Measurement DaqScan User Manual

USER’S MANUAL
DaqScan/2000 Series
1126-0901 rev 2.2
*327556B-01*
372556B-01
DaqScan/2001 DaqScan/2002 DaqScan/2004 DaqScan/2005
Measurement Computing
10 Commerce Way
Norton, MA 02766
(508) 946-5100
Fax: (508) 946-9500
info@mccdaq.com
www.mccdaq.com
DaqScan/2000 Series
Ethernet-Based Data Acquisition System Components
ii
Warranty Information
Contact Measurement Computing by phone, fax, or e-mail in regard to warranty-related issues: Phone: (508) 946-5100, fax: (508) 946-9500, e-mail: info@mccdaq.com
Many Measurement Computing products carry the CE marker indicating they comply with the safety and emissions standards of the European Community. When applicable these products have a Declaration of Conformity stating which specifications and operating conditions apply. You can view the Declarations of Conformity at
www.mccdaq.com/legal.aspx (CE Information page).
Refer all service to qualified personnel. This caution symbol warns of possible personal injury or equipment damage under noted conditions. Follow all safety standards of professional practice and the recommendations in this manual. Using this equipment in ways other than described in this manual can present serious safety hazards or cause equipment damage.
This warning symbol is used in this manual or on the equipment to warn of possible injury or death from electrical shock under noted conditions.
This ESD caution symbol urges proper handling of equipment or components sensitive to damage from electrostatic discharge. Proper handling guidelines include the use of grounded anti-static mats and wrist straps, ESD-protective bags and cartons, and related procedures.
This symbol indicates the message is important, but is not of a Warning or Caution category. These notes can be of great benefit to the user, and should be read.
In this manual, the book symbol always precedes the words “Reference Note.” This type of note identifies the location of additional information that may prove helpful. References may be made to other chapters or other documentation.
Tips provide advice that may save time during a procedure, or help to clarify an issue. Tips may include additional reference.
Limitation of Liability
Measurement Computing cannot be held liable for any damages resulting from the use or misuse of this product.
Copyright, Trademark, and Licensing Notice
All Measurement Computing documentation, software, and hardware are copyright with all rights reserved. No part of this product may be copied, reproduced or transmitted by any mechanical, photographic, electronic, or other method without Measurement Computing’s prior written consent. IOtech product names are trademarked; other product names, as applicable, are trademarks of their respective holders. All supplied IOtech software (including miscellaneous support files, drivers, and sample programs) may only be used on one installation. You may make archival backup copies.
CE Notice
Warnings, Cautions, Notes, and Tips
Specifications and Calibration
Specifications are subject to change without notice. Significant changes will be addressed in an addendum or revision to the manual. As applicable, the hardware is calibrated to published specifications. Periodic hardware calibration is not covered under the warranty and must be performed by qualified personnel as specified in this manual. Improper calibration procedures may void the warranty.
CAUTION
Using this equipment in ways other than described in this manual can cause personal injury or equipment damage. Before setting up and using your
equipment, you should read all documentation that covers your system.
Pay special attention to Warnings and Cautions.
Note:
During software installation, Adobe® PDF versions of user manuals will automatically install onto your hard drive as a part of product support. The default location is in the Programs group, which can be accessed from the Windows Desktop. Initial navigation is as follows:
Start [Desktop “Start” pull-down menu]
Programs
IOtech DaqX Software
You can also access the PDF documents directly from the data acquisition CD by using the <View PDFs> button located on the opening screen.
Refer to the PDF documentation for details regarding both hardware and software.
PDF
1126-0901
DaqScan2000.pdf
Contains hardware-related material for DaqScan/2000 Series devices. This pdf file plus the following constitute a complete set of documentation for these devices.
PDF
457-0909
DaqView_DaqViewXL.pdf
Discusses how to install and use these “out-of-the-box” data acquisition programs.
PDF
1086-0926 1086-0922
PostAcquisition Analysis.pdf
This pdf consists of two documents. The first discusses eZ-PostView, a post data acquisition analysis program. The application is included free as a part of product support. The second includes information regarding eZ-FrequencyView and eZ-TimeView. These two applications have more features than does eZ-PostView and are available for purchase. They can; however, be used freely during a 30-day trial period.
iv 947091 DaqScan/2000 Series
PDF
457-0905
DBK Options.pdf
The DBK Option Cards and Modules User’s Manual discusses each DBK product. In addition, the document includes chapters on signal management, system connections, pinouts, DBK set up in software, and troubleshooting.
PDF
1008-0901
ProgrammersManual.pdf
The Programmer’s Manual pertains to developing custom programs using Applications Program Interface (API) commands.
Programmers should check the readme.file on the install CD-ROM for the location of program examples included on the CD.
Reference Note: A reference to additional documents of importance begins on page iv.
Manual Layout
Chapter 1 – Daq Systems
Chapter 2 – DaqScan/2000 Series, Device Overviews
Chapter 3 – Block Diagrams
Chapter 4 – Connectors, Indicators, Cables, & Pinouts
Chapter 5 – Software
Chapter 6 – Configuring the System for Operation
Chapter 7 – CE Compliance and Noise Considerations
Chapter 8 – Taking Measurements
Chapter 9 – Calibration
Chapter 10 – Troubleshooting and Customer Support
Chapter 11 – Specifications
Glossary
DaqScan/2000 Series 947091 v
This page is intentionally blank.
vi 947091 DaqScan/2000 Series

Table of Contents

Quick Start
1 – Daq Systems 2 – DaqScan/2000 Series, Device Overviews
Overview …… 2-1 Product Features …… 2-2 System Applications …… 2-3
3 – Block Diagrams
4 – Connectors, Indicators, Cables, & Pinouts
DaqScan/2000 Series Hardware …… 4-1 Cables ……4-3 Pinouts ……4-5
5 – Software 6 – Configuring the System for Operation 7 – CE Compliance and Noise Considerations 8 – Taking Measurements
Introduction …… 8-1 Scanning …… 8-1 Triggering …… 8-2 Synchronous I/O Operations …… 8-3 Asynchronous I/O Operations …… 8-5 Synchronizing Multiple Units …… 8-6 Using DBK Cards and Modules for Signal Conditioning …… 8-8
9 – Calibration 10 – Troubleshooting and Customer Support
Electrostatic Discharge (ESD), Handling Notice…… 10-1 Product Care …… 10-1 ReadMe Files and the Install CD-ROM …… 10-2 Driver Support…… 10-2 DaqCOM Issues …… 10-2 Ethernet Problems …… 10-3 Frequently Asked Questions …… 10-4 Customer Support …… 10-5
11 - Specifications
Glossary
DaqScan/2000 Series 898095 vii
This page is intentionally blank.
viii 898095 DaqScan/2000 Series

DaqScan/2000 Series Quick Start

Daq
S
Ethernet-Based Data Acquisition Components
can/2001
Step 1 - Install Software
1. Close all running applications on the host PC.
2. Insert the Data Acquisition CD into your CD-ROM drive and wait for the CD to auto-run. An Opening
Screen will appear. As an alternative, you can download software from www.PFFGDT.com/VRIWZDUH
Before you get started
Verify that you have the following items.
DaqScan/2000 Series Device
AC Power Cord
Ethernet Patch Cable
Ethernet jack Data Acquisition CD
Monitor: SVGA, 1024 x 768 screen resolution
Windows 2000 and Windows XP users
PC with Intel™ Pentium, 1 GHz or equivalent; 512 MB memory; 10 GB disk space
Windows Vista users:
PC must be Windows Vista Premium Ready
[on PC or on a hub connected to the Ethernet]
:
3. Click the <
Note: If you are downloading software from our website, follow instructions provided there.
4. From the hardware selection screen [which follows a licensing agreement], select the
device from the drop-down list and follow the on-screen instructions.
ENTER SETUP> button.
Step 2 - Connect the DaqScan to the Ethernet
In the figure, the DaqScan is connected directly to an Ethernet port on a host computer. Please consult your user’s manual (located in PDF format on the CD) should you need information regarding network types other than this simple direct connection.
1. Connect the Ethernet cable to the Ethernet jack on the DaqScan rear panel.
2. Connect the other end of the Ethernet cable to the Ethernet jack on the host computer or
network hub.
Step 3 - Connect the DaqScan to Power
1. With the DaqScan’s power switch “OFF,” connect the power cord to the unit.
2. Connect the cord’s plug to a standard AC outlet; then turn the DaqScan Po wer switch to “ON.”
1126-0940, rev 1.0 324633A-01 Printed in Hungary
Step 4 - Configure Computer Network Settings
Measurement Computing
10 Commerce Way
Norton, MA 02766
(508) 946-5100
Fax: (508) 946-9500
info@mccdaq.com
www.mccdaq.com
Note: We recommend that you discuss this procedure with your Network Administrator before proceeding.
1. Open the Control Panel by navigating from the Windows Desktop as follows:
Start Menu Settings Control Panel.
2. Double-click the “Network Connections” icon.
3. Double-click the icon for the network that the 2000 Series device is connected to.
4. In the “Local Area Connection Status” box, click on the <Properties> button. The “Local Area Connection
Properties” box will appear.
5. Double-click the “Internet Protocol (TCP/IP)” component. The “Internet Protocol (TCP/IP) Properties” box
will appear.
6. Select the “Use the following IP Address” radio
button. (See figure at left).
7. Set the IP address field to 10.0.0.x
where x is some number from 1 to 254.
Make sure that each computer and each device on the dedicated network has a unique IP address.
8. Set the Subnet mask to 255.0.0.0. Note that the
remaining fields can be left as is. Click <OK>.
9. Click <OK> o n follow-up screens to exit.
Internet Protocol (TCP/IP) Properties
(Partial View)
Step 5 - Configure & Test the System
1. Open the Daq Configuration Applet.
a. Navigate from the Windows’ Desktop: Start Menu Settings Control Panel b. From the Control Panel, double-click the Daq Configuration icon.
2. Add the 2000 Series Device to the list of installed devices.
a. Select the Computer image in the Device Inventory configuration tree. b. Click the <Add Device> button. The “Select Device Type” box will appear. c. Select the DaqScan/2000 Series device from the list of devices. d. Click the <OK> button. The “Properties” box will appear for the selected device. e. En ter the Serial Number of the device. The number is on the MAC label (rear panel, upper left). The
serial number is located just below a barcode.
f. Select the “Auto IP Setting” radio button. The IP Address of the device will automatically be
calculated and displayed in the IP Address field
g. Click the <OK> button.
3. Test the system connections.
a. Make sure the device has been properly installed and is powered-on. b. Make sure all cables are properly and securely connected. c. Click the “Test Hardware” tab. d. Click the <TCP/IP Test> button. This tests the Transmission Control Protocol / Internet Protocol.
e. Upon completion of the TCP/IP test, click the <Resource Test> button.
When testing, if the unit does not respond within 30 seconds perform the following steps:
1) reboot the system, 2) upon power-up, re-open the Daq Configuration applet, 3)select another configuration setting, 4) reinitiate the test.
Step 6 - Connect Data Acquisition Signal Lines
Prior to making signal connections review the Specifications chapter of your user’s manual to ensure that your intended signal inputs do not exceed the specified limits. The manual is included in PDF format on your CD.
*324633A-01*
324633A-01
Printed in Hungary
Daq Systems 1
Daq Systems, the Modular Concept …… 1-1 DaqBooks, DaqBoards, and Daq PC-Cards ….. 1-2 Using DBK Cards and Modules for Signal Conditioning …… 1-4 Daq Software …… 1-5
This chapter provides a brief overview of Daq devices.
Users interested in jumping straight to DaqScan information should skip directly to chapter 2.

Daq Systems, the Modular Concept

Daq equipment and software form a modular, interrelated family of products that provide great flexibility in data acquisition system design. This flexibility allows for the development of custom systems that are unique to the user, and which can be optimized for his or her specific application needs. With the Daq product line, system expansion or redesign can typically be accomplished with relative ease.
Primary Acquisition Device. This is the main data acquisition device, e.g., a
DaqBook, DaqBoard, DaqLab, DaqScan, or Daq PC-Card. These devices provide a vital data conversion and communications link between the data source of transducers and signal conditioners and the data processor of the host computer.
DBK Option Cards and Modules. Over 35 DBK cards and modules (the number is
constantly growing) provide various types of signal conditioning and system expansion. Note that certain DBK modules exist for the purpose of supplying power to other members of the acquisition system. The DBK options are discussed in the DBK Option Cards and User’s Manual (p/n 457-0905).
Note: Only passive DBKs, such as the DBK1 BNC module, the DBK11A screw terminal
card, and the DBK40 BNC analog interface, can be used with a Daq PC-Cards.
Reference Note: DBK options are discussed in the DBK Option Cards and Modules User’s Manual (p/n 457-0905). As a part of product support, this manual is automatically loaded onto your hard drive during software installation. The default location is the Programs directory, which can be accessed through the Windows Desktop.
Software. DaqView out-of-the-box software provides a graphical user interface with
easy to read spreadsheet formats for viewing channel data, as well as a choice of analog, digital, and bar-graph meters. Waveform analysis can be performed, when applicable. A product support option, included on the data acquisition CD, provides a means of performing post data analysis. More information is included in the software-specific PDF documents that are installed on your hard-drive as a part of product support.
In addition to the included out-of-the-box software, Daq products can be controlled via user-written custom programs through Applications Program Interface (API). Several languages are supported, e.g., C/C++, and VisualBASIC.
Note that third party icon-driven software, such as DASYLab and LabView, can be used with the Daq devices.
DaqScan/2000 Series 949595 Daq Systems 1-1
DaqView and DASYLab can only be used with one DaqScan/2000 Series device at a time.
LabView can be used with multiple Daq/2000 Series devices.
For multiple device applications (via custom programming) refer to the Using Multiple Devices section of the Programmer’s Manual. During software installation from the data acquisition CD (p/n 1022-0601), a PDF version of the Programmer’s Manual is automatically loaded onto your hard drive as a part of product support. The default location is the Programs directory.
Reference Note:
Programming topics are covered in the Programmer’s User Manual (p/n 1008-0901). As a part of product support, this manual is automatically loaded onto your hard drive during software installation. The default location is the Programs directory, which can be accessed through the Windows Desktop.

