Your new Measurement Computing product comes with a fantastic extra —
Management committed to your satisfaction!
Thank you for choosing a Measurement Computing product—and congratulations! You own the finest, and you can now enjoy
the protection of the most comprehensive warranties and unmatched phone tech support. It’s the embodiment of our mission:
To provide PC-based data acquisition hardware and software that will save time and save money.
Simple installations minimize the time between setting up your system and actually making measurements. We offer quick and
simple access to outstanding live FREE technical support to help integrate MCC products into a DAQ system.
Limited Lifetime Warranty: Most MCC products are covered by a limited lifetime warranty against defects in materials or
workmanship for the life of the product, to the original purchaser, unless otherwise noted. Any products found to be defective in
material or workmanship will be repaired, replaced with same or similar device, or refunded at MCC’s discretion. For specific
information, please refer to the terms and conditions of sale.
Harsh Environment Program: Any Measurement Computing product that is damaged due to misuse, or any reason, may be
eligible for replacement with the same or similar device for 50% of the current list price. I/O boards face some harsh
environments, some harsher than the boards are designed to withstand. Contact MCC to determine your product’s eligibility for
this program.
30 Day Money-Back Guarantee: Any Measurement Computing Corporation product may be returned within 30 days of
purchase for a full refund of the price paid for the product being returned. If you are not satisfied, or chose the wrong product by
mistake, you do not have to keep it.
These warranties are in lieu of all other warranties, expressed or implied, including any implied warranty of merchantability or
fitness for a particular application. The remedies provided herein are the buyer’s sole and exclusive remedies. Neither
Measurement Computing Corporation, nor its employees shall be liable for any direct or indirect, special, incidental or
consequential damage arising from the use of its products, even if Measurement Computing Corporation has been notified in
advance of the possibility of such damages.
3
Page 4
Trademark and Copyright Information
TracerDAQ, Universal Library, Measurement Computing Corporation, and the Measurement Computing logo are either
trademarks or registered trademarks of Measurement Computing Corporation.
Windows, Microsoft, and Visual Studio are either trademarks or registered trademarks of Microsoft Corporation
LabVIEW is a trademark of National Instruments.
CompactFlash is a registered trademark of SanDisk Corporation.
XBee and XBee-PRO are trademarks of MaxStream, Inc.
All other trademarks are the property of their respective owners.
Information furnished by Measurement Computing Corporation is believed to be accurate and reliable. However, no
responsibility is assumed by Measurement Computing Corporation neither for its use; nor for any infringements of patents or
other rights of third parties, which may result from its use. No license is granted by implication or otherwise under any patent or
copyrights of Measurement Computing Corporation.
All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted, in any form by any
means, electronic, mechanical, by photocopying, recording, or otherwise without the prior written permission of Measurement
Computing Corporation.
Notice
Measurement Computing Corporation does not authorize any Measurement Computing Corporation product for use
in life support systems and/or devices without prior written consent from Measurement Computing Corporation.
Life support devices/systems are devices or systems which, a) are intended for surgical implantation into the body,
or b) support or sustain life and whose failure to perform can be reasonably expected to result in injury.
Measurement Computing Corporation products are not designed with the components required, and are not subject
to the testing required to ensure a level of reliability suitable for the treatment and diagnosis of people.
4
Page 5
Table of Contents
Preface
About this User's Guide ....................................................................................................................... 6
What you will learn from this user's guide ......................................................................................................... 6
Conventions in this user's guide ......................................................................................................................... 6
Where to find more information ......................................................................................................................... 6
Introducing the USB-2404-10 ............................................................................................................... 7
Overview: USB-2404-10 features ...................................................................................................................... 7
Software features ................................................................................................................................................ 8
Chapter 2
Installing the USB-2404-10 ................................................................................................................... 9
What comes with your USB-2404-10 shipment? ................................................................ ............................... 9
Unpacking the USB-2404-10 ............................................................................................................................. 9
Installing the software ...................................................................................................................................... 10
Installing the hardware ..................................................................................................................................... 10
Calibrating the USB-2404-10 ........................................................................................................................... 10
Screw terminals (CH0 to CH3) ........................................................................................................................................11
USB connector .................................................................................................................................................................12
LED .................................................................................................................................................................................12
Strain relief slot for USB cable ........................................................................................................................................12
Signal source connections................................................................................................................................. 12
Analog input circuitry ....................................................................................................................................... 13
Analog input ..................................................................................................................................................... 15
Power ................................................................................................................................................................ 16
Bus interface ..................................................................................................................................................... 16
Declaration of Conformity .................................................................................................................. 18
5
Page 6
Preface
About this User's Guide
What you will learn from this user's guide
This user's guide explains how to install, configure, and use the USB-2404-10, and refers you to related
documents available on our web site, and to technical support resources.
