency setting range: DC, 15 Hz to 1 kHz
Control in the power mode
Press “P” button on the calibrator and then select AC or DC mode by pressing „AC“ or „DC”
button. The display shows the following data:
* main value of set power in selected unit of measurement VA, W, VAr * voltage at VOLTAGE OUTPUT HI-LO terminals * current through CURRENT OUTPUT HI-LO terminals
* power factor or the phase shift between the voltage and current in °. * frequency, if AC power is selected * uncertainty of set power
Set desired value of power using numeric keyboard, potentiometer or cursor buttons. Output
power is not yet connected to the output terminals. Connect the instrument to be calibrated to VOLTAGE OUTPUT HI-LO and CURRENT
OUTPUT HI-LO terminals or short the CURRENT OUTPUT HI-LO terminals.
Press OUTPUT button. Green LED is lit above the “On” button to indicate the connection of simulated electrical
power to the output terminals. Connected terminals are also displayed in the „Outputs state“
window as green circles.
Desired power is connected to output terminals.
M133C Electric Power/Energy Calibrator MEATEST
Power factor polarity (Lead/Lag softkey) In case that phase shift between voltage and current is expressed as power factor you can change
its polarity using „Lead/Lag“ softkey. The Lead polarity presents capacity load (current before
voltage). The Lag polarity presents inductance load (voltage before current).
Power units (Units softkey) The calibrator can display AC power in one of three ways:
apparent power in VA active power in W reactive power in VAr
Use “Units” softkey to open the units selection menu. Select the required expression using cursor buttons or the rotary knob. Confirm the selection by pressing the rotary knob or the “Select”
softkey. The value of displayed power is recalculated together with new units selection. DC power can be expressed only in W.
The calibrator can display the phase relation of output voltage and current as power factor (–1 to
+1) or as phase shift in degrees (0 to 359.99 °). Main menu is used to change the method of
displaying the phase relation.
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 22
Power modes (Mode softkey) Output power can be generated in different power modes: AC power
Mode
Name
Description
Basic
Pac Basic
AC power basic mode. Phase shift between channels output
voltages are fixed (120º and 240º). All parameters are the same
for all channels. Number of active channels can be defined in Main menu (1, 2 or 3).
High I
Pac High I
In this mode current outputs of all channels are connected in parallel (using the M133C-01 High Current Adapter). Current range is increased three times.
Extended
Pac
Extended
Parameters for all calibrators outputs (3x voltage, 3x current) can be set quite independently in this mode. For each output you
can define amplitude (V or A) and phase shift (º). Phase shift is
relative to the internal reference.
Harmonic
(version M133C
only)
P
Harmonic
Mode is used for generation harmonic signal composed of up to 50 harmonic components. All components can be defined independently for all outputs (3x voltage, 3x current). Beside this mode allows modulation by sine or square signal.
Interharmonic
(version M133C
only)
P
Iharmonic
Mode is used for generation interharmonic signal. One interharmonic component can be add independently for all outputs (3x voltage, 3x current).
Dip/Swell
(version M133C
only)
P
Dip/Swell
Mode is used for generation dips or swells on the output signal. All parameters can be defined independently for all outputs (3x voltage, 3x current).
DC power
Mode
Name
Description
Basic
Pdc Basic
DC power basic mode. Only first channel 1 is controlled in this mode.
High I
Pdc High I
In this mode current outputs of all channels are connected in parallel (using the M133C-01 High Current Adapter). Current range is increased three times.
Extended
Pdc
Extended
Parameters for all calibrators outputs (3x voltage, 3x current) can be set quite independently in this mode. For each output you can define amplitude (V or A).
M133C Electric Power/Energy Calibrator MEATEST
Use „Mode“ softkey to open the mode selection menu. Select the required mode using cursor buttons or the rotary knob. Confirm the selection by pressing the rotary knob or the “Select” softkey. Output terminals are disconnected if the new mode is selected.
Setting the power in modes Pdc Basic and Pac Basic
The calibrator allows several ways of setting the value of generated power.
1. Setting the main power value The main value can be changed using numeric keyboard, changing the digit at the
cursor position after selecting the cursor position with <, > buttons or using the rotary knob.
Output power is changed by changing the value of output current.
2. Setting the voltage
The main power value can be changed by changing the voltage. Repeatedly press the “Sel” button (in the middle of cursor buttons) until the voltage
value appears in edit mode (blue background).
The value can be set using numeric keyboard and confirmed by pressing V softkey. The
value can be set using cursor buttons or rotary knob as well.
Main power value is recalculated using new set voltage and existing setting of current
and power factor.
Note: Read also chapter 4.3 Connection / disconnection of output terminals.
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 24
3. Setting the current
The main power value can be changed by changing the current. Repeatedly press the “Sel” button (in the middle of cursor buttons) until the current
value appears in edit mode (blue background).
The value can be set using numeric keyboard and confirmed by pressing mA or A
softkey. The value can be set using cursor buttons or rotary knob as well.
Main power value is recalculated using new set current and existing setting of voltage
and power factor.
4. Setting the power factor (AC power only) If W or Var is indicated, the main power value can be changed by changing the power
factor. Change of power factor does not change the output apparent power.
Repeatedly press the “Sel” button (in the middle of cursor buttons) until the power
factor (phase) value appears in edit mode (blue background).
The value can be set using numeric keyboard and confirmed by pressing Lead or Lag
softkey. The value can be set using cursor buttons or rotary knob as well.
Main power value is recalculated using new set power factor and existing setting of
current and voltage. The calculation is only made if active or reactive power is displayed.
M133C Electric Power/Energy Calibrator MEATEST
Setting the power in modes Pdc High I and Pac High I (Three phase version only)
In this mode current outputs of all channels are connected in parallel using optional adapter. Current range is increased three times.
Voltage terminal Channel 1 are used in high current modes.
Power setting is the same as in modes Pdc Basic a Pac Basic.
Setting the power in modes Pdc Extended and Pac Extended
Modes allow independent parameter setting for all calibrators outputs (3x voltage, 3x current) For
each output you can define amplitude (V or A) and phase shift (º). Phase shift is relative to the
internal reference. The main power value displays summed power calculated as sum of power of all connected
channels. It is not possible to change this value directly. Color of this value is black. Each calibrators output (U1, U2, U3, I1, I2, I3) is displayed as a folder. The folder U1 is active
after selecting the Extended mode. Next folders are activated by repeatedly pressing „Channel“ softkey.
Pressing „On/Off“ softkey disables (enables) the active output. Only enabled outputs will be connected after pressing the „On“ button. Enabled outputs are highlighted by blue rectangle above
the output name. After the AC power on all outputs of connected phase units (channels) are enabled.
Picture displays state after AC power on. Both outputs of main phase unit are enabled, active is output U1.
1. Setting the amplitude of voltage (current)
The main power value can be changed by changing the amplitude of voltage (current). Repeatedly press the “Sel” button (in the middle of cursor buttons) until the voltage
(current) value appears in edit mode (blue background).
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 26
The value can be set using numeric keyboard and confirmed by pressing V (mA, A)
softkey. The value can be set using cursor buttons or rotary knob as well.
Main power value is recalculated using new amplitude. Next folders (outputs) are activated by repeatedly pressing „Channel“ softkey.
Note: Read also chapter 4.3 Connection / disconnection of output terminals.
2. Setting the phase (AC power only)
If W or Var is indicated, the main power value can be changed by changing the power
factor. Change of power factor does not change the output apparent power.
Repeatedly press the “Sel” button (in the middle of cursor buttons) until the phase
value appears in edit mode (blue background).
The value can be set using numeric keyboard and confirmed by pressing ° softkey. The
value can be set using cursor buttons or rotary knob as well.
Main power value is recalculated using new set phase and existing setting of current
and voltage. The calculation is only made if active or reactive power is displayed.
Next folders (outputs) are activated by repeatedly pressing „Channel“ softkey.
3. Setting the frequency (AC power only)
Repeatedly press the “Sel” button (in the middle of cursor buttons) until the frequency
value appears in edit mode (blue background).
M133C Electric Power/Energy Calibrator MEATEST
The value can be set using numeric keyboard and confirmed by pressing Hz softkey.
The value can be set using cursor buttons or rotary knob as well.
Setting the power in P Harmonic mode (model M133C only)
Mode is used for generation harmonic signal composed of up to 50 harmonic components. All components can be defined independently for all outputs (3x voltage, 3x current). Beside this mode allows modulation by sine or square signal.
Each calibrators output (U1, U2, U3, I1, I2, I3) is displayed as a folder. The folder U1 is active
after selecting the Harmonic mode. Next folders are activated by repeatedly pressing „Channel“
softkey.
Pressing „On/Off“ softkey disables (enables) the active output. Only enabled outputs will be connected after pressing the „On“ button. Enabled outputs are highlighted by blue rectangle above
the output name. After the AC power on all outputs of connected phase units (channels) are enabled.
Harmonics defined as “% of RMS Harmonics defined as “% of Fundamental
Pressing „Info“ button displays informative window. There is displayed used range for every output. Range is important for uncertainty calculation, because the uncertainty is not displayed in mode P Harmonic.
Internal multimeter is not available in P Harmonic mode.
1. Setting the amplitude of voltage (current)
Amplitude can be defined as RMS value (harmonics components are defined as % of RMS) or Fundamental harmonic value (harmonics components are defined as % of Fundamental harmonic).
Repeatedly press the “Sel” button (in the middle of cursor buttons) until the voltage
(current) value appears in edit mode (blue background).
The value can be set using numeric keyboard and confirmed by pressing V (mA, A)
softkey. The value can be set using cursor buttons or rotary knob as well.
Output terminals are disconnected if new amplitude is set. Next folders (outputs) are activated by repeatedly pressing „Channel“ softkey.
Note: Read also chapter 4.3 Connection / disconnection of output terminals.
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 28
2. Setting the phase
Repeatedly press the “Sel” button (in the middle of cursor buttons) until the phase
value appears in edit mode (blue background).
The value can be set using numeric keyboard and confirmed by pressing ° softkey. The
value can be set using cursor buttons or rotary knob as well.
Next folders (outputs) are activated by repeatedly pressing „Channel“ softkey.
3. Setting amplitude and phase of harmonic components
First should be selected required harmonic component (1-50). Repeatedly press the
“Sel” button (in the middle of cursor buttons) until the parameter Harmonic appears in edit mode (blue background).
Number of harmonic component can be set using numeric keyboard and confirmed by
pressing „Enter“ button. The value can be set using cursor buttons or rotary knob as
well. Next to the harmonic component number is displayed amplitude and phase of this harmonic component.
Pressing the „Sel“ button appears the amplitude of selected harmonic component in
edit mode (blue background). This is not valid for the fundamental harmonic (1. harmonic component) if harmonics components are deifned as % of RMS. Amplitude of fundamental harmonic is calculated using the RMS value of output signal and amplitudes of harmonic components 2-50 in this case.
Amplitude of harmonic components 2-50 can be set using numeric keyboard and
confirmed by pressing „%“ softkey. The value can be set using cursor buttons or rotary knob as well. Amplitude can be set in range 0 to 30.000%. Amplitude is expressed as % of RMS value or % of Fundamental harmonic.
Phase of harmonic components 2-50 can be set using numeric keyboard and confirmed
by pressing „°“ softkey. The value can be set using cursor buttons or rotary knob as well. Phase can be set in range 0.00 to 359.99 °.
„Harm+“ („Harm-“) softkeys facilitate listing among harmonics components. Softkeys
displays amplitude of next (previous) harmonic component.
„Ampl. “ („Phase “) softkeys switch between Amplitude and Phase edit mode for
selected harmonic component.
All harmonic components 2-50 can be cleared by pressing „Clear“ softkey. „Clear“
softkey is available if the fundamental harmonic is displayed.
Output terminals are disconnected if new amplitude is set.
M133C Electric Power/Energy Calibrator MEATEST
4. Setting of modulation
After AC power on modulation is switched off (OFF). First should be selected shape of
modulation signal. Repeatedly press the “Sel” button (in the middle of cursor buttons) until the parameter Modulation appears in edit mode (blue background). Shape can be
selected using „SIN“, „RECT“ or „OFF“ sofkey. Duty cycle (1 to 99%) is available for
rectangular modulation signal.
All harmonic components are modulated by modulation signal. Frequency of modulation signal is determinated by parameter „Modulation“.
Setting the power in P Interharmonic mode (model M133C only)
Mode is used for generation signal composed of fundamental harmonic and one interharmonic component. All parameters can be defined independently for all outputs (3x voltage, 3x current).
Each calibrators output (U1, U2, U3, I1, I2, I3) is displayed as a folder. The folder U1 is active after selecting the Interharmonic mode. Next folders are activated by repeatedly pressing
„Channel“ softkey.
Pressing „On/Off“ softkey disables (enables) the active output. Only enabled outputs will be connected after pressing the „On“ button. Enabled outputs are highlighted by blue rectangle above
the output name. After the AC power on all outputs of connected phase units (channels) are enabled.
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 30
Pressing „Info“ button displays informative window. There is displayed used range for every
output. Value of used range is important for uncertainty calculation, because the uncertainty is not displayed in P Interharmonic mode.
Internal multimeter is not available in P Interharmonic mode.
1. Setting the amplitude (RMS value) of voltage (current)
Repeatedly press the “Sel” button (in the middle of cursor buttons) until the voltage
(current) value appears in edit mode (blue background).
The value can be set using numeric keyboard and confirmed by pressing V (mA, A)
softkey. The value can be set using cursor buttons or rotary knob as well.
Output terminals are disconnected if new amplitude is set. Next folders (outputs) are activated by repeatedly pressing „Channel“ softkey.
2. Setting the phase
Repeatedly press the “Sel” button (in the middle of cursor buttons) until the phase
value appears in edit mode (blue background).
The value can be set using numeric keyboard and confirmed by pressing ° softkey. The
value can be set using cursor buttons or rotary knob as well.
Next folders (outputs) are activated by repeatedly pressing „Channel“ softkey.
3. Setting the amplitude (RMS value) of interharmonic component
Repeatedly press the “Sel” button (in the middle of cursor buttons) until the
interharmonic value appears in edit mode (blue background).
The value can be set using numeric keyboard and confirmed by pressing V (mA, A)
softkey. The value can be set using cursor buttons or rotary knob as well.
Output terminals are disconnected if new amplitude is set. Next folders (outputs) are activated by repeatedly pressing „Channel“ softkey.
M133C Electric Power/Energy Calibrator MEATEST
Setting the power in P Dip/Swell mode (model M133C only)
Mode is used for generation dips or swells on the output signal. All parameters can be defined independently for all outputs (3x voltage, 3x current).
Each calibrators output (U1, U2, U3, I1, I2, I3) is displayed as a folder. The folder U1 is active
after selecting the Dip/Swell mode. Next folders are activated by repeatedly pressing „Channel“
softkey.