DaqBooks, DaqBoards and Daq PC-Cards

Daq products typically connect to one or more DBKs on their signal input side and to a host computer on their output side. The following table categorizes the various Daq devices and provides very basic information.
1-2 Daq Systems 949595 DaqScan/2000 Series
Daq Data Acquisition Devices
Category Device Description
Primary Data Acquisition Device
DBK Option Cards and Modules*
Software
DaqLab/2001 DaqLab/2005
DaqScan/2001 DaqScan/2002 DaqScan/2004 DaqScan/2005
DaqBook/2001 DaqBook/2005 DaqBook/2020
DaqOEM/2001 DaqOEM/2005
DaqBook/100 Series DaqBook/200 Series
DaqBook/2000A DaqBook/2000E DaqBook/2000X
DaqBoard/2000 Series
DaqBoard/2000c Series
DaqBoard (ISA types)*
Daq PC-Card*
Analog Signal Conditioning
Analog Output
Digital I/O and Control
Expansion Connections
Power Supply Included Software
Optional Software
Ethernet-Based Laboratory Data Acquisition 16-bit, 200 kHz devices with Ethernet connectivity.
Each can serve as host to an internal analog and/or digital DBK card. Each has a built-in AC-to-DC power circuit and can be plugged directly into a standard AC outlet.
Ethernet-Based Data Acquisition System Components 16-bit, 200 kHz devices with Ethernet connectivity.
Each has a built-in AC-to-DC power circuit and can be plugged directly into a standard AC outlet.
Portable Data Acquisition 16-bi
t, 200 kHz
Make use of an external power supply. Portable Data Acquisition
16-bit, 200 kHz These boards reside external to the host PC.
Portable Data Acquisition
12-
DaqBook/100, /112, /120
bit:
16-bit: DaqBook/200, /216, /260
Portable Data Acquisition 16-bi
t, 200 kHz
the DaqBook/2000E provides for a direct Ethernet connection. These devices include a 100-pin P4 connector. Each makes use of an external power supply.
Plug-In Boards for PCI Bus-Slots
16-bit ,
200 kHz. Six boards identified as /2000 through /2005
Plug-In Boards for Compact-PCI Bus-Slots
16-bit ,
200 kHz. Six boards identified as /2000c through
/2005c
Plug-In Boards for ISA Bus-Slots
12-bit:
DaqBoard/100A, /112A
16-bit: DaqBoard/200A, /216A, /2000
Plug-In PCMCI Card
12-
bit:
Daq/112B
16-bit: Daq/216B
Cards and DBK modules used to condition Analog Signals
DBK/ 4, 52, 53, 54, 55, 65, 80, 81, 82, 83, 84, 85, 90, 100, 207, 207/CJC
Cards used to modify Analog Output Signals
DBK/ 2, Cards and DBK modules used to condition Digital I/O DBK/ 20, 21, 23, 24, 25, 208, 210
Cards and DBK modules used to expand the acquisition system.
D 206, 209
DBKs: 30A, 32A, 33, 34 DaqView, Post Data Acquisition Analysis Program (actual
application not specified), Visual Basic ex Programming Interface (API)
DaqViewXL, DASYLab
7, 8, 9, 12, 13, 15, 17, 18, 19, 42, 43A, 44, 45, 50, 51,
5
BK/ 1,
10, 11A, 35, 40, 41, 60, 200, 201, 202, 203, 204, 205,
devices with Ethernet connectivity.
boards with Ethernet connectivity.
devices with Ethernet connectivity, though only
tensions, Application
* The list of DBK cards and modules continues to grow. For information regarding new and upcoming
products consult out website, your sales representative, or contact the factory directly.
DaqScan/2000 Series 949595 Daq Systems 1-3

Using DBK Cards and Modules for Signal Conditioning

The DBK signal-conditioning cards and modules are designed for use with DaqLab, DaqScan, DaqBook, LogBook, and various data acquisition boards, i.e., ISA, PCI, and compact-PCI types. The DBKs perform best when used with an acquisition device that can dynamically select channel, gain, and range. DBK cards and modules with dynamic channel and gain/range selection allow for high channel-to-channel scan rates with a variety of transducers.
Note: Only passive DBKs, such as the DBK1 BNC module, the DBK11A screw terminal card,
and the DBK40 BNC analog interface, can be used with Daq PC-Cards.
DBK output signals can be bipolar, e.g., -5 to +5 V, or unipolar, e.g., 0 to 10 V. The user can select a range of relevant values to correspond to the lowest signal (e.g., -5 or 0 V) and the highest signal (e.g., 5 or 10 V) signal. This type of range selection guarantees the highest resolution in 12-bit or 16-bit conversion.
DBK modules share the same footprint as the DaqBook and a typical notebook PCs; allowing for convenient stacking. The majority of these modules have their own power supply; however, several options exist for packaging and powering the DBKs.
Reference Note: DBK options are detailed in the DBK Option Cards and Modules User’s Manual (p/n 457-0905). As a part of product support, this manual is automatically loaded onto your hard drive during software installation. The default location is the Programs directory, which can be accessed through the Windows Desktop
.
1-4 Daq Systems 949595 DaqScan/2000 Series

Daq Software

The Daq devices have software options capable of handling most applications. Three types of software are available:
ready-to-use graphical programs, e.g., DaqView, DaqViewXL, and post acquisition
drivers for third-party, icon-driven software such as DASYLab and LabView
various language drivers to aid custom programming using API
Ready-to-use programs are convenient for fill-in-the-blank applications that do not require programming for basic data acquisition and display:
DaqView is a Windows-based program for basic set-up and data acquisition. DaqView
DaqViewXL allows you to interface directly with Microsoft Excel to enhance data
A post acquisition data analysis programs, e.g., eZ-PostView, typically allows you to
data analysis programs such as eZ-PostView
lets you select desired channels, gains, transducer types (including thermocouples), and a host of other parameters with a click of a PC’s mouse. DaqView lets you stream data to disk and display data in numerical or graphical formats. PostView is a post-acquisition waveform-display program within DaqView.
handling and display. Within Excel you have a full-featured Daq control panel and all the data display capabilities of Excel.
view and edit post-acquisition data.
The Daq Configuration control panel allows for interface configuration, testing, and
troubleshooting.
Each Daq system comes with an Application Programming Interface (API) via a subroutine library (DAQ) or an ActiveX/COM based library (DaqCOM). API-language drivers include: C/C++, and Visual Basic. The latest software is a 32-bit version API.
Reference Notes:
¾ The software document modules, DaqView, DaqViewXL, and Post Acquisition
Data Analysis User’s Guide, are not included as part of the hardcopy manual,
but are available in PDF version. See the PDF Note, below.
¾ Programming topics are covered in the Programmer’s User Manual (1008-
0901). As a part of product support, this manual is automatically loaded onto your hard drive during software installation. The default location is the Programs directory, which can be accessed through the Windows Desktop.
PDF
Note:
During software installation, Adobe® PDF versions of user manuals will automatically install onto your hard drive as a part of product support. The default location is in the Programs group, which can be accessed from the Windows Desktop. Refer to the PDF documentation for details regarding both hardware and software.
®
A copy of the Adobe Acrobat Reader provides a means of reading and printing the PDF documents. Note that hardcopy versions of the manuals can be ordered from the factory.
is included on your CD. The Reader
DaqScan/2000 Series 949595 Daq Systems 1-5
1-6 Daq Systems 949595 DaqScan/2000 Series
DaqScan/2000 Series 2
Device Overviews
Overview …… 2-1 Product Features …… 2-2 System Applications …… 2-3

Overview

DaqScan/2000 Series devices are high-speed, multi-function, data acquisition devices for use with Ethernet PCs and Ethernet networks. Each unit connects directly to a PC’s Ethernet port or network hub. All I/O is accessed via female DB37 connectors located on the rear of the
unit, making cabling easy from the DaqScan to the test device. The DaqScan/2000 Series includes four models, ranging from the DaqScan/2002, which offers
gital/frequency I/O,
di analog outputs, 40 digital I/O, 4 counter inputs and 2 timer outputs. Each features a 16-bit,
200 kHz A/D converter and digital calibration. In addition, all four units include an internal AC-to-DC power converter.
I/O Comparison Matrix
to the full-featured DaqScan/2001 which includes 16 analog inputs, 4
Model Analog Inputs
SE / DIFF
DaqScan/2001 16/8 40 4 2 4
DaqScan/2002 0 40 4 2 0
DaqScan/2004 0 40 4 2 4
DaqScan/2005 16/8 40 4 2 0
Digital
I/O
Counter
Inputs
Timer
Outputs
Analog
Outputs

Synchronous Operation Matrix

Operation

Synchronous Input

DaqScan/
2001
Analog Main Unit Inputs Yes No No Yes Analog Expansion Input Yes No No Yes Counter Inputs Yes Yes Yes Yes Digital Main Unit Inputs Yes Yes Yes Yes Digital Expansion Inputs Yes Yes Yes Yes

Synchronous Output

Analog D/A Waveform Output Yes No Yes No

Asynchronous IO

Main Unit Digital I/O Yes Yes Yes Yes Expansion Digital I/O Yes Yes Yes Yes Timer Output
Yes Yes Yes Yes
(Pulse Generation)
Analog Output Yes No Yes No
DaqScan/
2002
DaqScan/
2004
DaqScan/
2005
DaqLab/2000 Series 898095 Device Overviews 2-1
Each DaqScan/2000 Series model is packaged in 1U high (1.75”), 19” wide full-rack package, and includes a rackmount kit that can attach to either the front or the rear of the enclosure. Multiple DaqScan models can be combined in the same system. Up to 4 units can be synchronized [scan synchronous post-trigger] via rear panel SYNC connectors.
DaqScan/2000 Series devices include comprehensive drivers for popular Windows-based environments, including Visual Basic, C/C++, ActiveX/COM, LabVIEW, MATLAB, and DASYLab. The devices include DaqView, an interactive spreadsheet-style application that is ideal for verifying signal connections during system design.
¾ Block diagrams are presented in chapter 3. ¾

Product Features

• Can be used to add analog I/O, digital I/O and frequency I/O to Ethernet-based test
systems. Each DaqScan/2000 Series device has a different combination of features (see comparison matrix on page 2-1).
All I/O can be synchronous, enabling precise timing between various I/O functions
Up to 4 units can be synchronized [scan synchronous post-trigger] via rear panel
SYNC connectors.
8 differential, or 16 single-ended inputs, expandable up to 256 voltage or 896 TC
channels using signal conditioning and expansion options
Up to 40 built-in TTL-level digital I/O, expandable up to 256 channels of isolated I/O
using low-cost isolation modules
• Configuration through software only. Device settings are made via software. There are no hardware settings to be made, for example, there are no device DIP switches to be set or jumpers to be positioned. However, note that DBK expansion options may require s small degree of manual set up. Refer to the DBK Option Cards and Modules User’s Manual (457-0905) for details. A PDF version of the document is included on the installation CD.
Includes support for Visual Basic®, C/C++, ActiveX/COM, LabVIEW®, MATLAB®, and
DASYLab®
Includes Out-of-the-Box™ DaqView™ software to facilitate signal and wiring
verification with no programming required
Convenient 1U high (1.75”), 19” wide rack mount package minimizes rack space in
test systems
Reference Notes:
Connectors and pinouts are discussed chapter 4.
Four 16-bit counters. Each can accept frequency inputs up to 10 MHz. The counters can be cascaded into two 32-bit counters. Counter connection is via a removable screw terminal block on the front panel.
Two 16-bit timer outputs. Each can generate square waves from 16 Hz to 1 MHz. Connection to the timer outputs is via a removable screw terminal block on the front panel
Scan Sequencing. Each DaqLab/2000 Series device has a scan sequencer that can select up to 16384 channel/range combinations. Scan rates are programmable on a per channel basis to 5 µs/channel, 10 µs/channel, or 1000 µs/channel.
Note: Inputs can be measured at the 5 µs/channel rate. However, some DBK
expansion options require more settling time and can be measured at the slower rate of 10 µs/channel or 1000 µs/channel. All types of measurements can be mixed within the scan group.
2-2 Device Overviews 898095 DaqLab/2000 Series

System Applications

In addition to basic applications that make use of a DaqScan model’s built-in I/O, a wi de variety of signal conditioning and expansion options make additional applications possible. Each of the following sample systems are obtained by using a DaqScan along with one or more DBK options. The applications, which are only briefly discussed, include:
High Channel Count Thermocouple Measurements
High-Isolation Voltage & Thermocouple Measurements
Isolated Digital I/O
High Speed Voltage Measurement
Multifunction I/O
High Channel Count Thermocouple Measurements
When combined with 56-channel DBK90 thermocouple input modules, DaqScan can measure up to 896 channels of thermocouple (TC) input. Any TC type can be installed into any channel using standard mini-TC connectors. DBK90’s built-in cold junction compensation (CJC) coupled with the software’s TC conversion algorithms make temperature measurements easy. Note that thermocouples are measured at a rate of 1 ms/channel in DBK90 based systems.
High-Isolation Voltage & Thermocouple Measurements
DaqScan units can be combined with DBK207/CJC™ options to create an isolated system capable of measuring up to 256 channels of voltage, thermocouple, RTD and strain gage inputs. All input channels can be scanned up to 200 kHz, and are isolated by 500V from other channels and from system common. Any combination of input signals is possible by selecting the appropriate 5B signal conditioning module for the DBK207/CJC.
Isolated Digital I/O System
When the DaqScan/2002 is combined with a DBK208 or DBK210, up to 256 channels of isolated digital I/O are possible, capable of controlling AC or DC voltages up to 250V, or sensing the presence of AC or DC voltages. All channels are isolated from one another and from the system by up to 500V. Inputs can be scanned in 16-bit increments at speeds up to 200 Kreadings/s.
High-speed Voltage Measurement System
The DaqScan/2005 can be combined with DBK85 16-channel voltage input modules. This creates a 5 µs/channel voltage measurement system with up to 256 channels. Each input can have a range from 156 mV FS to 10V FS, which is programmable on a per-channel basis.
Multifunction I/O System
All of the aforementioned capabilities can be combined into one system, using one DaqScan/2001 as the heart of the system. Such a DaqScan/2001 system could provide 56 non-isolated TC inputs, 16 isolated voltage inputs, 16 low-voltage inputs, 4 analog outputs, 4 frequency inputs, and 32 isolated discrete high-voltage outputs.
Reference Note: For information regarding DBK options refer to the DBK Options User’s Manual (p/n 457-0905). A PDf version of the manual is included on your installation CD.
DaqLab/2000 Series 898095 Device Overviews 2-3
2-4 Device Overviews 898095 DaqLab/2000 Series
Block Diagrams 3
DaqScan/2001 Block Diagram
DaqScan/2000 Series 947495 Block Diagrams 3-1
DaqScan/2002 Block Diagram
DaqScan/2004 Block Diagram
3-2 Block Diagram 947495 DaqScan/2000 Series
DaqScan/2005 Block Diagram
DaqScan/2000 Series 947495 Block Diagrams 3-3
3-4 Block Diagram 947495 DaqScan/2000 Series
Connectors, Indicators, Cables, & Pinouts 4
DaqScan/2000 Series Hardware …… 4-1 Cables …… 4-3 Pinouts …… 4-5