Conventions in this user's guide
For more information on …
Text presented in a box signifies additional information and helpful hints related to the subject matter you are
reading.
Caution! Shaded caution statements present information to help you avoid injuring yourself and others,
damaging your hardware, or losing your data.
< : > Angle brackets that enclose numbers separated by a colon signify a range of numbers, such as those assigned
to registers, bit settings, etc.
bold text Bold text is used for the names of objects on the screen, such as buttons, text boxes, and check boxes. For
example:
1. Insert the disk or CD and click the OK button.
italic text Italic text is used for the names of manuals and help topic titles, and to emphasize a word or phrase. For
example:
The InstaCal installation procedure is explained in the Quick Start Guide.
Never touch the exposed pins or circuit connections on the board.
Where to find more information
For additional information relevant to the operation of your hardware, refer to the Documents subdirectory
where you installed the MCC DAQ software (C:\Program Files\Measurement Computing\DAQ by default), or
search for your device on our website at www.mccdaq.com.
Safety guidelines
You can connect hazardous voltages to the USB-2404-10 device's screw terminals. A hazardous voltage is a
voltage greater than 42.4 Vpk or 60 VDC to earth ground. Take the following precautions if you connect
hazardous voltages to the USB-2404-10:
Caution! Ensure that hazardous voltage wiring is performed only by qualified personnel adhering to local
electrical standards.
Caution! Do not mix hazardous voltage circuits and human-accessible circuits on the same device.
Caution! Make sure that devices and circuits connected to the USB-2404-10 are properly insulated from
human contact.
Caution! When device terminals are hazardous voltage LIVE (>42.4Vpk/60 VDC), ensure that devices and
circuits connected to the USB-2404-10 are properly insulated from human contact. Use the
ACC-160 connector backshell kit to ensure that the terminals are not accessible.
6
Page 7
USB
Microcontroller
USB
Hi-Speed
USB 2.0
Compliant
Interface
USB
CH0+
0-CH
Isolated
ADC
Amplifier
Overvoltage
Protection
+
-
Prefilter
Isolated
ADC
Amplifier
Overvoltage
Protection
+
Prefilter
Isolation barrier
CH1+
1-CH
Isolated
ADC
Amplifier
Overvoltage
Protection
+
-
Prefilter
Isolated
ADC
Amplifier
Overvoltage
Protection
+
Prefilter
Isolation barrier
CH2+
2-CH
Isolated
ADC
Amplifier
Overvoltage
Protection
+
-
Prefilter
Isolated
ADC
Amplifier
Overvoltage
Protection
+
Prefilter
Isolation barrier
CH3+
3-CH
Isolated
ADC
Amplifier
Overvoltage
Protection
+
-
Prefilter
Isolated
ADC
Amplifier
Overvoltage
Protection
+
Prefilter
Chapter 1
Introducing the USB-2404-10
Overview: USB-2404-10 features
This user's guide contains all of the information you need to connect the USB-2404-10 to your computer and to
the signals you want to measure.
The USB-2404-10 is a USB 2.0 high-speed device that is supported under popular Microsoft® Windows®
operating systems. The USB-2404-10 is fully compatible with both USB 1.1 and USB 2.0 ports.
The USB-2404-10 provides four channels of 24-bit simultaneous analog inputs.
The maximum sampling rate is 50 kS/s per channel, and the input voltage range is ±10 V, nominal.
An anti-alias filter removes noise that may be present in the signals prior to conversion. A 250 Vrms channel-
to-channel and channel-to-ground isolation protects the device and computer from ground spikes, and ensures a
reliable data stream. An on-board LED indicates the status of the device.
Field connections are made to four 2-position detachable screw terminals. Ten additional two-position
connectors are available with the optional ACC-102 connector kit.