Pressing „On/Off“ softkey disables (enables) the active output. Only enabled outputs will be connected after pressing the „On“ button. Enabled outputs are highlighted by blue rectangle above
the output name. After the AC power on all outputs of connected phase units (channels) are enabled.
Pressing „Trigger“ button starts generation of Dip/Swell shape. Generation can be started (triggered) also from input IN3. Dip/Swell shape is divided into five time periods. Length of these periods can be changed. Active period is during the generation shown in the box.
t1 Voltage (Current) amplitude is generated during period t1 after trigger t2 amplitude is fluently changed to Dip/Swell amplitude t3 Dip/Swell amplitude is generated t4 amplitude is fluently changed to Voltage (Current) amplitude t5 Voltage (Current) amplitude is generated
There are following trigger parameters in Setup menu: Dip/Swell repetition:
One Shot - trigger starts generation of one Dip/Swell shape Repeat - trigger starts repeated generation of Dip/Swell shapes
Dip/Swell synchronization: Sync Off - Dip/Swell shape starts immediately after trigger Sync On - start of Dip/Swell shape is synchronized with the phase 0°
Dip/Swell ext. trigger Input Off - Dip/Swell can be triggered only internal (keyboard or remote control) Input IN3 - Dip/Swell can be triggered internal or by external TTL falling edge signal applied
to input IN3
Internal multimeter is not available in P Dip/Swell mode.
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 32
1. Setting the amplitude (RMS value) of voltage (current)
Repeatedly press the “Sel” button (in the middle of cursor buttons) until the voltage
(current) value appears in edit mode (blue background).
The value can be set using numeric keyboard and confirmed by pressing V (mA, A)
softkey. The value can be set using cursor buttons or rotary knob as well.
Output terminals are disconnected if new amplitude is set. Next folders (outputs) are activated by repeatedly pressing „Channel“ softkey.
2. Setting the phase
Repeatedly press the “Sel” button (in the middle of cursor buttons) until the phase
value appears in edit mode (blue background).
The value can be set using numeric keyboard and confirmed by pressing ° softkey. The
value can be set using cursor buttons or rotary knob as well.
Next folders (outputs) are activated by repeatedly pressing „Channel“ softkey.
3. Setting the amplitude (RMS value) of Dip/Swell
Repeatedly press the “Sel” button (in the middle of cursor buttons) until the Dip/Swell
value appears in edit mode (blue background).
The value can be set using numeric keyboard and confirmed by pressing V (mA, A)
softkey. The value can be set using cursor buttons or rotary knob as well.
Output terminals are disconnected if new amplitude is set. Next folders (outputs) are activated by repeatedly pressing „Channel“ softkey.
4. Setting Dip/Swell time periods
Repeatedly press the “Sel” button (in the middle of cursor buttons) until reuiered time
period (t1, t2, t3, t4 or t5) appears in edit mode (blue background).
The value can be set using numeric keyboard and confirmed by pressing ms (s) softkey.
The value can be set using cursor buttons or rotary knob as well.
Output terminals are disconnected if new time is set. Next folders (outputs) are activated by repeatedly pressing „Channel“ softkey.
M133C Electric Power/Energy Calibrator MEATEST
4.5 Generation of electric energy
The calibrator can generate exact value of electric power and energy. Energy function modes provides output voltage at VOLTAGE OUTPUT HI – LO terminals and output current at CURRENT OUTPUT HI – LO terminals. Depending on the setting of the calibrator, voltage up to 600 Vef can be present at the terminals.
Power setting range: 0.005 VA to 18 kVA (each channel) Voltage setting range: 1 to 600 Vac, 1 to 280 Vdc Current setting range: 5 mA to 30 A Power factor setting range: -1 to +1 (phase 0 to +359.99 °) Time setting range: 1 to 10000 s Frequency setting range: DC, 15 Hz to 1 kHz
Energy pulse output (only for calibrators with this output) Energy pulse output is located on the rear panel (BNC connector in the right bottom corner).
Frequency of pulses is proportional to the energy generated by calibrator and is defined by the energy meter constant (can be changed in setup menu).
There are two types of output selectable in setup menu:
- open collector (up to 30V / 100mA)
- open collector with the pull-up resistor 150Ω connected to +5V Frequency range of energy pulse output: 0.02 Hz to 1 MHz
Control in the energy mode
Press “E” button on the calibrator and then select AC or DC mode by pressing „AC“ or „DC”
button. The display shows the following data:
* main value – running energy counter in selected unit of measurement VAs, Ws, VArs * deviation between energy meter and calibrator – calculation is based on pulses generated
by meter * voltage at VOLTAGE OUTPUT HI-LO terminals * current through CURRENT OUTPUT HI-LO terminals
* power factor or the phase shift between the voltage and current in °. * frequency, if AC power is selected * meter constant (number of pulses per kWh/kVAh/kVArh generated by energy meter)
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 34
* energy counting method (control mode) * uncertainty of energy and deviation
Set desired energy parameters using numeric keyboard, rotary knob or cursor buttons. Output
power is not yet connected to the output terminals.
Connect the instrument to be calibrated to VOLTAGE OUTPUT HI-LO and CURRENT
OUTPUT HI-LO terminals or short the CURRENT OUTPUT HI-LO terminals.
Select the Control mode. Press ON button. Green LED is lit above the “On” button to indicate the connection of simulated electrical
power to the output terminals. The calibrator starts energy dose generation. Main value displays running energy, control mode parameters displays running time, counted pulses or measured frequency. Current terminals are disconnected as soon as energy batch is finished. Voltage terminals are disconnected only if parameter “Maintain Voltage Signal” is set to “Off”. Otherwise the voltage terminals are disconnected after pressing “OFF” button.
Ways of energy counting (Control)
Calibration of Energy meter:
M133C Electric Power/Energy Calibrator MEATEST
There are different ways how to control the energy dose counting. Packet (Time counting)
The simplest (but less accurate) way is to count the time. Counter / Timer (Pulse counting)
The other way is to count pulses from energy meter. This method is called „rolling start“ method. Meter must be connected to input IN1 or IN2. Press the ON key on the calibrator. The calibrator
starts to count energy after counting “Warm-up” pulses from the meter (rolling start). Counting is finished after the set number of “Test” pulses is reached. The calibrator disconnects the output terminals and shows energy delivered between set number of “Test” pulses. Calibrator also
displays deviation of the meter in % (meter constant must be defined). In Counter mode is measuring interval defined by number of pulses. In Timer mode is measuring interval defined by time.
Free run (Frequency measuring) Calibrator measures frequency of pulses from measured meter. Meter must be connected to input IN1, IN2 or Meter. Deviation between meter and calibrator is continuously calculated and displayed (meter constant must be defined). Calibrator displays deviation between measured and internal "reference frequency" in %.
Use “Control” softkey to open the energy counting method selection menu. Select the required
method using cursor buttons or the rotary knob. Confirm the selection by pressing the rotary knob or the “Select” softkey.
Power factor polarity (Lead/Lag softkey) In case that phase shift between voltage and current is expressed as power factor you can change
its polarity using „Lead/Lag“ softkey. The Lead polarity presents capacity load (current before
voltage). The Lag polarity presents inductance load (voltage before current).
Power units (Units softkey) The calibrator can display AC power (energy) in one of three ways:
apparent power in VA (VAs) active power in W (Ws) reactive power in Var (VArs)
Use “Units” softkey to open the units selection menu. Select the required expression using cursor
buttons or the rotary knob. Confirm the selection by pressing the rotary knob or the “Select” softkey. The value of displayed power is recalculated together with new units selection. DC power (energy) can be expressed only in W (Ws).
The calibrator can display the phase relation of output voltage and current as power factor (–1 to
+1) or as phase shift in degrees (0 to 359.99 °). Main menu is used to change the method of
displaying the phase relation.
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 36
Energy modes (Mode softkey) Output energy can be generated in different energy modes: AC energy
Mode
Name
Description
Basic
Eac Basic
AC energy basic mode. Phase shift between channels output
voltages are fixed (120º and 240º). All parameters are the same for
all channels. Number of active channels can be defined in Main menu (1, 2 or 3).
High I
Eac High I
In this mode current outputs of all channels are connected in parallel (using the M133C-01 High Current Adapter). Current range is increased three times.
DC energy
Mode
Name
Description
Basic
Edc Basic
DC energy basic mode. Only first channel 1 is controlled in this mode.
High I
Edc High I
In this mode current outputs of all channels are connected in parallel (using the M133C-01 High Current Adapter). Current range is increased three times.
Use „Mode“ softkey to open the mode selection menu. Select the required mode using cursor buttons or the rotary knob. Confirm the selection by pressing the rotary knob or the “Select”
softkey. Output terminals are disconnected if the new mode is selected.
M133C Electric Power/Energy Calibrator MEATEST
Setting the energy in modes Edc Basic and Eac Basic
The calibrator allows several ways of setting the value of generated energy.
1. Setting the voltage
The energy (power) value can be changed by changing the voltage. Repeatedly press the “Sel” button (in the middle of cursor buttons) until the voltage
value appears in edit mode (blue background).
The value can be set using numeric keyboard and confirmed by pressing V softkey. The
value can be set using cursor buttons or rotary knob as well.
Energy (power) value is recalculated using new set voltage and existing setting of
current and power factor.
Note: Read also chapter 4.3 Connection / disconnection of output terminals.
2. Setting the current
The energy (power) value can be changed by changing the current. Repeatedly press the “Sel” button (in the middle of cursor buttons) until the current
value appears in edit mode (blue background).
The value can be set using numeric keyboard and confirmed by pressing mA or A
softkey. The value can be set using cursor buttons or rotary knob as well.
Energy (power) value is recalculated using new set current and existing setting of
voltage and power factor.
3. Setting the power factor (AC power only)
If Ws or VArs is indicated, the energy (power) value can be changed by changing the
power factor. Change of power factor does not change the output apparent energy.
Repeatedly press the “Sel” button (in the middle of cursor buttons) until the power
factor (phase) value appears in edit mode (blue background).
The value can be set using numeric keyboard and confirmed by pressing Lead or Lag
softkey. The value can be set using cursor buttons or rotary knob as well.
Energy (power) value is recalculated using new set power factor and existing setting of
current and voltage. The calculation is only made if active or reactive power is displayed.
4. Setting the frequency (AC energy only)
Repeatedly press the “Sel” button (in the middle of cursor buttons) until the frequency
value appears in edit mode (blue background).
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 38
The value can be set using numeric keyboard and confirmed by pressing Hz softkey.
The value can be set using cursor buttons or rotary knob as well.
5. Setting the Meter constant
Meter constant defines number of pulses sent by the meter for selected energy unit.
Repeatedly press the “Sel” button (in the middle of cursor buttons) until the Meter
constant value appears in edit mode (blue background).
The value can be set using numeric keyboard and confirmed by pressing Enter softkey.
The value can be set using cursor buttons or rotary knob as well.
6. Setting the time (test, warm-up counters)
Repeatedly press the “Sel” button (in the middle of cursor buttons) until the time value
(or “Test” / “Warm-up” values) appears in edit mode (blue background).
The value can be set using numeric keyboard and confirmed by pressing Enter softkey.
The value can be set using cursor buttons or rotary knob as well.
Setting the power in modes Edc High I and Eac High I (Three phase version only)
In this mode current outputs of all channels are connected in parallel (using the M133C-01 High Current Adapter). Current range is increased three times.
Output voltage terminals are on the main unit (Channel 1).
Power setting is the same as in modes Edc Basic a Eac Basic.
4.6 Generation of calibrated voltage
The calibrator can generate calibrated voltage. Output terminals for voltage ranges are VOLTAGE OUTPUT HI – LO terminals. Depending on the setting of the calibrator, voltage up to 600 Vef can be present at the terminals.
Voltage setting range: 1 to 600 Vac, 1 to 280 Vdc
Control in the voltage mode
Press “U” button on the calibrator and then select AC or DC mode by pressing „AC“ or „DC”
button. The display shows the following data:
* main voltage value at VOLTAGE OUTPUT HI-LO terminals * frequency, if AC voltage is selected * uncertainty of voltage
M133C Electric Power/Energy Calibrator MEATEST
Set desired value of voltage using numeric keyboard, potentiometer or cursor buttons. Output
voltage is not yet connected to the output terminals.
Connect the instrument to be calibrated to VOLTAGE OUTPUT HI-LO. Press ON button. Green LED is lit above the “On” button to indicate the connection of output terminals.
Note: Read also chapter 4.3 Connection / disconnection of output terminals.
Overloading of output terminals
If the output terminals are overloaded or short-circuited in the voltage mode, the calibrator disconnects the signal from the output terminals and reports „Output Overload“.
Setting the voltage in modes Udc Basic and Uac Basic
The calibrator allows several ways of setting the value of generated voltage.
1. Setting the voltage
Repeatedly press the “Sel” button (in the middle of cursor buttons) until the voltage
value appears in edit mode (blue background).
The value can be set using numeric keyboard and confirmed by pressing V softkey. The
value can be set using cursor buttons or rotary knob as well.
2. Setting the frequency (AC voltage only)
Repeatedly press the “Sel” button (in the middle of cursor buttons) until the frequency
value appears in edit mode (blue background).
The value can be set using numeric keyboard and confirmed by pressing Hz softkey.
The value can be set using cursor buttons or rotary knob as well.
Note: Read also chapter 4.3 Connection / disconnection of output terminals.
4.7 Generation of calibrated current
The calibrator can generate calibrated current. Output terminals for current ranges are CURRENT OUTPUT HI – LO terminals.
Current setting range: 5 mA to 30 A
When 25/50-turn coil (option 140-50) is used, AC current range is to 1000 A.
Control in the current mode
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 40
Press “I” button on the calibrator and then select AC or DC mode by pressing „AC“ or „DC”
button. The display shows the following data:
* main current throught the CURRENT OUTPUT HI-LO terminals * frequency, if AC current is selected * uncertainty of current
Set desired value of current using numeric keyboard, rotary knob or cursor buttons. Output
current is not yet connected to the output terminals.
Connect the instrument to be calibrated to CURRENT OUTPUT HI-LO. Press ON button. Green LED is lit above the “On” button to indicate the connection of output terminals. If Coil x50 or Coil x25 function is activated (see below – Main menu), the optional current coil
must be connected to the output terminals. The calibrator can be used to calibrate clamp Ameters up to 1500 A. The calibrator generates AC and DC current within the range up to 20 A with current coil.
Overloading of output terminals
When external circuit connected to current output terminals is disconnected or there is higher voltage at the load than permitted, the calibrator disconnects the output terminals and displays „Output Overload“ message. The same message can be displayed when 50-turn coil is used for AC current output at frequencies above 80 Hz. It depends on the set current and the type of ammeter connected.