DaqScan/2000 Series Hardware

DaqScan Front Panel
Front Panel
The DaqScan/2000 Series front panels contain a power indicator LED. The power switch and
connectors are on the rear panel.
DaqScan/2001, Rear Panel
Rear Panel
The DaqScan/2000 Series rear panels each contain a Power Switch, AC power cord connector,
2 Synchronization jacks, a MAC Address Label, Ethernet connector, 5 LEDs [ 2 of which are on the Ethernet connector], and a set of DB37 connectors. All four DaqScan models have P2 and P3 connectors; but in addition, DaqScan/2001 and DaqScan/2005 have P1 connectors for analog I/O.
P1 - DB37 Connector: P1 applies to DaqScan/2001 and DaqScan/2005 only. P1 is the analog input port for 16 single-ended analog inputs or 8 differential, programmable on a per­channel basis as single-ended or differential and as uni-polar or bipolar. The 3 programmable ranges are from ±10 V to ±156 mV full scale. P1 can be used to connect the DaqScan to an optional analog DBK card.
P2 - DB37 Connector: P2 is a general purpose digital I/O port with 24 (3 x 8-bit) lines on P2,
or digital I/O expansion port controlling up to 272 external lines. Programmable as input or
output. P2 can be used to connect to an optional digital DBK card.
P3 - DB37 Connector: 16-bit digital I/O port with 4 counter inputs and 2 timer outputs. In
addition, DaqScan/2001 and DaqScan/2004 each have a P3 connection to 4 analog outputs
(DAC0, DA1, DAC2, and DAC3).
DaqScan/2000 Series 898095 Connectors, Indicators, Cables, & Pinouts 4-1
POWER Switch and Power LED: A rocker-type switch with a “0” label for Power Off, and a “1” for Power On. When the unit is on and has power, the POWER LED will light. An AC to DC power converter resides within each DaqScan/2000 series device.
MAC Address Label: The Media Access Control (MAC) label shows the device serial number in
three formats: (1) Barcode, (2) base 10, and (3) hexadecimal. If prompted to enter a serial
number in software, use the base 10 number. Conversion to a hexadecimal number for use in
addressing will be automatic.
ETHERNET: The 10/100BaseT Ethernet jack can be used to connect to the Ethernet port of
the host PC, or to an Ethernet network. Either of two Ethernet patch cables may be used to
make the connection. CA-242 is a 1.5 foot cable. CA-242-7 is a 7-foot cable.
The Ethernet connector has two built in LEDs (Tx and Rx) that indicate traffic flow. These are
defined in the following LED section. Note that the Ethernet cable length must be <10m in
order for the system to be CE Compliant.
LEDs: There are 5 ETHERNET Status LEDS. Two rectangular LEDs, Tx
and Rx are built right into the frame of the Ethernet jack. The other
three, located just to the right of the jack, are round and are labeled L, T,
and D.
Tx – “ON” indicates traffic is being transmitted (see figure at right).
Rx – “ON” indicates that the port is receiving traffic. L (Link) “ON” indicates a link exists. “OFF” indicates no link. T (Transmit) “ON” indicates 100 M-Bits, “OFF” indicates
10 M-Bits. D (Duplex) “ON” indicates full duplex, which allows
simultaneous two-way data traffic. “OFF” indicates half­duplex, which only allows one-way data traffic at any given time.
Tx and Rx LEDs
SYNC (Qty of 2) – Two “synchronization ports” provide a means of synchronizing up to four
DaqScan/2000 Series units in regard to post post-trigger scanning. The ports accept CA-74-1
and CA-74-5 RJ-11, 6 conductor type cables. The total combined length of the SYNC cables is
not to exceed 15 feet (4.57 m).
POWER IN: 90 to 250 VAC, 3-prong socket for power cord.
4-2 Connectors, Indicators, Cables, & Pinouts 898095 DaqScan/2000 Series

Cables

Should you need help in regard to cables or have any other product inquiries please contact the factory.
Cables for use with DaqScan/2000 Series Devices
Description Part No.
1. Molded expansion cable. Connects a DBK module to a DaqScan, 2 in.
2. Molded expansion cable. Connects a DBK module to a DaqScan, 4 in.
3. DBK expansion cable. Connects a DBK module to a DaqScan, 2.5 in.
4. DBK expansion cable. Connects a DBK module to a DaqScan, 4.5 in.
5. Expansion-card cable. Connects DBK series expansion card options. The “x” in the part no. indicates the number of cards that can be connected via the cable.
6. Ethernet patch cable, shielded, 18 in. CA-242
7. Ethernet patch cable, shielded, 7 ft. CA-242-7
8. Sync cable, RJ-11, 6 conductor, 1 ft. CA-74-1
9. Sync cable, RJ-11, 6 conductor, 5 ft. CA-74-5
CA-255-2T
CA-255-4T
CA-37-1T
CA-37-3T
CA-37-x
The total combined length of the SYNC cables is not to exceed 15 feet (4.57 m).
DaqScan/2000 Series 898095 Connectors, Indicators, Cables, & Pinouts 4-3
4-4 Connectors, Indicators, Cables, & Pinouts 898095 DaqScan/2000 Series

Pinouts

P1 – Analog I/O (Applies to DaqScan/2001 and DaqScan/2005 only) …… 4-6 P2 – Digital I/O …… 4-7 P3 – Pulse/Frequency/Digital I/O…… 4-8
CAUTION
Turn off power to all devices connected to the system before connecting cables or setting configuration jumpers and switches. Electrical shock or damage to equipment can result even under low-voltage conditions.
CAUTION
The discharge of static electricity can damage some electronic components. Semiconductor devices are especially susceptible to ESD damage. You should always handle components carefully, and you should never touch connector pins or circuit components unless you are following ESD guidelines in an appropriate ESD controlled area. Such guidelines include the use of properly grounded mats and wrist straps, ESD bags and cartons, and related procedures.
Each of the four DaqScan/2000 Series devices has 37-pin connectors designated as P2 and P3. In addition, DaqScan/2001 and DaqScan/2005 have a third DB37 connector, P1.
Signal assignment to connectors is as follows.
P1 is used for Analog I/O (DaqScan/2001 and DaqScan/2005 only)
P2 is used for Digital I/O
P3 is used for Pulse/Frequency and Digital I/O; in addition, for DaqScan/2001 and
DaqScan/2004, P3 includes active pins for 4 analog output (DAC) channels.
To connect a DBK option to a DaqScan connect an appropriate cable (see table page 4-2) to the appropriate DB37 connector [P1, P2, or P3] on the DaqScan and connect the free end of the cable to the DBK card or module. If you are not familiar with DBK options, or need refreshed, refer to the DBK Options Manual, p/n 457-0905.
DaqScan/2000 Series 898095 Connectors, Indicators, Cables, & Pinouts 4-5
P1 (Applies only to DaqSCan/2001 and DaqScan/2004) Analog I/O
P1 Pin Signal Name Description
1 +5 Volts Expansion +5 V power 2 -15 Volts Expansion -15 VDC power 3 Expansion 7 Digital OUT, external ADDRESS select bit 3 4 Expansion 9 Digital OUT, external ADDRESS select bit 1 5 Expansion 5 Digital OUT, external GAIN select bit 1 6 Expansion 6 Digital OUT, external GAIN select bit 0 7 DGND Digital Ground 8 Negative Reference Analog, -5 V reference
9 Positive Reference Analog +5V reference 10 Not Connected N/A 11 CH 15 (SE), or CH 7 LO DIFF Ch 15 HI IN (single-ended mode) / Ch 7 LO IN (differential mode) 12 CH 14 (SE), or CH 6 LO DIFF Ch 14 HI IN (single-ended mode) / Ch 6 LO IN (differential mode) 13 CH 13 (SE), or CH 5 LO DIFF Ch 13 HI IN (single-ended mode) / Ch 5 LO IN (differential mode) 14 CH 12 (SE), or CH 4 LO DIFF Ch 12 HI IN (single-ended mode) / Ch 4 LO IN (differential mode) 15 CH 11 (SE), or CH 3 LO DIFF Ch 11 HI IN (single-ended mode) / Ch 3 LO IN (differential mode) 16 CH 10 (SE), or CH 2 LO DIFF Ch 10 HI IN (single-ended mode) / Ch 2 LO IN (differential mode) 17 CH 9 (SE), or CH 1 LO DIFF Ch 9 HI IN (single-ended mode) / Ch 1 LO IN (differential mode) 18 CH 8 (SE), or CH 0 LO DIFF Ch 8 HI IN (single-ended mode) / Ch 0 LO IN (differential mode) 19 Signal Ground (SGND) Sense Common (SGND) 20 A/I Clock External ADC Pacer Clock Input / Internal ADC Pacer Clock Output 21 +15 Volts Expansion +15 V power 22 Expansion 8 Digital OUT, external ADDRESS select bit 2 23 Expansion 10 Digital OUT, external ADDRESS select bit 0 24 Not Connected N/A 25 TTL Trigger Digital IN, External TTL Trigger Input 26 Expansion 11 Digital OUT, Simultaneou s sample and hold (SSH) 27 Not Connected N/A 28 AGND Analog Ground 29 AGND Analog Ground 30 CH 7 (SE), or CH 7 HI DIFF Ch 7 IN (single-ended mode) / Ch 7 HI IN (differential mode) 31 CH 6 (SE), or CH 6 HI DIFF Ch 6 IN (single-ended mode) / Ch 6 HI IN (differential mode) 32 CH 5 (SE), or CH 5 HI DIFF Ch 5 IN (single-ended mode) / Ch 5 HI IN (differential mode) 33 CH 4 (SE), or CH 4 HI DIFF Ch 4 IN (single-ended mode) / Ch 4 HI IN (differential mode) 34 CH 3 (SE), or CH 3 HI DIFF Ch 3 IN (single-ended mode) / Ch 3 HI IN (differential mode) 35 CH 2 (SE), or CH 2 HI DIFF Ch 2 IN (single-ended mode) / Ch 2 HI IN (differential mode) 36 CH 1 (SE), or CH 1 HI DIFF Ch 1 IN (single-ended mode) / Ch 1 HI IN (differential mode) 37 CH 0 (SE), or CH 0 HI DIFF Ch 0 IN (single-ended mode) / Ch 0 HI IN (differential mode)
4-6 Connectors, Indicators, Cables, & Pinouts
898095 DaqScan/2000 Series
P2 Digital I/O
P2 Pin Signal Name Description
1 Not Connected N/A 2 Not Connected N/A 3 Port B - B7 Digital I/O: P2 Digital Port B, Bit 7; or, P2 Expansion Address Bit 0 Out 4 Port B - B6 Digital I/O: P2 Digital Port B, Bit 6; or, P2 Expansion Address Bit 1 Out 5 Port B - B5 Digital I/O: P2 Digital Port B, Bit 5; or, P2 Expansion Address Bit 2 Out 6 Port B - B4 Digital I/O: P2 Digital Port B, Bit 4; or, P2 Expansion Address Bit 3 Out 7 Port B - B3 Digital I/O: P2 Digital Port B, Bit 3; or, P2 Expansion Address Bit 4 Out 8 Port B - B2 Digital I/O: P2 Digital Port B, Bit 2; or, P2 Expansion RESET Output
9 Port B - B1 Digital I/O: P2 Digital Port B, Bit 1; or, P2 Expansion WRITE Output 10 Port B - B0 Digital I/O: P2 Digital Port B, Bit 0; or, P2 Expansion READ Output 11 DGND Digital Ground 12 Not Connected N/A 13 DGND Digital Ground 14 Not Connected N/A 15 DGND Digital Ground 16 Not Connected N/A 17 DGND Digital Ground 18 + 5 Volt Supply Expansion +5 Volt Power 19 DGND Digital Ground 20 + 5 Volt Supply Expansion +5 Volt Power 21 DGND Digital Ground 22 Port C - C7 Digital I/O: P2 Digital Port C, Bit 7; or, P2 Expansion Data Bit 7 23 Port C - C6 Digital I/O: P2 Digital Port C, Bit 6; or, P2 Expansion Data Bit 6 24 Port C - C5 Digital I/O: P2 Digital Port C, Bit 5; or, P2 Expansion Data Bit 5 25 Port C - C4 Digital I/O: P2 Digital Port C, Bit 4; or, P2 Expansion Data Bit 4 26 Port C - C3 Digital I/O: P2 Digital Port C, Bit 3; or, P2 Expansion Data Bit 3 27 Port C - C2 Digital I/O: P2 Digital Port C, Bit 2; or, P2 Expansion Data Bit 2 28 Port C - C1 Digital I/O: P2 Digital Port C, Bit 1; or, P2 Expansion Data Bit 1 29 Port C - C0 Digital I/O: P2 Digital Port C, Bit 0; or, P2 Expansion Data Bit 0 30 Port A - A7 Digital I/O: P2 Digital Port A, Bit 7 31 Port A - A6 Digital I/O: P2 Digital Port A, Bit 6 32 Port A - A5 Digital I/O: P2 Digital Port A, Bit 5 33 Port A - A4 Digital I/O: P2 Digital Port A, Bit 4 34 Port A - A3 Digital I/O: P2 Digital Port A, Bit 3 35 Port A - A2 Digital I/O: P2 Digital Port A, Bit 2 36 Port A - A1 Digital I/O: P2 Digital Port A, Bit 1 37 Port A - A0 Digital I/O: P2 Digital Port A, Bit 0
DaqScan/2000 Series
898095 Connectors, Indicators, Cables, & Pinouts 4-7
P3 Pulse/Frequency/Digital I/O
P3 Pin Signal Name Description
1 DGND Digital Ground 2 DGND Digital Ground 3 Digital 7 P3 Digital Port Bit 7 4 Digital 6 P3 Digital Port Bit 6 5 Digital 5 P3 Digital Port Bit 5 6 Digital 4 P3 Digital Port Bit 4 7 Digital 3 P3 Digital Port Bit 3 8 Digital 2 P3 Digital Port Bit 2
9 Digital 1 P3 Digital Port Bit 1 10 Digital 0 P3 Digital Port Bit 0 11 DGND Digital Ground 12 Expansion 2 Reserved for XC/D­13 Expansion 3 Reserved for XWR- 14 Expansion 4 Reserved for XRD- 15 Timer 0 P3 Timer 0 Output 16 Timer 1 P3 Timer 1 Output 17 Counter 2 P3 Counter 2 Input 18 Counter 0 P3 Counter 0 Input 19 + 15 Volts Expansion, +15 VDC 20 + 5 Volt Supply Expansion, +5 Volt Power 21 Analog Out Clock External DAC Pacer Clock Input/Internal DAC Pacer Clock Output 22 Digital 15 Digital I/O; P3 Digital Port Bit 15 23 Digital 14 Digital I/O; P3 Digital Port Bit 14 24 Digital 13 P3 Digital Port Bit 13 25 Digital 12 P3 Digital Port Bit 12 26 Digital 11 P3 Digital Port Bit 11 27 Digital 10 P3 Digital Port Bit 10 28 Digital 9 P3 Digital Port Bit 9 29 Digital 8 P3 Digital Port Bit 8 30 AGND Analog Ground 31 Analog Out DAC 3 Analog DAC 3 Output 32 Analog Out DAC 2 Analog DAC 2 Output 33 Analog Out DAC 1 Analog DAC 1 Output 34 Analog Out DAC 0 Analog DAC 0 Output 35 Counter 3 P3 Counter 3 Input 36 Counter 1 P3 Counter 1 Input 37 - 15 Volts Expansion, - 15 VDC Power
Note: DAC0, DAC1, DAC2, and DAC3 apply to DaqScan/2001 and DaqScan/2004. They do not apply to DaqScan/2002 or DaqScan/2005.
4-8 Connectors, Indicators, Cables, & Pinouts 898095 DaqScan/2000 Series
Software 5
Overview …… 5-1 Out-of-the-Box Drivers for Third-party, Icon-driven Software …… 5-2 Language Drivers …… 5-2