The USB 2.0 high-speed driver transfers data at rates up to 480 Mbps. The USB-2404-10 is powered by the
+5 volt USB supply from your computer. No external power is required.
USB-2404-10 block diagram
USB-2404-10 functions are illustrated in the block diagram shown here.
Figure 1. USB-2404-10 functional block diagram
7
Page 8
USB-2404-10 User's Guide Introducing the USB-2404-10
Software features
For information on the features of InstaCal and the other software included with your USB-2404-10, refer to the
Quick Start Guide that shipped with your device. The Quick Start Guide is also available in PDF at
Check www.mccdaq.com/download.htm for the latest software version.
8
Page 9
Installing the USB-2404-10
What comes with your USB-2404-10 shipment?
As you unpack your USB-2404-10, verify that the following components are included.
Hardware
USB-2404-10
USB cable (2 meter length)
Chapter 2
Additional documentation
In addition to this hardware user's guide, you should also receive the Quick Start Guide (available in PDF at
www.mccdaq.com/PDFmanuals/DAQ-Software-Quick-Start.pdf). This booklet supplies a brief description of
the software you received with your USB-2404-10 and information regarding installation of that software.
Please read this booklet completely before installing any software or hardware.
Optional accessories
ACC-102 — Two-position screw terminal connector blocks (quantity ten). Additional details are available
on our web site at www.mccdaq.com/daq-accessory/acc-102.aspx.
ACC-160 — Backshell for use with the ACC-102 two-position screw terminal connector blocks. Provides
strain relief and operator protection from high-voltage signals (quantity six). Additional details are
available on our web site at www.mccdaq.com/daq-accessory/acc-160.aspx.
Unpacking the USB-2404-10
As with any electronic device, you should take care while handling to avoid damage from static
electricity. Before removing the USB-2404-10 from its packaging, ground yourself using a wrist strap or by
simply touching the computer chassis or other grounded object to eliminate any stored static charge.
If your USB-2404-10 is damaged, notify Measurement Computing Corporation immediately by phone, fax, or
e-mail.
Phone: 508-946-5100 and follow the instructions for reaching Tech Support.
Fax: 508-946-9500 to the attention of Tech Support
Email: techsupport@mccdaq.com
For international customers, contact your local distributor where you purchased the USB-2404-10. Click on this
link www.measurementcomputing.com/sales.asp to locate your distributor.
9
Page 10
USB-2404-10 User's Guide Installing the USB-2404-10
Installing the software
Refer to the Quick Start Guide for instructions on installingthe software on the Measurement Computing Data
Acquisition Software CD. This booklet is available in PDF at www.mccdaq.com/PDFmanuals/DAQ-Software-
Quick-Start.pdf.
Installing the hardware
Install the MCC DAQ software before you install your board
The driver needed to run your board is installed with the MCC DAQ software. Therefore, you need to install the
MCC DAQ software before you install your board. Refer to the Quick Start Guide for instructions on installing
the software.
Be sure you are using the latest system software
Before you install your USB-2404-10, run Windows Update to update your operating system with the latest
USB drivers.
To connect the USB-2404-10 to your system, turn your computer on, and connect the USB cable to a USB port
on your computer or to an external USB hub that is connected to your computer. The USB cable provides power
and communication to the USB-2404-10.
When you connect the USB-2404-10 for the first time, a series of Found New Hardware popup balloons
(Windows XP) or dialogs (other Windows versions) open as the USB-2404-10 is detected by your computer.
When the last balloon or dialog closes, the installation is complete. The LED on the USB-2404-10 blinks
steadily to indicate that the device is initialized and receiving power.
Caution! Do not disconnect any device from the USB bus while the computer is communicating with the
USB-2404-10, or you may lose data and/or your ability to communicate with the device.
Allow the USB-2404-10 to operate for at least 30 minutes before using the device. This warm up time is
required to achieve the specified rated accuracy of measurements.
Calibrating the USB-2404-10
The USB-2404-10 is shipped fully calibrated. Calibration coefficients are stored in EEPROM. Return the device
to Measurement Computing Corporation when calibration is required. The normal calibration interval is once
per year.