Current modes (Mode softkey) Output current can be generated in different current modes: AC current
Mode
Name
Description
Basic
Iac Basic
AC current basic mode. Only first channel 1 is controlled in this mode.
High I
Iac High I
In this mode current outputs of all channels are connected in parallel (using the M133C-01 High Current Adapter). Current range is increased three times.
M133C Electric Power/Energy Calibrator MEATEST
DC current
Mode
Name
Description
Basic
Idc Basic
DC current basic mode. Only first channel 1 is controlled in this mode.
High I
Idc High I
In this mode current outputs of all channels are connected in parallel (using the M133C-01 High Current Adapter). Current range is increased three times.
Use „Mode“ softkey to open the mode selection menu. Select the required mode using cursor buttons or the rotary knob. Confirm the selection by pressing the rotary knob or the “Select” softkey. Output terminals are disconnected if the new mode is selected.
Setting the current in modes Idc Basic and Iac Basic
The calibrator allows several ways of setting the value of generated current.
1. Setting the current
Repeatedly press the “Sel” button (in the middle of cursor buttons) until the current
value appears in edit mode (blue background).
The value can be set using numeric keyboard and confirmed by pressing mA or A
softkey. The value can be set using cursor buttons or rotary knob as well.
2. Setting the frequency (AC current only)
Repeatedly press the “Sel” button (in the middle of cursor buttons) until the frequency
value appears in edit mode (blue background).
The value can be set using numeric keyboard and confirmed by pressing Hz softkey.
The value can be set using cursor buttons or rotary knob as well.
Setting the current in modes Idc High I and Iac High I (Three phase version only)
In this mode current outputs of all channels are connected in parallel (using the M133C-01 High Current Adapter). Current range is increased three times.
Current setting is the same as in modes Idc Basic a Iac Basic.
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 42
5 Multimeter
The calibrator includes a built-in multimeter which can measure DC voltage, DC current and frequency. Measured signal must be connected to the “METER INPUT” terminals. Terminals mA, COM are for current measurements. Terminals V, COM are for voltage and frequency measurements.
5.1 Function selection
Measurement funkction should be selected in Main menu (item Meter). Press UP, DOWN display buttons or use the rotary knob to select one of the following
functions: * Voltage – DC voltage range 10V
* Current – DC current range 20mA * Frequency – frequency range 10 kHz
Notes:
* The multimeter can be used as a DC multimeter with ranges 10V and 20mA. * Frequency measurement is possible up to 15 kHz. Input signal must be within 0.2 to 5 V
range. Square wave or pulse shape of the input signal is expected.
The manufacturer does not recommend connection of multimeter inputs to
calibrator outputs.
Such connection can result in presence of high voltage at multimeter inputs,
which can damage the multimeter.
M133C Electric Power/Energy Calibrator MEATEST
6 Calibrator setup menu (Main menu)
The calibrator allows many other, less frequently used parameters to be set. Setup menu is used to set these parameters. Setup menu is opened by pressing MENU button. If output terminals are connected, they will be disconnected and the following display appears:
Use or cursor button or the rotary knob to browse the menu options. Press „Select“ softkey or
press in on the rotary knob to select highlighted item. Press „Exit“ softkey to leave the selected
menu level. New settings are retained when the calibrator is switched off. The Main menu contains items:
* General – general instruments settings (display, keyboard, time, date) * Interface – parameters of instruments remote control * Calibrator – parameters of signal generation * Meter – parameters of multimeter * Calibration – instruments calibration data
6.1 General Menu
Submenu contains basic parameters of display and keyboard.
1. Volume
This parameter sets the beeper volume. Cursor buttons, rotary knob or numeric keyboard allow to set the value in the range of 0 to 15. Press the Write softkey to set the beeper volume and return to the
General menu.
2. Brightness
This parameter sets the display brightness. Cursor buttons, rotary knob or numeric keyboard allow to set the value in the range of 0 to 7. Press the Write softkey to set the display’s brightness and return to the General menu.
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 44
3. Beeper
This parameter enables / disables the signalization of pressed key. Possible states are „Beep On“ and „Beep Off“. Press the Select softkey or press in on the rotary knob to set the selected state.
4. Calibration Password
This parameter sets the calibration password. Calibration password is a five-digit number, which must be entered to access the calibration mode. If the calibration password is set to 0, this information is displayed in the General menu. Other values are displayed as „Secret“.
You will be prompted to enter the present calibration password (use the numeric keybord and přes Enter). Now you can change the calibration password. Cursor buttons, rotary knob or numeric keyboard allow to set the value in the range of 0 to 99999. Press the Write softkey to set the display’s brightness and return to the General menu.
Note: It is advisable to write down actual calibration code if changed. If you forget the calibration code, you have to send the calibrator to the manufacturer.
5. Time
Setting the real time. The parameter can be changed using cursor buttons, rotary knob or numeric keyboard.
6. Date
Setting the date. The parameter can be changed using cursor buttons, rotary knob or numeric keyboard.
7. Device Information
Viewing the device information. Displays the serial number and software version.
6.2 Interface Menu
Submenu contains remote control parameters.
1. Active interface
Sets the type of interface used to calibrator’s remote control. By selecting an item from the list
IEEE488, RS232 or Ethernet type can be selected. The calibrator can be remotely controlled only using the selected interface.
2. IEEE488 address
Sets the calibrator’s IEEE488 (GPIB) address. Cursor buttons, rotary knob or numeric keyboard allow to set the value in the range of 0 to 30. Press the Write softkey to set the address and return to the
Interface menu. Address 02 is set by the manufacturer.
3. Baud rate
Sets the communication speed of RS232 bus. By selecting an item from the list 1200, 2400, 4800, 9600, 19200, 38400, 76800 or 115200 Bd can be selected.
M133C Electric Power/Energy Calibrator MEATEST
4. Ethernet settings
Unique fix IP address.
* IP address * Local network mask * Local network gateway
6.3 Calibrator Menu
Submenu contains parameters that affect generated signal.
1. Voltage sources GND
This parameter allows to connect Lo terminals of all voltage channels to GND. In practice this means that Lo voltage terminals are grounded. By selecting an item from the list GndU Off and GndU On type can be selected.
2. Current sources GND
This parameter allows to connect Lo terminals of all current channels to GND. In practice this means that Lo current terminals are grounded. By selecting an item from the list GndI Off and GndI On type can be selected.
It is recommended to ground all output channels (GndU On, GndI On). If the meter to be calibrated has Lo terminal grounded, it is recommended to unground the corresponding
calibrator’s output to exclude ground loops. Note
If neither the calibrator’s output, nor the meter’s inputs are grounded, signal/noise ratio can arise at the calibrator’s output.
3. Phase unit
Sets the unit used for presentation the phase shift between the voltage and current output in the
power and energy generation modes. By selecting an item from the list Deg (º) and Cos (Lead,
Lag) unit can be selected.
4. Phase rotation
Phase rotation setting. By selecting an item from the list POS (standard phase) and NEG (reversed phase) rotation can be selected.
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 46
5. 280V range lock
- OFF position (default state)
140V range is used for values between 70.001V and 140.000V 280V range is used for values between 140.001V and 280.000V
- ON position
280V range is used for values between 70.001V and 280.000V
Feature is active for single harmonic modes only.
6. Current coil
Sets the calibrator for connection the 25/50-turn current coil (clamp ammeters calibration). The coil multiplies the output current either 25 times or 50 times depending on set parameter. By selecting an item from the list Coil Off, Coil x25 and Coil x50 type of connected current coil can be selected.
7. Synchronization
Sets the external synchronization of the output signal. By selecting an item from the list Internal (internal synchronization), Power line (synchronization to the power supply), External IN1 (synchronization to the passive signal on input 1 – output colector or relay contact), External IN2 (synchronization to the active signal on input 2) a External IN3 (synchronization to the active signal on input 3) type of frequency synchronization coil can be selected.
Note:
External synchronized signal has larger distortion than internal synchronized signal because of fine frequency tuning.
8. Active channels (Pac, Eac)
Sets the number of controlled channels in the three phase configuration. This parameter is valid only for Pac Basic and Eac Basic function modes. By selecting an item from the list 1 (output is from the main channel only), 1-2 (output from Channel 1 and Channel 2) and 1-2-3 (output from all three channels) number of used channels can be selected.
9. Harmonic components
Harmonic components can be entered in two ways:
1) % of RMS value
RMS value is constant (isn’t changed by changing harmonic components value). New RMS value changes amplitude of all harmonic components.
2) % of fundamental
Fundamental value is constant (isn’t changed by changing harmonic components value).
RMS value is changed by entering new harmonic components value.
M133C Electric Power/Energy Calibrator MEATEST
10. Energy
Submenu contains parameters that affect Energy generation.
8.1. Energy units
Sets the unit used for energy. By selecting an item from the list Ws and Wh unit can be selected.
8.2. Energy - output
Contains parameters for setting the energy pulse output (BNC connector on the rear panel).
8.2.1. Internal pull-up
This parameter allows to connect the internal pull-up resistor (150 Ω) to the Energy pulse output.
8.2.2. Output constant
Contains parameters for definition the constant of Energy pulse output.
8.2.2.1. Value
This parameter sets number of pulses per selected unit (pulses generated on Energy pulse output on the rear panel BNC).
8.2.2.2. Unit
This parameter sets unit of Output constant value (pulses generated on Energy pulse output on the rear panel BNC).
8.3. Energy – input IN1
Contains parameters for setting the energy pulse input IN1.
8.3.1. Internal pull-up
This parameter allows to connect the internal pull-up resistor (150 Ω or 1 kΩ) to the Energy pulse input.
8.3.2. Input filter
This parameter allows to add input filter for contact debouncing. Input frequency with the input filter is limited to 400 Hz.
8.4. Maintain Voltage Signal
This parameter allows to maintain the voltage signal on test completion in energy counting modes (Packet / Counter / Timer).
11. Dip-Swell
Submenu contains parameters that affect Dip-Swell generation.
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 48
9.1. Dip-Swell repetition
Sets the repetition of Dip/Swell shape in P Dip/Swell mode. By selecting an item from the list One Shot (without repetition) and Repeat can be repetiotion selected. If Repeat is selected, Dip/Swell shape is generated after trigger repeatedly until disconnecting output terminals.
9.2. Dip-Swell synchronization
Sets the synchronization of Dip/Swell shape with internal phase 0°. By selecting an item from the list Sync Off and Sync On can be synchronization selected. If Sync Off is selected, Dip/Swell generation starts immediately after trigger. If Sync On is selected, Dip/Swell is synchronized with the internal phase 0° (channels with phase shift 0°).
9.3. Dip-Swell ext. trigger
Enables external triggering in P Dip/Swell mode. By selecting an item from the list Input Off and Input IN3 can be external trigger selected. If Input Off is selected, Dip/Swell generation can be started only internal (keyboard or remote control). If Input IN3 is selected, Dip/Swell generation can be started internal or by falling edge applied to the input IN3. The input must remain low for 10µs after the falling edge to be recognized properly.
6.4 Meter Menu
Submenu contains parameters that affect measured signal.
1. Function
Sets the internal multimeter measuring function and range. By selecting an item from the list Voltage (DC voltage 0-12V), Current (DC current 0-24mA) and Frequency (Frequency 1Hz­15kHz) function can be selected.
6.5 Calibration Menu
Calibration password is required to access the Calibration Menu. Default calibration password is „0”. For more detail about calibration see the chapter 7.
M133C Electric Power/Energy Calibrator MEATEST
7 Calibration mode
The calibrator includes calibration procedure, which allows recalibration of the calibrator. Recalibration can be performed using the buttons and menu in the calibrator.
7.1 Calibration principles
The instrument can be calibrated:
completely, i.e. all functions are calibrated in all recommended points partially, i.e. only selected functions are calibrated in all recommended points partially, i.e. only selected functions are calibrated in selected points
Complete calibration consists of all partial calibrations performed in the order defined by the calibration menu. If an item of the calibration menu, e.g. „Voltage DC#1” is selected, it is not necessary to calibrate all ranges defined by the calibration algorithm. If new calibration of all ranges is not possible (e.g. the required standard is not available), old calibration data can be confirmed, i.e. current step of the calibration can be skipped.
AC voltage (Voltage AC#x) calibration is performed by setting 2 fixed points for all voltage ranges. The frequency is 55 Hz.
AC current (Current AC#x) calibration is performed by setting 2 fixed points for all current ranges. The frequency is 55 Hz.
DC voltage (Voltage DC#x) calibration is performed by setting 2 fixed points for all voltage ranges and for both polarities (+ and -).
DC current (Current DC#x) calibration is performed by setting 2 fixed points for all current ranges and for both polarities (+ and -).
AC modulation voltage (Voltage MOD#x, model M133C only) calibration is performed by setting 2 fixed points for all voltage ranges. The frequency is 55 Hz. This calibration values are used in P Harmonic and P Interharmonic function modes.
AC modulation current (Current MOD#x, model M133C only) calibration is performed by setting 2 fixed points for all current ranges. The frequency is 55 Hz. This calibration values are used in P Harmonic and P Interharmonic function modes.
Multimeter (Meter) calibration consist in calibration of ranges 10V DC, 20mA DC and 10 kHz.
Calibration interruption can be performed in any point of the calibration
procedure. However this particular calibration influences parameters of the
calibrator.
Accuracy of the calibrator is guaranteed when full calibration was done.
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 50
7.2 Access to the calibration procedure
Calibration password is required to access the calibration procedure.
Press Menu to open the Main menu. Select the Calibration submenu and press the „Select“ softkey. Calibrator requests the entry of the calibration password.
Enter the correct calibration code using numeric keyboard and press ENTER. If incorrect calibration code is entered, an error message appears on the display for
approximately 3 seconds: Bad calibration code!
If correct calibration code is entered, calibration menu with following list appears.
Use and cursor buttons to move the cursor through the list of calibration items.
1. Channel #1
Calibration data of the phase channel 1.
2. Channel #2
Calibration data of the phase channel 2.
3. Channel #3
Calibration data of the phase channel 3.
M133C Electric Power/Energy Calibrator MEATEST
Every item from the list above has following submenu:
Voltage AC# x - AC voltage calibration
Current AC# x - AC current calibration
Voltage DC# x - DC voltage calibration
Current DC# x - DC current calibration
Voltage MOD# x - AC voltage in modes P Harmonic a P Interharmonic calibration
Current MOD# x - AC current in modes P Harmonic a P Interharmonic calibration
Where x is ordinal number of the phase unit (channel).
4. Meter
Calibration data of internal meter.
7.3 Selection of calibration type
After the calibration menu is displayed, one of partial calibrations can be selected. Use and cursor buttons to move the cursor through the list. Having selected the required function to be calibrated, press Select softkey. The following data are shown (the following example is valid for Voltage AC#1):
The table lists recommended calibration points. Having selected the required calibration point using Select softkey, the following data are shown.