Overview

The Daq devices have software options capable of handling most applications. Three types of software are available:
Out-of the-box graphical programs, e. data analysis programs such as eZ-PostView, eZ-TimeView, and eZ-FrequencyView.
drivers for third-party, icon-driven software such as DASYLab and LabVIEW.
various language drivers to aid custom programming using API; includes
DaqCOM

Out-of-theBoxTM Software

Out-of-the-Box programs are convenient for fill-in-the-blank applications that do not require programming for basic data acquisition and display:
DaqView is a Windows-based program for basic set-up and data acquisition.
TM
Software …… 5-1
g., DaqView, DaqViewXL, and post acquisition
TM
ActiveX/COM support.
DaqView lets you select desired channels, gains, transducer types (including thermocouples), and a host of other parameters with a click of a PC’s mouse. DaqView lets you stream data to disk and display data in numerical or graphical formats.
DaqViewXL allows you to interface directly with Microsoft Excel to enhance data
handling and display. Within Excel you have a full-featured Daq control panel and all the data display capabilities of Excel.
eZ-PostView provides a simple method of graphically viewing acquired waveforms
from the acquisition device. Up to 8 windows can be displayed on one screen, with up to 16 channels overlaid on each window. eZ-PostView makes it simple to visually inspect acquired waveforms from multiple channels within seconds of acquiring the data. Documentation, in Adobe PDF format, is provided on the CD. Refer to the PostAcquisition Analysis.PDF.
eZ-TimeView & eZ-FrequencyView
eZ-TimeView and eZ-FrequencyView are optional post-acquisition analysis packages, which are related to eZ-PostView, but include more features. eZ-TimeView is targeted at time-domain analysis, including min/max, peak-peak, mean, RMS, plus a wide variety of plotting and waveform viewing capabilities. eZ-FrequencyView is targeted at post-acquisition frequency-domain analysis, including FFT’s, octave ­analysis, plus dozens of other analysis features. Documentation, in Adobe PDF format, is provided on the CD. Refer to the PostAcquisition Analysis.PDF.
The Daq Configuration control panel allows for interface configuration, testing,
and troubleshooting.
DaqScan/2000 Series 948895 Software 5-1
Reference Notes:
¾ The software documentation for DaqView, DaqViewXL, and Post
Acquisition Data Analysis are not included as part of the hardcopy
manual, but are available in PDF format. See the PDF Note, below.
¾ Programming topics are covered in the Programmer’s User Manual
(1008-0901). As a part of product support, that manual is automatically loaded onto your hard drive during software installation. The default location is the Programs directory, which can be accessed through the Windows Desktop.
PDF
Note:
During software installation, Adobe® PDF versions of user manuals will automatically install onto your hard drive as a part of product support. The default location is in the Programs group, which can be accessed from the Windows Desktop. Refer to the PDF documentation for details regarding both hardware and software.
Note that the PDF documents can be read directly from the CD by using the <View PDFs> button located on the opening install screen.
®
A copy of the Adobe Acrobat Reader provides a means of reading and printing the PDF documents. Note that hardcopy versions of the manuals can be ordered from the factory.
is included on your CD. The Reader

Drivers for Third-party, Icon-driven Software

LabVIEW® Support
We provide LabVIEW® support for expansion and signal conditioning modules. Our data acquisition VIs for LabVIEW are more than just simple hardware access VIs, they are complete with engineering data conversion, data display and logging capabilities.
DASYLab®
If your application requirements go beyond the scope of DaqView, you may want to consider DASYLab software, as the latter offers a greater degree of flexibility and customizing.

Language Drivers

Each Daq system comes with an Application Programming Interface (API). API-language drivers include: C/C++, and Visual Basic. The latest software is a 32-bit version API.
DaqScan/2000 Series devices are supplied with DaqX Subroutine API Libraries. These provide complete support for all of the functionality available on each data acquisition device in Visual Basic®, and C/C++. In addition, DaqX is supported under all versions of Windows®. Over 100 program examples are included with DaqX Subroutine API Libraries. API documentation is provided in a Programmer’s Manual (p/n 1008-0901), which is included on the installation CD.
The DaqCOM™ suite of programming allows applications developers to rapidly develop and deploy custom systems by leveraging COM (Component Object Model) technology. DaqCOM does this by providing a powerful easy-to-use interface to most programming languages including, Visual Basic®, VBA, C++, and J++. In additi on, DaqCOM supports the new Windows.NET architecture and includes examples for VisualBasic.NET and C++. Support for VisualStudio.NET is accomplished via the COMInterop feature within VisualStudio.NET.
5-2 Software 948895 DaqScan/2000 Series
Configuring the System for Operation 6
Connecting a DaqLab/2000 Series or DaqScan/2000 Series Device to the Ethernet
CAUTION
Turn off power to the system devices and externally connected equipment before connecting cables. Electric shock or damage to equipment can result even under low­voltage conditions.
Take ESD precautions (packaging, proper handling, grounded wrist strap, etc.)
Before you get started
Verify that you have the following items.
DaqLab/2000 or DaqScan/2000 Series Device DBK option cards, as applicable AC Power Cord Ethernet Patch Cable Ethernet jack [on PC or on a hub connected to the Ethernet] Data Acquisition CD Monitor: SVGA, 1024 x 768 screen resolution Windows 2000 SP4 and Windows XP users:
PC with Intel™ Pentium, 1 GHz or equivalent; 512 MB memory; 10 GB disk space
Windows Vista users:
PC must be Windows Vista Premium Ready
Reference Note: Adobe PDF versions of user manuals will automatically install onto your hard drive as a part of product support. The default location is in the Programs group, which can be accessed from the Windows Desktop. You can also access documents directly from the data acquisition CD via the <View PDFs> button located on the CD’s opening screen.
Step 1 - Install the Software ….. 6-2 Step 2 - Determine the type of Network Connection …… 6-2 Step 3 - Connect System Components …… 6-4 Step 4 - Power-up the System …… 6-5 Step 5 - Configure the Computer’s Network Settings …… 6-6 Step 6 - Configure Device Network Settings using DaqIPConfig …… 6-9 Step 7 - Configure and Test the System using the Daq Configuration …… 6-10
The DaqLab/2000 Series and DaqScan/2000 Series acquisition devices connect directly to an Ethernet port on a PC or network hub via a built-in 10/100BaseT Ethernet interface. An Ethernet patch cable CA-242 (1.5 foot) or CA-242-7 (7 foot) cable is used to make the connection.
DaqLab/2000 Series & DaqScan/2000 Series 947091 Configuring the System for Operation 6-1

STEP 1 – Install the Software

1. Remove previous version Daq drivers, if present. You can do this through Microsoft’s Add/Remove Programs feature.
2. Place the Data Acquisition CD into the CD-ROM drive. Wait for PC to auto-run the CD. This may take a few moments, depending on your PC. If the CD does not auto-run, use the Desktop’s Start/Run/Browse feature to locate and run Setup.exe [from the applicable CD-ROM drive].
3. After the intro-screen appears, follow the screen prompts.

STEP 2 – Determine the type of Network Connection

To properly connect and configure the device you must determine the type of network it will become part of. This is because the type of network used has a direct bearing on the IP address of the device.
Briefly, the four network types are as follows:
Dedicated Network - with a direct cable connection from the PC to the
device
Dedicated Network - making use of a network hub or switch LAN with a DHCP server
(Local Area Network with a Dynamic Host Configuration Protocol)
LAN without a DHCP server
(Local Area Network with no Dynamic Host Configuration Protocol)
Brief descriptions and illustrations follow.
Dedicated Network - with a direct cable connection from the PC to the device
In this scenario a DaqLab/2000 Series device is connected directly to an Ethernet jack on a host computer. The following figure could just as easily portray a DaqScan/2000 Series device.
Dedicated Network using a Direct Cable Connection
6-2 Configuring the System for Operation 947091 DaqLab/2000 Series & DaqScan/2000 Series
Some network devices such as a wireless access point may act as a DHCP server. If this is the case, follow the instructions for the LAN with a DHCP server. For detailed information consult the documentation that is specific to your network device.
Dedicated Network - making use of a network hub or switch
In the following figure a DaqLab/2000 Series device is connected to the Ethernet through a network hub/switch. At least one computer is also connected to the hub. DaqScan/2000 series devices can be connected directly to the hub/switch in the same manner.
Dedicated Network using a Hub/Switch
LAN with a DHCP Server
(Local Area Network with a Dynamic Host Configuration Protocol server)
Many corporations use the LAN/Server with DHCP arrangement for their networks. In this type of setup several computers are typically connected to a network that makes use of a DHCP server. In addition, the primary data acquisition device, e.g., DaqLab/2001, is connected to the network hub/switch.
LAN with a DHCP Server
DaqLab/2000 Series & DaqScan/2000 Series 947091 Configuring the System for Operation 6-3
Reference Note:
For information regarding connections, refer to Chapter 4, Connectors, Indicators, Cables, & Pinouts.
Notes:
Using a DaqLab/2000 Series or DaqScan/2000 Series device on a typical LAN may
affect the speed of the network and internet data transfer. Because of this we recommend adding a network card to the computer and using one of the two dedicated network configurations.
Contact your network administrator before connecting your device to a corporate
network.
LAN with no DHCP Server
(Local Area Network with no Dynamic Host Configuration Protocol server)
This scenario looks the same as that shown in the previous illustration, except there is no Dynamic Host Configuration Protocol (DHCP). In this type of setup, one or more computers are connected to a network; and each computer has a static IP address.

STEP 3 – Connect the System Components

What you will need to connect your DaqLab/2000 Series or DaqScan/2000 Series
device to the Ethernet:
An available connection to the Ethernet. The connection can be either
- an Ethernet jack on a computer or
- an Ethernet jack on a hub that is connected to the Ethernet.
An Ethernet patch cable, e.g., a CA-242 (1.5 foot cable) or a
CA-242-7 (7-foot cable).
1. Connect the Ethernet cable to the Ethernet jack on the Daq device.
2. Connect the other end of the Ethernet cable to the Ethernet jack on the host computer or network hub.
6-4 Configuring the System for Operation 947091 DaqLab/2000 Series & DaqScan/2000 Series

