IOtech 652u Quick Start
Before you get started
verify that you have the following items and meet or exceed the
minimum requirements listed.
IOtech 652u
TR-2U Power Supply
USB2.0 port [on PC] and USB Cable
Dynamic Signal Analysis CD
License Keys for purchased software;
e.g., eZ-Analyst, eZ-TOMAS
Monitor: SVGA, 1024 x 768 screen resolution
Windows 2000 SP4 and Windows XP users:
Intel
™
Pentium [or equivalent], 1 GHz
512 MB memory; 10 GB disk space
Windows Vista users:
PC must be Windows Vista Premium Ready
Step 1 - Install Software
1. Close all running applications on the host PC.
2. Insert the Dynamic Signal Analysis CD into your CD-ROM drive and wait for the CD to auto-run. An Opening
Screen will appear. As an alternative, you can download software from: www.iotech.com/ftp.html
3. Click the <ENTER SETUP> button.
Note: If you are downloading software from our website, follow instructions provided there.
4. From the hardware selection screen [which follows a licensing agreement], select the 652u
from the drop-down list and follow the on-screen instructions.
Supply power from the TR-2U to the 652u before connecting the USB cable to the computer. This allows the
652u to inform the host computer (upon connection of the USB cable) that the unit requires minimal power
from the computer’s USB port. When disconnecting a 652u from the PC, unplug the USB cable before
unplugging the TR-2U power cable.
10-Channel, USB2.0 Dynamic Signal Analyzer for Vibration Analysis & Monitoring
Step 2 - Connect the 652u to Power and to the Computer
1. Connect the TR-2U to a standard AC outlet.
2. Connect the TR-2U to the External Power connector on the 652u.
3. Using a USB cable, connect the 652u to a USB2.0 port on the computer. USB2.0 port is required.
4. Follow the computer screen prompts as directed to allow the computer to detect your new hardware.
1170-0940 rev 1.0 324996A-01

IOtech
25971 Cannon Road
Cleveland, OH 44146-1833
Phone: (440) 439-4091
Fax: (440) 439-4093
E-mail: sales@iotech.com
E-mail: productsupport@iotech.com
Internet: www.iotech.com
LED Notes: The “Power” LED blinks during device detection and initialization; then remains on solid as long as
Turn off power to the system devices and externally connected equipment before
connecting cables. Electric shock or damage to equipment can result even under
low-voltage conditions.
Take ESD precautions (packaging, proper handling, grounded wrist strap, etc.)
Reference Notes:
Adobe Acrobat PDF versions of documents pertaining to IOtech 652u are automatically installed onto
your PC’s hard-drive as a part of product support at the time of software installation. The default
location is the Programs group. It can be accessed via the Windows Desktop Start Menu.
Digital I/O
Channel
Designation
the module has power. If there is insufficient power the LED will go off. The “Active” LED is on whenever active
communication is taking place between 652u and the host PC. It will be on solid during data acquisitions.
Step 3 - Connect Data Acquisition Signal Lines
The 652u has 10 analog channel inputs (CH1 through CH10) via front panel BNC connectors and 8 digital I/O
lines via rear panel DB9 connector.
Prior to making signal connections review the Specifications chapter of your user’s manual to ensure that the input
signals do not exceed the specified limits. The manual is included in PDF format on the CD.