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 52
Softkeys have the following meaning: Write new calibration value is entered into the memory, old value is lost. The calibrator
returns to the previous menu.
Skip current calibration step is skipped, old value is retained in the memory. The calibrator
returns to the previous menu.
Moreover, the display shows the range which is being calibrated (Range 10Vac), and the value to be set at the external standard multimeter (value in brackets 1V).
Setting the new calibration data Use cursor buttons, rotary knob or numeric keyboard to set such main data on the display, when
the output signal measured by external standard multimeter reaches the required calibration point. When the standard output value is reached, press Write softkey to write new calibration value to the calibration memory. If you press Exit softkey, the calibrator ignores the new value and old value is retained.
The procedure is repeated for all calibration points of the selected function. If you press Exit softkey the calibrator returns back to the previous menu level.
Termination of calibration Calibration can be terminated by repeated press the Exit softkey until the calibrator returns to the
Main menu.
Calibration points Each function of the calibrator has assigned fixed calibration points which have to be set during
the calibration. The calibrator needs no calibration of the following parameters:
frequency phase relationship (power factor) of output voltage and current in AC power and AC
energy generation modes
DC and AC power and energy
M133C Electric Power/Energy Calibrator MEATEST
Voltage AC function (Voltage AC#x)
Nominal value [V]
Set limits [V]
Range [V]
1
100 u
10
10
500 u
10
10
1 m
30
30
2 m
30
30
4 m
70
70
4 m
70
70
7 m
140
140
7 m
140
140
15 m
280
280
15 m
280
280
40 m
600
600
40 m
600
Table AC voltage
Current AC function (Current AC#x)
Nominal value[A]
Set limits [A]
Range [A]
30 m
10 u
300 m
300 m
15 u
300 m
300 m
30 u 1 1
50 u 1 1
100 u 2 2
100 u 2 2
200 u 5 5
250 u 5 5
500 u
10
10
500 u
10
10
1 m
30
30
1.5 m
30
Table AC current
Voltage DC function (Voltage DC#x)
Nominal value [V]
Set limits [V]
Range [V]
1
100 u
10
10
500 u
10
-1
100 u
-10
-10
500 u
-10
10
1 m
30
30
2 m
30
-10
1 m
-30
-30
2 m
-30
30
4 m
70
70
4 m
70
-30
4 m
-70
-70
4 m
-70
70
7 m
140
140
7 m
140
-70
7 m
-140
-140
7 m
-140
140
15 m
280
280
15 m
280
-140
15 m
-280
-280
15 m
-280
Table DC voltage
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 54
Current DC function (Current DC#x)
Nominal value [A]
Set limits [A]
Range [A]
30 m
10 u
300 m
300 m
15 u
300 m
-30 m
10 u
-300 m
-300 m
15 u
-300 m
300 m
30 u 1 1
50 u
1
-300 m
30 u
-1
-1
50 u
-1
1
100 u
2
2
100 u
2
-1
100 u
-2
-2
100 u
-2 2 200 u 5 5
250 u 5 -2
200 u
-5
-5
250 u
-5 5 500 u
10
10
500 u
10
-5
500 u
-10
-10
500 u
-10
10
1 m
30
30
1.5 m
30
-10
1 m
-30
-30
1.5 m
-30
Table DC current
Voltage AC modulation function (Voltage MOD#x)
Nominal value [V]
Set limits [V]
Range [V]
1
100 u
10
10
500 u
10
10
1 m
30
30
2 m
30
30
4 m
70
70
4 m
70
70
7 m
140
140
7 m
140
140
15 m
280
280
15 m
280
Table MOD AC voltage
M133C Electric Power/Energy Calibrator MEATEST
Current AC modulation function (Current MOD#x)
Nominal value[A]
Set limits[A]
Range [A]
30 m
10 u
300 m
300 m
15 u
300 m
300 m
30 u 1 1
50 u 1 1
100 u 2 2
100 u 2 2
200 u 5 5
250 u 5 5
500 u
10
10
500 u
10
10
1 m
30
30
1.5 m
30
Table MOD AC current
Multimetr (Meter)
Nominal value [-]
Set limits [-]
Range [-]
0 V
50 u
10 V
10 V
200 u
10 V
0 mA
50 nA
20 mA
20 mA
500 nA
20 mA
10 kHz
200 mHz
10 kHz
Table Multimeter
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 56
7.4 Full calibration procedure
Required instruments
Following pages describe procedure of the full calibration. Following instruments are required for calibration:
81/2 digit multimeter type HP3458A or Fluke 8508A, or other type with accuracy
0.001 % on DC voltage
Calibrator Meatest M142 or M140 with DC ranges 10V, 20mA a AC range 10V /
10kHz.
Resistance shunt 10 m, 100 m Burster 1280, or other type with accuracy 0.01% Counter HP 53181A, HO 53130, BM 642 or other with accuracy 0,001 % Powermeter 0.02- 0.05 %, Zimmer LG95 or better
HP8903A Distortion analyzer and scope with bandwidth min. 20 MHz are recommended for THD measuring of AC signals.
Calibration procedure
1. Connect the calibrator and the multimeter to the mains and let them switched on for at least
three hours in a laboratory at 231 oC.
2. Press Menu button to call up the Main menu, select Calibration end press Select softkey to
enter the calibration menu.
3. Enter the calibration code and press Enter button (default calibration code is “0”).
4. AC voltage ranges calibration a) Connect the voltage input terminals of the multimeter to the VOLTAGE OUTPUT
HI - LO terminals of the calibrator.
b) Select Voltage AC#1 from the calibration menu and confirm by pressing Select
softkey. Switch calibrator’s output terminals ON.
c) Follow the instructions provided on the calibrator’s display and the ACU table to
adjust the calibrator’s output in the calibration points.
d) To adjust the calibrator’s output in the calibration points, press Select softkey and
use , , , cursor buttons, rotary knob or numeric keyboard to adjust the output voltage. Confirm correctly set value by pressing Write softkey. If you want to skip the calibration point whose calibration you have already entered, press Exit softkey.
e) Switch output terminals OFF. Disconnect multimeter and calibrator. f) Repeat steps a) to e) also for Voltage AC#2 and Voltage AC#3.
M133C Electric Power/Energy Calibrator MEATEST
5. AC current ranges calibration a) Select Current AC#1 from the calibration menu and confirm by pressing Select
softkey. Set the same function on external multimeter.
b) Follow the instructions provided on the calibrator’s display and the ACI table to
adjust the calibrator’s output in the calibration points.
c) To adjust the calibrator’s output in the calibration points, press Select softkey and
use , , , cursor buttons, rotary knob or numeric keyboard to adjust the output voltage. Confirm correctly set value by pressing Write softkey. If you want to skip the calibration point whose calibration you have already entered, press Exit softkey.
d) Resistance shunt should be used on 2, 5, 10 and 30 A ranges, if standard multimeter
does not cover this range.
e) Repeat steps a) to d) also for Current AC#2 and Current AC#3.
6. DC voltage ranges calibration a) Select Voltage DC#1 from the calibration menu and confirm by pressing Select
softkey. Switch calibrator’s output terminals ON.
b) Follow the instructions provided on the calibrator’s display and the DCU table to
adjust the calibrator’s output in the calibration points.
c) To adjust the calibrator’s output in the calibration points, press Select softkey and
use , , , cursor buttons, rotary knob or numeric keyboard to adjust the output voltage. Confirm correctly set value by pressing Write softkey. If you want to skip the calibration point whose calibration you have already entered, press Exit softkey.
d) Switch output terminals OFF. Disconnect multimeter and calibrator. e) Repeat steps a) to d) also for Voltage DC#2 and Voltage DC#3.
7. DC current ranges calibration a) Select Current DC#1 from the calibration menu and confirm by pressing Select
softkey. Set the same function on external multimeter.
b) Follow the instructions provided on the calibrator’s display and the DCI table to
adjust the calibrator’s output in the calibration points.
c) To adjust the calibrator’s output in the calibration points, press Select softkey and
use , , , cursor buttons, rotary knob or numeric keyboard to adjust the output voltage. Confirm correctly set value by pressing Write softkey. If you want to skip the calibration point whose calibration you have already entered, press Exit softkey.
d) Resistance shunt should be used on 2, 5, 10 and 30 A ranges, if standard multimeter
does not cover this range.
e) Repeat steps a) to d) also for Current DC#2 and Current DC#3.
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 58
8. AC voltage modulation ranges calibration (P Harmonic, P Interharmonic modes,
model M133C only)
a) Select Voltage MOD#1 from the calibration menu and confirm by pressing Select
softkey. Switch calibrator’s output terminals ON.
b) Follow the instructions provided on the calibrator’s display and the MOD ACU
table to adjust the calibrator’s output in the calibration points.
c) To adjust the calibrator’s output in the calibration points, press Select softkey and
use , , , cursor buttons, rotary knob or numeric keyboard to adjust the output voltage. Confirm correctly set value by pressing Write softkey. If you want to skip the calibration point whose calibration you have already entered, press Exit softkey.
d) Switch output terminals OFF. Disconnect multimeter and calibrator. e) Repeat steps a) to d) also for Voltage MOD#2 and Voltage MOD#3.
9. AC current modulation ranges calibration (P Harmonic, P Interharmonic modes,
model M133C only) a) Select Current MOD#1 from the calibration menu and confirm by pressing Select
softkey. Set the same function on external multimeter.
b) Follow the instructions provided on the calibrator’s display and the MOD ACI table
to adjust the calibrator’s output in the calibration points.
c) To adjust the calibrator’s output in the calibration points, press Select softkey and
use , , , cursor buttons, rotary knob or numeric keyboard to adjust the output voltage. Confirm correctly set value by pressing Write softkey. If you want to skip the calibration point whose calibration you have already entered, press Exit softkey.
d) Resistance shunt should be used on 2, 5, 10 and 30 A ranges, if standard multimeter
does not cover this range.
e) Repeat steps a) to d) also for Current MOD#2 and Current MOD#3.
10. Multimeter calibration
Built-in multimeter calibration consists of the calibration of 10 V voltage range, 20 mA current range and 10 kHz frequency range.
External calibrator of higher precision class should be used to calibrate the built-in multimeter (Meatest M142). If accurate external calibrator is not available an 81/2 digit multimeter is recommended for calibration.
a) 10 V DC voltage range calibration
Select Meter from the calibration menu. Press Select softkey to select the first
calibration point, 0 V.
Make short on calibrators input terminals METER INPUT V-COM. Main value presents the calibration point. Leave value 0.000 and press Write
softkey.
Select calibration point 10 V.
M133C Electric Power/Energy Calibrator MEATEST
Connect external voltage 10 V to the calibrator input terminals METER INPUT V-
COM. Exact value can be measured by paralel standard voltmeter.
Adjust main value on the display. It should be equal to the value measured by
standard multimeter. Use , , , cursor buttons, rotary knob or numeric keyboard to adjust the main value. Confirm new value by pressing Write softkey.
b) 20 mA DC current range calibration
Select Meter from the calibration menu. Press Select softkey to select the first
calibration point, 0 mA.
Make short on calibrators input terminals METER INPUT mA-COM. Main value presents the calibration point. Leave value 0.000 and press Write
softkey.
Select calibration point 20 mA. Connect external current source 20 mA to the calibrator input terminals METER
INPUT mA-COM. Exact value can be measured by serial standard ampmeter.
Adjust main value on the display. It should be equal to the value measured by
standard multimeter. Use , , , cursor buttons, rotary knob or numeric keyboard to adjust the main value. Confirm new value by pressing Write softkey.
c) 10 kHz frequency range calibration
Select calibration point 10 kHz. Connect external AC voltage 10 kHz (voltage between 5 to 10 V) to the calibrator
input terminals METER INPUT V-COM. Exact frequency value can be measured by paralel standard counter.
Adjust main value on the display. It should be equal to the value measured by
standard counter. Use , , , cursor buttons, rotary knob or numeric keyboard to adjust the main value. Confirm new value by pressing Write softkey.
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 60
8 Error messages
If an error occurs during the calibrator’s operation or control, error message is displayed on the display. Errors can be caused by:
Incorrect control using the front panel, i.e. attempts to force a prohibited mode, e.g. setting an
out-of-range value, overloading of output terminals etc.,
Incorrect control using GPIB or RS-232 interface. Calibrator’s fault.
Sample error message which appears when too large value is attempted to set up is shown bellow. All error messages are displayed in the frame overlapped main screen.
The following table lists all error messages, their meaning and simple troubleshooting.
ID
Error message
Description
-430
Deadlocked.
Remote interface error. A command was received which generates too much data to fit in the output buffer and the output buffer is full. Command execution continues but all data are lost.
-420
Unterminated.
Remote interface error. The calibrator was addressed to talk but a command has not been received which sends data to the output buffer.
-410
Interrupted.
Remote interface error. A command was received which sends data to the output buffer, but the output buffer contained data from a previous command. The output buffer is cleared when power has been off, or after reset command has been executed.
-363
Input buffer overrun
Remote interface error.
-220
Invalid parameter
Remote interface error. An invalid character string was received. Check to see if you have enclosed the character string in single or double quotes and that the string
M133C Electric Power/Energy Calibrator MEATEST
contains valid ASCII character.
-140
Character data
Remote interface error. Received command doesn’t contain valid character parameter.
-120
Numeric data
Remote interface error. Received command doesn’t
contain valid numeric parameter.
-110
Command header
Remote interface error. Received command isn’t valid.
501
Eeprom write.
Eeprom write failed.
502
Eeprom read.
Eeprom read failed.
503
Eeprom error.
Eeprom data lost. Check calibration data.
701,704
Output overload
Output signal over specified limits. Decrease signal level or decrease the load.
703
High temperature
Calibrator power stage overheated. Disconnect external load. Ambient temperature is too high or forced ventilation holes are blocked.
705
Input overload
Input signal over specified limits. Decrease signal level.
706
Current output overload
Current output terminals overloaded. Decrease signal level or decrease the load.
707
Voltage output overload
Voltage output terminals overloaded. Decrease signal level or decrease the load.
722
Unexpected crossing.
Error in internal communication.
721
Unknown function.
Error in internal communication.
730, 731
Calibrator not ready
Error in internal communication.
732
Internal cpu RESET
Calibrator will be restarted.
743
Interface receive
Error in internal communication.
745
Internal CPU timeout
Error in internal communication.
746, 747, 748
Slave error
Error in internal communication.
750
Harmonic U#1 over range
Amplitude of harmonic signal on voltage channel #1 is too high. Reduce amplitudes of individual harmonic components.
751
Harmonic U#2 over range
Amplitude of harmonic signal on voltage channel #2 is too high. Reduce amplitudes of individual harmonic components.