10
Page 11
Screw terminal
Signal
0
CH0+ (CH0 IN HI)
1
CH0– (CH0 IN LO)
0
CH1+ (CH1 IN HI)
1
CH1– (CH1 IN LO)
0
CH2+ (CH2 IN HI)
1
CH2– (CH2 IN LO)
0
CH3+ (CH3 IN HI)
1
CH3– (CH3 IN LO)
Functional Details
Components
The USB-2404-10 has the following external components:
Screw terminal connectors
USB port
LED
Strain relief slot for USB cable
Chapter 3
Figure 2. Front panel (Screw terminal connections)
Figure 3. Rear panel (USB connection and LED)
Screw terminals (CH0 to CH3)
The USB-2404-10 has four 2-terminal detachable screw terminals for connection to four isolated analog inputs.
The maximum sampling rate is 50 kS/s per channel, and the input voltage range is ±10 V, nominal. The
channels may be sampled individually or simultaneously. Signal assignments are listed in the following table.
Screw terminal pin assignments
Use 16 AWG to 28 AWG wires to connect signals to the device.
11
Page 12
USB-2404-10 User's Guide Functional Details
LED State
device status
Not lit
The device is not connected to a USB port or hub.
Continuous single-blink
The device is operating normally.
Continuous double-blink
The device is connected to a USB 1.1 Full-Speed port or hub, which may affect
performance.
Optimum performance requires connections to a USB 2.0 Hi-Speed host controller
(480 Mbps) and USB 2.0 high-speed hubs.
Signal
Source
+
CH+
CH-
USB-2404-10
USB connector
The USB connector provides +5 V power and communication. The voltage supplied through the USB connector
is system-dependent, and may be less than 5 V. No external power supply is required.
LED
The LED indicates the device status. When connected to a USB port, the LED blinks steadily to indicate that
the device is initialized and receiving power. Refer to the following table for the possible LED states.
LED States
Strain relief slot for USB cable
Use the strain relief slot to keep the USB cable from disconnecting from the device inadvertently. Feed a tie
wrap through the slot and secure to the USB cable when it is connected to the device.
High-voltage applications
For high voltage applications, we recommend using the ACC-160 backshell to ensure that the terminals are not
accessible. The backshell also provides strain relief to protect the screw terminals.
Figure 4. Connector backshell (ACC-160 accessory)
Additional details on this product are available on our web site at www.mccdaq.com/daq-accessory/acc-
160.aspx.
Signal source connections
You can connect ground-referenced or floating signal sources to the USB-2404-10.
Connect the positive signal of the signal source to the CH+ screw terminal.
Connect the negative signal of the signal source to the CH– screw terminal.
The following figures show the connections between the USB-2404-10 and both a grounded signal source and a
floating signal source.
Figure 5. Connecting a grounded signal source
12
Page 13
USB-2404-10 User's Guide Functional Details
Signal
Source
+
CH+
CH-
USB-2404-10
USB-2404-10
CH+
CH-
Overvoltage
Protection
Prefilter
ADC
Amplifier
When making a ground-referenced connection between the signal source and the USB-2404-10, make sure the
voltage on the CH+ and CH– connections are within the channel-to-earth safety voltage range. Refer to the
Specifications chapter for information about operating voltages and overvoltage protection.
Figure 6. Connecting a floating signal source
Don't connect to signals or use for measurements within category III or IV
Refer to the "Safety voltages" section in the Specifications chapter for more information about Measurement
Categories.
Analog input circuitry
The USB-2404-10 analog input channels are floating with respect to earth ground and to each other. The
incoming analog signal on each channel is conditioned, buffered, and then sampled by a 24-bit Delta-Sigma
ADC. Each channel provides an independent signal path and ADC, enabling you to sample all four channels
simultaneously. Figure 7 shows the circuitry for one analog input channel.
Figure 7. Analog input circuitry for one channel
Filtering
The USB-2404-10 uses a combination of analog and digital filtering to provide an accurate representation of
in-band signals while rejecting out-of-band signals. The filters discriminate between signals based on the
frequency range, or bandwidth, of the signal. The passband, stopband, and alias-free bandwidths are important.
The USB-2404-10 represents signals within the passband frequency, as quantified primarily by passband ripple
and phase nonlinearity. All signals within the alias-free bandwidth are either unaliased signals or signals that
have been filtered by at least the amount of the stopband rejection.