STEP 4 – Power-up the System

What you will need:
A 90 to 240 VAC power supply. Your unit’s power supply cord.
How to make the connection:
In the following steps we use the term “Daq Device” to mean a DaqLab/2000 Series device or DaqScan/2000 Series device as applicable to your system.
1. Using the unit’s power switch, turn the Daq device “OFF.” The switch will be in the “0” position and the power LED will be unlit.
2. Connect the power cord to the Power In connector on the Daq device.
3. Connect the plug to a standard AC outlet, or to the outlet of another reliable 90 to 240 VAC power supply.
4. If your power supply has a switch or breaker, close it to turn power “ON.”
5. Turn ON the Daq device by placing its power switch to the 1” position. The unit’s power led will light up.
DaqLab/2000 Series & DaqScan/2000 Series 947091 Configuring the System for Operation 6-5
If using a LAN (Local Area Network), which has a DHCP server, skip this section and continue with STEP 7 - Configure and Test the System using the Daq*Configuration Applet (page 6-10).
If using a LAN (Local Area Network), which has no DHCP server, skip this section and continue with STEP 6 - Configure Device Network Settings using
DaqIPConfig (page 6-9).
Local Area Connection Status
4. In the “Local Area Connection Status” box (previous figure), click on the <Properties> button. The “Local Area Connection Properties” box will appear (following figure).
STEP 5 – Configure the Computer’s Network Settings [Applies to “dedicated networks” only]
1. Open the Control Panel by navigating from the Windows Desktop as follows: Start Menu Settings Control Panel.
2. Double-click the “Network and Dial-up Connections” icon.
3. Double-click the “Network Connection” icon for the network that the Daq device is connected to.
6-6 Configuring the System for Operation 947091 DaqLab/2000 Series & DaqScan/2000 Series
Local Area Connection Properties
5. Double-click the “Internet Protocol (TCP/IP)” component (previous figure). The “Internet Protocol (TCP/IP) Properties” box will appear (following figure).
Internet Protocol (TCP/IP) Properties
DaqLab/2000 Series & DaqScan/2000 Series 947091 Configuring the System for Operation 6-7
Make sure that each computer on the dedicated network has a unique IP address. Make sure that the IP address of the acquisition device is also unique. The computer and the device cannot use the same IP address.
Internet Protocol (TCP/IP) Properties
Configure the Computer’s TCP/IP settings as follows.
6. Select the “Use the following IP Address” radio button.
7. Set the IP address field to 10.0.0.x where x is some number from 1 to 255.
Make sure that each computer on the dedicated network has a unique IP address!
8. Set the Subnet mask to 255.0.0.0. Note that the remaining fields can be left unchanged.
6-8 Configuring the System for Operation 947091 DaqLab/2000 Series & DaqScan/2000 Series
Never set the TCP/IP for the device to be the same as the host computer’s IP address! Each computer and each device on a dedicated network must have a unique IP address!
If using a LAN (Local Area Network), which has a DHCP server, skip this section and continue
with STEP 7 - Configure and Test the System using the Daq*Configuration Applet (page 6-10).
STEP 6 Configure Device Network Settings using DaqIPConfig
Applies only to a LAN (Local Area Network), which has a no DHCP server.
The DaqIPConfig applet is designed for 32-bit Windows/2000/XP/Vista systems. You can use DaqIPConfig to change the IP address of the device to be compatible with networks that require fixed IP addresses. As a precaution, you should always consult with your IT administrator before using the applet to ensure that each device and host computer connected to the network maintains a unique IP address. The applet is located in the program group for the associated device and can be accessed from the Windows Desktop via the start menu.
1. Locate the DaqIPConfig Applet.
Locate the DaqIPConfig applet by navigating from
the Windows’ Desktop as follows:
Start Menu Programs IOtech DaqX
Software DaqIPConfig
2. Open the DaqIPConfig Applet.
Click on the DaqIPConfig selection to open the applet.
3. Select the device that is to have the address
change.
Note: In the above figure there is only one device to select, i.e., DaqLab/2001.
4. Set the internet protocol (TCP/IP) settings to
be compatible with host computer.
(a) Select the radio button labeled “Use the
following IP address.”
(b) Enter the new internet protocol settings. If
needed, consult your network administrator for acceptable numbers.
Do not set the TCP/IP to the computer’s IP address!
(c) Click the <OK> button.
5. Reboot the device.
The new IP address will not take affect until the device has been powered-off, then powered back on.
6. Repeat steps 3, 4, and 5 for other devices in the system.
After configuring the network settings for all devices, proceed to Step 7.
DaqLab/2000 Series & DaqScan/2000 Series 947091 Configuring the System for Operation 6-9
STEP 7 – Configure and test the System using the Daq* Configuration Applet
The Daq* Configuration applet is located in the Windows Control Panel. It allows you to add or remove a device and change configuration settings. The included test utility provides feedback on the validity of current configuration settings and performance summaries.
1. Open the Daq* Configuration Applet.
a. Open the Control Panel by navigating from the Windows’ Desktop as follows:
Start Menu Settings Control Panel
b. From the Control Panel, double-click the Daq* Configuration icon.
2. Add the first-level device to the list of installed devices.
The first-level device is the device that will be connected directly to the Ethernet, via a host computer’s Ethernet jack or a jack on a network hub. DaqLab/2001 is an example of a first­level device.
a. Select the Computer image in the Device Inventory configuration tree (following
figure).
b. Click the <Add Device> button. The “Select Device Type” box will appear. c. Select the device from the list of devices. In our example we are selecting a
DaqLab/2001.
d. Click the <OK> button. The “Properties” box will appear for the selected device.
Using Daq Configuration’s “Device Inventory” and ”Select Device Type” to Add a Device
Note: Although the above figure indicates that a DaqLab/2001 was selected as the
first level device, we could similarly have selected another device, for example, a DaqLab/2005 or a DaqScan/2004.
3. Set the properties of the first-level device.
In this step you will set the device properties according to one of the following two methods,
depending on whether you have a “Dedicated Network” or a “LAN with DHCP Server Network.”
Illustrations of the network types are provided on page 6-3.
6-10 Configuring the System for Operation 947091 DaqLab/2000 Series & DaqScan/2000 Series
Users of Dedicated Networks follow these 2 steps.
Provide your network administrator with the information located on
the device’s MAC label.
Find out from the administrator if the IP Address will be changing. If
so, see if you can obtain a permanent IP Address that will be dedicated
specifically to your device.
For DEDICATED Networks
For LAN with DHCP Server Networks
Daq Configuration, Properties Dialog Boxes
Note: The above images are based on DaqLab/2001. We could similarly have selected another device.
a. Enter the Serial Number of the first-level device, e.g., DaqLab/2001. In the
following screen shots the Serial Number is 800069.
b. Select the “Auto IP Setting” radio button. Note that the IP Address of the
DaqLab/2001 will automatically be calculated and displayed in the IP Address field as indicated in the following left-hand figure.
c. Click the <OK> button.
Users of LAN with DHCP Server Networks follow these 3 steps.
The DaqIPConfig applet provides the Serial Number and the IP Address of the device. Users of LAN with DHCP Server Networks will need to enter both numbers in the Daq Configuration, Properties dialog boxes (previous right-hand figure). Page 6-9 includes instructions for accessing DaqIPConfig.
If needed, refer to the upper right-hand figure in regard to radio-button and data entry locations.
a. Enter the base 10 version of the Serial Number of the device, e.g., DaqLab/2001. b. Select the “Manual IP Setting” radio button. c. In the IP Address field, enter the IP address. d. Click the <OK> button.
4. Test the system connections.
a. Make sure the device has been properly installed and is powered-on. b. Make sure all cables are properly and securely connected. c. Click the “Test Hardware” tab. d. Click the <TCP/IP Test> button. This tests the Transmission Control Protocol /
Internet Protocol. The TCP/IP test has two components: Pinging Device and TCP Connection.
DaqLab/2000 Series & DaqScan/2000 Series 947091 Configuring the System for Operation 6-11
The Pinging Device test is an ICMP (Internet Control Message Protocol) ping test. In addition to indicating either “Passed” or “Failed,” the test displays the round-trip
time of the ping, for example, <10 ms; and Packet Loss expressed as a percent of
When testing a DaqLab/2000 Series or a DaqScan/2000 Series device, if the unit does not respond within 30 seconds perform the following steps:
1) reboot the system
2) upon power-up, re-open the Daq Configuration applet
3) select another configuration setting
4) reinitiate the test
Reference Notes:
Depending on your application and system setup, related information can be found in the following sources:
The DBK Options Manual (p/n 457-0905). The Programmer’s Manual, (p/n 1008-0901).
data lost.
A long ping round trip time, for example, >50 ms and/or any packet loss indicates a slow network that is not optimized.
If the Pinging Device Test fails a <Details> button will appear just above the <Resource Test> button. If this is the case, click the <Details> button to obtain possible causes and possible solutions.
The result of the TCP (Transmission Control Protocol) test will be either passed or failed. If the TCP test failed a <Details> button will appear just above the <Resource Test> button. If this is the case, click the <Details> button to obtain possible causes and possible solutions.
e. Upon completion of the TCP/IP test, click the <Resource Test> button. The
Resource Test consists of two components: Resource Tests and Performance Tests.
The Resource Test performs a quick communications check. If the host PC can communicate with the device a message that the test passed is displayed.
The Performance Tests obtain quantitative results using the current configuration of the acquisition device. The test results represent the maximum rates at which various operations can be performed, and will vary according to port hardware capabilities.
The ADC FIFO Input Speed part of the test results in a display of the maximum
rate at which data can be transferred from the tested device’s internal ADC FIFO
to computer memory. Results are given in samples/second, where a sample (2 bytes in length) represents a single A/D value.
This completes the procedure for connecting a DaqLab/2000 Series or a DaqScan/2000 Series device to the Ethernet.
Note: You can access PDF documents directly from the opening screen of the data acquisition
CD via the <View PDFs> button.
6-12 Configuring the System for Operation 947091 DaqLab/2000 Series & DaqScan/2000 Series
CE Compliance & Noise Considerations 7

Overview

Overview ……7-1 CE Standards and Directives …… 7-1 Safety Conditions ……7-2 Emissions/Immunity Conditions ……7-2 CE Enhancements for DBKs .…… 7-3 CE Compliance for System Expansion …… 7-3 Noise Considerations …… 7-4
CE-compliant products bear the “CE” mark and include a Declaration of Conformity stating the particular specifications and conditions that apply. The test records and supporting documentation that validate the compliance are kept on file at the factory.
The European Union established CE standards in 1985. The standards include specifications for safety, EMI emissions, and immunity from electromagnetic interference. Products that are intended for placement in the European Union must meet or exceed the standards and bear the "CE" mark.
Although not required in the USA, meeting or exceeding the CE standards is considered good engineering practice, since doing so enhances safety while reducing noise and ESD problems.
In contracted and in-house testing, most Daq products met the required specifications. In many cases, products that were not originally in compliance were redesigned accordingly. In noted instances, alternate product versions, shield plates, edge guards, special connectors, or add-on kits are required to meet CE compliance.

CE Standards and Directives

The electromagnetic compatibility (EMC) directives specify two basic requirements:
1. The device must not interfere with radio or telecommunications.
2. The device must be immune from electromagnetic interference from RF transmitters,
etc.
The standards are published in the Official Journal of European Union under direction of CENELEC (European Committee for Electrotechnical Standardization). The specific standards relevant to Daq equipment are listed on the product’s Declaration of Conformity.
The safety standard that applies to Daq products is EN 61010-1:1993 (Safety Requirements
for Electrical Equipment for Measurement, Control, and Laboratory Use, Part 1: General Requirements). Environmental conditions include the following:
indoor use
altitude up to 2000 m
temperature 5°C to 40°C (41°F to 104°F)
maximum relative humidity 80% for temperatures up to 31°C (87.8°F) decreasing
linearly to 50% relative humidity at 40°C (104°F)
mains supply voltage fluctuations not to exceed ±10% of the nominal voltage
other supply voltage fluctuations as stated by the manufacturer
transient overvoltage according to installation categories (overvoltage categories) I, II
and III. For mains supply, the minimum and normal category is II.
pollution degree I or II in accordance with IEC 664
DaqLab/2000 Series and DaqScan/2000 Series 959195 CE-Compliance & Noise Considerations 7-1
For clarification, terms used in some Declarations of Conformity include:
pollution degree: any addition of foreign matter, solid, liquid or gaseous (ionized
gases) that may produce a reduction of dielectric strength or surface resistivity. Pollution Degree I has no influence on safety and implies: the equipment is at operating temperature with non-condensing humidity conditions; no conductive particles are permitted in the atmosphere; warm-up time is sufficient to avert any condensation or frost; no hazardous voltages are applied until completion of the warm-up period. Pollution Degree II implies the expectation of occasional condensation.
overvoltage (installation) category: classification with limits for transient
overvoltage, dependent on the nominal line voltage to earth. Category I implies signals without high transient values. Category II applies to typical mains power lines with some transients.

Safety Conditions

Users must comply with all relevant safety conditions in the user’s manual and the Declarations of Conformity. This manual and Daq hardware make use of the following Warning and Caution symbols: If you see either of these symbols on a product, carefully read the related information and be alert to the possibility of personal injury.
This warning symbol is used in this manual or on the equipment to warn of possible injury or death from electrical shock under noted conditions.
This warning/caution symbol is used to warn of possible personal injury or equipment damage under noted conditions.
Daq products contain no user-serviceable parts; refer all service to qualified personnel. The specific safety conditions for CE compliance vary by product; but general safety conditions include:
The operator must observe all safety cautions and operating conditions specified in the
documentation for all hardware used.
The host computer and all connected equipment must be CE compliant.
All power must be off to the device and externally connected equipment before internal
access to the device is permitted.
Isolation voltage ratings: do not exceed documented voltage limits for power and signal
inputs. All wire insulation and terminal blocks in the system must be rated for the isolation voltage in use. Voltages above 30 Vrms or ±60 VDC must not be applied if any condensation has formed on the device.
Current and power use must not exceed specifications. Do not defeat fuses or other
over-current protection.

Emissions/Immunity Conditions

The specific immunity conditions for CE compliance vary by product; but general immunity conditions include:
Cables must be shielded, braid-type with metal-shelled connectors. Input terminal
connections are to be made with shielded wire. The shield should be connected to the chassis ground with the hardware provided.
The host computer must be properly grounded.
In low-level analog applications, some inaccuracy is to be expected when I/O leads are
exposed to RF fields or transients over 3 or 10 V/m as noted on the Declaration of Conformity.
7-2 CE-Compliance & Noise Considerations 959195 DaqLab/2000 Series and DaqScan/2000 Series

CE Enhancements for DBKs

The following CE enhancements are detailed in the DBK Cards and Modules User’s Manual (p/n 457-0905).
DBK41/CE
Edge Guard (for DBK5, DBK8, and DBK44)
Applicable cables and connectors

CE Compliance for System Expansion

The Daq devices are CE compliant at the time they leave the factory, and will remain in compliance providing that the conditions stated on their respective Declaration of Conformity cards continue to be met.
You may want to expand your data acquisition system while having it retain its “CE Compliant” status. If so, refer to the most recent Declaration of Conformity for each device, prior to making any efforts to expand the system.
Connecting a non-CE Compliant DBK module or card to your system will make the system non-compliant.
In order to remain CE Compliant, DaqLab/2000 Series and DaqScan/2000 Series devices require the use of ferrite inductive collars and a ground line, which are included in an Ethernet CE Kit. Installation instructions for the kit follow this chapter.
General rules of thumb regarding expansion, while remaining CE Compliant, are:
Review the most recent Declaration of Conformity for each device.
Use short cables.
When possible, use shielded cables.
When assembling or disassembling components, take ESD precautions, including the use of grounded wrist straps.
Ensure that the host computer is CE Compliant.
Ensure that expansion devices are CE Compliant
Ensure all system components are properly grounded.
DaqLab/2000 Series and DaqScan/2000 Series 959195 CE-Compliance & Noise Considerations 7-3

Noise Considerations

Controlling electrical noise is imperative because it can present problems even with the best measurement equipment. Most laboratory and industrial environments suffer from multiple sources of electrical noise. For example, AC power lines, heavy equipment (particularly if turned on and off frequently), local radio stations, and electronic equipment can create noise in a multitude of frequency ranges.
Local radio stations are a source of high frequency noise, while computers and other electronic equipment can create noise in all frequency ranges. Creating a completely noise-free environment for test and measurement is seldom practical. Fortunately, simple techniques such as using shielded/twisted pair wires, filtering, and differential voltage measurement are available for controlling the noise in our measurements. Some techniques prevent noise from entering the system; other techniques remove noise from the signal.
While many techniques for controlling noise in signals provide a means of removing the noise that is already present, the preferred solution is to prevent the occurrence of noise in the signal in the first place.
The following practices, some of which are required for CE compliance, should be employed to minimize noise.
Make a solid earth ground connection. Required for CE Compliance chassis of the primary data acquisition device, e.g., DaqLab/2001 is connected to earth ground. This practice: (a) keeps radiated emissions low by keeping the chassis electrically quiet, (b) keeps potential common-mode voltages low, (c) improves user safety, and (d) provides a safe path for Electrostatic Discharge energy back to earth ground. Use a shielded Ethernet cable, e.g., CA-242, to connect the DaqLab or DaqScan chassis to the host computer’s chassis.
Use short Ethernet cables. The use of short Ethernet cables will reduce noise. The shorter the cable the better.
Use shielded cables. Loose wires are effective antennae for radio frequency pickup and can form loops for inductive pickup. The use of properly connected shields will greatly reduce such noise.
Minimize ambient EMI. The lower the ambient EMI, the better. Sources of electromagnetic interference include solenoids, motors, computer equipment, high power distribution wiring, etc.
Distance cables. Power supply switch transients can vary in strength and frequency. Ethernet cables can radiate digital switching noise. For these reasons route the power supply and Ethernet cables such that they are as far as possible from all analog lines. The analog lines include those that connect to P1 on the front panel, as well as those that connect to the rear panel via an expansion port, or through the P4 connector.
Use ferrite inductive collars. A clamp-on ferrite collar can be secured at each end of the Ethernet cable and at both ends of the power supply cable. The collars on the Ethernet cable will reduce digital switching noise. Note that the collars will not reduce the integrity of the Ethernet channel if the system is making use of shielded Ethernet cables, part number CA-242. Ferrite collars on the power cable will reduce the effects of power transients.
. Ensure that the
This partial view of the host PC s hows the Ethernet cable with a ferrite collar. The PC’s connection to the ground-line is also visible.
7-4 CE-Compliance & Noise Considerations 959195 DaqLab/2000 Series and DaqScan/2000 Series
This partial view of the data acquisition device shows its connection to the ground-line. It also shows two ferrite collars, the foremost of which is clamped to the power cable. The background collar is clamped onto the Ethernet cable.
Taking Measurements 8
Introduction …… 8-1 Scanning …… 8-1 Triggering …… 8-2 Synchronous I/O Operations…… 8-3 Asynchronous I/O Operations …… 8-5 Synchronizing Multiple Units …… 8-6 Using DBK Cards and Modules for Signal Conditioning ….. 8-8

Introduction

With signal lines connected and a system up and running you can begin acquiring data. Specific connector details and pinouts are provided in chapter 4. You should refer to the software documentation applicable to your data acquisition system, for example, DaqView.