752
Harmonic U#3 over range
Amplitude of harmonic signal on voltage channel #3 is too high. Reduce amplitudes of individual harmonic components.
753
Harmonic I#1 over range
Amplitude of harmonic signal on current channel #1 is too high. Reduce amplitudes of individual harmonic components.
754
Harmonic I#2 over range
Amplitude of harmonic signal on current channel #2 is too high. Reduce amplitudes of individual harmonic components.
755
Harmonic I#3 over range
Amplitude of harmonic signal on current channel #3 is too high. Reduce amplitudes of individual harmonic components.
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 62
9 Maintenance
The multifunction calibrator is electronic instrument with microprocessor control. All blocks which are heavily loaded during the operation are cooled by a fan.
Rules for correct operation
Especially the following rules should be adhered to guarantee correct operation of the calibrator:
The calibrator can only be switched on and off by pressing the mains switch located at the rear
panel.
Do not connect the calibrator to other voltage than set by the voltage selector. Do not block the vent openings located at the rear panel and bottom panel. The calibrator must not be operated in dusty environment. It was designed to be used in a
laboratory.
No liquid or small objects can be permitted to enter the calibrator through the vent openings.. Do not switch the calibrator outside its operating temperature range. Connect the instruments to be calibrated to proper output terminals. There is no way of
protecting the calibrator from the damage caused by some improper connections.
Do not damage the output terminals by plugging in “bananas” thicker than the terminals were
designed for.
Whenever possible, use the setup menu to ground Lo output terminals (GndU On, GndI On). Do not overload the power stages by leaving the calibrator switched on with the load
connected for a long time, especially on 30 A current range and 140 V and 280 V voltage ranges.
If the instruments to be calibrated are not connected to calibrator’s output terminals using
original cables, ensure that cables suitable for the calibration voltage and current are used. Maximum output voltage can reach 600 V AC and the maximum output current can reach 30 A AC.
Regular maintenance
The calibrator does not require any special maintenance of electrical or mechanical parts. If is gets dirty, the case and the display can be cleaned by a wool rag moistened with alcohol.
The calibrator should be calibrated in the recommended 12-month intervals. A calibration center should perform the calibration.
M133C Electric Power/Energy Calibrator MEATEST
What to do in case of failure
If an obvious failure occurs during the operation (e.g. the display is not lit, the fan is not turning), the calibrator must be switched off immediately. First, check the fuse located in the power cord receptacle. Procedure is following:
Remove the end of power cord from the mains connector at the rear panel. Insert the blade of a flat screwdriver into the opening cut in the mains voltage selector and pry
out the fuse holder.
Remove the fuse. Replace it with new fuse of the same rating if the fuse was broken. Replace the fuse holder, reconnect the power cord and switch on the calibrator. If the problem
persists, contact the manufacturer.
If an obvious fault is evidenced, e.g. a measurement range or an operating mode is not functional, the user cannot correct the fault. Contact the manufacturer.
Hidden faults can cause different symptoms and be caused by different causes. Usually, they cause instability of some parameter. Hidden defects can be caused by unacceptable distortion, degraded insulation etc. In this case contact the manufacturer.
Sometimes it seems that the calibrator has hidden defect, when the rules for correct operation are not adhered to. In this case, the fault is caused by the operator. Most frequent cases of false “hidden defects”:
mains voltage out of tolerance limits or unstable wrong grounding of the measurement circuit (bad connection of the ground terminal of the
mains outlet, or several ground connection when grounding loops are formed)
proximity to sources of intensive influence, whose products are spread through the mains or
propagated by the electromagnetic field
strong electrostatic or electromagnetic field which can cause major instability during calibration
using higher impedance.
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 64
10 Verification test
Procedure recommended for verifying parameters of the calibrator is described in this chapter. During tests it is not necessary access to the interior of the instrument.
Required equipment Following instruments are required for performance verification test: 81/2 digit multimeter type HP3458A or Fluke 8508A, or other type with accuracy 0.01 % on
AC voltage
Multifuntion calibrator MEATEST M142 or M140 with DCranges 10V, 20mA and AC range
10V / 15kHz.
resistance shunt 10 m, 100 m Burster 1280, or other type with accuracy 0.01% with known
AC/DC difference to 1 kHz
counter HP 53181A, HO 53130, BM 642 or other with accuracy 0,001 % powermeter 0.02- 0.05 % like Zimmer LMG95REF, Applied Precision RS2110S
HP8903A Distortion analyzer and scope with bandwidth min. 20 MHz. 100 khz bandwidth is recommended for THD measuring of AC signals.
Configuration of the calibrator Calibrator should be tested directly from the front panel terminals To suppress influence of noise
or interference with power line frequency in measuring circuit it is recommended following setting of the calibrator (in Main menu):
1. Current coil Coil Off
2. Voltage sources GND GndU On
3. Current sources GND GndI On
Note: If it is grounded neither calibrator nor standard meter, higher level can occur on the
output terminals.
In general, when calibrator is connected to the standard meter, ground loops can be
arise through power line connection. Ground loops can result apparently worse noise, short term stability or non-harmonic distortion of output signal. If necessary use toroidal chokes to suppress this products.
Use sin waveform of the output signal on all AC tests.
M133C Electric Power/Energy Calibrator MEATEST
Performance verification may be performed after warm-up period i.e. 1 hour after switching on. Calibrator have to be in temperature stabilize condition at minimum 8 hours before performance verification test is started.
Basic steps of the performance verification test Verification procedure consists of following steps:
30 V DC voltage test with linearity check DC voltage internal ranges 10 V, 70 V, 140 V, 280 V test 30 V AC voltage test with linearity check AC voltage internal ranges 10 V, 70 V, 140 V, 280 V, 600 V test 1 A DC current test with linearity check DC current internal ranges 300 mA, 2 A, 5 A, 10 A, 30 A test AC current internal ranges 300 mA, 2 A, 5 A, 10 A, 30 A test AC/DC power 480W, 2400W (AC power with PF 0, +0.5, -0.5) test Frequency nominal value 1 kHz test Multimeter internal ranges 10 VDC, 20 mADC, frequency 10 kHz test Distortion checking of AC voltage, range 30 V.
Procedure Following part describes procedure of performance verification test. Recommended measuring
points are the same as the points in table of limits (see tables bellow).
1. Connect the calibrator to the mains and let them switched on for at least one hour in a
laboratory at 231 oC.
2. Connect voltage input of the standard multimeter to the voltage output terminals of the
calibrator. Set appropriate parameters on the standard multimeter to achieve its best accuracy.
3. Perform 30 VDC linearity, DC voltage, 30 VAC linearity, AC voltage tests according to the
tables I, II, III, IV. Deviations should not exceed specified limits.
4. Connect current input of standard multimeter to the current output terminals of the calibrator.
Set appropriate parameters on the standard multimeter to meet its best accuracy.
5. Perform 1 ADC linearity, DC current, AC current tests according to the tables V, VI, VII.
Deviations should not exceed specified limits.
6. Connect standard powermeter to the appropriate output and current terminals of the calibrator.
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 66
7. Perform AC/DC power test according to the table VIII. Deviations should not exceed specified
limits.
8. Connect output voltage terminals of the calibrator to the counter. Set output voltage 5 VAC,
frequency 1 kHz.
9. Perform frequency test according to the table IX. Deviation should not exceed specified limit.
10. Connect input voltage terminals of calibrator’s internal multimeter to the output terminals of
the standard calibrator. Set appropriate value of frequency and DC voltage according to the table X.
11. Perform multimeter test in points FREQUENCY 10000Hz and 10 VDC. Deviations should
not exceed limits in the tables bellow. Note : Because accuracy of calibrator output is not accurate enough in all points, use
external standard counter (frequency) or multimeter (DC voltage) connected in parallel to the output terminals to obtain exact value on the output.
12. Connect input current terminals of calibrator’s internal multimeter to the output terminals of
the standard calibrator. Set appropriate value of DC current according to the table X.
13. Perform multimeter test on range 20 mA DC according to the table X. Deviation should not
exceed specified limit.
14. Connect distortion meter to the voltage output terminals of the calibrator. Set output voltage 10
VAC, frequency 1000 Hz and sin waveform.
15. Check harmonic distortion of output signal. It should not exceed 0.05%.
If calibrator is out of limits is in some points of this test, appropriate function and range should be recalibrated. It is not necessary to recalibrate all functions, but only this one, which does not meet specification. See chapter Calibration mode, where recalibration procedure is described.
M133C Electric Power/Energy Calibrator MEATEST
Tables of limits
30 V DC Basic range with linearity test
Function
Range
Value (V)
Frequency (Hz)
Deviation allowed (mV)
V-DC
30.0 V
12.0
4.8
V-DC
30.0 V
14.0
5.1
V-DC
30.0 V
16.0
5.4
V-DC
30.0 V
18.0
5.7
V-DC
30.0 V
20.0
6.0
V-DC
30.0 V
22.0
6.3
V-DC
30.0 V
24.0
6.6
V-DC
30.0 V
26.0
6.9
V-DC
30.0 V
28.0
7.2
V-DC
30.0 V
30.0
7.5
V-DC
30.0 V
-12.0
4.8
V-DC
30.0 V
-14.0
5.1
V-DC
30.0 V
-16.0
5.4
V-DC
30.0 V
-18.0
5.7
V-DC
30.0 V
-20.0
6.0
V-DC
30.0 V
-22.0
6.3
V-DC
30.0 V
-24.0
6.6
V-DC
30.0 V
-26.0
6.9
V-DC
30.0 V
-28.0
7.2
V-DC
30.0 V
-30.0
7.5
Table I
DC voltage test
Function
Range
Value (V)
Frequency (Hz)
Deviation allowed (mV)
V-DC
10.0 V
10.0
2.5
V-DC
10.0 V
-10.0
2.5
V-DC
70.0 V
70.0
17.5
V-DC
70.0 V
-70.0
17.5
V-DC
140.0 V
140.0
35.0
V-DC
140.0 V
-140.0
35.0
V-DC
280.0 V
280.0
70.0
V-DC
280.0 V
-280.0
70.0
Table II
30 V AC Basic range with linearity test
Function
Range
Value (V)
Frequency (Hz)
Deviation allowed (mV)
V-AC
30.0 V
12.0
55
4.8
V-AC
30.0 V
14.0
55
5.1
V-AC
30.0 V
16.0
55
5.4
V-AC
30.0 V
18.0
55
5.7
V-AC
30.0 V
20.0
55
6.0
V-AC
30.0 V
22.0
55
6.3
V-AC
30.0 V
24.0
55
6.6
V-AC
30.0 V
26.0
55
6.9
V-AC
30.0 V
28.0
55
7.2
V-AC
30.0 V
30.0
55
7.5
Table III
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 68
AC voltage test
Function
Range
Value (V)
Frequency (Hz)
Deviation allowed (mV)
V-AC
10 V
10.0
55
2.5
V-AC
30 V
30.0
20
9.0
V-AC
30 V
30.0
55
7.5
V-AC
30 V
30.0
120
9.0
V-AC
30 V
30.0
1000
9.0
V-AC
70 V
70.0
55
17.5
V-AC
140 V
140.0
55
35.0
V-AC
280 V
280.0
55
70.0
V-AC
600 V
600.0
55
180.0
Table IV
1 A DC Basic range with linearity test
Function
Range
Value (A)
Frequency (Hz)
Deviation allowed (µA)
A-DC
1 A
0.4
200
A-DC
1 A
0.5
225
A-DC
1 A
0.6
250
A-DC
1 A
0.7
275
A-DC
1 A
0.8
300
A-DC
1 A
0.9
325
A-DC
1 A
1.0
350
A-DC
1 A
-0.4
200
A-DC
1 A
-0.5
225
A-DC
1 A
-0.6
250
A-DC
1 A
-0.7
275
A-DC
1 A
-0.8
300
A-DC
1 A
-0.9
325
A-DC
1 A
-1.0
350
Table V
DC current test
Function
Range
Value (A)
Frequency (Hz)
Deviation allowed (µA)
A-DC
300.0 mA
0.3
105 μ
A-DC
300.0 mA
-0.3
105 μ
A-DC
2.0 A
2.0
700 μ
A-DC
2.0 A
-2.0
700 μ
A-DC
5.0 A
5.0
1.75 m
A-DC
5.0 A
-5.0
1.75 m
A-DC
10.0 A
10.0
4.5 m
A-DC
10.0 A
-10.0
4.5 m
A-DC
30.0 A
30.0
15.0 m
A-DC
30.0 A
-30.0
15.0 m
Table VI
M133C Electric Power/Energy Calibrator MEATEST
AC current test
Function
Range
Value (A)
Frequency (Hz)
Deviation allowed (µA)
A-AC
300.0 mA
0.3
55
105 μ
A-AC
1.0 A
1.0
55
350 μ
A-AC
1.0 A
1.0
800
500 μ
A-AC
2.0 A
2.0
55
800 μ
A-AC
5.0 A
5.0
55
1.75 m
A-AC
10.0 A
10.0
55
4.5 m
A-AC
30.0 A
30.0
55
15.0 m
Table VII
AC/DC power test
Function
Range
Value (W)
Frequency (Hz)
Deviation allowed (W)
P 1
480W
480
55
0.216
P 0,5LA
480W
240
55
0.180
P 0,5LE
480W
240
55
0.180
P 1
2400W
2400
55
1.27
P 0,5LA
2400W
1200
55
0.972
P 0,5LE
2400W
1200
55
0.972
Table VIII Voltage = 240V
Frequency test
Function
Range
Value (Hz)
Frequency (Hz)
Deviation allowed (Hz)
FREQ
10 kHz
10000.0
0.5
Table IX
Multimeter test
Function
Range
Value
Frequency (Hz)
Deviation allowed
FREQ
10 kHz
10000.0 Hz
0.5 Hz
V-DC
10 V
10.0 V
1.3 mV
V-DC
10 V
-10.0 V
1.3 mV
A-DC
20 mA
0.02 A
3.3 μA
A-DC
20 mA
-0.02 A
3.3 μA
Table X
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 70
11 System control
The calibrator is equipped with RS232 serial line, IEEE488 bus and the TCP/IP Ethenet connection. System connectors are located at the rear panel. For the remote control to work properly, interface parameters must be set in the system menu. For RS232 line communication speed can be set (1200 to 115200 Bd). For IEEE488 bus, an address from range 0 to 30 is allowed. For Ethernet connection can be set IP address. The calibrator can be only controlled by one of interfaces at a time. It is therefore necessary to select one of the interfaces (RS232/Ethernet/IEEE488) using the system menu.