Passband bandwidth
The signals within the passband bandwidth have frequency-dependent gain or attenuation. The small amount of
variation in gain with respect to frequency is called the passband flatness. The device's digital filters adjust the
frequency range of the passband to match the data rate. Therefore, the amount of gain or attenuation at a given
frequency depends on the sample rate. Figure 8 shows typical passband flatness for the 50 kS/s sample rate.
13
Page 14
USB-2404-10 User's Guide Functional Details
Figure 8. Passband response (typical)
Stopband bandwidth
The filter significantly attenuates all signals above the stopband frequency (0.547 · , where is the sample
rate). The filter is used primarily to prevent aliasing. Therefore, the stopband frequency scales precisely with the
sample rate.
The stopband rejection (100 dB) is the minimum amount of attenuation applied by the filter to all signals with
frequencies within the stopband bandwidth.
Alias-free bandwidth
Any signal that appears in the alias-free bandwidth of the USB-2404-10 is not an aliased artifact of signals at a
higher frequency. The alias-free bandwidth (0.453 · is defined by the ability of the filter to reject frequencies
above the stopband frequency. The alias-free bandwidth equals the sample rate minus the stopband frequency.
Sample rates
A master timebase frequency () controls the USB-2404-10 sample rate ( ). The internal master timebase
frequency is 12.8 MHz.
The equation below provides the available sample rates of the USB-2404-10:
, where n is any integer from 1 to 31.
The sample rate must keep the device within the sample rate range. Refer to the Specifications chapter for more
information about the sample rate range.
Using the internal master timebase of 12.8 MHz results in sample rates of 50 kS/s, 25 kS/s, 16.67 kS/s and so on
down to 1.613 kS/s, depending on the value of n.
14
Page 15
Parameter
Conditions
Specification
Number of channels
4
A/D converter resolution
24-bit
A/D converter type
Delta-Sigma with analog pre-filtering
Sampling mode
Simultaneous
Sample rate range )
Minimum: 1.613 kS/s
Maximum: 50 kS/s (Note 1)
Sample rates )
Internal master timebase )
Frequency: 12.8 MHz
Accuracy: ±100 ppm maximum
Input voltage ranges
CH+ to CH–
±10 V, nominal
±10.52 V, typical
±10.3 V, minimum
Overvoltage protection
±100 V
Input coupling
DC
Input impedance
CH+ to CH–
1 MΩ
Input noise
70 µV rms
Gain drift
±5 ppm/ºC
Offset drift
±24 µV/ºC
Post calibration gain match
channel-to- channel, 20 kHz
0.22 dB maximum
Crosstalk
1 kHz
–130 dB
Phase mismatch
channel-to- channel
0.075º/kHz maximum
Phase nonlinearity
= 50 kS/s
0.11º maximum
Input delay
38.4 +3 µs
Passband frequency
0.453 ·
Passband flatness
= 50 kS/s
±100 mdB maximum
Stopband frequency
0.547 ·
Stopband rejection
100 dB
Alias-free bandwidth
0.453 ·
–3 dB pre-filter bandwidth
= 50 kS/s
24.56 kHz
Common mode rejection ratio
(CMRR)
= 60 Hz
126 dB
Spurious free dynamic range
(SFDR)
1 kHz, –60 dBFS
–128 dBFS
Total harmonic distortion
1 kHz, –1 dBFS
–99 dB
1 kHz, –20 dBFS
–105 dB
Specifications
All specifications are subject to change without notice.
Typical for the range 0 to 60 °C unless otherwise noted.
All voltages are relative to the CH– signal on each channel unless otherwise noted.
Analog input
Table 1. Analog input specifications
Chapter 4
Note 1: Full performance requires connections to a USB 2.0 Hi-Speed host controller and USB 2.0 high-
speed hubs.) The maximum sample rate may be lower on USB 1.1 ports.
15
Page 16
USB-2404-10 User's Guide Specifications
Measurement conditions
Percent of reading
(gain error)
Percent of range
(offset error) (Note 2)
Calibrated maximum (–0 to 60 ºC)
±0.13%
±0.05%
Calibrated typical (25 ºC, ±5 ºC)
±0.03%
±0.008%
Uncalibrated maximum
(–0 to 60 ºC)
±1.4%
±0.67%
Uncalibrated typical
(25 ºC, ±5 ºC)
±0.3%
±0.11%
Parameter
Specification
Current consumption from USB
500 mA, maximum
Parameter
Specification
USB specification
USB 2.0 Hi-Speed mode (480 Mbps) is recommended.