Scanning

DaqScan/2000 Series devices have an on-board scan sequencer that permits the user to select any combination of up to 16384 channel/range combinations. The sequencer scans all channels contained in the sequence at the rate of 5 µs/channel, 10 µs/channel, or 1000 µs/channel, depending on the speed of the DBK expansion options; thereby minimizing the time-skew from channel-to-channel. The user can also set the time between scan groups, from 0 to 6 hours. In addition to scanning analog inputs, the sequencer can scan digital inputs and counter inputs.
Digital and frequency inputs can be scanned using the same scan sequencer employed for analog inputs, enabling the time correlation of acquired digital data to acquired analog data.
5µs, 10µs, or 1000µs (See Note)
(See Note)
Scanning Example
Note: All inputs can be measured at the 5 µs/channel rate. Some DBK expansion options require more
DaqScan/2000 Series 898095 Taking Measurements 8-1
settling time and can be measured at the slower rate of 10 µs/channel or 1000 µs/channel. All types of measurements can be mixed within the scan group.

Triggering

Triggering can be the most critical aspect of acquiring data. DaqScan/2000 Series devices support a full complement of trigger modes to accommodate a wide variety of applications.
Hardware Analog Triggering. Many data acquisition products claim analog triggering, but rely on the PC to take readings and make a decision, which leads to uncertain and potentially long latencies. DaqScan/2000 Series devices use true analog triggering, whereby the trigger level programmed by the user, sets an analog DAC, which is then compared in hardware to the analog input level on the selected channel. The result is analog trigger latency that is guaranteed to be less than 5 µs, significantly shorter than most data acquisition devices. Any analog channel can be selected as the trigger channel. The user can program both the trigger level, as well as the edge (rising or falling).
Digital Triggering (P1). digital trigger input line is provided, allowing TTL-level triggering, again with latencies guaranteed to be less than 5 µs. Both the logic levels (1 or 0), as well as the edge (rising or falling), can be programmed for the discrete digital trigger input. The digital trigger input is labeled TTLTRG.
Digital Pattern Triggering (P2 and P3). whereby the user can designate any of the digital input ports as the trigger port. The programmed digital pattern, including the ability to mask or ignore specific bits, is then compared to the actual input until a match is detected, after which the sequencer begins the scan sequence.
Counter Triggering. r channels can be programmed as a trigger source.
Software-Based Triggering. Software-based triggering differs from the modes described above because the readings [analo the trigger event [not in the hardware as described above]. The advantage of this mode is that it permits triggering based on more complex situations. For example: to trigger on a specific temperature, which was derived from the acquisition of at least two analog measurements, plus the calculation of the measured temperature using linearization algorithms.
Normally software-based triggering results in long latencies from the time that a trigger c circumvents this undesirable phenomenon by use of pre-trigger data. Specifically, when software-based triggering is employed, and the PC detects that a trigger condition has occurred, (which may be thousands of readings later than the actual occurrence of the signal), the Daq driver automatically looks back to the location in memory where the actual trigger­causing measurement occurred. The acquired data that is presented to the user actually begins at that point. The latency in this mode is equal to one scan cycle.
Appli
es to DaqScan/2001 and DaqScan/2005 only. A separate
The devices support digi
Triggering can be programmed to occur when one of the counters
eaches, exceeds, or is within a programmed level. Any of the built-in counter/totalizer -
g, digital, or counter] are interrogated by the PC to detect
ond
ition is detected, until the actual capturing of data commences. However, circuitry
tal pattern triggering,
8-2 Taking Measurements 898095 DaqScan/2000 Series
Pre- and Post-Triggering Modes. Six modes of pre- and post-triggering are supported, providing a wide variety of options to accommodate any measurement requirement. When using pre-trigger, the user must use software-based triggering to initiate an acquisition.
No pre-trigger, post-trigger stop event. This, the simplest of modes, acquires
data upon receipt of the trigger, and stops acquiring upon receipt of the stop­trigger event.
Fixed pre-trigger with post-trigger stop event. In this mode, the user
specifies the number of pre-trigger readings to be acquired, after which, acquisition continues until a stop-trigger event occurs.
No pre-trigger, infinite post-trigger. No pre-trigger data is acquired in this
mode. Instead, data is acquired beginning with the trigger event, and is terminated when the operator issues a command to halt the acquisition.
Fixed pre-trigger with infinite post-trigger. The user specifies the amount of
pre-trigger data to acquire, after which the system continues to acquire data until the program issues a command to halt acquisition.
Variable pre-trigger with post trigger stop event (driver support only). Unlike
the previous pre-trigger modes, this mode does not have to satisfy the pre-trigger number of readings before recognizing the trigger event. Thus the number of pre­trigger readings acquired is variable and dependent on the time of the trigger event relative to the start. In this mode, data continues to be acquired until the stop trigger event is detected.
Variable pre-trigger with infinite post trigger (driver support only). This is
similar to the mode just described, except that the acquisition is terminated upon receipt of a command from the program.
Stop Trigger. Any of the software trigger modes described above can be used to
stop an acquisition. Thus an acquisition can be programmed to begin on one event, such as a temperature level, and then stopped on another, such as a digital pattern event.

Synchronous I/O Operations

Synchronous Input Operations

The DaqScan/2000 Series devices allow synchronous scanning and acquisition of Analog Input [for those models with a P1 connector], Digital Input and Counter Input Data at up to 200 kHz aggregate scanning rates. When applicable, the Analog Input data can be either main unit (DaqScan/2001 or DaqScan/2005) or expansion modules from P1-compatible analog input modules. The Digital Input data can be main unit 8-bit P2 (8255) digital input or from P2 compatible DBK digital input expansion modules.
Analog Input Channels
P1 – Channels 0 through 15 (DaqScan/2001 and DaqScan/2005 only)
DaqScan/2001 and DaqScan/2005 allow analog input configuration of themselves, as well as for any connected P1 compatible DBK analog input expansion modules.
Channel Selection and Mode Settings
DaqScan/2001 and DaqScan/2005 have P1 connectors that allow for expansion of up to 256 high-speed channels, or 896 TC channels. In regard to system expansion, refer to the DBK Option Cards & Modules User’s Manual (p/n 457-0905).
Channel Range and Polarity
The analog input channels of DaqScan/2001 and DaqScan/2005 can be programmed for gain settings of 1, 2, 4,
DaqScan/2000 Series 898195 Taking Measurements 8-3
8, 16, 32, and 64.
Channel Sampling Interval
In regard to analog input, DaqScan/2001 and DaqScan/2005 allow for programmable sampling i
ntervals of 5 us, 10 us, or 1000 us on a per channel basis. This mode allows some DBK expansion channels [that change slowly for a higher degree of accuracy] to be sampled at a longer interval, and channels that change more rapidly to be sampled at a shorter interval. Each 5 µs, 10 µs, or 1000 µs interval reduces the maximum aggregate acquisition rate for the entire scan by that amount.
Digital I/O Channels
There are 40 Digital I/O channels available to DaqScan/2000 Series devices. The channels are grouped as follows:
8 Channels, A0 through A7 from P2 Port A 8 Channels, B0 through B7 from P2 Port B
8 Channels, C0 through C7 from P2 Port C 16 Channels, Digital 0 through Digital 15 from P3
The devices allow for either synchronous scanning of digital input channels or asynchronous
/O o
perations for all configured digital channels.
I
Counter Input Channels
Counter 0, 1, 2, 3 (CTR0, CTR1, CTR2, CTR3)
DaqScan/2000 Series devices allow synchronous scanning of four16-bit counter input channels. The 4 channels can be cascaded into two 32-bit counter channels. For either cascaded or non-cascaded counter channels each channel can be configured for:
fie
Pulse Counting Mode – speci
s that each counter should be cleared upon being
read and placed into the input scan.
Totalize Counting Mode – specifies that each counter is to free-run and not be cleared during the input acquisition.
Synchronous Input Acquisition Clocking
A /I Clock
DaqScan/2000 Series devices allow clocking of the synchronized inputs either by an internal, programmable pacer clock or by external clocking. These products use a sequencer to implement a multiplexing approach to gathering the input data. This means that with either internal or external clocking the entire channel scan (including the samplin g time for each channel) may not exceed the maximum aggregate rate of 200 kHz.

Synchronous Output Operations

DaqScan/2001 and DaqScan/2004 allow synchronous output of any available D/A channels at up to 100 kHz for each channel. All D/A channels may have output clocked out synchronously.
Output Channel Configuration Analog Output Channels
Each of the four D/A channels for DaqScan/2001 and DaqScan/2004 can be configured for waveform output individually. If the D/A channel is not confi then is available for asynchronous output operations.
Synchronous Output Clocking (DaqScan/2001 and DaqScan/2004 only)
A / O Clock
DaqScan/2001 and DaqScan/2004 allow clocking of the synchronized output by the acquisition clock source, an internal, programmable pacer clock or by an external clock source. When the clock source generates a new clock signal all outputs are updated concurrently. Regardless of the clock source, the clock may not exceed the maximum update rate of 100 kHz.
(DaqScan/2001 and DaqScan/2004 only)
gured for waveform output it
8-4 Taking Measurements 898095 DaqScan/2000 Series
Synchronous Output Data Source (DaqScan/2001 and DaqScan/2004 only)
DaqScan/2001 and DaqScan/2004 allow the data source for synchronized DAC output operations to be that of an internal contained in the buffer.

Asynchronous I/O Operations

Each DaqScan/2000 Series device allows asynchronous input of any counter or digital channel that is not currently configured for synchronous acquisition. The circuitry allows for asynchronous output to any D/A channels that are not currently configured for waveform output. Likewise, the digital port can be used for both asynchronous input and output operations. In addition, the timer outputs can be programmed at any time regardless of the current state of synchronous or asynchronous operations on other channels.
Digital I/O Channels Local 8255 Channels (P2: Port A, Port B, Port C)
The devices have an implemented Intel 8255 core in the digital I/O logi c on the P2 port. With the Intel configuration purposes. The configuration port is used to configure the other three 8-bit ports for either input or output operations.
Local 16-bit P3 Port The 16-bit P3 digital port can be used as either an input or output port. With this port, no
onf
c
Expansion Digital I/O
The devices have the ability to expand their digital I/O capabilities via the P2 port and the conn DBK Option Cards & Modules User’s Manual (p/n 457-0905). When using the digital I/O expansion modules the local P2 Intel 8255 digital I/O becomes inaccessible in lieu of the expansion modules. These expansion modules provide additional Intel 8255 ports as well as input isolation for applications that require the expanded capabilities.
8255 there are three 8-bit wide ports avai
iguration is required. The port simply outputs when written to and inputs when read.
ection of applicable digital I/O expansion modules. These modules are discussed in the
memory-based buffer. The output data, for all channels, is
lable for I/O and one 8-bit wide port for
Pulse Stream Output Using Timers (Timer0, Timer1)
Each device allows for the generation of output pul These output timers can be set at any time regardless of the state of any synchronous or asynchronous operations which are currently taking place on other channels.
The timer outputs are through screw terminals on the front panel.
Analog Output Channels (DaqScan/2001 and DaqScan/2004 only)
DaqScan/2001 and DaqScan/2004 can output analog data to any of their four D/A channels. Each D/ currently being used for waveform output operations. The four analog outputs are via P3 (see pinout in chapter 4).

Counter Input Channels (CTR0, CTR1, CTR2, CTR3)

DaqScan/2000 Series devices have counter input capabilities and have the ability to read counter input provid As in the case of synchronous operations the four 16-bit counter input channels can be used individually or cascaded into two 32-bit counter channels. For either cascaded or non­cascaded counter channels each channel can be configured for:
A channel may be asynchronously updated by an application if the D/A channel is not
ing that the counter channel is not configured for synchronous acquisition.
Clear on R
upon being read.
Continuous Totalize Mode – specifie
cleared during the read operation.
ead Mode - specif
ies that each counter should be cleared (reset to 0)
ses based upon a programmable setting.
s that each counter is to free-run and not be
DaqScan/2000 Series 898195 Taking Measurements 8-5

Synchronizing Multiple Units

Up to four DaqScan/2000 Series devices can be synchronized to each other via their rear panel SYNC connectors and associated cables. Each device has two identical SYNC ports; either [or both] can be used to connect units. The DaqScan units can be scan-synchronized and triggered from any other SYNC-connected unit. Note that cables CA-74-1 and CA-74-5 are 1-foot and 5-foot in length, respectively. The total combined cable length of the Sync cables is not to exceed 15 feet (4.57 m).
Synchronization Concept Block Diagram
The preceding diagram shows how a DaqScan/2000 Series device can input or output synchronization timing and trigger events on either SYNC connector. A more detailed block diagram follows shortly.
Reference Notes: ¾ The SYNC features are programmable via software. For detailed information
see the Using Multiple Devices and the daqAdcSetClockSource sections of the Programmer’s Manual (p/n 1008-0901).
¾ During software installation, Adobe
automatically install onto your hard drive as a part of product support. The default location is in the Programs group, which can be accessed from the Windows Desktop. Refer to the PDF documentation for details regarding both hardware and software.
®
PDF versions of user manuals
Synchronization Model for DaqScan/2000 Series Devices (see note)
Note: DaqScan/2002 and DaqScan/2004 have no P1 connector; thus the pin 20
and pin 25 lines do not apply to those two devices.
8-6 Taking Measurements 898095 DaqScan/2000 Series
In regard to synchronization, the following applies:
One of three signals can be used to drive the DaqScan/2000 Series acquisition scan
rate. These are:
the internal acquisition pacer clock
an external acquisition pacer clock (for DaqScan/2001 and /2005 only)
the global sync (G-SYNC) input from the SYNC ports
Both the SYNC connector input and the external clock input can be divided down.
When a DaqScan is in the Master Mode, both the trigger condition and the scan timing
are output on the SYNC port. The global trigger (G-TRIG) is selected instead of the TTL Trigger Input.
Whether the DaqScan is using its internal scan clock, or external clock input, it can be programmed to output the clock on the SYNC connector. In either case, the unit is behaving like a synchronization master. Other DaqScan units [that are connected to the master, via a SYNC port] should be programmed as synchronization slaves.
When a DaqScan is programmed as a synchronization slave, it will derive its scan period from the SYNC port. If the slave unit must be triggered at the same time as the master unit, then the slave unit should use TTL Trigger as its trigger source.
Sync Notes:
When a DaqScan/2000 Series device is programmed as a synchronization slave, the TTL
trigger source is automatically derived from the SYNC port.
When a DaqScan/2000 Series device is programmed as a synchronization master, it will
output its trigger condition on the SYNC port.
When synchronizing two or more DaqScan/2000 Series devices, the slave units should have
at least 0.1 µs of dead time in the scan period. By “dead time,” we mean a duration in which no channels are sampled. This accommodates fundamental differences in the clocks [for each DaqScan].
All DaqScans that are connected via SYNC cables can be scan-synchronized to within
0.1 µs of each other.
A maximum of 4 units can be synchronized; scan-synchronous (post-trigger).
Sync cables are not to exceed a total combined length of 15 feet (4.57m).
Multi-unit synchronization via the SYNC ports does not support pre-trigger data. Only post-trigger data can be collected and time correlated between the units.
DaqScan/2000 Series 898195 Taking Measurements 8-7