11.1 IEEE488 bus properties
The instrument performs the following functions based on IEEE488 bus commands:
SH1, AH1, T5, L3, RL1, DC1, SR1
The instrument also recognizes the following general commands: DCL Device Clear reset the calibrator
SDC Selected Device Clear reset the calibrator EOI End or Identify Message Terminator close the message GTL Go To Local close remote control mode LLO Local Lock Out local control locked SPD Serial Poll Disable close serial message status SPE Serial Poll Enable release serial message
status
11.2 RS232 serial line setting
To transfer the data using RS232 bus, 8N1 data format is used, i.e. each data word includes 8 bits, no parity and one stop bit. The communication speed can be set using the system menu. Available values: 1200, 2400, 4800, 9600, 19200, 38400, 76800 and 115200 Bd.
RS-232 connector layout
Pin
Name
Direction
Description
2
TXD
Output
Transmitter
3
RXD
Input
Receiver
5
GND - Ground
9-pin connector D-SUB FEMALE
M133C Electric Power/Energy Calibrator MEATEST
Cable between the calibrator and PC (configuration 1:1)
PC
D-Sub 1
D-Sub 2
Calibrator
Receiver
2
2
Transmitter
Transmitter
3
3
Receiver
Ground
5
5
Ground
Note: It is not possible to send or receive data over RS232 or Ethernet when the calibrator is not in remote mode. You should use the command SYST:REM or SYST:RWL.
11.3 Command syntax
The commands described in this chapter can be issued through all buses (RS232/Ethernet/IEEE488).
All commands listed in this chapter are explained in two columns: KEYWORD and PARAMETERS.
KEYWORD column includes the name of the command. Each command includes one or more keywords. If a keyword is in brackets ( [ ] ), it is not mandatory. Non-mandatory commands are used only to achieve compatibility with language standard SCPI.
Capitals designate the abbreviated form of the commands; extended form is written in lowercase. Command parameters are in brackets (<>); each parameter is separated using a comma.
Parameters in brackets ( [ ] ) are not mandatory. Line ( | ) means “or” and is used to separate several alternative parameters.
Semicolon ‘;’ is used to separate more commands written on one line. E.g. :SAF:LR 100.5;:OUTP ON
Note (only for RS232 and Ethernet): Each command must end in <cr> or <lf>. Both codes <crlf> can be used at the same time. The calibrator performs all commands written on one line of the program after it receives <cr>, <lf> or <crlf> code. Without this code, the program line is ignored.
Description of abbreviations
<DNPD> = Decimal Numeric Program Data, this format is used to express decimal number with or without the exponent.
<CPD> = Character Program Data. Usually, it represents a group of alternative character
parameters. E.g. {ON | OFF | 0 | 1}.
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 72
<SPD> = String Program Data. String value consisting from more parts. It is used for date/time
setting.
? = A flag indicating a request for the value of the parameter specified by the command.
No other parameter than the question mark can be used.
(?) = A flag indicating a request for the parameter specified by the command. This
command permits a value to be set as well as requested. <cr> = carriage return. ASCII code 13. This code executes the program line. <lf> = line feed. ASCII code 10. This code executes the program line.
M133C Electric Power/Energy Calibrator MEATEST
INPut subsystem
This subsystem enables to change parameters of calibrator inputs.
Keyword Parameters
INPut :FILTer (?) <CPD> { OFF | ON | 0 | 1 } :PULLup (?) <CPD> { OFF | 150 | 1K }
INP:FILT (?) <CPD> { OFF | ON | 0 | 1 }
This command connects the internal filter to the IN1.
OFF or 0 – disconnects filter (frequency range 0 to 1 MHz) ON or 1 – connects filter (frequency range 0 to 400 Hz)
State of the function remains valid even after the calibrator is switched off and back on. If query is sent, calibrator returns state of the fucntion { 0 | 1 }.
INP:PULL (?) <CPD> { OFF | 150 | 1K }
This command connects the internal pull-up resistor (150 Ω or 1 kΩ) to the IN1.
OFF – disconnects pull-up resistor 150 – connects pull-up resistor (150 Ω) 1K – connects pull-up resistor (1k Ω)
State of the function remains valid even after the calibrator is switched off and back on. If query is sent, calibrator returns connects resistor { OFF | 150 | 1K }.
OUTPut subsystem
This subsystem enables to activate or deactivate the calibrator output.
Keyword Parameters
OUTPut [:STATe](?) <CPD> { ON | OFF } :LOWVoltage(?) <CPD> { FLOat | GROund } :LOWCurrent (?) <CPD> { FLOat | GROund } [:PHASe]:UNIT(?) <CPD> { DEG | COS } :CURCoil(?) <CPD> { OFF | X25 | X50} :SYNChronization(?) <CPD> { INT | LINE | IN1 | IN2 | IN3 } :LOCKed? <CPD> { 0 | 1 } :ENERgy :UNIT(?) <CPD> { WS | WH }
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 74
:MVOLtage(?) <CPD> { OFF | ON | 0 | 1 } :REFerence :UNIT(?) <CPD> { W | VA | VAR } :PULLup(?) <CPD> { 0 | 1 } :CONStant(?) <DNPD> :MHARmonics :UNIT(?) <CPD> { PRMS | PFUN } :CONFiguration(?) <CPD> { 1 | 12 | 123 } :L280(?) <CPD> { OFF | ON | 0 | 1 }
OUTP[:STAT](?) <CPD> { ON | OFF }
This command activates or deactivates output terminals.
ON - activates the output OFF - deactivates the output
If query is sent, calibrator returns ON if the output is active or OFF if it is inactive Example: OUTP ON - activates the output
OUTP? - the calibrator returns ON or OFF
OUTP:LOWV(?) <CPD> { FLO | GRO }
This command connects or disconnects the Lo terminals of all voltage outputs to/from GND terminal.
FLOat ungrounds voltage outputs GROund grounds voltage outputs
If query is sent, calibrator returns GRO when the output is grounded or FLO when floating. Example:
OUTP:LOWV GRO OUTP:LOWV? Calibrator returns FLO or GRO
OUTP:LOWC(?) <CPD> { FLO | GRO }
This command connects or disconnects the Lo terminals of all current outputs to/from GND terminal.
FLOat ungrounds current outputs GROund grounds current outputs
If query is sent, calibrator returns GRO when the output is grounded or FLO when floating. Example:
OUTP:LOWC GRO OUTP:LOWC? Calibrator returns FLO or GRO
OUTP[:PHAS]:UNIT(?) <CPD> { DEG | COS }
M133C Electric Power/Energy Calibrator MEATEST
This command sets the method used to specify the phase shift between the output voltage and current. DEG activates a mode where all entries are done using the angle, „“ in the range of 0.0 –
360.0 COS activates a mode where all entries are done using the power factor in the range of 1.000 to –1.000, LAG or LEAD ( LAG = 0-180, LEAD = 180-360 )
The unit of measurement remains valid even after the calibrator is switched off and back on. If query is sent, calibrator returns the set unit of measurement { DEG | COS }. Example: OUTP:UNIT DEG <cr> - sets the angle as the unit of phase shift
OUTP:UNIT? <cr> - the calibrator returns DEG
OUTP:CURC(?) <CPD> { OFF | X25 | X50 }
This command activates or deactivates the use of 25 or 50-turns current coil.
OFF - deactivates the current coil X25 - activates the 25 turns current coil X50 - activates the 50 turns current coil
If query is sent, calibrator returns X50 if the 50-turn coil is activated X25 if the 25-turn coil is activated or OFF if the current coil is deactivated.
Example: OUTP:CURC X50 activates 50-tuen current coil OUTP:CURC? Calibrator returns OFF, X50 or X25
OUTP:SYNC(?) <CPD> { INT | LINE | IN1 | IN2 | IN3 }
This command sets mode of output frequency synchronization.
INT – internal synchronization LINE – synchronization to power supply frequency IN1 – synchronization to external signal (input IN1) IN2 – synchronization to external signal (input IN2) IN3 – synchronization to external signal (input IN3)
If query is sent, calibrator returns INT, LINE, IN1, IN2 or IN3 according to selected synchronization mode.
Example: OUTP:SYNC INT selects frequency synchronization to internal reference OUTP:SYNC? Calibrator returns INT, LINE, IN1, IN2 or IN3
OUTP:SYNC:LOCK? <CPD> { 0 | 1 }
This query returns the actual state of frequency synchronization.
0 – output frequnecy is not locked (it is not possible to activate the output terminals) 1 – output frequency is locked to reference signal
Example: OUTP:SYNC:LOCK? calibrator returns 1 or 0
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 76
OUTP:ENER:UNIT(?) <CPD> { WS | WH }
This command sets unit used for energy.
WS - energy is displayed in Ws, VAs, VArs WH - energy is displayed in Wh, VAh, VArh
Energy unit remains valid even after the calibrator is switched off and back on. If query is sent, calibrator returns set energy unit { WS | WH }. Example: OUTP:ENER:UNIT WH <cr> - sets energy unit Wh
OUTP: ENER:UNIT? <cr> - the calibrator returns WH
OUTP:ENER:MVOL(?) <CPD> { OFF | ON | 0 | 1 }
This command activates or deactivates Maintain Voltage Signal function.
OFF or 0 – function is not active ON or 1 – fucntion is active
State of the function remains valid even after the calibrator is switched off and back on. If query is sent, calibrator returns state of the fucntion { 0 | 1 }. Example: OUTP:ENER:MVOL ON <cr> - activates Maintain Voltage Signal function
OUTP: ENER:MVOL? <cr> - the calibrator returns 1
OUTP:REF:PULL (?) <CPD> { 0 | 1 }
This command connects the internal pull-up resistor (150 Ω) to the Energy pulse output.
0 disconnects pull-up resistor 1 connects pull-up resistor
If query is sent, calibrator returns 0 or 1. Example: OUTP:REF:PULL 1 connects pull-up resistor
OUTP:REF:PULL? calibrator returns 0 or 1
OUTP:REF:CONS(?) <DNPD>
This command sets number of pulses per selected unit (pulses generated on Energy pulse output on the rear panel BNC).
<DNPD> Parameter represents the value number of pulses per selected unit.
If query is sent, calibrator returns set value using standard exponential format. Example: 400.0 i/kWh is returned as 4.000000e+002.
Example: Number of pulses 1000 i/kVArh: OUT:REF:CONS 1000.0 OUT:REF:UNIT VAR
M133C Electric Power/Energy Calibrator MEATEST
OUTP:REF:UNIT(?) <CPD> { W | VA | VAR }
This command defines unit of constant for pulses generated on Energy pulse output. <CPD>
If query is sent, calibrator returns selected unit. Example: OUTP:REF:UNIT VAR <cr> - sets energy pulse output unit VAR
OUTP: REF:UNIT? <cr> - the calibrator returns VAR
OUTP:MHAR:UNIT(?) <CPD> { PRMS | PFUN }
This command defines how are evaluated higher harmonic components. <CPD>
PRMS – % of RMS value  PFUN – % of Fundamental harmonic value
If query is sent, calibrator returns selected unit. Example: OUTP:MHAR:UNIT PRMS <cr> - harmonics are expressed as “% of RMS”
OUTP: MHAR:UNIT? <cr> - the calibrator returns PRMS
OUTP:CONF (?) <CPD> { 1 | 12 | 123 }
This command defines how many outputs is used in basic power and energy AC modes (Pac and Eac).
1 – active are channels U1, I1 only 12 – active are channels U1, U2, I1, I2 123 – active are all channels
If query is sent, calibrator returns active channels. Example: OUTP:CONF 123 <cr> - activates all channels
OUTP: CONF? <cr> - the calibrator returns 123
OUTP:L280(?) <CPD> { OFF | ON | 0 | 1 }
This command activates or deactivates „280V range lock” function.
OFF or 0 – 280V range is not locked ON or 1 – 280V range is locked
State of the function is switched off after power on. If query is sent, calibrator returns state of the fucntion { 0 | 1 }. Example: OUTP:L280 ON <cr> - lockes 280V range
OUTP:L280? <cr> - the calibrator returns 1
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 78
MEASure a CONFigure subsystem
This subsystem allows controling the internal multimeter. Keyword Parameters
MEASure? CONFigure(?) <CPD> { VOLTage | CURRent | FREQuency }
MEAS?