Otherwise USB 1.1 Full-Speed mode (12 Mbps)
Parameter
Specification
Operating temperature range
0 to 60 °C
Storage temperature range
–40 to 85 °C
Operating humidity
10 to 90% relative humidity, non-condensing
Storage humidity
5 to 95% relative humidity, non-condensing
Maximum altitude
2000 meters (6561.679 feet)
Pollution degree (IEC60664)
2
Parameter
Specification
Dimensions
4.5" L x 5.5" W x 1.5" H
Weight
1.2 lbs. (544 grams)
Accuracy
Table 2. Analog input accuracy
Note 2: The range is equal to 10.52 V.
Power
Table 3. Power specifications
Bus interface
Environmental
Mechanical
Table 4. Bus specifications
Table 5. Environmental specifications
Table 6. Mechanical specifications
16
Page 17
USB-2404-10 User's Guide Specifications
Parameter
Conditions
Specification
Channel-to-earth ground isolation
Continuous
250 Vrms, Measurement Category II (Note 4)
Withstand
2,300 Vrms, verified by a 5 sec dielectric withstand test
Channel-to-channel isolation
Continuous
250 Vrms, Measurement Category II (Note 4)
Withstand
1390 Vrms, verified by a 5 sec dielectric withstand test
Connector type
Screw terminal
Screw terminal wiring
16 to 28 AWG copper conductor wire with 7 mm (0.28 in.) of insulation stripped
from the end.
Backshell for use with the ACC-102 two-position screw terminal connector blocks. Provides
strain relief and operator protection from high-voltage signals (quantity four)
Safety voltages
Table 7. Safety specifications (Note 3)
Note 3: Connect only voltages that are within the limits specified in this table.
Note 4: Measurement Category II is for measurements performed on circuits directly connected to the
electrical distribution system. This category refers to local-level electrical distribution, such as
that provided by a standard wall outlet, for example 115 V for US or 230 V for Europe.
Caution! Do not connect the device to signals or use for measurements within Measurement Categories
III or IV.
Screw terminal connectors
Table 8. Screw terminal connector specifications
Accessory products
Table 9. Screw terminal assignments
Table 10. Screw terminal connector specifications
17
Page 18
Declaration of Conformity
Manufacturer: Measurement Computing Corporation
Address: 10 Commerce Way
Suite 1008
Norton, MA 02766
USA
Measurement Computing Corporation declares under sole responsibility that the product
USB-2404-10
to which this declaration relates is in conformity with the relevant provisions of the following standards or other
documents:
Category: Electrical equipment for measurement, control and laboratory use.
EC EMC Directive 2004/108/EC: General Requirements, EN 61326-1:2006 (IEC 61326-1:2005).
Emissions: IEC 61326-2-1:2005.
EN 55011 (1990)/CISPR 11 Radiated emissions: Group 1, Class A
EN 55011 (1990)/CISPR 11 Conducted emissions: Group 1, Class A
Immunity: EN 61326-1:2006, Table 3.
IEC 61000-4-2 (1995): Electrostatic Discharge immunity, Criteria B.
IEC 61000-4-3 (1995): Radiated Electromagnetic Field immunity, Criteria A.
To maintain compliance to the standards of this declaration, the following conditions must be met.
The host computer, peripheral equipment, power sources, and expansion hardware must be CE
compliant.
All I/O cables must be shielded, with the shields connected to ground.
I/O cables must be less than 3 meters (9.75 feet) in length.
The host computer must be properly grounded.
Equipment must be operated in a controlled electromagnetic environment as defined by Standards EN
61326:2006, or IEC 61326:2005.
Declaration of Conformity based on tests conducted by Chomerics Test Services, Woburn, MA 01801, USA in
September, 2008. Test records are outlined in Chomerics Test Report #EMI5135.08.
We hereby declare that the equipment specified conforms to the above Directives and Standards.
Carl Haapaoja, Director of Quality Assurance
Page 19
Measurement Computing Corporation
10 Commerce Way
Suite 1008
Norton, Massachusetts 02766
(508) 946-5100
Fax: (508) 946-9500
E-mail: info@mccdaq.com
www.mccdaq.com
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.