Using DBK Cards and Modules for Signal Conditioning

The DBK signal-conditioning cards and module are designed for use with DaqLabs, DaqScans, DaqBooks, LogBooks, and various types of data acquisition boards, i.e., ISA, PCI, and compact-PCI types. The DBKs perform best when used with an acquisition device that can dynamically select channel, gain, and range. DBK cards and modules with dynamic channel and gain/range selection allow for high channel-to-channel scan rates with a variety of transducers.
Note: Only passive DBKs, such as the DBK1 BNC module, the DBK11A screw terminal card,
and the DBK40 BNC analog interface, can be used with Daq PC-Cards.
DBK output signals can be bipolar, e.g., -5 to +5 V, or unipolar, e.g., 0 to 10 V. The user can select a range of relevant values to correspond to the lowest signal (e.g., -5 or 0 V) and the highest signal (e.g., 5 or 10 V) signal. This type of range selection guarantees the highest resolution in 12-bit or 16-bit conversion.
The majority of these modules have their own power supply; however, several options exist for packaging and powering the DBKs.
Reference Note: DBK options are detailed in the DBK Option Cards and Modules User’s Manual (p/n 457-0905). As a part of product support, the manual is automatically loaded onto your hard drive during software installation. The default location is the Programs group, which can be accessed through the Windows Desktop. A PDF version of the document can be read directly from the data acquisition CD via the introduction screen’s <View PDFs> button.
8-8 Taking Measurements 898095 DaqScan/2000 Series
Calibration 9
Overview …… 9-1 Using DaqCal.exe …… 9-1

Overview

Daqscan/2000 Series devices are digitally calibrated at the factory. The di gital method involves storing a correction factor on the circuit board for each range at the time of calibration. Whenever a particular range is selected, the appropriate calibration constant is automatically applied to a compensating DAC, thereby calibrating the specific range. The result is that readings generated by the A/D are already calibrated, and do not require additional processing.
This is significantly better than the calibration method in which readings are adjusted in software after they are transferred to the host the A/D
Daqscan/2000 Series devices include a user-cal mode, whereby the user can adjust the calibrati accomplished with two distinct calibration tables in the unit’s on-board EPROM, one which contains the factory calibration factors, and the other which is available for user calibration.
, and can adversely affect the speed at which the PC obtains a calibrated reading.
on of the unit’s board, without destroying the supplied factory calibration. This is
PC. That method reduces the dynamic range of

Using DaqCal.exe

When calibration adjustments are needed they should be completed in the following order:
1.
2.
3.
4.
5.
6.
7.
8.
A Windows-based program, DaqCal.exe, is used to cal expansion cards. DaqCal.exe is used in conjunction with:
a 4.5 digit, digital multi-meter
an adjustable voltage cal
an ambient temperature meter
To use the calib
1. Launch DaqCal.
2. W
PGA Input and Output Offset Sample/Hold Offset A/D Offset and Gain VDC Voltage Reference DAC0 Full-Scale DAC1 Full-Scale1 DAC2 Full-Scale DAC3 Full-Scale
1
DAC Full-Scale only applies to Daqscan/2001 and DaqScan/2004.
ration program:
Note:
DaqCal is installed automatically from your data acquisition CD as a part
of product support. This takes pl default location is the IOtech DaqX Software folder, in the Programs group.
hen DaqCal opens you will be prompted to select your device from a list. After
doin
g so, simply follow the illustrated on-screen instructions.
1
1 1
ibrate Daq systems, i
ibrator
ace during software installation. DaqCal’s
ncluding analog
Daqscan/2000 Series 948695 Calibration 9-1
9-2 Calibration 948695 Daqscan/2000 Series
Troubleshooting and Customer Support 10
Electrostatic Discharge (ESD), Handling Notice…… 10-1 Product Care …… 10-1 ReadMe Files and the Install CD-ROM ……10-2 DaqCOM Issues …… 10-2 Driver Support……10-2
Ethernet Problems …… 10-3 Frequently Asked Questions …… 10-4 Customer Support …… 10-5

Electrostatic Discharge (ESD), Handling Notice

The discharge of static electricity can damage some electronic components. Semiconductor devices are especially susceptible to ESD damage. You should always handle components carefully, and you should never touch connector pins or circuit components unless you are following ESD guidelines in an appropriate ESD-controlled area. Such guidelines include the use of properly grounded mats and wrist straps, ESD bags and cartons, and related procedures.

Product Care

Daq products and DBK options are essentially maintenance free and require a minimal amount of care. They should be treated much like any other high-tech equipment. In general:
Keep them clear of harsh chemicals and abrasive elements.
Avoid exposing the products to extreme heat, for example, avoid setting up a Daq system
near a boiler or furnace.
Avoid extreme shock and vibration.
Avoid subjecting the units to liquids and extremely fine air particulate, such as silica dust.
Never touch circuit boards without proper ESD (Electrostatic Discharge) measures in
place.
A “common-sense” approach to handling your data acquisition equipment will go a long way in protecting it from inadvertent damage.
Note that you can use lint-free rags and Isopropyl Alcohol (Rubbing Alcohol) to clean the outer plate surfaces of Daq devices and optional DBK modules.
DaqScan/2000 Series 898095 Troubleshooting & Customer Support 10-1

ReadMe Files and the Install CD-ROM

The Install CD-ROM includes ReadMe Files. These files often contain late-breaking information that may not appear in the user documentation. During installation you should review the ReadMe files when prompted to by the program.
The Install CD-ROM includes:
DaqBook/2000 Series
DaqX.DLL (32-bit API)
Microsoft C/C++ API support
Microsoft VB API support
Daq Configuration Control Panel Applet
User documentation in Adobe® PDF format
DaqView
Upon installation, the user needs to run the Daq Configuration Control applet is used to configure the Daq device for an appropriate Ethernet Port. This must be performed in order for the application to access the device through the Windows NT/2000/XP driver.
1
DaqBook/2000 Series drivers also apply to DaqScan/2000 Series devices.
1
Windows NT/2000/XP driver
Panel applet.
The

Driver Support

The daqX API can be used with WaveBook, DaqBook, DaqBoard, DaqLab, DaqScan, Daq PC Card and TempBook product lines. All daqX functions share the daq prefix.
The API support examples can be found in the installed directory:
<Install> ProgrammingLanguageSupport\DAQ
Refer to the Programmer’s Manual, p/n 1008-0901.

DaqCOM Issues

DaqCOM support is contained within the Help file support, which is installed with DaqCOM. It is located within the DaqCOM program group.
DaqCOM support will not be available unless it has been specifically installed with the product. If the D Windows desktop), then you will need to install it from the data acquisition CD, or download it from our website.
aqCOM program group is not li
sted under programs (which can be accessed via the
10-2 Troubleshooting & Customer Support 898095 DaqScan/2000 Series

Ethernet Problems

Make sure the Ethernet network is not experiencing technical problems. Check with your Network Administrator.
Ensure that your system is properly configured in the Daq Configuration Applet [located in the Windows’ Control Panel]. If a device is not configured correctly it will not be accessible from an application.
Verify that the correct MAC (Media Access Control) number is
If communications cannot be established with the Daq device or, if trying to connect causes the system to hang or crash, try the following:
- Veri
- Ensure that all physic
- Run the Daq Configuration applet from the Windows Control Panel. Make sure
If problems persist, notify your Network Admi problem, contact the factory.
fy that the device power switch is set to ON and that the POWER ON indicator
is lit.
al device interface connections are correct. This includes Ethernet connections from the Daq device to the host PC’s Ethernet jack, or Ethernet Network hub.
the device is config protocol. Run the device test and if it fails try other protocols for the device. Be sure to click on <Apply> to apply the settings, before re-running the test.
ured to the appropriate Ethernet port with the appropriate
nistrator. If he or she can’t solv
reflected by the software.
e the
DaqScan/2000 Series 898095 Troubleshooting & Customer Support 10-3

Frequently Asked Questions

(1) Topic: Environmental Factors
Question: What environments are the products intended for? Answer: Refer to the product specifications in regard to operating and storage temperature
rang
es, and relative humidity. Note that system components are not intended to be exposed to harsh environments and should always be protected from snow, rain, extreme dust, and harsh sun.
(2) Topic: DaqX Question: What is DaqX software?
Answer: DaqX software is low-level software support for application development. DaqX is
insta
lled during the installation of the DaqView application. DaqX supports Visual
C++ and Visual Basic programming environments.
Question: Where can I get the latest version of DaqX software? Answer: You can download the latest versions from our website. This service is free.
(3) Topic: Calibration
Question: Should I calibrate my Daq device when I receive it? Answer: No, because each Daq-related product is factory-calibrated to its rated accuracy
before it leaves the factory.
Question: When is calibration required? Answer: Depending on your operating guidelines, your equipment will require periodic
c
ation to ensure the original accuracy is maintained. We recommend that you
alibr
have your DaqScan system calibrated at least once a year.
(4) Topic: Computers for Data Acquisition
Question: What type of computer do you recommend for use with DaqScan/2000 Series
devices?
Answer: The PC should have built in Ethernet capabilities and use an operating system of
Windows NT, 2000, XP, or later.
10-4 Troubleshooting & Customer Support 898095 DaqScan/2000 Series

Customer Support

If you want to Expand or Enhance Your Daq System . . .

You can visit our internet site at ZZZPFFGDTFRP to find the latest accessories, expansion cards and modules. You can also call our sales department directly at () -.

If you need to Report Problems or Request Product Support

Note: Please do not return any equipment to the factory unless it has an RMA
number (Return Merchandise Authorization number). RMA numbers are issued by the factory.
To report problems or request support, contact our applications department or your service representative. Contact information follows shortl following information available, as applicable:
Hardware model
Operating system, type of computer, and device information i
control panel, such as interrupts and address settings for our hardware and others.
Resul
Hardware setup and software configuration.
Infor
ts of tests, such as the Daq Configuration control panel.
mation on input signals, including voltage ranges, signal impedance ranges,
noise content, and common mode voltages.
numbers and software version numbers.
The name of a contact person from your company who can discuss the problems
encountered.
Instructions for return shipping.
All troubleshooting notes and comments on tests performed, and all problem-
related conditions.
If your problem is Ethernet related, discuss the problem with your Network
Administrator prior to contacting the factory. Please provide us with your administrator’s comments, if applicable.
y. When you contact us, please have the
n the Windows
Note: Before calling for assistance, take a few minutes to read all parts of the manual that
may be relevant to the problem. Also, please review the troubleshooting material.
You can reach 0HDVXUHPHQW&RPSXWLQJ by one of the following means:
Phone: ()  Fax: ( E-mail (Product Information/Sales7HFKQLFDO6XSSRUW): LQIR@PFFGDT.com
,QWHUQHWZZZPFFGDTFRP
Mail: 0HDVXUHPHQW&RPSXWLQJ&RUS. z &RPPHUFH:D\ z 1RUWRQ0$ 
All equipment returned to the manufacturer must have an RMA (Return Material Authorization) number. You can obtain an RMA number by calling the Applications Department. When returning the equipment, use the original shipping container (or equivalent) to prevent damage. Put the RMA number on your shipping label to ensure that your shipment will be handled properly. After receiving your equipment, we will fax a confirmation form that summarizes the charges (if applicable) and expected return date.
DaqScan/2000 Series 898095 Troubleshooting & Customer Support 10-5
10-6 Troubleshooting & Customer Support 898095 DaqScan/2000 Series
Specifications 11

DaqScan/2000 Series

General

Supply Voltage Range: 90 to 250 VAC Power Required: 15W, assuming no DBK options
Power A Operating
Storage Temperature: -40°C to +80°C (-40° to 176°F) Relativ Vibration: MIL ST
Signal I/O Connec
Dimensions: 425 mm W x 220 mm D x 45 mm H (16.75” x 8.5” x 1.75”) Weight: 2.3 kg (5 lbs.)

A/D Specifications

Type: Successive approximation Resolution: 16 bit Conversion Time: 5 µs Maximum Sample Rate: 200 kHz Nonlinearity (Integral): ±1 LSB Nonlinearity (Differential): ± 2 LSB Missing Codes: None, over full operating temperature range
Analog Inputs (DaqScan/2001 and DaqScan/2005 only)
vailable for External Signal Conditioning & Expansion Options:
5V at 1A, ±15V at 500 mA, not to exceed 10W
Temperature: 0°C to + 50°C (32° to 122°F)
e Humidity: 0 to 95% non-condensing
D 810E, Categories 1 and 10
tors:
DaqScan/2001 and DaqScan/2005: DB37 male connectors: P1, P2, P3 on rear panel DaqScan/2002 and DaqScan/2004: DB37 male connectors: P2 and P3 on rear panel
Channels:
16 single-ended or 8 differential; programmable on a per-channel basis as single-ended or differential and as
unipolar or bipolar
Connectors:
37-pin DSUB (P1) on rear panel
Expansion:
Up to 896 TC channels when used with DBK90 expansion option [1 µs/channel], or up to 256 channels when used with other expansion options [5 µs/channel]
Bandw
idth: 500 kHz
Settling Time: 5 µs max
mperature Coefficient: ±(10ppm +0.3 LSB)/°C outside the range of 0° to 35°C
Te Input Impeda Bias Current: <1nA at 0° to 35°C Common Mode Reje Maximum Input Voltage: +11V relative to analog common
-Voltage Protection: ±35V
Over Crosstalk: -100 dB DC to 60 Hz; -86 dB @ 10 kHz
DaqScan/2000 Series 898095 Specifications 11-1
nce: 10 M for single-ended; 20 M for differential
imum to 1 LSB for full-scale step
ction: 86 dB, DC to 60 Hz for gains <
8; >100 dB for gains >16
Ranges: Software programmable via sequencer on a per-channel basis.
Notes:
Input Sequencer
Analog, digital and counter inputs can be scanned synchronously based either on an internal programmable timer, or an external clock source. Analog and digital outputs can be synchronized to either of these clocks.
Scan Clock Sources: 2
Internal, programmable from 5 µs to 5.96 hours maximum in 1 µs steps
External, TTL level input up to 200 kHz maximum
Programmable Parameters per Scan: channel (random order), gain, unipolar or bipolar Depth: 16,384 locations On-boa Expansion Channel Scan Rate: 5 µs,10 µs, or 1000 µs per channel, programmable
rd Channel-to-Channel Scan Rate: 5 or 10 µs per channel, programmable
Specifications are subject to change without notice.
Voltage Range (Note 1)
0 to +10 V 0.015 + 0.005 1 0 to +5 V 0.015 + 0.005 1 0 to +2.5 V 0.015 + 0.005 1 0 to +1.25 V 0.015 + 0.008 2 0 to +0.625 V 0.015 + 0.008 2 0 to +0.3125 V 0.015 + 0.008 2
-10 to +10 V 0.015 + 0.005 1
-5 to +5 V 0.015 + 0.005 1
-2.5 to +2.5 V 0.015 + 0.005 1
-1.25 to +1.25 V 0.015 + 0.005 2
-0.625 to +0.625 V 0.015 + 0.008 2
-0.3125 to +0.3125 V 0.015 + 0.008 3
-0.156 to +0.156 V 0.020 + 0.008 3
1) Specifications assume differential input single channel scan, 200 kHz scan rate, unfiltered.
2) Accuracy specification is exclusive of noise. Measurements w
3) Inputs shorted to Signal Ground (SGND). 8192 samples.
Accuracy (Note 2)
One Year, 0 to 35°C
±(% reading + % range)
Absolute Typical
Input Noise (Note 3)
(LSB rms)
10 Hz to 200 kHz
ere taken at P1.