This query returns value measured by the multimeter. calibrator returns the set value of power using standard exponential format. Example: 5.456 V is
returned as 5.456000e+000. Example: MEAS? Calibrator returns measured value
CONF(?) <CPD> { VOLT | CURR | FREQ }
This command sets the function of internal multimeter. If query is sent, calibrator returns VOLT, CURR or FREQ according to selected multimeter
function. Example: CONF VOLT sets the multimeter to function voltage DC
SOURce subsystem
This subsystem allows controlling the individual functions of the calibrator. Keyword Parameters
[SOURce] :MODE? { PAC | PDC | PACI | PDCI | PACE | PHAR | PIH | PDIP | PDCE | EAC | EDC |
EACI | EDCI | VAC | VDC | CAC | CDC | CACI | CDCI }
:PAC :POWer(?) <DNPD> [:POWer] :UNIT(?) <CPD> { W | VA | VAR } :VOLTage(?) <DNPD> :CURRent(?) <DNPD> [:CURRent] :PHASe(?) <DNPD> :POLarity(?) <CPD> { LEAD | LAG } :FREQuency(?) <DNPD >
:PDC :POWer(?) <DNPD>
M133C Electric Power/Energy Calibrator MEATEST
:VOLTage(?) <DNPD> :CURRent(?) <DNPD>
:PACI :POWer(?) <DNPD> [:POWer] :UNIT(?) <CPD> { W | VA | VAR } :VOLTage(?) <DNPD> :CURRent(?) <DNPD> [:CURRent] :PHASe(?) <DNPD> :POLarity(?) <CPD> { LEAD | LAG } :FREQuency(?) <DNPD >
:PDCI :POWer(?) <DNPD> :VOLTage(?) <DNPD> :CURRent(?) <DNPD>
:PACE :POWer? [:POWer] :UNIT(?) <CPD> { W | VA | VAR } :VOLTage<n>(?) <DNPD> :PHASe(?) <DNPD> :ENABle (?) <CPD> { ON | OFF } :CURRent<n>(?) <DNPD> :PHASe(?) <DNPD> :ENABle (?) <CPD> { ON | OFF } :FREQuency(?) <DNPD >
:PDCE :POWer? :VOLTage<n>(?) <DNPD> :ENABle (?) <CPD> { ON | OFF } :CURRent<n>(?) <DNPD> :ENABle (?) <CPD> { ON | OFF }
:PHAR :POWer? <DNPD>, <DNPD> :VOLTage<n>(?) <DNPD> :PHASe(?) <DNPD> :ENABle (?) <CPD> { ON | OFF } :HARMonic<y>(?) <DNPD> :PHASe(?) <DNPD> :MODulation(?) <DNPD> :SHAPe(?) <CPD> { SIN | RECT | OFF } :DUTY(?) <DNPD> :CURRent<n>(?) <DNPD> :PHASe(?) <DNPD> :ENABle(?) <CPD> { ON | OFF } :HARMonic<y>(?) <DNPD>
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 80
:PHASe(?) <DNPD> :MODulation(?) <DNPD> :SHAPe(?) <CPD> { SIN | RECT | OFF } :DUTY(?) <DNPD> :FREQuency(?) <DNPD > :MODulation(?) <DNPD >
:PIH :VOLTage<n>(?) <DNPD> :PHASe(?) <DNPD> :ENABle(?) <CPD> { ON | OFF } :IHARmonic(?) <DNPD> :CURRent<n>(?) <DNPD> :PHASe(?) <DNPD> :ENABle(?) <CPD> { ON | OFF } :IHARmonic(?) <DNPD> :FREQuency(?) <DNPD > :IHARmonic(?) <DNPD >
:PDIP :VOLTage<n>(?) <DNPD> :PHASe(?) <DNPD> :ENABle(?) <CPD> { ON | OFF } :CHANge (?) <DNPD> :CURRent<n>(?) <DNPD> :PHASe(?) <DNPD> :ENABle(?) <CPD> { ON | OFF } :CHANge (?) <DNPD> :FREQuency(?) <DNPD > :TIME(?) <DNPD >,<DNPD >,<DNPD >,<DNPD >,<DNPD > :TRIGger :INPut(?) <CPD> { OFF | IN3 } :SYNChronization(?) <CPD> { OFF | ON } :REPetition (?) <CPD> { ONE | REPeat } :STARt
:EAC :ENERgy? :DEViation? :POWer? [:POWer] :UNIT(?) <CPD> { W | VA | VAR } :VOLTage(?) <DNPD> :CURRent(?) <DNPD> [:CURRent] :PHASe(?) <DNPD> :POLarity(?) <CPD> { LEAD | LAG } :FREQuency(?) <DNPD > :CONStant(?) <DNPD> :TIME(?) <DNPD>
M133C Electric Power/Energy Calibrator MEATEST
:CONTrol(?) <CPD> { PACK | CNT1 | CNT2 | TIM1 | TIM2 | FR1 | FR2 | FR3 } :TEST :TIME(?) <DNPD> :COUNt(?) <DNPD> :FREQ? <DNPD> :WUP :TIME(?) <DNPD> :COUNt(?) <DNPD>
:EDC :ENERgy? :DEViation? :POWer? :VOLTage(?) <DNPD> :CURRent(?) <DNPD> :CONStant(?) <DNPD> :TIME(?) <DNPD> :CONTrol(?) <CPD> { PACK | CNT1 | CNT2 | TIM1 | TIM2 | FR1 | FR2 | FR3 } :TEST :TIME(?) <DNPD> :COUNt(?) <DNPD> :FREQ? <DNPD> :WUP :TIME(?) <DNPD> :COUNt(?) <DNPD>
:EACI :ENERgy? :DEViation? :POWer? :VOLTage(?) <DNPD> :CURRent(?) <DNPD> [:CURRent] :PHASe(?) <DNPD> :FREQuency(?) <DNPD > :CONStant(?) <DNPD> :TIME(?) <DNPD> :CONTrol(?) <CPD> { PACK | CNT1 | CNT2 | TIM1 | TIM2 | FR1 | FR2 | FR3 } :TEST :TIME(?) <DNPD> :COUNt(?) <DNPD> :FREQ? <DNPD> :WUP :TIME(?) <DNPD> :COUNt(?) <DNPD>
:EDCI :ENERgy?
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 82
:DEViation? :POWer? :VOLTage(?) <DNPD> :CURRent(?) <DNPD> :CONStant(?) <DNPD> :TIME(?) <DNPD> :CONTrol(?) <CPD> { PACK | CNT1 | CNT2 | TIM1 | TIM2 | FR1 | FR2 | FR3 } :TEST :TIME(?) <DNPD> :COUNt(?) <DNPD> :FREQ? <DNPD> :WUP :TIME(?) <DNPD> :COUNt(?) <DNPD>
:VAC :VOLTage(?) <DNPD> :FREQuency(?) <DNPD >
:VDC :VOLTage(?) <DNPD>
:CAC :CURRent(?) <DNPD> :FREQuency(?) <DNPD >
:CDC :CURRent(?) <DNPD>
:CACI :CURRent(?) <DNPD> :FREQuency(?) <DNPD >
:CDCI :CURRent(?) <DNPD>
M133C Electric Power/Energy Calibrator MEATEST
[SOUR]:MODE?
This command returns selected mode of the calibrator { PAC | PDC | PACI | PDCI | PACE | PDCE | PHAR | PIH | PDIP | EAC | EDC | EACI | EDCI | VAC | VDC | CAC | CDC | CACI | CDCI }.
PAC – Power AC mode PDC – Power DC mode PACI – Power AC high current mode PDCI – Power DC high current mode PACE – Power AC extended mode PDCE – Power DC extended mode PHAR – Power harmonic mode PIH – Power interharmonic mode PDIP – Power Dip/Swell mode EAC – Energy AC mode EDC – Energy DC mode EACI – Energy AC high current mode EDCI – Energy DC high current mode VAC – Voltage AC mode VDC – Voltage DC mode CAC – Current AC mode CDC – Current DC mode CACI – Current AC high current mode CDCI – Current DC high current mode
Example: MODE?
[SOUR]:PAC:POW(?) <DNPD>
This command sets the power value in Power AC mode. <DNPD>
Parameter represents the value of the power expressed in actual units (W, VA, VAr). If query is sent, calibrator returns the set value of power using standard exponential format.
Example: 100.60 W is returned as 1.006000e+002. Example:
AC power 1000W: PAC:POW 1000
Note: Command switches the calibrator in Power AC mode.
[SOUR]:PAC[:POW]:UNIT(?) <CPD> { W | VA | VAR }
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 84
This command selects units for power in the Power AC mode. If query is sent, calibrator returns actual units in Power AC mode. Example:
Select unit W for Power AC mode: PAC:UNIT W
Note: Command switches the calibrator in Power AC mode.
[SOUR]:PAC:VOLT(?) <DNPD>
This command sets the output voltage value. <DNPD>
Parameter represents the value of the output voltage expressed in V. If query is sent, calibrator returns the value of output voltage using standard exponential format.
Example: 110.12 V is returned as 1.101200e+002. Example:
Voltage 230.5V: PAC:VOLT 230.5
Note: Command switches the calibrator in Power AC mode.
[SOUR]:PAC:CURR(?) <DNPD>
This command sets the output current value in Power AC mode. <DNPD>
Parameter represents the value of the output current expressed in A. If query is sent, calibrator returns the value of output current using standard exponential format.
Example: 11.012 A is returned as 1.101200e+001. Example:
Current 23.05A: PAC:CURR 23.05
Note: Command switches the calibrator in Power AC mode.
[SOUR]:PAC[:CURR]:PHAS(?) <DNPD>
This command sets the phase angle between voltage and current. <DNPD>
Represents the phase angle between the output voltage and current in degrees (PHAS:UNIT=DEG), or specified as a value of the power factor (PHAS:UNIT=COS).
M133C Electric Power/Energy Calibrator MEATEST
If query is sent, calibrator returns the set value of the phase shift between the output voltage and current using standard exponential format. Example: 156.3 is returned as 1.563000e+002.
Example: PAC:PHAS 250.2 - sets 250.2 phase shift between the output voltage and current (PHAS:UNIT=DEG) PAC:PHAS? - the calibrator returns 2.502000e+002
Example: PAC:PHAS 0.554 – sets power factor 0.554 (PHAS:UNIT=COS) PAC:PHAS? - the calibrator returns 5.540000e-001,LAG
Note: Command switches the calibrator in Power AC mode.
[SOUR]:PAC[:CURR]:POLarity(?) <CPD> { LEAD | LAG }
This command sets the polarity of phase angle between voltage and current. Command is valid only for PHAS:UNIT=DEG.
<CPD> Represents the polarity of phase angle between the output voltage and current.
If query is sent, calibrator returns the polarity of the phase angle between the output voltage and current.
Example: PAC:POL LEAD sets the phase angle polarity to Lead
Note: Command switches the calibrator in Power AC mode.
[SOUR]:PAC:FREQ(?) <DNPD>
This command sets the frequency in Power AC mode. <DNPD>
Parameter represents the value of frequency expressed in Hz. If query is sent, calibrator returns the value of frequency using standard exponential format.
Example: 50Hz is returned as 5.000000e+001. Example:
Frequency 60Hz: PAC:FREQ 60
Notes: Command switches the calibrator in Power AC mode.
[SOUR]:PDC:POW(?) <DNPD>
This command sets the power value in Power DC mode.
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 86
<DNPD> Parameter represents the value of the power expressed in W.
If query is sent, calibrator returns the set value of power using standard exponential format. Example: 100.60 W is returned as 1.006000e+002.
Example: DC power 1000W: PDC:POW 1000
Note: Command switches the calibrator in Power DC mode.
[SOUR]:PDC:VOLT(?) <DNPD>
This command sets the output voltage value in Power DC mode. <DNPD>
Parameter represents the value of the output voltage expressed in V. If query is sent, calibrator returns the value of output voltage using standard exponential format.
Example: 110.12 V is returned as 1.101200e+002. Example:
Voltage 230.5V: PDC:VOLT 230.5
Note: Command switches the calibrator in Power DC mode.
[SOUR]:PDC:CURR(?) <DNPD>
This command sets the output current value in Power DC mode. <DNPD>
Parameter represents the value of the output current expressed in A. If query is sent, calibrator returns the value of output current using standard exponential format.
Example: 11.012 A is returned as 1.101200e+001. Example:
Current 23.05A: PDC:CURR 23.05
Note: Command switches the calibrator in Power DC mode.
[SOUR]:PACI:POW(?) <DNPD>
This command sets the power value in Power AC high current mode.
M133C Electric Power/Energy Calibrator MEATEST
<DNPD> Parameter represents the value of the power expressed in actual units (W, VA, VAr).
If query is sent, calibrator returns the set value of power using standard exponential format. Example: 100.60 W is returned as 1.006000e+002.
Example: AC power 1000W: PACI:POW 1000
Note: Command switches the calibrator in Power AC high current mode.
[SOUR]:PACI[:POW]:UNIT(?) <CPD> { W | VA | VAR }
This command selects units for power in the Power AC high current mode. If query is sent, calibrator returns actual units in Power AC high current mode. Example:
Select unit W for Power AC high current mode: PACI:UNIT W
Note: Command switches the calibrator in Power AC high current mode.
[SOUR]:PACI:VOLT(?) <DNPD>
This command sets the output voltage value. <DNPD>
Parameter represents the value of the output voltage expressed in V. If query is sent, calibrator returns the value of output voltage using standard exponential format.
Example: 110.12 V is returned as 1.101200e+002. Example:
Voltage 230.5V: PAC:VOLT 230.5
Note: Command switches the calibrator in Power AC high current mode.
[SOUR]:PACI:CURR(?) <DNPD>
This command sets the output current value in Power AC high current mode. <DNPD>
Parameter represents the value of the output current expressed in A. If query is sent, calibrator returns the value of output current using standard exponential format.
Example: 11.012 A is returned as 1.101200e+001.
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 88
Example: Current 23.05A: PACI:CURR 23.05
Note: Command switches the calibrator in Power AC high current mode.
[SOUR]:PACI[:CURR]:PHAS(?) <DNPD>
This command sets the phase angle between voltage and current. <DNPD>
Represents the phase angle between the output voltage and current in degrees (PHAS:UNIT=DEG), or specified as a value of the power factor (PHAS:UNIT=COS).
If query is sent, calibrator returns the set value of the phase shift between the output voltage and current using standard exponential format. Example: 156.3 is returned as 1.563000e+002.
Example: PACI:PHAS 250.2 - sets 250.2 phase shift between the output voltage and current (PHAS:UNIT=DEG) PACI:PHAS? - the calibrator returns 2.502000e+002
Example: PACI:PHAS 0.554 – sets power factor 0.554 (PHAS:UNIT=COS) PACI:PHAS? - the calibrator returns 5.540000e-001,LAG
Note: Command switches the calibrator in Power AC high current mode.
[SOUR]:PACI[:CURR]:POLarity(?) <CPD> { LEAD | LAG }
This command sets the polarity of phase angle between voltage and current. Command is valid only for PHAS:UNIT=DEG.
<CPD> Represents the polarity of phase angle between the output voltage and current.
If query is sent, calibrator returns the polarity of the phase angle between the output voltage and current.
Example: PACI:POL LEAD sets the phase angle polarity to Lead
Note: Command switches the calibrator in Power AC high current mode.
[SOUR]:PACI:FREQ(?) <DNPD>
This command sets the frequency in Power AC high current mode. <DNPD>
Parameter represents the value of frequency expressed in Hz.
M133C Electric Power/Energy Calibrator MEATEST
If query is sent, calibrator returns the value of frequency using standard exponential format. Example: 50Hz is returned as 5.000000e+001.
Example: Frequency 60Hz: PACI:FREQ 60
Notes: Command switches the calibrator in Power AC high current mode.
[SOUR]:PDCI:POW(?) <DNPD>
This command sets the power value in Power DC high current mode. <DNPD>
Parameter represents the value of the power expressed in W. If query is sent, calibrator returns the set value of power using standard exponential format.
Example: 100.60 W is returned as 1.006000e+002. Example:
DC power 1000W: PDCI:POW 1000
Note: Command switches the calibrator in Power DC high current mode.
[SOUR]:PDCI:VOLT(?) <DNPD>
This command sets the output voltage value in Power DC high current mode. <DNPD>
Parameter represents the value of the output voltage expressed in V. If query is sent, calibrator returns the value of output voltage using standard exponential format.
Example: 110.12 V is returned as 1.101200e+002. Example:
Voltage 230.5V: PDCI:VOLT 230.5
Note: Command switches the calibrator in Power DC high current mode.
[SOUR]:PDCI:CURR(?) <DNPD>
This command sets the output current value in Power DC high current mode. <DNPD>
Parameter represents the value of the output current expressed in A.
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 90
If query is sent, calibrator returns the value of output current using standard exponential format. Example: 11.012 A is returned as 1.101200e+001.
Example: Current 23.05A: PDCI:CURR 23.05
Note: Command switches the calibrator in Power DC high current mode.
[SOUR]:PACE:POW?
This query returns the set value of power using standard exponential format. Example: 100.60 W is returned as 1.006000e+002.
Note: Query switches the calibrator in Power AC extended mode.
[SOUR]:PACE[:POW]:UNIT(?) <CPD> { W | VA | VAR }
This command selects units for power in the Power AC extended mode. If query is sent, calibrator returns actual units in Power AC extended mode. Example:
Select unit W for Power AC extended mode: PACE:UNIT W
Note: Command switches the calibrator in Power AC extended mode.