External Acquisition Scan Clock Input

Maximum Rate: 200 kHz Clock Signal Range: 0V to +5V Minimum Pulse Width: 50 ns high, 50 ns low

External Sync Ports

Number of External Sync Ports: 2, on rear panel Maximum Number of Units to be Synchronized: 4 units, scan-synchronous (post trigger)
mum Length of Sync Cables: 15 feet (4.57 m), total for all cables
Maxi

Triggering

Trigger Sources: 6, individually selectable for starting and stopping an acquisition. Stop acquisition can occur on a different channel than start acquisition; stop acquisition can be triggered via modes 2, 4, 5, or 6 described below. Pre-trigger is supported with fixed or variable pre-trigger periods.
1. Single-Channel Analog Hardware Trigger
Any analog input channel, including any of the analog expansion channels, can be software programmed as the analog trigger channel.
Input Signa Trigger Level: Programmable, 11-bit resolution Hy Latency: 5 µs max
11-2 Specifications
steresis: Programmable, 11-bit resolution
l Range: -10V to +
10V max
898095 DaqScan/2000 Series
Specifications are subject to change without notice.
2. Single-Channel Analog Software Trigger.
Any analog input channel, including any of the analog expansion channels, can be selected as the software trigger channel. If the trigger channel involves a calculation, such as w automatically compensates for the delay required to calculate the reading, resulting in a maximum latency of one scan period.
Input Signal Range: Any Trigger Level: Programmable, 16-bit resolution, including “w Hysteresis: Programmable, 16-bit resolution Latency: One scan period max.
3. Single-Channel Digital Trigger: A separate digital input is provided for digital triggering.
Input Signal Range: -15V to +15V max Trigger Le
vel: TTL
Minimum Pulse Width: 50 ns high, 50 ns low Latenc
y: 5 µs max.
4. Digital Pa
ttern Triggering: 8 or 16-bit pattern triggering on any digital input. Programmable for trigger on
equal, above, below, or within or outside of a window. Individual bits can be masked for “don’t care” condition.
Latency: One scan period maximum.
5. Counter/Totalizer Triggering: Counter/totalizer inputs can trigger an acquisition. User can select to trigger
on a frequency or on total counts that are equal, above, below, or w
Latency: One scan period maximum.
6. Software Triggering: Trigger can be initiated under program control.
where wi
ith temperature, then the driver
thin the range of the selected trigger channel
indow triggering”
.
ithin/outside of a window.

Analog Outputs (DaqScan/2001 and DaqScan/2004 only)

For DaqScan/2001 and DaqScan/2004, four analog output channels are updated synchronously relative to scanned inputs, and clocked from either an internal onboard clock, or an external clock source. Analog outputs can also be updated asynchronously, independent of any other scanning in the system.
Channels: 4 DAC channels; numbered 0, 1, 2, and 3 Connectors: pins 31 (DAC3), 32 (DAC2), 33 (DAC1), and 34 (DAC0) on the P3 DB37 connector Resolution: 16 bits Data Buffer: 256 Ksample Output voltage range: ±10 V Output current: ±10 mA Offset error: ±0.0045 V maximum Gain error: ±0.01% Digital Feedthru: 50 mV when updated Update rate: 100 kHz maximum, 1.5 Hz minimum (no minimum with external clock) Settling Time Clock So
1. Onboard D/A clock, independent of scanning input clock
2. Onboard scanning input clock
3. External D/A input clock, independent of external scanning input clock
4. External scanning input clock; available via screw-terminal on front panel removable block
: 10 µs maximum to 1 LSB for full-scale step
urces: 4 programmable
DaqScan/2000 Series 898095 Specifications 11-3

Digital I/O

Channels: 40, expandable up to 272 with external digital DBK options Input Scanning Modes: 2 programmable
Ports: 3 x 8-bit (82C55 emulation), and 1 x 16-bit; each port is programmable as input or output. Connectors:
Input Characteristics: See following Digital I/O figures Input protection: ±8 kV ESD clamp diodes parallel I/O levels: TTL Sampling/Update rate: 200 kHz maximum Output Characteristics: See following Digital I/O figures
Specifications are subject to change without notice.
1. Asynchronous, under program control at any time relative to input scanning
2. Synchronous with input scanning
37-pin DSUB (P2) on rear panel: for three 8-bit ports (24-bits) 37-pin DSUB (P3) on rear panel for DIGITAL I/O channels 0 through 15 (16-bit port)

Frequency/Pulse Counters

Counter inputs can be scanned synchronously along with analog and digital scanned inputs, based either on internal programmable timer, or an external clock source. Bus mastering DMA provides CPU and system­independent data transfers, ensuring data acquisition performance irrespective of other system activities. Counters can be configured to clear when read, or to totalize and clear under program control.
Channels: 4 x 16-bit, cascadable as 2 x 32-bit Frequency Measurement Rate: 10 MHz maximum Connectors: 37-pin DSUB (P3) on rear panel; pin 18 (CTR0), pin 36 (CTR1), pin 17 (CTR2), pin 35 (CTR 3) Input Signal Range: -15 V to +15 V
Trigger Level: TTL Input Characteristics: 2.7 k series in parallel with 20 pF to common and 10 k to +5 V Input protection: ±8 kV ESD clamp diodes parallel Minimum pulse width: 50 ns high, 50 ns low

Timer Outputs

Channels: 2 x 16-bit Connectors: 37-pin DSUB (P3) on rear panel; pin 15 (TMR0), pin 16 (TMR1) Output Waveform: Square wave Output rate: 1 MHz base rate divided by 1 to 65535 (programmable) High level output voltage: 2.0 V minimum @ -3.75 mA, 3.0 V minimum @ -2.5 mA Low level output voltage: 0.4 V maximum @ 2.5 mA
11-4 Specifications
898095 DaqScan/2000 Series
Specifications are subject to change without notice.

Accessories and Cables

Cables for use with DaqScan/2000 Series Devices
Description Part No.
1. Molded expansion cable, 2 in. CA-255-2T
2. Molded expansion cable, 4 in. CA-255-4T
3. DBK expansion cable, not-molded, 2.5 in. CA-37-1T
4. DBK expansion cable, not-molded, 4.5 in. CA-37-3T
5. Expansion-card cable. Connects DBK series expansion card options. The “x” in the part no. indicates the number of cards that can be connected via the cable.
6. Ethernet patch cable, shielded, 18 in. CA-242
7. Ethernet patch cable, shielded, 7 ft. CA-242-7
8. SYNC cable, RJ-11, 6 conductor, 1 ft. CA-74-1
9. SYNC cable, RJ-11, 6 conductor, 5 ft. CA-74-5
CA-37-x
The total combined length of the SYNC cables is not to exceed 15 feet (4.57 m).
Manuals
Manual Set (Set #1126-0900). This is a hardcopy set of manuals that is available for purchase.
The set includes the following three manuals:
DaqScan/2000 Series User’s Manual (p/n 1126-0901) Programmer’s Manual for developing custom programs using API commands.
(p/n 1008-0901)
DBK Option Cards & Modules User’s Manual. (p/n 457-0905)

Software

DaqViewXL for seamless execution of DaqView from within Microsoft Excel’s tool palette.
DASYLab icon-based data acquisition graphics, control, and analysis software.
eZ-TimeView post-acquisition time-domain analysis software
eZ-FrequencyView post-acquisition frequency domain analysis software
DaqScan/2000 Series 898095 Specifications 11-5
Specifications are subject to change without notice.
This page is intentionally blank
.
11-6 Specifications
898095 DaqScan/2000 Series

Glossary

Acquisition A collection of scans acquired at a specified rate as controlled by the sequencer.
Analog A signal of varying voltage or current that communicates data.
Analog-to-Digital
Convert
API Application Program Interface. The interface program within the Daq system’s driver that
Bipolar A range of analog signals with positive and negative values (e.g., -5 to +5 V); see unipolar.
Buffer Buffer refers to a circuit or device that allows a signal to
Buffer Amplifier An amplifier used primarily to match two different impedance points, and isolate one stage from
Channel In reference to Da
er (ADC)
t or device that converts analog values into digital values, such as binary bits, for use in
A circui
digital computer processing.
includes f (several languages supported).
isolation, or another function, without altering the signal. Buffer usually refers to:
(a) A devi
(b) A fol
(c) An amplifier which accepts high so
a succeeding stage in order to prevent an undesirable interaction between the two stages. (Also see, Buffer
In a broader sense, an input channel is a signal path between the transducer at the point of
measurement and the data acquisition system. A channel can go through various stages (buf periodically sampled for readings.
unction calls specific to
ce or circuit that al Such storage can compensate for differences in data flow rates. In a FIFO (First In - First Out) buffer, the data that is stored first is also the first data to leave the buffer.
lower stage used to drive a number of gates without overloading the preceding stage.
impedance output (effectively, an impedance buffer).
).
q devices, channel simply refers to a si
fers, multiplexers, or signal conditioning amplifiers and filters). Input channels are
Daq hardware and can be used with user-written programs
pass through it, while pr
lows for the temporary storage of data during data transfers.
ce impedance input and results in low source
ur
ngle input, or output entity.
oviding
An output channel from a devi
way in response to an input channel signal.
Common mode Common mode pertains to signals that are identical in amplitude and dur
Common mode
volt
age
Crosstalk An undesired transfer of signals between systems or system components. Crosstalk causes
Digital A digital signal is one of discrete val
Digital-to-Analog
Convert
DIP switch A DIP switch is a gr oup of miniature switches in a small Dual In-line Package (DIP). Typically,
Differential mode The differential mode measures a voltage between 2 si
Glossary 959395 G-1
er (DAC)
in reference to signal components.
Common mode voltage refers to a vol
shared by two or more signals. Example: referenced to common, Signal 1 is +5 VDC and Signal 2 is +6 VDC. The common mode voltage for the two signals is +5.5 VDC [(5 + 6)/2].
si
gnal in
terference, more commonly referred to as noise.
digits (0s and 1s) represent digital data.
A circuit or
users set these switches to configure their
single-ended mode).
device that converts digital values (binary bits), into analog signals.
ce can be digital
tage magnitude (referenced to a common point) that is
ue, in
or analog. Outputs can vary in a programmed
ation; also
contrast to a varying signal. Combinations of binary
particular application.
gnal li
nes for a single channel. (Also see
can be used
Differential mode
voltage
Differential mode voltage refers to a voltage difference between two signals that are referenced
to a common point. Example: Signal 1 is +5 VDC referenced to common. Signal 2 is +6 VDC referenced to common.
If the +5 VDC signal is used as the reference, the differential mode voltage is +1 VDC
(+ 6 VDC - +5 VDC = +1 VDC).
If the +6 VDC signal is used as the reference, the di
(+ 5 VDC - +6 VDC = -1 VDC).
ESD Electrostatic discharge (ESD) is the transfer of an electrostatic charge between bodies having
Excitation Some transducers [e.g. strain gages, thermistors, and resistance temperature detectors
Gain The degree to which an input signal is amplified (or attenuated) to allow greater accuracy and
Isolation The arrangement or operation of a circuit so that si
Linearization Some transducers produce a voltage in linear proportion to the condition measured. Other
Multiplexer (MUX) A device that collects signals from several inputs and outputs them on a single channel.
Sample (reading) The value of a signal on a channel at an instant in time. W
different electro when induced by an electrostatic field. ESD energy can damage an integrated circuit (IC).
(R
TDs)] requir
transducer corresponds to the condition measured.
resolution; can be expressed as ×n or ±dB.
affect the isolated circuit.
In reference to Daq devices, iso
the signal source and the analog-to-digital converter (ADC). Isolation is necessary when measuring high common-mode voltage.
transducers (e.g., thermocouples) have a nonlinear response. To convert nonlinear signals i
nto accurate readin
interpolate values between these points.
channel and conver
static potentials. This transfer occurs duri
e a known voltage or current. Typically, the variation of this signal through the
lation usual
gs requires software to calibrate several points in the range used and then
ts the sampled value into a 12- or 16-bit value.
ly refers to a separation of the direct link between
fferential
gnals from another circui
mode voltage is -1 VDC
ng direct contact of the bodies, or
hen triggered, the ADC reads the
t or device do not
Scan A series of measurements across a pre-selected sequence of channels.
Sequencer A programmable device that manages channels and channel-specif
Simult
aneous
Sample-and-Hold
Single-ended mode The single-ended mode measures a voltage between a signal line and a common reference that
Trigger An event to start a scan or mark an instant during an acquisition. The event can be defined in
TTL Transistor-Transistor Logic (TTL) is a circuit in which a multiple-emitter transistor has replaced
Unipolar A range of analog signals that is always zero or positive (e.g., 0 to 10 V). Evaluating a signal in
An o
peration that gathers samples from multiple channels at the same instant and holds these
values until all are sequentially converted to digital values.
may be shared wi
v
ous ways; e.g., a TTL signal, a specified voltage level in a monitored channel, a button
ari manually or mechanically engaged, a software command, etc. Some applications may use pre- and post-triggers to gather data around an instant or based on signal counts.
the multiple diode cluster (of the diode-transistor logic circuit); typically used to communicate logic signals at 5
the ri
ght range (unipolar or bipolar) allows greater resolution by using the full-range of the
corresponding digital value. See bipolar.
th other channels. (Al
V.
so see differential mode).
ic settings.
G-2 959395 Glossary
Loading...