[SOUR]:PACE:VOLT<n>(?) <DNPD>
This command sets the output voltage value in channel <n> <DNPD>
Parameter represents the value of the output voltage expressed in V. <n>
Parameter represents the output channel (1, 2 or 3). If query is sent, calibrator returns the value of output voltage using standard exponential format.
Example: 110.12 V is returned as 1.101200e+002. Example:
Voltage 230.5V in channel 1: PACE:VOLT1 230.5
Note: Command switches the calibrator in Power AC extended mode.
M133C Electric Power/Energy Calibrator MEATEST
[SOUR]:PACE:VOLT<n>:PHAS(?) <DNPD>
This command sets the phase angle between voltage output in channel <n> and internal reference phase.
<DNPD> Represents the phase angle in degrees.
If query is sent, calibrator returns the set value of the phase angle using standard exponential format. Example: 156.3 is returned as 1.563000e+002.
Example: PACE:VOLT1:PHAS 0 PACE:VOLT2:PHAS 120 PACE:VOLT3:PHAS 240
Note: Command switches the calibrator in Power AC extended mode.
[SOUR]:PACE:VOLT<n>:ENAB(?) <CPD> { ON | OFF }
This command enables (disables) the voltage output in channel <n>. Disabled outputs are OFF also during OUTP ON.
ON - enables the output  OFF - disables the output
If query is sent, calibrator returns ON if the output is enabled or OFF if it is disabled. Example:
PACE:VOLT1:ENAB ON Note:
Command switches the calibrator in Power AC extended mode.
[SOUR]:PACE:CURR<n>(?) <DNPD>
This command sets the output current value in channel <n> <DNPD>
Parameter represents the value of the output current expressed in A. <n>
Parameter represents the output channel (1, 2 or 3). If query is sent, calibrator returns the value of output current using standard exponential format.
Example: 11.012 V is returned as 1.101200e+001. Example:
Current 23.05A in channel 1: PACE:CURR1 23.05
Note: Command switches the calibrator in Power AC extended mode.
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 92
[SOUR]:PACE:CURR<n>:PHAS(?) <DNPD>
This command sets the phase angle between current output in channel <n> and internal reference phase.
<DNPD> Represents the phase angle in degrees.
If query is sent, calibrator returns the set value of the phase angle using standard exponential format. Example: 156.3 is returned as 1.563000e+002.
Example: PACE:CURR1:PHAS 0 PACE: CURR 2:PHAS 120 PACE: CURR 3:PHAS 240
Note: Command switches the calibrator in Power AC extended mode.
[SOUR]:PACE:CURR<n>:ENAB(?) <CPD> { ON | OFF }
This command enables (disables) the current output in channel <n>. Disabled outputs are OFF also during OUTP ON.
ON - enables the output  OFF - disables the output
If query is sent, calibrator returns ON if the output is enabled or OFF if it is disabled. Example:
PACE:CURR1:ENAB ON Note:
Command switches the calibrator in Power AC extended mode.
[SOUR]:PACE:FREQ(?) <DNPD>
This command sets the frequency in Power AC extended mode. <DNPD>
Parameter represents the value of frequency expressed in Hz. If query is sent, calibrator returns the value of frequency using standard exponential format.
Example: 50Hz is returned as 5.000000e+001. Example:
Frequency 60Hz: PACE:FREQ 60
Notes: Command switches the calibrator in Power AC extended mode.
M133C Electric Power/Energy Calibrator MEATEST
[SOUR]:PDCE:POW?
This command sets the power value in Power DC extended mode. This query returns the set value of power using standard exponential format. Example: 100.60 W
is returned as 1.006000e+002. Note:
Query switches the calibrator in Power DC extended mode.
[SOUR]:PDCE:VOLT<n>(?) <DNPD>
This command sets the output voltage value in channel <n> <DNPD>
Parameter represents the value of the output voltage expressed in V. <n>
Parameter represents the output channel (1, 2 or 3). If query is sent, calibrator returns the value of output voltage using standard exponential format.
Example: 110.12 V is returned as 1.101200e+002. Example:
Voltage 230.5V in channel 1: PDCE:VOLT1 230.5
Note: Command switches the calibrator in Power DC extended mode.
[SOUR]:PDCE:VOLT<n>:ENAB(?) <CPD> { ON | OFF }
This command enables (disables) the voltage output in channel <n>. Disabled outputs are OFF also during OUTP ON.
ON - enables the output  OFF - disables the output
If query is sent, calibrator returns ON if the output is enabled or OFF if it is disabled. Example:
PDCE:VOLT1:ENAB ON Note:
Command switches the calibrator in Power DC extended mode.
[SOUR]:PDCE:CURR<n>(?) <DNPD>
This command sets the output current value in channel <n>
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 94
<DNPD> Parameter represents the value of the output current expressed in A.
<n> Parameter represents the output channel (1, 2 or 3).
If query is sent, calibrator returns the value of output current using standard exponential format. Example: 11.012 V is returned as 1.101200e+001.
Example: Current 23.05A in channel 1: PDCE:CURR1 23.05
Note: Command switches the calibrator in Power DC extended mode.
[SOUR]:PDCE:CURR<n>:ENAB(?) <CPD> { ON | OFF }
This command enables (disables) the current output in channel <n>. Disabled outputs are OFF also during OUTP ON.
ON - enables the output  OFF - disables the output
If query is sent, calibrator returns ON if the output is enabled or OFF if it is disabled. Example:
PDCE:CURR1:ENAB ON Note:
Command switches the calibrator in Power DC extended mode.
[SOUR]:PHAR:POW?
This query returns active and reactive power values using standard exponential format. Example: 100.60 W, 1.1 VAr is returned as 1.006000e+002, 1.100000e+000. Note:
Query switches the calibrator in Power Harmonic mode.
[SOUR]:PHAR:VOLT<n>(?) <DNPD>
This command sets the output voltage value in channel <n> <DNPD>
Parameter represents the value of the output voltage expressed in V. <n>
Parameter represents the output channel (1, 2 or 3). If query is sent, calibrator returns the value of output voltage using standard exponential format.
Example: 110.12 V is returned as 1.101200e+002.
M133C Electric Power/Energy Calibrator MEATEST
Example: Voltage 230.5V in channel 1: PHAR:VOLT1 230.5
Note: Command switches the calibrator in Power Harmonic mode.
[SOUR]:PHAR:VOLT<n>:PHAS(?) <DNPD>
This command sets the phase angle between voltage output in channel <n> and internal reference phase.
<DNPD> Represents the phase angle in degrees.
If query is sent, calibrator returns the set value of the phase angle using standard exponential format. Example: 156.3 is returned as 1.563000e+002.
Example: PHAR:VOLT1:PHAS 0 PHAR:VOLT2:PHAS 120 PHAR:VOLT3:PHAS 240
Note: Command switches the calibrator in Power Harmonic mode.
[SOUR]:PHAR:VOLT<n>:ENAB(?) <CPD> { ON | OFF }
This command enables (disables) the voltage output in channel <n>. Disabled outputs are OFF also during OUTP ON.
ON - enables the output  OFF - disables the output
If query is sent, calibrator returns ON if the output is enabled or OFF if it is disabled. Example:
PHAR:VOLT1:ENAB ON Note:
Command switches the calibrator in Power Harmonic mode.
[SOUR]:PHAR:VOLT<n>:HARM<y>(?) <DNPD>
This command sets the level of <y> harmonic in output voltage in channel <n> <DNPD>
Parameter represents the level of <y> harmonic expressed in %. <n>
Parameter represents the output channel (1, 2 or 3).
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 96
<y> Parameter represents the harmonic order (1 to 50). Number 1 is for query only.
If query is sent, calibrator returns the level of the harmonic order using standard exponential format. Example: 10% is returned as 1.000000e+001.
Example: Set harmonic order 2 to 5.3% in voltage channel 1: PHAR:VOLT1:HARM2 5.3
Note: Command switches the calibrator in Power Harmonic mode.
[SOUR]:PHAR:VOLT<n>:HARM<y>:PHAS(?) <DNPD>
This command sets the phase shift of <y> harmonic component in channel <n> <DNPD>
Parameter represents the phase shift of <y> harmonic component expressed in °. <n>
Parameter represents the output channel (1, 2 or 3). <y>
Parameter represents the harmonic order (2 to 50). Order 1 has zero phase shift and can’t be changed.
If query is sent, calibrator returns the phase shift of harmonic component using standard exponential format. Example: 10° is returned as 1.000000e+001.
Example: Set phase shift of harmonic component 2 to 5.3° in voltage channel 1: PHAR:VOLT1:HARM2:PHAS 5.3
Note: Command switches the calibrator in Power Harmonic mode.
[SOUR]:PHAR:VOLT<n>:MOD(?) <DNPD>
This command sets the modulation level for voltage output in channel <n>. <DNPD>
Represents the modulation level in %. If query is sent, calibrator returns the modulation level using standard exponential format.
Example: 12.45 is returned as 1.145000e+001. Example:
PHAR:VOLT1:MOD 12.45 Note:
Command switches the calibrator in Power Harmonic mode.
M133C Electric Power/Energy Calibrator MEATEST
[SOUR]:PHAR:VOLT<n>:MOD:SHAP(?) <CPD> { SIN | RECT | OFF }
This command sets type of modulation for voltage output in channel <n>. <CPD>
SIN – modulation using sinusoidal wave shape  RECT – modulation using rectangular wave shape  OFF - disables modulation
If query is sent, calibrator returns SIN, RECT or OFF. Example:
PHAR:VOLT1:MOD:SHAP SIN Note:
Command switches the calibrator in Power Harmonic mode.
[SOUR]:PHAR:VOLT<n>:MOD:DUTY(?) <DNPD>
This command sets the duty cycle of rectangular modulation for voltage output in channel <n>. <DNPD>
Represents the duty cycle in %. If query is sent, calibrator returns the duty cycle using standard exponential format. Example: 50%
is returned as 5.000000e+001. Example:
PHAR:VOLT1:MOD:DUTY 50 Note:
Command switches the calibrator in Power Harmonic mode.
[SOUR]:PHAR:CURR<n>(?) <DNPD>
This command sets the output current value in channel <n> <DNPD>
Parameter represents the value of the output current expressed in A. <n>
Parameter represents the output channel (1, 2 or 3). If query is sent, calibrator returns the value of output current using standard exponential format.
Example: 11.012 A is returned as 1.101200e+001. Example:
Voltage 23.05A in channel 1: PHAR:CURR1 23.05
Note: Command switches the calibrator in Power Harmonic mode.
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 98
[SOUR]:PHAR:CURR<n>:PHAS(?) <DNPD>
This command sets the phase angle between current output in channel <n> and internal reference phase.
<DNPD> Represents the phase angle in degrees.
If query is sent, calibrator returns the set value of the phase angle using standard exponential format. Example: 156.3 is returned as 1.563000e+002.
Example: PHAR:CURR1:PHAS 0 PHAR:CURR2:PHAS 120 PHAR:CURR3:PHAS 240
Note: Command switches the calibrator in Power Harmonic mode.
[SOUR]:PHAR:CURR<n>:ENAB(?) <CPD> { ON | OFF }
This command enables (disables) the current output in channel <n>. Disabled outputs are OFF also during OUTP ON.
ON - enables the output  OFF - disables the output
If query is sent, calibrator returns ON if the output is enabled or OFF if it is disabled. Example:
PHAR:CURR1:ENAB ON Note:
Command switches the calibrator in Power Harmonic mode.
[SOUR]:PHAR:CURR<n>:HARM<y>(?) <DNPD>
This command sets the level of <y> harmonic in output current in channel <n> <DNPD>
Parameter represents the level of <y> harmonic expressed in %. <n>
Parameter represents the output channel (1, 2 or 3). <y>
Parameter represents the harmonic order (1 to 50). Number 1 is for query only. If query is sent, calibrator returns the level of the harmonic order using standard exponential
format. Example: 10% is returned as 1.000000e+001.
M133C Electric Power/Energy Calibrator MEATEST
Example: Set harmonic order 2 to 5.3% in voltage channel 1: PHAR:CURR1:HARM2 5.3
Note: Command switches the calibrator in Power Harmonic mode.
[SOUR]:PHAR:CURR<n>:HARM<y>:PHAS(?) <DNPD>
This command sets the phase shift of <y> harmonic component in channel <n> <DNPD>
Parameter represents the phase shift of <y> harmonic component expressed in °. <n>
Parameter represents the output channel (1, 2 or 3). <y>
Parameter represents the harmonic order (2 to 50). Order 1 has zero phase shift and can’t be changed.
If query is sent, calibrator returns the phase shift of harmonic component using standard exponential format. Example: 10° is returned as 1.000000e+001.
Example: Set phase shift of harmonic component 2 to 5.3° in current channel 1: PHAR:CURR1:HARM2:PHAS 5.3
Note: Command switches the calibrator in Power Harmonic mode.
[SOUR]:PHAR:CURR<n>:MOD(?) <DNPD>
This command sets the modulation level for current output in channel <n>. <DNPD>
Parameter represents the modulation level in %. If query is sent, calibrator returns the modulation level using standard exponential format.
Example: 12.45 is returned as 1.145000e+001. Example:
PHAR:CURR1:MOD 12.45 Note:
Command switches the calibrator in Power Harmonic mode.
[SOUR]:PHAR:CURR<n>:MOD:SHAP(?) <CPD> { SIN | RECT | OFF }
This command sets type of modulation for current output in channel <n>.
MEATEST M133C Electric Power/Energy Calibrator
Operation manual 100
<CPD>
SIN – modulation using sinusoidal wave shape  RECT – modulation using rectangular wave shape  OFF - disables modulation
If query is sent, calibrator returns SIN, RECT or OFF. Example:
PHAR:CURR1:MOD:SHAP SIN Note:
Command switches the calibrator in Power Harmonic mode.
[SOUR]:PHAR:CURR<n>:MOD:DUTY(?) <DNPD>
This command sets the duty cycle of rectangular modulation for current output in channel <n>. <DNPD>
Represents the duty cycle in %. If query is sent, calibrator returns the duty cycle using standard exponential format. Example: 50%
is returned as 5.000000e+001. Example:
PHAR:CURR1:MOD:DUTY 50 Note:
Command switches the calibrator in Power Harmonic mode.
[SOUR]:PHAR:FREQ(?) <DNPD>
This command sets the frequency in Power Harmonic mode. <DNPD>
Parameter represents the value of frequency expressed in Hz. If query is sent, calibrator returns the value of frequency using standard exponential format.
Example: 50Hz is returned as 5.000000e+001. Example:
Frequency 60Hz: PHAR:FREQ 60
Notes: Command switches the calibrator in Power Harmonic mode.
[SOUR]:PHAR:FREQ:MOD(?) <DNPD>
This command sets the modulation frequency in Power Harmonic mode. <DNPD>
Parameter represents the value of modulation frequency expressed in Hz.